DRYPIX Lite Quality Control Guide E
DRYPIX Lite Quality Control Guide E
DRYPIX Lite Quality Control Guide E
Implementation
of the QC
Function 2
DRYPIX 2000
Appendix
Index
This Quality Control Guide provides operational procedure of the DRYPIX Lite
Quality Control and Precautions thereof. Be sure to read the Quality Control Guide
prior to using this product. After reading the Quality Control Guide, please store it
in a safe place for handy reference. Please also read the “DRYPIX Lite Operation
Manual” which provides detailed information on the operation of the DRYPIX Lite
and the “FCR Quality Assurance Program Plus – For Medical Institutions – Image
Recorder Guidebook” (optional).
897N102177A
ii DRYPIX Lite (DRYPIX 2000) Quality Control Guide 897N102177A
Introduction
This Quality Control Guide is applicable to the following software.
This Quality Control Guide describes the quality control (QC) functions of the Fuji Medical Dry
Imager DRYPIX Lite (DRYPIX 2000). The QC function can only be used if the QC function is
set to “Available” during installation. The QC is a function to implement the quality control of
the images printed by the DRYPIX Lite (DRYPIX 2000).
For the concepts behind the quality control, refer to the following guidebook.
“FCR Quality Assurance Program Plus – For Medical Institutions – Image Recorder
Guidebook” (sold separately, hereafter indicated as “The guidebook”)
CAUTIONS
1. No part or all of this manual may be reproduced in any form without prior permission.
2. The information contained in this manual may be subject to change without prior notice.
3. FUJIFILM Corporation shall not be liable for malfunctions and damages resulting from
installation, relocation, remodeling, maintenance, and repair performed by other than dealers
specified by FUJIFILM Corporation.
4. FUJIFILM Corporation shall not be liable for malfunctions and damages of FUJIFILM
Corporation products due to products of other manufacturers not supplied by FUJIFILM
Corporation.
5. FUJIFILM Corporation shall not be liable for malfunctions and damages resulting from
remodeling, maintenance, and repair using repair parts other than those specified by
FUJIFILM Corporation.
6. FUJIFILM Corporation shall not be liable for malfunctions and damages resulting from
negligence of precautions and operating methods contained in this manual.
7. FUJIFILM Corporation shall not be liable for malfunctions and damages resulting from use
under environment conditions outside the range of using conditions for this product such as
power supply, installation environment, etc. contained in this manual.
8. FUJIFILM Corporation shall not be liable for malfunctions and damages resulting from natural
disasters such as fires, earthquakes, floods, lightning, etc.
Trademark
FCR and FDR are trademarks or registered trademarks of FUJIFILM Corporation.
Chapter 1 Overview 1
This chapter provides an overview of the quality control of the DRYPIX Lite’s images as
well as an overview of the QC functions to implement quality control.
Appendix
A worksheet is attached for a recording of the results of the QC functions. Copy
the worksheet to use it.
Index
Chapter 1 Overview
1.1 The Quality Control (QC) Function.............................................. 1-2
1.2 Precautions When Implementing the Uniformity Test Using
the QC Function............................................................................. 1-3
Appendix
Index
Overview
Page
1 In order to maintain the quality of the images printed by the DRYPIX Lite, this device is subjected to an uniformity
test. With the DRYPIX Lite keep the following points in mind when implementing the QC function for the
uniformity test.
Overview
1. Use the DRYPIX Lite to print the “QC test patterns” used for quality control.
2. Print a QC test pattern and check that an error code is not displayed on the DRYPIX Lite display
panel.
The indicated measured result and the measured results besides those of the QC test pattern are recorded on
the worksheet.
The uniformity test is periodically implemented and its results are recorded each time.
Check the results measured by the uniformity test do not exceed the standard value.
Contact our official dealer should the measured results exceed the standard value. In this case, the worksheet
with a record of the results measured by the uniformity test is required.
Quality control can be implemented using other means than the uniformity test. In addition, the uniformity test
can be implemented without the use of the QC function, however the use of the uniformity test will make quality
control easier and more reliable.
Overview
2.1.1 Preparations
2 ● External densitometer
● Plastic ruler
● Worksheet
Implementation of the QC Function
For the external densitometer, use a densitometer calibrated with the standard density step tablet that complies with
the expiration date. The use of a standard density step tablet manufactured based on a density step tablet (NIST :
National Institute of Standards and Technology (US)) is recommended. The recommended density step tablet and
densitometer are as follows.
■ Densitometer
Manufacturer: FUJIFILM Medical Co., Ltd.
Product name: 301RS
Use a general straight ruler enabling measurement down to the millimeter. The use of a transparent ruler with an easy
to check indication is recommended.
(3) Worksheet
Make a copy of the worksheet in the appendix of this Quality Control Guide to use it.
2
Printing the QC test pattern
Checking artifacts
Visually check the QC test pattern for irregularities
and disorderly patterns.
Record the visual check results in the worksheet.
you confirm the presence of artifacts.
The “Guidebook” provides details of the standard values (basic values) and the setting standard.
The table below shows the basic Utility menu operations available when the printer has switched to Utility mode.
Menu button
Set button
You can display the next function and option.
Set button
You can confirm or execute the current option.
Menu button
If outputting a QC test pattern, the Utilities are terminated automatically, and the device returns to a status where
you can print.
CAUTION
Note that you will have to wait several minutes to 15 minutes to output a QC test pattern
depending on usage conditions.
2
To b e u s e d f o r t h e
measurement by the
Format B
external densitometer
of density 0, 1, and 2. Measure format B
Format A
Measure format A using a plastic ruler.
<26 × 36 cm>
Format B
Format A
Format B Format B
Format A
Format A
2
Implementation of the QC Function
Approx. 10 mm
(0.4 in.)
NOTE
One filter is already put in the cover. Do not place
the densitometer upside down when removing
the cover. Photodiode
(installed on the
internal circuit board)
Photodiode cover
NOTE
Make sure that the filter is not placed at a slant.
Measurement results of the print density: Record the measurement values at density 0-4 with the judgment result. (*Density 4 is the service setting)
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21
Density 4 3.30
3.25
3.20
3.15
3.10
3.05
3.00
2.95
2.90
2.85
2.80
2.75
2.70
Measurement value
Determination of PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS
the measurement FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21
Density 3 2.20
2.15
2.10
2.05
2.00
1.95
1.90
1.85
1.80
1.75
1.70
1.65
Measurement value
Determination of PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS
the measurement
FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21
Density 2 1.45
1.40
1.35
Appendix
1.30
1.25
1.20
1.15
1.10
1.05
1.00
0.95
Measurement value
Determination of PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS
the measurement
FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL
0.50
0.45
0.40
0.35
Measurement value
Determination of PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS
the measurement
FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21
0.30
Density 0
0.25
0.20
0.15
0.10
Measurement value
Determination of PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS
the measurement
FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL
Format Measurement Result: Record the measurement value of format A and B along with the determination of the measurement in the table below.
Format A
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21
Measured
value (mm)
Determination of PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS
the measurement FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL
Format B
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21
Measured
value (mm)
Determination of PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS
the measurement
FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL
Artifact Determination of the Measurement: Record the determination of the measurement of the artifact in the table below.
Artifact
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21
Appendix
Determination of PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS PASS
the measurement FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL FAIL
A
Artifact.......................................................2-10
B
Basic values and setting standards..............2-2
Initial setting values...................... 2-8, 2-9
D
Densitometer...............................................2-2
Density 4.....................................................2-8
Density step tablet.......................................2-2
Determination of the measurement results
.................................................. 2-8, 2-9, 2-10
E
External densitometer.......................... 2-2, 2-8
F
Format A.....................................................2-9
Format B.....................................................2-9
P
Plastic ruler.......................................... 2-2, 2-9
Q
QC function
Initiating...............................................2-4
Terminating..........................................2-5
QC test patterns..........................................2-7
Printing................................................2-6
Quality control.............................................1-2