Download as pdf or txt
Download as pdf or txt
You are on page 1of 4

2018/12/28

Clinical Chemistry A93A01307CEN

A11A01627

56 mL
ABX Pentra ALT CP 14 mL

■ Pentra C400

Diagnostic reagent for quantitative in vitro determination of Alanine


AminoTransferase (ALT) in serum or plasma by colorimetry.

Application Release Method (3, 4)

Optimized UV-test according to IFCC (International


Serum, plasma: ALT
Federation of Clinical Chemistry) modified method without
1.xx pyridoxal phosphate.
ALT
L-Alanine + 2-Oxoglutarate ◀————▶ L-Glutamate + Pyruvate
Intended Use LDH
Pyruvate + NADH + H+ ◀————▶ D-Lactate + NAD+
ABX Pentra ALT CP reagent is intended for the
quantitative in vitro diagnostic determination of Alanine (ALT = Alanine Aminotransferase, LDH = Lactate
AminoTransferase (ALT) in serum or plasma by Dehydrogenase)
colorimetry.
Alanine amino transferase measurements are used in the
diagnosis and treatment of certain liver diseases (e.g., Reagents
viral hepatitis and cirrhosis) and heart diseases.
ABX Pentra ALT CP is ready-to-use.

Clinical Interest (1, 2) Reagent 1:


TRIS pH 7.15 140 mmol/L
Alanine Aminotransferase (ALAT/ALT), formerly called L-Alanine 700 mmol/L
Glutamic Pyruvic Transaminase (GPT) and Aspartate
LDH (lactate dehydrogenase) ≥ 2300 U/L
Aminotransferase (ASAT/AST), formerly called Glutamic
Oxalacetic Transaminase (GOT) are the most important Sodium azide < 1 g/L
representatives of a group of enzymes, the
aminotransferases or transaminases, which catalyze the Reagent 2:
conversion of α-keto acids into amino acids by transfer of 2-Oxoglutarate 85 mmol/L
amino groups. As a liver specific enzyme ALT is only
NADH 1 mmol/L
significantly elevated in hepatobiliary diseases. Increased
AST levels, however, can occur in connection with Sodium azide < 1 g/L
damages of heart or skeletal muscle as well as of liver
parenchyma. Parallel measurement of ALT and AST is ABX Pentra ALT CP should be used according to this
therefore applied to distinguish liver from heart or skeletal notice. The manufacturer cannot guarantee its
muscle damages. The AST/ALT ratio is used for performance if used otherwise.
differential diagnosis in liver diseases. While ratios < 1
QUAL-QA-WORI-5542 Rev.1

indicate mild liver damage, ratios > 1 are associated with


severe, often chronic liver diseases.

1/4
S.A.S. au capital de 23.859.980 € - RCS Montpellier 328 031 042 - SIRET 328 031 042 000 42 - APE 332 B Latest version of documents on www.horiba.com
Clinical Chemistry

ABX Pentra ALT CP

Handling Specimen (5)

1. Remove both caps of the cassette. ■ Serum.


■ Plasma in lithium heparin.
2. If present, remove foam by using a plastic pipette.
3. Place the cassette into the refrigerated Pentra C400 Anticoagulants other than those listed have not been
reagent compartment. tested by HORIBA Medical and are therefore not
recommended for use with this assay.

Stability:
Calibrator
■ At 20-25°C: 3 days
■ At 4-8°C: 7 days
For calibration, use:
■ At -20°C: 7 days
ABX Pentra Multical (A11A01652) (not included)
10 x 3 mL (lyophilisate)
Reference Range (4)
Control a Each laboratory should establish its own reference
ranges. The values given here are used as guidelines only.
For internal quality control, use:
Women: ≤ 34 U/L (37°C)
■ ABX Pentra N Control / ABX Pentra N MultiControl Men: ≤ 45 U/L (37°C)
(A11A01653 / 1300054414) (not included)
10 x 5 mL (lyophilisate)
■ ABX Pentra P Control / ABX Pentra P MultiControl Storage and Stability
(A11A01654 / 1300054415) (not included)
10 x 5 mL (lyophilisate)
Stability before opening:
Each control should be assayed daily and/or after a Stable up to the expiry date on the label if stored at
calibration. 2-8°C.
The frequency of controls and the confidence intervals
should correspond to laboratory guidelines and country- Stability after opening:
specific directives. You should follow federal, state and
local guidelines for testing quality control materials. The Refer to the paragraph "Performance on Pentra C400".
results must be within the range of the defined confidence
limits. Each laboratory should establish a procedure to Do not freeze.
follow if the results exceed these confidence limits.
Waste Management
Materials Required but not Provided a
■ Please refer to local legal requirements.
■ This reagent contains less than 0.1% of sodium azide
■ Automated clinical chemistry analyzer: Pentra C400
as a preservative. Sodium azide may react with lead
■ Calibrator: ABX Pentra Multical (A11A01652)
and copper to form explosive metal azides.
■ Controls:
ABX Pentra N Control / ABX Pentra N MultiControl
(A11A01653 / 1300054414)
ABX Pentra P Control / ABX Pentra P MultiControl
(A11A01654 / 1300054415)
■ Standard laboratory equipment.

aModification: new control.

2/4
S.A.S. au capital de 23.859.980 € - RCS Montpellier 328 031 042 - SIRET 328 031 042 000 42 - APE 332 B Latest version of documents on www.horiba.com
Clinical Chemistry

ABX Pentra ALT CP

General Precautions b Accuracy and Precision

Repeatability (within-run precision)


■ This reagent is for professional in vitro diagnostic use
only. Repeatability according to the recommendations found in
■ For prescription use only. the Valtec protocol (7) with samples tested 20 times:
■ This reagent is classified as non-hazardous in
compliance with regulation (EC) N°.1272/2008. ■ 2 controls
■ Reagent 1 (R1): ■ 3 specimens (low / medium / high levels)
Warning: This reagent is obtained from substances of
animal origin. Consequently, it should be treated as Mean value U/L CV %
potentially infectious and handled with the appropriate Control specimen 1 39.73 1.00
cautions in accordance with good laboratory practices
Control specimen 2 126.09 1.19
(6).
■ Do not pipette by mouth. Specimen 1 17.43 3.07
■ Do not replenish the reagents. Specimen 2 28.44 2.28
■ Do not swallow. Avoid contact with skin and mucous Specimen 3 127.88 0.59
membranes.
■ Observe the standard laboratory precautions for use.
■ The reagent cassettes are disposable and should be Reproducibility (total precision)
disposed of in accordance with the local legal Reproductibility according to the recommendations found
requirements. in the CLSI (NCCLS), EP5-A protocol (8) with samples
■ Please refer to the SDS associated with the reagent. tested in duplicate for 20 days (2 series per day):
■ Do not use the product if there is visible evidence of
biological, chemical or physical deterioration. ■ 2 controls
■ It is the user's responsibility to verify that this ■ 2 specimens (low / high levels)
document is applicable to the reagent used.
Mean value U/L CV %
Control specimen 1 39.79 2.53
Performance on Pentra C400
Control specimen 2 124.91 1.77
Serum, plasma Specimen 1 31.49 6.00
Specimen 2 87.60 2.48
The performance data listed below are representative of
performance on HORIBA Medical Systems. Measuring Range
Number of tests: 250 tests The assay confirmed a measuring range from 4.0 U/L to
If the number of tests requested is low and the 600.0 U/L.
Pentra C400 user intends to utilise the cassette to the The measuring range is extended up to 1800.0 U/L with
maximum on board stability, it is the recommendation of the automatic post-dilution.
HORIBA Medical, to utilise the consumable part XEC232 The reagent linearity has been assessed up to 600.0 U/L
(Kit membrane) to achieve the number of tests stated in according to the recommendations found in the CLSI
this notice. (NCCLS), EP6-A protocol (9).

On Board Reagent Stability Correlation


Once opened, the reagent cassette placed in the
Patient samples: Serum and plasma
refrigerated Pentra C400 compartment is stable for 42
Number of patient samples: 100
days.
Specimens are correlated with a commercial reagent
Sample volume: 20 µL/test taken as reference according to the recommendations
found in the CLSI (NCCLS), EP9-A2 protocol (10).
Detection Limit
The detection limit is determined according to the Valtec
protocol (7) and equals 4 U/L.

bModification: general precautions modification.

3/4
S.A.S. au capital de 23.859.980 € - RCS Montpellier 328 031 042 - SIRET 328 031 042 000 42 - APE 332 B Latest version of documents on www.horiba.com
Clinical Chemistry

ABX Pentra ALT CP

The equation for the allometric line obtained using 4. IFCC Primary Reference Procedures for the
Passing-Bablock regression procedure (11) is: Measurement of Catalytic Activity Concentrations of
Y = 1.00 X + 5.00 (U/L) Enzymes at 37°C; Part 4; Clin. Chem. Lab. Med.
with a correlation coefficient r2 = 0.9964. (2002) 40 (7): 718-724.
5. Guder WG, Zawta B. The Quality of Diagnostics
Interferences c Samples. Samples: From the Patient to the
Laboratory. 1st Ed. Guder WG, Narayanan S, Zawta
Haemoglobin: No significant influence is observed up B. (WHILEY-VCH, Darmstadt, Germany) (2001): 43.
to 195 µmol/L (336 mg/dL). 6. Council Directive (2000/54/EC). Official Journal of the
Triglycerides: No significant influence is observed up European Communities. No. L262 from October 17,
to an Intralipid® concentration 2000: 21-45.
(representative of lipemia) of 7. Vassault A, Grafmeyer D, Naudin C et al. Protocole
100.0 mg/dL. de validation de techniques (document B). Ann. Biol.
Total Bilirubin: No significant influence is observed up Clin. (1986) 44: 686-745.
to 450 µmol/L (26.3 mg/dL). 8. Evaluation of Precision Performance of Clinical
Direct Bilirubin: No significant influence is observed up Chemistry Devices. Approved Guideline, CLSI
to 891 µmol/L (52.1 mg/dL). (NCCLS) document EP5-A (1999) 19 (2).
The presence of Sulfasalzine or Sulfapyridine in serum/ 9. Evaluation of the Linearity of Quantitative Analytical
plasma can cause false results. Methods. Approved Guideline, CLSI (NCCLS)
Other limitations are given by Young as a list of drugs and document EP6-A (2003) 23 (16).
preanalytical variables known to affect this methodology 10. Method Comparison and Bias Estimation Using
(12, 13).
Patient Samples. Approved Guideline, 2nd ed., CLSI
(NCCLS) document EP9-A2 (2002) 22 (19).
Calibration Stability
11. Passing H, Bablock W. A new biometrical procedure
The reagent is calibrated on Day 0. The calibration for testing the equality of measurements from two
stability is checked by testing 2 control specimens. different analytical methods. J. Clin. Chem. Clin.
The calibration stability is 8 days. Biochem. (1983) 21: 709-20.
Note: A recalibration is recommended when reagent lots 12. Young DS. Effects of Drugs on Clinical Laboratory
change, and when quality control results fall outside the Tests. 4th Edition, Washington, DC, AACC Press
range established. (1997) 3: 143-163.
13. Young DS. Effects of Preanalytical Variables on
Clinical Laboratory Tests. 2nd Edition, Washington,
Reference DC, AACC Press (1997) 3: 120-132.
1. Thomas L. Alanine aminotransferase (ALT), Aspartate
aminotransferase (AST). In: Thomas L, editor. Clinical
Laboratory Diagnostics. 1st ed. Frankfurt: TH-Books
Verlagsgesellschaft (1998): 55-65.
2. Panteghini M, Bais R. Enzymes. In: Tietz Textbook of
Clinical Chemistry and Molecular Diagnostics. 4th
Ed., Burtis CA, Ashwood ER, Bruns DE, (Elsevier
Saunders eds. St Louis, USA) (2006): 604-607.
3. Bergmeyer HU, Horder M, Rej R. International
Federation of Clinical Chemistry (IFCC) Scientific
Committee, Analytical section: approved
recommendation (1985) on IFCC methods for the
measurement of catalytic concentration of enzymes.
Part 3. IFCC method for alanine aminotransferase (L-
alanine: 2-oxoglutarate aminotransferase, EC
2.6.1.2). J. Clin. Chem. Clin. Biochem. (1986) 24:
481-495.

cModification: modification of interferences.

4/4
S.A.S. au capital de 23.859.980 € - RCS Montpellier 328 031 042 - SIRET 328 031 042 000 42 - APE 332 B Latest version of documents on www.horiba.com

You might also like