DVVR
LTD, PUNE
PROTOCOL
STATUS
SR NO DRS ID REQUIREMENT DESCRIPTION TEST AIM PROCEDURE APPLICABLE EXPECTED BEHAVIOUR ACTUALHAVIOUR REMARKS/EVIDENCE
YES NO PASS FAIL
The Quantino laser system and its accessories are The Quantino laser system and its accessories should be used
The Quantino laser system and its accessories is able to be
intended to be used in surgical procedures using open, in surgical procedures using open, laparoscopic, and used in surgical procedures using open, laparoscopic, and
laparoscopic, and endoscopic incision, excision, endoscopic incision, excision, resection, ablation,
resection, ablation, vaporization, coagulation and To check whether Thulium Laser functions according to 1. Send the system to User/Surgeorn. vaporization, coagulation and haemostasis of soft tissue and endoscopic incision, excision, resection, ablation, vaporization,
1 DRS-FR-01 2.verify the received Clinical evaluation feedback form YES coagulation and haemostasis of soft tissue and lithotripsy of Pass
haemostasis of soft tissue and lithotripsy of stones in its intended use. of Thulium fiber laser system. lithotripsy of stones in use in medical specialties including stones in use in medical specialties including Urology, Thoracic Refer:Clinical evaluation report
use in medical specialties including Urology, Thoracic Urology, Thoracic and Pulmonary, ENT, Neurosurgery,
and Pulmonary, ENT, Neurosurgery, Gastroenterology, Gastroenterology, Gynecology and General Surgery as per its and Pulmonary, ENT, Neurosurgery, Gastroenterology,
Gynecology and General Surgery as per its intended use.
Gynecology and General Surgery. intended use.
2 DRS-OC-01 Medical fiber probe should be used for optical power To test that the laser system generates output power Inspect the fenestrated cup design and geometry as NO Medical fiber probe should be used for optical power delivery The fenestrated cup design and geometry is as per drawing. Pass Refer Drawing No:
delivery from laser system to end use in surgery. of 60W at 1940nm wavelength. per drawings. from laser system to end use in surgery. Batch manufacturing Record
3 DRS-OC-02 The Thulium Fiber Laser should generate 60W of optical To inspect the output optical power at 1940nm YES The Thulium Fiber Laser should generate 75W of optical The Thulium Fiber Laser is able to generate 75W of optical Pass
power at 1940nm wavelength. Wavelength. power at 1940nm wavelength. power at 1940nm wavelength.
Go to Procedure tab.
The system should allow the surgeon to have flexibility To test whether the system allows the surgeon to have 1.Adjust the Power, Frequency, aming beam intensity The system should allow the surgeon to have flexibility to The system allows the surgeon to flexibily adjust the
7 DRS-UR-04 by sliding the slider to and fro. YES Pass Usability Validation Report : 504-UVR-FL60
to adjust the parameters with ease. flexibility to adjust the parameters with ease. 2.Adjust the pulse modes, standyby & ready modes, adjust the parameters with ease. parameters with ease.
presets, auto manual etc by pressing the tabs.
To check whether the system allows the admin to 1. Select Login as Admin.
The system should allow the admin to create an account 2. Enter the Admin Name and Proceed. The system should allow the admin to create an account of The system allows the admin to create an account of the
12 DRS-UR-09 of the surgeon by adding details such as account name, create an account of the surgeon by adding details such 3. Select New Surgeons details and enter details such YES the surgeon by adding details such as account name, hospital surgeon by adding details such as account name, hospital Pass Usability Validation Report : 504-UVR-FL60
as account name, hospital name, Registration no,
hospital name, Registration no, password etc. password etc. as account name, hospital name, Registration no, name, Registration no, password etc. name, Registration no, password etc.
password etc. and press proceed.
Go to Procedure tab.
UI should facilitate the adjustment of critical To test whether the UI facilitates the adjustment of 1.Adjust the Power, Frequency, aming beam intensity
parameters for the surgery such asPower, Energy, critical parameters for the surgery such asPower, by sliding the slider to and fro. UI should facilitate the adjustment of critical parameters for UI facilitates the adjustment of critical parameters for the
13 DRS-UR-10 YES the surgery such asPower, Energy, Frequency, aiming beam surgery such asPower, Energy, Frequency, aiming beam Pass Usability Validation Report : 504-UVR-FL60
Frequency, aiming beam intensity, Lasing technique, Energy, Frequency, aiming beam intensity, Lasing 2.Adjust the pulse modes, standyby & ready modes, intensity, Lasing technique, Type, Fiber identification etc. intensity, Lasing technique, Type, Fiber identification etc.
Type, Fiber identification etc. technique, Fiber identification. presets, Lasing techniques auto manual etc by
pressing the tabs.
UI should have auto selection - pre-set program options To test whether the UI have auto selection - pre-set 1.Go to procedure tab UI should have auto selection - pre-set program options with UI have auto selection - pre-set program options with fixed
2.Select the Auto mode.
14 DRS-UR-11 with fixed value of energy, pulse mode & duty cycle for program options with fixed value of energy, pulse 3.By default fixed value of energy, pulse mode & duty YES fixed value of energy, pulse mode & duty cycle for particular value of energy, pulse mode & duty cycle for particular surgical Pass Usability Validation Report : 504-UVR-FL60
particular surgical procedure. mode& duty cycle for particular surgical procedure. surgical procedure. procedure.
cycle for particular surgical procedure appreas.
The system should have controls to facilitate the To inspect whether the system have controls to 1.Go to setting Tab. The system should have controls to facilitate the adjustment The system have controls to facilitate the adjustment of Audio,
16 DRS-UR-13 facilitate the adjustment of Audio, Lasing sound, 2.Adjust the Audio, Lasing sound, Brightness etc as per YES Pass Usability Validation Report : 504-UVR-FL60
adjustment of Audio, Lasing sound, Brightness etc. Brightness etc. requirement. of Audio, Lasing sound, Brightness etc. Lasing sound, Brightness etc.
System should have an electronic version of the To inspect whether the system have an electronic 1.Go to Help Tab. System should have an electronic version of the Instructions System have an electronic version of the Instructions for Use
17 DRS-UR-14 version of the Instructions for Use (IFU) and the 2.Click on the Instructions for Use (IFU) and YES Pass Usability Validation Report : 504-UVR-FL60
Instructions for Use (IFU) and the “Installation Manual" “Installation Manual" Installation Manual. for Use (IFU) and the “Installation Manual" (IFU) and the “Installation Manual"
System information should be displayed in one of the To check whether the system information is displayed 1.Go to Help Tab. System information should be displayed in one of the screens
18 DRS-UR-15 YES System information is displayed in one of the screens in the UI. Pass Usability Validation Report : 504-UVR-FL60
screens in the UI. in one of the screens in the UI 2.Click on the Machine Status. in the UI.
To test whether the display creates strain on User's 1.Go to setting Tab.
19 DRS-UR-16 The display should not create strain on User's eyes. 2.Adjust the Brightness usling slider bar as YES The display should not create strain on User's eyes. The display does not create strain on User's eyes. Pass Usability Validation Report : 504-UVR-FL60
eyes. requirement so as to not create strain on User's eyes.
Take the foot switch and inspect the color and coding.
There should be a color identification to foot switch To check if there is a color identification to foot switch 1.Grey colour marked as "A" is for Dusting and There should be a color identification to foot switch according
21 DRS-UR-18 according to dusting/fragmentation and according to dusting/fragmentation and Ablation. YES to dusting/fragmentation and Ablation/Coagulation There is a color identification to foot switch according to Pass Usability Validation Report : 504-UVR-FL60
dusting/fragmentation and Ablation/Coagulation techniques.
Ablation/Coagulation techniques. Ablation/Coagulation techniques. 2.Black colour marked as "B" is for Fragmentation and techniques.
Coagulation.
The system should have physical key switch to ON and To check if the system have physical key switch to ON 1. Switch On the system. The system should have physical key switch to ON and OFF the The system have physical key switch to ON and OFF the laser
23 DRS-UR-20 OFF the laser system and secure the laser system and OFF the laser system and secure the laser system 2.Set all the parameters for lasing. YES laser system and secure the laser system against unauthorized Pass Usability Validation Report : 504-UVR-FL60
against unauthorized use. against unauthorized use. 3.Turn On the key switch and then the lasing will start. use. system and secure the laser system against unauthorized use.
The system should run auto-calibration before letting To test if the system run auto-calibration before letting
25 DRS-UR-22 YES
the system ready for surgery. the system ready for surgery.
40 DRS-UR-38 The system should have label indications of safety, To check if the system have label indications of safety, 1.Check the system for safety, warning, cautions Yes The system should have label indications of safety, caution, The system have label indications of safety, caution, warning Pass Usability Validation Report : 504-UVR-FL60
caution, warning on the system for easy identification caution, warning on the system for easy identification labels. warning on the system for easy identification on the system for easy identification User Manual:
42 DRS-MR-02 The enclosure shall be durable and rugged construction. To inspect whether the enclosure is durable and rugged 1.Conduct Packaging and Transportation Validation Yes The enclosure shall be durable and rugged construction. The enclosure is durable and rugged construction. Pass Packaging validation Report:
construction. test and Verify the test report as per the standard Transportation Validation report:
44 DRS-MR-04 The enclosure shall have an arrangement to facilitate To inspect if the enclosure have an arrangement to 1.Check the position and dimensions of the USB port NO The enclosure shall have an arrangement to facilitate The enclosure have an arrangement to facilitate placement of Pass Refer drawing :
placement of USB port. facilitate placement of USB port. provision on the front bezel as per drawing. placement of USB port. USB port. Refer BMR
The system shall have an arrangement to facilitate a To inspect if the system have an arrangement to 1.Check the position and dimensions of the foot pedal The system shall have an arrangement to facilitate a provision The system have an arrangement to facilitate a provision for Refer drawing :
45 DRS-MR-05 connection provision on the front bezel as per NO Pass
provision for foot pedal connection. facilitate a provision for foot pedal connection. drawing. for foot pedal connection. foot pedal connection. Refer BMR
The system shall have an arrangement to facilitate an To inspect if the system have an arrangement to 1.Check the position and dimensions of the foot pedal The system shall have an arrangement to facilitate an The system have an arrangement to facilitate an arrangement Refer drawing :
46 DRS-MR-06 connection provision on the front bezel as per NO Pass
arrangement for fiber probe connection. facilitate an arrangement for fiber probe connection. drawing. arrangement for fiber probe connection. for fiber probe connection. Refer BMR
The system shall have a provision to facilitate an To inspect if the system have a provision to facilitate an 1.Check the position and dimensions of the air vents The system shall have a provision to facilitate an arrangement The system have a provision to facilitate an arrangement for Refer drawing :
47 DRS-MR-07 provision on the enclosure assembly and back cover as NO Pass
arrangement for air vents. arrangement for air vents. per drawing. for air vents. air vents. Refer BMR
The system shall have an arrangement to facilitate To inspect if the system have an arrangement to 1.Check the position and dimensions of the 10 inch The system shall have an arrangement to facilitate placement The system have an arrangement to facilitate placement of 10 Refer drawing :
48 DRS-MR-08 placement of 10 inch with capacitive touchscreen facilitate placement of 10 inch with capacitive capacitive touchscreen display provision on the front NO Pass
display. touchscreen display. bezel as per drawing. of 10 inch with capacitive touchscreen display. inch with capacitive touchscreen display. Refer BMR
The Enclosure design must have an arrangement to To inspect if the Enclosure design have an arrangement 1.Check the position and dimensions of the dust The Enclosure design must have an arrangement to facilitate a The Enclosure design have an arrangement to facilitate a dust Refer drawing :
49 DRS-MR-09 facilitate a dust protector for the Fiber Optic SMA905 to facilitate a dust protector for the Fiber Optic protector for the Fiber Optic SMA905 connector NO Pass
connector. SMA905 connector. provision on the front bezel as per drawing. dust protector for the Fiber Optic SMA905 connector. protector for the Fiber Optic SMA905 connector. Refer BMR
The enclosure shall have an arrangement to facilitate To inspect if the enclosure have an arrangement to 1.Check the position and dimensions of the key switch The enclosure shall have an arrangement to facilitate The enclosure have an arrangement to facilitate placement of Refer drawing :
50 DRS-MR-10 provision on the enclosure assembly and back cover as NO Pass
placement of key switch. facilitate placement of key switch. per drawing. placement of key switch. key switch. Refer BMR
The enclosure shall have an arrangement to facilitate To inspect if the enclosure have an arrangement to 1.Check the position and dimensions of the EMI/EMC The enclosure shall have an arrangement to facilitate The enclosure have an arrangement to facilitate placement of Refer drawing :
51 DRS-MR-11 placement of EMI/EMC switch to connect the power facilitate placement of EMI/EMC switch to connect the NO placement of EMI/EMC switch to connect the power cable and EMI/EMC switch to connect the power cable and turn on the Pass
cable and turn on the switch. power cable and turn on the switch. switch provision on the back cover as per drawing. turn on the switch. switch. Refer BMR
The system shall have an arrangement to facilitate To inspect if the system have an arrangement to
The system shall have an arrangement to facilitate placement The system have an arrangement to facilitate placement of a
placement of a blast shield to check any detection of facilitate placement of a blast shield to check any 1.Check the position and dimensions of the blast Refer drawing :
53 DRS-MR-13 efficiency decrease in laser emission or failures of the detection of efficiency decrease in laser emission or sheild placement provision as per drawing. NO of a blast shield to check any detection of efficiency decrease blast shield to check any detection of efficiency decrease in Pass Refer BMR
in laser emission or failures of the connected optical fibers. laser emission or failures of the connected optical fibers.
connected optical fibers. failures of the connected optical fibers.
To inspect if the enclosure have space to affix temper Refer drawing :
54 DRS-MR-14 Enclosure shall have space to affix temper proof labels. proof labels. 1.Check the labels positions as per drawings. NO Enclosure shall have space to affix temper proof labels. Enclosure have space to affix temper proof labels. Pass Refer BMR
The system shall have an arrangement to facilitate easy To inspect if the system have an arrangement to 1.Assemble the system as per drawings. The system shall have an arrangement to facilitate easy The system have an arrangement to facilitate easy assembly Refer drawing :
55 DRS-MR-15 assembly and disassembly .of the components for facilitate easy assembly and disassembly of the 2.Reverse the assembly procedure to disassemble the NO assembly and disassembly of the components for Pass
maintenance components for maintenance system. maintenance and disassembly .of the components for maintenance Refer BMR
Enclosure must meet Ingress protection standard of To test if the enclosure must meet Ingress protection of 1.Conduct the Ingress Protection test and Verify the
56 DRS-MR-16 IP20 IP20 as per IEC 60529 standard test report as per the standard IEC 60529. Yes Enclosure must meet Ingress protection standard of IP20 Enclosure meets Ingress protection standard of IP20 Pass Refer ingress Protection Report :
64 DRS-OR-03 Optical engine should have four numbers of fan To inspect if the optical engine have four numbers of 1.Check the Material data sheet. NO Optical engine should have four numbers of fan attached to Optical engine have four numbers of fan attached to heat sink Pass Refer drawing:
attached to heat sink with 56.5 CFM fan attached to heat sink with 56.5 CFM heat sink with 56.5 CFM with 56.5 CFM Refer MSDS
Optical engine should be kept safe in controlled To check if the optical engine is kept safe and not 1.Check the temperature and humidity of the
65 DRS-OR-04 NO Optical engine should be kept safe in controlled environment. Optical engine is kept safe in controlled environment. Pass Refer Temperature and humidity report
environment. exposed to heat or dust. controlled environment.
The Laser system shall use a Thulium Doped Fiber Laser To test the Laser system uses a Thulium Doped Fiber The Laser system shall use a Thulium Doped Fiber Laser The Laser system uses a Thulium Doped Fiber Laser working at
66 DRS-OR-05 working at 1.94-micron wavelength Laser working at 1.94-micron wavelength 1.Check the Wavelength Characteritisation report. YES working at 1.94-micron wavelength 1.94-micron wavelength Pass Refer Wavelength Characteritisation report
Laser shall be configured for both Dusting/ To test if the laser is configured for both Dusting/ Laser shall be configured for both Dusting/ Fragmentation and Laser is configured for both Dusting/ Fragmentation and
67 DRS-OR-06 Fragmentation and ablation/coagulation mode i.e., Fragmentation and ablation/coagulation mode i.e., 1.Check the temporal profile of output optical pulse. YES ablation/coagulation mode i.e., pulsed(QCW) and CW modes ablation/coagulation mode i.e., pulsed(QCW) and CW modes Pass Refer Temporal Profile Report
pulsed(QCW) and CW modes respectively. pulsed(QCW) and CW modes respectively. respectively. respectively.
68 DRS-OR-07 Laser system shall accommodate different sizes of To inspect if laser system accommodate different sizes 1. Check the compatibility of various Delivery fibers YES Laser system shall accommodate different sizes of Delivery Laser system accommodate different sizes of Delivery fibers Pass Refer Compatibility report
Delivery fibers e.g., 200, 365 & 550microns. of Delivery fibers e.g., 200, 365 & 550microns. with different core diameters. fibers e.g., 200, 365 & 550microns. e.g., 200, 365 & 550microns.
System shall have a provision for an aiming beam at To check if the system have a provision for an aiming System shall have a provision for an aiming beam at 520nm System have a provision for an aiming beam at 520nm (guide
69 DRS-OR-08 520nm (guide laser) to allow surgeon to position the beam at 520nm (guide laser) to allow surgeon to 1.Check the green guide laser beam through medical YES (guide laser) to allow surgeon to position the laser probe laser) to allow surgeon to position the laser probe before Pass Refer report
probe on white surface.
laser probe before active lasing. position the laser probe before active lasing. before active lasing. active lasing.
Optical characteristics and optical power shall be To verify the optical characteristics and optical power Optical characteristics and optical power shall be verified for Optical characteristics and optical power is verified for
70 DRS-OR-09 verified for optimum Dusting and Fragmentation and for optimumDusting and Fragmentation and ablation 1.Check reports of In-vitro trials. YES optimum Dusting and Fragmentation and ablation and optimum Dusting and Fragmentation and ablation and Pass Refer In-Vitro reports
ablation and coagulation in a Saline Solution and coagulation in a Saline Solution. coagulation in a Saline Solution coagulation in a Saline Solution
Laser shall not produce unwanted heat that can damage To test if the laser produce unwanted heat that can Laser shall not produce unwanted heat that can damage the Laser does not produce unwanted heat that can damage the
71 DRS-OR-10 the tissue during lithotripsy procedure. damage the tissue during lithotripsy procedure or not. 1.Check reports of In-vitro trials. YES tissue during lithotripsy procedure. tissue during lithotripsy procedure. Pass Refer In-Vitro reports
72 DRS-OR-11 Optical engine should give desired output power with To test optical engine gives desired output power with 1.Check output power Vs Current report. YES Optical engine should give desired output power with Optical engine gives desired output power with controlled Pass Refer output power Vs Current report
controlled input current. controlled input current. controlled input current. input current.
Optical engine should deliver output power after To test optical engine delivers output power after Optical engine should deliver output power after medical
73 DRS-OR-12 medical probe with minimum loss in coupler box to medical probe with minimum loss in coupler box to 1.Check output power Vs Current report. YES probe with minimum loss in coupler box to have a efficient Optical engine delivers output power after medical probe with Pass Refer output power Vs Current report
minimum loss in coupler box to have a efficient laser system.
have a efficient laser system. have a efficient laser system. laser system.
74 DRS-OR-13 There should be no overshoot and under shoot of the To test there is no overshoot and under shoot of the 1.Check the temporal profile of output optical pulse. NO There should be no overshoot and under shoot of the optical There is no overshoot and under shoot of the optical output in Pass Refer Temporal Profile Report
optical output in temporal profile of optical engine. optical output in temporal profile of optical engine. output in temporal profile of optical engine. temporal profile of optical engine.
Adequate driver and controller of TEC required to To check if adequate driver and controller of TEC are
89 DRS-ER-02 maintain the temperature stability of the pump laser used to maintain the temperature stability of the pump
assembly. laser assembly.
91 DRS-ER-04 Ethernet connectivity should be there in the system. To check if ethernet connectivity is there in the system. 1.Connect the system with ethernet port to laptop
2.Check the IP address.
1.Check the MSDS for components of
92 DRS-ER-05 Medical grade controlling system should be there. To check if medical grade controlling system is there. NO Medical grade controlling system should be there. Medical grade controlling system is there. Pass Refer MSDS
motherboard/SOM
Dual Foot Pedal Selectable switching circuit is required To check if dual Foot Pedal Selectable switching circuit 1.Test the footpedal connector with the system to Dual Foot Pedal Selectable switching circuit is required to Dual Foot Pedal Selectable switching circuit switches between
93 DRS-ER-06 to switch between CW and QCW modes. is required to switch between CW and QCW modes. change between CW and QCW modes. YES switch between CW and QCW modes. CW and QCW modes. Pass Usability Validation Report : 504-UVR-FL60
10.1", 1280*800 pixels, 16:9 colour TFT Display driving To inspect if 10.1", 1280*800 pixels, 16:9 colour TFT Connect all the power cables and display with the 10.1", 1280*800 pixels, 16:9 colour TFT Display driving circuit 10.1", 1280*800 pixels, 16:9 colour TFT Display driving circuit
94 DRS-ER-07 circuit is required along with driving circuit with Display driving circuit is used along with driving circuit motherboard and make the power ON. YES is required along with driving circuit with attached capacitive is required along with driving circuit with attached capacitive Pass Usability Validation Report : 504-UVR-FL60
attached capacitive touch panel with attached capacitive touch panel. 1. Go to UI and check the touch. touch panel touch panel
96 DRS-ER-09 EEPROM circuit arrangement is required to verify To inspect if EEPROM circuit arrangement is used to 1.Connect the probe to the system and check the YES EEPROM circuit arrangement is required to verify authorized EEPROM circuit arrangement verifies authorized Laser probes Pass Usability Validation Report : 504-UVR-FL60
authorized Laser probes verify authorized Laser probes probe indication on procedure tab. Laser probes
USB connectivity and supporting circuit is required for To inspect if USB connectivity and supporting circuit is 1.Connect the device to the USB port and check the USB connectivity and supporting circuit is required for USB connectivity and supporting circuit is able to diagnostic
97 DRS-ER-10 YES Pass Usability Validation Report : 504-UVR-FL60
diagnostic data transfer. used for diagnostic data transfer. auto detection of the device. diagnostic data transfer. data transfer.
Electronic circuit arrangement is required to ring a To inspect electronic circuit arrangement used to ring a 1.Connect the footpedal to the system and check the Electronic circuit arrangement is required to ring a buzzer Electronic circuit arrangement is able to ring a buzzer during
98 DRS-ER-11 buzzer during the time of foot pedal operation. buzzer during the time of foot pedal operation. beep sound on pressing the foot pedal. YES during the time of foot pedal operation. the time of foot pedal operation. Pass Usability Validation Report : 504-UVR-FL60
Wi-Fi connectivity should be enable for mirroring screen To verifiy if Wi-Fi connectivity is able to help for Wi-Fi connectivity should be enable for mirroring screen and Wi-Fi connectivity gets enable for mirroring screen and
1.Scan QR code on mobile device to get the UI control
99 DRS-ER-12 and controls on any authenticated mobile system(by mirroring screen and controls on any authenticated YES controls on any authenticated mobile system(by scanning controls on any authenticated mobile system(by scanning Pass Usability Validation Report : 504-UVR-FL60
scanning respective QR code) mobile system(by scanning respective QR code on mobile screen. respective QR code) respective QR code)
Key switch should be implemented to control the access To check if key switch is able to control the access of 1.Turn the key switch ON and OFF and check the Key switch should be implemented to control the access of
100 DRS-ER-13 YES Key switch is implemented to control the access of system Pass Usability Validation Report : 504-UVR-FL60
of system system. lasing. system
101 DRS-ER-14 EMI/EMC switch should be implemented for main To test if EMI/EMC switch is implemented for main 1.Connect the power cable to EMI/EMC switch and YES EMI/EMC switch should be implemented for main power EMI/EMC switch is implemented for main power control Pass Usability Validation Report : 504-UVR-FL60
power control power control. turn on the switch to power up the system. control
Thermally regulated exhaust fan should be To verify if thermally regulated exhaust fan is Thermally regulated exhaust fan should be implemented with Thermally regulated exhaust fan is implemented with low
102 DRS-ER-15 implemented with low noise temperature implemented with low noise temperature. 1.Check the noise with dB meter. NO low noise temperature noise temperature Pass Refer dB meter report
105 DRS-SR-01 System optics and photonics shall be in accordance with To test if system optics and photonics is in accordance 1.Conduct the test and Verify the test report as per Yes System optics and photonics shall be in accordance with Laser System optics and photonics is in accordance with Laser Pass Refer Report:
Laser standard iso 11145. with Laser standard iso 11145. the standard. standard iso 11145. standard iso 11145.
System shall be tested in accordance with IEC 60601-2- To test if the system is in accordance with IEC 60601-2- System shall be tested in accordance with IEC 60601-2-22,
22, Medical Electrical Equipment–Part 2-22: Particular 22, Medical Electrical Equipment–Part 2-22: Particular Medical Electrical Equipment–Part 2-22: Particular System is in accordance with IEC 60601-2-22, Medical
1.Conduct the test and Verify the test report as per Electrical Equipment–Part 2-22: Particular Requirements for
106 DRS-SR-02 Requirements for Basic Safety and Essential Requirements for Basic Safety and Essential the standard. Yes Requirements for Basic Safety and Essential Performance of Basic Safety and Essential Performance of Surgical, Cosmetic, Pass Refer Report:
Performance of Surgical, Cosmetic, Therapeutic and Performance of Surgical, Cosmetic, Therapeutic and Surgical, Cosmetic, Therapeutic and Diagnostic Laser
Diagnostic Laser Equipment standards. Diagnostic Laser Equipment standards. Equipment standards. Therapeutic and Diagnostic Laser Equipment standards.
System shall be tested in accordance with IEC 60601-1- To test if the system is in accordance with IEC 60601-1- System shall be tested in accordance with IEC 60601-1-2,
2, Medical Electrical Equipment – Part 1-2: General 2, Medical Electrical Equipment – Part 1-2: General Medical Electrical Equipment – Part 1-2: General System is in accordance with IEC 60601-1-2, Medical Electrical
1.Conduct the test and Verify the test report as per Equipment – Part 1-2: General Requirements for Basic Safety
107 DRS-SR-03 Requirements for Basic Safety and Essential Requirements for Basic Safety and Essential the standard. Yes Requirements for Basic Safety and Essential Performance – and Essential Performance – Collateral Standard: Pass Refer Report:
Performance – Collateral Standard: Electromagnetic Performance – Collateral Standard: Electromagnetic Collateral Standard: Electromagnetic Disturbances –
Disturbances – Requirements and Tests Disturbances – Requirements and Tests. Requirements and Tests Electromagnetic Disturbances – Requirements and Tests
System shall be tested in accordance with IEC 60601-1- To test is the system is in accordance with IEC 60601-1- System shall be tested in accordance with IEC 60601-1-6, System is in accordance with IEC 60601-1-6, Medical Electrical
6, Medical Electrical Equipment – Part 1-6: General 6, Medical Electrical Equipment – Part 1-6: General 1.Conduct the test and Verify the test report as per Medical Electrical Equipment – Part 1-6: General
108 DRS-SR-04 Requirements for Basic Safety and Essential Requirements for Basic Safety and Essential the standard. Yes Requirements for Basic Safety and Essential Performance – Equipment – Part 1-6: General Requirements for Basic Safety Pass Refer Report:
and Essential Performance – Collateral Standard: Usability
Performance – Collateral Standard: Usability Performance – Collateral Standard: Usability Collateral Standard: Usability
System software shall be validated in accordance with To test if the system software is in accordance with IEC 1.Conduct the test and Verify the test report as per System software shall be validated in accordance with IEC System software is validated in accordance with IEC 62304,
109 DRS-SR-05 IEC 62304, Medical Device Software – Software Life 62304, Medical Device Software – Software Life Cycle Yes 62304, Medical Device Software – Software Life Cycle Pass Refer Report:
Cycle Processes. Processes. the standard. Processes. Medical Device Software – Software Life Cycle Processes.
System shall be tested in accordance with IEC 60825- To test if the system is in accordance with IEC 60825- 1.Conduct the test and Verify the test report as per System shall be tested in accordance with IEC 60825-1:2014, System is in accordance with IEC 60825-1:2014, Edition 3.0,
110 DRS-SR-06 1:2014, Edition 3.0, Safety of Laser Products – Part 1: 1:2014, Edition 3.0, Safety of Laser Products – Part 1: Yes Edition 3.0, Safety of Laser Products – Part 1: Equipment Safety of Laser Products – Part 1: Equipment Classification and Pass Refer Report:
Equipment Classification and Requirements. Equipment Classification and Requirements. the standard. Classification and Requirements. Requirements.
111 DRS-SR-07 System shall be tested in accordance with IEC 60529 To test if the system is in accordance with IEC 60529 1.Conduct the test and Verify the test report as per Yes System shall be tested in accordance with IEC 60529 Degrees System is in accordance with IEC 60529 Degrees of protection Pass Refer Report:
Degrees of protection provided by enclosures (IP code). Degrees of protection provided by enclosures (IP code). the standard. of protection provided by enclosures (IP code). provided by enclosures (IP code).
The system should meet the biocompatibility The system should meet the biocompatibility requirements as The system meets the biocompatibility requirements as per
112 DRS-SR-08 requirements as per ISO 10993-1 ,ISO 10993-5 , ISO To test the system as per ISO 10993-1 ,ISO 10993-5 , 1.Conduct the Biocomptability test and Verify the test Yes per ISO 10993-1 ,ISO 10993-5 , ISO 10993-7 , ISO 10993-10 , ISO 10993-1 ,ISO 10993-5 , ISO 10993-7 , ISO 10993-10 , ISO Pass Refer Biocompstibility Report:
ISO 10993-7 , ISO 10993-10 , ISO 10993-11. report as per the standard
10993-7 , ISO 10993-10 , ISO 10993-11. ISO 10993-11. 10993-11.
The system should be packed in box with foam backed To test whether the system packed in box with foam The system should be packed in box with foam backed heavy The system is packed in box with foam backed heavy duty
1.Conduct Packaging Validation test and Verify the test
113 DRS-PL-01 heavy duty plastic shipper box to protect it from backed heavy duty plastic shipper box protect it from Yes duty plastic shipper box to protect it from damage during Pass Refer Packaging Validation report:
damage during shipping. damage during shipping.
report as per the standard
shipping. plastic shipper box to protect it from damage during shipping.
Label Should be attached to the system containing all To inspect if the Labels are attached on the system Label Should be attached to the system containing all Label are attached to the system containing all important data Refer Drawing:
114 DRS-PL-02 important data for the identification of laser system and containing all inportant data for the identification of 1.Check the system for all labels as per drawings. Yes important data for the identification of laser system and for the identification of laser system and specifications of the Pass
specifications of the mains supply. lasre system and specifications of the mains supply. specifications of the mains supply. mains supply. Refer BMR:
Take the foot switch and inspect the color and coding.
1.Grey colour marked as "A" is for Dusting and
116 DRS-PL-04 Foot switch shall be marked for Dusting/ Fragmentation To inspect whether the foot switch is marked for Ablation. Yes Foot switch shall be marked for Dusting/ Fragmentation and Foot switch is marked for Dusting/ Fragmentation and Pass Refer Drawing:
and Ablation/coagulation or should be colour coded. Dusting and Fragmentation or colour coded. Ablation/coagulation or should be colour coded. Ablation/coagulation and also colour coded. Refer BMR:
2.Black colour marked as "B" is for Fragmentation and
Coagulation.
Label should be affixed on the foot switch containing To inspect if the label is affixed on the foot switch
necessary information such as Part Name/Code, Brand, containing necessary information such as Part Label should be affixed on the foot switch containing Label are affixed on the foot switch containing necessary Refer Drawing:
117 DRS-PL-05 1.Check the footswitch for all labels as per drawings. Yes necessary information such as Part Name/Code, Brand, information such as Part Name/Code, Brand, Connection Pass
Connection details, rating, and symbols for Name/Code, Brand, Connection details, rating, and Connection details, rating, and symbols for identification. details, rating, and symbols for identification. Refer BMR:
identification. symbols for identification.
119 DRS-PL-07 The packaging should meet the packaging requirements To inspect that the system packaging meets ISO 11607- 1.Conduct the packaging validation test and Verify the Yes The packaging should meet the packaging requirements as per The packaging meets the packaging requirements as per ISO Pass Refer Packaging Validation report:
as per ISO 11607-1 and ISO 11607-2. 1 and ISO 11607-2 requirements. test report as per the standard ISO 11607-1 and ISO 11607-2. 11607-1 and ISO 11607-2.
120 DRS-PL-08 The labelling should comply the regulatory To inspect labels as pet the requirements. 1.Check the system for all labels as per standards Yes The labelling should comply the regulatory requirements (Ref The labelling complies the regulatory requirements (Ref Pass Refer Drawing:
requirements (Ref standard BS EN ISO 15223-1:2020) standard BS EN ISO 15223-1:2020) standard BS EN ISO 15223-1:2020) Refer BMR:
When transporting and storing the laser system, the To test if whether while transporting and storing the When transporting and storing the laser system, the When transporting and storing the laser system, the
1.Conduct transportation Validation test and Verify the
121 DRS-ST-01 temperature should be between -15 °C to +50 °C, the laser system, the temperature is between -15 °C to +50 test report as per the standard. Yes temperature should be between -15 °C to +70 °C, the relative temperature is between -15 °C to +70 °C, the relative humidity Pass Refer Transportation Validation report:
relative humidity 35% to 60% (non-condensing) °C, the relative humidity 35% to 60% (non-condensing) humidity 35% to 60% (non-condensing) 35% to 60% (non-condensing)
122 DRS-ST-02 The system should meet the transportation To inspect that the system meets transportation 1.Conduct transportation Validation test and Verify the
Yes The system should meet the transportation requirements as The system meets the transportation requirements as per Pass Refer Transportation Validation report:
requirements as per ASTM D4169. requirements as per ASTM D4169. test report as per the standard. per ASTM D4169. ASTM D4169.