Professional Documents
Culture Documents
Dyonics Users Manual
Dyonics Users Manual
DYONICS™ RF System
Glossary of Symbols
Preface
This equipment has been tested and found to comply with the limits for medical devices to IEC/EN 60601-1-2:2001.
These limits are designed to provide reasonable protection against harmful interference in a typical medical
installation. This equipment generates, uses, and can radiate radiofrequency energy and, if not installed and
used in accordance with the instructions, may cause harmful interference to other devices in the vicinity. If this
equipment does cause harmful interference to other devices, which can be verified by turning the equipment off
and on, the user is encouraged to try to correct the interference by one or more of the following measures:
• Reorient or relocate the receiving device
• Increase the separation between the affected equipment and the generator
• Connect the affected equipment to an outlet or circuit different from that to which the generator is connected
• Consult the manufacturer or field service technician for help
High frequency surgical unit with respect to electrical shock, fire and mechanical hazards only in accordance with:
IEC/EN 60601-1/CAN/CSA C22.2 No. 601.1, IEC/EN 60601-1-2, IEC/EN 60601-1-4, IEC/EN 60601-2-2, and IEC/EN
60601-2-18.
Table of Contents
Preface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 Maintenance and Troubleshooting . . . . . . . . . . . . . . . . . . . . . . 13
Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 Maintenance. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Indications for Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 Fuse Replacement. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Contraindications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 Alarms. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Principle of Operation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 Troubleshooting Guide . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Warnings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 Error and Fault Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Precautions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 Product Specifications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Adverse Reactions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 Technical Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Unpacking. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 Input Power Requirements. . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Assembly and System Check. . . . . . . . . . . . . . . . . . . . . . . . . . . 8 Output Power. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
System Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 Generator Dimensions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Front Panel Layout. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 Foot Pedal. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Rear Panel Layout. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11 Environmental Conditions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Instructions for Use. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12 Generator Output Graphs . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Operator Training Requirements. . . . . . . . . . . . . . . . . . . . . . . . . 12 Generator Classification and Safety Verification. . . . . . . . . . . 19
General System Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12 Classification. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Voltage Outputs. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12 Safety Verification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
System Preparation and Care. . . . . . . . . . . . . . . . . . . . . . . . . . . 12 Parts List. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
System Preparation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12 Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
RF Probe Selection. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13 Service Replacement Units Warranty. . . . . . . . . . . . . . . . . . . . 20
System Shut Down . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13 Service Replacement Program. . . . . . . . . . . . . . . . . . . . . . . . . 20
Cleaning and Sterilization. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13 Repair Service Program. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Surface Cleaning & Disinfection of Generator and Foot Pedal. . 13
DYONICS RF Probe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
energy to the treatment site via a DYONICS RF Probe • Plica Removal All Joints
with an integrated cable. The generator is activated • Scar Tissue All Joints
by the foot pedal. The disposable DYONICS RF • Soft Tissue All Joints
Probe is available in various single or multi-electrode • Synovial Membrane All Joints
configurations and is supplied separately. • Tendon All Joints
• Triangular Fibrocartilage (TFCC) Wrist
• The Smith & Nephew DYONICS™ RF System is 2. Turn the power switch on the front of the generator
designed to be operated exclusively as a unit. to the “ON” position. The system performs a
• Smith & Nephew fully warrants the safety and Power-On Self-Test routine. A series of digits
efficacy of its devices when used as intended for identifies the software version.
indications for which they are approved. Smith 3. Attach the foot pedal receiver to the tan receptacle
& Nephew cannot verify the safety of single-use (black outline) on the front of the generator. The
DYONICS™ RF Probes that have been reprocessed Foot Pedal Connected Indicator on the front of the
or reused. generator illuminates.
• The DYONICS RF System should not be used 4. Depress the Ablation function on the foot pedal.
adjacent to or stacked with other equipment. If the The red Warning Indicator on the front of the
system is used adjacent to or stacked with other generator illuminates and the generator emits an
equipment, the system should be verified that it is intermittent monotone alarm. The yellow Ablation
operating in its intended configuration. Activation Indicator on the front panel is off.
• Do not use other Smith & Nephew foot pedals. Note: Use only the foot pedal authorized by Smith
Use only the foot pedal provided with the & Nephew for use with the DYONICS RF System.
DYONICS RF System 5. Attach the DYONICS RF Probe to the cable
• When endoscopes are used with necessary receptacle (orange outline).The green RF Probe
accessories, the patient leakage currents may Indicator illuminates. The DYONICS RF System
be additive. defaults to a recommended power setting for
• The DYONICS RF System contains electronic each DYONICS RF Probe style to serve as a
printed circuit board assemblies and should not guide for safe and effective operation.
be disposed in any waste container. It should be Note: Initial and maximum settings are suggested
disposed of in accordance with any applicable settings. Proper activation of the DYONICS RF
national or institutional policies relating to obsolete Probe should always be confirmed.
electronic equipment. Contact Smith & Nephew
In order to adjust the settings, the DYONICS RF
for return of the DYONICS RF System for proper
Probe cable must be connected to the generator.
disposal. Dispose of the RF probe according
to normal institutional practices for potentially WARNING: Electrosurgical accessories
contaminated items. which are activated or hot from use can
cause a fire.
Adverse Reactions WARNING: Accessory tips may remain
As a consequence of electrosurgery, damage hot enough to cause burns after the
to surrounding tissue through iatrogenic injury electrosurgical current is deactivated.
could occur. WARNING: Localized burns to the patient or
physician may result from electrosurgical
Unpacking current carried through other instruments and
conductive objects.
Verify that all items have been received and are not
damaged. Report damage immediately to Customer CAUTION: Do not contact metal objects with an
Service. Save all containers and packaging material; activated RF probe.
they are required if it is necessary to return the CAUTION: Do not place active accessories near or in
equipment. contact with flammable materials (such as gauze or
surgical drapes).
Assembly and System Check 6. Taking care not to touch the DYONICS RF Probe
1. Connect the power cord to the receptacle on the end, depress the Ablation (yellow) foot pedal. The
rear panel of the generator. Connect the other end yellow Ablation Activation Indicator illuminates. DO
of the power cord to the electrical outlet. NOT use the generator if the alarms during system
check are not audible. If the generator does not
Note: If a power cord other than the DYONICS
function as described above, do not use. Please
power cord is used, ensure that the power cord
contact Customer Service immediately.
meets local requirements and complies with the
voltage and current rating listed on the back panel
of the generator.
System Overview
2 6
1
7 9 11
8
3 4 5 10
Figure 1. Generator front panel
13 17
Cable Receptacle Warning Indicator
20
Accepts Smith & Nephew DYONICS™ RF Probes Illuminates and alarms when a generator-specific
with orange mating ends. failure or malfunction occurs.
Clean with detergents and disinfectants according Intermittent dual tone Power limit alarm or
to standard practices. Disinfect with liquid chemical overload
disinfectants such as chlorine solutions, iodophors, Intermittent monotone Connection fault alarm
glutaraldehydes and hydrogen peroxides. Follow (DYONICS RF Probe, Over
manufacturer guidelines for concentration and length Temperature Condition,
of exposure. Over Power Condition)
DYONICS RF Probe
The DYONICS RF Probe is supplied sterile and
is intended for single use only. Do not clean,
resterilize, or reuse the DYONICS RF Probe as this
may result in product malfunction, failure, or patient
injury, which may also expose the patient to the risk
of transmitting infectious diseases. Please refer to
the Instructions for Use packaged with the DYONICS
RF Probe for specific information concerning DYONICS
RF Probe use.
Troubleshooting Guide
If you are experiencing problems with the Smith &
Nephew DYONICS™ RF System, use the following
troubleshooting guide to help identify or eliminate the
problem before contacting Customer Service.
Green Foot Pedal Check that the foot pedal is properly connected to the generator and that the cord is
Connected Indicator light not nicked, cut, or frayed. Do not use the foot pedal if the cord has been damaged.
does not illuminate If the problem persists, change the foot pedal. If the Foot Pedal Connected Indicator
light is still not illuminated, return the system for service.
Green RF Probe Make sure that the DYONICS RF Probe is properly connected to the generator. If the
Connected Indicator light problem persists, first change the DYONICS RF Probe. If the RF probe indicator light
does not illuminate when is still not illuminated, return the system for service.
a DYONICS RF Probe is
connected.
Nothing happens Verify that both the Foot Pedal Connected Indicator light and the RF Probe
when one of the device Connected Indicator light are illuminated when the foot pedal is depressed.
activation functions Check if the setting has been adjusted to a level appropriate for operation
on the foot pedal is (usually 1 or greater). Confirm that the DYONICS RF Probe tip and shaft are
depressed. covered by a conductive irrigant. Make sure that the DYONICS RF Probe is
properly connected to the generator. If the problem persists, replace the
DYONICS RF Probe. If the system still fails to operate, return the system for service.
The DYONICS RF Probe Generally indicates that the DYONICS RF Probe is not properly connected to the
does not activate, an generator. Check all connections. If the alarm continues to sound when the foot
intermittent monotone pedal is depressed, replace the DYONICS RF Probe. If the problem continues, return
alarm sounds, and a red the system for service.
warning light illuminates.
A dual tone alarm sounds This is a safety feature of the Smith & Nephew DYONICS RF System and may occur
and a red warning light if the DYONICS RF Probe is activated for an extended time without contacting tissue.
illuminates when the To reset the unit, step off the foot pedal. Make sure that the DYONICS RF Probe is
DYONICS RF Probe is close to or in contact with the target tissue and depress the foot pedal again. If the
activated. alarm continues, replace the DYONICS RF Probe. If the problem persists, return the
system for service.
Fault Foot pedal or front panel switch pressed during power on F/6
Generator Dimensions
Weight (max): < 5 kg (< 11 lbs)
Height: 10.2 cm (4.0 inches)
Width: 30.4 cm (12.0 inches)
Length: 40.2 cm (15.8 inches)
350
300
250
200
Power (Wrms)
150
100
50
0
C 1 2 3 4 5 6 7 8 9
Set Point
The graph below shows the output power (at full and half settings) versus load resistance
(per IEC/EN 60601-2-2, subclause 6.8.3).
350
300
250
200
Power (wrms)
Half Power
Full Power
150
100
50
0
0 200 400 600 800 1000 1200
Load Resistance (Ohms)
The following graph illustrates the relationship between instrument settings and output RMS voltage.
350
300
250
200
Voltage (Vrms)
150
100
50
0
C 1 2 3 4 5 6 7 8 9
Set Point
- Foot pedal meets requirements of IEC/ 72202141 DYONICS RF ABLATION 90° Probe
EN 60601-2-2, subclause 44.6, watertight with Integrated Cable
construction (IPX8). 72202144 DYONICS RF HEDGEHOG Probe
• Equipment not suitable for use in the presence with Integrated Cable
of a flammable anesthetic mixture. 72202143 DYONICS RF CURVE 45° Probe
• Mode of operation: capable of continuous with Integrated Cable
operation. 72202145 DYONICS RF HOOK 30° Probe
with Integrated Cable
Safety Verification 72202142 DYONICS RF BROADNOSE 90°
The Smith & Nephew DYONICS™ RF System meets the Probe with Integrated Cable
requirements of IEC/EN 60601-1, IEC/EN 60601-1-2, 72202146 DYONICS RF SLICE 20° Probe
IEC/EN 60601-1-4, IEC/EN 60601-2-2, CSA 22.2 with Integrated Cable
No. 601.1, and IEC/EN 60601-2-18. It is recommended
that the biomedical engineering department test 72202147 DYONICS RF CONTRACT 30° Probe
the system to ensure that it meets the following with Integrated Cable
leakage levels: 72202148 DYONICS RF ROUND 60° Probe
• Leakage current <100μA at 100-120/220-240 V~, with Integrated Cable
50/60 Hz isolated patient connections
• Leakage current <500μA at 100-120/220-240 V~,
50/60 Hz non-patient applied parts
If the system fails to meet the specifications listed
above, please contact Smith & Nephew Customer
Service for a return merchandise authorization.
The Smith & Nephew DYONICS RF System is covered by the following U.S. Patents: 5,697,909; 5,697,281; 5,697,536;
5,697,882; 5,683,366; 5,681,282; 5,766,153; 5,810,764; 5,843,019; 5,871,469. Additional patents issued and pending.
™Trademarks of Smith & Nephew. Certain marks registered U.S. Patent & Trademark Office.