Calibrations OD 38

You might also like

Download as pdf
Download as pdf
You are on page 1of 5
CALIBRATION House @) A Calibration Laboratory Calibration Certificate BUS Certificate No. | CH23/17011-10 | Date of Issue 17/01/2023 Pages | 1 of 1 Calibration Date. 17/01/2023 Next Recommended Due Date | 16/01/2024 ULR No. | CC230423000001445F_ Saaea Tes A Aline; 1 un Ge Rag esriag a | 3l-Sectot *C* Industral Aera, Sanwer Road, Indore — 452 015 (MP. ‘Customer Reference No.: | 3800 [Dated 7 16.01.2023 | Receipt Date 1 16.01.2023 Location of Calibration | At Calibration Laboraiony Work Order No. 1701 Description of UUC 1 PRESSURE GAUGE CMake Wael Seno Range Teast Count Excel ENES7-I EXLGII3A O=0.6kglent O01 kglem? Condition of UUC when Feeipt Visually O& Calibration o be dane [ wisdjustinent without adusiment Calibration Method | Our Calibration Procedure No, CH/CP/32; Based on DKD-R6-1; By Comparison Method ‘with Calibration Conditions: Used Pressure Transmitting Medium is Coolant and Mounting Position of Standard & UUC is Vertical at same Altitude. Environmental Conditions [Temperature:=23 1.5°C, Humidity 50 15% Rh ‘Master Equipment’s used for Calibration are traceable to National / International Standards through NABL Accredited Laboratory. SN. Deseription 1d. No. Certificate No. Uncertainty (2) [Calibrated By [Validity | (01 _| Digital Pressure Gauge | CH/DPG/O7 | CS1/LB/MP/704-01_| 0.0025/0.0053 bar_| NABL CC-2828 | 36.02.2025, ‘Unceriainty of Measurement (is Expanded Uncertainty at 95.45 % Confidence Level with coverage factor k= 2): Calibration Results for Mechanical - Pressure: (all readings are in ky/em*) ‘uc Value | Observed Standard Value” [Mean Valve | Observed UUC | sy cueaisn | Expanded sn. | sate ind (Ps) (Mi) Error 4p may | UaeeRanty Mi (Up) | M2 (Down) | (Mi+M2y2 | (Pind-Mi) @ 1 0.00 0.000 0.000 0.0000 0.0000 0.0000 2 0.10 ‘101 0.098. 0.9995 0.0005 0.0030 0.036 3 (020 (0.203 0.198. 0.2005 0.0005 0.0050, 0.031 4 0.30 ‘03310 0.304 03070 0.0070 0.0060 (0.025 3 0.40 O41 ‘0.405 0.4080 0.0080 0.0060. 0.021 6 0.60 0.612 0.605 0.6085 0.0085 0.0070 0.018 The SL nit of pres pascal (Pa). Above Given usm unit Hon ad is equal 98068 Pa GEER NO Me epee ! = Sy Aiton By? Seana 5 i Njan ne Tee Mang RRA R End HC GHicate HAR Cewn’ CHICCIPGI - Rev No. 0401 092022 ‘Note ¥ This certificate refers only Yo the fem calibrated.» A Calibration Sticker has been also affixed on UC. * Refer to Centficate before use. * UUC:- Unit Under Calibration. Reading has been rounded off, where ever applicable. 18 + Next recommended due date is decided by customer & mentioned as pe eustomer’s requisition. ‘The calibration results reported inthis certffeate are valid atthe time of & the stated condition of measurement. * This certificate shall not be reproduced except in full, without written approval of the issuing laboratory head, CCALIGRATION HOUSE 7, Trani Colony. A. Road, Dewas 455 001 (MP) | Ph #817272 967443, Mob - 094240 4051, 94066 61926, 68889 99400 Email: calbratoonouse@agmall com, cal@calbratonhouse.org CALIBRATION HOUSE e A Calibration Laboratory Calibration Certificate BEWAS Cerificate No. | CH/23/17011-11 _| Date of Issue _[ivoi2023 | __ Pages | Toft Calibration Date | 17/01/2023 Next Recommended Due Date _ 16/01/2024 ULR No. | CC230423000001446F aia Pana ao 1 Missa Engieetig '31-Sectot ‘C” Industrial Aera, Sanwer Road, | Indore ~ 452 015 (MP). ‘Customer Reference Nox [3800 [Dated [1601-2023 | Receipt Date] 16.01.2023 Location of Calibration | At Calibration Laboratory [ Work Order Ne._| 17011 Description oF UUC PRESSURE GAUGE Make Modal Sane. | ___1aNo. Range | __Leest- Count Excel ENSS7-1 EXLISIIOIA TEPGED T=1.06 kajem?™ 0.02 kg/en™ Condition of TUG when receipt Visually Of Calibration to be done ‘wicadjustment 7 without eustnent Calbration Method | Our Callbration Prosedure No. CHICDI32, Based on DKD-R6.1; By Comparison Method with Calibration Conditions : Used Pressure Transmitting Medium is Coolant and Mounting Standard & UC is Vertical at seme Altitude. ion of Environmental Conditions | Temperature:-23= 1.5°C, Humidity= 50 15% Rh ‘Master Equipment’s used_for Calibration are traceable to National / Intemational Standards through NABL Accredited Laboratory. SN. Deseription Td. No. Certificate No, Uncertainty () [Calibrated By | Validity (01 | Digital Pressure Gauge | CH/DPG/O7 | CS/21/LB/MP/70s-01_| 0.0025/0.0053 bar | NABL CC-2828 | 26.02.2025 Uncertainty of Measurement (is Expanded Uncertainty at 95.45 % Confidence Level with coverage factor k: Calibration Results for Mechanical - Pressure: (all readings are in kg/cm? ) ‘uC Value | Observed Sandard Vale [Mean Value | Obsered UC | ysis | Exnanded SN, | Setto Pind) a ne but imam] | Uneetinty a Mi @p) | M2 (Down) | i+w | (Pind-Miw) ) 1 0.00 0.000 0.000 0.0000 0.0000 0.0000 as 2 0.20 0.206 0.190 0.1980 0.0020 0.0160. 0.096 3 0.40 0.408 0.389 0.3985, 0.0015 0.0190. 0.061 4 [040 0.614 0.596 0.6080 0.0050 0.0180 0.002 5 0.80 0.818 0.789 0.8035 -0.0035, 0.0290. 0.055. 6 1.06 1.079 1.012 1.0455 0.0145 0.0670 0.108 The SI unit of pressure is pascal (Pa). Above Given results is in unit kg/em? and is equal to 98066.5 Pa. gM) SU Aitorised By = Mr, Nayan Jain:- Technical Manager CHICCIPG! —Rev No. 0401.092022 ‘Note = * Ths ceiicate refers only To the ilem calibrated, * A Calibration Susker has been also afixed on UUC. * Refer to Cemificate before use. * UC: Unit Under Calibration. Reading has been rounded off, where ever applicable. + Next recommended due date is decided by customer & mentioned as per customer's requisition. * The calibration results repored in tis certificate are valida the time of & the sated condition of measurement This corificate shall not be reproduced except in ful, without written approval of the issuing laboratory he CCaLiBRATION HOUSE 7 Tarani Colony, A.B. Road, Dowss 455001 (MP) | Ph +1 7272 -367443, Mob. -084240 45081, 94086 61326, 68689 99490 E-mai calbraloorouse@gmal.con,calfcalbrsionhouse org CALIBRATION House A Calibration Laboratory be Calibration Certificate EHeS SHABITOIIO2 | Date of sue [712023 | Pages [Toft ‘Next Recommended Due Date | 16/01/2024 | ULRNo. | CC230423000001437F | cxeaats ea cakes 7 Ms Jah Engineering Li. | |e St sesee otal Asa, Sucer Read, |___ Indore 452 015 (M.P.) Customer Reference Noz | 3800 [Dated [16.01.3023 | Receipt Date [16.01.2003 (Location of Calibration | At_ Calibration Laboratory ‘Work Order Na._| 17011 Description of UUC T FEELER GAUGE ‘Make Model ___|_____StNo. __IdNo, Range [—“‘Niat7 Globe = = G10 mm (Condition of UUC when receipt | Visually O& (Calibration to be done “witheadjustment 7 without adjustment tion Procedure No. CHICP/I2; Based on 1S 3179-1990. | Temperature:-20:+ °C, Humidity:- 502 10% Ri ‘Master Equipment’s used for Calibration are traceable to National/ International Standards through NABL Acsredited Laboratory. SN. ‘Description Td. No. Cerificate No. Calibrated B) ‘Validity 1 | Length Measuring Machine | CH/LMMIOT | UMM21/5268 NABL CC-2421 28.02.2023 (Metroscope) Unceriainty of Measurement (is Expanded Uncertainty at 95.45 % Confidence Level with coverage factor k= 2):= 0.0018 mm Calibration Results for Mechanical - Dimensional: all readings are in mm) TENo, | Nominal | Observed Standard Value Size of UUC [Minimum [Maximum TEC | 0.10 0.1002 0.1008: JEECAS | 0.10 0.1006 0.1018. TOR — mae (Oma ‘CHICCIFO —Rey No. 0401.092022 Note = * Ths ceificate refers only to the item calibrated. » A Calibration Sticker has been also affixed on UC. Refer to Centficate before use. * UUC:- Unit Under Calibration, Reading has been rounded off, where ever applicable. fe) Next recommended due date is decided by customer & mentioned as per customer's requisition. The calibration results reported in this certificate are valid atthe time of & the stated condition of measurement, e * This certificate shall not be reproduced except in fll, without writen epproval of the issuing laboratory head. a ‘Cauiararion HOUSE — 71, Tari Colony A.B. Road, Dewas 455 001 (MP) | Ph. +91 7272 -357483, Mob, -094240 45051, 94086 61326, 8899 93400 small: cabvabanhouse@gmal com, cal@ealibralonnouse org CALIBRATION House A Calibration Laboratory be Calibration Certificate DSwES Ceciiaie a | earBTTOTTAT | Due ote [nova [Page [ori Calibration Date | 17/01/2023 ‘Next Recommended Due Date | 16/01/2024 | ULRNo. | CC230423000001436F Coane Go Biter an epee 31-Sectot ‘C’ Industrial Aera, Sanwer Road, Indore — 452 015 (MP.) Customer Reference No | 3800 [Dated [160123023 | Receipt Date | 16.012025 Location of Calibration | At Calibration Laboraiony ‘Work Order No._| 17011 VERNIER CALIPER, cs ~_[___Model ENO] _1aNo. | __‘Range Teast Count Mituto 330-118 30864670, TENE? 0=200 mm (0.02 mm Condition of UUC when receipt Visually OF Calibration to be done iad tment without adjustment ‘Callration Method | Our Calibration Procedure No, CHICPIOS ; Based on IS 16491 (Part 12016 Environmental Conditions Temperature:-20=2°C, Humidity:- 50: 10% Rh “Master Equipment’ used for Calibration are traceable to National / International Standards through NABL Accredited Laboratory. (sx. ‘Description 1d. No. ‘Cerificate No. Calibrated By [Validity [Co Caliper Cheexer CHICCION | CHTTISOIC-O1 NABL CC-2304 | 18.04.2023, 02_| Slip Gauge Set ‘CHISL02 | SCL21-22/03/108-001 ‘NABL CC2282, | 10.03.2004 (03_| External Micrometer CHIMM/OI_| CH?23/08014-06 NABL CC-2304 | 07.01.2024 (04 | Thread Measuring Wire CHTMWAT | CH/21/0103A-05, ‘NABL CC2304 | 28.02.2023 05 _| Surface Plate (G) CHISP/O_| CH/22/I9124-01 ‘NABL CC-2304 | 18.12.2024 Uncertainty of Measurement (is Expanded Uncertainiy at 95.45 % Confidence Level with coverage factor €=2):+ 0.0174 mm. Calibration Results for Mechanical - Dimensional: (all readings are in mm) ‘Extemal Measurement Tnternal Measurement Standard Value WUC Value. WUC Error Standard Valoe ‘WUC Value WUC Error 0.00 0.00 0.00 20.00 19.90 0.10 21.02 21.02, 0.00. 700.00 99.90. 0.10 31.10 51.10, 0.00 200.00 199.88 0.12 700.00) 100.00 0.00 Partial Surface Contact Error | 0.000 200.00 7200.00 0.00 Tine Contact Error of External Faces 0.020 Depth Measurement Parallelism of Internal Jews 0,080 20.00 20.00 0.00, Seale Shift Error |__ 0.100) 100.00 100.00 0.00) TWote= Readings are taken at Center of he Jaws, Te above resulls are average of ig fvge observations, eS SR ore: Mc. Nayan Jain - Technical Manager amare CCHICCIVCI — Rew No. woL.082022 Nowe This ceificae refers only tothe lem callbraved. >A Calibration Sticker has been also affixed on UC. * Refer to Certificate before use. * UUC:- Unit Under Calibration. Reading has been rounded off, where ever applicable, * Next recommended due date is decided by customer & mentioned as per customer's requisition * The ealibraton results reported inthis certificate are valid atthe time of & the stated condition of measurement. * This certificate shall not be reproduced excoptin full, without written approval ofthe issuing laboratory head. (CAUBRATION HOUSE — 71, Tarai Colony, A.B. Rosd, Dewes 455.001 (MP) | Ph. 291 7272 - 357449, Mob. - 094240 45051, 94066 61326, 88899 93490 Eri: cabbraioohouse@gma.com,cal@ealbraionhovse org CALIBRATION House A Calibration Laboratory b d Calibration Certificate ee Cc - 2304 Cenifcate No.__| CHRS/ITONI-O6 | Date of issue Towe07s | __Pages | of Calibration Date | 177017023 ___| Next Reconsmended Due Date | 16/01/2024 | ULRINo. | CC230423000001441F Customer's Name & Address: | Mis Jash Engineering Lid. 31-Sectot ‘C” Industrial Aera, Sanwer Road, |__Indore ~ 452 015 (MP), Cisiomer Reference Nov [3800 ~ | Dated] 16.0130 | Receipt Date | V60T 205 Location of Calibration | At Calibration Laboraton” ‘Work Order No 17011 Deseipion of UUC ‘MEASURING TAPE Gee) [ie ano, Range J — Freemans SEMTAT 0-5 mw Condition of UUC when receipt | Visually Ot Calibration to be done I jtment / without adjustment, (Calibration Method | Our Calibration Procedure No. CH/CP/31; Based on IS 1269 (Part 2 for Stee) Environmental Conditions [Temperature:-202°C, Humidity- 50 10% Rh ‘Master Equipment’s used for Calibration are traceable to National International Standards through NABL Acoredited Laboratory. SN. Description 1d. No. Certfieate No. Calibrated By [Validity O1 | Tape & Seale Messuring Machine | CHIMSTCIOT | CH22/18048-01, NABL 6¢.2304_| 17.04.2023 Unceriinty of Measurement (is Expanded Uncertainty at 95.45 % Confidence Level wih coverage factor k= 2): = 0.714 mm Calibration Results for Mechani = Dimensional: _(all readings are in mm) ‘Sr. | Nominal UUC ‘Observed No.| Value ‘Standard Value 1 500 300.816 2 7000. 1000.524 3 2000. 200.416 4 3000. 300.629 a 3000. 000.849 ‘Authorised By > (Calibrated By : Ms. Avani Jain Mr, Nayan Jain - Technical Manager CHICCIMET ~Rev Ne, 0401092022 TNote:=* This ceificate refers only (othe Vem calibrated. * A Calibration Sticker has been also aixed on UC. * Reler to Certificate before use. * UC:- Unit Under Calibration. Reading has been rounded off, where ever applicable. * Next recommended due date is decided by customer & mentioned as per custome’ requisition. * The calibration results reported inthis certificate are valid atthe time of & the stated condition of measurement. * This certificate shall not be reproduced except in ful, without writen approval of the issuing laboratory head. CALIBRATION HOUSE 71, Taoni Colony. A 8. Rad, Dewae 485 001 (MP) | Ph, +91 7272-35743, Mob. 094240 46051, 84066 61528, 88899 93480 Ema: calfbratonhouse@gmal.com,cal@esibvatonhouse org

You might also like