Download as pdf or txt
Download as pdf or txt
You are on page 1of 5

Placenta (2005), 26, 353e357

doi:10.1016/j.placenta.2004.07.002

SHORT COMMUNICATION

Dissolved Oxygen Concentration in Culture Medium: Assumptions


and Pitfalls

D. Newby, L. Marks and F. Lyall*


Maternal and Fetal Medicine Section, Institute of Medical Genetics, Yorkhill, Glasgow G3 8SJ, UK
Paper accepted 8 July 2004

Oxygen is a key factor in the regulation of cytotrophoblast differentiation, proliferation and invasion in early pregnancy.
Abnormalities in oxygen concentration have also been linked to a number of pregnancy disorders. Cell culture models have been
used to study the effect of oxygen on cytotrophoblast behaviour in vitro, however, there is often little or no validation of oxygen
levels in these cell culture systems. In this study, dissolved oxygen levels in culture medium maintained in standard culture
conditions (18% O2) measured 18%. On transfer to a low oxygen environment (2% O2), oxygen levels decreased to 6e8% after
4 h and reached 2% only after 24 h in culture. Culture medium pre-gassed with nitrogen to remove dissolved oxygen quickly
absorbed oxygen when exposed to ambient air during dispensing and required further incubation in a 2% oxygen environment
before dissolved oxygen levels equilibrated to 2%. Thus, cultured cells placed in a low oxygen environment would be exposed to
varying levels of oxygen before the desired level of oxygen exposure is reached. This study highlights the importance of validation
of oxygen levels and potential problems associated with in vitro studies on the regulatory effects of oxygen.
Placenta (2005), 26, 353e357 Ó 2004 Elsevier Ltd. All rights reserved.

INTRODUCTION others make no mention of oxygen levels in their culture


systems other than in the incubator environment [6,9,10]. The
A number of important clinical conditions such as pre- aim of this study was to define the oxygen environments
eclampsia, fetal growth restriction and spontaneous mis- encountered in our cell culture systems and to address some of
carriage have been linked to the placenta being exposed to the issues and potential problems associated with studies of
abnormal oxygen concentrations [1]. Women living at high this nature.
altitude are exposed to chronic hypoxia and they too have
increased risk of complications of pregnancy [2], underscoring METHODS
the involvement of oxygen. Cytotrophoblast proliferation and Cell culture incubators
differentiation appears, at least in part, to be regulated by
oxygen tension. These observations have been the driving Studies were carried out using two Forma Scientific water-
force for investigators to develop in vitro models to determine jacketed incubators (Forma Scientific, Inc. USA), one set at
the effects of different oxygen concentrations on cell culture 5% CO2 in air for standard culture conditions with the sensor
models such as primary trophoblast cultures and placental on the incubator reading approximately 18.2% O2 and one set
explant culture [3e7]. One of the key issues in studies of this at 5% CO2/93% N2 with the oxygen sensor reading
nature is the actual oxygen levels which cultured cells are approximately 2% O2 for low oxygen conditions. Oxygen
exposed to during incubation in different oxygen environ- levels in each incubator environment were measured using
ments. Some authors report specific methods to control the a Fyrite Gas Analyser.
oxygen environments to which the cells are exposed and/or
Measurement of dissolved oxygen levels
the measurement of oxygen in culture medium [3,5,7,8], while
There are many commercially available oxygen probes and
* Corresponding author. Tel.: C44 141 201 0657; fax: C44 141 357
meters, of varying specifications, for the measurement of
4277. dissolved oxygen in a liquid environment but few are ideal for
E-mail address: f.lyall@udcf.gla.ac.uk (F. Lyall). the measurement of dissolved oxygen in cell culture systems
0143e4004/$esee front matter Ó 2004 Elsevier Ltd. All rights reserved.
354 Placenta (2005), Vol. 26

such as those used in trophoblast and explants cultures. A 20


Jenway 970 Portable Dissolved Oxygen/(C Meter and

Dissolved oxygen (%)


Electrode (Jencons Scientific Ltd, UK) was chosen for the 15

measurement of dissolved oxygen in this study due to the cost


10
and the suitability of the technical specifications, in particular
the integral temperature compensation. Issues relating to the 5
measurement of oxygen in cell culture systems and the factors
that influenced the choice of oxygen probe used in this study
0 5 10 15 20 25
are discussed in more detail later in this communication.
Time (h)
The oxygen meter was calibrated, according to the
manufacturer’s instructions, to 21% oxygen in water saturated Figure 1. Dissolved oxygen concentration (% O2) in test culture medium
incubated at 37 (C in hypoxic conditions for up to 24 h. Oxygen levels were
air and to 0% oxygen using the zero salts supplied with the measured using a Jenway Model 970 Dissolved Oxygen Meter and Electrode
meter. In order to satisfy the requirements of the oxygen (Jencons Scientific Ltd, UK).
electrode, experiments were carried out using approximately
125 ml of M199 (supplemented with 10% fetal bovine serum
then dispensed into smaller volumes (5 ! 25 ml) to mimic
and 1% antimycotic/antibiotic solution) (GibcoBRL Life
culture medium changes (total dispensing time 4 min) and
Technologies, UK) in a T75 vented-lid cell culture flask. This
then returned to the T75 flask. Dissolved oxygen in the culture
was necessary so that the depth of the culture medium in the
medium was measured immediately in the 125 ml of culture
upright flask was sufficient to cover the temperature compen-
medium and the flask placed in the low oxygen incubator at
sating element of the electrode and allowed stirring of the
37 (C. Dissolved oxygen levels were measured at intervals up
culture medium during measurement. The flask was placed flat
to 16 h.
in the incubators to allow maximum surface area for gaseous
exchange. The surface area/volume in the T75 flask was
RESULTS
75 cm2/125 ml and this compares to 8 cm2/4 ml in 35 mm
culture dishes, 1.9 cm2/0.4e1 ml in 24 well plates and 0.6 cm2/
Oxygen levels, measured using a Fyrite Gas Analyser, were
200 ml in the Millicell inserts used by many investigators for
18e20% in the standard incubator and 2e3% in the low
trophoblast and explant culture experiments. All experiments
oxygen incubator.
were performed in triplicate and repeated three times.
Experiment 1
Experiment 1
The dissolved oxygen concentration in culture medium
Culture medium was warmed to 37 (C in T75 flasks in maintained in standard culture conditions measured approx-
standard culture conditions (5% CO2 in air). Dissolved oxygen imately 18% (pO2 140 mmHg). When culture medium was
levels were measured and the flasks were then transferred to transferred to the low oxygen incubator, oxygen levels
low oxygen conditions at 37 (C for up to 24 h. Dissolved decreased gradually to measure 6e8% (46e61 mmHg) after
oxygen levels in the culture medium were measured at 4 h, 7% (53 mmHg) after 8 h, 3% (23 mmHg) after 16 h and
intervals during the incubation period. Separate flasks were 2e3% (15e23 mmHg) after 24 h of incubation (Figure 1).
used for each time point to prevent the medium being affected The pH was unaffected by the changes in oxygen concentra-
by atmospheric oxygen. tion (pH 7.2e7.4).

Experiment 2 Experiment 2
Nitrogen gas (Cryoservice, UK) was bubbled through culture When nitrogen gas was bubbled through culture medium,
medium (2e3 psi) (125 ml) in a T75 flask for 30 min to dissolved oxygen levels fell from 18e20% to approximately
eliminate dissolved oxygen in the culture medium. The 1.5% (11 mmHg) after 15 min and to 0% after 30 min. When
opening to the flask was sealed with parafilm through which pre-gassed culture medium (0%) was placed directly into the
there was an inlet for the pipette delivering the nitrogen gas incubator with the 2% oxygen environment, oxygen levels in
and an outlet to allow gas to escape. The flask was immediately the culture medium equilibrated to approximately 2% after 1 h
transferred to a 2% oxygen environment at 37 (C. Dissolved and remained at 2e3% (15e23 mmHg) following overnight
oxygen levels in the culture medium were measured after 1 h incubation.
and following overnight incubation.
Experiment 3
Experiment 3
When culture medium pre-gassed with nitrogen to eliminate
Nitrogen gas was bubbled through culture medium (125 ml) in oxygen (0%) had been dispensed in ambient air, dissolved
a T75 flask for 30 min to eliminate dissolved oxygen in the oxygen levels increased to 8.5% (65 mmHg). Further in-
culture medium, as described above. The culture medium was cubation of the cell culture medium in a hypoxic environment
Newby et al.: Dissolved Oxygen Concentration in Culture Medium 355

of 2% oxygen for approximately 16 h was required for prior exposure to atmospheric oxygen or standard culture
dissolved oxygen levels in the culture medium to decrease to conditions decreased gradually to 2% over an incubation
2% (15 mmHg). period of 24 h. During this time cultured cells would be
exposed to varying oxygen concentrations and degrees of
DISCUSSION hypoxia at the cell/culture medium interface depending on
the length of time in culture and before the desired oxygen
There are many factors that must be given careful consider- concentration, in this case 2%, is reached. These experi-
ation when measuring dissolved oxygen levels in a cell culture ments were carried out using a large volume of culture
system. Although many oxygen electrodes and meters are medium due to the requirements of the oxygen electrode in
available, not all are suitable or ideal for use with cell cultures. use. Presumably, with the much smaller volumes of culture
Many probes are simply too large for the direct measurement medium used in our actual cell culture experiments the rate
of oxygen in cell cultures maintained in, for example, 35 mm of gaseous exchange between the culture medium and the
dishes, 24-well plates or Millicell inserts. Smaller probes, such environment would be greater and hypoxic conditions would
as the ISO2 electrode (World Precision Instruments, Inc, be established more rapidly. However, problems associated
USA) with a 2 mm diameter tip, are available but often at with culturing cells in a static environment, which may limit
much greater expense. Temperature is of vital importance as the rate of gaseous exchange, still remain. A rocking
temperature effects the concentration of dissolved oxygen in platform placed inside the incubator may be useful in
a liquid. The MI-730 electrode (Microelectrodes, Inc, USA) is instances where cells are firmly attached to substrate to allow
quoted by the manufacturers as having a change in probe diffusion. However, the authors should ensure that effects
response of 2.2% per degree change in temperature. Thus, observed are not due to flow/shear stress per se.
unless an electrode has an integral temperature compensation One way of circumventing the delay in establishing the
function (e.g. Jenway 970, Jencons Scientific Ltd, UK), care appropriate culture conditions may be to pre-gas the culture
must be taken to ensure that all measurements, including medium, either with nitrogen to eliminate oxygen from the
calibration, should be carried out at the same temperature (i.e. medium or with a gas of the desired mix and oxygen level. As
37 (C in the case of cultured cells). This is not always possible we have shown, after pre-gassing with nitrogen to 0% oxygen,
for a variety of reasons depending on the method used for dissolved oxygen levels will equilibrate to 2% and remain at
calibration. For example, many manufacturers recommend that level when maintained within a 2% oxygen environment.
calibration in ambient air, while pre-gassing liquids, either However, manipulation and microscopic monitoring of
for calibration or to establish defined oxygen levels, results in cultured cells is often carried out in ambient air, thus
changes in liquid temperature that are difficult to control. exposing the culture medium and cells to atmospheric oxygen.
Trophoblast cells and villous explants are most commonly We have demonstrated that culture medium that has been
cultured in a static environment. However, dissolved oxygen pre-gassed with nitrogen to eliminate oxygen quickly absorbs
measurements cannot accurately be carried out in a static sys- atmospheric oxygen on exposure to ambient air. Further
tem due to oxygen starvation at the membrane of the electrode. incubation in a 2% oxygen environment for 16 h was required
This would result in false low oxygen readings, while agitation for dissolved oxygen levels in the culture medium to
of cultures may dislodge cells or explants. Thus it would be equilibrate back to 2%, thus limiting the advantage of pre-
difficult to obtain accurate measurements in many cell culture gassing the culture medium. Gassing cultures directly to
systems. Some authors have reported using a blood gas analyser avoid uptake of atmospheric oxygen also has inherent
to measure dissolved oxygen in culture medium sampled from difficulties particularly when millicell dishes are used. Such
cell cultures [7]. We initially used a Bayer 288 blood gas difficulties include dislodging the cells or explants during
analyser (Bayer, Germany) in our experiments. However, this gassing, the cooling effect of the gas on the culture medium
was not suitable as the analyser could only accurately measure and the time involved in gassing individual cultures in an
oxygen levels in whole blood (correspondence with manufac- experiment. Furthermore, cultures maintained in a low
turer) and gave false readings with culture medium. Kilani oxygen environment would be exposed to increased oxygen
et al. [11] have reported their experience using a blood gas levels during handling and during the associated time delay
analyser and the requirement for addition of whole blood to before the cell culture environment equilibrates to appropriate
the medium to improve the accuracy of readings. levels on return to the incubator, thus disrupting the exposure
In order to characterise oxygen levels within the oxygen of the cells to the required oxygen concentration. The only
controlled cell culture environments used in our laboratory way to avoid uptake of atmospheric oxygen would be to use
in a way that best compensated for the factors discussed a chamber flushed with the appropriate gas mix to allow
above, we chose an oxygen electrode with automatic temper- manipulation of cultures in the same oxygen environment as
ature compensation and measured dissolved oxygen levels in the incubator atmosphere [3] and then to be able to transfer
test culture medium maintained within the same environ- the dishes to the incubator without exposure to any
ments as our cultured cells, rather than directly in cell cul- atmospheric air.
tures themselves. Dissolved oxygen levels in culture medium This work highlights the importance of, and difficulties
incubated in a hypoxic environment of 2% oxygen following associated with, the validation of oxygen levels in vitro cell
356 Placenta (2005), Vol. 26

culture studies investigating the effects of oxygen on 4. If the culture dishes are to be removed for examination
placental development and function. Our studies focused under the microscope, authors should state how this
on plastic dishes since this is what most workers use. aected the oxygen concentration.
However, studies with glass dishes may, because of differ-
ences in oxygen retention, reveal other issues. The issues
highlighted are particularly important with regard to Group 2 recommendations
reproducibility and the comparison of data between different
5. Measurement of the oxygen concentration in the culture
research laboratories.
medium should be made using an oxygen electrode with a
Based on our experiences we propose a number of
tip size which can be fully immersed in the culture medium.
recommendations when investigators perform culture experi-
A number of companies including Jencons (http://
ments using varying oxygen conditions. Clearly, different
www.jenway.com) and Diamond General (http://www.
groups will be asking different questions in their model
diamondgeneral.com) sell a range of electrodes with dif-
systems and therefore the recommendations can only be used
fering tip sizes. The smaller tip electrodes are more ex-
as a guideline. These have been sub-divided into two
pensive. These must be calibrated before use according to
categories; Group 1: what this paper’s results have provided
the manufacturer’s instructions. The electrodes must have
evidence to support and Group 2: what the authors’ ‘‘expert
an integral temperature compensation function as temper-
opinion’’ adds to the recommendations.
ature affects the dissolved oxygen in the culture medium.
6. We do not recommend shaking the culture dish contain-
ing cells to measure oxygen concentration when villous
Group 1 recommendations
explants are cultured on matrigel since this may dislodge
1. The oxygen concentration in the incubator or chamber them slightly or completely thus altering their invasive
should be validated using a suitable device such as the potential. However, gentle shaking of a control dish which
Fyrite Gas Analyser used in the present study. contains medium but no explant will give the researcher
2. Authors should state the volume of medium per area of an idea of how the oxygen concentration has changed as
culture dish and how long it took to reach the required a result of the electrode being in a static environment.
oxygen concentration before the experiment in question 7. Authors should specify the oxygen concentration of the
started. culture medium that the placental villous tissue was
3. Authors should specify if the culture medium was collected into after delivery of the placenta. The culture
changed during the course of the experiment and how medium should be gassed to the appropriate level in the
this aected the oxygen concentration. Ideally medium laboratory in a universal container and then the sample
should be changed in a chamber flushed with the desired could be quickly dropped into it at collection. They
oxygen concentration and with medium that has been should state how long it took from collection to
pre-bubbled to the required oxygen concentration. We do preparation and the conditions used to dissect the piece
not recommend bubbling directly into culture dishes of tissue. Ideally all procedures should be performed in
where cells are growing. A gloved box cabinet would be the starting oxygen concentration of the experiment. We
required with the pipette and pre-gassed medium placed feel it would be unrealistic and technically very dicult to
inside the box before the box was flushed to obtain the isolate single trophoblast cells in a low oxygen environ-
desired oxygen concentration. Pipettes or tips should be ment because of the many steps involved.
flushed once with medium to remove air trapped inside 8. After consideration of the above manipulations in their
the pipette. At this point the lid of the culture dish can be experiment it would be helpful to other readers if the
removed. However, authors should bare in mind that authors then defined whether, on balance, their culture
even with the lid on air exchange will still be taking place, system represents a model of hypoxia or a model or
albeit more slowly. re-perfusion.

ACKNOWLEDGEMENT
We are grateful to Action Research and the British Heart Foundation for funding.

REFERENCES vitro and models the placental defects that occur in preeclampsia. J Clin
Invest 1996;97:540e50.
[1] Kingdom JCP, Kaufmann P. Oxygen and placental villous development: [4] Genbacev O, Zhou Y, Ludlow JW, Fisher SJ. Regulation of human
origins of fetal hypoxia. Placenta 1997;18:623e4. placental development by oxygen tension. Science 1997;277:1669e72.
[2] Zamudio S, Palmer SK, Regensteiner JG, Moore LG. High altitude and [5] Watson AL, Skepper JN, Janiaux E, Burton GJ. Susceptibility of human
hypertension during pregnancy. Am J Human Biol 1995;7:182e93. placenta syncytiotrophoblastic mitochondria to oxygen-mediated damage
[3] Genbacev O, Joslin R, Damsky CH, Polliotti BM, Fisher SJ. Hypoxia in relation to gestational age. J Clin Endocrinol Metab 1998;83:
alters early gestation human cytotrophoblast differentiation/invasion in 1697e705.
Newby et al.: Dissolved Oxygen Concentration in Culture Medium 357

[6] Caniggia I, Homa M, Winter J, Gassmann M, Lye SJ, Kuliszewski M, [9] Kudo Y, Boyd CAR, Sargent IL, Redman CWG. Hypoxia alters expression
et al. Hypoxia-inducible factor-1 mediates the biological effects of and function of syncytin and its receptor during trophoblast cell fusion
oxygen on human trophoblast differentiation through TGFb3. J Clin of human placental BeWo cells: implications for impaired tropho-
Invest 2000;105:577e87. blast syncytialisation in pre-eclampsia. Biochim Biophys Acta 2003;1638:
[7] Huppertz B, Kingdom J, Caniggia I, Desoye G, Black S, Korr H, et al. 63e71.
Hypoxia favours necrotic versus apoptotic shedding of placental [10] Nelson DM, Smith SD, Furesz TC, Sadovsky Y, Ganapathy V,
syncytiotrophoblast into the maternal circulation. Placenta 2003;24: Parvin CA, et al. Hypoxia reduces expression and function of system
181e90. amino acid transporters in cultured term human trophoblasts. Am J
[8] Fitzpatrick TE, Graham CH. Stimulation of plasminogen activator Physiol Cell Physiol 2003;284:C310e5.
inhibitor-1 expression in immortalized human trophoblast cells cultured [11] Kilani RT, MacKova M, Davidge ST, Guilbert LJ. Placenta 2003;24:
under low levels of oxygen. Exp Cell Res 1998;245:155e62. 826e34.

You might also like