Professional Documents
Culture Documents
WTW Oxygen Handbook 2019
WTW Oxygen Handbook 2019
SECTION 1
Fundamentals6
1.1 Dissolved Oxygen Sensors ��������������������������������������������������� 7
SECTION 2
Calibration and analytical quality assurance 11
2.1 Calibration ����������������������������������������������������������������������������� 11
2.1.1 Calibration in watervapor-saturated air �������������������� 11
2.1.2 Calibration in air-saturated water ������������������������������ 13
2.1.3 The calibration process in the measuring meter ����� 13
2.2 Sensor function check ��������������������������������������������������������� 14
2.2.1 Test in water-vaporsaturated air ��������������������������������� 15
2.2.2 Test in air-saturated water ������������������������������������������� 15
2.2.3 Test using zero solution ����������������������������������������������� 15
SECTION 3
Measurement and analytical quality assurance 17
3.1 Cleaning the sensor ������������������������������������������������������������� 17
3.2 Regenerating the sensor ����������������������������������������������������� 17
3.3 Unit and display of the measuring result �������������������������� 19
3.3.1 Display as a concentration value ������������������������������� 19
3.3.2 Display as oxygen saturation percentage ���������������� 19
3.4 Potential error sources ���������������������������������������������������� 20
3.4.1 Polarization times (running-in time) before
measurement ���������������������������������������������������������������� 20
3.4.2 Drift check (AUTOREAD) �������������������������������������������� 20
3.4.3 Importance of flow ������������������������������������������������������ 20
3.4.4 Salt content correction ������������������������������������������������ 22
3.4.5 Impact of interferring gases ��������������������������������������� 22
3.4.6 Solubility function �������������������������������������������������������� 23
SECTION 4
Optical oxygen measurement 24
4.1 The principle of optical measurement ������������������������������ 24
4.2 Characteristics of the optical dissolved oxygen
sensors ����������������������������������������������������������������������������������� 26
4.3 Calibration of optical dissolved oxygen sensors ������������� 27
4.4 Cleaning of optical dissolved oxygen sensors ����������������� 27
4.5 Impact of interferring gases ����������������������������������������������� 28
Bibliography29
Further handbooks 30
About us 31
Information around the clock! ��������������������������������������������������� 31
Oxygen Handbook
SECTION 1 tial pressure in the further discus-
sion. In dry atmospheric air the
Fundamentals oxygen partial pressure is 20.95%
of the barometric pressure. This
Oxygen is not just a component percentage is reduced over a wa-
of the air; it also appears in fluids ter surface because water vapor
as a dissolved gas. An equilibrium also has a vapor pressure and
state is then reached when the thus a partial pressure (Fig. 2).
oxygen partial pressure, that is,
the part of the overall pressure Under saturation conditions it fol-
caused by the oxygen, is the same lows from above:
in both air and fluid (Fig. 1). The
pO2 (T) = 0,2095 · (pair - pW (T))
fluid is then oxygen-saturated. In
relation to physical-chemical cor-
rectness, it should be added that with as the oxygen partial
the partial pressure in the fluid is pressure, pair as the barometric
actually fugacity. The equating of pressure and as the water
both values is, however, permissi- vapor pressure. highlights
ble in the relevant pressure range temperature-dependent values.
for the measurements, whereby In most cases however statements
we can limit ourselves to the par- of oxygen concentration at
O22
N2
Barometric
pressure
O2
O2 O2
CO2
Water vapor
O2 Rare gases pressure
6
are desirable. This is proportion- For water this effect plays a subor-
ally linked to the oxygen partial dinate, negligible role.
pressure and of course to the
The effect of dissolved materials
type of fluid, reflected by the Bun-
is somewhat different. They can
sen coefficient .
both reduce and increase the
solubility of oxygen. A salt content
(sodium chloride) of one percent
in water leads to a saturation con-
with as the molar mass of the
centration of 8.54 mg/L instead of
oxygen and as the mole vol-
9.09 mg/L at 20°C.
ume. For the measurement of the
oxygen concentration the tem- Organic materials, on the other
perature must be known. If the hand, normally increase the ab-
result is desired in % saturation, sorption of oxygen into water. The
the current barometric pressure is maximum saturation concentra-
also required. tion increases with the proportion
of the organic substance. Pure
These comparisons demonstrate
ethanol, for example, dissolves
that water can dissolve more oxy-
40 mg/L of oxygen.
gen at a high rate of barometric
pressure than under low baromet-
ric pressure conditions. 1.1 Dissolved oxygen
sensors
As the temperature rises the
water vapor pressure rises, i.e. the
oxygen partial pressure falls. To
emphasize this impact here is a
comparison between 20°C and
40°C at an barometric pressure of
1013 hPa. While at 20°C 9.09 mg/L
of oxygen dissolves in water, at
40°C the amount is only 6.41 mg/l.
8
The equations of the anode reac- lead or silver coating, fouling is
tions highlight the effect of the a biological coating of the gold
electrolyte solution. The compo- cathode which would take place if
nents of the electrolyte solution the ions were not captured.
bind the metal ions that result
Amperometric sensors can be
from the anode reactions.
operated as a three-electrode
The electrolyte solutions must be cell with an additional silver/silver
suitable for the type of electrode. bromide electrode. They no lon-
The CellOx® 325 or StirrOx® G ger have any anode in the tradi-
galvanic sensors require a ELY/G tional sense. One of the silver/sil-
solution and amperometric ver bromide electrodes takes on
sensors such as the TriOxmatic® the task of the counter electrode
require a ELY/N solution. (current dissipation), and the
other the task of an independent
Ag+ + Br- → AgBr ↓
reference electrode. This is cur-
or rentless and shows a fundamen-
Pb2+ + 2 OH- → 2 Pb(OH)2 ↓ tally better potential constance
Pb(OH)2 → PbO + H2O than a conventional electrode.
Current
be carried out for the dissolved
Water-vapor-
oxygen measurement at certain saturated air
Every linear calibration function The reason for this is that in equi-
is set through at least two points. librium state the oxygen partial
In the case of a dissolved oxygen pressure is the same in fluid and
measurement with WTW® me- in air.
ters, one point is the straight line 2.1.1 Calibration in
of the sensor’s zero point. At the water vapor saturated air
absence of oxygen, the sensor This condition is met over a large
surface of water, such as a lake or
11
Oxygen Handbook
even the aeration tank of a waste- The water vapor partial pres-
water treatment plant. sure is merely a function of
temperature if air humidity is at
100%. In order to determine this
value, the sensor is additionally
equipped with a temperature
measurement probe.
Water-vapor
In line with this relationship the saturated air
12
It is sufficient if the sponge in the The option of calibration in water
OxiCal® is moist (Fig. 7). It should, vapor saturated air is supported
however, never be wet. The user by every WTW® meter and is,
should only moisten the sponge as indicated by the points given
with distilled water and squeeze above, certainly to be preferred in
it. The dampness is fully sufficient. comparison to a calibration in air-
2.1.2 Calibration in air- saturated water. It has also to be
saturated water applied when working according
to the Standard (ISO 5814).
The water is vented until the
oxygen partial pressure in water The relative, not the absolute
and air is the same. However, this slope (as is the case for pH mea-
method conceals some risks: suring meters) is displayed in %
for the calibration result.
• The barometric pressure in
the ventilation hose is al- 2.1.3 The calibration
ways a little greater than the process in the measuring
normal barometric pressure meter
and thus the water is always a The oxygen meter takes the elec-
little over-saturated following trode signal (current!) and com-
venting. pares it with the oxygen partial
pressure, which is obtained using
• The temperature in the water equation (1) for the current air
decreases as a result of the conditions of barometric pres-
venting (cooling by evapora- sure pair and temperature . The
tion). meter thus obtains a slope that
• If a temperature equalization allows measured currents to be
is waited for, the water will be converted into the oxygen partial
a little over-saturated. value . At the same time a
comparison is made between the
• The point of complete satura- actual measured value and the
tion is hard to estimate. There average saturation current of the
is the risk of under-saturation. electrode type following the refill
• Oxygen-consuming substanc- (nominal current). The result rep-
es lead to under-saturation resents the relative slope S.
13
Oxygen Handbook
With reference to the average tion, there is the visual check and
saturation current, slopes greater three characteristic measurement
than 1.00 are also possible. A points.
slope of 0.81 following a calibra-
The gold cathode is examined
tion means that the slope is 81%
for the visual check. If it is no
of the nominal value. It does not
longer gold in color, but instead
indicate anything about the preci-
covered with lead or silver, the
sion of the measurement, but
sensor shows high values and
rather is a reference for estimating
is generally no longer free of a
the remaining operating time, i.e.
zero-current. A remedy to this
the time until the next change of
can be found through regenerat-
the electrolyte solution is recom-
ing the dissolved oxygen sen-
mended.
sor according to the operating
This also applies to the mean- instructions. The gold cathode
ing of the sensor symbols in the should be polished only using a
display of the modern oxygen special, moistened polishing strip
meters which evaluate the service with circular movements and little
life of the sensor in the same way. pressure. Only this strip has to be
used in any circumstances as a
The calculation of the oxygen
scratched and unpolished elec-
concentration is correctly carried
trode surface damages the sensor
out with the internally stored
and negatively affects the measur-
absolute slope in nA/hPa.
ing reliability.
2.2 Sensor function Caution: The anodes, regardless
check of whether they are made from
For pH measurements, calibration lead or silver, must not be pol-
data offer a direct possibility to ished in any case.
assess the quality of the measur- More comprehensive than the
ing set-up of meter, electrode subjective, visual check is an
and standard buffer solution. evaluation at the three special
This is not possible for oxygen measurement points:
measurement due to the relative
slope. Nevertheless, in order to
gain results from the sensor func-
14
1st in water vapor saturated air, This is also the initial reason for
2nd in air saturated water and which we succesfully looked for
3rd in oxygen free water. an alternative to the calibration
in air-saturated water which was
2.2.1 Test in water vapor
saturated air common beforehand. If no value
is found in this tolerance interval,
In water vapor saturated air
the sensor should be sent to the
should provide a value between
plant to be checked.
100 and 104% oxygen satura-
tion. If the values are below this 2.2.3 Test using zero
benchmark, it is likely that the solution
membrane has become wet dur- This test proves the absence
ing calibration, and there may be of zero current in the sensors.
too much water in the calibration At a oxygen content of 0 mg/L
vessel. The target value of above a sensor should only display a
100% saturation results from the maximum of the resolution of the
different viscosities of water and meter (1 digit). The check is car-
air as well as the tension of the ried out with a sodium sulfite solu-
water surface. Expressed in a sim- tion. Sulfite reacts with dissolved
plified manner, it is easier for the oxygen to sulphate, whereby
oxygen molecules to permeate dissolved oxygen is removed
the membrane in air than in water. from the water. The production is
The normal measurement mode simple, as might be expected. A
assumes a liquid sample, so on air teaspoon of sodium sulfite is dis-
the reading exceeds 100%. solved in 100 ml tap water. After
2.2.2 Test in air saturated 15 minutes of leaving the solution
water undisturbed it is free of oxygen. It
must be kept undisturbed as oth-
After calibration the value in
erwise oxygen from the surround-
air-saturated water should show
ings will once more be stirred in.
between 97 and 102% satura-
tion. The theoretical value is at One minute after the immersion
100% and is difficult to repeat. of the sensors the measuring
The cause of the relatively wide meter should display a value of
tolerance is not the sensor, but maximum 2% and of a maximum
rather the saturation procedure. 0.4% after 15 minutes. If not, the
15
Oxygen Handbook
sensor is no longer free of zero
current and must be cleaned
again or sent to the service. After
the test, thoroughly rinse the sen-
sor with distilled water, in order
to remove the rest of the sodium
sulfite solution.
16
SECTION 3
Measurement and analytical quality assurance
19
Oxygen Handbook
For many meters the oxygen measured value is considered to
partial pressure can also be dis- be the actual measured value.
played in hPa.
∆β/∆t
21
Oxygen Handbook
In the case of the installation in tion tank is non-critical due to the
pipes the sample bypasses the high sample volume. In an open
sensor head and provides suffi- beaker glass, on the other hand,
cient flow. WTW provides station- the oxygen concentration can be
ary measurement systems with easily changed through stirring.
specific mounting fixtures for
In the case of optical sensors,
pipes.
which, as a matter of principle, do
Alternatively the sensor can be not require any flow to function,
moved in the measured medium. the stirring does still help to reach
For example, stirring the sensor in the end value faster.
a beaker glass or by slewing it in
3.4.4 Salt content correction
a lake. For measurements at great
depths, armatures for measure- The temperature-dependent
ments up to 100 m are available. Bunsen coefficient (see Equation
2) changes if substances are dis-
It must be ensured that the stir- solved in the water. This effect is
ring causes no falsification of the taken into account by the input of
measured values. This may par- salinity. The salinity can be deter-
ticularly be the case if the sample mined with a conductivity meter
examined is over- or under-sat- and, to some extent, the salt
urated with oxygen and oxygen content of seawater in g/kg. This
can be expelled or stirred in. Oxy- function can be also be used for
gen over-saturation is, for exam- other waters, as the deviations are
ple, to be found in still waters in often small.
summer if the photosynthesis of
3.4.5 Impact of interferring
rampant algae produces oxygen.
gases
An example of oxygen under-sat-
The membrane of the dissolved
uration is the BOD determination
oxygen sensor can also be per-
where bacteria lower the oxygen
meated by other gases alongside
concentration by respiration in the
oxygen (Fig. 15).
Karlsruhe Bottles.
Phase
Fig. 16 Luminescence excitation
using light of a short wavelength
Fig. 17 Simplified representation
(green here), emission of a red
of modulated stimulus and emit-
and so low-energy fluorescent
ted fluorescence beam
beam.
Green graph: light periodi-
What does this mean for the dis-
cally beamed in, red graph: light
solved oxygen measurement?
emitted from the dye at a shifted
There are dyes which are excited
phase resulting from the changed
by visible light and react on oxy-
decay ratio.
gen with a high selectivity. The
oxygen molecules serve as the
quenchers described above. They
extinguish or change the light
resulting from the dye depending
on the oxygen partial pressure. By
25
Oxygen Handbook
4.2 Characteristics of optical dissolved oxygen sensor,
there is diffusion of the dissolved
the optical dissolved
oxygen in the water depending
oxygen sensors on the concentration into or out
The significant difference from of the membrane. This process
the classical dissolved oxygen naturally runs faster the better
sensors is that optical sensors the exchange takes place at the
consume no oxygen. There are interface. A stirring supports this
no electrochemical reactions. In exchange.
addition, there is a further charac-
There is no polarization time
teristic: The optical sensor does
for optical sensors as it has no
not require any flow to replace
electrodes which have to reach an
“consumed” oxygen. Stirring sys-
operating condition.
tems are then superfluous. In ad-
dition, there is no necessity for the Optical dissolved oxygen sensors
electrolyte change or cleaning the are sensors which have active
electrode system. The life span of components (LEDs, detector, pro-
the optical sensor caps with the cessor). In contrast to the electro-
dye-containing membrane usually chemical sensors, they are more
is at least one year. power consuming which leads
to a markedly reduced operating
Electrochemical sensors form in
time in the case of portable sys-
water a 10 µm thick layer of water
tems. Additionally optical sensors
molecules that can not be stirred.
need different meters.
As consequence this layer serves
as an additional diffusion barrier. The measurement range is notice-
On air these sensors show a satu- ably restricted in comparison
ration signal of about 102 %. The to the electrochemical sensors
optical sensor measures 100 % air above. The reason for this is the
saturation as this barrier does not quenching. Simplified, no oxygen
exist here. leads to a strong signal, a large
amount of oxygen leads to a weak
However it may be reasonable
signal. In general the measure-
to apply flow to an optical sen-
ment range does not exceed ap-
sor. As with all systems with
interfaces, even when using the
26
prox. 200% air-saturation respec- Of course a calibration carried out
tively 20 mg/l. by the user is also possible. It is
implemented analogously to the
4.3 Calibration of optical electrochemical sensors in water-
dissolved oxygen vapor-saturated air in the calibra-
sensors tion vessel. As already mentioned
above, the resulting saturation
An advantage of the optical
display is at 100%.
dissolved oxygen sensors is
that very stable drift behaviour. 4.4 Cleaning of optical
The changes in the membrane
dissolved oxygen
are mainly caused through the
irradiated light. A decrease in sensors
sensitivity results, triggered by the The sensitive surface of an opti-
so-called bleaching to a decrease cal oxygen sensor is protected
in intensity and a change of the by a light proof top layer. This
phase shift. top layer is very thin and must
therefore not be scratched or
These changes are generally
damaged as otherwise this could
much smaller than 5% of the initial
lead to malfunctions. In relation
value over the course of a year. If
to the cleaning the same mea-
this tolerance can be accepted,
sures apply as to the conventional
the sensor does not need to be
oxygen membranes. In the case
calibrated.
of easily adhesive contamination
We apply the dye to changing careful cleaning with a soft cloth is
caps that the user simply puts recommended, further notes can
onto the sensor. These exchange be found in the operating instruc-
caps contain a chip with the tions.
specifications of the individual
membrane determined in the pro-
duction process (“factory calibra-
tion”). These data are automati-
cally transferred to the sensor and
used for the correct calculation of
the oxygen concentration.
27
Oxygen Handbook
4.5 Impact of hydrogen sulfide, which results
from the putrefaction processes
interferring gases
in many waters, that here dissem-
In contrast to the amperomet- bles an oxygen signal. This signal
ric sensors, carbon dioxide and is reversible, the sensor does not
hydrogen sulfide play no role undergo any damage.
in relation to the operation life
of or damage to the sensors. It
must solely be noted that, for
example, hydrogen sulfide also
functions as a quencher due to its
molecular mass. An example from
practice would be the decreasing
oxygen content in limnological
depth measurements at increas-
ing depths, which seem to un-
dergo an increase if the sensor
is immersed into soft sediment.
However the reason for this is the
28
Bibliography
29
Oxygen Handbook
Further handbooks
Meters
Lab undand ProLab
Electrodes
Titration
Titrations
pH,
Fibel Serien
Handbook
pH,ISE,
ISE,LF
LFUND
UNDOO22MESSEN
MESSEN——GENAU,
GENAU,ZUVERLÄSSIG,
ZUVERLÄSSIG,SELEKTIV
SELEKTIV
Titrations Fibel
pH, ISE, COND AND O2 | PRECISE - RELIABLE - SELECTIVE
IN LABAND
AND FIELD OFTITRATION
IN LAB AND
THEORIE
THEORY FIELD DER
UNDPRACTICE
PRAXIS TITRATION
30
About us
New products
Click here. Xylem Analytics pres-
ents you with new products, de-
velopments, innovative measure-
ment and analysis meters, helpful
accessories, useful system exten-
sions and much more.
The place to go 24 hours a day
Applications
At Xylem you will find the solution
to your measurement tasks in re-
search, analysis and quality con-
trol – and a great deal of usage tips
along with it.
Everything custom-made.
Contact addresses
Looking for a contact partner in
your area? Here you can find our
national and international contact
partners, their addresses and
representatives etc.
Simply click and browse through.
www.WTW.com
31
Xylem |ˈzīləm|
1) The tissue in plants that supplies the water from the roots up through the plant;
2) a leading global water technology enterprise.
We are a global team with the shared goal of creating innovative solutions to provide for our
world’s water needs. The focus of our work is the development of new technologies that
improve the type and manner of water use and re-use in the future. We move, treat and
analyze water and lead it back into the environment, and help people to efficiently use water
in their homes, buildings, factories and agricultural operations. Through the takeover of
Sensus in October 2016, Xylem expanded its portfolio with intelligent measuring meters,
network technologies, and advanced services for data analysis in the water, gas and electric-
ity industry. We have strong long-term relationships with customers in more than 150 coun-
tries, among whom we are famed for our high-performance combination of leading product
brands and expertise, supported by a tradition of innovation.
Additional information on how Xylem can help you can be found at www.xylem.com