Measurement of Pectoralis Minor Muscle Length in Women Diagnosed With Breast Cancer Eliability, Validity, and Clinical Application

You might also like

Download as pdf or txt
Download as pdf or txt
You are on page 1of 9

Original Research

Measurement of Pectoralis Minor S.E. Harrington, PT, PhD, Department


of Exercise Science, Physical Therapy
Program, University of South Carolina,
Muscle Length in Women Diagnosed Blatt PE Center, 101G, Columbia, SC

Downloaded from https://academic.oup.com/ptj/advance-article-abstract/doi/10.1093/ptj/pzz174/5733064 by University of Liverpool user on 16 February 2020


29208 (USA). Address all

With Breast Cancer: Reliability, correspondence to Dr Harrington at:


sharring@mailbox.sc.edu.
J. Hoffman, PT, DPT, Department of
Validity, and Clinical Application Physical Therapy, Creighton University,
Omaha, Nebraska.
Shana E. Harrington, Julie Hoffman, Dimitrios Katsavelis D. Katsavelis, PhD, Department of
Exercise Science, Creighton University.
[Harrington SE, Hoffman J, Katsavelis
Background. Decreased pectoralis minor muscle length is common after primary breast D. Measurement of pectoralis minor
cancer treatment and can result in an abnormal position of the scapula. This position muscle length in women diagnosed
can contribute to shoulder pain and pathomechanics and can lead to problems such as with breast cancer: reliability, validity,
impingement syndrome, rotator cuff tears, and frozen shoulder. Currently, there are limited and clinical application. Phys Ther.
reliable methods for measuring pectoralis minor length. 2020;100:1–9.]
© 2020 American Physical Therapy
Objective. The objective of this study was to examine the reliability and validity of Association
measuring pectoralis minor length in women diagnosed with breast cancer.
Accepted: September 23, 2019
Submitted: January 3, 2019
Design. This was a cross-sectional reliability and validity study.
Methods. Bilateral pectoralis minor length (in centimeters) was assessed using a
palpation meter in women (N = 29) diagnosed with breast cancer by 2 licensed physical
therapists who were masked to the measures. Bilateral pectoralis minor length was also
measured using a motion capture system to assess validity.

Results. Intratester reliability (intraclass correlation coefficient, ICC [3,k] = 0.971; 95%
confidence interval [CI] = 0.939–0.986; standard error of measurement [SEM] = 0.16 cm)
and intertester reliability (ICC[3,k] = 0.915; 95% CI = 0.81–0.962; SEM = 0.31 cm) were
excellent for the palpation meter on the affected side and the unaffected side (intratester
reliability: ICC[3,k] = 0.951; 95% CI = 0.897–0.977; SEM = 0.19 cm; intertester reliability:
ICC[3,k] = 0.945; 95% CI = 0.877–0.975; SEM = 0.22 cm). Significant correlations were
found between the motion capture system and the palpation meter on the affected side
(r = 0.87) and the unaffected side (r = 0.81). Bland-Altman plots between the palpation
meter and the motion capture system demonstrated that all the measures fell within the
limits of agreement.

Limitations. This study encountered possible errors with the accuracy of the motion
capture system tracking because of the proximity of the markers and inherent volumetric
restrictions.

Conclusions. The palpation meter is a reliable, valid, easily administered, and cost-
effective tool for assessing pectoralis minor length in women with breast cancer.

Post a comment for this


article at:
https://academic.oup.com/ptj

2020 Volume 100 Number 3 Physical Therapy 1


Pectoralis Minor Length in Breast Cancer

I
n 2019, it was estimated that over 268,000 women in The importance of pectoralis minor length and its
the United States would be diagnosed with contribution to upper extremity function are well
breast cancer.1 Currently, 3.1 million women have been documented in the scientific literature. However, there are

Downloaded from https://academic.oup.com/ptj/advance-article-abstract/doi/10.1093/ptj/pzz174/5733064 by University of Liverpool user on 16 February 2020


diagnosed with and treated for breast cancer in the United only a few reported methods in the literature describing
States.1 Because of advancements in detection and how to clinically assess pectoralis minor length. These
treatment, approximately 89% of these women will survive methods have several limitations, including limited
for 5 years or longer.2 Traditionally, management of breast psychometric properties (eg, reliability, validity, and
cancer often primarily focuses on tumor response to responsiveness), and may not be feasible on women who
treatment and survival rate.3 However, with improved have been treated for breast cancer.25 A common clinical
survivorship, the focus on function and quality of life method to examine pectoralis minor muscle length is to
related to breast cancer treatment has now gained use a flexible tape measure from the coracoid process to
significantly more attention.4 the fourth intercostal space.26 However, this method could
prove difficult to apply to women who have undergone
The majority of women diagnosed with breast cancer will surgery for breast cancer because of breast tissue
undergo some form of lymph node dissection (sentinel deformities and irregularities that often result from
lymph node dissection or axillary lymph node dissection) common surgical techniques used to remove the cancer.
and most will receive either breast-conserving or The most common clinical method to assess pectoralis
mastectomy surgery.5 It is well documented that upper minor length is to measure the distance from the
extremity impairments may develop within weeks or posterolateral angle of the scapula to the examination
months after these surgeries for breast cancer and can be table of a patient in supine.25 However, measurements
difficult to treat.6,7 These impairments include decreased obtained using this method have been shown to be poorly
shoulder active and passive range of motion, decreased correlated with a normalized measure of pectoralis minor
upper extremity strength, and pain; these have been length27 and have been found to have poor diagnostic
shown to negatively affect upper extremity function and accuracy.28 Measuring resting scapula position, as
quality of life.6,8–10 Because of its proximity to the surgical described with the Lennie Test, is an additional technique
and radiation fields, the pectoralis minor muscle has been often used to indirectly measure the length of the
linked to many of these impairments.11 The pectoralis pectoralis minor; however, poor to moderate reliability
minor muscle originates at the medial coracoid process of and validity have been reported with this method.29–31
the scapula and inserts on ribs 2 through 5. It is situated Using MRI scans to examine cross-sectional size of the
in a key location to be adversely affected by a variety of pectoralis minor has also been reported in the literature
breast cancer treatments, including surgery12 and radiation although this is another form of an indirect technique that
therapy.13,14 A variety of factors could influence pectoralis has a variety of limitations. A cross-sectional study
minor muscle length, although literature to explain this is conducted by Shamley et al revealed that the pectoralis
sparse in women diagnosed with breast cancer. The minor was significantly smaller on the affected side
pectoralis minor muscle may be compromised when it is compared with the unaffected side in 74 women treated
used to fill a lumpectomy cavity as described by Manaswi for unilateral breast cancer.11 Unfortunately, the
and Mehrotra.15 Additionally, in women who received methodology used to discover this decrease in pectoralis
radiotherapy for breast cancer, a significantly smaller minor length included the use of MRI scans, which are
(bilateral cross-sectional area using magnetic resonance extremely costly and time-consuming, are not feasible in
imaging [MRI]) pectoralis minor muscle on the affected daily clinical practice, and cross-sectional size does not
side was found compared with the unaffected side.11 directly examine muscle length.11 Because of these
Decreased pectoralis minor length is common after limitations, what is needed is a simple and cost-effective
primary breast cancer treatment and can result in an tool that can be used in clinical and research settings that
abnormal position of the scapula (protracted and can objectively measure pectoralis minor length in women
depressed).12 Alterations in scapular position are not only treated for breast cancer.
undesirable but are also believed to contribute to shoulder
pain and pathomechanics, leading to problems such as The PALpation Meter (PALM; Performance Attainment
impingement syndrome, rotator cuff tears, frozen Associates, St Paul, MN) (Fig. 1) combines a caliper and an
shoulder, and glenohumeral instability.16–23 The cause of inclinometer into 1 tool that measures distance. In a
pectoral minor tightness in women diagnosed with breast recent study by Rondeau et al, the PALM was used to
cancer is multifactorial.12 Pain resulting from breast cancer measure pectoralis minor length in 29 overhead athletes.
surgery may cause the pectoralis minor muscle to exhibit Intrarater reliability was found to be reliable and precise
increased tone,12 whereas radiation may cause fibrosis of for both the dominant (intraclass correlation coefficient
the pectoralis minor muscle sheaths and tendons, (ICC)[2,k] = 0.980; standard error of measurement
producing further tightness.24 Additionally, women may [SEM] = 0.32 cm) and nondominant (ICC[2,k] = 0.990;
try to protect their surgical sites through increased flexion SEM = 0.29 cm) arms.32 Additionally, the PALM has been
of the thoracic spine and protraction of the scapula, which shown to be a valid and reliable instrument for assessing
may intensify pectoralis minor shortening.12 scapular orientation and postural alignment, including

2 Physical Therapy Volume 100 Number 3 2020


Pectoralis Minor Length in Breast Cancer

Procedures
Prior to the start of the study, 2 licensed physical
therapists with a combined 34 years of clinical experience

Downloaded from https://academic.oup.com/ptj/advance-article-abstract/doi/10.1093/ptj/pzz174/5733064 by University of Liverpool user on 16 February 2020


underwent a training session lasting approximately 1 hour
to familiarize themselves with the testing procedures. This
session instructed the examiners in performing an
accurate measurement of the pectoralis minor length
using the PALM, including bony landmark palpation and
pilot testing on 6 people who were healthy. Participants
reported to the Rehabilitation Science Research
Laboratory at Creighton University and data collection
lasted approximately 45 minutes. Participants wore either
a tank top or sports bra so their shoulders could be
appropriately exposed for preparation of measurements to
be taken. After signing the institutionally approved
informed consent form, 3 trials of bilateral pectoralis
Figure 1. minor length were assessed using the PALM by a licensed
PALpation Meter (PALM; Performance Attainment Associates, St physical therapist with 16 years of clinical experience.
Paul, MN) placed on the participant to measure pectoralis minor Pectoralis minor length was obtained by placing the
length. caliper tips of the PALM on the medial coracoid process
and the fourth intercostal space (central location) adjacent
to the sternum on the participants’ affected and unaffected
arms (Fig. 1).25,27,32 The “central vector” as described
pelvic height and leg length discrepancies.33–35 Therefore, elsewhere25 was measured in centimeters and was defined
it is well suited for use as an objective measurement of as the pectoralis minor length for this study. First, the
pectoralis minor length in women receiving treatment for researcher demonstrated a standing resting position to the
breast cancer because it is both cost effective (costing participants. Participants were then asked to stand still in
approximately $225) and user-friendly.36 We are unaware their preferred resting position, with arms relaxed, while 3
of any literature establishing the reliability and validity of consecutive measurements, lasting approximately 5
the PALM for measuring pectoralis minor length in women seconds each, were performed with the PALM for each
diagnosed with breast cancer. Determining the reliability shoulder. Participants were given a 30-second break
and validity of the PALM would enable clinicians and between each measurement during which they were
researchers the capacity to objectively assess pectoralis asked to rest, and the PALM was reset to 0. The physical
minor length in these women, which could help guide the therapist using the PALM was masked regarding the actual
rehabilitation process. Therefore, the purpose of this study pectoralis minor length measures while another trained
is to examine the reliability and validity of the PALM when clinician stood adjacent to the tester, lifting the paper that
measuring pectoralis minor length in women diagnosed covered the measure and recording each pectoralis minor
with breast cancer. length. This process was then replicated by a second
licensed physical therapist who had 18 years of clinical
experience. Because we are unaware of a gold standard
Methods reference test for the measurement of pectoralis minor
Participants length28 we chose to use a 6-camera optoelectronic
This cross-sectional study enrolled 29 women who had a motion capture system (Oqus 5+; Qualisys AB,
breast cancer diagnosis from 2015 to 2016. This sample Gothenburg, Sweden) (hereafter referred to as Qualisys)
size was based on previous literature to assess the as our reference standard.
reliability and validity of the PALM.32–35 Eligible
participants included adult women who had been treated After the PALM measures were completed, therapist 1
for breast cancer in the Omaha, Nebraska, and Council placed 2 lightweight reflective markers (12 mm) at the
Bluffs, Iowa, regions of the United States. The eligibility origin and insertion of each pectoralis minor muscle,
criteria included women who had a diagnosis of stage 0 to which corresponds to the same areas as where the PALM
III breast cancer within the past 3 years, who were 19 to calipers were placed as described above.
85 years old, who had received treatment for unilateral Three-dimensional coordinates were acquired at 120 Hz
breast cancer, who had completed primary breast cancer using the Qualisys (Fig. 2). This specific system has
treatment (surgery, chemotherapy, and/or radiation), and precision in detecting the marker with a mean error of
who had no current recurrence of cancer. Prior to <0.15 mm. Participants were once again instructed to
participation, all participants signed an informed consent stand still in their preferred resting position, with arms
form approved by the Creighton University Institutional relaxed within the field of view of the motion capture
Review Board. system. Three 5-second trials were recorded and stored on

2020 Volume 100 Number 3 Physical Therapy 3


Pectoralis Minor Length in Breast Cancer

determine the criterion validity of the PALM pectoralis


minor length with an a priori alpha level of .05.
Bland-Altman analysis with graphic plots followed by

Downloaded from https://academic.oup.com/ptj/advance-article-abstract/doi/10.1093/ptj/pzz174/5733064 by University of Liverpool user on 16 February 2020


linear regression was used to study agreement and
proportional bias between the PALM and Qualisys
measures.38 ,39 Several steps occurred to conduct the
analysis using Bland-Altman. First, the difference between
each value measured by the PALM and Qualisys for each
participant was calculated with a 1-sample t test. Next, an
average of the values measured by the PALM and Qualisys
was calculated for every participant on the affected and
unaffected arm. Both the difference and the average for all
participants were then placed in a scatterplot, with the
average values obtained by the PALM and Qualisys on the
Figure 2. y-axis and the between-method difference on the x-axis.
Qualisys Tracking Manager (Qualisys AB, Gothenburg, Sweden) The mean difference between the PALM and the Qualisys
marker placement. and the 95% CIs were marked. The mean difference ±1.96
times the SD of the mean difference was added to the plot
to derive the limits of agreement. These limits defined the
a PC for further analysis. The Qualisys Tracking Manager interval in which differences between the PALM and the
(Qualisys AB, Gothenburg, Sweden) software calculated Qualisys could be expected for 95% of future
the average distance during the entire 5-second interval. measurements in comparable people.38 The figure plots
Five-second intervals are commonly used with motion the amount of agreement or disagreement between the
capture to avoid inconsistency that could be caused by PALM and the Qualisys and illustrates how the agreement
slight motion.37 Additionally, during the training sessions or disagreement is related to the magnitude of the
of the examiners, approximately 5 seconds were required measured values.40 A mean difference close to y = 0 is a
for the third examiner to review (because the output was good indicator of agreement. A linear regression was
hidden) and record the measurement from the PALM. calculated using the difference calculated between the
Participant demographics—age, side involved, stage of PALM and Qualisys to examine whether there is
cancer, type of surgery, date of surgery, whether a lymph proportional bias. If P >.05, there is no proportional bias
node dissection was performed, type of lymph node in the measures.41 The minimal detectable change (MDC)
based on a 95% CI (MDC95 ) was calculated using the
dissection, number of positive lymph nodes, adjuvant √
following equation: MDC95 = SEM × 1.96 × 2.42 The
treatment (chemotherapy, radiation therapy, hormone √
therapy), hand dominance, height, and weight—were SEM was calculated as follows: SEM = s 1 − ICC, where s
is the pooled SD of the measurements. Statistical analyses
collected after the pectoralis minor measures were
were performed with IBM SPSS Statistics Version 25.0
completed to minimize bias of the examiners knowing the
(IBM SPSS, Chicago, IL, USA).
participants’ breast cancer history and treatment.
Role of the Funding Source
Data Reduction This work was supported by the Dr George F. Haddix
The “central vector” of the pectoralis minor, which is
President’s Faculty Research Fund of Creighton University
defined as the 3-dimensional distance from the origin to
(to S.E.H.). The funder played no role in the design,
the insertion points of the muscle, was recorded with the
conduct, or reporting of this study.
Qualisys Tracking Manager software. The same software
was used to process and calculate the distance for each
5-second interval. The 3 trials were averaged for each Results
participant and each side separately so they could be used A total of 29 women who were 61.6 (±10.6) years old and
in the ICC analysis for the affected and nonaffected sides. had a diagnosis of unilateral breast cancer (14.8 [±6.63]
months since surgery) participated in the intrarater
Data Analysis reliability and validity portions of this study. Because of
Intratester and intertester reliability and precision of the schedule conflicts for therapist 2, 26 of the 29 women
PALM measuring pectoralis minor length were determined participated in the interrater reliability assessment; those
by calculating the ICC(3,k) and SEM values. Bivariate 26 women were 61.5 (±10.6) years old and had a
Pearson product moment correlation analysis was used to diagnosis of breast cancer (14.8 [±6.50] months since
examine the relationship between the PALM (clinical surgery). Participant demographics are shown in Table 1.
assessment) and the motion capture system (laboratory
and reference standard) for both the affected and Intratester reliability was excellent43 for the affected arm
unaffected arms. The correlation coefficients were used to (ICC[3,k] = 0.971; 95% CI = 0.939–0.986; SEM = 0.16 cm)

4 Physical Therapy Volume 100 Number 3 2020


Pectoralis Minor Length in Breast Cancer

Table 1. Figure 4 shows the Bland-Altman plot for the pectoralis


Participant Demographic Dataa minor length data. The mean difference between the
measures was −0.61 cm. The limits of agreement for the
PALM and Qualisys were 0.48 to −1.70 cm. The plots show

Downloaded from https://academic.oup.com/ptj/advance-article-abstract/doi/10.1093/ptj/pzz174/5733064 by University of Liverpool user on 16 February 2020


Characteristic Value
Age, y a random scatter of points above and below the mean
difference line, showing agreement. Linear regression
Mean (SD) 61.6 (10.6)
(P > .05) revealed that there was no proportional bias.
Range 41–84 The calculated MDC95 for the PALM was 0.80 cm.
Height, cm
Mean (SD) 162.2 (7.45)
Discussion
Range 150–174
The purpose of this study was to examine the reliability
Weight, kg and validity of the PALM when measuring pectoralis
Mean (SD) 76.1 (15.5) minor length in women diagnosed with breast cancer. The
Range 53–124
results indicate that the PALM was able to accurately
measure pectoralis minor length with marginal error
BMI, kg/m2 , mean (SD) 29.1 (6.12)
when the arm is in a resting position in women diagnosed
Average time since surgery (SD), mo 14.8 (6.63) with breast cancer. The ability to objectively measure
Type of surgery pectoralis minor length using the PALM in women
diagnosed with breast cancer may allow rehabilitation
Mastectomy 7 (24)
professionals to determine restriction and/or changes in
Breast conserving 22 (76) the muscle and gain valuable information to help prevent
Stage of breast cancer and/or reduce shoulder impairments.
0 1 (3)
A variety of methods for measuring pectoralis minor
I 22 (76)
muscle length are commonly reported in the literature,
II 4 (14) but these are indirect measures of the muscle that either
III 2 (7) lack reported interrater reliability and validity44 or use
Type of lymph node dissection
expensive equipment to obtain the measure.11 As reported
by Host, pectoralis minor length was assessed in
SLND 28 (97)
participants with shoulder impingement by measuring the
None 1 (3) distance between the posterior acromion and a table with
Chemotherapy 3 (10) the person lying supine,44 a procedure described by
Kendall45 and Sahrmann.46 This methodology was also
Radiation 16 (55)
used in another study with 45 participants who had
Chemotherapy and radiation 6 (21) shoulder symptoms and 45 who were “healthy.”28
Hand dominance Although this study reported excellent intrarater reliability
Right 26 (90) for dominant and nondominant sides of the participants
without symptoms and for the pain-free and painful sides
Left 3 (10)
of participants with symptoms, interrater reliability and
Breast cancer on dominant side 15 (52) validity of this indirect measure was not reported.28 We
Reported lymphedema 3 (10) are aware of only 1 other study that measures pectoralis
Reconstruction after mastectomy 3 (43)
minor in women with breast cancer.11 Shamley et al used
MRI to determine muscle cross-sectional area of the
a
Data are reported as numbers (percentages) of participants unless pectoralis minor muscle in 74 women with unilateral
otherwise indicated. BMI = body mass index; SLND = sentinel lymph node
dissection. carcinoma of the breast.11 Although the interrater
reliability for the MRI measures was excellent (r = 0.89)
and MRI was able to detect a significant decrease in the
size of the pectoralis minor muscle on the affected side
and the unaffected arm (ICC[3,k] = 0.951; 95% CI = 0.897– (P = .026),11 this method is another indirect measure of
0.977; SEM = 0.19 cm). Intertester reliability was the muscle and the use of MRI is very expensive and not
also excellent43 for the affected arm (ICC[3,k] = 0.915; 95% clinically feasible for physical therapists.
CI = 0.81–0.962; SEM = 0.31 cm) and the unaffected arm
(ICC[3,k] = 0.945; 95% CI = 0.877–0.975; SEM = 0.22 cm). There have been a handful of published studies
There were significant correlations between pectoralis measuring pectoralis minor length using the PALM with
minor length measures between the PALM and the motion the same methodology as this current study, although
capture system for the affected (r = 0.87; P < .001) participants in these studies have all been overhead
and unaffected (r = 0.81; P < .001) arms (Fig. 3). athletes.32,47 To the best of our knowledge, a study by

2020 Volume 100 Number 3 Physical Therapy 5


Pectoralis Minor Length in Breast Cancer

Downloaded from https://academic.oup.com/ptj/advance-article-abstract/doi/10.1093/ptj/pzz174/5733064 by University of Liverpool user on 16 February 2020


Figure 3.
Values of pectoralis minor length measures using the PALpation Meter (PALM; Performance Attainment Associates, St Paul, MN, USA) and
Oqus 5+ (Qualisys AB, Gothenburg, Sweden) on the affected and unaffected arms.

Figure 4.
Bland-Altman plot for pectoralis minor muscle length (PML) data.

Rondeau et al is the only one that examines the reliability (ICC[2,k] = 0.990; SEM = 0.29 cm) arms.32 The excellent
and validity of the PALM.32 They measured pectoralis intrarater reliability reported by Rondeau et al is similar to
minor length in 29 overhead athletes and found significant what was found in this study. However, the present study
correlations between measures using the PALM and an reported higher correlations when examining validity
electromagnetic motion analysis system on the dominant between the PALM and the motion capture system
(r = 0.695; P ≤ .005) and nondominant (r = 0.837; (Qualisys). It is important to note that the participants in
P ≤ .005) arms.32 Although interrater reliability was not the study conducted by Rondeau et al used both males
assessed, intrarater reliability found the PALM to be and females, whereas we had only female participants
reliable and precise for both the dominant because emerging evidence suggests length differences
(ICC[2,k] = 0.980; SEM = 0.32 cm) and nondominant among genders.27 ,32 Additionally, the participants in the

6 Physical Therapy Volume 100 Number 3 2020


Pectoralis Minor Length in Breast Cancer

study by Rondeau et al were significantly younger system because of the proximity of the markers and
compared with those in the current study as inclusion inherent volumetric restrictions.
criteria reported participants who were healthy,
Conclusions

Downloaded from https://academic.oup.com/ptj/advance-article-abstract/doi/10.1093/ptj/pzz174/5733064 by University of Liverpool user on 16 February 2020


college-aged overhead athletes.32 Finally, in the study by
Rondeau et al, 3-dimensional coordinates were recorded The PALM is an easy-to-administer and cost-effective tool
by an electromagnetic tracking system that uses sensors to that can be used to measure pectoralis minor length in
detect position in space,32 whereas a camera system was women diagnosed with breast cancer. Because of the
used in the current study that is considered more accurate variety of breast cancer treatments that can affect the
because of its updated technology and resolution.25,27 pectoralis minor, measuring pectoralis minor length may
be important for rehabilitation professionals. Because the
The level of agreement between the PALM and the PALM was determined to be reliable, valid, and precise,
Qualisys was limited according to the Bland-Altman plot. we recommend that this tool be used to measure
However, as Giavarina reported, “the B&A [Bland-Altman] pectoralis minor length in women diagnosed with breast
plot method only defines intervals of agreements, it does cancer. Finally, future research could use the PALM to
not say whether those limits are acceptable or not.”41 This examine whether deficits in pectoralis minor length in
is significant, especially when examining clinical women who have undergone treatment for breast cancer
measures. To understand our results in clinical terms, we affects function and quality of life as well as whether
examined both the reported MDC for pectoralis minor differences occur because of cancer treatments such as
length in the literature and calculated our sample-specific radiation and surgery type.
MDC95 . Rosa et al calculated the MDC for between-day
measurements in people with and without shoulder
impingement.26 They found a MDC95 for asymptomatic Author Contributions and Acknowledgments
individuals was 1.13 cm, whereas if participants were
symptomatic the MDC was 1.14 cm.26 The mean difference Concept/idea/research design: S.E. Harrington, D. Katsavelis
found in the present study of −0.61 falls well below this Writing: S.E. Harrington, D. Katsavelis
reported MDC95 by Rosa et al. Additionally, the calculated Data collection: S.E. Harrington, J. Hoffman, D. Katsavelis
MDC95 of 0.80 cm in the present study is larger than the Data analysis: S.E. Harrington, J. Hoffman, D. Katsavelis
Project management: S.E. Harrington
mean difference found when using the Bland-Altman
Fund procurement: S.E. Harrington
plot, again supporting the clinical applicability of the Providing participants: S.E. Harrington
measures. Providing facilities/equipment: S.E. Harrington
Consultation (including review of manuscript before submitting):
To further emphasize the clinical meaning of J. Hoffman
Bland-Altman, all of the measures fell within the limits of
The authors thank Stacy Fritz, PT, PhD, for her contribution to the
agreement, and there is no relationship between
Bland-Altman analysis and interpretation.
discrepancy and level of measurement as described by
Bunce.48 It is worth mentioning that, when comparing the
PALM and Qualisys, measured differences occurred; the Ethics Approval
Qualisys may underestimate the pectoralis minor length
when compared to the PALM or the PALM could Prior to participation, all participants signed an informed consent
form approved by the Creighton University Institutional Review
overestimate the pectoralis minor length. This could be
Board.
due to a variety of limitations when using small reflective
markers near each other, as occurred in some cases for
the placement at the fourth intercostal space, as well as Funding
having to place these markers on top of bony
prominences. Additionally, breast tissue and differences in This work was supported by the Dr George F. Haddix President’s
Faculty Research Fund of Creighton University (to S.E.H.).
breast tissue volume (eg, a unilateral mastectomy) could
cause 1 of the markers to be angled differently. Because of
these limitations, the PALM may be a more valid measure Disclosures
than the Qualisys when examining pectoralis minor
length in women with a history of breast cancer. The authors completed the ICMJE Form for Disclosure of Potential
Conflicts of Interest and reported no conflicts of interest.
There are limitations to consider when interpreting the DOI: 10.1093/ptj/pzz174
results of this study. Possible errors with palpation may
occur in placing the reflective markers on bony landmarks
when using the Qualisys and when placing the PALM’s References
caliper points on the landmarks on participants with 1 American Cancer Society. How common is breast cancer?
https://www.cancer.org/cancer/breast-cancer/about/how-
higher amounts of subcutaneous fat could exist.25,32,33 common-is-breast-cancer.html. Published 2019. Accessed
Another limitation may be the accuracy of the tracking October 10, 2019.

2020 Volume 100 Number 3 Physical Therapy 7


Pectoralis Minor Length in Breast Cancer

2 National Cancer Institute. Cancer stat facts: female breast 21 Ogston JB, Ludewig PM. Differences in 3-dimensional
cancer. https://seer.cancer.gov/statfacts/html/breast.html. shoulder kinematics between persons with multidirectional
Published 2018. Accessed November 20, 2019. instability and asymptomatic controls. Am J Sports Med.
3 Paim CR, de Paula Lima EDR, Fu MR, de A, Cassali GD. 2007;35:1361–1370.

Downloaded from https://academic.oup.com/ptj/advance-article-abstract/doi/10.1093/ptj/pzz174/5733064 by University of Liverpool user on 16 February 2020


Postlymphadenectomy complications and quality of life 22 Vermeulen H, Stokdijk M, Eilers P, Meskers C, Rozing P,
among breast cancer patients in Brazil. Cancer Nurs. Vlieland TV. Measurement of three dimensional shoulder
2008;31:302–309. movement patterns with an electromagnetic tracking device
4 Gordon LG, Scuffham P, Battistutta D, Graves N, Tweeddale in patients with a frozen shoulder. Ann Rheum Dis.
M, Newman B. A cost-effectiveness analysis of two 2002;61:115–120.
rehabilitation support services for women with breast cancer. 23 Weiser WM, Lee TQ, McMaster WC, McMahon PJ. Effects of
Breast Cancer Res Treat. 2005;94:123–133. simulated scapular protraction on anterior glenohumeral
5 American Cancer Society. Surgery for breast cancer. https:// stability. Am J Sports Med. 1999;27:801–805.
www.cancer.org/cancer/breast-cancer/treatment/surgery-for- 24 Kim SM, Park JM. Normal and abnormal US findings at the
breast-cancer.html. Published 2018. Accessed November 20, mastectomy site. Radiographics. 2004;24:357–365.
2019. 25 Borstad JD. Measurement of pectoralis minor muscle length:
6 Harrington S, Padua D, Battaglini C, et al. Comparison of validation and clinical application. J Orthop Sports Phys Ther.
shoulder flexibility, strength, and function between breast 2008;38:169–174.
cancer survivors and healthy participants. J Cancer Surviv. 26 Rosa DP, Borstad JD, Pires ED, Camargo PR. Reliability of
2011;5:167–174. measuring pectoralis minor muscle resting length in subjects
7 Kuehn T, Klauss W, Darsow M, et al. Long-term morbidity with and without signs of shoulder impingement. Braz J Phys
following axillary dissection in breast cancer patients: clinical Ther. 2016;20:176–183.
assessment, significance for life quality and the impact of 27 Borstad JD, Ludewig PM. The effect of long versus short
demographic, oncologic and therapeutic factors. Breast pectoralis minor resting length on scapular kinematics in
Cancer Res Treat. 2000;64:275–286. healthy individuals. J Orthop Sports Phys Ther.
8 Harrington S, Michener LA, Kendig T, Miale S, George SZ. 2005;35:227–238.
Patient-reported upper extremity outcome measures used in 28 Lewis JS, Valentine RE. The pectoralis minor length test: a
breast cancer survivors: a systematic review. Arch Phys Med study of the intra-rater reliability and diagnostic accuracy in
Rehabil. 2014;95:153–162. subjects with and without shoulder symptoms. BMC
9 Harrington S, Padua D, Battaglini C, Michener LA. Upper Musculoskelet Disord. 2007;8:64.
extremity strength and range of motion and their relationship 29 Gibson MH, Goebel GV, Jordan TM, Kegerreis S, Worrell TW.
to function in breast cancer survivors. Physiother Theory A reliability study of measurement techniques to determine
Pract. 2013;29:513–520. static scapular position. J Orthop Sports Phys Ther.
10 Perry S, Kowalski TL, Chang C-H. Quality of life assessment in 1995;21:100–106.
women with breast cancer: benefits, acceptability and 30 Peterson DE, Blankenship KR, Robb JB, et al. Investigation of
utilization. Health Qual life Outcomes. 2007;5:24. the validity and reliability of four objective techniques for
11 Shamley DR, Srinanaganathan R, Weatherall R, et al. measuring forward shoulder posture. J Orthop Sports Phys
Changes in shoulder muscle size and activity following Ther. 1997;25:34–42.
treatment for breast cancer. Breast Cancer Res Treat. 31 Sobush DC, Simoneau GG, Dietz KE, Levene JA, Grossman
2007;106:19–27. RE, Smith WB. The Lennie test for measuring scapular
12 Cheville AL, Tchou J. Barriers to rehabilitation following position in healthy young adult females: a reliability and
surgery for primary breast cancer. J Surg Oncol. validity study. J Orthop Sports Phys Ther. 1996;23:39–50.
2007;95:409–418. 32 Rondeau MW, Padua DA, Thigpen CA, Harrington SE.
13 Lee T, Kilbreath S, Refshauge K, Pendlebury S, Beith J, Lee M. Precision and validity of a clinical method for pectoral minor
Pectoral stretching program for women undergoing length assessment in overhead-throwing athletes. Athl Train
radiotherapy for breast cancer. Breast Cancer Res Treat. Sports Health Care. 2012;4:67–72.
2007;102:313–321. 33 Carlson M, Wilkerson J. Are differences in leg length
14 Yap KP, McCready DR, Narod S, Manchul LA, Trudeau M, predictive of lateral patello-femoral pain? Physiother Res Int.
Fyles A. Factors influencing arm and axillary symptoms after 2007;12:29–38.
treatment for node negative breast carcinoma. Cancer. 34 da BR, Armijo-Olivo S, Gadotti I, Warren S, Reid DC, Magee
2003;97:1369–1375. DJ. Reliability of scapular positioning measurement procedure
15 Manaswi A, Mehrotra N. Use of pectoralis major with or using the palpation meter (PALM). Physiotherapy.
without pectoralis minor muscle flap to fill lumpectomy in the 2010;96:59–67.
breast. Ann Plast Surg. 2010;65:23–27. 35 Petrone MR, Guinn J, Reddin A, Sutlive TG, Flynn TW,
16 Fu FH, Harner CD, Klein AH. Shoulder impingement Garber MP. The accuracy of the palpation meter (PALM) for
syndrome: a critical review. Clin Orthop Relat Res. measuring pelvic crest height difference and leg length
1991;269:162–173. discrepancy. J Orthop Sports Phys Ther. 2003;33:319–325.
17 Hébert LJ, Moffet H, McFadyen BJ, Dionne CE. Scapular 36 Performance Attainment Associates and MedNet
behavior in shoulder impingement syndrome. Arch Phys Med Technologies, Inc. PALM: PALpation Meter. Performance
Rehabil. 2002;83:60–69. Attainment Associates website. http://www.spineproducts.
18 Lin J-J, Lim HK, Yang J-L. Effect of shoulder tightness on com/paa_product_listing.php#palm. Published 2010.
glenohumeral translation, scapular kinematics, and Accessed November 20, 2019.
scapulohumeral rhythm in subjects with stiff shoulders. 37 Davis Hammonds AL, Laudner KG, McCaw S, McLoda TA.
J Orthop Res. 2006;24:1044–1051. Acute lower extremity running kinematics after a hamstring
19 Lin J-J, Hanten WP, Olson SL, et al. Functional activity stretch. J Athl Train. 2012;47:5–14.
characteristics of individuals with shoulder dysfunctions. 38 Bland JM, Altman DG. Agreement between methods of
J Electromyogr Kinesiol. 2005;15:576–586. measurement with multiple observations per individual.
20 Mell AG, LaScalza S, Guffey P, et al. Effect of rotator cuff J Biopharm Stat. 2007;17:571–582.
pathology on shoulder rhythm. J Shoulder Elbow Surg. 39 Mantha S, Roizen MF, Fleisher LA, Thisted R, Foss J.
2005;14:S58–S64. Comparing methods of clinical measurement: reporting

8 Physical Therapy Volume 100 Number 3 2020


Pectoralis Minor Length in Breast Cancer

standards for Bland and Altman analysis. Anesth Analg. 44 Host HH. Scapular taping in the treatment of anterior
2000;90:593–602. shoulder impingement. Phys Ther. 1995;75:803–812.
40 Igelström H, Emtner M, Lindberg E, Åsenlöf P. Level of 45 Kendall FP, McCreary EK, Provance PG, Rodgers M,
agreement between methods for measuring moderate to Muscles RW. Testing and Function, With Posture and

Downloaded from https://academic.oup.com/ptj/advance-article-abstract/doi/10.1093/ptj/pzz174/5733064 by University of Liverpool user on 16 February 2020


vigorous physical activity and sedentary time in people with Pain. 5th ed. Philadelphia, PA: Lippincott Williams &
obstructive sleep apnea and obesity. Phys Ther. 2013;93:50–59. Wilkins; 2005.
41 Giavarina D. Understanding Bland Altman analysis. Biochem 46 Sahrmann S. Diagnosis and treatment of
Med (Zagreb). 2015;25:141–151. movement impairment syndromes. Elsevier Health
Sciences, 1st ed. 2002;221:338–340.
42 van der N, Ostelo RW, Bekkering GE, van MW, de HC.
Minimal clinically important change for pain intensity, 47 Harrington S, Meisel C, Tate A. A cross-sectional
functional status, and general health status in patients with study examining shoulder pain and disability in
nonspecific low back pain. Spine. 2006;31:578–582. division I female swimmers. J Sport Rehabil. 2014;
23:65–75.
43 Portney LG, Watkins MP. Validity of measurements. In:
Alexander, J, ed. Foundations of Clinical Research: 48 Bunce C. Correlation, agreement, and Bland–Altman analysis:
Applications to Practice. 2nd ed. Upper Saddle River, NJ: statistical analysis of method comparison studies. Am J
Prentice Hall; 2000;35–42. Ophthalmol. 2009;148:4–6.

2020 Volume 100 Number 3 Physical Therapy 9

You might also like