IQon Spectral CT
U Acid [HU※]
Imágenes espectrales
Z eff [EAN]
basadas en Detector
permiten resultados VNC [HU※]
convencionales y espectrales
Iodine [mg/ml]
no water
100%
del tiempo sin necesidad de
Conv
120kVp
[HU]
CT Convencional
MonoE
Conv
120kVp
[HU]
[HU]
protocolos especiales de
escaneo. IQon Spectral CT
© Koninklijke Philips N.V.
Detector de doble capa
Yttrium based • La capa superior absorbe los fotones de
Low-energy detector baja energía.
GOS based • También actúa como filtro para la capa
High-energy inferior, que absorbe los fotones de alta
detector energía.
Photo
Diode
Photo Diode
Integrated ASIC
Detector: tiene la capacidad de
simultáneamente distinguir entre fotones de
diferentes energías.
© Koninklijke Philips N.V.
© Koninklijke Philips N.V.
Técnicas Basadas en las proyecciones son la ruta a la
verdadera Imagen Cuantitativa
Image Based
Image Based +
Projection Based
(Alvarez method)
“Best results are obtained by the rawdata-based Alvarez method. However, this method raises the
requirement that completely matched projection data are acquired by the CT scanner”
© Koninklijke Philips N.V.
Maaß, et. al. Medical Physics, 36(8), 2009:
© Koninklijke Philips N.V.
Spectral CT Basics
IQon Detector
Compton Photo
scatter Electric
Compton Photo
scatter Electric
Photo Compton
Electric scatter
7.4 20 53 Z Eff
Water Calcium Iodine
© Koninklijke Philips N.V.
IQon Spectral CT
Resultados
Imagen muestra Muestra la imagen Imágenes de Imágenes de Número Atómico Imágenes que Los valores
atenuación como si removiendo el densidad de densidad de efectivo muestran los presentados en la
se hubiera usado componente de material que material que valores de HU sin imagen son
una sola energía iodo permiten permiten • Permite la contribución del relativos a la
(keV) para el scan cuantificar el iodo cuantificar el iodo caracterizar las calcio en ellas. densidad
• Uso cuando no estructuras en • Evaluación de electronica del
• Aumento de la se realizan • Realce de • Realce de base a sus hernia agua (3.34x1029
señal del iodo fases sin medio de medio de materiales intervertebral. electrones x m-3)
• Mejora en contraste contraste (con contraste componentes. • Visualización en unidades %
Contrast / Noise • Manejo de el calcio (Calcio de edema óseo
• Reduce Dosis de presente) removido) • Estima la
artefacto de radiación. • Aislar el iodo • Visualización & densidad
endurecimiento cuantificación Electrónica de
• Reduce del iodo cada voxel
artefactos
metálicos
A1
A2
x A2(70keV) A1(70keV)
x
+
17
Image(E=70keV) © Koninklijke Philips N.V.
Imágenes Monoenergéticas CT
Bajo keV
Medio keV
Alto keV
Convencional 40 keV
Convencional 70 keV
Courtesy of University Hospitals Case Western Medical Center, Cleveland, OH, USA
© Koninklijke Philips N.V.
Certeza en el diagnóstico
Clinical benefit
– High MonoE spectral results and OMAR
applied to bony structures that contain
metal help improve beam hardening
artifacts that can interfere with a soft
tissue evaluation.
Case summary
– Follow-up scan was ordered on the
IQon Spectral CT to evaluate the
stabilization of a fracture.
Providing non-enhanced tissue info similar to true non contrast images with better IQ.
Material Decomposition allowing the suppression of low atomic number tissue and
quantification of contrast enhanced structures
© Koninklijke Philips N.V.
Sample images acquired in a clinical study of the IQon system at Hadassah Medical Center. Investigational device limited by law to investigational use
Evaluación de vasos en presencia de material de
embolización.
Permite la remoción de todas las estructuras no contrastadas y así poder cuantificar las
estructuras que realzan.
© Koninklijke Philips N.V.
Sample images acquired in a clinical study of the IQon system at Hadassah Medical Center. Investigational device limited by law to investigational use
Effective Atomic Number Imaging
Conventional Z Iodine
Eff No Water (mg/ml)
Spectral reconstruction of images based ONLY on the effective atomic number attenuation.
Voxels are now presented as effective atomic number value
Caracterización de cálculos
Conventional UA sobreposición
de imágenes
Clinical benefit
– Bone marrow edema is a medical condition
that causes decreased fat and increased fluid
contents within the bone marrow.
– This is normally not visualized on a typical CT
scan.
Case summary
– 18-year-old received a CT scan after he fell to
rule out if there was a fracture.
Conventional Calcium Suppressed MRI – He then had an MRI 24 hours later to rule out
any pathology.
Tools used
– Spectral CT Viewer
Images courtesy of Dr. Gopal Punjabi, Hennepin County Medical Center © Koninklijke Philips N.V.
Sudden onset of abdominal pain
Conventional CT, coronal plane. Iodine map: Absent uptake in Iodine overlay shows finding
The optic nerves are right optic nerve (blue arrow). nicely!
unremarkable. The left optic nerve shows
good iodine uptake (red arrow).
Images courtesy of Dr. Ben Hoffman, Hennepin County Medical Center © Koninklijke Philips N.V.
SPECTRAL CT
IQon Spectral CT
with IntelliSpace Portal
SPECTRAL VIEWER SPECTRAL TUMOR SPECTRAL SPECTRAL AVA SPECTRAL MAGIC
TRACKING COMPREHENSIVE GLASS ON PACS
CARDIAC (CCA)
OTHER NEURO/ED-TRAUMA
CARDIOVASCULAR
SPECTRAL
APPLICATIONS
CT TAVI COMPREHENSIVE CALCIUM SCORING DYNAMIC CARDIAC PLAQUE CT/AMI FUSION EP PLANNING CARDIAC VIEWER AVA & STENT
CARDIAC MYOCARDIAL ASSESSMENT PLANNING
PERFUSION © Koninklijke Philips N.V.
Powerful advancements that fit your workflow
Brain
Clinical benefit
– Spectral Magic Glass on IQon Spectral
CT allows for simultaneous comparison
of multiple spectral results for a specific
region of interest.
Case summary
– Patient presented to ED with symptoms
of an acute stroke, and was referred to
CT.
– Patient was scanned on the IQon
Spectral CT, and spectral results were
available in less than three minutes
post scan.
– Low MonoE improved the visualization
of the perfusion defect in the pons.
Tools used
– Spectral CT Viewer
– Spectral Magic Glass
3 © Koninklijke Philips N.V.
Images courtesy of: AUH, Aarhus, Denmark Results from case studies are not predictive of results in other cases. Results in other cases may vary.
Fase Sin Supresión de Densidad
MonoEnergéticas Iodo / NO Agua Densidad de Iodo Z Efectivo
Contraste Virtual Calcio electrónica
Imagen muestra Muestra la imagen Imágenes de Imágenes de Número Atómico Imágenes que Los valores
atenuación como si removiendo el densidad de densidad de efectivo muestran los presentados en la
se hubiera usado componente de material que material que valores de HU sin imagen son
una sola energía iodo permiten permiten • Permite la contribución del relativos a la
(keV) para el scan cuantificar el iodo cuantificar el iodo caracterizar las calcio en ellas. densidad
• Uso cuando no estructuras en • Evaluación de electronica del
• Aumento de la se realizan • Realce de • Realce de base a sus hernia agua (3.34x1029
señal del iodo fases sin medio de medio de materiales intervertebral. electrones x m-3)
• Mejora en contraste contraste (con contraste componentes. • Visualización en unidades %
Contrast / Noise • Manejo de el calcio (Calcio de edema óseo
• Reduce Dosis de presente) removido) • Estima la
artefacto de radiación. • Aislar el iodo • Visualización & densidad
endurecimiento cuantificación Electrónica de
• Reduce del iodo cada voxel
artefactos
metálicos