De / From Bien-Air Dental SA, Switzerland

Service après vente / After sales service


À / to Réparateurs Bien-Air / Bien-Air’s repairs
Date 23.02.2011

Technical Information N° 73
For an appropriate use of the ultrasonic scaler POWERCARE.

Purpose
Ensure that the manufacturer’s recommendations for an intermittent function are applied during the
use of the ultrasonic scaler POWERCARE. Be sure that you inform the practitioner of this point in the
“User guide”

REF Designation See on …


1700343-001 Set POWERCARE: http://www.bienair.com/dental_products_b.asp?familly=506
2100188/12.09 User guide page 50, Technical data
2100190/12.09 Installation instruction page 53, Technical data
Download instructions: http://www.bienair.com/dental_support_instruction.asp

Important advice of use!


The ultrasonic scaler POWERCARE is an intermittent function device. Please refer to “User
guide” and read the following extracts.
Function with irrigation: ON use for 60 seconds, then OFF 30 sec.
Utilisation without irrigation: ON use for 30 seconds, then OFF 120 sec.

Caution: Use of the device outside of these recommended limits, may result in damage to the device
and subsequent voiding of the warranty.

Validity since delivery of the device.


For further information, write us to the following e-mailbox: aftersales-ch@bienair.com