Download as pdf or txt
Download as pdf or txt
You are on page 1of 6

Acta Oncologica

ISSN: 0284-186X (Print) 1651-226X (Online) Journal homepage: https://www.tandfonline.com/loi/ionc20

Magnetic resonance only workflow and validation


of dose calculations for radiotherapy of prostate
cancer

Rasmus L. Christiansen, Henrik R. Jensen & Carsten Brink

To cite this article: Rasmus L. Christiansen, Henrik R. Jensen & Carsten Brink (2017) Magnetic
resonance only workflow and validation of dose calculations for radiotherapy of prostate cancer,
Acta Oncologica, 56:6, 787-791, DOI: 10.1080/0284186X.2017.1290275

To link to this article: https://doi.org/10.1080/0284186X.2017.1290275

View supplementary material

Published online: 20 Feb 2017.

Submit your article to this journal

Article views: 1279

View related articles

View Crossmark data

Citing articles: 10 View citing articles

Full Terms & Conditions of access and use can be found at


https://www.tandfonline.com/action/journalInformation?journalCode=ionc20
ACTA ONCOLOGICA, 2017
VOL. 56, NO. 6, 787–791
http://dx.doi.org/10.1080/0284186X.2017.1290275

ORIGINAL ARTICLE

Magnetic resonance only workflow and validation of dose calculations for


radiotherapy of prostate cancer
Rasmus L. Christiansena, Henrik R. Jensena and Carsten Brinka,b
a
Laboratory of Radiation Physics, Department of Oncology, Odense University Hospital, Odense, Denmark; bInstitute of Clinical Science,
University of Southern Denmark, Odense, Denmark

ABSTRACT ARTICLE HISTORY


Background: Current state of the art radiotherapy planning of prostate cancer utilises magnetic reson- Received 29 December 2016
ance (MR) for soft tissue delineation and computed tomography (CT) to provide an electron density Accepted 29 January 2017
map for dose calculation. This dual scan workflow is prone to setup and registration error. This study
evaluates the feasibility of an MR-only workflow and the validity of dose calculation from an MR
derived pseudo CT.
Material and methods: Thirty prostate cancer patients were CT and MR scanned. Clinical treatment
plans were generated on CT using a single 18 MV arc volumetric modulated arc therapy (VMAT) with a
prescription of 78 Gy/39 fractions. Dose was recalculated on pseudo CT and assuming uniform water
density. Pseudo CT and uniform density based dose calculations were compared to CT dose calcula-
tions by gamma analysis. One patient was treated with a plan based solely on MR and pseudo CT
including daily image guided radiotherapy (IGRT) performed by manual match of implanted gold
markers.
Results: A pseudo CT was generated for 29 of the 30 patients. Median gamma pass rates for 1%/1 mm
passing criteria for dose calculated on pseudo CT when compared to CT were 100% for most evaluated
structures. Dose calculated on uniform density also yielded high median pass rates, but with a higher
occurrence of pass rates below 95%. Cases of pass rate below 95% on pseudo CT proved to originate
from the presence of rectal air on CT, not represented by the pseudo CT. Treatment based on MR
alone was successfully delivered to one patient, including manually performed daily IGRT.
Conclusions: Median gamma pass rates were high for pseudo CT and proved superior to uniform
density. Local differences in dose calculations were concluded not to have clinical relevance. Feasibility
of the MR-only workflow was demonstrated through successful delivery of a treatment course planned
based on MR alone.

Background recommended (or indeed available) for clinical use. However,


Ko€hler et al. [10] report on the same pseudo CT algorithm as
Magnetic resonance (MR) images provide the gold standard
evaluated in the current study published by the manufac-
for defining soft tissue structures during radiotherapy plan-
turer, investigating only the high dose volume and not spe-
ning for prostate cancer [1–3]. However, in order to calculate
cific organs at risk (OAR). A consensus on evaluating and
a dose distribution the electron density map of a computed
reporting these evaluations does not currently exist.
tomography (CT) scan is required. This necessitates a dual
modality workflow and image co-registration which is cur- Furthermore, different institutions apply different treatment
rently standard in many clinics. However, this workflow intro- techniques which may impact on the dose calculations,
duces the risk of setup error between the scans as well as depending on the given pseudo CT. Therefore, a validation
registration error. An MR-only workflow eliminates set-up and of dose calculated on the specific pseudo CT in concordance
registration error while also reducing workload and strain on with the treatment technique used is necessary. To the
the patient. The use of MR guided treatment delivery in the extent of our knowledge, previous studies have until now
years to come provides further argument for use of an MR- only been performed for the target region and the whole
only workflow [4,5]. Therefore, several groups have under- body volume (with a low-dose threshold), thus not sufficient
taken work to create an electron density map from MR; a for clinical implementation. This study investigates the valid-
pseudo CT [6–11]. These studies evaluate different ity of dose calculations based on the first commercially avail-
approaches to pseudo CT generation and report high able pseudo CT derived from MR in comparison to CT
degrees of similarities of dose calculations compared to CT. without any modifications to the pseudo CT and with respect
Most of these are still under development and not explicitly to target volumes and specific OAR. Furthermore, the

CONTACT Rasmus L€ ubeck Christiansen Rasmus.Lubeck.Christiansen@rsyd.dk Laboratory of Radiation Physics, Department of Oncology, Odense University
Hospital, Odense, Denmark
Supplemental data for this article can be accessed here.
ß 2017 Acta Oncologica Foundation
788 R. L. CHRISTIANSEN ET AL.

planning and treatment workflow were validated end-to-end were equipped for radiotherapy with a flat, indexed table
to evaluate the feasibility of MR-only planning. top, knee-and foot fixation device (Bionix, Toledo, OH) and
external laser positioning system (LAP, Lu€neburg, Germany).
All acquired MR sequences are listed in Table S1.
Material and methods Small FOV T2w MR scans were rigidly co-registered with
Patients CT for the purpose of delineating the target structures and
selected OAR. The CT was used to delineate the remaining
Thirty prostate cancer patients referred for localised, curative OAR.
radiotherapy (78 Gy in 39 fractions) were included in the Treatments were planned on CT in Pinnacle v9.10 (Philips,
study. All patients included in the study were referred for CT Milwaukee, WI, USA) as single arc volumetric modulated arc
and MR scans on the same day as part of the standard pro- therapy (VMAT) with an 18 MV beam. All dose calculation
cedure. Written informed consent to acquire the additional was based on the standard CT to density conversion table in
MRCAT sequence during the MR scan and approval for use in our clinical treatment planning system, thus no correction of
the current study was obtained from all patients. this table was performed for the pseudo CT data.

Pseudo CT generation Comparison of dose calculations

Magnetic Resonance for Calculating Attenuation (MRCAT) To minimise errors in dose calculation related to setup uncer-
(Philips, Helsinki, Finland) is a commercially available solution tainties, the planning CT was rigidly co-registered to the
for generation of pseudo CT for dose calculation purposes pseudo CT using Elastix [12] and reformatted to fit identically
for prostate cancer. MRCAT consists of a T1 weighted (T1w), in terms of voxel size, slice positioning and field of view
dual echo mDixon scan sequence providing the source (FOV) of the pseudo CT. In this process, the voxel dimensions
images for the subsequent reconstruction algorithm which of the CT were changed from 0.976 mm  0.976 mm  3 mm
outputs a pseudo CT. to 1.04 mm  1.04 mm  2.5 mm.
The MRCAT pseudo CT image is not identical to a CT as it Beam parameters, points and regions of interest (ROI) and
comprises five discrete Hounsfield Units (HU) ( 968, 86, 42, dose grid of the clinical plan were transferred to three image
198 and 949) to discriminate between air, fat, water/muscle, sets: pseudo CT, the reformatted clinical CT and a version of
the pseudo CT where the body was assigned to a uniform
spongious bone and cortical bone (Figure 1). The MRCAT
density of 1 g/cm3. The dose was recalculated with a dose
reconstruction algorithm utilises the information of the
grid resolution of 3 mm on each image set, yielding three
mDixon MR images to adapt a generic anatomical model to
dose distributions for each patient. For each patient, the simi-
the individual patients. The resulting pseudo CT does not
larity of dose distribution calculated based on both pseudo CT
represent non-tissue densities within the patient outline, e.g.,
and uniform density, respectively, were evaluated against that
volumes of air or metallic implants, such as prosthetics or
of CT using pass rate of the 3D Gamma index, as described
fiducial markers.
by Low et al. [13]. The dose acceptance criterion was eval-
uated as a 2% or 1% percentage of the prescribed dose and
Planning procedure distance criteria of 2 mm and 1 mm, respectively. The calcula-
tions were performed by in-house developed MatLab script
The planning workflow for included patients was as per the (Matlab version 2014a, The MathWorks Inc., Natick, MA). The
standard protocol, except addition of the MRCAT scan proto- calculation was performed for the entire scanned volume,
col to MR. Dose comparisons were performed retrospectively. thus no dose threshold was used in the current study.
CT scans were performed on a Philips Big Bore Brilliance 16
slice CT scanner (Philips, Best, The Netherlands) with 140 kVp
Verification of workflow end-to-end
and reconstructed to a slice thickness of 3 mm, 512  512
matrix. MR scans were performed on a Philips Ingenia 1.5 T To verify that the ultimate goal of delivering a full course of
MR scanner (Philips, Best, The Netherlands). Both scanners treatment planned solely on MR is feasible, the entire

Figure 1. CT (left) comprises a range of 4000 HU, whereas the MRCAT generated pseudo CT represents the range of attenuations with five HU bins. A gold fiducial
marker and rectal air are visual on CT in contrast to pseudo CT.
ACTA ONCOLOGICA 789

planning process was performed by MR alone for one on pseudo CT. Daily online verification of patient setup was
patient. A full FOV T2 weighted MR provided the basis for performed by CBCT prior to treatment delivery. Gold seeds
OAR delineations. Prior to treatment delivery, the clinical implanted in the prostate visual on the CBCT were manually
dose calculation based on pseudo CT was verified against CT matched to the corresponding delineations of the markers
by setting up beam parameters, etc. and recalculating the from the planning MR (Figure S1).
dose, similar to the workflow described above. Fiducial
markers were identified on MR and delineated for daily setup
verification with cone beam CT (CBCT). Discussion
This study has evaluated the quality of dose calculation
based on an unmodified commercially available pseudo CT
Results
to that of CT. Evaluation was performed by analysis of the
Of the 30 patients included, 29 had successfully generated gamma index for the entire scanned volume as well as indi-
pseudo CTs which were included in the analysis. vidual target and OAR to assess clinical impact of MR-only
Results of gamma analysis are summarised for the patient radiotherapy planning. Subsequently, one patient was treated
cohort in Tables 1 and 2. Median values were 100% for all based on MR images alone to prove the feasibility of an MR-
evaluated structures under 2% and 2 mm passing criteria only planning workflow.
both for pseudo CT and uniform density based calculations. A pseudo CT could not be calculated for one patient but
Under the stricter passing criteria of 1% and 1 mm, median it was not possible to investigate why this was so within the
values were also 100% for all structures calculated on the current study as we did not have access to the pseudo CT
pseudo CT. A few individual structures yielded pass rates source code. However, this was most likely related to a
below 95%. These incidents were linked to rectal air being ‘sanity check’ performed in the software, evaluating whether
present on CT, but not on the pseudo CT (see Figure 2). the obtained pseudo CT seems plausible. If not no output is
Dose calculations based on uniform density resulted in generated. The cause of the non-successful generation of
median gamma pass rates of 97.5% and above for all eval- pseudo CT was thus likely an atypical anatomy of that
uated structures under 1% and 1 mm criteria. However, a patient.
large portion of evaluated structures had lower pass rates Very high median pass rates were seen for both pseudo
than for the pseudo CT data, as illustrated by the 16.7–83.3 CT and uniform density for 2% and 2 mm passing criteria,
percentiles in Table 1. although outliers were less prominent for pseudo CT. This
Visual inspection of gamma index overlay on CT (as tendency was more pronounced when the stricter 1%/1 mm
shown in Figure 2) for each patient also revealed that gener- passing criteria were applied; where pseudo CT still showed
ally areas near the skin surface were prone to failure. consistently high pass rates for all patients, uniform density
For one patient, all delineations of target and OAR were yielded several gamma index failures, as illustrated by the
performed on MRI for dose calculation and plan optimisation ranges of 16.7–83.3 percentiles of Table 1. The small ranges

Table 1. Median gamma pass rates shown for selected structures for CT vs pseudo CT and CT vs uniform density dose calculations.
16.7–83.3 percentiles mimicking 1 standard deviation, had data been normally distributed are shown in brackets.
2% and 2 mm 1% and 1 mm
Pseudo CT Uniform density Pseudo CT Uniform density
PTV78 100 (99.7–100) 100 (98.4–100) 99.9 (96.7–100) 97.1 (56.5–100)
Prostate 100 (100–100) 100 (99.9–100) 100 (98.7–100) 98.5 (23.7–100)
Seminal vesicles 100 (100–100) 100 (100–100) 100 (99.3–100) 100 (71.1–100)
Rectum 100 (99.1–100) 100 (98.2–100) 100 (94.9–100) 100 (89.9–100)
Small bowel incl. sigmoid 100 (99.6–100) 100 (99.9–100) 100 (97.5–100) 100 (97.6–100)
Bladder 100 (100–100) 100 (100–100) 100 (99.9 –100) 100 (95.8–100)
Penile bulb 100 (100–100) 100 (100–100) 100 (100–100) 100 (100–100)
Femoral heads 100 (100–100) 100 (100–100) 100 (100–100) 100 (99.1–100)
Body 99.7 (99.3–99.8) 99.7 (99.4–99.9) 98.9 (98.3–99.2) 98.9 (98.2–99.3)

Table 2. Mean gamma pass rates (and their standard deviation) shown for selected structures for CT vs pseudo CT and CT vs uniform
density dose calculations for comparison with results of other groups, although data are not normally distributed.
2% and 2 mm 1% and 1 mm
Pseudo CT Uniform density Pseudo CT Uniform density
PTV78 99.8 (0.5) 98.9 (2.7) 97.3 (7.1) 82.5 (24.3)
Prostate 99.9 (0.3) 97.6 (7.7) 97.4 (10.4) 74.3 (35.5)
Seminal vesicles 99.9 (0.3) 99.8 (0.7) 96.7 (10.0) 88.7 (19.3)
Rectum 99.2 (2.7) 98.7 (3.2) 96.8 (7.0) 95.4 (8.6)
Small bowel incl. sigmoid 99.7 (1.0) 98.2 (8.8) 98.9 (2.3) 95.9 (17.8)
Bladder 100.0 (0.0) 100.0 (1.8) 98.9 (4.8) 96.2 (9.9)
Penile bulb 100.0 (0.0) 100.0 (0.0) 100.0 (0.0) 99.0 (5.5)
Femoral heads 100.0 (0.0) 99.9 (0.6) 100.0 (0.2) 97.6 (10.2)
Body 99.5 (0.4) 99.5 (0.7) 98.7 (0.9) 98.1 (3.0)
790 R. L. CHRISTIANSEN ET AL.

Figure 2. Gamma index (1%/1 mm) shown as coloured overlay on CT.

of 16.7–83.3 percentiles of pseudo CT based calculations indi- pass rate of D75% for 1% and 1 mm criteria. In their study,
cate that calculations are accurate on patient level, not just differences in outer contour between CT and pseudo CT
the cohort. were overwritten with water density to minimise the effect
For pseudo CT calculation, gamma failures were seen pri- of setup and registration error [10]. Similarly, Dowling et al.
marily in conjunction with rectal air being present on CT but [11] added 1 mm to the patient circumference to adjust a
not pseudo CT. Given that the full treatment course is deliv- drop in signal on the skin surface [11]. In contrast to these
ered over 39 fractions, identical occupation of the rectal vol- studies, no changes in the obtained pseudo CT were made
ume by air is unlikely. Therefore, it is likely that it is at least in the present study, as the purpose was to test the validity
as good as an approximation to overwrite these moving air of the dose calculation based on the pseudo CT generated
pockets with water density rather than assuming stationary by the commercially available software package. For the
air throughout the treatment course. same reason, the gamma index was calculated and reported
Failures of the gamma index in the skin areas occurred pri- for the entire scanned volume.
marily due to the inevitable setup error between CT and MR. This validation of the dose calculation utilising electron
Furthermore, the volumes of failure are small in comparison density information from the pseudo CT was undertaken
to the entire body volume (5%), and so is the absolute for a specific treatment technique; 18 MV full arc VMAT,
dose deposited in these areas. Therefore these gamma fail- which will de-emphasise local HU errors, since only a small
ures are not considered to lessen the validity of the MRCAT- part of the dose is delivered in a specific direction. The
CT based dose calculation. On the other hand, when a thresh- high dose volume is placed distant from the skin, thus this
old is applied, such as Korhonen et al. [9], Dowling et al. [11] will not be as sensitive to small differences in the defin-
and Ko €hler et al. [10] do, the effect is not at all evaluated. ition of the outer contour of the patient. The same argu-
There are variations in the method of reporting gamma ments to some extent explain why homogeneous water
values among studies. Some groups use mean instead of performs as well as it does in the comparison. Other treat-
median gamma pass rates although data are unlikely to be ment sites and techniques will require individual
normally distributed. For the purpose of comparison to those validations.
results, mean pass rates of the gamma indexes in this study For a daily online setup verification to be practical, an
are given in Table 2. The dual model HU conversion of automatic match of CBCT to the planning image should be
Korhonen et al. [9] yielded mean 2D gamma pass rates for possible. Currently, the vendor does not provide such a solu-
VMAT plans of 99.9% for 2% and 2 mm passing criteria and tion. To manage this, an in-house developed solution is thus
93.2% for 1% and 1 mm for the D10% volume. The mean required and has now been developed locally to facilitate an
pass rate of 98.7% for 1% and 1 mm found in the current automatic match in the future workflow.
study for the entire body is superior to this. However, accord- In conclusion, dose calculated based on MRCAT pseudo
ing to Pulliam et al. [14], their use of a 2D gamma analysis is CT was practically indistinguishable from those of CT and as
likely to result in slightly lower pass rates than if a 3D ana- such considered approved for clinical application for prostate
lysis had been performed, such as in the current study. cancer with the described treatment technique. Gamma pass
Dowling et al. [11] evaluated dose calculations based on rates were consistently higher for pseudo CT than uniform
pseudo CT generated from a standard T2w sequence and density when passing criteria of 1% and 1 mm are applied.
reported mean 3D gamma pass rates of 100% for the D90% One patient was successfully treated with a dose plan gener-
volume (corresponding roughly to the PTV78 in the present ated from MR only, demonstrating the practical feasibility of
study) under 2% and 2 mm as well as 1% and 1 mm criteria. an MR only work flow for prostate cancer treatment.
Mean pass rates for D10% were reported to be 95% (2% and However, automatic matching between CBCT and pseudo CT
2 mm) and 93% (1% and 1 mm). As in the present study, a was not possible due to implanted fiducials not being shown
global dose was used as reference for the gamma index cal- by the pseudo CT.
culation, although it is unclear whether they also chose the
prescription dose.
Disclosure statement
The MRCAT generated pseudo CT was evaluated by the
manufacturer on clinical data resulting in 99.83% gamma The authors report no conflicts of interest.
ACTA ONCOLOGICA 791

References [8] Dowling JA, Lambert J, Parker J, et al. An atlas-based electron


density mapping method for magnetic resonance imaging (MRI)-
[1] Rasch C, Barillot I, Remeijer P, et al. Definition of the prostate in alone treatment planning and adaptive MRI-based prostate radi-
CT and MRI: a multi-observer study. Int J Radiat Oncol Biol Phys. ation therapy. Int J Radiat Oncol Biol Phys. 2012;83:e5–e11.
1999;43:57–66. [9] Korhonen J, Kapanen M, Keyril€ainen J, et al. A dual model HU
[2] Khoo VS, Joon DL. New developments in MRI for target volume conversion from MRI intensity values within and outside of bone
delineation in radiotherapy. Br J Radiol. 2006;79:S2–S15. segment for MRI-based radiotherapy treatment planning of pros-
[3] Dirix P, Haustermans K, Vandecaveye V. The value of magnetic tate cancer. Med Phys. 2014;41:011704.
resonance imaging for radiotherapy planning. Semin Radiat [10] Ko€hler M, Vaara T, van Grootel M, et al. MR-only simulation for
Oncol. 2014;24:151–159. radiotherapy planning. Koninklijke Philips N V. 2015;1–15.
[4] Lagendijk JJW, Raaymakers BW, van Vulpen M. The magnetic res- [11] Dowling JA, Sun J, Pichler P, et al. Automatic substitute com-
onance imaging-linac system. Semin Radiat Oncol. puted tomography generation and contouring for magnetic res-
2014;24:207–209. onance imaging (MRI)-alone external beam radiation therapy
[5] Mutic S, Dempsey JF. The ViewRay system: magnetic resonance- from standard MRI sequences. Int J Radiat Oncol Biol Phys.
guided and controlled radiotherapy. Semin Radiat Oncol. 2015;93:1144–1153.
2014;24:196–199. [12] Klein S, Staring M, Murphy K, et al. Elastix: a toolbox for intensity-
[6] Chen L, Price RA, Wang L, et al. MRI-based treatment planning for based medical image registration. IEEE Trans Med Imaging.
radiotherapy: dosimetric verification for prostate IMRT. Int J 2010;29:196–205.
Radiat Oncol Biol Phys. 2004;60:636–647. [13] Low DA, Harms WB, Mutic S, et al. A technique for the quantita-
[7] Johansson A, Karlsson M, Nyholm T. CT substitute derived from tive evaluation of dose distributions. Med Phys. 1998;25:656–661.
MRI sequences with ultrashort echo time. Med Phys. 2011;38: [14] Pulliam KB, Huang JY, Howell RM, et al. Comparison of 2D and
2708–2714. 3D gamma analyses. Med Phys. 2014;41:021710.

You might also like