VWR MU 6100L Multi Meter Operating Manual
VWR MU 6100L Multi Meter Operating Manual
VWR MU 6100L Multi Meter Operating Manual
EU cat. no NA cat. no
MU 6100 L S1 665-0309 76460-494
MU 6100 L S2 665-0310 76460-496
MU
610
0L
ba77076e06 05/2021
MU 6100 L
Importer to UK:
2 ba77076e06 05/2021
MU 6100 L Contents
MU 6100 L - Contents
1 Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
1.1 Safety information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
1.1.1 Safety information in the operating manual . . . . . 7
1.1.2 Safety signs on the meter . . . . . . . . . . . . . . . . . . 7
1.1.3 Further documents providing safety information . 7
1.2 Safe operation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
1.2.1 Authorized use . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
1.2.2 Requirements for safe operation . . . . . . . . . . . . . 8
1.2.3 Unauthorized use . . . . . . . . . . . . . . . . . . . . . . . . . 8
2 Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
2.1 MU 6100 L meter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
2.2 Sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
3 Technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
3.1 Measuring ranges, resolution, accuracy . . . . . . . . . . . . . 10
3.1.1 pH/ORP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
3.1.2 Oxi . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
3.1.3 Cond. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
3.2 General data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
4 Commissioning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
4.1 Scope of delivery. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
4.2 Power supply. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
4.3 Initial commissioning . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
4.3.1 Inserting the batteries. . . . . . . . . . . . . . . . . . . . . 16
4.3.2 Connect the power pack . . . . . . . . . . . . . . . . . . 16
4.3.3 Stand . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
5 Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
5.1 General operating principles . . . . . . . . . . . . . . . . . . . . . . 18
5.1.1 Keypad . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
5.1.2 Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
5.1.3 Status information (meter) . . . . . . . . . . . . . . . . . 19
5.1.4 Connectors. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
5.2 Switching on the meter . . . . . . . . . . . . . . . . . . . . . . . . . . 20
5.3 Switching off . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
5.4 Navigation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
5.4.1 Operating modes . . . . . . . . . . . . . . . . . . . . . . . . 21
ba77076e06 05/2021 3
Contents MU 6100 L
6 pH value . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
6.1 Measuring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
6.1.1 Measuring the pH value . . . . . . . . . . . . . . . . . . . 28
6.1.2 Measuring the temperature. . . . . . . . . . . . . . . . . 29
6.2 Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
6.2.1 Why calibrate? . . . . . . . . . . . . . . . . . . . . . . . . . . 30
6.2.2 When do you have to calibrate? . . . . . . . . . . . . . 30
6.2.3 Automatic calibration (AutoCal) . . . . . . . . . . . . . 30
6.2.4 Manual calibration (AnyCal) . . . . . . . . . . . . . . . . 33
6.2.5 Calibration points . . . . . . . . . . . . . . . . . . . . . . . . 36
6.2.6 Calibration data. . . . . . . . . . . . . . . . . . . . . . . . . . 37
7 ORP voltage. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
7.1 Measuring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
7.1.1 Measuring the ORP . . . . . . . . . . . . . . . . . . . . . . 39
7.1.2 Measuring the temperature. . . . . . . . . . . . . . . . . 40
7.2 ORP calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
8 Ion concentration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
8.1 Measuring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
8.1.1 Measuring the ion concentration . . . . . . . . . . . . 41
8.1.2 Measuring the temperature. . . . . . . . . . . . . . . . . 42
8.2 Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
8.2.1 Why calibrate? . . . . . . . . . . . . . . . . . . . . . . . . . . 43
8.2.2 When to calibrate? . . . . . . . . . . . . . . . . . . . . . . . 43
8.2.3 Calibration (ISE Cal) . . . . . . . . . . . . . . . . . . . . . . 43
8.2.4 Calibration standards . . . . . . . . . . . . . . . . . . . . . 46
8.2.5 Calibration data. . . . . . . . . . . . . . . . . . . . . . . . . . 46
9 Dissolved oxygen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
9.1 Measuring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
9.1.1 Measuring D.O.. . . . . . . . . . . . . . . . . . . . . . . . . . 49
9.1.2 Measuring the temperature. . . . . . . . . . . . . . . . . 50
9.2 Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
9.2.1 Why calibrate? . . . . . . . . . . . . . . . . . . . . . . . . . . 51
9.2.2 When to calibrate? . . . . . . . . . . . . . . . . . . . . . . . 51
4 ba77076e06 05/2021
MU 6100 L Contents
10 Conductivity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
10.1 Measuring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
10.1.1 Measuring the conductivity. . . . . . . . . . . . . . . . . 56
10.1.2 Measuring the temperature . . . . . . . . . . . . . . . . 57
10.2 Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
10.2.1 Why calibrate? . . . . . . . . . . . . . . . . . . . . . . . . . . 58
10.2.2 When to calibrate? . . . . . . . . . . . . . . . . . . . . . . . 58
10.2.3 Determining the cell constant (calibration in
control standard) ...................... 58
10.2.4 Calibration data . . . . . . . . . . . . . . . . . . . . . . . . . 59
11 Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
11.1 Measurement settings. . . . . . . . . . . . . . . . . . . . . . . . . . . 61
11.1.1 Settings for pH measurements. . . . . . . . . . . . . . 61
11.1.2 Buffer sets for calibration . . . . . . . . . . . . . . . . . . 62
11.1.3 Calibration interval . . . . . . . . . . . . . . . . . . . . . . . 64
11.1.4 Settings for ORP measurements . . . . . . . . . . . . 64
11.1.5 Settings for D.O. sensors . . . . . . . . . . . . . . . . . . 65
11.1.6 Calibration interval . . . . . . . . . . . . . . . . . . . . . . . 66
11.1.7 Settings for conductivity measuring cells . . . . . . 67
11.1.8 Calibration interval . . . . . . . . . . . . . . . . . . . . . . . 69
11.1.9 Temperature compensation . . . . . . . . . . . . . . . . 69
11.1.10 Setting the TDS factor . . . . . . . . . . . . . . . . . . . . 70
11.2 Sensor-independent settings . . . . . . . . . . . . . . . . . . . . . 71
11.2.1 System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71
11.2.2 Data storage. . . . . . . . . . . . . . . . . . . . . . . . . . . . 72
11.2.3 Automatic Stability control . . . . . . . . . . . . . . . . . 72
11.3 Reset. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72
11.3.1 Resetting the measurement settings. . . . . . . . . 72
11.3.2 Resetting the system settings . . . . . . . . . . . . . . 74
12 Data memory . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75
12.1 Manual storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75
12.2 Automatic data storage at intervals .............. 76
12.3 Measurement data memory . . . . . . . . . . . . . . . . . . . . . . 78
12.3.1 Editing the measurement data memory . . . . . . . 78
12.3.2 Erasing the measurement data memory . . . . . . 80
12.3.3 Measurement dataset . . . . . . . . . . . . . . . . . . . . 80
12.3.4 Memory locations . . . . . . . . . . . . . . . . . . . . . . . . 80
ba77076e06 05/2021 5
Contents MU 6100 L
15 What to do if... . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 90
15.1 pH/ORP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 90
15.2 Dissolved oxygen. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 91
15.3 Conductivity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 93
15.4 General information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 94
16 Firmware update . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 95
17 Glossary. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 96
18 Index. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 99
20 Warranty. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 108
6 ba77076e06 05/2021
MU 6100 L Safety
1 Safety
1.1 Safety information
WARNING
indicates a possibly dangerous situation that can lead to seri-
ous (irreversible) injury or death if the safety instruction is not
followed.
CAUTION
indicates a possibly dangerous situation that can lead to
slight (reversible) injury if the safety instruction is not fol-
lowed.
NOTE
indicates a possibly dangerous situation where goods might be damaged
if the actions mentioned are not taken.
Note all labels, information signs and safety symbols on the meter and in the
battery compartment. A warning symbol (triangle) without text refers to safety
information in this operating manual.
ba77076e06 05/2021 7
Safety MU 6100 L
This meter is authorized exclusively for pH, ISE, ORP, dissolved oxygen and
conductivity measurements in a laboratory environment.
Only the operation and running of the meter according to the instructions and
technical specifications given in this operating manual is authorized (see
section 3 TECHNICAL DATA, page 10).
Any other use is considered unauthorized.
8 ba77076e06 05/2021
MU 6100 L Overview
2 Overview
2.1 MU 6100 L meter
The MU 6100 L meter enables you to perform measurements (pH, U, conduc-
tivity, dissolved oxygen) ) quickly and reliably.
The MU 6100 L provides the maximum degree of operating comfort, reliability
and measuring certainty for all applications.
2.2 Sensors
A measuring system ready to measure consists of the MU 6100 L meter and a
suitable sensor.
ba77076e06 05/2021 9
Technical data MU 6100 L
3 Technical data
3.1 Measuring ranges, resolution, accuracy
3.1.1 pH/ORP
10 ba77076e06 05/2021
MU 6100 L Technical data
3.1.2 Oxi
ba77076e06 05/2021 11
Technical data MU 6100 L
3.1.3 Cond
12 ba77076e06 05/2021
MU 6100 L Technical data
ba77076e06 05/2021 13
Technical data MU 6100 L
14 ba77076e06 05/2021
MU 6100 L Commissioning
4 Commissioning
4.1 Scope of delivery
MeterMU 6100 L
4 batteries 1.5 V Mignon type AA
Power pack
Stand
Stand holder
Short instructions
CD-ROM with
– USB drivers
– detailed operating manual
– Software MultiLab Importer
ba77076e06 05/2021 15
Commissioning MU 6100 L
1 Battery compartment
CAUTION
Make sure that the poles of the batteries are positioned cor-
rectly.
The ± signs on the batteries must correspond to the ± signs in
the battery compartment.
You can operate the meter either with normal batteries or with
rechargeable batteries (Ni-MH). In order to charge the batteries, an
external charging device is required.
CAUTION
The line voltage at the operating site must lie within the input
voltage range of the original power pack (see section 3 TECH-
NICAL DATA, page 10).
16 ba77076e06 05/2021
MU 6100 L Commissioning
CAUTION
Use original power packs only.
1. Connect the plug of the power pack to the socket for the power pack on
the MU 6100 L.
2. Connect the original power pack to an easily accessible power outlet.
4.3.3 Stand
The stand base can be mounted at the right side of the meter.
ba77076e06 05/2021 17
Operation MU 6100 L
5 Operation
5.1 General operating principles
5.1.1 Keypad
18 ba77076e06 05/2021
MU 6100 L Operation
5.1.2 Display
3 4
pH
2
6.093
25.0 °C 5
1 AutoCal TEC 6
HOLD AR 7
15.03.2014
Menu 8
08:00
ba77076e06 05/2021 19
Operation MU 6100 L
5.1.4 Connectors
4 1 2 3 5 6 7
1 pH electrode
2 Reference electrode / temperature sensor
3 Reference electrode / temperature sensor
4 D.O. sensor or conductivity measuring cell
5 Mini USB-B interface
6 Power pack
7 Service interface
CAUTION
Only connect sensors to the meter that cannot return any volt-
ages or currents that are not allowed (> SELV and > current
circuit with current limiting).
Almost all customary sensors fulfill these conditions.
pH
6.949
25.0 °C
15.02.2014
Menu 14:15
20 ba77076e06 05/2021
MU 6100 L Operation
Automatic shut-off The instrument has an automatic shut-off function in order to save the batteries
function (see section 11.2.1 SYSTEM, page 71). The automatic shut-off function switches
off the meter if no key is pressed for an adjustable period.
The automatic shut-off function is not active
if the communication cable is connected
if the power pack is connected
if the Automatic data storage function is active, or with automatic data trans-
mission
Display illumination The meter automatically switches off the display illumination if no key is
pressed for 30 seconds. The illumination is switched on with the next keystroke
again.
You can also generally switch the display illumination on or off (see section
11.2.1 SYSTEM, page 71).
5.4 Navigation
The principles of navigation in menus and dialogs are explained in the following
sections.
Operating Explanation
mode
Measuring The measurement data of the connected sensor are shown
in the measured value display
Calibration The course of a calibration with calibration information, func-
tions and settings is displayed
Storing in The meter stores measuring data automatically or manually
memory
Transmit- The meter transmits measuring data and calibration records
ting data to a USB-B interface automatically or manually.
Setting The system menu or a sensor menu with submenus, set-
tings and functions is displayed
Only those displays and functions are available in the active operating mode
that are currently being required.
ba77076e06 05/2021 21
Operation MU 6100 L
In the measured value display, open the setting menus with <OK>. The current
functions of the softkeys are shown on the display.
Use <OK> (short pressure) to open the menu for calibration and measure-
ment settings for the displayed measured parameter.
Use <OK _
> (long keystroke (approx. 2 s) to open the Storage & config
menu with the sensor-independent settings.
Use the keys of the keypad to carry out further functions such as storage or cali-
bration (see section 5.1.1 KEYPAD, page 18). These functions are not available
in other operating situations.
The menus for settings and dialogs in procedures contain further subelements.
To select a subelement, use the < >< > keys. The current selection is
displayed with a frame.
To make further settings, switch to the next higher menu level with
<F1>[Back].
Use <MODE> to return to the measured value display.
Submenus
The name of the submenu is displayed at the upper edge of the frame. Sub-
menus are opened by confirming with <OK>. Example:
System
General
Interface
Clock
Service information
Reset
15.02.2014
Back
14:15
Settings
Settings are indicated by a colon. The current setting is displayed on the
right-hand side. The setting mode is opened with <OK>. Subsequently, the
setting can be changed with < >< > and <OK>. Example:
22 ba77076e06 05/2021
MU 6100 L Operation
General
Language: Deutsch
Beep: Off
Illumination: On
Contrast: 50 %
Switchoff time: 1h
Temperature unit: °C
Stability control: On
15.02.2014
Back
14:15
Functions
Functions are designated by the name of the function. They are immediately
carried out by confirming with <OK>.
Example: Display the Calibration record function.
pH
Calibration record
Calibration data storage
Buffer: TEC
One point calibration: Yes
Calibration interval: 7d
Unit for slope: mV/pH
Messages
Information is marked by the i symbol. It cannot be selected. Example:
pH
Calibration record
Calibration data storage
Buffer: TEC
One point calibration: Yes
Calibration interval: 7d
Unit for slope: mV/pH
ba77076e06 05/2021 23
Operation MU 6100 L
pH
6.949
25.0 °C
15.02.2014
Menu
14:15
15.02.2014
Back
14:15
System
General
Interface
Clock
Service information
Reset
Back 15.02.2014
14:15
24 ba77076e06 05/2021
MU 6100 L Operation
General
Language: Deutsch
Beep: Off
Illumination: On
Contrast: 50 %
Switchoff time: 1h
Temperature unit: °C
Stability control: On
15.02.2014
Back
14:15
General
Language: Deutsch
Beep: Off
Illumination: On
Contrast: 50 %
Switchoff time: 1h
Temperature unit: °C
Stability control: On
15.02.2014
Back
14:15
The meter has a clock with a date function. The date and time are shown in the
measured value display.
When storing measured values and calibrating, the current date and time are
automatically stored as well.
The correct setting of the date and time and date format is important for the fol-
lowing functions and displays:
Current date and time
Calibration date
Identification of stored measured values.
After a fall of the supply voltage (empty batteries), the date and time
are reset.
ba77076e06 05/2021 25
Operation MU 6100 L
The date format can be switched from the display of day, month, year
(dd.mm.yyyy) to the display of month, day, year (mm/dd/yyyy or mm.dd.yyyy).
Clock
Date format: dd.mm.yyyy
Date: 15.02.2014
Time: 14:15:25
15.02.2014
Back
14:15
4. Change and confirm the setting with < >< > and <OK>.
The minutes are highlighted.
5. Change and confirm the setting with < >< > and <OK>.
The seconds are highlighted.
6. Change and confirm the setting with < >< > and <OK>.
The time is set.
7. If necessary, set the Date and Date format. The setting is made similarly
to that of the time.
8. To make further settings, switch to the next higher menu level with
[Back]<F1>.
or
Switch to the measured value display with <MODE>.
The instrument is in the measuring mode.
26 ba77076e06 05/2021
MU 6100 L Operation
Channel display
Display of the plug position
pH for the respective parameter
The filled bar indicates for
6.949 each connected sensor to
which plug position of the
24.8 °C meter it is connected.
15.02.2014
14:15 USB output
The measured values of the connected sensors can be displayed in the follow-
ing ways:
Clear display of all connected sensors
Detailed display of one sensor
With the softkey you can very easily switch between the two display types. The
suitable softkey is displayed depending on the operating situation.
pH 6.966 pH
25.0 °C 6.966
ϰ 0.840 1/cm 25.0 °C
25.0 °C
15.02.2014 15.02.2014
USB output USB output
14:15 14:15
ba77076e06 05/2021 27
pH value MU 6100 L
6 pH value
6.1 Measuring
NOTE
When connecting an grounded PC/printer, measurements cannot be
performed in grounded media as the values would be incorrect.
The USB interface is not galvanically isolated.
pH
6.949
24.8 °C
15.02.2014
Menu 14:15
Stability control The stability control function (AutoRead) continually checks the stability of the
(AutoRead ) measurement signal. The stability has a considerable impact on the reproduc-
ibility of measured values (see section 11.2.3 AUTOMATIC STABILITY CONTROL,
page 72).
Criteria for a stable The Stability control function checks whether the measured values are stable
measured value
28 ba77076e06 05/2021
MU 6100 L pH value
- 1 °C Manual
ba77076e06 05/2021 29
pH value MU 6100 L
6.2 Calibration
pH electrodes age. This changes the zero point (asymmetry) and slope of the
pH electrode. As a result, an inexact measured value is displayed. Calibration
determines and stores the current values of the zero point and slope of the
electrode.
Thus, you should calibrate at regular intervals.
Make sure that in the sensor menu, Buffer menu, the buffer set is correctly
selected (see 11.1.1 SETTINGS FOR PH MEASUREMENTS, PAGE 61).
Use any one to five buffer solutions of the selected buffer set in ascending or
descending order.
Below, calibration with Technical buffers (TEC) is described. When other buffer
sets are used, other nominal buffer values are displayed. Apart from that, the
procedure is identical.
30 ba77076e06 05/2021
MU 6100 L pH value
pH Buffer 1
-180.0 mV
24.8 °C
AutoCal TEC
15.02.2014
14:15
pH Buffer 10.011
-180.0 mV
24.8 °C
AutoCal TEC
AR
15.02.2014
14:15
9. Wait for the end of the measurement with stability control or accept the
calibration value with <OK>.
The calibration display for the next buffer appears (voltage display).
10. If necessary, finish the calibration procedure as a single-point calibration
with <MODE>.
The calibration record is displayed.
ba77076e06 05/2021 31
pH value MU 6100 L
pH Buffer 7.000
0.0 mV
24.8 °C
AutoCal TEC
AR
15.02.2014
14:15
15. Wait for the measurement with stability control to be completed or termi-
nate the stability control and take over the calibration value with <OK>.
The calibration display for the next buffer appears (voltage display).
16. If necessary, finish the calibration procedure as a two-point calibration
with <MODE>.
The calibration record is displayed.
Continuing with
17. Thoroughly rinse the electrode with deionized water.
three- to five-point
calibration 18. Immerse the electrode in the next buffer solution.
19. When measuring without temperature sensor:
– Temper the buffer solution, or measure the current temperature.
– Enter the temperature value in the menu.
20. Start the measurement with <OK>.
The measured value is checked for stability (stability control). The [AR]
status indicator is displayed. A progress bar is displayed and the display
of the measured parameter flashes.
pH Buffer 4.010
180.0 mV
24.8 °C
AutoCal TEC
AR
15.02.2014
14:15
32 ba77076e06 05/2021
MU 6100 L pH value
21. Wait for the measurement with stability control to be completed or termi-
nate the stability control and take over the calibration value with <OK>.
The calibration display for the next buffer appears (voltage display).
22. If necessary, use <MODE> to finish calibration or
Continue calibrating using the next buffer with <OK>.
Make sure that in the sensor menu, Buffer menu, the AnyCal buffer set is
correctly selected (see section 11.1.1 SETTINGS FOR PH MEASUREMENTS, page
61).
Use any one to five buffer solutions in ascending or descending order.
pH Buffer 1
0.0 mV
24.8 °C
AnyCal
15.02.2014
14:15
ba77076e06 05/2021 33
pH value MU 6100 L
pH Buffer 7.000
0.0 mV
24.8 °C
AnyCal
AR
15.02.2014
14:15
9. Wait for the end of the measurement with stability control or accept the
calibration value with <OK>.
The calibration display for the setting of the nominal buffer value appears.
pH Buffer 1
7.000 V
W
24.8 °C
AnyCal
15.02.2014
14:15
10. Set the nominal buffer value for the measured temperature with
< >< >.
11. Accept the set calibration value with <OK>.
The calibration display for the next buffer appears (voltage display).
12. If necessary, finish the calibration procedure as a single-point calibration
with <MODE>.
The calibration record is displayed.
34 ba77076e06 05/2021
MU 6100 L pH value
pH Buffer 2
4.035 V
W
24.8 °C
AnyCal
15.02.2014
14:15
18. Set the nominal buffer value for the measured temperature with
< >< >.
19. Accept the set calibration value with <OK>.
The calibration display for the next buffer appears (voltage display).
20. Finish the calibration procedure as a two-point calibration with <MODE>.
The calibration record is displayed.
Continuing with
21. Thoroughly rinse the electrode with deionized water.
three- to five-point
calibration 22. Immerse the electrode in the next buffer solution.
23. When measuring without temperature sensor:
– Temper the buffer solution, or measure the current temperature.
– Enter the temperature value in the menu.
24. Start the measurement with <OK>.
The measured value is checked for stability (stability control). The [AR]
status indicator is displayed. A progress bar is displayed and the display
of the measured parameter flashes.
25. Wait for the measurement with stability control to be completed or termi-
nate the stability control and take over the calibration value with <OK>.
The calibration display for the setting of the nominal buffer value appears.
ba77076e06 05/2021 35
pH value MU 6100 L
pH Buffer 3
9.958 V
W
24.8 °C
AnyCal
15.02.2014
14:15
26. Set the nominal buffer value for the measured temperature with
< >< >.
27. Accept the set calibration value with <OK>.
The calibration display for the next buffer appears (voltage display).
28. Use <MODE> to finish calibration or
Continue calibrating using the next buffer with <OK>.
After the fifth buffer has been measured the calibration is automat-
ically finished. Then the calibration record is displayed.
The calibration line is determined by linear regression.
Calibration can be performed using one to five buffer solutions in any order
(single-point to five-point calibration). The meter determines the following
values and calculates the calibration line as follows:
You can display the slope in the units, mV/pH or % (see section
11.1.1 SETTINGS FOR PH MEASUREMENTS, page 61).
36 ba77076e06 05/2021
MU 6100 L pH value
Displays the calibra- The calibration record of the last calibration is to be found under the menu item,
tion data <OK> / Calibration / Calibration record. To open it, press the <CAL__> key in
the measured value display.
Subsequently, you can transmit the displayed calibration data to the interface,
e.g. to a PC, with the <F2>/[USB output] key.
Displaying the cali- The calibration records of the last calibrations are to be found in the menu,
bration data memory <OK> / Calibration / Calibration data storage.
Calibration evalua- After calibrating, the meter automatically evaluates the calibration. The zero
tion point and slope are evaluated separately. The worse evaluation of both is taken
into account. The evaluation appears on the display and in the calibration
ba77076e06 05/2021 37
pH value MU 6100 L
record.
Calibration record
MU 6100 L
(example) Ser. no. 11292113
CALIBRATIONpH
15.02.2014 15:55
AutoCal TEC
Buffer 1 4.01
Buffer 2 7.00
Buffer 3 10.01
Voltage 1 184.0 mV
Voltage 2 3.0 mV
Voltage 3 -177.0 mV
Temperature 1 24.0 °C
Temperature 2 24.0 °C
Temperature 3 24.0 °C
Slope -60.2 mV/pH
Asymmetry 4.0 mV
Sensor +++
etc...
38 ba77076e06 05/2021
MU 6100 L ORP voltage
7 ORP voltage
7.1 Measuring
NOTE
When connecting an grounded PC/printer, measurements cannot be
performed in grounded media as the values would be incorrect.
The USB interface is not galvanically isolated.
U
157.0 mV
24.8 °C
15.02.2014
Menu 14:15
Stability control The stability control function (AutoRead) continually checks the stability of the
(AutoRead ) measurement signal. The stability has a considerable impact on the reproduc-
ibility of measured values (see section 11.2.3 AUTOMATIC STABILITY CONTROL,
page 72).
Criteria for a stable The Stability control function checks whether the measured values are stable
measured value within the monitored time interval.
ba77076e06 05/2021 39
ORP voltage MU 6100 L
Freezes the mea- With the HOLD function, you can freeze the current measured value. The dis-
sured value (HOLD played measured value stops changing until you switch the HOLD function off.
function)
1. Freeze the measured value with <HOLD>.
The [HOLD] status indicator is displayed.
2. Release the frozen measured value again with <HOLD> or <MODE>.
The HOLD function is switched off.
The [HOLD] status display disappears.
- 1 °C Manual
ORP electrodes are not calibrated. You can, however, check ORP
electrodes by measuring the ORP of a test solution and comparing
the value with the nominal value.
40 ba77076e06 05/2021
MU 6100 L Ion concentration
8 Ion concentration
8.1 Measuring
NOTE
When connecting an grounded PC/printer, measurements cannot be
performed in grounded media as the values would be incorrect.
The USB interface is not galvanically isolated.
ISE
0.157 mg/l
25.0 °C
Ag
15.02.2014
Menu 12:12
Stability control The stability control function (AutoRead) continually checks the stability of the
(AutoRead ) measurement signal. The stability has a considerable impact on the reproduc-
ibility of measured values (see section 11.2.3 AUTOMATIC STABILITY CONTROL,
ba77076e06 05/2021 41
Ion concentration MU 6100 L
page 72).
Criteria The AutoRead criteria affect the reproducibility of the measured values. The
following criteria can be adjusted:
high: highest reproducibility
medium: medium reproducibility
low: lowest reproducibility
Freezes the mea- With the HOLD function, you can freeze the current measured value. The dis-
sured value (HOLD played measured value stops changing until you switch the HOLD function off.
function)
1. Freeze the measured value with <HOLD>.
The [HOLD] status indicator is displayed.
2. Release the frozen measured value again with <HOLD>.
The HOLD function is switched off.
The [HOLD] status display disappears.
- 1 °C Manual
42 ba77076e06 05/2021
MU 6100 L Ion concentration
8.2 Calibration
ba77076e06 05/2021 43
Ion concentration MU 6100 L
ISE Standard 1
0.0 mV
24.8 °C
Ag 0.010 mg/l
15.02.2014
12:12
0.0 mV
24.8 °C
AR
15.02.2014
12:12
11. Wait for the end of the AutoRead measurement or accept the calibration
value with <OK>.
The calibration display for the next standard solution appears.
44 ba77076e06 05/2021
MU 6100 L Ion concentration
ISE Standard 2
0.0 mV
24.8 °C
Ag 0.020 mg/l
15.02.2014
12:12
8.9 mV
24.8 °C
AR
15.02.2014
12:12
16. Wait for the end of the AutoRead measurement or accept the calibra-
tion value with <OK>.
The calibration display for the next standard solution appears.
ISE Standard 3
20.5 mV
24.8 °C
Ag 0.050 mg/l
15.02.2014
12:12
ba77076e06 05/2021 45
Ion concentration MU 6100 L
Continuing with Repeat the steps 12 to 17 in the same way with the third and further standard
three- to seven-point solutions as necessary. The new calibration values are displayed after the last
calibration calibration step was completed.
Use two to seven different standard solutions. The standard solutions have to
be selected in either increasing or decreasing order.
Select the unit of the standard solution and measurement result in
the ISE setup/Unit menu.
Displays the calibra- The calibration record of the last calibration is to be found under the menu item,
tion data <OK> / Calibration / Calibration record. To open it, press the <CAL__> key in
the measured value display.
46 ba77076e06 05/2021
MU 6100 L Ion concentration
Subsequently, you can transmit the displayed calibration data to the interface,
e.g. to a PC, with the <F2>/[USB output] key.
Displaying the cali- The calibration records of the last calibrations are available in the menu, <OK>
bration data memory / Calibration / Calibration data storage.
Calibration evalua- After calibrating, the meter automatically evaluates the calibration.
tion
Display Calibration Magnitude of the slope [mV]
record
+++ 50.0 ... 70.0 or 25.0 ... 35.0
CALIBRATIONISE
18.01.2013 08:09:10
ba77076e06 05/2021 47
Ion concentration MU 6100 L
48 ba77076e06 05/2021
MU 6100 L Dissolved oxygen
9 Dissolved oxygen
9.1 Measuring
NOTE
When connecting an grounded PC/printer, measurements cannot be
performed in grounded media as the values would be incorrect.
The USB interface is not galvanically isolated.
O2
7.9 mg/l
25.1 °C
[OXY 11]
15.02.2014
USB output
08:00
Salinity correction When measuring the concentration of solutions with a salt content of more than
1 g/l, a salinity correction is required. For this, you have to measure and input
the salinity of the measured medium first. When the salinity correction is
switched on, the [SAL] indicator is displayed in the measuring window.
You can switch the salinity correction on or off and enter the salinity
in the menu for calibration and measurement settings (see section
11.1.5 SETTINGS FOR D.O. SENSORS, page 65).
Stability control The stability control function (AutoRead) continually checks the stability of the
(AutoRead ) measurement signal. The stability has a considerable impact on the reproduc-
ibility of measured values (see section 11.2.3 AUTOMATIC STABILITY CONTROL,
ba77076e06 05/2021 49
Dissolved oxygen MU 6100 L
page 72).
Criteria for a stable The Stability control function checks whether the measured values are stable
measured value within the monitored time interval.
The minimum duration until a measured value is assessed as stable is the mon-
itored time interval. The actual duration is mostly longer.
Freezes the mea- With the HOLD function, you can freeze the current measured value. The dis-
sured value (HOLD played measured value stops changing until you switch the HOLD function off.
function)
1. Freeze the measured value with <HOLD>.
The [HOLD] status indicator is displayed.
2. Release the frozen measured value again with <HOLD>.
The HOLD function is switched off.
The [HOLD] status display disappears.
50 ba77076e06 05/2021
MU 6100 L Dissolved oxygen
9.2 Calibration
D.O. sensors age. This changes the slope of the D.O. sensor. Calibration deter-
mines the current slope of the sensor and stores this value in the instrument.
For this calibration procedure, the Comparison meas. setting must be set to Off
in the Calibration menu.
Proceed as follows to calibrate the instrument:
The sponge in the air calibration vessel must be moist (not wet).
Leave the sensor in the air calibration vessel for a time long enough
to adjust.
ba77076e06 05/2021 51
Dissolved oxygen MU 6100 L
O2
0.88
25.1 °C
OxiCal
15.02.2014
08:00
For this calibration procedure, the Comparison meas. setting must be set to On
in the Calibration menu.
O2
7.9 mg/l
25.1 °C
OxiComp
15.02.2014
08:00
52 ba77076e06 05/2021
MU 6100 L Dissolved oxygen
O2
7.9 mg/l
0.973 V
W
OxiComp
15.02.2014
08:00
6. Using <V > <W >, set the correction factor to adjust the displayed con-
centration value to the nominal value (value of the comparison mea-
surement). Subsequently, accept the correction factor with <OK>.
The meter switches to the measured value display.
The status display [Factor] is active.
ba77076e06 05/2021 53
Dissolved oxygen MU 6100 L
Displays the calibra- The calibration record of the last calibration is to be found under the menu item,
tion data <OK> / Calibration / / Calibration record. To open it, press the <CAL__> key
in the measured value display.
Subsequently, you can transmit the displayed calibration data to the interface,
e.g. to a PC, with the <F2>[USB output] key.
Displaying the cali- The calibration records of the last calibration procedures are available in the
bration data memory menus, <OK>/Calibration / / Calibration data storage and <OK>/Storage &
config / Data storage/ / Calibration data storage.
or
Data storage /
Calibration data stor-
age / Output to USB
Calibration evalua- After calibration, the meter automatically evaluates the current status of the
tion calibration. The evaluation appears on the display and in the calibration record.
54 ba77076e06 05/2021
MU 6100 L Dissolved oxygen
Calibration record
MU 6100 L
(example) Ser. no. 11292113
CALIBRATION OXY 11
15.02.2014 16:13:33
S = 0.88 25.0 °C
Sensor +++
ba77076e06 05/2021 55
Conductivity MU 6100 L
10 Conductivity
10.1 Measuring
NOTE
When connecting an grounded PC/printer, measurements cannot be
performed in grounded media as the values would be incorrect.
The USB interface is not galvanically isolated.
The selection of the measuring cell and the setting of the cell con-
stant is done in the measurement settings menu for conductivity
(see section 11.1.7 SETTINGS FOR CONDUCTIVITY MEASURING CELLS,
page 67). The cell constant to be set must either be taken from the
operating manual of the measuring cell or is printed on the measur-
ing cell.
ϰ
1417 µS/cm
25.0 °C
[Tr25][nLF][ 0.840 1/cm]
15.02.2014
USB output
08:00
56 ba77076e06 05/2021
MU 6100 L Conductivity
The factor to calculate the total dissolved solids is set to 1.00 in the
factory. You can adjust this factor to meet your requirements in the
range 0.40 ... 1.00. The factor is set in the Measurement menu for
the parameter, TDS.
Stability control The stability control function (AutoRead) continually checks the stability of the
(AutoRead ) measurement signal. The stability has a considerable impact on the reproduc-
ibility of measured values (see section 11.2.3 AUTOMATIC STABILITY CONTROL,
page 72).
Criteria for a stable The Stability control function checks whether the measured values are stable
measured value within the monitored time interval.
The minimum duration until a measured value is assessed as stable is the mon-
itored time interval. The actual duration is mostly longer.
Freezes the mea- With the HOLD function, you can freeze the current measured value. The dis-
sured value (HOLD played measured value stops changing until you switch the HOLD function off.
function)
1. Freeze the measured value with <HOLD>.
The [HOLD] status indicator is displayed.
2. Release the frozen measured value again with <HOLD>.
The HOLD function is switched off.
The [HOLD] status display disappears.
ba77076e06 05/2021 57
Conductivity MU 6100 L
10.2 Calibration
Aging slightly changes the cell constant, e.g. due to coatings. As a result, an
inexact measured value is displayed. The original characteristics of the cell can
often be restored by cleaning the cell. Calibration determines the current value
of the cell constant and stores this value in the meter.
Thus, you should calibrate at regular intervals (we recommend: every 6
months).
You can determine the actual cell constant of the conductivity measuring cell
by calibrating with the control standard in the following ranges:
In the default condition, the calibrated cell constant of the meter is set to
0.840 cm-1 (conductivity measuring cell CO 11).
For this calibration procedure, the Measuring cell setting must be set to cal in
the Measurement menu. Proceed as follows to determine the cell constant:
58 ba77076e06 05/2021
MU 6100 L Conductivity
ϰ
0.839 1/cm
25.1 °C
15.02.2014
08:00
Displays the calibra- The calibration record of the last calibration is to be found under the menu item,
tion data <OK> / Calibration / Calibration record. To open it, press the <CAL__> key in
the measured value display.
Subsequently, you can transmit the displayed calibration data to the interface,
e.g. to a PC, with the <F2>[USB output] key.
Displaying the cali- The calibration records of the last calibration procedures are available in the
bration data memory <OK>/Calibration / Calibration data storage and <OK>/Storage & config /
Data storage / Calibration data storage menu.
ba77076e06 05/2021 59
Conductivity MU 6100 L
or
Data storage /
Calibration data stor-
age / Output to USB
Calibration evalua- After calibration, the meter automatically evaluates the current status of the cal-
tion ibration. The evaluation appears on the display and in the calibration record.
Calibration record
MU 6100 L
(example) Ser. no. 11292113
CALIBRATION Cond
15.02.2014 16:13:33
60 ba77076e06 05/2021
MU 6100 L Settings
11 Settings
11.1 Measurement settings
The settings for pH measurements are made in the menu for calibration and
measurement settings of the pH/ORP measurement. To open the settings, dis-
play the required measured parameter in the measured value display and
press the <OK> key. After completing the settings, switch to the measured
value display with <MODE>.
In the following table, only those settings are listed that concern the pH mea-
surement.
Default settings are printed in bold.
ba77076e06 05/2021 61
Settings MU 6100 L
You can use the buffer sets quoted in the table for an automatic calibration. The
pH values are valid for the specified temperature values. The temperature
dependence of the pH values is taken into consideration during calibration.
62 ba77076e06 05/2021
MU 6100 L Settings
2 TEC US
Technical buffers US
4 NIST/DIN 1.679 25 °C
DIN buffers according to DIN 4.006
19266 and NIST Traceable 6.865
Buffers 9.180
12.454
5 TEC 2 4.010 25 °C
Technical buffers 7.000
10.011
10 Merck 5 * 4.010 25 °C
7.000
10.000
* Brand names or trade names are trademarks of their respective owners protected
by law.
ba77076e06 05/2021 63
Settings MU 6100 L
Setting the calibra- The calibration interval is set to 7 days in the factory.
tion interval You can change the interval (1 ... 999 days):
The settings for ORP measurements are made in the menu for calibration and
measurement settings of the pH/ORP measurement. To open the settings, dis-
play the required measured parameter in the measured value display and
press the <OK>key. After completing the settings, switch to the measured
value display with <MODE>.
In the following table, only those settings are listed that influence the ORP mea-
surement.
Default settings are printed in bold.
64 ba77076e06 05/2021
MU 6100 L Settings
The settings are available in the menu for measurement and calibration set-
tings. To open the settings, display the required measured parameter in the
measured value display and press the <OK> key. After completing the set-
tings, switch to the measured value display with <MODE>.
Default settings are printed in bold.
ba77076e06 05/2021 65
Settings MU 6100 L
Setting the calibra- The calibration interval is set to 14 days in the factory.
tion interval You can change the interval (1 ... 999 days):
66 ba77076e06 05/2021
MU 6100 L Settings
The settings are made in the Measurement menu for the measured parameter,
conductivity. To open the settings, display the required measured parameter in
the measured value display and press the <OK> key. After completing the set-
tings, switch to the measured value display with <MODE>.
Default settings are printed in bold.
ba77076e06 05/2021 67
Settings MU 6100 L
68 ba77076e06 05/2021
MU 6100 L Settings
Setting the calibra- The calibration interval is set to 150 days in the factory.
tion interval You can change the interval (1 ... 999 days):
ba77076e06 05/2021 69
Settings MU 6100 L
Application tips Set the temperature compensation suitable for your test sample:
The factor to calculate the total dissolved solids is set to 1.00 in the delivery
condition.
You can adjust this factor to meet your requirements in the range 0.40 ... 1.00.
The factor is set in the menu for the parameter TDS (Measurement / TDS fac-
tor).
70 ba77076e06 05/2021
MU 6100 L Settings
11.2.1 System
To open the Storage & config menu, press the <OK _ > key in the measured
value display. After completing the settings, switch to the measured value dis-
play with <MODE>.
Default settings are printed in bold.
ba77076e06 05/2021 71
Settings MU 6100 L
This menu contains all functions to display, edit and erase stored measured
values and calibration records (see section 12 DATA MEMORY, page 75).
You can activate or switch off the automatic Stability control function (see sec-
tion 11.2.1 SYSTEM, page 71).
11.3 Reset
You can reset (initialize) all sensor settings and sensor-independent settings
separately from each other.
The calibration data are reset to the default settings together with
the measuring parameters. Recalibrate after performing a reset.
The following settings for pH measurements are reset to the default settings
72 ba77076e06 05/2021
MU 6100 L Settings
The sensor settings are reset under the Reset menu item in the menu for cali-
bration and measurement settings. To open it in the measured value display,
press the <OK> (or <F1>/[Menu]) key.
Oxi
The calibration data are reset to the default settings together with
the measuring parameters. Recalibrate after performing a reset.
The following settings for D.O. measurements are reset to the default settings
with the Reset function:
The sensor settings are reset under the Reset menu item in the menu for cali-
bration and measurement settings. To open it in the measured value display,
press the <OK> (or <F1>/[Menu]) key.
Cond The following settings for conductivity measurements are reset to the default
ba77076e06 05/2021 73
Settings MU 6100 L
The sensor settings are reset under the Reset menu item in the menu for cali-
bration and measurement settings. To open it in the measured value display,
press the <OK> (or <F1>/[Menu]) key.
The system settings are reset in the menu, Storage & config / System / Reset.
To open the Storage & config menu in the measured value display, press the
<OK _ > (or <F1__>/[Menu]) key.
74 ba77076e06 05/2021
MU 6100 L Data memory
12 Data memory
You can transmit measured values (datasets) to the data memory:
Manual data memory (see section 12.1 MANUAL STORAGE, page 75)
Automatic storage at intervals (see section 12.2 AUTOMATIC DATA STORAGE
AT INTERVALS, page 76)
With each data storage process, the current datasets of the sen-
sors indicated on the display are transmitted to the interface at the
same time.
15.02.2014 11:24:16
pH 7.0 24.8 °C AR +++
O2 7.80 mg/l 24.8 °C AR +++
ID number: 1
Continue
15.02.2014
Back
14:15
ba77076e06 05/2021 75
Data memory MU 6100 L
If the memory is full The following window appears if all 200 storage locations are occupied:
Warning
Yes
No
Back 15.02.2014
14:15
Configuring the
automatic memory
1. Press the <STR _
> key.
The menu for automatic data storing appears.
function
ID number 1
1
Interval 30 s
Duration 180 min
Continue
0d03h00min 2
0 1d17h33m
in 15.02.2014
Back 14:15
3
76 ba77076e06 05/2021
MU 6100 L Data memory
Settings You can configure the automatic data storing function with the following
settings:
Starting the auto- To start the automatic storing function, select Continue with < >< > and
matic storing func- confirm with <OK>. The meter switches to the measured value display.
tion
pH 0 14
1
6.949
25.0 °C
0d03h00min
2
15.02.2014
Eco mode 14:15
The active automatic storage function can be recognized by the progress bar
in the function display. The progress bar indicates the remaining storage dura-
tion.
Energy saving mode If the automatic storing function is active, the meter provides an energy saving
([Eco mode]) mode ([Eco mode]) to avoid unnecessary energy consumption. The energy
ba77076e06 05/2021 77
Data memory MU 6100 L
saving mode switches off functions of the meter that are not required for the
automatic storage of measurement data (such as the display). By pressing any
key the energy saving mode is switched off again.
Terminating the Proceed as follows to switch off the automatic data storing function before the
automatic memory adjusted storing duration has expired:
function prema-
turely
1. Press the <STR _ > key.
The following window appears.
Warning
Yes
No
Back 15.02.2014
14:15
2. Using < >< >, select Yes and confirm with <OK>.
The meter switches to the measured value display.
The automatic data storing function is terminated.
The settings are explained here using the manual data memory as
an example. The same settings and functions are available for the
automatic data memory.
78 ba77076e06 05/2021
MU 6100 L Data memory
15.02.2014
Back
14:15
15.02.2014 10:56:20
MU 6100 L
Ser. no. 08502113
ID number 2
Quitting the display To quit the display of stored measurement datasets, you have the following
ba77076e06 05/2021 79
Data memory MU 6100 L
options:
Switch directly to the measured value display with <MODE>.
Quit the display and move to the next higher menu level with <F1>/[Back].
Erasing the measurement data memory (see section 12.3.1 EDITING THE
MEASUREMENT DATA MEMORY, page 78).
The MU 6100 L meter has two measurement data memories. The measured
values recorded either manually or automatic are stored separately in indi-
vidual measurement data memories.
80 ba77076e06 05/2021
MU 6100 L Transmitting data (USB interface)
The following rule applies: With the exception of the menus, shortly
pressing the <F2>[USB output] key generally outputs the display
contents to the interface (displayed measured values, measure-
ment datasets, calibration records).
ba77076e06 05/2021 81
Transmitting data (USB interface) MU 6100 L
13.2 Connecting a PC
Connect the MU 6100 L to the PC via the USB interface.
NOTE
When connecting an grounded PC/printer, measurements cannot be
performed in grounded media as the values would be incorrect.
The USB interface is not galvanically isolated.
Installation of the System requirements of the PC for installation of the USB driver:
USB driver on the
PC
1. Insert the supplied installation CD in the CD drive of your PC.
2. Install the driver from the CD.
Follow the Windows installation instructions as necessary.
3. Connect the MU 6100 L to the PC via the USB interface.
The meter is listed as a virtual COM interface among the connections in
the Windows instrument manager.
82 ba77076e06 05/2021
MU 6100 L Maintenance, cleaning, disposal, accessories
1 Battery compartment
CAUTION
Make sure that the poles of the batteries are positioned cor-
rectly.
The ± signs on the batteries must correspond to the ± signs in
the battery compartment.
You can operate the meter either with normal batteries or with
rechargeable batteries (Ni-MH). In order to charge the batteries, an
external charging device is required.
ba77076e06 05/2021 83
Maintenance, cleaning, disposal, accessories MU 6100 L
14.2 Cleaning
Occasionally wipe the outside of the measuring instrument with a damp, lint-
free cloth. Disinfect the housing with isopropanol as required.
CAUTION
The housing is made of synthetic material (ABS). Thus, avoid
contact with acetone or similar detergents that contain sol-
vents. Remove any splashes immediately.
14.3 Packing
This meter is sent out in a protective transport packing.
We recommend: Keep the packing material. The original packing protects the
meter against damage during transport.
14.4 Disposal
This equipment is marked with the crossed out wheeled bin symbol to indicate
that this equipment must not be disposed of with unsorted waste.
Instead it's your responsibility to correctly dispose of your equipment at lifecycle
-end by handling it over to an authorized facility for separate collection and
recycling. It's also your responsibility to decontaminate the equipment in case
of biological, chemical and/or radiological contamination, so as to protect from
health hazards the persons involved in the disposal and recycling of the equip-
ment.
84 ba77076e06 05/2021
MU 6100 L Maintenance, cleaning, disposal, accessories
For more information about where you can drop off your waste of equipment,
please contact your local dealer from whom you originally purchased this
equipment.
By doing so, you will help to conserve natural and environmental resources and
you will ensure that your equipment is recycled in a manner that protects
human health.
Thank you
14.5 Accessories
ba77076e06 05/2021 85
Maintenance, cleaning, disposal, accessories MU 6100 L
14.5.2 pH / ORP
86 ba77076e06 05/2021
MU 6100 L Maintenance, cleaning, disposal, accessories
ba77076e06 05/2021 87
Maintenance, cleaning, disposal, accessories MU 6100 L
Solutions EU NA
Buffer pH 4 AVS TITRINORM, 100 ml 32095.184 -
Buffer pH 7 AVS TITRINORM, 100 ml 32096.187 -
Buffer pH 10 AVS TITRINORM, 100 ml 32040.185 -
Buffer NIST pH 4.01, 30 x 30 ml 1.99001.0001 -
Buffer NIST pH 7, 30 x 30 ml 1.99002.0001 -
Buffer NIST pH 10, 30 x 30 ml 1.99004.0001 -
Buffer pH 1.68, 500 mL - BDH5006-500ML
Buffer pH 4, 500 mL - BDH5024-500ML
Buffer pH 7, 500 mL - BDH5052-500ML
Buffer pH 10, 500 mL - BDH5078-500ML
Buffer pH 12.45, 500 mL - BDH5096-500ML
Buffer NIST pH 4.006, 500 mL - BDH5018-500ML
Buffer NIST pH 6.865, 500 mL - BDH5040-500ML
Buffer NIST pH 9.18, 500 mL - BDH5066-500ML
Buffer NIST pH 10.012, 500 mL - BDH5072-500ML
Storage Solution (3 moles/l KCl), 100 ml 83605.180 BDH7296-0
Cleaning Solution Pepsine/Hydrochloric 83603.180 89207-612
acid, 100 ml
14.5.4 Conductivity
Order no. (Catalogue No.)
Measuring cells EU NA
pHenomenal CO 11 663-0147 76470-822
CONDUCTIVITY SENSOR PHENOMENAL
1,5M 8PIN
Solutions EU NA
KCl 0.01 mol/L: 1.413 mS/cm, 100 mL 83607.180 -
KCl 0.01 mol/L: 1.413 mS/cm, 500 mL 83607.290 89236-544
KCl 0.1 mol/L: 12.8 mS/cm, 500 mL 83608.260 89236-546
88 ba77076e06 05/2021
MU 6100 L Maintenance, cleaning, disposal, accessories
14.5.5 D.O.
Order no. (Catalogue No.)
Measuring cells EU NA
pHenomenal OXY 11 664-0042 76470-820
OXYGEN SENSOR PHENOMENAL 3M 8
PIN
Accessories EU NA
MAINTENANCE KIT OXYGEN 664-0049 76460-466
ba77076e06 05/2021 89
What to do if... MU 6100 L
15 What to do if...
15.1 pH/ORP
90 ba77076e06 05/2021
MU 6100 L What to do if...
Test sample:
– pH value not stable – Measure with air excluded if
necessary
– Temperature not stable – Temper if necessary
ba77076e06 05/2021 91
What to do if... MU 6100 L
92 ba77076e06 05/2021
MU 6100 L What to do if...
15.3 Conductivity
ba77076e06 05/2021 93
What to do if... MU 6100 L
94 ba77076e06 05/2021
MU 6100 L Firmware update
16 Firmware update
Available firmware updates are provided on the Internet.
With the firmware update program and a PC you can update the firmware of the
MU 6100 L to the newest version.
For the update you have to connect the meter to a PC.
For the update via the USB interface, the following is required:
a free USB interface (virtual COM port) on the PC
the driver for the USB interface (on the enclosed CD-ROM)
the USB cable (included in the scope of delivery of the MU 6100 L).
After switching the meter off and on you can check whether the meter has taken
over the new software version (see YOU WANT TO KNOW WHICH SOFTWARE
VERSION IS IN THE METER, PAGE 94).
ba77076e06 05/2021 95
Glossary MU 6100 L
17 Glossary
pH/ORP/ISE
Dissolved oxygen
D.O. partial pressure Pressure caused by the oxygen in a gas mixture or liquid.
D.O. saturation Short name for the relative D.O. saturation.
OxiCal® Name for a procedure to calibrate D.O. measuring systems in water
vapor-saturated air.
Salinity The absolute salinity SA of seawater corresponds to the relationship
of the mass of dissolved salts to the mass of the solution (in g/Kg). In
practice, this dimension cannot be measured directly. Therefore, the
practical salinity according to IOT is used for oceanographic monitor-
ing. It is determined by measuring the electrical conductivity.
Salt content General designation for the quantity of salt dissolved in water.
Slope (relative) The relative slope expresses the relation of the slope value to the
value of a theoretical reference sensor of the same construction type.
96 ba77076e06 05/2021
MU 6100 L Glossary
Conductivity
Temperature compen- Name of a function that considers the temperature influence on the
sation measurement and converts it accordingly. Depending on the measured
parameter to be determined, the temperature compensation functions
in different ways. For conductimetric measurements, the measured
value is converted to a defined reference temperature. For potentiomet-
ric measurements, the slope value is adjusted to the temperature of the
test sample but the measured value is not converted.
General topics
Adjusting To manipulate a measuring system so that the relevant value (e.g. the
displayed value) differs as little as possible from the correct value or a
value that is regarded as correct, or that the difference remains within
the tolerance.
AutoRange Name of the automatic selection of the measuring range.
Calibration Comparing the value from a measuring system (e.g. the displayed
value) to the correct value or a value that is regarded as correct. Often,
this expression is also used when the measuring system is adjusted at
the same time (see adjusting).
ba77076e06 05/2021 97
Glossary MU 6100 L
98 ba77076e06 05/2021
MU 6100 L Index
18 Index
ba77076e06 05/2021 99
Index MU 6100 L
Z
Zero point of pH electrode . . . . . . . . . . . . . 30
19 Technical service
Web resources
Contact us:
For information or technical assistance contact your local VWR repre-
sentative or visit www.vwr.com.
20 Warranty
VWR warrants that this product will be free from defects in material and
workmanship for a period of three (3) years from date of delivery. If a
defect is present, VWR will, at its option and cost, repair, replace, or
refund the purchase price of this product to the customer, provided it is
returned during the warranty period. This warranty does not apply if the
product has been damaged by accident, abuse, misuse, or misapplica-
tion, or from ordinary wear and tear. If the required maintenance and
inspection services are not performed according to the manuals and
any local regulations, such warranty turns invalid, except to the extent,
the defect of the product is not due to such non performance.