Obstetric monitor
STAR5000F (C20)

Standard Configuration 1: Fetal and Maternal monitor: C31 module (3-lead


Fetal monitor : FHR, TOCO, FM, AFM, printer ECG,NIBP,TEMP,RESP,Masimo SPO2, EtCO2 interface), FHR,
Standard Configuration 2: TOCO, FM, AFM, printer
Fetal and Maternal monitor: C30 module (4.3inch led Standard Configuration 5:
screen,3-lead ECG,NIBP,TEMP,RESP,Masimo SPO2, EtCO2 Fetal and Maternal monitor: C31 module (3-lead
interface), FHR, TOCO, FM, AFM, printer ECG,NIBP,TEMP,RESP,Comen SPO2, EtCO2 interface), FHR,
Standard Configuration 3: TOCO, FM, AFM, printer
Fetal and Maternal monitor: C30 module (4.3inch led Optional Configuration:
screen,3-lead ECG,NIBP,TEMP,RESP,Comen SPO2, EtCO2 Twins monitoring, Fetal stimulator, ETCO2(ETCO2 is just
interface), FHR, TOCO, FM, AFM, printer available in Fetal and maternal monitor )
Standard Configuration 4: Safety Standards
ISO 13485:2003 approved, CE marking according to
MDD93/42/EEC
Physical Characteristics rate:
Product Size: 400 ㎜×300mm×95 mm Ultrasound: 1.0MHz ±10%
Net Weight: 5kg FHR Range: 50bpm- 210bpm
Display: 12.1’’ Color TFT LCD touch screen Resolution: 1bpm
Resolution: 800X600 Accuracy: ± 2 bpm
Method: folded TOCO
Operation Environment TOCO range: 0-100 Relative(%)
working 0-40℃ Resolution: ±1%
Temperature: Zero Mode: Manual
Humidity: 15-85% Non-linear error: ≤±8%
Power Supply AC100-250V, 50/60Hz Fetal Movement(FM)
Battery Type: Rechargeable lithium-ion battery Mode: Auto; manual
Battery Capacity: 14.8V DC 4400mAh Technique: Pulsed Doppler Ultrasound
Battery Recharging Maximum 6.5 hours for charging; Range: 0-100(%)
Time: Resolution: 1%
Battery working 4 hours for continuous working Fetal Stimulator
period: Method: Vibration
Interface: USB(data input or output, upgrade) Mode: Manual(3 times), continuing
AC power supply socket Voltage: DC 3V, (AA battery)
Parameter cable interface Respiration
RS232 port Method: impedance method
RJ45 port RR measurement Adult: 0-120bpm
Data: Input/output(patient’s case) range:
Network: Central Monitor System Pediatric/Neonate: 0 -150bpm
Date storage Resolution: ±1 rpm
patient case 3000 groups Accuracy: 6-150rpm:±1rpm or 1%, whichever is
Alarm event recall: 300 groups greater
NIBP Recall: 2000 groups 0-6rpm: unspecified
Trend Graph: 120 hours RESP Apnea Alarm: 10s-40s
Trend Table: 120 hours Alarm: Audible and visual alarm; alarm
freezed wave recall 240s for Adult, 60 hours for fetal events recallable
Recorder Sweep Speed: 6.25,12.5,25,50 mm/s
Record width: 150mm Gain Selection: X0.25, X0.5, X1, X2
Recording parper: 150mm*90mm or ECG
150mm*100mm , Z-fold, thermo Lead Type Cardio TecTM 12-leads ECG Analysis, 5
sensitive Lead and 3 lead selectable
Fast print speed: 25mm/s(stored traces) Lead selection: 3-lead: I; II; III
Paper Advance 1/2/3cm/min selectable 5-lead: I; II; III; aVR; aVL
Speed: 12-Lead I; II; III; aVR; aVL;aVF; V1-V6.
Record Data,time, TOCO type,paper Gain Selection: X 0.125, X0.25, X0.5, X1, X2, X4,auto
information : speed,FHR type,bed no.,etc. Sweep Speed: 6.25,12.5, 25, 50mm/s
Printer: Built-in thermal recorder; external Bandwidth : MON Mode:0.5 Hz~40 Hz
laser printer DIA Mode: 0.05Hz~130 Hz
Ultrasound OPE Mode:1 Hz~20 Hz
Technique: Ultrasound Pluse Doppler with ST Mode:0.2Hz~40Hz
autocorrelation
Pulse repetition 1 MHz
Defib. Protection: Withstand 4000VAC/50Hz voltage in Comen SPO2
isolation ,Again electrosurgical Measurement & 1~100%
interference and defibrillation alarm range
Recovery time : 3s Resolution: 1%
ST SEGMENT -2.0mV - +2.0Mv (Automatic) Accuracy: ±2% (70~100%,Ped, non-motion+
detection: ±3% (70-100%, Neo, non-motion);
Arrhythmia Yes 0-69% unspecified
Analysis : Alarm range 0~100%
pacemaker detectable Pulse rate: Range: 20~300bpm
detection: Resolution:1bpm
Alarm: Yes, audible and visual alarm, alarm Accuracy: ±3bpm(non-motion)
events recallable ±5bpm(motion)
NIBP Alarm range:20~250bpm
Method Automatic oscillometric EtCO2(Sidestream)
Work mode: Manual / Automatic/Continuous CO2 Measurement 0 -150 mm Hg, 0 to 79%, 0 to 20kPa
Measurement Unit: mmHg / Kpa selectable Range: (at 760mmHg)
Measurement Systolic, Diastolic, Mean Accuracy: ± 2 mm Hg (0 – 40 mm Hg)
types: ± 5% of reading (41 – 70 mm Hg)
Range of systolic Adult Mode:40-270mmHg ± 8% of reading (71 –100 mm Hg)
pressure: Pediatric Mode:40-200mmHg ± 10% of reading (101 –150 mm Hg)
Neonate Mode 40-135mmHg awRR range 0-150rpm
Range of diastolic Adult Mode:10-215mmHg RR accuracy: ±2rpm (0-70rpm)
pressure: Pediatric Mode:10-150mmHg ±5rpm (>70rpm)
Neonate Mode 10-100mmHg Response time: <240msec (10% to 90%)
Range of mean Adult Mode:20-235mmHg Delay time: <2s
pressure: Pediatric Mode:20-165mmHg EtCO2(mainstream)
Neonate Mode 20-110mmHg CO2 Measurement 0 -150 mm Hg, 0 to 79%, 0 to 20kPa
Over-pressure both Hardware and software over Range: (at 760mmHg)
protection: pressure protection Accuracy: ± 2 mm Hg (0 – 40 mm Hg)
Accuracy: Less than ±5mmHg ± 5% of reading (41 – 70 mm Hg)
Resolution 1mmHg ± 8% of reading (71 –100 mm Hg)
PR from NIBP: 40-240bpm ± 10% of reading (101 –150 mm Hg)
Temperature (Dual Channel) awRR range 0-150rpm
Measurement 0-50°C (32 to 122F) awRR precision ±2rpm (0-70rpm)
range: ±5rpm (>70rpm)
Resolution: 0.1°C Delay time: <2s
Accuracy: ±0.1°C (without probe) *Notice: Specifications subject to changes without prior
Channel: 2 channels, T1, T2, TD (Temperature notice. All rights reserved by Comen
Difference)