33K6-4-1362-1 Rev 30 April 2018 DUROMETER PDF - Ret
33K6-4-1362-1 Rev 30 April 2018 DUROMETER PDF - Ret
33K6-4-1362-1
TECHNICAL MANUAL
CALIBRATION PROCEDURE
FOR
DUROMETER
TYPES A, B, C, D, DO, O, OO
Distribution Statement C - Distribution authorized to U. S. Government agencies and their contractors for
official use or for administrative or operational purposes only, 30 September 1988. Requests for this
document shall be referred to AFMETCAL, 813 Irving-Wick Dr W, Heath, OH 43056-1199.
Destruction Notice - For unclassified, limited documents, destroy by any method that will prevent
disclosure of the contents or reconstruction of the document.
Published under Authority of the Secretary of the Air Force
30 APRIL 2018
T.O. 33K6-4-1362-1
DUROMETER
TYPES A, B, C, D, DO, O, OO
1 CALIBRATION DESCRIPTION:
Table 1.
Indentor Extension
Accuracy: 0.001 in
Accuracy: 0.001 in
Hardness
Accuracy: 0.075 N
(7.65 g)
Accuracy: 0.075 N
(7.65 g)
Accuracy: 0.4445 N
(45.33 g)
1
T.O. 33K6-4-1362-1
Table 1. (Cont.)
Hardness (Cont.)
Accuracy: 0.4445 N
(45.33 g)
Accuracy: 0.4445 N
(45.33 g)
Accuracy: 0.0176 N
(1.79 g)
Accuracy: 0.075 N
(7.65 g)
*1
Type OO Range: 0 to 100 points
(0.203 to 1.111 N)
(20.70 to 113.29 g)
Accuracy: 0.0182 N
(1.856 g)
*2 P/N 80-130 and 80-135 certified at the 80, 100 and 135 points only. Refer to step 3.5.
2
T.O. 33K6-4-1362-1
2 EQUIPMENT REQUIREMENTS:
Accuracy: N/A
Accuracy: Class 6
Accuracy: N/A
*2 Items (2.4), (2.5) and (2.6) are optional items for (2.1) Durometer Calibrator.
3 PRELIMINARY OPERATIONS:
3.1 Review and become familiar with the entire procedure before beginning the Calibration Process.
WARNING
3.3 Ensure the TI is clean and free from defects or damage capable of affecting use or calibration.
3
T.O. 33K6-4-1362-1
3.4 As applicable, connect test equipment to an appropriate power source. Set POWER switches to ON and allow
warm-up as required by the manufacturer.
3.5 Type A, B, C, D, DO, O and OO TIs are certified from 20 to 90 points only. P/N 80-130 and 80-135 TIs are
certified at 80, 100 and 135 points only. A Limited Certification label will be required, annotated with the applicable
limitation.
3.6 To minimize friction error, lightly tap the mechanical/analog TI with finger before reading.
3.7 Applied values listed in Table 2 were determined using the following equations derived from ASTM D2240:
3.8 Perform only those portions of the Calibration Process which pertain to the TI being calibrated.
4 CALIBRATION PROCESS:
NOTE
Unless otherwise specified, verify the results of each test and take corrective
action whenever the test requirement is not met, before proceeding.
CAUTION
Use care not to damage the indentor when calibrating Type B and D TIs. If not
strictly observed, could result in damage to, or destruction of, equipment or loss
of mission effectiveness.
4.1.1 With the TI indicating zero, compare indentor to Gage Block stack, verify extension is between 0.096 and
0.100 inch, for P/N 307L verify extension is between 0.098 and 0.100 inch, for Type M TIs verify extension is
between 0.048 and 0.050 inch.
4
T.O. 33K6-4-1362-1
4.2.1 Determine which type of TI is to be tested and set up the test equipment as shown in Figure 1. Ensure
appropriate Slide/Counter Weights and Contact Head are installed onto Durometer Calibrator Beam.
NOTE
For Types D, B TIs use Brass Contact Head. For Types A, C, DO TIs use
Aluminum Contact Head.
0
90 10
80 20
TI 70 30
60 40
50
PINION
KNOB
SLIDE CONTACT
WEIGHT HEAD
100 90 80 70 60 50 40 30 20 10 0D 0 0
BEAM COUNTER
WEIGHT
READER
LEVEL
DUROMETER
CALIBRATOR
RIGID TABLE
Figure 1.
4.2.4 Ensure the TI is secured onto the Durometer Calibrator or Holder (external stand) as required and the TI
indentor is aligned to and near the Durometer Calibrator Contact Head. Zero the TI as necessary.
NOTE
4.2.5 Position Durometer Calibrator Slide Weight to the first value listed in the TI Point column of Table 2.
4.2.6 Lower TI on the Contact Head until the Durometer Calibrator balance line lines up to zero. Lightly tap on the
black Pinion Knob to ensure TI is seated.
4.2.7 The TI must indicate within the values listed in the Point Limits column of Table 2. If using a Balance and
Weights instead of a Durometer Calibrator, verify that applied Weight indicates within values listed in the Load
Limits column of Table 2.
5
T.O. 33K6-4-1362-1
4.2.8 Repeat steps 4.2.5 through 4.2.7 for each remaining value listed in the TI Point column of Table 2.
Table 2.
6
T.O. 33K6-4-1362-1
Table 2. (Cont.)
4.2.9 Annotate and attach a Limited Certification Label as per step 3.5.
4.2.10 Calibration complete. As applicable, set all POWER switches to OFF, disconnect and secure all equipment.
Not Required
7/(8 Blank)