X-Ray Fluorescence Analysis of Zinc and Zinc-Iron Alloy Coated Steel Sheet
X-Ray Fluorescence Analysis of Zinc and Zinc-Iron Alloy Coated Steel Sheet
Table 1 shows analyzing depths of zinc–iron alloy intensity ratios of Zn-Kα (take off angle: 20 degrees)/
layer obtained by theoretical calculation for different Fe-Kα (take off angle: 20 degrees) and Zn-Kβ1 (take off
analysis lines and take off angles. The thin film angle: 40 degrees) / Fe-Kα (take off angle: 40 degrees).
FP method determines both iron concentration and Table 2 shows differences of measured intensities and
coating weight by solving simultaneous equations from analyzed results between using intensity ratios and
two measured intensities. The larger the difference in direct intensities when a sample height is increased
analyzing depth between the two analysis lines, the more by 0.25 mm. This table clearly shows that using the
accurate the analytical value of iron concentration and intensity ratio reduces errors due to the sample height.
coating weight. It is also important that the analyzing
depth of analysis lines used for the thin film FP 3. Measurement examples of hot-dip zinc and
method is larger than the thickness of the plating layer. zinc–iron alloy coated steel sheet
Combinations of Zn-Kβ1 (take off angle: 40 degrees) 3.1. Measurement example of GI
and Zn-Kα (take off angle: 20 degrees) satisfy these Figure 6 shows the relationship of coating weight
conditions sufficiently and have been shown to be most between chemical analysis values and XRF analysis
suitable for analysis of GA. values obtained by using Simultix 15. Measurement
When actually measuring the coating weight and conditions are shown in Table 3. Chemical analysis
composition by the thin film FP method using analysis values and X-ray analysis values are in good agreement,
lines with different take off angles, there are two and the accuracy of coating weight is 4.57 g/m2 in the
potential sources of error. First, large change in the range of 150 to 450 g/m2.
result of coating weight and composition occur by Table 4 shows a measurement repeatability of a
small variation in measured intensity and second, a sample. The coefficient of variation (C.V.) of coating
sample height variation can influence coating weight and weight (402.2 g/m2) is 0.26%, and good measurement
composition of results. repeatability is obtained.
In order to reduce these errors, in Simultix 15,
intensities of Zn-Kα and Zn-Kβ1 are divided by 3.2. Measurement example of GA
intensities of Fe-Kα having the same take off angles Figure 7 shows the relationship of coating weight and
respectively, and the thin film FP method uses the
Zn-Kα 32 227
Fe-Kα 32 165
Zn-Lα 32 10
Zn-Kα 40 264
Zn-Kβ1 40 325
Zn-Kα 20 159
Fe-Kα 40 190
Fe-Kα 20 117
Table 2. Influence by change of sample height on X-ray Fig. 6. Relationship of coating weight between chemical
intensity/X-ray intensity ratio and analysis value (a analysis value and XRF analysis value.
sample height is increased by 0.25 mm).
Fig. 7. Relationship of (a) coating weight and (b) iron concentration between chemical analysis value and
XRF analysis value.
Table 4. Result of repeatability for the same GI sample. same as previous subsection. Chemical analysis values
and X-ray analysis values are in good agreement, and
Coating weight
Number of measurements the accuracy of coating weight and iron concentration in
g/m2
plating layer are 0.71 g/m2 and 0.27 mass%, respectively.
1 400.6 Table 5 shows a measurement repeatability of a sample.
2 401.8 The coefficient of variation (C.V.) of coating weight
3 401.7 (50.4 g/m2) and iron concentration (12.1 mass%) are
4 402.3 0.07% and 0.15%, respectively, and good measurement
5 402.5 repeatability is obtained.
6 401.8
7 402.8 4. Conclusion
8 404.5 Analysis examples of Simultix 15 are shown for hot-
9 401.3 dip zinc and zinc–iron alloy coated steel sheet. In case
10 402.5 of GI, it is possible to accurately measure thick plating
Average 402.2 of 400 g/m2 or more by optimizing analysis line and take
Standard deviation 1.04 off angle. In case of GA analysis by using Simultix 15,
C.V. (%) 0.26 since it is unnecessary to dissolve samples like chemical
analysis and possible to simultaneously measure coating
weight and iron concentration in the plating layer by
Table 5. Result of repeatability for the same GA sample. using Simultix 15, it greatly shortens analysis time in
batch analysis in which the number of measurement
Number of Coating weight Iron concentration
samples is expected to increase dramatically as the
measurements g/m2 mass%
production volume increases.
1 50.36 12.07 Currently, in steel sheets for automotive body, the
2 50.38 12.11 usage rate of steel sheet with high specific strength
3 50.40 12.09 such as advanced high strength steel sheet is increasing
4 50.41 12.10 in order to achieve both weight reduction for the
5 50.37 12.11 purpose of reducing fuel consumption and securing of
6 50.42 12.06 safety. In order to realize a high specific strength, the
7 50.35 12.08 structure of the steel sheet is controlled by adding alloy
8 50.43 12.10 elements such as Ni, Mn, Si and Nb(13). Depending on
9 50.34 12.08 the steel type, alloying elements added is 20 mass% or
10 50.32 12.11 more. When such a steel sheet is galvannealed, alloy
Average 50.38 12.09 elements as well as iron diffuse into the plating layer.
Standard deviation 0.036 0.018 Since certain conditions are defined for the composition
C.V. (%) 0.07 0.15 of steel sheet and elements in the plating layer in FP
calculation, sufficient accuracy may not be obtained if a
single calibration for thin film FP calculation is applied
iron concentration between chemical analysis values and to galvannealed steel sheets with different steel plate
XRF analysis values obtained by using Simultix 15 with composition. In this case, sufficient accuracy can be
the thin film FP method. Measurement conditions are obtained by dividing the thin film FP calibration for each
steel type for the analysis of galvannealed steel sheets. ferrous materials—Gravimetric determination of the mass per
unit area (1992).
(7) JIS G 3302, Hot-dip zinc-coated steel sheet and strip (2010).
References (8) Rigaku Corporation, X-ray fluorescence spectrometer, United
(1) M. Urai, J. Iwaya, M. Iwai and M. Arimura: Kobe Steel States Patent 7356114, 2008. 4. 8.
Engineering Reports, 47 (1997), 29–32. (9) Rigaku Corporation, X-ray fluorescence spectrometer, Korea
(2) M. Sakurai, L. W. Zhang, Y. Tajiri and T. Kondo: Tetsu-to- Patent 101046018, 2011. 7. 1.
Hagane, 77 (1991), 979–986. (10) Rigaku Corporation, X-ray fluorescence analyzer, China Patent
(3) H. Takahashi: Journal of The Surface Finishing Society of 1932493, 2012. 1. 4.
Japan, 68 (2017), 610–616. (11) Rigaku Corporation, X-ray fluorescence spectrometer and x-ray
(4) J. Kawabe, H. Hashimoto, R. Ohta and S. Takahashi: Tetsu-to- fluorescence measurement method, EU Patent 1764612, 2015.
Hagane, 73 (1987), S1187. 4. 15.
(5) K. Nishifuji, H. Tanabe, K. Imai, H. Kato and A. Honda: (12) Y. Kataoka, H. Kohno, E. Furusawa and M. Mantler: X-Ray
Bulletin of the Japan Institute of Metals, 30 (1991), 307–309. Spectrometry, 36 (2007), 221–225.
(6) ISO 1460, Metallic coatings—Hot dip galvanized coatings on (13) M. Takahashi: Tetsu-to-Hagane, 100 (2014), 82–93.