Download as pdf or txt
Download as pdf or txt
You are on page 1of 28

WHITE PAPER

Reproducible
Quantification
in PET•CT: Clinical
Relevance and
Technological
Approaches

Partha Ghosh, MD

www.siemens.com/mi
Table of Contents

The Influence of Quantification on Clinical Decision Making 1

The Factors Influencing the Accuracy and Reproducibility of Quantitative Results 2

Patient Preparation and Protocol 2

Patient Size 2

Post Injection Delay Times 2

Plasma Glucose Levels 2

Acquisition and Reconstruction Parameters 2

Instrumentation 5

Detector System 5

Calibration Method 6

Registration and Attenuation Correction Methods 7

Quantification Software 11

PET Oncology Follow-up with SUVpeak 11

PET Cardiac Perfusion with Myocardial Blood Flow 13

Conclusion 16

References 16
INDICATIONS AND USAGE IMPORTANT SAFETY INFORMATION
Fludeoxyglucose F18 Injection is indicated for positron emis- Radiation Risks
sion tomography (PET) imaging in the following settings: Radiation-emitting products, including 18F FDG, may increase
the risk for cancer, especially in pediatric patients. Use the
• Oncology: For assessment of abnormal glucose metabolism smallest dose necessary for imaging and ensure safe handling
to assist in the evaluation of malignancy in patients with to protect the patient and health care worker.
known or suspected abnormalities found by other testing
modalities, or in patients with an existing diagnosis of cancer. Blood Glucose Abnormalities
In the oncology and neurology setting, suboptimal imaging
• C ardiology: For the identification of left ventricular may occur in patients with inadequately regulated blood
myocardium with residual glucose metabolism and glucose levels. In these patients, consider medical therapy
reversible loss of systolic function in patients with coro- and laboratory testing to assure at least two days of normo-
nary artery disease and left ventricular dysfunction, glycemia prior to 18F FDG administration.
when used together with myocardial perfusion imaging.
Adverse Reactions
• Neurology: For the identification of regions of abnormal glucose Hypersensitivity reactions with pruritus, edema and rash have
metabolism associated with foci of epileptic seizures (1). been reported; have emergency resuscitation equipment and
personnel immediately available.

Full prescribing information may be found at the end of this


white paper.

The Influence of Quantification on


Clinical Decision Making
Metabolic imaging with PET/CT using 18 F Fludeoxyglucose Accurate and reproducible measurement of SUV is critical to the
( 18F FDG) * as well as other imaging biomarkers has achieved clinical efficacy of PET/CT imaging since the quantitative parameter
wide acceptance in oncology, cardiology and neurology not only is often used to guide clinical decision making. This is especially
because of the unique metabolic information generated by this true for characterization of equivocal lesions like lung nodules
modality, but also because of its ability to quantify biological where an SUV of 2.5 or higher has been traditionally regarded as
processes. Such quantitative parameters generated by PET/CT the criteria to label a lesion as malignant and subject it to biopsy1.
imaging act as valuable tools for characterization of lesions as A prospective study1 attempted to differentiate malignant from
well as for monitoring response to therapy. benign lung nodules in 87 patients. When a mean SUV of greater
than or equal to 2.5 was used for detecting malignancy, the
The fundamental quantitative parameter in PET is the Standard- sensitivity, specificity and accuracy were 97 percent, 82 percent
ized Uptake Value (SUV), which is calculated as follows: and 92 percent, respectively. In addition, there was a significant
correlation between the doubling time of tumor volume and
the SUV.
Activity Concentration (kBq/ml)
SUV = SUV-based quantification is also critical for proper evaluation of
Inj. dose (MBq)
BodyWt (kg) response to therapy of tumors using PET/CT, an approach which
is being increasingly adopted in oncology for monitoring chemo-
therapy and chemoradiation therapy, as well as to modify therapy
regimes in cases of non-response which may be ascertained early
This value can be normalized to body weight as in the above using sequential PET/CT imaging during therapy. Initial tumor
equation or normalized to body surface area or lean body mass. response to effective chemotherapy is usually a decrease in tumor
Traditionally the SUV max (pixel with the highest SUV within a metabolic activity with subsequent decrease in tumor size based
given region of interest (ROI) in a tumor) has been used for tumor on the degree of cell killing and resulting intratumoral necrosis
characterization and follow up although the reproducibility of and fibrosis. Often the time lag between metabolic response
this measurement is subject to significant variation. and morphological response is significant, measured in weeks. A
series of studies involving response evaluation of various tumors
using PET/CT has shown the advantage of PET/CT in demonstra-
tion of early response or non-response to therapy much before
decrease in tumor volume could be detected on CT. Although

*Important safety information on Fludeoxyglucose 18F injection can be found on the previous page.
The full prescribing information can be found on pages 18-25.
1
the RECIST criteria based on morphological parameters of tumor The Factors Influencing the
has been widely used, it is very often inappropriate in assessing
therapy response in certain tumors. Several newer cancer thera- Accuracy and Reproducibility of
pies have a more cytostatic than cytocidal effect and good tumor Quantitative Results
response may be associated predominantly with a decrease in
metabolism, even in the absence of a major shrinkage in tumor
size. Quantitative evaluation of tumor metabolism, as indicated SUV is subject to several sources of variability arising from the
by PET SUV, is increasingly being adopted as an adjunct to RECIST patient and the PET/CT system, including acquisition and recon-
for clinical trials involving therapy response evaluation, especially struction parameters, as well as statistical variations related to
those related to such cytostatic and targeted therapies. Therapy system calibration and evaluation methodology.
response evaluation using molecular imaging biomarkers corre-
lates better with patient outcome than do size measurements, Patient Preparation and Protocol
which could significantly reduce the costs of drug development
and, thereby, eventually decrease drug costs in clinical practice. Patient Size
Body habitus is a source of variability in SUV since conventional
The percentage decline of SUV that fits into the profile of a SUV normalizes for body weight. However, fat has a much lower
responding tumor differs among cancer types. In lung, gastric 18F FDG uptake compared to muscle and organs, such as the liver.

and esophageal cancers, declines in SUVmax of 20 to 35 percent Several tissues show progressive increase in SUV with increasing
after one to two doses of therapy have been shown to be body weight due to the proportionally higher fat content7. SUV
predictive of favorable outcomes 2 . However, in lymphomas normalized to lean body mass (SUL) reduces body fat percentage
and ovarian carcinoma, a larger drop in SUVmax following initial related variability. However, not all centers may use standardized
therapy is observed in responding lesions3,4. The levels of SUV SUV normalization criteria and thus intersystem variability may
change required to conform to the clinical profile of a favorable persist.
response varies widely and depends primarily on tumor biology.
Thus a relatively modest decrease in SUVmax may constitute a Post Injection Delay Times
good response in some tumors. Cytostatic chemotherapy agents The waiting period between 18F FDG injection and PET/CT acquisi-
may block tumor growth but the SUV max may remain constant tion has been shown to have an impact on SUV8. Several studies
or decrease slightly. Studies have demonstrated significant vari- have demonstrated progressive increase in SUV in malignant lung
ability of SUV max values in test-retest measurements, ranging tumors subjected to dual time point acquisitions9. These observa-
between 6 to 10 percent5 with increase in variability with higher tions suggest that SUV can vary widely depending on delay in
SUVmax values. Due to such variability, the level of SUVmax change acquisition post injection, and standardization of the post injec-
which is to be considered a positive indicator for therapy response tion delay is important for reproducible and reliable results, as
is subject to controversy. The European Organization for Research has been adopted in most current clinical trials. The European
and Treatment of Cancer (EORTC) PET study group 6 suggested a Association of Nuclear Medicine (EANM) procedure guidelines for
reduction of a minimum of 15 percent to 25 percent in tumor oncologic PET imaging10 recommend the interval between 18F
SUVmax or SUVmean to reflect a partial metabolic response (PR) FDG administration and the start of acquisition to be 60 minutes.
after one cycle of chemotherapy and greater than 25 percent Recording of the actual interval and time of injection is also
after more than one treatment cycle. Decrease in test-retest recommended. Adherence to same post injection delay for repeat
variability of SUV secondary to technological innovations and imaging of the same patient with a tolerance of +/- 5 minutes is
improved standardization of data acquisition and reconstruction recommended in the context of therapy response assessment.
should improve criteria for therapy response assessment. On
the road to reliable response assessment based on quantitative Plasma Glucose Levels
results, it is therefore desirable to define true thresholds for The plasma glucose level also has a major effect on SUV. Hyper-
progressive disease and partial response. As a prerequisite, scan- glycemia has been shown to cause a significant reduction in tissue
ners would have to be accurate to within a small margin. FDG uptake and consequently a reduction in SUV11. Such variations
demand strict monitoring and management of plasma glucose level
Tumor volume estimation based on PET as well as delinea- prior to PET/CT studies in order to reduce SUV variability.
tion of gross tumor volume (GTV) for PET/CT-based radiation
therapy planning usually involves determining a threshold as a Acquisition and Reconstruction Parameters
percentage value of the SUVmax of a tumor. Thus the accuracy of Accurate measurement of injected dose is key to accurate and
the highest SUV of a tumor is a key component for accurate delin- reproducible SUV calculation. Attention to calibration of the dose
eation of tumor margins if threshold-based Volume of Interest calibrator as well as measurement of residual activity in syringe
(VOI) generation tools are used. after injection is of utmost importance. SUV calculation depends
on the measurement of activity concentration in kBq/ml and is
In view of the clinical relevance of accurate and reproducible SUV thus affected by factors such as recovery coefficients (RC) and
in PET imaging, it is important to look at the variables that affect partial volume effects (PVE). The recovery coefficient of a PET/
SUV and patient-specific, as well as system- and technology- CT system can be defined as the ratio of the measured activity in
specific measures, that could reduce such variability. a lesion divided by the true activity. It decreases when the struc-

2
Patient- and system-specific
factors influence the accuracy
and reproducibility of SUV.
Improvements in image quality,
attenuation correction method­
ology, system calibration and
quality control, as well as adequacy
of quantitative software tools,
together influence the ultimate
SUV accuracy and diagnostic
confidence.

ture is smaller than a few times the practical image resolution or Such variability can be significantly improved by technological
when such structure occupies a portion of the voxel due to inad- developments including higher performance PET detector mate-
equate sampling. In both situations, the recovery coefficient is rial (LSO) with higher count rate efficiencies; higher resolution
typically less than 100 percent and SUV may be calculated lower scanners with smaller crystal elements and voxel sizes (HI-REZ
than the actual value. Hoffman et al 12 evaluated this problem crystals) as well as newer reconstruction algorithms that improve
and assessed the recovery coefficient for images of cylinder with lesion contrast and resolution throughout the entire FOV (Point
uniform radioactivity distribution. He demonstrated a relation- Spread Function (PSF) reconstruction such as HD•PET). More-
ship between system resolution, cylinder diameter and recon- over, acquisition and reconstruction with time of flight (TOF)
struction parameters, especially the Full Width at Half Maximum (Siemens ultraHD•PET) dramatically reduce background noise
(FWHM) of the reconstruction filter. He concluded that RC of and increases lesion contrast, which can impact small lesion
a cylinder with diameter equal to the resolution FWHM of the detectability. Higher system resolution with smaller voxel size
imaging system was only 50 percent. Such RC equally affects the reduces partial volume effects, improving the accuracy of SUV.
SUV calculation. Typically, Gaussian filters of 5 to 10 mm FWHM
are applied resulting in a clinical image resolution of 7 to 12 mm. Attempts are being made to improve standardization of PET/
PVE increases with decreasing image resolution. PVE becomes CT-based clinical trials in order to enhance the reliability of SUV
mainly important for lesions smaller than three times the FWHM, across multiple systems in order to reduce variability. A PET
such as lesions 20 mm or less. Recovery coefficient and partial imaging working group was formed by the Netherlands Society
volume effects also strongly depend on object geometry with of Haemato-Oncology 13. The group specifically aimed at stan-
maximum effect in irregular shaped tumors. dardization of 18F FDG PET studies in order to allow inter-institute
interchangeability of SUV. This group recommended well-defined
With iterative reconstruction methods, a sufficient number of guidelines for patient preparation, data acquisition, image recon-
iterations need to be applied to ensure sufficient convergence struction and inter-institutional standardization of image quality
of the algorithm. SUV increases of up to 70 percent have been parameters based on standardized phantom studies. Several of
shown when using higher numbers of iterations. the recommendations are discussed below.

Data acquisition settings such as time per bed position, the The Netherlands study group recommended a strict adherence
amount of overlap between subsequent bed positions, acquisi- to injected dose, post injection delay and iterative reconstruc-
tion mode (2D or 3D) and 18F FDG dose affect scan statistics and/ tion guidelines among centers in order to improve SUV reliability.
or the noise equivalent count rate of PET studies. Poorer scan Matching of image resolution as closely as possible among
statistics and lower image signal-to-noise ratio (SNR) result in an centers in multi-centre trials to avoid differences in SUV was
upward bias of SUV, especially when using the SUV max as final recommended. The group also recommended guidelines for
outcome parameter. CT attenuation correction (CT AC) can influ- ROI selection in lesions to help standardization. Fixed size ROI
ence SUV in cases with patient motion (e.g., breathing), which providing ‘SUVpeak’, maximum pixel values providing ‘SUV max’ or
may cause a mismatch between PET and CT and thus result in the average ROI value within a 2D or 3D ROI providing ‘SUVmean’
incorrect attenuation correction. The latter may be minimized by were all put into standardized guidelines. Recommendations
breathing instructions (breath holding at mid-inspiration volume for patient preparation were specified in order to minimize
or shallow breathing) during CT scanning. Truncation of CT due patient-related or other physiological effects on SUV accuracy
to smaller CT field of view (FOV) may lead to faulty attenuation and reproducibility. Moreover, guidelines aimed at optimizing
correction and SUV errors and should therefore be avoided.

3
18F FDG uptake in the tumor, minimizing uptake in surrounding evidence, the Netherlands group advocates phantom-based
tissues (muscle, brown fat) and minimizing SUV variability were standardization of PET/CT systems as a prerequisite for partici-
put forward. Recommendations for the administration procedure pation in clinical trials. The guidelines mandate the maximum
were given to ensure that the net dose given to the patient is voxel value or SUV max to be always reported. However, use of
exactly known, primarily by avoiding unknown remaining larger VOI may provide SUV estimates with better precision.
activities during preparation and administration. 18F FDG dose Therefore, in response monitoring settings, use of 3D VOI with a
was specified as a function of patient weight, scanning mode, threshold of 41 percent, 50 percent or 60 percent of the SUVmax
percentage bed overlap and acquisition time per bed position. is recommended to be used consistently in all sequential scans of
a patient13. When such threshold-based VOIs are not adequate,
Interchangeability of SUV between acquisitions in different scan- smaller VOI obtained with higher percentage threshold values
ners and in different institutions is affected by the overall spatial can be applied. The guidelines included quality control (QC)
resolution of the PET images after reconstruction, filtering or measurements using torso abdominal phantom with multiple 18F
smoothing and processing. Comparison of data from different FDG-filled spheres of various sizes with a lesion-to-background
scanners requires standardized reconstruction and filtering in ratio of 8 to be used to calculate reproducibility of SUVmax and
order to match image resolution across scanners so as to allow SUVmean at different threshold-based VOIs. VOIs based on adap-
for SUV interchangeability. SUV outcome is also determined by tive threshold of 41 percent of maximum SUV provide VOI with a
data analysis procedures. Common ROI strategies are the use of volume close to real sphere volumes used.
fixed sized 2D or 3D ROI, manually defined ROI in one or more
axial slices and 3D ROI based on region growing procedures while As reflected by the studies and observations of working groups
applying a user-specified threshold. for standardization of PET-based therapy clinical trials, the vari-
ability of SUV between sequential observations on the same
Boellaard et al 14 performed a simulation study to determine system or different systems exert considerable influence on
the effects of noise, image resolution and ROI definition on the the interpretation of such trials. This highlights the need for
accuracy of SUVs. Experiments and simulations were based on adequate PET/CT hardware and software optimization to reduce
thorax phantoms with tumors of 10, 15, 20 and 30 mm diameter system-related variability of quantification parameters. Such
and tumor-to-background ratios (TBRs) of 2, 4 and 8. Fifty sino- variability may be related to non-optimized quality control and
grams were generated at three noise levels. All sinograms were calibration procedures, photomultiplier drifts due to tempera-
reconstructed using ordered subset expectation maximization ture and other factors, system sensitivity fluctuations, PET and
(OSEM) with two iterations and 16 subsets, with or without a CT misregistrations related to minor patient motion, etc. Special
6 mm Gaussian filter. For each tumor, the maximum pixel value focus on PET/CT hardware design and engineering to reduce or
and the SUVmean (from ROI based on isocontour and on adaptive eliminate such variability introducing factors by optimizing and
threshold of 50 percent and 70 percent), as well as 3D ROI of automating quality control and calibration procedures, as well as
one cubic centimeter volume (SUVpeak) placed in the zone of the software-based autoregistration of PET and CT and automated
highest uptake were calculated. The results indicated that the use of appropriate quantitative measures like SUVpeak in order
maximum pixel value within an object increases with image noise to reduce statistical variability, would possibly increase the reli-
level and with increasing object size. Based on this experimental ability of SUV for response monitoring.

Figure 1: PET Detector Materials Properties

PET Detector Material Properties


Property Characteristic Desired Value LSO BGO GSO Nal

Density (g/cc) High 7.4 7.1 6.7 3.7


Defines detection efficiency of detector and
scanner sensitivity
Effective atomic number High 65 75 59 51

Decay time (nsec) Defines detector dead time and randoms rejection Low 40 300 60 230

Relative light output (%) Impacts spacial and energy resolution High 75 15 35 100

Energy resolution (%) Influences scatter rejection Low 10.0 10.1 9.5 7.8

Nonhygroscopic Yes Yes Yes Yes No


Simplifies manufacturing, improves reliability
and reduces service costs
Ruggedness Yes Yes Yes No No

4
Instrumentation Crystal Geometry
The size of individual crystal elements in a detector block in
As discussed in the preceding chapters, the accuracy of quantita- the PET gantry is a key factor that influences the resolution of
tive values, like SUV derived from PET/CT imaging, depends not acquired PET image. High resolution images not only improve
only on patient-specific variables such as injected dose and blood lesion detection and diagnostic accuracy, but also improve the
glucose level but, more importantly, on system-specific factors accuracy of quantification due to lower partial volume effects,
like detector sensitivity and resolution, as well as consistency of as mentioned previously in this paper. Biograph mCT achieves
system calibration, PET and CT coregistration and appropriate the industry’s finest volumetric resolution* of 87 cubic mm with
algorithms for image reconstruction. a maximum image matrix of 400x400 due to the OptisoHD detec-
tion system, with an arrangement of 13x13 crystals per detector
Detector System block of 4x4x20 mm crystal elements. Such high resolution
imaging decreases partial volume effects compared to conven-
The detector system is at the heart of the PET scanner. Its design tional detector designs and achieves the highest NEMA and
determines the key performance characteristics. The elements of volumetric resolution for clinical PET.* Due to the high resolution
the detector system are the crystal, the photomultiplier tube and and optimum count rate efficiency of Biograph mCT, visualization
the processing electronics. Opposing crystals in a PET detector of small lesions is enhanced along with the quantitative accuracy
ring receive the coincident photons emitted secondary to a posi- of the SUV generated.
tron annihilation and convert them into a light signal, which is
then received and amplified by a photomultiplier tube. In order
for the opposing photons to be recognized as a single coincident
event, they must strike the detectors located opposite each other
in the PET gantry within a small duration of time. This is called
13x13 LSO array
the coincidence window. The detector material is key to count
rate efficiency. The detector performance characteristics are
Crystals
determined by the material, crystal size, detector geometry and
the quantity of crystal elements in a PET/CT system. Light guide

Crystal Material
Count rate efficiency of a PET crystal depends on factors like scin-
2x2 PMT array
tillation decay time and light output. Crystal and photomultiplier
tubes, along with system electronics, need to optimally perform
to obtain very short coincidence windows. A short coincidence
window leads to higher true count rates and lower scatter and Voltage divider
random events, which constitute noise. Crystal decay time and
light output contribute to the ability to achieve shorter coinci-
dence windows. Coincidence windows for current generation OptisoHD detector design with 13x13 LSO crystal matrix and
photomultiplier tube assembly.
PET detector materials are in the range of 4.1 nanoseconds.
Conventional PET scintillator materials like bismuth germanium
oxide (BGO), gadolinium oxyorthosilicate (GSO) or sodium iodide Crystal Volume
doped with thallium (Nal) are characterized by larger coincidence The volume of crystal material available in a gantry not only deter-
windows. Adoption of newer crystal material, like lutetium oxyor- mines the FOV per PET bed position and the coverage, but also
thosilicate (LSO), pioneered by Siemens, has increased count rate the count rate during 3D acquisition. As the amount of crystal
efficiency and made 3D scanning possible, which has significantly increases, the amount of coincident photons also increases along
improved acquisition speeds and image quality. with an increase in the true count rate capability. Siemens has
pioneered the concept of increasing the amount of crystal mate-
As shown in the comparative chart of various PET detector rial within a system by increasing the number of detector rings
materials (Figure 1), LSO has high density and a high atomic in order to create a larger solid angle for PET acquisition with
number for fast scintillation decay time and a high light output consequent increase in true count rate capability in 3D mode.
for improved energy and position determination. This enables As a result, the number of bed positions and the time per bed
Siemens Biograph™ PET•CT systems to achieve high count rate decrease at the same time for a compounded effect on scan time
efficiencies even with low injected dose. reduction. Therefore, a unique four ring detector design (TrueV)
on the Biograph system is available to enable scans at half
With LSO, seemingly opposing requirements are achieved at the the time or half the dose compared to conventional three ring
same time: for oncology imaging, superb image quality is demon- designs. When combined with TOF technology (ultraHD•PET),
strated even at fast acquistion speeds, while for dynamic perfu- imaging can be performed at half the time and half the dose
sion imaging, high count rates are recorded without detector compared to conventional three ring designs. The increase in
saturations due to rapid decay times. count rate efficiency also helps achieve optimum counting statis-

*Based on competitive literature available at time of publication; data on file. 5


tics even at very high resolutions in order to achieve superb quan- Individual crystal efficiencies, like light output, may change
titative accuracy even with significant reductions in acquisition within a detector block and may lead to inaccuracies in counting
time. Quantitative accuracy is compromised with low count rate efficiency. The block noise check analyzes the efficiencies of
and high noise levels. Thus improved count rate capabilities due each crystal within a detector block. The efficiency of each
to crystal and detector-related innovations help improve quantita- crystal is compared to the average efficiency of all crystals in
tive accuracy of the system. the same location in all blocks. Any crystals with large deviation
are flagged as defects while others are effectively compensated
Calibration Method through calibration. This ensures consistent performance among
all crystal elements within a detector block.
As mentioned earlier, the reproducibility and accuracy of a quan-
titative parameter like SUV is influenced by the overall system With multiple detector blocks within a detector ring, inconsisten-
performance. Factors such as inherent detector drift, shifting cies within individual detector blocks can increase noise and arti-
PMT gains due to temperature and voltage variations and other facts. The block efficiency checks the efficiency of entire detector
factors may affect quantification accuracy. To compensate for blocks against each other. Deviations >20 percent are flagged as
these effects and ensure consistent results, appropriate calibra- failed; those below the threshold can effectively be calibrated to
tion mechanisms need to be employed. Simpler methods use ensure a uniform image.
line sources with spread out calibration intervals that measure
a subset of performance parameters. However, more compre- 3D PET acquisition benefits from short coincidence windows
hensive methods are required to more closely replicate the char- to decrease random coincidences which may severely increase
acteristics of a patient15 and also to allow for more advanced image noise. Thus the objective of every PET detector design and
acquisition technologies, like TOF. In order to optimize the calibration method is to improve true count rates and eliminate
procedure for consistent system calibration and to ensure repro- randoms. The measured randoms test compares the measured
ducibility and accuracy of quantitative measurements, Biograph amount of randoms from the phantom against the calculated
mCT incorporates a daily phantom-based quality control and number given by the singles rate squared. A variation above 3
calibration protocol termed as Quanti•QC. Quanti•QC measures percent is classified as a failed test. The obvious source of error
and calibrates a large array of system parameters on a regular is an improperly set coincidence window. Thus the measured
basis in order to fine tune the system for optimum performance randoms test helps ensure the proper operation of the coinci-
and reproducible quantitative measurements and to eliminate dence detection system.
system-related variables. In addition, the tightly regulated water-
cooled gantry ensures system temperature stability for consistent The scanner efficiency, also called sensitivity, is defined as the
performance. counts per second per Becquerel per cubic centimeter (cps/Bq/
cc). Sensitivity is primarily determined by the amount of LSO in
Quanti•QC the scanner and should not change over time. However, sensi-
The process employs daily system calibration using a cylindrical tivity may vary due to the phantom activity or volume being
20 cm diameter, 27 cm long phantom with approximately two entered incorrectly or hardware failures leading to failure to
mCi of 68Ge and, among other steps, performs 11 automated detect true coincidence events. Measurement and calibration for
steps to achieve normalization of the PET scanner: optimized sensitivity can be performed only with a phantom, since
a volume reference is required. The phantom activity in Bq/cc is
1. Block Noise required to calculate the sensitivity in cps/Bq/cc from the corrected
2. Block Efficiency cps rate. The test is passed with a deviation < 30 percent from the
3. Measured Randoms expected sensitivity. Utilization of a daily QC protocol based on a
4. Scanner Efficiency 68Ge phantom calibrating for system sensitivity, as well as all other

5. Scatter Ratio parameters mentioned earlier, marks the key difference of Biograph
6. Scanner Efficiency Correction Factor mCT from the conventional line source-based calibration approach
7. Image Plane Efficiency and defines the commitment of Siemens to reproducible quantita-
8. Block Timing Offset tive accuracy.
9. Block Timing Width
10. Time Alignment Residual
11. Time Alignment Fit (x/y)

6
Scatter of photons is an essential component of PET acquisition The time alignment fit and residual analyzes the consistency
and can degrade image quality and add to image noise. The between the spatial and temporal dimensions of the TOF sino-
scatter ratio check, which is a part of the daily calibration, also gram. In order to improve image quality using TOF reconstruction
requires a phantom, since only volumes exhibit scattering. The 20 methods, the information in the TOF sinogram must be consis-
cm diameter 68Ge polyethylene phantom used for daily calibra- tent between the time and spatial domains of the sinogram.
tion in Biograph mCT has a known scatter fraction of 32 percent
on a 4-ring PET system. Based on the relationship between scatter Phantom-based daily Quanti•QC based on measurement, calibra-
fraction and tube drift in the PET photomultiplier tubes (PMT), tion and optimization of such a wide range of parameters is key
the scatter test is a highly sensitive indicator of PMT drift. If the to the consistent and reproducible quantitative accuracy and
scatter test fails, a detector setup is required. image quality performance of Biograph mCT. Such a comprehen-
sive calibration approach performed as a daily routine is a key
Further checking of the accuracy of calibration is performed by differentiator of Biograph mCT from other approaches.
the scanner efficiency correction factor check which determines
if the absolute image calibration has been computed correctly.
Again, a volumetric phantom is required to compare the fully Registration and Attenuation Correction Methods
corrected PET image to the actual current radioactivity in the
phantom. The test is passed with a deviation <30 percent from Correct attenuation correction (AC) of PET is of crucial impor-
the expected value. The test is similar to the scanner efficiency tance not only for optimum image quality but also for the accu-
test, but based on actual reconstructed images rather than the racy of measured quantitative values. In addition to relevance
raw sinogram. to oncologic imaging, PET quantification expands to parameters
like myocardial blood flow relevant to PET myocardial perfusion
The image plane efficiency test is an additional validation to imaging. Since CT-based attenuation correction is key for proper
check if normalization is achieved in all image planes. In the image generation for both whole-body PET/CT for oncology as
world of 3D PET scanning, there are many more counts in the well as PET in cardiology and neurology, any approach to improve
center of each ring than between the rings and at the ends of the attenuation correction also helps quantitative accuracy.
FOV. The normalization corrects for these statistical differences
by computing a plane efficiency for each axial slice in the image. The key to correct CT attenuation correction (CT AC) is accurate
If these corrections are not accurate, there would be streaks in coregistration of PET and CT. For whole-body PET/CT acquisition,
the coronal view of a patient. inaccurate coregistration can be based on several factors: system-
related CT and PET misregistrations may be related to patient
TOF acquisition involves determination of the time interval bed deflection and misregistration due to minor patient motion
between two coincident photons and using that information between PET and CT acquisitions. Specific artifacts related to
to determine the origin of the coincidence photons. This tech- such misregistrations have been defined in the literature and are
nique reduces background activity and improves lesion contrast recognized by clinicians. However, the impact of such misregis-
significantly. The block timing offset and width test is related trations on quantitative accuracy have not yet been clearly delin-
to TOF calibration. Since the test is performed with the 20 cm eated. Similar misregistrations between CT and PET in cardiac
phantom in the center of the FOV, it is expected that the blocks PET myocardial perfusion or reversibility studies have been more
on opposite sides of the gantry would have the majority of their widely studied since the artifacts can often be significant and
emission events in the center time bin of the TOF sinogram. This may lead to clinical interpretation errors. In the era of quantita-
phenomenon can be observed by histogramming all the emission tive myocardial blood flow imaging with PET, improper attenua-
events into the 13 time bins provided in the TOF sinogram. These tion correction due to CT and PET misregistration due to different
histograms should also have a characteristic width based on the breathing patterns and diaphragmatic positions during CT and
phantom size. Regular calibration for such TOF specific factors PET acquisition may cause inaccurate flow values and create
ensures optimal image quality since introduction of TOF adds attenuation correction-generated perfusion abnormalities16.
additional variables to the PET acquisition process.

7
Figure 2: Conventional vs Cantilever Patient Handling System

Conventional SMART PHS The conventional patient bed design


(left) shows significant deflection
between the CT and PET which leads
to systematic misregistration between
the CT and PET. The cantilever design of
the SMART PHS of Biograph mCT (right)
shows no deflection between the two
gantries and the resulting image also
shows perfect coregistration.
CT PET CT PET

Cantilever Patient Handling System Design The most common correction method used in clinical practice is
Biograph mCT eliminates system-introduced misregistration a manual checking and alignment after the data is acquired20,
by ensuring absence of any patient bed deflection between though this can be subject to inter and intra-user variability. It
the CT and PET by a unique cantilever patient handling system is also time-consuming21. A different approach is to acquire a
(SMART PHS). Although PET/CT systems are integrated, the PET number of fast, low-dose CT scans during free breathing, and then
and CT scanner components are still placed in sequential order. choose the best-aligned CT in order to perform AC22. However, it
Changes in patient position due to mechanical variations within is still time-consuming to consider each CT in turn, and leads to
the patient handling system need to be avoided between the additional imaging time and radiation dose to the patient.
PET and CT scans. Conventional designs, where the imaging
pallet alone supports the patient within the CT and PET imaging An automated process of registration of PET and CT data has the
field, incur various degrees of deflection, which may lead to an potential to significantly improve quality and accuracy of cardiac
offset between the PET and CT acquisitions. A cantilever design PET by eliminating faulty CT AC-related artifacts and consequent
where the patient handling system moves as one avoids deflec- quantitative inaccuracies and also by eliminating time consuming
tion, resulting in consistency between the two scans, thereby and often inaccurate manual coregistration required for correct
improving registration accuracy. alignment. SMART Auto Cardiac Registration, a feature on
Biograph mCT, is an automatic translation algorithm to align the
Auto Cardiac Registration PET image with a single CT in order to reduce time spent manu-
Misregistration in PET/CT in cardiac imaging is common since ally aligning datasets, improve reproducibility, and reduce dose,
the position of the heart and diaphragm can be vastly different by only requiring one CT.
during the short CT acquisition compared to the longer PET acqui-
sition. Respiratory cycles can cause the left ventricle (LV) to be For the Auto Cardiac Registration on Biograph mCT, a translation-
translated up to 22 mm17, and heart motion during the CT acqui- only registration algorithm has been implemented that uses the
sition can also lead to stepping artifacts in the CT AC images18. Mutual Information similarity function23 and a Powell optimizer24
This may cause parts of the LV in the emission image, typically in to determine the optimal translation. Mutual Information is
the lateral mid-apical region, to invade the lung parenchyma in a popular image similarity measure for registration of multi-
the CT AC and lead to incorrect µ values. The resulting emission modality images. It is a measure of the statistical dependence
image will then have a decrease of counts in such areas, which between two random variables or the amount of information
may be misread as a real perfusion defect caused by cardiovas- that one variable contains about the other. It can be qualitatively
cular disease19. explained as a measure of how well one image explains the other.

8
Figure 3: Conventional vs Auto Cardiac Registration

Improvement in anterolateral wall uptake (arrow) following correct alignment of cardiac perfusion PET and CT prior to attenuation correction. Inferior wall
attenuation correction is evident in spite of slight misalignment of CT and PET in the upper row. However, subtle decrease in anterolateral wall in the corrected
image due to misalignment (upper row) is absent when the CT and PET are correctly aligned (bottom row).

Misaligned CT and PET

Uncorrected PET PET and CT Fusion Attenuation-


Corrected PET

Correctly aligned CT and PET

Uncorrected PET PET and CT Fusion Attenuation-


Corrected PET

Given that both PET and CT images show certain clearly discern- and the CT dataset is then registered to it. While the registra-
able similarities as in landmarks like clearly defined organs, tion is calculated mainly using the information inside the LV, the
discernable differences in uptake and Hounsfield Unit (HU) levels whole CT image volume is translated with respect to the emission
between various organs and surrounding tissues like heart and image prior to reconstruction.
lungs, liver and abdomen, etc., an image intensity histogram is
able to measure the amount of information available in the two Auto Cardiac Registration has been validated to perform at least
images combined and the similarities between the two in order as well as a clinical manual registration for the purpose of AC of
to achieve coregistration. The similarity function is modified by PET/CT when separate free breathing CT scans are acquired for
means of a weighting function that ensures that the algorithm both rest and stress, and for healthy and diseased patients25.
focuses on the alignment of the LV, not on other structures.

The location of the LV is identified during an automated pre-


processing step, where the LV is detected and cropped in the
axial, coronal and sagittal dimensions of the PET non-AC dataset,

9
Figure 4: Conventional vs Neuro AC Attenuation Correction

Conventional CT AC
18
F FDG brain PET following
CT mu-map Attenuation-Corrected PET Neuro AC without CT appears
identical to the same data
attenuation corrected using
CT. Note absence of any
discernable difference in
uptake patterns and intensity
between the two corrected
PET images.

SMART Neuro AC

Calculated mu-map Attenuation-Corrected PET

PET-only Attenuation Correction The SMART Neuro AC algorithm on Biograph mCT takes advan-
Recent focus on neurological applications of PET, including tage of the brain’s fairly simple and consistent geometry and
epileptic foci determination with 18F FDG as well as potential new tissue distribution. It approximates the mu-map for a PET only
applications like brain amyloid imaging have led to innovations brain scan based on the ability to find the tissue/air boundary
in attenuation correction for PET brain acquisition. Attenuation around head from a non-attenuation corrected (NAC) PET scan.
correction approaches for PET brain without a CT is desirable due
to avoidance of radiation dose as well as eliminating radiation to To accommodate all possible tracers in brain imaging, an adap-
sensitive organs like the lens of the eye. However, such attenua- tive thresholding of the edge map is required, using a method of
tion correction algorithms need not only maintain image quality polar boundary processing. Once the perimeter of the head has
but also ensure quantitative accuracy similar to brain PET with been determined, bone inside the air/skin boundary is modeled
CT AC.

10
using image morphology. At 4 mm from the boundary, a 4 mm Quantification Software
segment of bone is introduced. The assigned mu values for the
three classes of mu are: PET Oncology Follow-up with SUVpeak
Current practice incorporates SUVmax (the maximum value of SUV
air = 0.000 cm-1 in a single voxel within a tumor ROI) as the major quantitative
tissue = 0.093 cm-1 parameter. However, since SUV max is essentially a single voxel
bone = 0.140 cm-1 ROI, it is subject to larger statistical variation compared to other
SUV measurement criteria. Average SUV within a tumor ROI has
Testing was performed on a total of 129 patient scans using both been used in several studies but suffers from a lack of reproduc-
qualitative and quantitative methods. The patient population ibility, since it depends on the tumor ROI, which may be subject
covers a variety of tracers 3, scanner types 2 and sites 4. Results to individual variations based on physician preferences regarding
indicate that 60 percent of the brain pixels are below 5 percent thresholding or manual margin drawings. Thus, for the majority
error and almost 99 percent of brain pixels are below 10 percent of studies and clinical trials, SUVmax has remained the default
error (absolute value) as defined to be the percent difference choice in spite of its limitations. SUVmax is highly dependent on
between the reconstructed PET values using the calculated the statistical quality of the images and the voxel size depending
mu-map vs. the values using the CT-based mu-map. The algo- on the acquisition matrix. The SUVmax variability increases as the
rithm also demonstrated good timing performance by creating a lesion matrix size is increased from 128x128 to 256x256. Vari-
mu-map from an uncorrected PET in about six seconds. ability increases with fewer counts as the patient size increases

Figure 5: Conventional SUVmax vs SUVpeak quantification

Conventional Siemens syngo.via


1 Pixel (SUVmax) 1 cm3 (SUVpeak)

Graphical representation of individual voxels with SUV involving a lung tumor after (top) and before (bottom) therapy. Pre-therapy and post-therapy
SUVmax evaluation (left) shows a 17.6 percent decrease in metabolic activity, which reflects inadequate tumor response. SUVpeak calculation (right)
based on an average of multiple voxels involving the region with highest uptake shows a far more significant tumor response with a 30.3 percent
decrease in metabolic activity.

Since SUVmax is subject to significant statistical variation, SUVpeak promises to be more reflective of true response as well as more robust to statistical variations.

11
Figure 6: Comparison of SUVmax, SUVpeak and tumor volume across three time points

May 8, 2008 May 20, 2008 June 4, 2008

SUVmax 20.73 SUVmax 15.19 SUVmax 9.12


SUVpeak 16.30 SUVpeak 8.81 SUVpeak 5.4
Tumor Vol 20.7 cm3 Tumor Vol 6.22 cm3 Tumor Vol 2.81 cm3

and the count statistics decrease 5 . The measured activity comparable dose and post injection delays. SUVmax and SUVmean
concentration in a small volume of a PET image depends on the were determined for tumor ROI drawn on the first study and
activity in neighboring voxels. Activity levels in the neighboring transferred onto the second study. SUVmax changes between two
voxels may cause a spill-in or spill-out of activity in any voxel studies were significantly higher than SUVmean and in some cases
being evaluated for SUV. This partial volume effect, combined it increased by 1.5 SUV units in which the SUVmax was above 7.5.
with sensitivity to image noise, is the reason SUVmax in a single Thus the variability in hottest tumors was higher than the ones
voxel within a tumor ROI is subject to inaccuracies. SUV max will with lower 18F FDG uptake. This calls into question the reliability
increase with a higher level of image noise, which may be due to of depending on SUVmax for therapy response evaluation in varie-
shorter acquisition time. Lower injected dose may also contribute gated tumors with fluctuating metabolic levels. SUVmean, on the
to higher noise due to suboptimal count statistics for a standard other hand, was more reproducible and the variation from large
acquisition time. ROIs did not differ more than 0.5 units. SUVpeak measurements
(a fixed sphere of one cubic cm volume centered on the hottest
SUVs obtained from larger, fixed ROIs are more reproducible than area of the tumor) are more reproducible since they involve
single pixel SUVs as is usual for calculation of SUVmax. Nahmias representing the mean value of a few voxels representing the
and Wahl et al 26 studied 26 patients with 18F FDG PET in two hottest tumor area. This makes it close to SUVmax in terms of its
separate occasions with a mean interval of 3+/-2 (SD) days with representation of the maximum tumor metabolism, yet it could

12
avoid the statistical fluctuations of SUVmax due to incorporation Dynamic PET myocardial perfusion imaging using list mode
of a larger number of voxels within the hottest tumor area. Velas- acquisition can be used to calculate absolute myocardial blood
quez et al27 assessed the repeatability of SUVmax, SUVpeak, and flow (MBF) values in different myocardial segments as well as
SUV mean in 62 patients in a multicenter phase I oncology trial Coronary Flow Reserve (CFR) which is defined as the ratio of
who underwent double baseline 18F FDG PET studies. The varia- MBF at peak stress to that at rest. In normally perfused healthy
tion of SUVmax between two baseline studies was much larger myocardium, blood flow at peak stress is 3 to 4 times higher than
than the other parameters including SUV peak and SUVmean. This that at rest. A CFR of 2.5 or higher is usually associated with
suggests that values obtained from regions with multiple voxels absence of significant coronary artery or microvascular disease 29.
are more statistically reproducible and dependable for clinical Unlike standard 82Rb and 13N NH3 MPI acquisitions that start
trials. Although SUVpeak promises to have less statistical variation, around 90 seconds and three minutes post-radionuclide admin-
the parameters used to calculate SUV peak, like sphere diameter, istration respectively, dynamic list mode acquisition starts with
impacts the reproducibility. A recent study28 evaluated the impact radionuclide administration to obtain data upon radionuclide
of different volumes for SUVpeak estimation on the reproducibility arrival in the right and left ventricle. The List Mode (LM) acquisi-
of SUV peak and tumor response using pre- and post-therapy tion continues for 6-8 minutes and 6-10 minutes for 82Rb and 13N
18F FLT ** PET/CT studies in a series of 17 patients with various NH3, respectively. Rest and stress radiopharmaceutical admin-
solid tumors. In this study, 24 different SUVpeaks were determined istration consist of approximately 1110 MBq (30 mCi) 82Rb or
for each FLT avid tumor. There was substantial variation in both 370-740 MBq (10-20 mCi) 13NH3. Time activity curves are gener-
SUVpeak and percentage tumor response with changes in SUVpeak ated for myocardial segments as well as for a voxel in the LV for
sphere diameter. SUVpeak of a lesion was calculated to be up to 49 generation of the arterial input function. MBF is calculated from
percent higher or 46 percent lower than the mean when values the time activity curves using tracer kinetic modeling based on a
obtained using 24 different sphere volumes were compared. one tissue compartment (82Rb) or two tissue compartment (13N
Similar variation (from 49 percent above to 35 percent below the NH3) models.
mean) was also seen for percentage tumor response. The study
supported the recommendation of adopting a 1.2 cm diameter
sphere with a volume of 1 cubic cm as a standard for SUV peak
definition for tumors larger than 2 cm in diameter. AMMONIA N 13 INJECTION

The sequential 18F FDG PET study in Figure 6 illustrates the INDICATIONS AND USAGE
variation between SUVmax and SUVpeak for evaluation of therapy
response. The lung mass shows decrease in 18F FDG uptake and Ammonia N 13 Injection is a radioactive diagnostic agent
SUVmax as well as SUVpeak. SUVmax and SUVpeak varies between 21 for Positron Emission Tomography (PET) indicated for
to 40 percent for individual acquisitions. Decline in SUVmax is less diagnostic PET imaging of the myocardium under rest or
pronounced compared to SUVpeak with 56 percent vs. 66 percent, pharmacologic stress conditions to evaluate myocardial
respectively. This is inherent to the SUVmax metric of the highest perfusion in patients with suspected or existing coronary
SUV in a VOI and could be more influenced by statistical variation. artery disease.

PET Cardiac Perfusion with Myocardial Blood Flow IMPORTANT SAFETY INFORMATION
Myocardial perfusion imaging at peak stress and at rest using
PET/CT with 13N NH3 or 82Rb has been widely accepted as a reli- Ammonia N 13 Injection may increase the risk of cancer.
able method of detection of myocardial ischemia secondary to Use the smallest dose necessary for imaging and ensure
coronary artery disease29. Standard methods of evaluation based safe handling to protect the patient and health care worker.
on relative perfusion where the image is normalized to the region
of highest perfusion may occasionally underestimate the degree ADVERSE REACTIONS
of ischemia and in some cases of balanced ischemia (secondary
to triple vessel disease), miss the ischemia completely. A normal No adverse reactions have been reported for Ammonia
relative perfusion in patients with multivessel balanced disease N 13 Injection based on a review of the published litera-
is not uncommon. ture, publicly available reference sources, and adverse drug
reaction reporting system.

**18F FLT is an investigational drug and is not approved for clinical use in the U.S. 13
Figure 7: Conventional relative perfusion vs. myocardial blood flow31

Conventional Siemens syngo.via


Relative Perfusion Absolute Flow

Graphical representation of perfusion imaging in patient with global ischemia (top) and in a normal patient (bottom). In conventional relative perfusion
imaging, the polar plot display results are normalized to the highest activity observed. The relative perfusion pattern on the polar map in the upper left
illustration would appear normal since the region with the highest perfusion itself is already ischemic. In cases of overall depressed perfusion, an adequately
perfused reference section is missing, which may result in misleading information.

With myocardial blood flow, an absolute measure is available that can be compared to established normal values. In the top right illustration, the peak
stress values in all segments range from 0.76 to 0.99 ml/min/g while the lower limit of normal is at around 2.0 ml/min/g. This suggests significant ischemia
in all myocardial segments (balanced ischemia) which reflects uniform levels of stenosis in all coronary arteries (triple vessel disease). This helps reduce
inconclusive results even in cases of balanced disease.

In comparison, a polar plot of stress myocardial perfusion in normal patients (bottom left) again shows a uniform perfusion pattern and the stress
myocardial blood flow values (bottom right) are also in the normal range with values between 2.39 and 3.11 ml/min/g29,31.

14
Figure 8: Relative perfusion compared to myocardial blood flow in case of balanced ischemia.

The syngo® MBF software offers a simplified method of calcu- Reduced stress MBF response is also seen in microvascular
lation of MBF and CFR values for both 13N NH3 and 82Rb with disease such as that associated with diabetes. Thus MBF needs
automated delineation of myocardial margins and placement of to be interpreted with caution in the absence of clearly defined
the LV ROI for the arterial input function generation. MBF and suspicion of coronary artery disease.
CFR values of individual myocardial voxels and arterial territories
as well as multiple distinct myocardial segments are depicted as Several factors affect the accuracy of MBF. These include
numerical values as well as parametric polar plots with appro- injected dose, volume and speed of injection, patient motion
priate scaling. and adequacy of CT attenuation correction. Since the cardiac
and diaphragm positions during the fast CT may be different
The mean +/- standard deviation (SD) of MBF and CFR in normal compared to that in the PET acquisition acquired for longer time,
controls using 13N NH3 and syngo MBF is 0.86 +/-0.29 ml/g/min the potential for misregistration and artifacts related to faulty
(rest MBF), 2.05 +/-0.73 ml/g/min (stress MBF), and 2.65 +/- 1.17 attenuation correction is significant and this may impact MBF
(CFR). Rubidium-82 values in normal controls also have similar accuracy as well. Biograph mCT incorporates automated coreg-
stress and resting flow values29,31. istration for CT and PET for accurate attenuation correction of
cardiac PET which improves the accuracy for CT attenuation
Inducible ischemia due to coronary artery disease is commonly correction and also the reliability of MBF.
associated with reduced MBF response to stress. A stress MBF
of 1.8-2.0 ml/g/min is usually regarded as the lower limit of
normal 29,31 . Resting flow rates have a larger normal range
although flow rates below 0.40 ml/gm/min are usually associated
with resting ischemia.

15
Several studies have established the value of MBF measurements References
in the evaluation of coronary artery disease as well as the charac-
terization of luminal stenosis and myocardial microvasculature. 1D
 uhaylongsod, FG et al. (1995 Jul). Detection of primary and
A recent study30 evaluated MBF using 13N NH3 and correlated recurrent lung cancer by means of F-18 fluorodeoxyglucose
with stenosis on CT angiography. Myocardium supplied by vessels positron emission tomography (FDG PET). J Thorac Cardiovasc
with significant stenosis showed significantly lower stress MBF Surg, 110 (1), 130-139.
(1.63 +/- 0.51 ml/g/min) compared to those with normal coronary
arteries (2.13 +/ 0.54 ml/g/min). 2W
 eber, WA et al. (2001 June 15). Prediction of response
to preoperative chemotherapy in adenocarcinomas of the
The clinical example depicted in Figure 8 illustrates the value of esophagogastric junction by metabolic imaging. J Clin Oncol,
MBF in diagnosis of balanced ischemia related to triple vessel 19 (12), 3058-3065.
disease in a patient with suspected coronary artery disease where
visual interpretation of relative perfusion did not clearly demon- 3L
 in, C et al. (2007 Oct). Early 18F-FDG PET for prediction
strate any segmental ischemia. of prognosis in patients with diffuse large B-cell lymphoma:
SUV-based assessment versus visual analysis. J Nucl Med,
Figure 8 shows a 13N NH3 myocardial perfusion PET•CT of a 48 (10), 1626-1632.
65-year-old woman with progressive dyspnea and signs of heart
failure. The stress rest PET images do not show well-defined hypo- 4A
 vril, N et al. (2005 Oct 20). Prediction of response to neoad-
perfused areas. However, the MBF values show grossly decreased juvant chemotherapy by sequential F-18-fluorodeoxyglucose
MBF throughout the entire LV at peak stress with normal MBF at positron emission tomography in patients with advanced-stage
rest, suggestive of balanced ischemia probably related to triple ovarian cancer. J Clin Oncol, 23 (30), 7445-7453.
vessel disease. Anterior wall MBF at peak stress is 0.92 ml/g/min
(lower limit of normal 2 ml/g/min) while resting MBF is 0.75 ml/g/ 5W
 ahl, RL et al. (2009 May). From RECIST to PERCIST:
min which is within normal limits. Evolving Considerations for PET response criteria in solid
tumors. J Nucl Med, 50 Suppl 1, 122S-150S.

6Y
 oung, H et al. (1999 Dec). Measurement of clinical and
Conclusion subclinical tumor response using 18F fluorodeoxyglucose
When evaluating the factors influencing the accuracy and repro- and positron emission tomography: review and 1999 EORTC
ducibility of quantitative results on PET, it becomes obvious that recommendations. Eur J Cancer, 35, 1773-1782.
all elements in the imaging workflow need to be controlled and
optimized in order to avoid variations across devices, patients and 7Z
 asadny, KR et al. (1993 Dec). Standardized uptake values
time as postulated by the RSNA Quantitative Imaging Biomarkers of normal tissues at PET with 2-[flourine-18]-fluoro-2-deoxy-
Alliance (QIBA). D-glucose: variations with body weight and a method for
correction. Radiology, 189 (3), 847-850.
Biograph mCT together with syngo.via quantification software***
helps minimize system-based sources of variation by effectively 8H
 amberg, LM et al. (1994 Aug). The dose uptake ratio as
reducing manual interaction and automating steps, such as an index of glucose metabolism: useful parameter or over­
quality control and image quantification. simplification? J Nucl Med, 35 (8), 1308-1312.

Now, for the first time, Biograph mCT gives you quantifiable 9M
 acDonald, K et al. (2011 Mar). The role of dual time point
results that are accurate and reproducible over time, along FDG PET imaging in the evaluation of solitary pulmonary
with high count rate efficiency and superior resolution. With its nodules with an initial standard uptake value less than 2.5.
unique combination of intelligent software, daily calibration and Clin Radiol, 66 (3), 244-250.
precise anatomical and functional co-registration, Biograph mCT
makes a quantifiable difference in diagnostic confidence, therapy 10 B
 oellaard, R et al. (2010 Jan). FDG PET and PET/CT: EANM
planning and treatment monitoring. procedure guidelines for tumor PET imaging: version 1.0.
Eur J Nucl Med Mol Imaging, 37, 181-200.

11 L
 indholm, P et al. (1993 Jan). Influence of the blood
glucose concentration on FDG uptake in cancer—a PET study.
J Nucl Med, 34 (1), 1-6.

16 *** syngo.PET MBF is not yet commercially available in the U.S.


12 H
 offman, EJ et al. (1979 June). Quantitation in positron 26 N
 ahmias, C et al. (2008 Nov). Reproducibility of standard-
emission computed tomography: 1. Effect of object size. ized uptake value measurements determined by 18F FDG PET
J Comut Assist Tomogr, 3 (3), 299-308. in malignant tumors. J Nucl Med, 49 (11), 1804-1808.

13 B
 oellaard, R et al. (2008 Dec). The Netherlands protocol 27 V
 elasquez, LM et al. (2009 Oct). Repeatability of 18F
for standardization and quantification of FDG whole body FDG PET in a multi-centre phase I study of patients with
PET studies in multi-centre trials. Eur J Nucl Med Mol advanced gastrointestinal malignancies. J Nucl Med,
Imaging, 35 (12), 2320-2333. 50 (10), 1646-1654.

14 B
 oellaard, R et al. (2004 Sep). Effects of noise, image reso- 28 V
 anderhoek, M et al. (2012 Jan). Impact of the definition
lution, and ROI definition on the accuracy of standard uptake of peak standardized uptake value on quantification of
values: a simulation study. J Nucl Med, 45 (9), 1519-1527. treatment response. J Nuc Med, 53, 4-11.

15 L
 ockhart, C. M. et al. (2011 Feb). Quantifying and reducing 29 P
 aolo, G et al. (2009). The Clinical Value of Myocardial
the effect of calibration error on variability of PET/CT standard- Blood Flow Measurement. J Nucl Med, 50, 1076-1087.
ized uptake value measurements, J Nuc Med, Vol. 52, No. 2.
30 L
 iga, R et al. (2011 Nov). Structural Abnormalities of the
16 M
 oller et al. (2007) Artifacts form misaligned CT in cardiac Coronary Arterial Wall—in Addition to Luminal Narrowing—
perfusion PET/CT studies: Frequency, effects and potential Affect Myocardial Blood Flow Reserve. J Nucl Med, 52 (11),
solutions. J Nucl Med, 48, 188-193. 1704-1712.

17 G
 oetze, S et al. (2007). Attenuation Correction in Myocardial 31 H
 utchins, GD et al. (1990). Noninvasive quantification of
Perfusion SPECT/CT: Effects of Misregistration and Value of regional blood flow in the human heart using 13N ammonia
Reregistration. J Nucl Med, 48, 1090-1095. and dynamic positron emission tomographic imaging.
J Amer Coll Cardio, 5, 1032-1042.
18 H
 amill, J et al. (2008). Real-Time MRI for Assessment of
PET/CT Attenuation Correction Protocols. IEEE Nuclear
Science Symposium Conference Record, 3761-3768.

19 G
 ould, K et al. (2007). Frequent Diagnostic Errors in Cardiac
PET/CT Due to Misregistration of CT Attenuation and Emission
PET Images: A Definitive Analysis of Causes, Consequences,
and Corrections. J Nucl Med, 48, 1112-1121.

20 G
 oetze, S et al. (2007). Attenuation Correction in Myocardial
Perfusion SPECT/CT: Effects of Misregistration and Value of
Reregistration. J Nucl Med, 48, 1090-1095.

21 M
 artinez-Moller, A et al. (2007). Artifacts from Misaligned
CT in Cardiac Perfusion PET/CT Studies: Frequency, Effects
and Potential Solutions. J Nucl Med, 48, 188-193.

22 S
 treeter, J et al. (2007). Attenuation correction of stress
PET 82Rb with ultrafast CT images. J Nucl Med, 48, 446P.

23 V
 iola P, et al. (1997). Alignment by Maximization of Mutual
Information. Intl J Computer Vision, 24, 137-154.

24 P
 ress, WH et al. (1992). Numerical recipes in C (2nd ed.).
University Press, Cambridge, England.

25 B
 ond, S et al. (TO COME) Automatic Registration of
Emission and Transmission Images in Cardiac PET/CT and
Cardiac SPECT/CT.

17
HIGHLIGHTS OF PRESCRIBING INFORMATION ----------------DOSAGE FORMS AND STRENGTHS----------------
These highlights do not include all the information needed Multi-dose 30mL and 50mL glass vial containing 0.74 to 7.40
to use Fludeoxyglucose F 18 Injection safely and effectively. GBq/mL (20 to 200 mCi/mL) Fludeoxyglucose F18 Injection and
See full prescribing information for Fludeoxyglucose F 18 4.5mg of sodium chloride with 0.1 to 0.5% w/w ethanol as a
Injection. stabilizer (approximately 15 to 50 mL volume) for intravenous
administration (3).
Fludeoxyglucose F 18 Injection, USP
For intravenous use ------------------------CONTRAINDICATIONS------------------------
Initial U.S. Approval: 2005 None

----------------------RECENT MAJOR CHANGES--------------------- -------------------WARNINGS AND PRECAUTIONS------------------


Warnings and Precautions (5.1, 5.2) 7/2010 • Radiation risks: use smallest dose necessary for imaging (5.1).
Adverse Reactions (6) 7/2010 • Blood glucose abnormalities: may cause suboptimal imaging (5.2).

----------------------INDICATIONS AND USAGE--------------------- ----------------------ADVERSE REACTIONS----------------------


Fludeoxyglucose F18 Injection is indicated for positron emission Hypersensitivity reactions have occurred; have emergency resus-
tomography (PET) imaging in the following settings: citation equipment and personnel immediately available (6).
• Oncology: For assessment of abnormal glucose metabolism to To report SUSPECTED ADVERSE REACTIONS, contact PETNET
assist in the evaluation of malignancy in patients with known Solutions, Inc. at 877-473-8638 or FDA at 1-800-FDA-1088 or
or suspected abnormalities found by other testing modalities, www.fda.gov/medwatch.
or in patients with an existing diagnosis of cancer.
• Cardiology: For the identification of left ventricular myocar- ------------------USE IN SPECIFIC POPULATIONS------------------
dium with residual glucose metabolism and reversible loss • P regnancy Category C: No human or animal data. Consider
of systolic function in patients with coronary artery disease alternative diagnostics; use only if clearly needed (8.1).
and left ventricular dysfunction, when used together with • Nursing mothers: Use alternatives to breast feeding (e.g., stored
myocardial perfusion imaging. breast milk or infant formula) for at least 10 half-lives of radio-
• Neurology: For the identification of regions of abnormal glucose active decay, if Fludeoxyglucose F 18 Injection is administered
metabolism associated with foci of epileptic seizures (1). to a woman who is breast-feeding (8.3).
• Pediatric Use: Safety and effectiveness in pediatric patients have
-----------------DOSAGE AND ADMINISTRATION----------------- not been established in the oncology and cardiology settings (8.4).
Fludeoxyglucose F18 Injection emits radiation. Use procedures
to minimize radiation exposure. Screen for blood glucose See 17 for PATIENT COUNSELING INFORMATION
abnormalities.
• In the oncology and neurology settings, instruct patients to fast Revised: 1/2011
for 4 to 6 hours prior to the drug’s injection. Consider medical
therapy and laboratory testing to assure at least two days of
normoglycemia prior to the drug’s administration (5.2).
• In the cardiology setting, administration of glucose-containing
food or liquids (e.g., 50 to 75 grams) prior to the drug’s injec-
tion facilitates localization of cardiac ischemia (2.3).
Aseptically withdraw Fludeoxyglucose F18 Injection from its
container and administer by intravenous injection (2).
The recommended dose:
• for adults is 5 to 10 mCi (185 to 370 MBq), in all indicated
clinical settings (2.1).
• for pediatric patients is 2.6 mCi in the neurology setting (2.2).
Initiate imaging within 40 minutes following drug injection;
acquire static emission images 30 to 100 minutes from time of
injection (2).

18
FULL PRESCRIBING INFORMATION: CONTENTS*

1 INDICATIONS AND USAGE 8 USE IN SPECIFIC POPULATIONS


1.1 Oncology 8.1 Pregnancy
1.2 Cardiology 8.3 Nursing Mothers
1.3 Neurology 8.4 Pediatric Use
2 DOSAGE AND ADMINISTRATION 11 DESCRIPTION
2.1 Recommended Dose for Adults 11.1 Chemical Characteristics
2.2 Recommended Dose for Pediatric Patients 11.2 Physical Characteristics
2.3 Patient Preparation 12 CLINICAL PHARMACOLOGY
2.4 Radiation Dosimetry 12.1 Mechanism of Action
2.5 Radiation Safety – Drug Handling 12.2 Pharmacodynamics
2.6 Drug Preparation and Administration 12.3 Pharmacokinetics
2.7 Imaging Guidelines 13 NONCLINICAL TOXICOLOGY
3 DOSAGE FORMS AND STRENGTHS 13.1 Carcinogenesis, Mutagenesis, Impairment of Fertility
4 CONTRAINDICATIONS 14 CLINICAL STUDIES
5 WARNINGS AND PRECAUTIONS 14.1 Oncology
5.1 Radiation Risks 14.2 Cardiology
5.2 Blood Glucose Abnormalities 14.3 Neurology
6 ADVERSE REACTIONS 15 REFERENCES
7 DRUG INTERACTIONS 16 HOW SUPPLIED/STORAGE AND DRUG HANDLING
17 PATIENT COUNSELING INFORMATION
* Sections or subsections omitted from the full prescribing
information are not listed.

FULL PRESCRIBING INFORMATION

1 INDICATIONS AND USAGE 2.1 Recommended Dose for Adults


Fludeoxyglucose F18 Injection is indicated for positron Within the oncology, cardiology and neurology settings, the
emission tomography (PET) imaging in the following recommended dose for adults is 5 to 10 mCi (185 to 370
settings: MBq) as an intravenous injection.

1.1 Oncology 2.2 Recommended Dose for Pediatric Patients


For assessment of abnormal glucose metabolism to assist Within the neurology setting, the recommended dose for
in the evaluation of malignancy in patients with known or pediatric patients is 2.6 mCi, as an intravenous injection.
suspected abnormalities found by other testing modalities, The optimal dose adjustment on the basis of body size or
or in patients with an existing diagnosis of cancer. weight has not been determined [see Use in Special Popula-
tions (8.4)].
1.2 Cardiology
For the identification of left ventricular myocardium with 2.3 Patient Preparation
residual glucose metabolism and reversible loss of systolic • To minimize the radiation absorbed dose to the bladder,
function in patients with coronary artery disease and left encourage adequate hydration. Encourage the patient to
ventricular dysfunction, when used together with myocar- drink water or other fluids (as tolerated) in the 4 hours
dial perfusion imaging. before their PET study.
• E ncourage the patient to void as soon as the imaging
1.3 Neurology study is completed and as often as possible thereafter for
For the identification of regions of abnormal glucose at least one hour.
metabolism associated with foci of epileptic seizures. • S creen patients for clinically significant blood glucose
abnormalities by obtaining a history and/or laboratory
2 DOSAGE AND ADMINISTRATION tests [see Warnings and Precautions (5.2)]. Prior to
Fludeoxyglucose F18 Injection emits radiation. Use proce- Fludeoxyglucose F 18 PET imaging in the oncology and
dures to minimize radiation exposure. Calculate the final neurology settings, instruct patient to fast for 4 to 6 hours
dose from the end of synthesis (EOS) time using proper prior to the drug’s injection.
radioactive decay factors. Assay the final dose in a prop- • I n the cardiology setting, administration of glucose-
erly calibrated dose calibrator before administration to the containing food or liquids (e.g., 50 to 75 grams) prior to
patient [see Description (11.2)]. Fludeoxyglucose F 18 Injection facilitates localization of
cardiac ischemia

19
2.4 Radiation Dosimetry
The estimated human absorbed radiation doses (rem/mCi) show that there are slight variations in absorbed radiation
to a newborn (3.4 kg), 1-year old (9.8 kg), 5-year old (19 dose for various organs in each of the age groups. These
kg), 10-year old (32 kg), 15-year old (57 kg), and adult (70 dissimilarities in absorbed radiation dose are due to devel-
kg) from intravenous administration of Fludeoxyglucose F opmental age variations (e.g., organ size, location, and
18 Injection are shown in Table 1. These estimates were overall metabolic rate for each age group). The identified
calculated based on human 2 data and using the data critical organs (in descending order) across all age groups
published by the International Commission on Radiological evaluated are the urinary bladder, heart, pancreas, spleen,
Protection 4 for Fludeoxyglucose 18F. The dosimetry data and lungs.

Table 1. Estimated Absorbed Radiation Doses (rem/mCi) After


Intravenous Administration of Fludeoxyglucose F 18 Injectiona

Newborn 1-year old 5-year old 10-year old 15-year old Adult
Organ
(3.4 kg) (9.8 kg) (19 kg) (32 kg) (57 kg) (70 kg)

Bladder wallb 4.3 1.7 0.93 0.60 0.40 0.32


Heart wall 2.4 1.2 0.70 0.44 0.29 0.22
Pancreas 2.2 0.68 0.33 0.25 0.13 0.096
Spleen 2.2 0.84 0.46 0.29 0.19 0.14
Lungs 0.96 0.38 0.20 0.13 0.092 0.064
Kidneys 0.81 0.34 0.19 0.13 0.089 0.074
Ovaries 0.80 0.8 0.19 0.11 0.058 0.053
Uterus 0.79 0.35 0.19 0.12 0.076 0.062
LLI wall* 0.69 0.28 0.15 0.097 0.060 0.051
Liver 0.69 0.31 0.17 0.11 0.076 0.058
Gallbladder wall 0.69 0.26 0.14 0.093 0.059 0.049
Small intestine 0.68 0.29 0.15 0.096 0.060 0.047
ULI wall** 0.67 0.27 0.15 0.090 0.057 0.046
Stomach wall 0.65 0.27 0.14 0.089 0.057 0.047
Adrenals 0.65 0.28 0.15 0.095 0.061 0.048
Testes 0.64 0.27 0.14 0.085 0.052 0.041
Red marrow 0.62 0.26 0.14 0.089 0.057 0.047
Thymus 0.61 0.26 0.14 0.086 0.056 0.044
Thyroid 0.61 0.26 0.13 0.080 0.049 0.039
Muscle 0.58 0.25 0.13 0.078 0.049 0.039
Bone surface 0.57 0.24 0.12 0.079 0.052 0.041
Breast 0.54 0.22 0.11 0.068 0.043 0.034
Skin 0.49 0.20 0.10 0.060 0.037 0.030
Brain 0.29 0.13 0.09 0.078 0.072 0.070
Other tissues 0.59 0.25 0.13 0.083 0.052 0.042

a MIRDOSE 2 software was used to calculate the radiation absorbed dose. Assumptions on the biodistribution based on data from Gallagher et al.1 and
Jones et al.2

b The dynamic bladder model with a uniform voiding frequency of 1.5 hours was used. *LLI = lower large intestine; **ULI = upper large intestine

20
2.5 Radiation Safety – Drug Handling 5.1 Radiation Risks
• U se waterproof gloves, effective radiation shielding, Radiation-emitting products, including Fludeoxyglucose F
and appropriate safety measures when handling Flude- 18 Injection, may increase the risk for cancer, especially
oxyglucose F18 Injection to avoid unnecessary radiation in pediatric patients. Use the smallest dose necessary for
exposure to the patient, occupational workers, clinical imaging and ensure safe handling to protect the patient
personnel and other persons. and health care worker [see Dosage and Administration
• R adiopharmaceuticals should be used by or under the (2.5)].
control of physicians who are qualified by specific training
and experience in the safe use and handling of radio- 5.2 Blood Glucose Abnormalities
nuclides, and whose experience and training have been In the oncology and neurology setting, suboptimal imaging
approved by the appropriate governmental agency autho- may occur in patients with inadequately regulated blood
rized to license the use of radionuclides. glucose levels. In these patients, consider medical therapy
• Calculate the final dose from the end of synthesis (EOS) and laboratory testing to assure at least two days of normo-
time using proper radioactive decay factors. Assay the glycemia prior to Fludeoxyglucose F 18 Injection adminis-
final dose in a properly calibrated dose calibrator before tration.
administration to the patient [see Description (11.2)].
• The dose of Fludeoxyglucose F18 used in a given patient 6 ADVERSE REACTIONS
should be minimized consistent with the objectives of Hypersensitivity reactions with pruritus, edema and rash
the procedure, and the nature of the radiation detection have been reported in the post-marketing setting. Have
devices employed. emergency resuscitation equipment and personnel imme-
diately available.
2.6 Drug Preparation and Administration
• Calculate the necessary volume to administer based on 7 DRUG INTERACTIONS
calibration time and dose. The possibility of interactions of Fludeoxyglucose F 18
• Aseptically withdraw Fludeoxyglucose F18 Injection from Injection with other drugs taken by patients undergoing
its container. PET imaging has not been studied.
• Inspect Fludeoxyglucose F18 Injection visually for particu-
late matter and discoloration before administration, 8 USE IN SPECIFIC POPULATIONS
whenever solution and container permit.
• Do not administer the drug if it contains particulate matter 8.1 Pregnancy
or discoloration; dispose of these unacceptable or unused Pregnancy Category C
preparations in a safe manner, in compliance with appli- Animal reproduction studies have not been conducted
cable regulations. with Fludeoxyglucose F 18 Injection. It is also not known
• Use Fludeoxyglucose F 18 Injection within 12 hours from whether Fludeoxyglucose F 18 Injection can cause fetal
the EOS. harm when administered to a pregnant woman or can
affect reproduction capacity. Consider alternative diag-
2.7 Imaging Guidelines nostic tests in a pregnant woman; administer Fludeoxyglu-
• Initiate imaging within 40 minutes following Fludeoxyglu- cose F 18 Injection only if clearly needed.
cose F 18 Injection administration.
• Acquire static emission images 30 to 100 minutes from 8.3 Nursing Mothers
the time of injection. It is not known whether Fludeoxyglucose F 18 Injection is
excreted in human milk. Consider alternative diagnostic
3 DOSAGE FORMS AND STRENGTHS tests in women who are breast-feeding. Use alternatives to
Multiple-dose 30mL and 50mL glass vial containing 0.74 breast feeding (e.g., stored breast milk or infant formula)
to 7.40 GBq/mL (20 to 200 mCi/mL) of Fludeoxyglucose for at least 10 half-lives of radioactive decay, if Fludeoxy-
F 18 Injection and 4.5 mg of sodium chloride with 0.1 to glucose F 18 Injection is administered to a woman who is
0.5% w/w ethanol as a stabilizer (approximately 15 to 50 breast-feeding.
mL volume) for intravenous administration.
8.4 Pediatric Use
4 CONTRAINDICATIONS The safety and effectiveness of Fludeoxyglucose F 18 Injec-
None tion in pediatric patients with epilepsy is established on the
basis of studies in adult and pediatric patients. In pediatric
5 WARNINGS AND PRECAUTIONS patients with epilepsy, the recommended dose is 2.6 mCi.
The optimal dose adjustment on the basis of body size or
weight has not been determined. In the oncology or cardi-
ology settings, the safety and effectiveness of Fludeoxyglu-
cose F 18 Injection have not been established in pediatric
patients.

21
11 DESCRIPTION Table 3. Radiation Attenuation of 511
keV Photons by lead (Pb) shielding
11.1 Chemical Characteristics
Fludeoxyglucose F 18 Injection is a positron emitting Shield thickness Coefficient of
radiopharmaceutical that is used for diagnostic purposes (Pb) mm attenuation
in conjunction with positron emission tomography (PET) 0 0.00
imaging. The active ingredient 2-deoxy-2-[ 18F]fluoro-D- 4 0.50
glucose has the molecular formula of C6H1118FO5 with
a molecular weight of 181.26, and has the following 8 0.25
chemical structure: 13 0.10
26 0.01
39 0.001
52 0.0001

For use in correcting for physical decay of this radionuclide,


the fractions remaining at selected intervals after calibra-
tion are shown in Table 4.

Table 4. Physical Decay Chart for


Fludeoxyglucose F 18 Injection is provided as a ready to
Fluorine F 18
use sterile, pyrogen free, clear, colorless solution. Each
mL contains between 0.740 to 7.40GBq (20.0 to 200 mCi) Minutes Fraction Remaining
of 2-deoxy-2-[18F]fluoro-D-glucose at the EOS, 4.5 mg of 0* 1.000
sodium chloride and 0.1 to 0.5% w/w ethanol as a stabilizer.
15 0.909
The pH of the solution is between 4.5 and 7.5. The solu-
tion is packaged in a multiple-dose glass vial and does not 30 0.826
contain any preservative. 60 0.683
110 0.500
11.2 Physical Characteristics
Fluorine F 18 decays by emitting positron to Oxygen O 16 220 0.250
(stable) and has a physical half-life of 109.7 minutes. The *calibration time
principal photons useful for imaging are the dual 511 keV
gamma photons, that are produced and emitted simultane-
ously in opposite direction when the positron interacts with 12 CLINICAL PHARMACOLOGY
an electron (Table 2).
12.1 Mechanism of Action
Table 2. Principal Radiation Emission Fludeoxyglucose F 18 is a glucose analog that concentrates
Data for Fluorine F 18 in cells that rely upon glucose as an energy source, or in
cells whose dependence on glucose increases under patho-
Radiation/Emission % Per Disintegration Mean Energy physiological conditions. Fludeoxyglucose F 18 is trans-
Positron(ß+) 96.73 249.8 keV ported through the cell membrane by facilitative glucose
Gamma(±)* 193.46 511.0 keV transporter proteins and is phosphorylated within the cell
to [18F] FDG-6-phosphate by the enzyme hexokinase. Once
*Produced by positron annihilation phosphorylated it cannot exit until it is dephosphorylated
From: Kocher, D.C. Radioactive Decay Tables DOE/TIC-I 1026,
89 (1981) by glucose-6-phosphatase. Therefore, within a given tissue
or pathophysiological process, the retention and clearance
The specific gamma ray constant (point source air kerma of Fludeoxyglucose F 18 reflect a balance involving glucose
coefficient) for fluorine F 18 is 5.7 R/hr/mCi (1.35 x 10 -6 transporter, hexokinase and glucose-6-phosphatase activi-
Gy/hr/kBq) at 1 cm. The half-value layer (HVL) for the 511 ties. When allowance is made for the kinetic differences
keV photons is 4 mm lead (Pb). The range of attenuation between glucose and Fludeoxyglucose F 18 transport and
coefficients for this radionuclide as a function of lead phosphorylation (expressed as the “lumped constant” ratio),
shield thickness is shown in Table 3. For example, the Fludeoxyglucose F 18 is used to assess glucose metabolism.
interposition of an 8 mm thickness of Pb, with a coef-
ficient of attenuation of 0.25, will decrease the external In comparison to background activity of the specific organ
radiation by 75 percent. or tissue type, regions of decreased or absent uptake of
Fludeoxyglucose F 18 reflect the decrease or absence of
glucose metabolism. Regions of increased uptake of Flude-
oxyglucose F 18 reflect greater than normal rates of glucose
metabolism.
22
12.2 Pharmacodynamics to Fludeoxyglucose F 18 and would be expected to result
Fludeoxyglucose F 18 Injection is rapidly distributed to all in intracellular formation of 2-deoxy-2-chloro-6-phospho-
organs of the body after intravenous administration. After D-glucose (ClDG-6-phosphate) and 2-deoxy-2-chloro-
background clearance of Fludeoxyglucose F 18 Injection, 6phospho-D-mannose (ClDM-6-phosphate). The phos-
optimal PET imaging is generally achieved between 30 to 40 phorylated deoxyglucose compounds are dephosphorylated
minutes after administration. and the resulting compounds (FDG, FDM, ClDG, and ClDM)
presumably leave cells by passive diffusion. Fludeoxyglu-
In cancer, the cells are generally characterized by enhanced cose F 18 and related compounds are cleared from non-
glucose metabolism partially due to (1) an increase in activity cardiac tissues within 3 to 24 hours after administration.
of glucose transporters, (2) an increased rate of phosphoryla- Clearance from the cardiac tissue may require more than 96
tion activity, (3) a reduction of phosphatase activity or, (4) a hours. Fludeoxyglucose F 18 that is not involved in glucose
dynamic alteration in the balance among all these processes. metabolism in any tissue is then excreted in the urine.
However, glucose metabolism of cancer as reflected by
Fludeoxyglucose F 18 accumulation shows considerable Elimination: Fludeoxyglucose F 18 is cleared from most
variability. Depending on tumor type, stage, and location, tissues within 24 hours and can be eliminated from the
Fludeoxyglucose F 18 accumulation may be increased, body unchanged in the urine. Three elimination phases
normal, or decreased. Also, inflammatory cells can have the have been identified in the reviewed literature. Within 33
same variability of uptake of Fludeoxyglucose F 18. minutes, a mean of 3.9% of the administrated radioactive
dose was measured in the urine. The amount of radiation
In the heart, under normal aerobic conditions, the myocar- exposure of the urinary bladder at two hours post-adminis-
dium meets the bulk of its energy requirements by oxidizing tration suggests that 20.6% (mean) of the radioactive dose
free fatty acids. Most of the exogenous glucose taken was present in the bladder.
up by the myocyte is converted into glycogen. However,
under ischemic conditions, the oxidation of free fatty acids Special Populations: The pharmacokinetics of Fludeoxy-
decreases, exogenous glucose becomes the preferred glucose F 18 Injection have not been studied in renally-
myocardial substrate, glycolysis is stimulated, and glucose impaired, hepatically impaired or pediatric patients. Flude-
taken up by the myocyte is metabolized immediately instead oxyglucose F 18 is eliminated through the renal system.
of being converted into glycogen. Under these conditions, Avoid excessive radiation exposure to this organ system and
phosphorylated Fludeoxyglucose F 18 accumulates in the adjacent tissues.
myocyte and can be detected with PET imaging.
The effects of fasting, varying blood sugar levels, condi-
In the brain, cells normally rely on aerobic metabolism. In tions of glucose intolerance, and diabetes mellitus on
epilepsy, the glucose metabolism varies. Generally, during Fludeoxyglucose F 18 distribution in humans have not been
a seizure, glucose metabolism increases. Interictally, the ascertained [see Warnings and Precautions (5.2)].
seizure focus tends to be hypometabolic.
13 NONCLINICAL TOXICOLOGY
12.3 Pharmacokinetics
Distribution: In four healthy male volunteers, receiving an 13.1 Carcinogenesis, Mutagenesis, Impairment of Fertility
intravenous administration of 30 seconds in duration, the Animal studies have not been performed to evaluate the
arterial blood level profile for Fludeoxyglucose F 18 decayed Fludeoxyglucose F 18 Injection carcinogenic potential,
triexponentially. The effective half-life ranges of the three mutagenic potential or effects on fertility.
phases were 0.2 to 0.3 minutes, 10 to 13 minutes with a
mean and standard deviation (STD) of 11.6 (±) 1.1 min, 14 CLINICAL STUDIES
and 80 to 95 minutes with a mean and STD of 88 (±) 4 min.
14.1 Oncology
Plasma protein binding of Fludeoxyglucose F 18 has not The efficacy of Fludeoxyglucose F 18 Injection in positron
been studied. emission tomography cancer imaging was demonstrated
in 16 independent studies. These studies prospectively
Metabolism: Fludeoxyglucose F 18 is transported into cells evaluated the use of Fludeoxyglucose F 18 in patients with
and phosphorylated to [ 18F]FDG-6- phosphate at a rate suspected or known malignancies, including non-small
proportional to the rate of glucose utilization within that cell lung cancer, colo-rectal, pancreatic, breast, thyroid,
tissue. [F 18]-FDG-6-phosphate presumably is metabolized melanoma, Hodgkin’s and non-Hodgkin’s lymphoma, and
to 2-deoxy-2-[F 18]fluoro-6-phospho-D-mannose([F 18] various types of metastatic cancers to lung, liver, bone, and
FDM-6-phosphate). axillary nodes. All these studies had at least 50 patients and
used pathology as a standard of truth. The Fludeoxyglucose
Fludeoxyglucose F 18 Injection may contain several impuri- F 18 Injection doses in the studies ranged from 200 MBq to
ties (e.g., 2-deoxy-2-chloro-D-glucose (ClDG)). Biodistribu- 740 MBq with a median and mean dose of 370 MBq.
tion and metabolism of ClDG are presumed to be similar

23
In the studies, the diagnostic performance of Fludeoxyglu- 14.3 Neurology
cose F 18 Injection varied with the type of cancer, size of In a prospective, open label trial, Fludeoxyglucose F 18
cancer, and other clinical conditions. False negative and Injection was evaluated in 86 patients with epilepsy. Each
false positive scans were observed. Negative Fludeoxyglu- patient received a dose of Fludeoxyglucose F 18 Injec-
cose F 18 Injection PET scans do not exclude the diagnosis tion in the range of 185 to 370 MBq (5 to 10 mCi). The
of cancer. Positive Fludeoxyglucose F 18 Injection PET mean age was 16.4 years (range: 4 months to 58 years;
scans can not replace pathology to establish a diagnosis of of these, 42 patients were less than 12 years and 16
cancer. Non-malignant conditions such as fungal infections, patients were less than 2 years old). Patients had a known
inflammatory processes and benign tumors have patterns diagnosis of complex partial epilepsy and were under
of increased glucose metabolism that may give rise to evaluation for surgical treatment of their seizure disorder.
false-positive scans. The efficacy of Fludeoxyglucose F 18 Seizure foci had been previously identified on ictal EEGs
Injection PET imaging in cancer screening was not studied. and sphenoidal EEGs. Fludeoxyglucose F 18 Injection PET
imaging confirmed previous diagnostic findings in 16%
14.2 Cardiology (14/87) of the patients; in 34% (30/87) of the patients,
The efficacy of Fludeoxyglucose F 18 Injection for cardiac Fludeoxyglucose F 18 Injection PET images provided new
use was demonstrated in ten independent, prospective findings. In 32% (27/87), imaging with Fludeoxyglucose
studies of patients with coronary artery disease and chronic F 18 Injection was inconclusive. The impact of these
left ventricular systolic dysfunction who were scheduled to imaging findings on clinical outcomes is not known.
undergo coronary revascularization. Before revasculariza-
tion, patients underwent PET imaging with Fludeoxyglucose Several other studies comparing imaging with Flude-
F 18 Injection (74 to 370 MBq, 2 to 10 mCi) and perfusion oxyglucose F 18 Injection results to subsphenoidal EEG,
imaging with other diagnostic radiopharmaceuticals. Doses MRI and/or surgical findings supported the concept that
of Fludeoxyglucose F 18 Injection ranged from 74 to 370 the degree of hypometabolism corresponds to areas of
MBq (2 to 10 mCi). Segmental, left ventricular, wall-motion confirmed epileptogenic foci. The safety and effectiveness
assessments of asynergic areas made before revasculariza- of Fludeoxyglucose F 18 Injection to distinguish idiopathic
tion were compared in a blinded manner to assessments epileptogenic foci from tumors or other brain lesions that
made after successful revascularization to identify myocar- may cause seizures have not been established.
dial segments with functional recovery.
15 REFERENCES
Left ventricular myocardial segments were predicted to
have reversible loss of systolic function if they showed 1. G allagher B.M., Ansari A., Atkins H., Casella V.,
Fludeoxyglucose F 18 accumulation and reduced perfusion Christman D.R., Fowler J.S., Ido T., MacGregor R.R.,
(i.e., flow-metabolism mismatch). Conversely, myocardial Som P., Wan C.N., Wolf A.P., Kuhl D.E., and Reivich M.
segments were predicted to have irreversible loss of systolic “Radiopharmaceuticals XXVII. 18F-labeled 2-deoxy-
function if they showed reductions in both Fludeoxyglucose 2-fluoro-d-glucose as a radiopharmaceutical for
F 18 accumulation and perfusion (i.e., matched defects). measuring regional myocardial glucose metabolism
in vivo: tissue distribution and imaging studies in
Findings of flow-metabolism mismatch in a myocardial animals,” J Nucl Med, 1977; 18, 990-6.
segment may suggest that successful revascularization 2. Jones S.C., Alavi, A., Christman D., Montanez, I., Wolf,
will restore myocardial function in that segment. However, A.P., and Reivich M. “The radiation dosimetry of 2
false-positive tests occur regularly, and the decision to have [F-18] fluoro-2-deoxy-D-glucose in man,” J Nucl Med,
a patient undergo revascularization should not be based 1982; 23, 613-617.
on PET findings alone. Similarly, findings of a matched 3. Kocher, D.C. “Radioactive Decay Tables: A handbook of
defect in a myocardial segment may suggest that myocar- decay data for application to radiation dosimetry and
dial function will not recover in that segment, even if it is radiological assessments,” 1981, DOE/TIC-I 1026, 89.
successfully revascularized. However, false-negative tests 4. ICRP Publication 53, Volume 18, No. l-4,1987, pages
occur regularly, and the decision to recommend against 75-76.
coronary revascularization, or to recommend a cardiac
transplant, should not be based on PET findings alone. The
reversibility of segmental dysfunction as predicted with
Fludeoxyglucose F 18 PET imaging depends on successful
coronary revascularization. Therefore, in patients with a
low likelihood of successful revascularization, the diag-
nostic usefulness of PET imaging with Fludeoxyglucose F
18 Injection is more limited.

24
16 HOW SUPPLIED/STORAGE AND DRUG HANDLING

Fludeoxyglucose F 18 Injection is supplied in a multi-dose,


capped 30 mL and 50 mL glass vial containing between
0.740 to 7.40GBq/mL (20 to 200 mCi/mL), of no carrier
added 2deoxy-2-[F 18] fluoro-D-glucose, at end of synthesis,
in approximately 15 to 50 mL. The contents of each vial are
sterile, pyrogen-free and preservative-free.

NDC 40028-511-30; 40028-511-50

Receipt, transfer, handling, possession, or use of this


product is subject to the radioactive material regulations
and licensing requirements of the U.S. Nuclear Regula-
tory Commission, Agreement States or Licensing States as
appropriate.

Store the Fludeoxyglucose F 18 Injection vial upright


in a lead shielded container at 25°C (77°F); excursions
permitted to 15-30°C (59-86°F).

Store and dispose of Fludeoxyglucose F 18 Injection in


accordance with the regulations and a general license, or
its equivalent, of an Agreement State or a Licensing State.

The expiration date and time are provided on the container


label. Use Fludeoxyglucose F 18 Injection within 12 hours
from the EOS time.

17 PATIENT COUNSELING INFORMATION

Instruct patients in procedures that increase renal clearance


of radioactivity. Encourage patients to:
• drink water or other fluids (as tolerated) in the 4 hours
before their PET study.
• void as soon as the imaging study is completed and as
often as possible thereafter for at least one hour.

Manufactured by: PETNET Solutions Inc.


810 Innovation Drive
Knoxville, TN 37932

Distributed by: PETNET Solutions Inc.


810 Innovation Drive
Knoxville, TN 37932

PN0002262 Rev. A
March 1, 2011

25
Trademarks and service marks used in this
material are property of Siemens Medical
Solutions USA or Siemens AG.

Siemens Medical Solutions USA, Inc. Global Business Unit Address


© 2012 Siemens Medical Solutions USA, Inc.
All rights reserved. Siemens Medical Solutions USA, Inc.
Molecular Imaging
All photographs © 2012 Siemens Medical 2501 N. Barrington Road
Solutions, USA, Inc. All rights reserved. Hoffman Estates, IL 60192-2061
USA
Note: Original images always lose a certain Telephone: +1 847 304 7700
amount of detail when reproduced. www.siemens.com/mi

Global Siemens Headquarters Global Siemens Headquarters Address of legal manufacturer


Healthcare Headquarters
Siemens AG Siemens Medical Solutions USA, Inc.
Wittelsbacherplatz 2 Siemens AG Molecular Imaging
80333 Munich Healthcare Sector 810 Innovation Drive
Germany Henkestrasse 127 Knoxville, TN 37932-9702
91052 Erlangen USA
Germany Telephone: +1 865 218 2000
Telephone: +49 9131 84-0 www.siemens.com/mi
www.siemens.com/healthcare

www.siemens.com/mi

Order No. A91MI-10117-1T-7600 | Printed in USA | All rights reserved | MI-216.NW/TM.JV.TW.1M | © 2.2012 Siemens Medical Solutions USA, Inc.

You might also like