Professional Documents
Culture Documents
Saep 150
Saep 150
Contents
1 Introduction ........................................................ 2
2 Conflicts and Deviations .................................... 2
3 Applicable Documents ....................................... 2
4 Definitions .......................................................... 3
5 Analyzer Accuracy Assessment ........................ 3
6 Analyzer-Transducer Accuracy Assessment ..... 4
Revision Summary................................................... 7
1 Introduction
1.1 Purpose
1.2 Scope
The scope of this procedure is to provide a method that can be used to ensure
acceptable accuracy and functionality of company owned spectrum analyzers,
sensors and associated cables. This document presents a two-stage procedure
which is essential to ensure vibration data collection accuracy. The two main
checks, both of which must be performed, are:
Accuracy and functionality assessment of all types of vibration spectrum
analyzers.
Accuracy and functionality assessment of spectrum analyzers in conjunction
with seismic sensors and cables.
Note: In case the spectrum analyzer found to be out of calibration, it must be sent to
the vendor or certified third party to perform actual calibration. Record of the
calibration test date shall be saved for future reference, a test label shall be
placed at the back of the analyzer indicating the calibration check date and final
results.
2.1 Any conflicts between this document and other applicable Mandatory Saudi
Aramco Engineering Requirements (MSAERs) shall be addressed to the EK&RD
Coordinator.
2.2 Any deviation from the requirements herein shall follow internal company
procedure SAEP-302.
3 Applicable Documents
4 Definitions
Amplifier: An electronic device that increases the voltage, current, or power of a signal.
Analyzer or Data Collector: A device that uses the Fast Fourier Transform (FFT)
algorithm to calculate a spectrum from a time domain signal.
Calibration: A test during which known values of the measured variables are applied
to the transducer or readout instrument and corresponding output readings are verified
and adjusted as necessary.
Frequency Generator: A device used to test electronic equipment by generating a
signal whose frequency, wave shape, and amplitude are independently adjustable over a
wide range of settings.
Linearity: (Calibration) The closeness of a calibration curve to a specified straight
line, expressed as the maximum deviation of any calibration point on a specified
straight line.
Shaker: An electromechanically device that provides a controllable vibration motion
for testing applications.
Vibration Transducer: Device that translates measured vibratory motion into an
electrical signal. Three of the most common transducers used in vibration
measurements are accelerometer, velocity transducer, and displacement probe.
Page 3 of 13
©Saudi Aramco 2019. All rights reserved.
Document Responsibility: Vibration and Lubrication Standards Committee SAEP-150
Issue Date: 7 August 2019
Next Planned Update: 16 February 2022 Vibration Spectrum Analyzer and Sensors Accuracy Assessment
This section sets a procedure to check the accuracy of the data collector using an
independent signal generator. The signal generator will be connected directly to
the analyzer as shown in Figure 1 (Appendix A). The signal generator is set to a
known constant amplitude, and the frequency can be varied so that the signal
input to the analyzer is recorded at each frequency step. The results deviation
shall be calculated for both frequency and amplitude, the maximum allowable
deviation is ±5% (ISO 2954 5.6).
The equipment used for the analyzer calibration consists of frequency generator,
connection cable, and the analyzer. Figure 1 (Appendix A) shows the
equipment connection layout.
Analyzer Settings
Frequency Range Amplitude Range No. of Spectral Lines
2, 5, 10, 20, 40, 80, 160, 500, 1,000, 2,000, 4,000 Hz 1.0 Volt RMS
Page 4 of 13
©Saudi Aramco 2019. All rights reserved.
Document Responsibility: Vibration and Lubrication Standards Committee SAEP-150
Issue Date: 7 August 2019
Next Planned Update: 16 February 2022 Vibration Spectrum Analyzer and Sensors Accuracy Assessment
This section sets a procedure to check the accuracy of the data collector and
associated transducer when using a shaker. The equipment used for the
analyzer-transducer assessment consists of:
a) Frequency generator
b) Analyzer
c) Reference transducer
d) Actual transducer to be assessed (accelerometer or velometer)
e) Shaker
f) Amplifier
g) Connecting cables as shown in Figure 2 (Appendix A).
The shaker produces vibration movement with a known value measured using
a reference (i.e., accurately-calibrated) accelerometer or velocity transducer.
A signal generator is connected to the shaker via a power amplifier (Figure 2 of
Appendix A) and data from the reference and actual transducers is stored for
subsequent analysis. The percentage deviation in amplitude between the two
data sets is calculated and the maximum allowable deviation is set at ±10%.
Analyzer and frequency-generator settings values for the linearity and frequency
response tests will depend upon the transducer type as shown in Tables 2 to 7
below.
The settings shown in Table 2 apply for both the linearity and
frequency-response tests.
Table 2 - Spectrum Analyzer Settings
Page 5 of 13
©Saudi Aramco 2019. All rights reserved.
Document Responsibility: Vibration and Lubrication Standards Committee SAEP-150
Issue Date: 7 August 2019
Next Planned Update: 16 February 2022 Vibration Spectrum Analyzer and Sensors Accuracy Assessment
The settings shown in Table 5 apply for both the linearity and
frequency-response tests.
Page 6 of 13
©Saudi Aramco 2019. All rights reserved.
Document Responsibility: Vibration and Lubrication Standards Committee SAEP-150
Issue Date: 7 August 2019
Next Planned Update: 16 February 2022 Vibration Spectrum Analyzer and Sensors Accuracy Assessment
Revision Summary
16 February 2017 New Saudi Aramco Engineering Procedure that established a method that can be used to
ensure acceptable accuracy and functionality of data collectors, sensors and associated
cables. This document presents a two-stage procedure which is essential to ensure vibration
data collection accuracy.
07 August 2019 Editorial revision to:
- Change the revision cycle from three years to five years
- Modify section 2 “Conflicts and Deviations” to comply with SAEP-301
- Reflect document’s references in the proper section(s)/paragraph(s) to comply with
SAEP-301
- Remove references not referred to within the content
Page 7 of 13
©Saudi Aramco 2019. All rights reserved.
Document Responsibility: Vibration and Lubrication Standards Committee SAEP-150
Issue Date: 7 August 2019
Next Planned Update: 16 February 2022 Vibration Spectrum Analyzer and Sensors Accuracy Assessment
SIGNAL
GENERATOR
DATA
COLLECTOR
POWER
SUPPLY
Transducer
TO BE
ASSESSED REFRENCE
(ACCURATELY
CALIBRATED)
VELOMITOR DATA
COLLECTOR
AMPLIFIER SIGNAL
GENERATOR
SHAKER
Page 8 of 13
©Saudi Aramco 2019. All rights reserved.
Document Responsibility: Vibration and Lubrication Standards Committee SAEP-150
Issue Date: 7 August 2019
Next Planned Update: 16 February 2022 Vibration Spectrum Analyzer and Sensors Accuracy Assessment
Appendix A (Continued)
Example of Data-Collector/Analyzer Accuracy Assessment
Table 8 below shows example of an analyzer accuracy check (Amplitude and Frequency).
The frequency is varied and the amplitude is observed for deviations. The maximum % deviation
in frequency is 0.0% and amplitude is 2.9% which meets the ±5% acceptable deviation.
2 2 0% 0.971 -2.9
5 5 0% 1.001 0.10
10 10 0% 1.004 0.40
20 20 0% 1.007 0.70
40 40 0% 1.007 0.70
80 80 0% 1.007 0.70
Page 9 of 13
©Saudi Aramco 2019. All rights reserved.
Document Responsibility: Vibration and Lubrication Standards Committee SAEP-150
Issue Date: 7 August 2019
Next Planned Update: 16 February 2022 Vibration Spectrum Analyzer and Sensors Accuracy Assessment
Appendix A (Continued)
Page 10 of 13
©Saudi Aramco 2019. All rights reserved.
Document Responsibility: Vibration and Lubrication Standards Committee SAEP-150
Issue Date: 7 August 2019
Next Planned Update: 16 February 2022 Vibration Spectrum Analyzer and Sensors Accuracy Assessment
The procedures outlined in Section 5 were applied in laboratory for the cases of
Analyzer-Accelerometer and Analyzer-velometer. The results are presented
below.
Page 11 of 13
©Saudi Aramco 2019. All rights reserved.
Document Responsibility: Vibration and Lubrication Standards Committee SAEP-150
Issue Date: 7 August 2019
Next Planned Update: 16 February 2022 Vibration Spectrum Analyzer and Sensors Accuracy Assessment
Accelerometer
Frequency Response Check
1.02
1.015
1.01
1.005
0.995
0.99
100 200 500 1K 2K 4K
Hz
Reference Actual Level
Page 12 of 13
©Saudi Aramco 2019. All rights reserved.
Document Responsibility: Vibration and Lubrication Standards Committee SAEP-150
Issue Date: 7 August 2019
Next Planned Update: 16 February 2022 Vibration Spectrum Analyzer and Sensors Accuracy Assessment
0.25
0.2
0.15
0.1
30 50 70 100 200 400 1K
Frequency (Hz)
Page 13 of 13
©Saudi Aramco 2019. All rights reserved.