User Traing 2021

You might also like

Download as pdf or txt
Download as pdf or txt
You are on page 1of 20

1.

Products description
2.Components description
3.Maintence and precautions
4.FAQ
Control and display output temperature, safety control.

Alarm
Display
temp

Set
temp

Up/down
temp

Heating mode
select /confirm
Control panel
temp
Hook to table
side rail
1.Installation

Hook on table side rail

Clamp on Iv pole
Connect mattress/blanket

No arrow direction
Pay attention!
2.Plug in and turn on it

Turn on
switch2
Height Weight
Model Length(mm) Width(mm)
(mm) (kg)

Control unit Warm 6100 One channel control unit


Warm 5190(Standard,
1900 520 30 3.8
operating table)
Warm 5120(child) 1200 520 30 2.6
Mattress
Warm 5100(special
operation, like 1070 520 30 2.3
gynecology, etc)

Control unit can work with any one of the mattress.


Release air if needed

Operate manual valve like:


Warm6100 control unit can work
with any one of the blanket.

Length
Model Width ( mm )
Height(mm)Weight(kg)
(mm)

Warm 8160 full body


blanket,before and after
1660 850 15 2.3
operation, ICU, inhospital
transfer.
Warm 8180 shoulder
blanket, special
Blankets 1800 400 15 1.3
operation, covering
shoulders.

Warm 8100regional blanket,


child blanket, special 1000 700 15 1.3
operation.
Accessories

3m power
2A Fuse cable
(1)The user should read user manual carefully before
operation;
(2)clean and disinfect product with 75% alcohol;
(3)Acid and strong alkali detergent is not allowed;
(4)Pay attention to the plugs of control unit and
mattress when plugging;
(5)Theory working life of product is 10 years;
(6)Avoid covering small parts of the mattress or
folding mattress in the operation process;
(7)Pay attention to the right patient direction and
be sure that the arrow area is covered by the
patient.
 (8)Reset new temperature but not realize:
after reset to the new target temperature,
be sure to press function key to confirm
it.
 (9) Mattress/blanket aviation plug to
control unit should be done exactly the
same with guide, match the direction arrow
and plug in.Do not rotate it.
(1)Small scratches on mattress will not effect usage
or damage to user and patient;
(2)Product working failure caused by any program ,
hardware will not lead to burn to patient, since there
are many preventing measures;
(3)Ultraviolet light disinfection is not recommended,
or it will effect the mattress working life;
(4)Manual mode and auto mode can be considered as
MT and AT of heating mode;
(5)The product is not home electric blanket;the
working theory, accuracy, raw material and art are
totally different.
Service engineer:Han Zhao
Tel:+86-10-52895298
Web:www.eternal-med.com
Email:pur@eternl-med.com
export@eternal-med.com
hanzhao0313@sina.com

You might also like