IDL-11098-01-A, USB - RS232 Driver Installation

You might also like

Download as pdf or txt
Download as pdf or txt
You are on page 1of 8

IDL-11098-01-A

I.S.E. S.r.l.
Via delle Driadi, 45 – 00133 Roma Rev. A del 26/02/13

pag. 1 / 8
USB-RS232 Driver installation

USB-RS232 Driver installation

Autoanalyzers Miura family

Il contenuto di questo documento è di esclusiva proprietà della I.S.E. s.r.l.


Sono pertanto vietate distribuzioni e fotocopie non espressamente autorizzate di tutto o di parte del contenuto.
I.S.E. s.r.l. tutela i propri diritti a norma di legge.
IDL-11098-01-A
I.S.E. S.r.l.
Via delle Driadi, 45 – 00133 Roma Rev. A del 26/02/13

pag. 2 / 8
USB-RS232 Driver installation

Document Identification

Emmision: Rev. A
Codifica: IDL-11098-01
Date: 26/03/2013
Id. file: IDL-11098-01-A, USB-RS232 Driver installation.doc
Rating:

Modification register

Rev. Date Description


A 26/02/2013 First release
B
C
D
E
F
G

Author and Authorizations

Name Date Signature


Author Ruggeri Claudio 26/02/2013

This document is property of I.S.E. s.r.l.


Unauthorized copies or alterations of its contents could cause legal actions in order
to preserve owner’s interests.
IDL-11098-01-A
I.S.E. S.r.l.
Via delle Driadi, 45 – 00133 Roma Rev. A del 26/02/13

pag. 3 / 8
USB-RS232 Driver installation

LIST OF THE DEFINITIONS AND THE ABBREVIATIONS

AD Applicable Document
CV Coefficient of Variation
FW Firmware
HW Hardware
SD Standard Deviation
SW Software
TBC To Be Confirmed
TBD To Be Defined
TBV To Be Verified
DUT Device Under Test
N.A. Not Applicable
RD Reference document
SDT Test station
SW Software
Ver Version
Pc Personal computer

This document is property of I.S.E. s.r.l.


Unauthorized copies or alterations of its contents could cause legal actions in order
to preserve owner’s interests.
IDL-11098-01-A
I.S.E. S.r.l.
Via delle Driadi, 45 – 00133 Roma Rev. A del 26/02/13

pag. 4 / 8
USB-RS232 Driver installation

Index of the contents


Index of the contents ....................................................................................................................................... 4
1. Purpose and Applicability....................................................................................................................... 5
2. Documentation ........................................................................................................................................ 5
2.1. Applicable documents .................................................................................................................. 5
2.2. Reference Documents ................................................................................................................... 5
3. USB-RS232 converter ................................................................................................................................ 6
3.1. Driver installation ............................................................................................................................. 6
4. Connectivity check ................................................................................................................................. 8

This document is property of I.S.E. s.r.l.


Unauthorized copies or alterations of its contents could cause legal actions in order
to preserve owner’s interests.
IDL-11098-01-A
I.S.E. S.r.l.
Via delle Driadi, 45 – 00133 Roma Rev. A del 26/02/13

pag. 5 / 8
USB-RS232 Driver installation

1. Purpose and Applicability


The purpose of this document is to describe the procedure of installation of the FTDI drivers for the
USB-RS-232 converter on a pc running Windows as operative system
This procedure has been tested on Windows 7 64 bit, Home Premium.
This procedure has to be executed only in case you have to reinstall the operative system (or in
case a pc is not provided with the instrument), else the device drivers are installed at the factory.
This document applies to Miura (single and double Arms), Miura 200 and Miura One.

2. Documentation

2.1. Applicable documents


The following documents are considered as applicable:
AD1 -

2.2. Reference Documents


The following documents are of reference:
RD1 -

This document is property of I.S.E. s.r.l.


Unauthorized copies or alterations of its contents could cause legal actions in order
to preserve owner’s interests.
IDL-11098-01-A
I.S.E. S.r.l.
Via delle Driadi, 45 – 00133 Roma Rev. A del 26/02/13

pag. 6 / 8
USB-RS232 Driver installation

3. USB-RS232 converter
Miura family instruments are now equipped with an USB-RS-232 converter which requires a specific
driver to be installed on the operative system.
In order to correctly install it, see the chapter below before to connect the instrument to the pc.

3.1. Driver installation

1. Select the installation package called CDM20824_Setup.exe


2. Click on the right mouse button and choose “Run as Administrator”.
3. The command prompt window should appear to notify the installation progress. (On some
system has been observed that the point number 2 has to be executed twice)
4. At the end of the installation process, connect the instrument to the pc using the USB cable
provided.
5. Wait that the system recognizes the new hardware and declares it as “Ready to use”
6. Go to Control Panel -> Hardware and Sound -> Device Manager, as shown in figure 1

Fig. 1 – Control Panel – Device Manager

7. Select “Ports (COM & LPT)”


8. Select USB Serial Port (COMx) as shown in figure 2

This document is property of I.S.E. s.r.l.


Unauthorized copies or alterations of its contents could cause legal actions in order
to preserve owner’s interests.
IDL-11098-01-A
I.S.E. S.r.l.
Via delle Driadi, 45 – 00133 Roma Rev. A del 26/02/13

pag. 7 / 8
USB-RS232 Driver installation

9. Click on the right mouse button and select “Properties”


10. Choose “Port settings” tab of the shown window
11. Click on “Advanced” button
12. Assign, if free, the COM number 1, selecting it on the list box instead of the default one
(usually COM3, as shown in the figure 3)

This document is property of I.S.E. s.r.l.


Unauthorized copies or alterations of its contents could cause legal actions in order
to preserve owner’s interests.
IDL-11098-01-A
I.S.E. S.r.l.
Via delle Driadi, 45 – 00133 Roma Rev. A del 26/02/13

pag. 8 / 8
USB-RS232 Driver installation

13. Click on the “OK” button


14. Click on the “OK” button
15. Restart the system

4. Connectivity check

After the reboot of the operative system, you should check the connectivity among the Miura
software and the instrument.
Close instrument cover, then start the software and wait for the initial hardware reset.
In case of an alarm belonging a poor connectivity through the serial cable, go to “System Config”
menu and select the correct serial port number, then click on “Save” button before restarting the
software.

This document is property of I.S.E. s.r.l.


Unauthorized copies or alterations of its contents could cause legal actions in order
to preserve owner’s interests.