Specification: S5

SHENZHEN COMEN MEDICAL INSTRUMENTS CO.,LTD


No.2 of FIYATA Timepiece Building, Nanhuan Avenue, Gongming
sub-district, Guangming District, Shenzhen, P.R. China
Tel: +86-755-26408879
Fax:+86-755-26431232
Email: info@szcomen.com
Web: www.comen.com

2020 V 1.1
Defibrillator/ Monitor
S5

Standard Configuration: Defib Mode: 210 times (360J


Manual defibrillation, AED, 3/5-lead ECG, RESP, Thermal charge at intervals of 1minute
Recorder without recording);
Optional: Pacing Mode: 4.5 hours (Load:50
Pacer, NIBP, PR, EtCO2, SpO2 Ω, frequency: 80bpm, current:
Safety Standards: 60mA, without recording)
5000mAh Battery:
Monitoring Mode: no less than 4
Physical Characteristics
Size: 295mm×252mm×316mm hours

Weight 5.6kg (Including 1 Defib Mode: 120times (360J charge

battery);5.384(Main unit) at intervals of 1minute without

Screen Size: 7” TFT screen recording);

Resolution 800 × 480 Pacing Mode: 3hours (Load:50Ω,

Waveforms: Max 4 waveforms frequency: 80bpm, current: 60mA,


without recording)
Operation Environment
Temperature: Brightness: Manual from X to 100, X refers to
0~45℃ the darkest brightness (X is 10 by
Humidity: 10%~95%, non-condensation
Atmosphere Pressure: 700hPa~1060hPa default)

Ingress Protection: IP44 Indicator


Power requirement: 100-240V~, 50/60Hz±3Hz Two alarm indicators

Battery type: Rechargeable Lithium-ion battery Power indicator

Battery capacity: 7500mAh, d.c.14.8V Battery indicator

5000mAh, d.c.14.8V Maintain indicator

Battery number: 1 QRS beep and alarm sound

Battery recharging 7500mAh Battery: Less than 2 Operating key sound

Time: hours to 80% and less than 3 hours Interfacing


to 100% with equipment power off USB interface
5000mAh Battery: Less than 1.5 RJ45 interface
hours to 80% and less than 2.5 AC power input
hours to 100% with equipment Multi-functional connector
power off Date storage
Battery backup: 7500mAh Battery: Alarm Event: 200 groups
Monitoring Mode: no less than 6 Patient profiles: 100 groups
hours Patient Events: 1000 groups
Wave Review: 10min
NIBP Review: 2000 groups
Trend Graph: 160 hours ±2J or 15% of setting, whichever is
Trend Table: 160 hours greater, while 25Ω, 75Ω, 100Ω,
Voice recording: Max 240 min in total; 125Ω, 150Ω, 175Ω impedance
(Up to 60 min for each patient) Patient Impedance 20~300Ω(External defibrillation);
Marked events Available Range:
Power-off storage: Yes Defibrillation proof: Type CF: ECG, RESP, SpO2, NIBP,
Alarm: User-adjustable High and Low 3- PR;
level Limits; Type BF: EtCO2
Prioritized audible and visual alarm Manual Mode
Network: Connected to Central Monitoring External defibrillators: 1J~360J, 25 types
System by hardwire/wireless (1/2/3/4/5/6/7/8/9/10/15/20/30/5
Recorder 0/70/100/120/150/170/200/220/2
Type: Built-in; Thermal array 50/270/300/360J)
Channel: Max 3 channel waveforms Synchronous Energy transfer begins within 60ms
Real-time recording: 3s, 5s, 8s, 16s, 32s, Continual Cardioversion: of the R wave from internal Sync
Speed: 6.25mm/s, 12.5mm/s, 25mm/s, signal
50mm/s Energy transfer begins within 25ms
Record width: 50mm of the External Sync signal
Resolution: 8dot/mm (Horizontal and vertical) AED
Output Energy: Adujustable:100-360J
Background grid: Configurable Number of electric Adjustable: once, twice, 3 times
External printer: Yes shocks
Types can be AED: VF & VT
Defibrillation AED maximum time Battery power supply: 18s
Operating mode: Manual Mode, AED Mode, , required for cardiac AC power supply: 21s
Synchronous Defibrillation rhythm analysis to be
Waveform: Biphasic truncated exponential ready for discharge:
waveform, with impedance Noninvasive Pacing
compensation Waveform: Monophasic square wave pulse
Defibrillation pathway: External defibrillation Pulse Width: 20ms or 40ms
Electrode type: External defibrillation paddles, Accuracy: ±5%
multifunctional electrode Pacing Mode: On-demand or fixed
External defibrillation Supports charging, discharging and Pacing frequency: 30 ppm to 210 ppm
electrode paddles: energy selection; Charging Accuracy: ±1ppm or ±1.5% (whichever is
completion indicator greater)
Charge Time: Less than 3 seconds to 200 Joules Pacing output: 0 mA to 200 mA
(Battery power) with a new, fully charged battery Accuracy: ±5% or ±5mA, whichever is greater
Less than7 seconds to 360 Joules Speed-down pacing: Pacing pulse frequency reduced to
with a new, fully charged battery 25% of original value.
Charge Time: Less than4 seconds to 200 Joules; Monitoring
(AC power) Less than 8 seconds to 360 Joules ECG (leads)
Energy accuracy: ±1.5J or ±10% of setting, whichever Lead Type: 3 leads ECG, 5 leads ECG, AUTO
is greater, while 50Ω impedance Lead selection: 5-lead: I; II; III; aVR; aVL; aVF; V
3-lead: I; II; III
Multi-lead Bandwidth: 0.5Hz~40Hz
synchronization Available Delay: ≤35ms
analysis: ST Detection: -2.0mV~+2.0mV (-
ECG sensitivity: Auto, 1.25 mm/mV (×0.125), 20.0mm~+20.0mm)
2.5 mm/mV (×0.25), 5 mm/mV Resolution: 0.01mV
(×0.5), 10 mm/mV (×1), Accuracy: -0.8mV ~ +0.8mV: ±0.02mV or
20 mm/mV (×2), 40 mm/mV (×4), ±10%;
Accuracy: Less than ±5% Others: Unspecified
Sweep speed: 6.25 mm/s, 12.5 mm/s, 25 mm/s, ST analysis review 20 groups
50 mm/s System noise: Less than 25μV
Accuracy: Less than ±10% Calibration voltage 1 mV; Accuracy: ±5%
Heart Rate: Adult: 15~300bpm Arrhythmia Analysis: 26 Types
Pediatric:15~350bpm Pacemaker detection: Detectable
Accuracy: ±1bpm or ± ECG (paddle)
1%(whichever is greater) Lead Type: Single lead ECG
Alarm limit range Adult: Heart Rate Adult: 15~300bpm
High limit: (low limit+2bpm) ~ measurement & alarm Pediatric:15~350bpm
300bpm range:
Low limit: 15bpm~ (high limit- Resolution: 1 bpm
2bpm) Accuracy: ±1% or ±1bpm (whichever is
Pediatric: greater)
High limit:(low limit+2bpm) ~ Bandwidth: Defib: 1~20Hz (-3dB~+0.4dB)
350bpm CMRR: Defib: >105dB
Low limit: 15bpm~(high limit- Input Impedance: ≥5MΩ
2bpm) Input signal range: ±8mV
Resolution: 1 bpm HR trigger value 200μV
Accuracy: ±1bpm Arrhythmia Analysis: 5 Types, ASY, VF, VT, PNC, and PNP
Bandwidth: Monitoring: 0.5~40Hz (- Respiration
3.0dB~+0.4dB) Method: Thoracic Impedance Method
Diagnosis: 0.05~150Hz (- RR measurement Adult: 0~120bpm
3.0dB~+0.4dB) range: Pediatric: 0 ~150bpm
Surgery: 1~20Hz (-3.0dB~+0.4dB) Accuracy: 7~150bpm: ±2bpm or ±2%
ST: 0.05~40Hz(-3.0dB~+0.4dB) (whichever is greater)
CMRR: Monitoring: >105dB 0~6bpm: unspecified
Diagnosis: >90dB Apnea Alarm: Adult: 10s~60s Ped: 10s~40s
Surgery: >105dB Accuracy: ±5s
ST: >105dB Alarm: Audible and visual alarm; alarm
Input Impedance: ≥5MΩ events reviewable
Input signal range: ±8mV COMEN NIBP
HR trigger threshold 200μV Method Automatic oscillometric
Lead off detection Measuring electrode: <0.1μV Work mode: Manual / Automatic/Continuous
current: Driving electrode: <1μV Interval Time: Adjustable
Pacemaker pulse Manual selection when the 1/2/2.5/3/4/5/10/15/30/60/90/12
suppression switch: pacemaker is turned on 0/180/240/480/720 min
Analog output: Magnification: 1:1000; Continuous: 5min
Accuracy: ±5%
Maximum Adu/Ped: 120s Pediatric: 5.3~21.3kPa
measurement cycle (40~160mmHg)
Measurement Unit: mmHg / kPa selectable Range of diastolic Adult: 2.7~26.7kPa (20~200mmHg)
Pressure types: Systolic, Diastolic, Mean pressure: Pediatric: 2.7~16kPa
Range of systolic Adult Mode: 5.3~36kPa (20~120mmHg)
pressure: (40~270mmHg) Range of mean Adult:3.5~29.3kPa (26~220mmHg)
Pediatric Mode: 5.3~26.7kPa pressure: Pediatric: 3.5~17.7kPa
(40~200mmHg) (26~133mmHg)
Range of diastolic Adult Mode:1.3~28.7kPa PR from NIBP 30~220bpm
pressure: (10~215mmHg) Accuracy: ±2% or ±3bpm, whichever is
Pediatric Mode: 1.3~20kPa greater
(10~150mmHg) Nellcor SpO2
Range of mean Adult Mode: 2.7~31.3kPa Measurement range: 0~100%
pressure: (20~235mmHg) Resolution: 1%
Pediatric Mode: 2.7~22kPa Accuracy: ±2% (70~100%, Adu/Ped, non-
(20~165mmHg) motion)
Over pressure Adult: 39.6kPa (297mmHg) 1~69% unspecified
protection: Pediatric: 32kPa (240mmHg) Alarm range: 20~100%
Tolerance: ±0.4kPa (±3mmHg) PR Measurement
Accuracy: ±±0.667kPa (±5mmHg), if Range: 20~300bpm
exceeds the above range, the Resolution: 1bpm
monitor can still display normally, Accuracy: ±3bpm (20~250bpm)
but the accuracy is not considered Unspecified (251~300bpm)
Alarm range: 20~350bpm
Alarm limit: Same as the range of measurement MASIMO SpO2
PR from NIBP: 40~240bpm Measurement & alarm
Resolution: 1bpm range 1~100%
Accuracy: ±3% or ±3bpm, whichever is Resolution: 1%
greater Accuracy: ±2% (70~100%, Ped/Adu, non-
SunTech NIBP motion)
Regulatory YY 0670-2008 ±3% (70~100%, motion);
compliance: 1~69% unspecified
Initial inflation range: Adult: 16~37.3kPa Alarm range 1~100%
(120~280mmHg) PR Measurement
Pediatric: 10.7~22.7kPa Range 25~240bpm
(80~170mmHg) Resolution: 1bpm
Maximum Adult: 130s Accuracy: ±3bpm (non-motion)
measurement cycle: Pediatric: 90s ±5bpm (motion);
Over pressure Adult/Pediatric: 40.0kPa Alarm range: 20~350bpm
protection: (300mmHg) PI value: 0.02~20%
Static pressure 0kPa~40.0kPa (0mmHg~300mmHg) Resolution: 0.01% (0.02~9.99%)
measurement range: 0.1% (10~20%)
Resolution: ±0.4kPa (±3mmHg) SIQ: Available
Range of systolic Adult: 5.3~34.7kPa (40~260mmHg) COMEN SpO2
pressure: 0~100%
Measurement & alarm CapnoTrak:
range: 0~99mmHg, 0~13.03%, 0~13.2kPa
Resolution: 1% (at 760mmHg)
Accuracy: ±2% (70~100%, Ped/Adu, non- Accuracy: Loflow:
motion) ± 2mmHg (0~40mmHg)
0~69% unspecified ± 5% of reading (41 – 70mmHg)
PR Measurement ± 8% of reading (71 –100mmHg)
Range: 20~254bpm ±10% of reading (101~150mmHg)
Resolution: 1bpm (In 25℃, if RR>80rpm, accuracy is
Accuracy: ±2bpm 12% of reading)
Alarm range: 20~350bpm CapnoTrak:
PI value: 0.05~20% ±2mmHg (0~38mmHg)
Resolution: 0.01% (0.05%~9.99%) ±10% of reading (38~99mmHg)
0.1% (10.0%~20.0%) RR influence to EtCO2
Accuracy: unspecified (0~99mmHg):
SIQ: Available -2~0.5mmHg (0-40bpm)
MASIMO EtCO2 (Sidestream) (-6% of reading)~0.5mmHg (41-
Measurement range: 0~190mmHg, 0~25vol% 70bpm)
(at 760mmHg) (-14% of reading)~0.5mmHg
Accuracy: Standard environment 22±5℃, (71~100bpm)
1013±40kPa: Resolution: 1mmHg
a) 0~114mmHg: Loflow: 2~150rpm
awRR range
±(1.52mmHg+reading×2%) CapnoTrak: 0, 2~100rpm
b) 114~190mmHg: not defined awRR accuracy: ±1rpm
All environment:
a) 0~114mmHg:
±(2.25mmHg+reading×4%)
b) 114~190mmHg: not defined
Resolution: 1mmHg or 0.1% or 0.1kPa
awRR range: 0~150rpm
awRR accuracy: ±1rpm

Response time: <3 s

Respironics EtCO2 (Sidestream)


Measurement range: Loflow:
0~150mmHg, 0~19.7%, (0~20kPa)
(at 760mmHg)

*Notice: Specifications subject to changes without prior


notice. All rights reserved by COMEN