Professional Documents
Culture Documents
Lim - Asme V
Lim - Asme V
The highlighted points are as per outline in the body of knowledge API 510 Pressure Vessel Inspector
Certification examination issued September 2022, January 2023, and May 2023 (Replaces May 2021)
ASME V
NONDESTRUCTIVE EXAMINATION
ARTICLE 1 GENERAL
REQUIREMENTS
T-120 GENERAL
(a) Subsection A describes the methods of nondestructive examination to be used if referenced by other
Code Sections or referencing documents.
(b) Subsection B lists Standards covering nondestructive examination methods which have been
accepted as standards.
T-150 PROCEDURE
T-160 CALIBRATION
a) The organization shall assure that all equipment calibrations required by Subsection A and/or
Subsection B are performed
T-221 Procedure Requirements. Each procedure shall include at least the following information, as
applicable:
(a) material type and thickness range
(b) isotope or maximum X-ray voltage used
(c) source-to-object distance (D in T-274.1)
(d) distance from source side of object to film (d in T-274.1)
(e) source size (F in T-274.1)
(f) film brand and designation
(g) screens used
T-221.2 Procedure Demonstration. Demonstration of the density and image quality indicator (IQI) image
requirements of the written procedure on production or technique radiographs shall be considered
satisfactory evidence of compliance with that procedure.
T-223 Backscatter Radiation. A lead symbol “B”, with minimum dimensions of 7 / 1 6 i n . (11 mm) in
height and 1/16 in. (1.5 mm) in thickness, shall be attached to the back of each film holder during each
exposure to determine if backscatter radiation is exposing the film.
T-232 Intensifying Screens. Intensifying screens may be used when performing radiographic examination
T-233.1 Standard IQI Design. IQIs shall be either the hole type or the wire type.
Table T-233.2 WIRE IQI DESIGNATION, WIRE DIAMETER, AND WIRE IDENTITY
T-234 Facilities for Viewing of Radiographs. The viewing conditions shall be such that light from around
the outer edge of the radiograph or coming through low-density portions of the radiograph does not interfere
with interpretation.
T-260 CALIBRATION
T-261.1 Verification of Source Size. The equipment manufacturer’s or supplier’s publications, such as
technical manuals, decay curves, or written statements documenting the actual or maximum source size
or focal spot, shall be acceptable as source size verification.
T-261.2 Determination of Source Size. When manufacturer’s or supplier’s publications are not
available, source size may be determined as follows:
(a) X-Ray Machines. For X-ray machines operating at 1.000 kV and less, the focal spot size may be
determined in accordance with SE-1165, Standard Test Method for
Measurement of Focal Spots of Industrial X-Ray Tubes by Pinhole Imaging.
(b) Iridium-192 Sources. For Iridium-192, the source size may be determined in accordance with SE-
1114, Standard Test Method for Determining the Focal Size of Iridium-192 Industrial
Radiographic Sources.
T-262.1 Densitometers. Densitometers shall be calibrated at least every 3 months during use as
follows:
(a) Traceable to a national standard last year by comparison with a national standard.
(b) The densitometer manufacturer’s step-by-step instructions for the operation of the densitometer
shall be followed.
(c) The density steps closest to 1.0, 2.0, 3.0, and 4.0 on the national standard step tablet or step wedge
calibration film shall be read.
(d) The densitometer is acceptable if the density readings do not vary by more than ± 0.05 density units
from the actual density stated on the national standard step tablet or step wedge calibration film.
T-262.4 Documentation
(a) Densitometers. Densitometer calibrations required by T-262.1 shall be documented, but the actual
readings for each step do not have to be recorded. Periodic densitometer verification checks required
by T-262.3(a) do not have to be documented.
(b) Step Wedge Calibration Films. Step wedge calibration film verifications required by T-262.1(a)
shall be documented, but the actual readings for each step do not have to be recorded.
(c) Step Wedge Comparison Films. Step wedge comparison film verifications required by T-262.2 and T-
262.3(b) shall be documented, but the actual readings for each step do not have to be recorded.
T-270 EXAMINATION
T-271.1 Single-Wall Technique. In the single-wall technique, the radiation passes through only one wall
of the weld (material), which is viewed for acceptance on the radiograph.
T-271.2 Double-Wall Technique. When it is not practical to use a single-wall technique, one of the
following double-wall techniques shall be used.
(a) Single-Wall Viewing. For materials and for welds in components, a technique may be used in which
T-272 Radiation Energy. The radiation energy employed for any radiographic technique shall achieve
the density an IQI image requirements of this Article.
T-273 Direction of Radiation. The direction of the central beam of radiation should be centered on the
area of interest whenever practical.
Material Ug
Thickness, in. (mm) Maximum, in. (mm)
Under 2 (50) 0.020 (0.51)
2 through 3 (50–75) 0.030 (0.76) Over
3 through 4 (75–100) 0.040 (1.02)
Greater than 4 (100) 0.070 (1.78)
Shall be placed on the part, not on the exposure holder/cassette. Their locations shall be permanently
marked on the surface of the part being radiographed when permitted, or on a map, in a manner permitting
the area of interest on a radiograph to be accurately traceable to its location on the part, for the required
retention period of the radiograph. In addition, their locations do not limit the area of interest or the area to
be interpreted. Evidence shall also be provided on the radiograph that the required coverage of the region
being examined has been obtained.
Material is from same alloy material group or grade per SE-1025 for hole type or SE-747 for wire type, or from
an alloy material group or grade with less radiation absorption than the material being radiographed.
IQI thickness based on nominal single wall thickness plus estimated reinforcement.
IQI selection of welds joining dissimilar materials or with dissimilar filler metals is based on weld metal.
(d) IQI Placement for Welds — Wire IQIs. The IQI(s) shall be placed on the weld so that the length of the
wires are transverse to the longitudinal axis of the weld.
T-277.2 Number of IQIs. When one or more film holders are used for an exposure, at least one IQI
image shall appear on each radiograph.
When a single exposure of a complete circumference is made only three IQI’s are required at 120
degrees apart.
T-280 EVALUATION
Must be free from mechanical, chemical or other blemishes that do not mask and are not confused with
the image of any discontinuity in the area of interest.
T-282.1 Density Limitations. Shall be 1.8 minimum for single film viewing for radiographs made with an X-
ray source and 2.0 minimum for radiographs made with a gamma ray source. For composite viewing of
multiple film exposures, each film of the composite set shall have a minimum density of 1.3. The maximum
density shall be 4.0 for either single or composite viewing. A tolerance of 0.05 in density is allowed for
variations between densitometer readings.
T-282.2 Density Variation. Density in area of interest must be within – 15% and + 30% from density
through the penetrameter otherwise an additional penetrameter shall be used.
If a light image of the “B,” appears on a darker background of the radiograph, protection from
backscatter is insufficient and the radiograph shall be considered unacceptable. A dark image of the “B” on
a lighter background is not cause for rejection.
T-290 DOCUMENTATION
The Manufacturer shall prepare and document the radiographic technique details. As a minimum, the
following information shall be provided.
(a) the requirements
(b) identification
(c) the dimensional map (if used) of marker placement
(d) number of radiographs (exposures)
(e) X-ray voltage or isotope type used
(f) source size
(g) base material type and thickness, weld thickness, weld reinforcement thickness, as applicable
(h) source-to-object distance
(i) distance from source side of object to film
(j) film manufacturer and their assigned type/designation
(k) number of films in each film holder/cassette
(l) single or double wall exposure
(m) single or double wall viewing
An effective means for detecting discontinuities which are open to the surface of nonporous metals and
other materials. Typical discontinuities detectable by this method are cracks, seams, laps, cold shuts,
laminations, and porosity.
Chlorine and fluorine content for austenitic stainless steel and titanium.
(a) Surface preparation by grinding, machining, or other methods may be necessary where surface
irregularities could mask indications.
(b) Prior to each liquid penetrant examination, the surface to be examined and all adjacent areas
within at least 1 in. (25 mm) shall be dry and free of all contaminant
For temperatures less than 40oF perform examination at proposed lower temperature and compare to the
40oF to 125oF range. This qualifies the procedure from lower temperature to 40oF.
o
For temperatures above 125 F the upper and lower limits shall be established and the procedure
qualified at these temperatures per Appendix III of this Article.
T-660 Calibrations
Light meters, both visible and fluorescent shall be calibrated once a year or whenever repaired.
T-670 Examination
Penetration (dwell) time is critical. The minimum penetration time shall be as required in Table T-672
The magnetic particle examination method is applied to detect cracks and other discontinuities on the
surfaces of ferromagnetic materials.
(c) Temperature Limitations. Particles shall be used within the temperature range limitations set by the
manufacturer of the particles.
T-750 TECHNIQUE
T-752.3 Prod Spacing. Prod spacing shall not exceed 8 in. (200 mm).
For this technique, alternating or direct current electromagnetic yokes, or permanent magnet yokes, shall
be used.
T-760 CALIBRATION
(a) Frequency. Magnetizing equipment with an ammeter shall be calibrated at least once a year, or
whenever the equipment has been subjected to major electric repair, periodic overhaul, or damage. If
equipment has not been in use for a year or more, calibration shall be done prior to first use.
(b) Procedure. The accuracy of the unit’s meter shall be verified annually by equipment traceable to
a national standard. Comparative readings shall be taken for at least three different current output
levels encompassing the usable range.
(c) Tolerance. The unit’s meter reading shall not deviate by more than ±10% of full scale, relative to the
actual current value as shown by the test meter.
(a) The magnetizing power of all yokes shall be verified whenever the yoke has been damaged or repaired.
(b) electromagnetic yoke shall have a lifting power of at least 10 lb (4.5 kg) at the maximum pole spacing
that will be used.
(c) permanent magnetic yoke shall have a lifting power of at least 40 lb (18 kg) at the maximum pole
spacing that will be used.
(d) Each weight shall be weighed with a scale from a reputable manufacturer and stenciled with the
applicable nominal weight prior to first use. A weight need only be verified again if damaged in a manner
that could have caused potential loss of material.
T-770 EXAMINATION
At least two separate examinations shall be performed on each area. During the second examination, the
lines of magnetic flux shall be approximately perpendicular to those used during the first examination.
T-777 Interpretation
Nonfluorescent particle examination shall be performed with a minimum light intensity of 100 foot
candles (fc) or 1076 lx.
(b) Examiners shall be in a darkened area for at least 5 min prior to performing examinations to enable
their eyes to adapt to dark viewing
T-778 Demagnetization
When residual magnetization interferes with subsequent processing or usage demagnetize any time after
completion of examination.
1. Scope
Thickness of materials using the contact pulse-echo method at temperatures not to exceed 200°F (93°C).
9. Technical Hazards
9.5 High-temperature materials. A rule of thumb often used is as follows: The apparent thickness reading
obtained from steel walls having elevated temperatures is high (too thick) by a factor of about
1% per 55°C (100°F). Thus, if the instrument was standardized on a piece of similar material at 20°C (68°F),
and if the reading was obtained with a surface temperature of 460°C (860°F), the apparent reading should
be reduced by 8%.