1. Check the pressure: 1.

Check A, B, C & L eluent, if they are enough;


H8 Operation SOP Flow Chart 2. Check the Waste liquid bucket, if it was
3-12Mpa.
2. Check the ADC value: empty;
415nm:16,000~18,000; 3. Check the reagent bottle, if any condensate
500nm:8,000~10,000. Check before start up the machine water inside, it need shaking well.

Prepare Calibrator:
Prepare QC sample: 1. Take out calibrator from refrigerator, wait
1. Take out QC from refrigerator, wait it Start manually Start automatically it until room temperature.
until room temperature. 2. Use pipette inject 100ul distilled water
2. Use pipette inject 100ul distilled or de-ionized water into calibrator bottle.
water or de-ionized water into QC 3. Wait 30 minutes until it fully re-dissolve.
bottle. 4. Aliquot the re-dissolve sample, 10ul in
3. Wait 30 minutes until it fully re- each EP tube, 1 bottle can aliquot in 10 EP
dissolve. tubes.
Confirm machine statement
4. Aliquot the re-dissolve sample, 10ul in
each EP tube, 1 bottle can aliquot in 10 Doing Calibration:
EP tube.
1. Use pipette inject 1,500ul L eluent into
EP tube, mixed it well;
Doing QC: 2. Input calibration information in
1. Use pipette inject 1,500ul L eluent calibration interface;
into EP tube, mixed it well; 3. Place the samples inside sample rack, the
Doing calibration
2. Input QC information in QC interface; sequence is: WB, C1, WB and C2.
3. Place the samples inside sample rack, 4. Doing the calibration in Cal. interface.
the sequence is: WB, QC1, WB and QC2. TIPS:
4. Doing the Auto QC in QC interface. 1. The aliquot sample needs to storage in -
TIPS: 20℃, it can storage for 14 days;
1. The aliquot sample needs to storage in 2. Calibrator bottle need to storage at
-20℃, it can storage for 14 days; Doing daily QC 2~8℃.
2. QC bottle need to storage at 2~8℃.

Mixed the whole blood sample, place the


Daily Maintenance blood sample in the sample rack, and click the
1. Check the liquid level of each bottle of analysis interface "Start "to start the test;
reagent and waste bottle, replace and Doing sample test
empty the waste liquid if necessary. TIPS:
2. Check remaining number of columns, 1. Samples taken from the refrigerator need to
filters and replace them if necessary. stand at room temperature for 30 minutes.
Test after mixing
Weekly Maintenance 2. Select the anticoagulated fresh blood
1. Cleaning sample rack and instrument sample of the day, in normal temperature, it
surface. should not exceed 48 hours;
Add more sample inside
2. Maintain the cleanliness of the 3. If testing diluted blood, in first position it
warehouse. must put a whole blood sample, EP tube hat
3. Cleaning screen. needs to be cut off.
All of maintenance suggest to wipe with a
soft, disposable towel or fabric.

Monthly Maintenance Wait until the test is finished, replace the


Result checking
1. Clean the Diluting Pool once sample rack
Use a cotton swab with 0.5% sodium
hypochlorite to manually clean the
diluting pool
2. Sampling probe cleaning Check if the graph is good and no baseline
Clean and maintain the sampling probe drift (This is a standard graph):
once a month. Open the right board and Maintenance operation
wipe the surface with a soft cloth or
sponge dipped with 5‰ sodium
hypochlorite solution.

Note: Check the status of the sampling


probe, and replace it if the sampling is
inaccurate or obviously bent. Shutdown the machine

: Necessary steps

: Optional steps