Professional Documents
Culture Documents
Pxie 4309
Pxie 4309
PXIe-4309
32 Ch (8 ADC), 2 MS/s, 18 - 28 bit, Flexible Resolution PXI Analog Input
Module
This document lists specifications for the PXIe-4309 data acquisition module. All specifications
are subject to change without notice. Refer to ni.com/manuals for the most current
specifications and product documentation.
Note To maintain forced air cooling in the PXIe system, refer to the Maintain
Forced-Air Cooling Note to Users.
Terminology
Maximum and minimum specifications characterize the warranted performance of the instrument
within the recommended calibration interval and under the stated operating conditions. These
specifications are subject to production verification or guaranteed by design.
Typical specifications are specifications met by the majority of the instruments within the
recommended calibration interval and under the stated operating conditions, based on
measurements taken during production verification and/or engineering development. The
performance of the instrument is not warranted.
Supplemental specifications describe the basic function and attributes of the instrument
established by design and are not subject to production verification. They provide information that
is relevant for the adequate use of the instrument that is not included in the previous definitions.
Noise*, †, ‡
Residual
Offset*, † Linearity† 10 S/s 10 kS/s 2 MS/s
10 V 50 9.6 5.8 84
15 V 75 10 14 8.7 125
*Source Impedance ≤50 Ω.
†Residual Offset, Linearity and Noise are included in the Absolute Accuracy values in the DC voltage
specifications table for Sample Rate ≤10 S/s.
‡Noise for Single Channel per ADC. For Multiple Channel per ADC refer to the Noise versus Sampling
Rate section.
Noise*, †, ‡
Residual
Offset*, † Linearity† 10 S/s 10 kS/s 2 MS/s
10 V 50 9.6 5.8 84
15 V 75 10 14 8.7 125
*Source Impedance ≤50 Ω.
†Residual Offset, Linearity and Noise are included in the Absolute Accuracy values in the DC voltage
specifications table for Sample Rate ≤10 S/s.
‡Noise for Single Channel per ADC. For Multiple Channel per ADC refer to the Noise versus Sampling
Rate section.
Noise*, †, ‡
Residual
Offset*, † Linearity† 10 S/s 10 kS/s
15 V 10 3.9 11
*Source Impedance ≤50 Ω.
†Residual Offset, Linearity and Noise are included in the Absolute Accuracy values in the DC voltage
specifications table for Sample Rate ≤10 S/s.
‡Noise for Single Channel per ADC. For Multiple Channel per ADC refer to the Noise versus Sampling
Rate section.
Noise*, †, ‡
Residual
Offset*, † Linearity† 10 S/s 10 kS/s
15 V 10 2 9.2
*Source Impedance ≤50 Ω.
†Residual Offset, Linearity and Noise are included in the Absolute Accuracy values in the DC voltage
specifications table for Sample Rate ≤10 S/s.
‡Noise for Single Channel per ADC. For Multiple Channel per ADC refer to the Noise versus Sampling
Rate section.
100n
Input Voltage (V)
50n
–50n
0
Temperature Change (°C)
–1
–2
–3
–4
–5
0 12 24 36 48 60 72
Time (hr)
–50n
0
Temperature Change (°C)
–1
–2
–3
–4
–5
0 12 24 36 48 60 72
Time (hr)
1µ
Spectral Noise Density (V/ Hz )
100n
10n
1n
10µ 100µ 1m 10m 100m 1.00
Frequency (Hz)
10µ
1µ
100n
100 1.0k 10.0k 100k 2.0M
Sample Rate (S/s)
±0.1V - 2 CH/ADC
10µ ±0.1V - 1 CH/ADC
1µ
100n
100 1.0k 10.0k 100k 2.0M
Sample Rate (S/s)
1µ
100n
10n
1.0 10.0 100 1.00k 10.0k
Sample Rate (S/s)
±0.1V - 2 CH/ADC
1µ ±0.1V - 1 CH/ADC
100n
10n
1.0 10.0 100 1.0k 10.0k
Sample Rate (S/s)
100n
10n
1n
1.0 10.0 100 1.0k 10.0k
Sample Rate (S/s)
100n
10n
1n
1.0 10.0 100 1.0k 10.0k
Sample Rate (S/s)
–5
–10
Amplitude (dB)
–15
–20
–25
–30
–35
–40
0.1 1 10
Frequency (Hz) / Sample Rate (S/s)
1 Applies to sampling rates ≤ 1 MS/s for all configurations that use a single channel per ADC.
2 Applies to sampling rates ≤ 200 kS/s (aggregate) for all configurations that use multiple channels per ADC.
3 Does not apply to Hardware-Timed Single Point, On-Demand, and External Sample Clock modes.
10 V .......................................................... 94 nV/ Hz
1µ
±1.0V
±0.1V
100n
10n
3n
100 1.0k 10.0k 100k 1.0M
Frequency (Hz)
100n
10n
3n
1.0 100 1.00k 10.0k
Frequency (Hz)
±10V - 1 CH/ADC
±0.1V - 4 CH/ADC
±0.1V - 2 CH/ADC
1µ ±0.1V - 1 CH/ADC
100n
10n
3n
10.0 100.0 1.0k 10.0k
Frequency (Hz)
100n
10n
3n
1.0 10.0 100 1.0k
Frequency (Hz)
±10V - 1 CH/ADC
±0.1V - 4 CH/ADC
±0.1V - 2 CH/ADC
1µ ±0.1V - 1 CH/ADC
100n
10n
3n
1.0 10.0 100 1.0k
Frequency (Hz)
100n
10n
3n
1.0 10.0 100 1.0k
Frequency (Hz)
±10V - 1 CH/ADC
±0.1V - 4 CH/ADC
±0.1V - 2 CH/ADC
1µ ±0.1V - 1 CH/ADC
100n
10n
3n
1.0 10.0 100 1.0k
Frequency (Hz)
100n
10n
3n
1.0 10.0 100 1.0k
Frequency (Hz)
±0.1V - 2 CH/ADC
±0.1V - 1 CH/ADC
100n
10n
3n
1.0 10.0 100 1.0k
Frequency (Hz)
110
CMRR (dB)
100
90
80
70
60
50
40
10.0 100 1.0k 10.0k 100k 1.0M
Frequency (Hz)
1 To maintain crosstalk performance use separation and/or shielding between signal cables.
2 Inputs shorted at terminal block screw terminals.
3 Inputs shorted at SCB-68 screw terminals using 2 m, 68-pin cable.
0
Normalized Signal Ampliltude (dB)
–1
–2
–3
–4
–5
–6
–7
–8
–9
10.0 100 1.0k 10.0k 100k 1.0M
Frequency (Hz)
Flatness
DC - 20 kHz...................................................... -6.5 mdB
DC - 80 kHz...................................................... -100 mdB
0.00
–0.02
–0.04
–0.06
–0.08
–0.10
–0.12
10.0 100 1.0k 100k 1.0M
Frequency (Hz)
1E+4 10 kΩ
20 kΩ
50 kΩ
1E+3 100 kΩ
1E+2
1E+1
1E+0
100 1.0k 10.0k 100k 400k
Aggregate Sample Rate (S/s)
100 Ω
1 kΩ
1E+5
2 kΩ
5 kΩ
1E+4 10 kΩ
20 kΩ
50 kΩ
1E+3 100 kΩ
1E+2
1E+1
1E+0
100 1.0k 10.0k 100k 400k
Aggregate Sample Rate (S/s)
1 Applies to all configurations that use multiple channels per ADC, Auto Zero Every Sample or Chopping.
2 Applies to all configurations that use multiple channels per ADC.
3 Applies to all configurations that use AI- connected to GND. Higher impedance between AI- and GND
may cause substantially higher error.
Digital Triggers
Purpose............................................................... Start trigger, reference trigger, pause trigger
Source................................................................. PFI 0, PFI 1, PXI_Trig <0..7>, PXI_Star,
PXIe_DStar A, PXIe_DStar B
Polarity.............................................................. Software-selectable
Debounce filter settings...................................... Disable, 90 ns, 5.12 µs, custom interval
General Specifications
Bus Interface
Form factor ........................................................x1 PXI Express peripheral module,
specification rev 1.0 compliant
Slot compatibility ...............................................PXI Express or PXI Express hybrid slots
DMA channels....................................................1, analog input
Power Requirements
+12 V..................................................................2 A max
+3.3 V.................................................................1 A max
Caution Do not use the PXIe-4309 for connections to signals or for measurements
within Categories II, III, or IV.
Caution Clean the hardware with a soft, nonmetallic brush. Make sure that the
hardware is completely dry and free from contaminants before returning it to service.
Environmental Specifications
Operating Environment
Ambient temperature range................................ 0 °C to 55 °C
(Tested in accordance with IEC 60068-2-1
and IEC 60068-2-2. Meets MIL-PRF-28800F
Class 3 low temperature limit and
MIL-PRF-28800F Class 2 high temperature
limit.)
Relative humidity range ..................................... 10% to 90%, noncondensing
(Tested in accordance with IEC 60068-2-56.)
Maximum altitude .............................................. 2,000 m (800 mbar)
Pollution Degree ............................................... 2
Indoor use only.
1 Measurement Categories CAT I and CAT O are equivalent. These test and measurement circuits are not
intended for direct connections to the MAINS building installations of Measurement Categories CAT II,
CAT III, CAT IV.
Calibration
You can obtain the calibration certificate and information about calibration services for the
PXIe-4309 at ni.com/calibration.
Self-calibration..................................................On software command, the module computes
gain, offset, and linearity corrections relative to
the high-precision internal voltage reference.
Self-calibration interval ....................................Depending on required absolute accuracy,
self-calibration is recommended whenever the
current device temperature differs by more than
the specified temperature range from the device
temperature at which the last self-calibration
was performed.
Calibration interval............................................2 years
Warm-up time.....................................................15 minutes
Note For UL and other safety certifications, refer to the product label or the Online
Product Certification section.
Electromagnetic Compatibility
This product meets the requirements of the following EMC standards for electrical equipment for
measurement, control, and laboratory use:
• EN 61326-1 (IEC 61326-1): Class A emissions; Basic immunity
• EN 55011 (CISPR 11): Group 1, Class A emissions
• EN 55022 (CISPR 22): Class A emissions
• EN 55024 (CISPR 24): Immunity
• AS/NZS CISPR 11: Group 1, Class A emissions
• AS/NZS CISPR 22: Class A emissions
• FCC 47 CFR Part 15B: Class A emissions
• ICES-001: Class A emissions
Note In the United States (per FCC 47 CFR), Class A equipment is intended for use
in commercial, light-industrial, and heavy-industrial locations. In Europe, Canada,
Australia and New Zealand (per CISPR 11) Class A equipment is intended for use
only in heavy-industrial locations.
Note Group 1 equipment (per CISPR 11) is any industrial, scientific, or medical
equipment that does not intentionally generate radio frequency energy for the
treatment of material or inspection/analysis purposes.
Note For EMC declarations and certifications, and additional information, refer to
the Online Product Certification section.
Note Group 1 equipment (per CISPR 11) is any industrial, scientific, or medical
equipment that does not intentionally generate radio frequency energy for the
treatment of material or inspection/analysis purposes.
Note For EMC declarations and certifications, and additional information, refer to
the Online Product Certification section.
Environmental Management
NI is committed to designing and manufacturing products in an environmentally responsible
manner. NI recognizes that eliminating certain hazardous substances from our products is
beneficial to the environment and to NI customers.
For additional environmental information, refer to the Minimize Our Environmental Impact web
page at ni.com/environment. This page contains the environmental regulations and
directives with which NI complies, as well as other environmental information not included in
this document.
Visit ni.com/services to learn about NI service offerings such as calibration options, repair,
and replacement.
377030B-01 Feb22