KODAK 8000 PCI Conversion to KODAK 8000 Ethernet

Kit Guide
Notice

The conversion kit guide for the KODAK 8000 PCI to KODAK 8000 Ethernet includes information on the
procedures for the conversion.

In order to do the conversion in the most efficient way, we advise you to follow the chapter order for the
procedures.

The information contained in this Guide may be subject to modification without notice, justification or notification
to the persons concerned.

No part of this Guide may be reproduced without the express permission of Carestream Health, Inc.

The US Federal law restricts this device to sale by or on the order of a physician.

This document is originally written in English.

Manual Name: KODAK 8000 PCI Conversion to KODAK 8000 Ethernet Upgrade Kit Guide
Part Number: SM815
Revision Number: 01
Print Date: 05-2012

The brand names and logos reproduced in this Guide are copyright.
KODAK is a trademark of KODAK used under License.
KODAK 8000 Extraoral Imaging System, complies with Directive 93/42/EEC relating to medical equipment.

Manufacturer

Care stream Hea lth, Inc.


150 Verona Street
Roche ster, NY 1 4 60 8, USA

Authorized Representative in the European Community

EC REP
TROPHY
4, Rue F. Pelloutier, Croissy-Beaubourg
77435 Marne la Vallée Cedex 2, France
Contents
1—ABOUT THIS GUIDE
Conventions in this Guide. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1–1
Note to the User. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1–2
Warning and Safety Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1–2
Warning and Safety Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1–3
Hygiene and Disinfection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1–4
Marking and Labeling Symbols. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1–5

2—CONVERSION KIT DESCRIPTION

3—PRE-CONVERSION PROCEDURES
Minimum Computer System Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3–1
Saving the Current System Parameters. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3–2
Installing the KODAK Dental Imaging Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3–3
Installing the Masking Files for Panoramic Ethernet Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3–9

4—CONVERSION PROCEDURE
Installing the Sensor Power Supply Board JZ030 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4–1
Replacing the PCI Sensor with Ethernet Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4–4
Replacing the CJ664 with a New CJ 664 Version (if Needed) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4–8
Installing New EEPROMs on the CJ664 Board (if Needed). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4–10
Adding the RJ45 Ethernet cable to the Column . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4–11

5—POST-INSTALLATION CONTROL
Restoring the New System Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5–1
Controlling of the X-Ray Beam . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5–2
Checking the Panoramic Image . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5–3

6—FITTING THE COVERS


Fitting the KODAK 8000 Covers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6–1

KODAK 8000 PCI Conversion to KODAK 8000 Ethernet (SM815)_Ed 01 iii


Contents

iv
Chapter 1
ABOUT THIS GUIDE

Conventions in this Guide

The following special messages emphasize information or indicate potential risk to


personnel or equipment:

WARNING
Warns you to avoid injury to yourself or others by following the
safety instructions precisely.

CAUTION
Alerts you to a condition that might cause serious damage.

IMPORTANT
Alerts you to a condition that might cause problems.

NOTE
Emphasizes important information.

TIP
Provides extra information and hints.

KODAK 8000 PCI Conversion to KODAK 8000 Ethernet (SM815)_Ed 01 1–1


Note to the User

Note to the User

WARNING
X-rays can be harmful and dangerous if not used properly. The
instructions and warnings contained in this guide must be
therefore carefully followed.

As a manufacturer of radiology units that conform to stringent radiological protection


standards in force throughout the world, we guarantee as low as reasonably achievable
degree of protection against radiation hazards. Nonetheless, you are handling a radiology
unit specially designed to emit x-ray doses in order to carry out a medical diagnosis.

The room in which your radiology unit is to be installed must comply with all official
regulations applicable to protection against radiation. You must install your radiology unit
in a room protected against x-ray emission. This room must reduce the frequency
interferences of the 30MHz to 1GHz band to at least 12db.

Your local representative will assist you in the initial use of your radiology unit and will
supply any relevant information you may require.

To use and operate your panoramic unit and your digital imaging software you must follow
the instructions contained in this guide.

Warning and Safety Instructions


When operating the radiology unit, observe the following warning and safety instructions:

DANGER OF ELECTRIC SHOCK

This is an electrical unit. DO NOT expose it to water spray. Such


action may cause an electric shock or a malfunction of the unit.

LASER WARNING

For maximum safety, advise the patient not to look at the


beam. Before turning on the beams, lower the Frankfurt
plane beam to the lowest level. While making adjustments,
ensure that the beam is not directed into the eyes of the
patient.

CAUTION

Use of controls, adjustments or performance of procedures


other than those specified herein may result in hazardous
radiation exposure.

1–2 ABOUT THIS GUIDE


Warning and Safety Instructions

Warning and Safety Instructions


WARNINGS

Unit:

 Read and understand this Safety Information before using the KODAK 8000
Extraoral Imaging Systems.
 You are responsible for the operation and maintenance of this unit. Only legally
qualified persons can operate this unit. They MUST have training to use the
radiological equipment. DO NOT open the cover of the unit. When necessary, have a
trained authorized service technician carry out inspection and maintenance
operations.
 Install this unit in an X-ray room that complies with current installation standards.
From this location, you must be able to maintain visual or audio communication
with the patient and be able to access the Acquisition interface module during
exposure.
 This unit must be permanently connected to the ground with a fixed power supply
cable. To avoid the risk of electric shock, this equipment must ONLY be connected
to a mains supply with protective earth.
 DO NOT operate the unit if there is the threat of an earthquake. Following an
earthquake, ensure that the unit is operating satisfactorily before using it again.
Failure to observe this precaution may expose patients to hazards.
 X-ray equipment is hazardous to patients and the operator if you do not observe the
exposure safety factors and operating instructions.
 DO NOT place objects within the field of operation of the unit.
 The patient should wear a protective lead-lined shoulder apron, unless other
Radiation Protection Protocols apply locally.
 While adjusting the height of the unit, ensure that the patient is kept clear of the
mechanism.
 When the unit is not in use, ensure that the ON/OFF switch is set to OFF (O).
 If the unit develops a fault, switch it to off (O), display an “Unserviceable” notice and
contact a service technician.
 To dispose of the unit or its components, contact a service technician.
 Ask the patient to refrain from moving during the entire period of exposure.
 Ask the patient to remain still until the unit arm has stopped moving and the RESET
movement has completed.
 DO NOT use this unit in conjunction with oxygen-rich environments. This unit is not
intended for use with flammable anesthetics or flammable agents.
 DO NOT hang from the cephalostat.
 Using accessories other than those specified in this document with the exception of
those sold by Carestream Health may result in a lower level of security for the entire
system.

Computer:

 DO NOT place the computer and the peripheral equipment connected to it in the
immediate vicinity of the patient in the unit. Leave at least 1.5 m distance between

KODAK 8000 PCI Conversion to KODAK 8000 Ethernet (SM815)_Ed 01 1–3


Hygiene and Disinfection

the patient and the unit. The computer and the peripheral equipment must conform
to the IEC60950 standard.
 See your computer installation guide for details of the data processing system and
screen. Leave a sufficient amount of clear space around the CPU to ensure that it is
properly ventilated.
 To obtain maximum image quality and visual comfort, position the screen to avoid
direct light reflections from internal or external lighting.

Hygiene and Disinfection


WARNINGS

 Disinfect any parts of the unit that come into contact with the patient and the
operator after each patient has been exposed to X-rays.
 To prevent cross-contamination, use a new hygienic sleeves for each new patient.

1–4 ABOUT THIS GUIDE


Marking and Labeling Symbols

Marking and Labeling Symbols

Type B device symbol complying with the IEC 60601-1


standard

In the European Union, this symbol indicates: DO NOT


discard this product in a trash receptacle; use an
appropriate recovery and recycling facility.
Contact your local sales representative for additional
information on the collection and recovery programs
available for this product
WARNING and IONIZING RADIATION symbols warn
you about radiation dangers.

LASER WARNING
Laser radiation. DO NOT stare into the beam.
Class 2 laser product.
Maximum output power: 1mW, 650 nm (IEC 60825-1
standard)
This unit emits laser radiation.

N Connecting point to the neutral conductor


L Connecting point to the live conductor
Protection ground (G)

Functional ground

KODAK 8000 PCI Conversion to KODAK 8000 Ethernet (SM815)_Ed 01 1–5


Marking and Labeling Symbols

1–6 ABOUT THIS GUIDE


Chapter 2
CONVERSION KIT DESCRIPTION
The KODAK 8000 PCI conversion to KODAK 8000 Ethernet kit is composed of
necessary components (see Table 2-1).

In order to do the conversions in the most efficient way, follow the chapter orders
for the conversion procedures.

The following table lists the content of the KODAK 8000 PCI conversion to KODAK
8000 Ethernet kit.

Table 2–1 KODAK 8000PCI to Ehernet Sensor Conversion Kit Content

Component Description Quantity


Ethernet Sensor 1

Sensor Ethernet cable (J5, 1


J10 & power supply)

RJ45 Ethernet Sensor 1


Cable

JZ030 sensor power 1


supply board

JZ030 metallic support 1


with screws 4 x M 3x10

M3x10 x4

Power supply cable 1


(21/CJ699/J12) for JZ030

KODAK 8000 PCI Conversion to KODAK 8000 Ethernet (SM815)_Ed 01 2–1


Table 2–1 KODAK 8000PCI to Ehernet Sensor Conversion Kit Content (Continued)

Component Description Quantity


Double-sided adhesive 1
tape for JZ030

CJ664 board v5 or higher 1


C80 K1 DZ1 J1
L1
J4 L2
1
C7
1
1
Q1 R19
C1 R1 R2 DZ2
J2 L3
R4 D1 D2 R5 R6 D3 R7 D4 C5
LD1
J3 R10 R11 C9
C4 E1 C8
P1
U3
C64 C12 R12 C11 C2
U5 ST1
C82
C6 U4 U6 Q 5 C13 D7 C3 R17
U39 C10 LD17
D8 D9
R18 P3 R20 LD15 R22

LD16 LD14 D12 C14 R16


C16 C15 R31 R33 C18 C19
Q2 P2
R23 R24 D5 J7
E3 R21 R25
U1 U11 C21 R27 R28 R29 R30 C17 R34 R35
J20 R37 R38 U15
C20 R43 C25
R42 C23 C31 C32 C26 R45 R46 C29
X31 R39 U12
R15 B1
U10 C30 U9 C24
R9
J 6 J5R8 D6 Z1 D14
R13 R47 C27 C28
R201 D 17 R49 D18 R48 U19B X8
X7 D21 D22 C37
C35 C36 R56
D19 R54
D10 R26 U16
J8
R3 D23 C34 R52 R50
R14 R200 D16
R32 R55
R63 R64 C38 U19 C40
C22 D13 U8 R66 C41
R67
C42
U14 R69
R72 C44 U20 C39 D35 D36
C46
R71 C45 Y1 C47 D26
R41 PT1 R70 R76 C43
C33 D15 C49
R91 U25 R75
U13
SW1 R80 R78 C81 C48
C50
U23
R57 R58 R82 R84
R81 R74
U24 C79
J9 D31
J11 LD19 LD18 DZ3 R83 R85
D29 D30 Q3
R86 U22 Q4
C52 U27 U28
R62 R96
C53
BUZ1 C54
D32 D33 R90
U26 R87
R88 R89 R97 R98
U30 R92 DZ5
E2
D45 C51
J10 D24 R59 D37
U17 R108

D25 R118
X5
D46 R123
R100 R124 R137 R120
U36 DZ4 C65
U21 C73
C57 R122
D28 R110 R126 R136 R130 R131
R127 R135 R128 R129 R132 SW2
C 67 D58
C83 R121
D27
R208 C66
R79 U29 U31 U33 U34 U32
U43 J12
1 D43 C63 C69 D44 LD13 D52 C76 D48 C68
PT4 LD9 R141 R205 Q9 C59 C60 C61 C84 D53
R93 LD6 U41
U38 R105 U42
D38 LD11 R204
C56 LD12 C75 LD8 R146 C70 R148 R149 R150
PT5 LD 7 R115 R138 R134 R144 X26
Q8
Q6 PT2
PT6 U37
D11 R203
PT3 R 140 R133
LD10 D51 R139 D47 R143 R202 U40 J 15 D42 D54
C74

J 19 D34 C55 J13 C78 C77 R207 J14 R142 J17 J18 R151 J16 C62 D39 D41 D57
D56

EEPROMs as of v3.04 or 2
higher for CJ664 board

Junction connector 2

Cable tie 20

Masking files for 1


panoramic Ethernet
sensor

KODAK Dental Image 1


Software v6.12.9 or higher

2–2 CONVERSION KIT DESCRIPTION


Chapter 3
PRE-CONVERSION PROCEDURES
Minimum Computer System Requirements
IMPORTANT
It is MANDATORY to check that the computer system configuration is
compatible with the computer system requirements for the KODAK 8000
Ethernet System software. If necessary you MUST update your computer
system configuration. KODAK 8000 System MUST be connected to the
computer via a point-to-point Ethernet link and not via a LAN..

Table 3–1 Minimum Computer System Requirements

Item Viewing Acquisition Comments

CPU 2 GHz Intel Duo Core 2 GHz Intel Duo Core


RAM 2 GB (Recommended 4GB) 2 GB (Recommended 4GB) RAM has a major impact on
system performance.
Hard disk drive  1.2 GB for software  4 GB for software installation
installation  80 GB free space to use the
 80 GB free space to use the software 2 GB
software (Recommended 500 GB)
Graphic board Graphic board supporting Open GL Nvidia based board with minimum The video RAM has major
1.2 with 256 MB of video RAM 512 MB of video RAM impact on system performance.
(Recommended Nvidia / ATI) (Recommended Nvidia Quadro FX
graphic boards)
Monitor  1 monitor  1 monitor Your monitor is a vital
component in displaying quality
 17" or larger  17" images. Low-quality screens will
prevent you from proper
 1024 x 800 minimum screen  1280 x 1024 minimum screen diagnoses and treatment.
resolution - 32 bits color resolution 1/1000
mode
Operating system  Windows 2000 SP4 (32 bits)  Windows XP Home / Pro The operating system
 Windows XP Home / Pro edition SP3 (32 bits) specifications are valid from
edition SP2 (32 bits)  Windows Vista (32 bits)
KODAK Dental Imaging
(Recommended SP3) (Recommended SP1) Software KDIS 6.12.9 and
above.
 Windows Vista (32 bits)  Windows 7 (32 or 64 bits)
(Recommended SP1)
 Windows 7 (32 or 64 bits)

Ethernet interface 100 Mbits for LAN 100 Mbits for LAN If there is a LAN connection a
second Ethernet interface is
required.
CD/DVD drive A DVD-BURNER drive is required. A DVD-BURNER drive is required.
Backup Media Removable/portable, external hard Removable/portable, external hard We strongly recommend a daily
disk drive disk drive. backup of x-ray images and
patient records.
Mouse A mouse with 2 buttons and a scroll You need a mouse with 2
wheel is required buttons and a scroll wheel to
view and use the 3D acquired
image.

KODAK 8000 PCI Conversion to KODAK 8000 Ethernet (SM815)_Ed 01 3–1


Saving the Current System Parameters

Saving the Current System Parameters


To save the system parameters, follow these steps:

1. On the control panel, press and hold , at the same time and turn ON the
unit.

2. Access the Panoramic Window. If an error message is displayed, click Retry button
until the screen is clear.

3. In the Panoramic Window, click and select Maintenance > Save / Restore.
The Save / Restore dialog box appears.

4. Click Save to save the current calibration and system parameters.

3–2 PRE-CONVERSION PROCEDURES


Installing the KODAK Dental Imaging Software

Installing the KODAK Dental Imaging Software


Before installing the KODAK Dental Imaging Software v6.12.9 or higher, check that:

 The computer has all the PC system requirements


 You have the software CD

To install the KODAK Dental Imaging Software, follow these steps:

1. Insert the software CD in the CD-ROM drive of the computer.

Wait for the installation program to start. If the program does not start automatically,
click Start > Run and enter D:\setup.exe if D is the letter for the CD-ROM drive, or
the letter of the relevant drive on the computer.

2. The Choose Setup Language dialog box is displayed. Select the installation
language and click OK.

3. The Kodak Dental Imaging Software welcome page and the InstallShield wizard
are displayed.

KODAK 8000 PCI Conversion to KODAK 8000 Ethernet (SM815)_Ed 01 3–3


Installing the KODAK Dental Imaging Software

4. The Welcome to KODAK Dental Imaging Software Installation dialog box is


displayed. Click Next to launch the installation.

5. The License Agreement dialog box is displayed. Accept and click Yes.

3–4 PRE-CONVERSION PROCEDURES


Installing the KODAK Dental Imaging Software

6. The Choose Destination Location dialog box is displayed. Click Next if you accept
the default destination folder (c:\program files\Kodak\Kodak Dental Imaging
Software) or browse to choose another destination folder.

NOTE
If you are upgrading the software, the default destination folder is
the previously selected destination folder.

7. The KODAK Dental Imaging Software dialog box is displayed. The Patient file is
selected by default but you must select the radiological device.

KODAK 8000 PCI Conversion to KODAK 8000 Ethernet (SM815)_Ed 01 3–5


Installing the KODAK Dental Imaging Software

8. To select the desired radiological device, do the following:

 Click on the drop-down list of No Pano/Ceph Installation and select the


appropriate option.
 Check that you have made the right selections.
 Click Next to begin the installation.

9. The InstallShield Wizard is displayed. Let the installation process run automatically
until the Installation Completer dialog box is displayed indicating the end of
installation.

3–6 PRE-CONVERSION PROCEDURES


Installing the KODAK Dental Imaging Software

This message will display if your Firewall blocks the installation of java. Click Unblock
to continue.

The Installation Complete dialog box is displayed when the installation is finished.
Click Yes I want to restart my computer now.

KODAK 8000 PCI Conversion to KODAK 8000 Ethernet (SM815)_Ed 01 3–7


Installing the KODAK Dental Imaging Software

10. Double-click to open the Patient Window.

11. Create a patient record. From the toolbar, click and enter the required patient

information. Click to access the Imaging Window.

12. The Enter the licence number dialog box is displayed. Enter the licence number and
click Validate if you have the licence number or click Cancel to continue.

3–8 PRE-CONVERSION PROCEDURES


Installing the Masking Files for Panoramic Ethernet Sensor

Installing the Masking Files for Panoramic


Ethernet Sensor
Before installing the masking files, check that you have the software CD.

To install the cephalometric masking files, follow these steps:

1. Insert the masking files CD in the CD-ROM drive of the computer. The installation will
run automatically.

2. The K8000MaskCopier dialog box appears when the installation is finished. Click
OK.

KODAK 8000 PCI Conversion to KODAK 8000 Ethernet (SM815)_Ed 01 3–9


Installing the Masking Files for Panoramic Ethernet Sensor

3–10 PRE-CONVERSION PROCEDURES


Chapter 4
CONVERSION PROCEDURE
Before starting, make sure that:

 The unit is powered OFF.


 The unit head, sensor and generator covers are removed.
 You have the required tools.
 You must follow the order of conversion in this chapter.

WARNING
You MUST switch off the unit before removing the covers. ONLY an
approved Trophy technician is qualified to inspect or maintain the unit
while it is switched on and while the covers are removed. In this case,
NO unqualified person must approach the unit.

Installing the Sensor Power Supply Board JZ030


To install the sensor power supply board JZ030, follow these steps:

1. Make sure that the unit is off.

2. Disconnect the following cables:

 PCI sensor power supply cableCJ669/J2


 Cable CJ664/J10 on J10 of the CJ664 board
 RS422

3. Remove the JZ016, PCI sensor power supply.

4. Attach the JZ030 metal support with the 4 screws to the Ethernet power supply
JZ030 board.

x4

2.5 mm

KODAK 8000 PCI Conversion to KODAK 8000 Ethernet (SM815)_Ed 01 4–1


Installing the Sensor Power Supply Board JZ030

5. Attach the double-sided adhesive tape to the back of the Ethernet power supply
board JZ030.

x2

6. Remove the adhesive and position the power supply board with the cable 21 on the
input CN1 facing the column.

1 : +15V
2 : +5.2V
3 : +5.2V
4 : GND
5 : GND
6 : -15V

2.5 mm
A

7. Connect the following connectors:

 Cable 21/CJ699/J12 on J12


 Ground cable between JZ030 and the ground connector.

4–2 CONVERSION PROCEDURE


Installing the Sensor Power Supply Board JZ030

8. Use a voltmeter to check the following voltage:

DANGER OF ELECTRIC SHOCK


The power supply adjustment must be made while the unit is
switched ON. Hazardous voltage is present in the power
supply JZ030 board. Unqualified persons MUST NOT
approach the unit.

 +5.2VDC check and adjust, if necessary


 +15VDC
 -15VDC

KODAK 8000 PCI Conversion to KODAK 8000 Ethernet (SM815)_Ed 01 4–3


Replacing the PCI Sensor with Ethernet Sensor

Replacing the PCI Sensor with Ethernet Sensor


To replace the PCI sensor with Ethernet sensor, follow these steps:

1. Disconnect the power supply cable from the mains.

2. Cut the 2 sensor cables of the PCI sensor.

3. Unscrew and remove the PCI sensor from its support.

3 mm
x3

4. Cut all the cable ties attached to the PCI sensor cables on the unit head.

5. Disconnect and remove the following PCI sensor cables from the unit head:

 CJ664/31/J19 from RS23


 RS232 from J15 on the CJ664 and remove from column (computer connection)

6. If there is no CephV2, then disconnect and remove RS422 from CJ664.

4–4 CONVERSION PROCEDURE


Replacing the PCI Sensor with Ethernet Sensor

7. Cut the cable ties that attache cable CJ669/17/J1 either on the CJ669 or CJ699
board.

8. Attach the Ethernet cable connector with adhesive tape to cable CJ669/17/J1. Make
sure that the Ethernet cable connector is well protected by the adhesive tape to
avoid it being damaged.

9. Insert from above, the cables into the cable opening on the unit head and pull from
below to help pass the Ethernet cable through the opening.

10. Once the Ethernet cable has gone completely through the opening, remove the
adhesive from the cables.

11. Pull back the cable CJ669/17/J1.

KODAK 8000 PCI Conversion to KODAK 8000 Ethernet (SM815)_Ed 01 4–5


Replacing the PCI Sensor with Ethernet Sensor

12. Attache the sensor cables with 2 connectors with adhesive tape to cable
CJ669/17/J1. Make sure that the attached connectors are attached at different
level to avoid thickness.

13. Insert from above, the cables into the cable opening on the unit head and pull from
below to help pass them through the opening.

14. Once the cables have gone completely through the opening, remove the adhesive
from the cables.

15. Pull back the cable CJ669/17/J1.

16. Pass all the cables through the hole above the rotative arm and tie with cable tie.

4–6 CONVERSION PROCEDURE


Replacing the PCI Sensor with Ethernet Sensor

17. Position and screw the Ethernet sensor to the support.

x2

B
A

x2

3 mm
5 mm

18. Connect the Ethernet sensor cables:

 Supply cable J10


 Ethernet Pano cable
 Synchro cable J5

19. Connect the following cables:

 Sensor power supply “Alim sensor” to JZ030 on CM1connector


 Cable CJ664/J10 on the CJ664 to J10 connector
 Cable CJ664/J15 on the CJ664 to J15 connector

KODAK 8000 PCI Conversion to KODAK 8000 Ethernet (SM815)_Ed 01 4–7


Replacing the CJ664 with a New CJ 664 Version (if Needed)

Replacing the CJ664 with a New CJ 664 Version


(if Needed)
WARNING
You must replace the CJ664 with the new CJ664 if:

 The CJ664 is lower than v5


 The EEPROMS version is lower than v3.04. In this
case, make sure that you do the necessary calibrations.

Before replacing the CJ664 board make sure that the unit is turned off.

To replace the CJ664 with a new version, follow these steps:

1. Make sure that the unit is turned off.

2. Disconnect all the connectors.

3. Unscrew the fixing screws of the board.

4. Remove the current CJ664 board.

5. Make sure that the 2 EEPROMS are mounted on the CJ664 board.

J3 J4 J2 J1
C80 K1 DZ1 J1
L1
J4 L2
1
C7
1
1
Q1 R19
R1 R2 DZ2
C1 L3
J2 D3
R4 D1 D2 R5 R6 R7 D4 C5
LD1
J3 R10 R11 C4 C8 C9
E1
P1
C64
U3 C12 R12 C11 C2
U5 ST 1
C82
C6 U4 U6 Q5 C13 D7 C3 R17
U39 C10 LD17
D8 D9
R18 P3 R20 LD15 R 22

LD16 D12 C14 R16


C16 C15 LD14
R31 R33 C18 C19
Q2
P2
R23 R24 D5 J7

J5
E3 R21 R25
U1 U11 C21 R27 R28 R29 R30 C17 R34 R35
J20 R37 R38 U15
C20 R43 C25
R42 C23 C31 C32 C26 R45 R46 C29
X31 R39 U12
R15 B1
R8 R9 U10 C30 U9 C24
J 6 J5 D6 Z1 D14
R13 R47 C27 C28
R49D18 R48 U19B X8

J6
R201 D 17
X7 C37
D21 D22 R56
D19 C35 C36 R54
D10 R26 U16
J8
R3 D23 C34 R52 R50
R14 R200 D16
R32 R55
R63 R64 C38 U19 C40
C22 D13 U8 R66 C41
R67
C42
U14 R 69

J8 R72 C 4 4 U20 C39 D35 D36


C46
R71 C45 Y1 C47 D26
R41 PT 1 R70 R 76 C 43

C33 D15 C49


R91 U25 R75
U13
SW1 R80 R78 C81 C48
C50
U23
R82 R84

J9
R57 R58 R74
R81
J9 U24 C79
D31
J 11 LD19 LD18 DZ3 R83 R85
D29 D30 U22 Q3 Q4
R86 U27
C52 U28 R96
R62 C53
BUZ1 C54

11
R90
D32 D33 U26 R87
R88 R89 R97 R98 U30 R92 DZ5
E2
D45 C51
J10 D24 R59 D37 R 108
U17

J10
D25 X5 R118
D46 R123 R120
R100 R124 R137
U36 DZ4 C65
U21 R122 C73
R110 R126 R136 R130 R131 R127 C57
D28 R135 R129 SW2
R128 R132
C 67 D58
C83 R121
D27
R208 C66
R79 U29 U31 U33 U34 U32
U43 J 12
1 D43 C63 C69 D44 LD13D52 C76 D48 C68
PT 4 R141 R205 Q9 C59 C60 C61 C84 D53
R93 LD6 LD9 U41

U38 R105 U42


C56 D38 LD11
LD12 C75 LD8
R204
R146 C70 R148 R149 R150
PT 5 LD 7 R115 R138 R134 R144 X26
Q8
Q6 PT 2
PT 6 U37
D11 R203
P T3 R 140 R 133 LD10 D51 R139 D47 R 143 R202 U40 J 15 D 42 D 54
C74

J19 D34 C55 J13 C78 C77 R207 J14 R142 J17 J 18 R151 J16 C62 D39 D41 D57
D56

J13 J17 J16 J15 J12


J19 J14 J18
R90
x6
R92

D45

SW2 D46

R122

R132 SW2
2.5 mm
CJ664

4–8 CONVERSION PROCEDURE


Replacing the CJ664 with a New CJ 664 Version (if Needed)

6. Position the DIP switch (SW2) in the same position as the replaced board.

Table 4–1 DIP Switch (SW2) Positions and Descriptions

DIP Switch DIP Switch


Function
Position State
S1 Emission of x-ray during the exposure OFF
No emission of x-ray during the exposure (Demo mode) ON
S2 Communication speed between the CJ732 and the fixed arm
control panel
9600 baud OFF
38400 baud ON
19200 baud OFF
57600 baud ON
S3 Communication speed between the CJ732 and the fixed arm
control panel
9600 baud OFF
38400 baud OFF
19200 baud ON
57600 baud ON
S4 NA
S5 NA
S6 NA
S7 NA
S8 NA

7. Position the CJ664 board with the connections J25, J10 and J11 facing the column
and screw the 6 fixing screws.

8. Connect all the connectors in their respective locations.

KODAK 8000 PCI Conversion to KODAK 8000 Ethernet (SM815)_Ed 01 4–9


Installing New EEPROMs on the CJ664 Board (if Needed)

Installing New EEPROMs on the CJ664 Board (if


Needed)

WARNING
You must replace the EEPROMS with the new EEPROMS if
the version is lower than v3.4. In this case, make sure that you
do the following calibrations:

 Reinitialize manufacturing parameters


 kV
 mA
 Heating voltage calibration
 Tube head calibration ?

1. Insert the tip of a screw driver gently under each side of the EEPROM to raise it.

WARNING
You MUST remove the EEPROM from both sides without
breaking the integrated circuits.

2. Remove the 2 current EEPROMs.

CJ732
x2

LOW HIGH

U28 U27
1

3. Position correctly and plug in the EEPROM U27 with the indentation facing “HIGH”
and the EEPROM U28 with the indentation facing “LOW”.

4–10 CONVERSION PROCEDURE


Adding the RJ45 Ethernet cable to the Column

Adding the RJ45 Ethernet cable to the Column


To add the RJ45 Ethernet cable to the column, follow these steps:

1. Make sure that the unit is turned off and disconnected from the mains.

2. Insert one side of the Ethernet cable RJ45 from above the opening in the column.

3. Make sure that it slides down all the way to the base of the column.

4. Pull the cable through the opening at the base of the column.

5. Attach the cable with the cable tie to the other cables in the column near the unit
head.

KODAK 8000 PCI Conversion to KODAK 8000 Ethernet (SM815)_Ed 01 4–11


Adding the RJ45 Ethernet cable to the Column

4–12 CONVERSION PROCEDURE


Chapter 5
POST-INSTALLATION CONTROL
Before the post-installation control, check that:

 The installation of the unit is complete.


 The installation of the KODAK Dental Imaging Software is complete.
 You have the test tool.

Restoring the New System Parameters

WARNING
ONLY if you have not changed the EEPROMS.

To restore the system parameters, follow these steps:

1. On the control panel, press and hold , at the same time and turn ON the
unit.

2. Error message E54 or E56 is displayed, click Restore 2 times to clear the screen.
The saved system and calibration parameters are restored in the new EEPROMS.

3. Access the Panoramic Window. In the Panoramic Window, click and select
Maintenance > Manufacturing Parameters. The Manufacturing Parameters dialog
box appears.

4. Click Save.

KODAK 8000 PCI Conversion to KODAK 8000 Ethernet (SM815)_Ed 01 5–1


Controlling of the X-Ray Beam

Controlling of the X-Ray Beam

You need to check that the x-ray beam is correctly centered on the CCD after replacing
the collimator.

To check the x-ray beam, follow these steps:

1. Switch on the unit in technical mode.

2. Access the Panoramic Acquisition Window. If an error message is displayed, click


Retry button until the screen is clear.

3. In the Panoramic Acquisition Window, click and select Maintenance >


Calibration > Control of the x-ray beam.

4. Launch an x-ray using the remote control. Check that the x-ray beam represented in
black is centered correctly in the white frame.

If the x-ray beam is not centered correctly, you must adjust the collimator and relaunch an
x-ray until the x-ray beam is centered correctly.

5–2 POST-INSTALLATION CONTROL


Checking the Panoramic Image

Checking the Panoramic Image

To check the Panoramic image, follow these steps:

1. Switch ON the PC. Wait for the connection to be established between the PC and
the unit.

2. Place the tool base on the chin rest base. Position the ball phantom on the tool base.
Raise the chin rest base to the maximum height.

3. Access the Imaging Window. Click in the Imaging Window to access the
Panoramic Acquisition Window.

 In the Patient pane, select .


 In the Program pane, select 60 kV and 2 mA.

KODAK 8000 PCI Conversion to KODAK 8000 Ethernet (SM815)_Ed 01 5–3


Checking the Panoramic Image

4. Adjust the ball phantom to the mid-sagittal and canine laser beams. On the control

panel, press :

 The mid-sagittal laser beam must be positioned on the central line (A).
 The canine laser beam must be positioned on the horizontal line (B).
 Make sure that (A) and (B) are perpendicular.

A B

5. Leave the x-ray room and using the x-ray remote control, launch an x-ray.

6. Use the Control Panel of the Imaging Window for an appropriate contrast of the
image. Check for the following results:

 The screen geometry is correct.


 The central ball (A) must be round and without any distortion.
 The 2 end balls must be symmetric from the central axis (B = C).

B C

5–4 POST-INSTALLATION CONTROL


Checking the Panoramic Image

7. If the result is unsatisfactory, check to see if the mid-sagittal laser beam is correctly
positioned on the ball phantom, then redo the operation.

If the result is still unsatisfactory, align the chin rest, mid-sagittal laser and the canine
laser and redo the operation until satisfaction (see KODAK 8000 Service Guide
(SM724):

 “Aligning the Chin Rest”


 “Aligning the Mid-sagittal Laser”
 “Aligning the Canine Laser”

KODAK 8000 PCI Conversion to KODAK 8000 Ethernet (SM815)_Ed 01 5–5


Checking the Panoramic Image

5–6 POST-INSTALLATION CONTROL


Chapter 6
FITTING THE COVERS
WARNING
Before fitting the covers, you MUST switch off the unit.

Fitting the KODAK 8000 Covers


Before fitting the KODAK 8000 covers, check that:

 All the screws in the wall and the floor have been tightened.
 All the screws attaching the head and the fixed arm to the column have been
tightened.

D
2x M5x20
B

2x M5x50

2x M5x15

C
E
2x M5x50

2x
M5x12

3 mm

KODAK 8000 PCI Conversion to KODAK 8000 Ethernet (SM815)_Ed 01 6–1


Fitting the KODAK 8000 Covers

To fit the covers, follow these steps:

1. Fit the upper cover (A) on the column. Fit the cover on the wall-bracket (B). Fit the
cover on the column connecting terminal (C) and tighten the fixing screws.

2. Fit the head cover (D) of the unit to the head chassis and tighten the fixing screws.

3. Fit under the head cover (E) and tighten the fixing screws.

4. Fit the inside column cover (F) and position and attach securely the ON/OFF button
to the cover with the fixing screws. Screw and tighten the fixing screws of the column
cover.

You can fit the cephalometric covers now.

6–2 FITTING THE COVERS