Kit Guide
Notice
The conversion kit guide for the KODAK 8000 PCI to KODAK 8000 Ethernet includes information on the
procedures for the conversion.
In order to do the conversion in the most efficient way, we advise you to follow the chapter order for the
procedures.
The information contained in this Guide may be subject to modification without notice, justification or notification
to the persons concerned.
No part of this Guide may be reproduced without the express permission of Carestream Health, Inc.
The US Federal law restricts this device to sale by or on the order of a physician.
Manual Name: KODAK 8000 PCI Conversion to KODAK 8000 Ethernet Upgrade Kit Guide
Part Number: SM815
Revision Number: 01
Print Date: 05-2012
The brand names and logos reproduced in this Guide are copyright.
KODAK is a trademark of KODAK used under License.
KODAK 8000 Extraoral Imaging System, complies with Directive 93/42/EEC relating to medical equipment.
Manufacturer
EC REP
TROPHY
4, Rue F. Pelloutier, Croissy-Beaubourg
77435 Marne la Vallée Cedex 2, France
Contents
1—ABOUT THIS GUIDE
Conventions in this Guide. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1–1
Note to the User. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1–2
Warning and Safety Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1–2
Warning and Safety Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1–3
Hygiene and Disinfection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1–4
Marking and Labeling Symbols. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1–5
3—PRE-CONVERSION PROCEDURES
Minimum Computer System Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3–1
Saving the Current System Parameters. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3–2
Installing the KODAK Dental Imaging Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3–3
Installing the Masking Files for Panoramic Ethernet Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3–9
4—CONVERSION PROCEDURE
Installing the Sensor Power Supply Board JZ030 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4–1
Replacing the PCI Sensor with Ethernet Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4–4
Replacing the CJ664 with a New CJ 664 Version (if Needed) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4–8
Installing New EEPROMs on the CJ664 Board (if Needed). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4–10
Adding the RJ45 Ethernet cable to the Column . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4–11
5—POST-INSTALLATION CONTROL
Restoring the New System Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5–1
Controlling of the X-Ray Beam . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5–2
Checking the Panoramic Image . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5–3
iv
Chapter 1
ABOUT THIS GUIDE
WARNING
Warns you to avoid injury to yourself or others by following the
safety instructions precisely.
CAUTION
Alerts you to a condition that might cause serious damage.
IMPORTANT
Alerts you to a condition that might cause problems.
NOTE
Emphasizes important information.
TIP
Provides extra information and hints.
WARNING
X-rays can be harmful and dangerous if not used properly. The
instructions and warnings contained in this guide must be
therefore carefully followed.
The room in which your radiology unit is to be installed must comply with all official
regulations applicable to protection against radiation. You must install your radiology unit
in a room protected against x-ray emission. This room must reduce the frequency
interferences of the 30MHz to 1GHz band to at least 12db.
Your local representative will assist you in the initial use of your radiology unit and will
supply any relevant information you may require.
To use and operate your panoramic unit and your digital imaging software you must follow
the instructions contained in this guide.
LASER WARNING
CAUTION
Unit:
Read and understand this Safety Information before using the KODAK 8000
Extraoral Imaging Systems.
You are responsible for the operation and maintenance of this unit. Only legally
qualified persons can operate this unit. They MUST have training to use the
radiological equipment. DO NOT open the cover of the unit. When necessary, have a
trained authorized service technician carry out inspection and maintenance
operations.
Install this unit in an X-ray room that complies with current installation standards.
From this location, you must be able to maintain visual or audio communication
with the patient and be able to access the Acquisition interface module during
exposure.
This unit must be permanently connected to the ground with a fixed power supply
cable. To avoid the risk of electric shock, this equipment must ONLY be connected
to a mains supply with protective earth.
DO NOT operate the unit if there is the threat of an earthquake. Following an
earthquake, ensure that the unit is operating satisfactorily before using it again.
Failure to observe this precaution may expose patients to hazards.
X-ray equipment is hazardous to patients and the operator if you do not observe the
exposure safety factors and operating instructions.
DO NOT place objects within the field of operation of the unit.
The patient should wear a protective lead-lined shoulder apron, unless other
Radiation Protection Protocols apply locally.
While adjusting the height of the unit, ensure that the patient is kept clear of the
mechanism.
When the unit is not in use, ensure that the ON/OFF switch is set to OFF (O).
If the unit develops a fault, switch it to off (O), display an “Unserviceable” notice and
contact a service technician.
To dispose of the unit or its components, contact a service technician.
Ask the patient to refrain from moving during the entire period of exposure.
Ask the patient to remain still until the unit arm has stopped moving and the RESET
movement has completed.
DO NOT use this unit in conjunction with oxygen-rich environments. This unit is not
intended for use with flammable anesthetics or flammable agents.
DO NOT hang from the cephalostat.
Using accessories other than those specified in this document with the exception of
those sold by Carestream Health may result in a lower level of security for the entire
system.
Computer:
DO NOT place the computer and the peripheral equipment connected to it in the
immediate vicinity of the patient in the unit. Leave at least 1.5 m distance between
the patient and the unit. The computer and the peripheral equipment must conform
to the IEC60950 standard.
See your computer installation guide for details of the data processing system and
screen. Leave a sufficient amount of clear space around the CPU to ensure that it is
properly ventilated.
To obtain maximum image quality and visual comfort, position the screen to avoid
direct light reflections from internal or external lighting.
Disinfect any parts of the unit that come into contact with the patient and the
operator after each patient has been exposed to X-rays.
To prevent cross-contamination, use a new hygienic sleeves for each new patient.
LASER WARNING
Laser radiation. DO NOT stare into the beam.
Class 2 laser product.
Maximum output power: 1mW, 650 nm (IEC 60825-1
standard)
This unit emits laser radiation.
Functional ground
In order to do the conversions in the most efficient way, follow the chapter orders
for the conversion procedures.
The following table lists the content of the KODAK 8000 PCI conversion to KODAK
8000 Ethernet kit.
M3x10 x4
D25 R118
X5
D46 R123
R100 R124 R137 R120
U36 DZ4 C65
U21 C73
C57 R122
D28 R110 R126 R136 R130 R131
R127 R135 R128 R129 R132 SW2
C 67 D58
C83 R121
D27
R208 C66
R79 U29 U31 U33 U34 U32
U43 J12
1 D43 C63 C69 D44 LD13 D52 C76 D48 C68
PT4 LD9 R141 R205 Q9 C59 C60 C61 C84 D53
R93 LD6 U41
U38 R105 U42
D38 LD11 R204
C56 LD12 C75 LD8 R146 C70 R148 R149 R150
PT5 LD 7 R115 R138 R134 R144 X26
Q8
Q6 PT2
PT6 U37
D11 R203
PT3 R 140 R133
LD10 D51 R139 D47 R143 R202 U40 J 15 D42 D54
C74
J 19 D34 C55 J13 C78 C77 R207 J14 R142 J17 J18 R151 J16 C62 D39 D41 D57
D56
EEPROMs as of v3.04 or 2
higher for CJ664 board
Junction connector 2
Cable tie 20
Ethernet interface 100 Mbits for LAN 100 Mbits for LAN If there is a LAN connection a
second Ethernet interface is
required.
CD/DVD drive A DVD-BURNER drive is required. A DVD-BURNER drive is required.
Backup Media Removable/portable, external hard Removable/portable, external hard We strongly recommend a daily
disk drive disk drive. backup of x-ray images and
patient records.
Mouse A mouse with 2 buttons and a scroll You need a mouse with 2
wheel is required buttons and a scroll wheel to
view and use the 3D acquired
image.
1. On the control panel, press and hold , at the same time and turn ON the
unit.
2. Access the Panoramic Window. If an error message is displayed, click Retry button
until the screen is clear.
3. In the Panoramic Window, click and select Maintenance > Save / Restore.
The Save / Restore dialog box appears.
Wait for the installation program to start. If the program does not start automatically,
click Start > Run and enter D:\setup.exe if D is the letter for the CD-ROM drive, or
the letter of the relevant drive on the computer.
2. The Choose Setup Language dialog box is displayed. Select the installation
language and click OK.
3. The Kodak Dental Imaging Software welcome page and the InstallShield wizard
are displayed.
5. The License Agreement dialog box is displayed. Accept and click Yes.
6. The Choose Destination Location dialog box is displayed. Click Next if you accept
the default destination folder (c:\program files\Kodak\Kodak Dental Imaging
Software) or browse to choose another destination folder.
NOTE
If you are upgrading the software, the default destination folder is
the previously selected destination folder.
7. The KODAK Dental Imaging Software dialog box is displayed. The Patient file is
selected by default but you must select the radiological device.
9. The InstallShield Wizard is displayed. Let the installation process run automatically
until the Installation Completer dialog box is displayed indicating the end of
installation.
This message will display if your Firewall blocks the installation of java. Click Unblock
to continue.
The Installation Complete dialog box is displayed when the installation is finished.
Click Yes I want to restart my computer now.
11. Create a patient record. From the toolbar, click and enter the required patient
12. The Enter the licence number dialog box is displayed. Enter the licence number and
click Validate if you have the licence number or click Cancel to continue.
1. Insert the masking files CD in the CD-ROM drive of the computer. The installation will
run automatically.
2. The K8000MaskCopier dialog box appears when the installation is finished. Click
OK.
WARNING
You MUST switch off the unit before removing the covers. ONLY an
approved Trophy technician is qualified to inspect or maintain the unit
while it is switched on and while the covers are removed. In this case,
NO unqualified person must approach the unit.
4. Attach the JZ030 metal support with the 4 screws to the Ethernet power supply
JZ030 board.
x4
2.5 mm
5. Attach the double-sided adhesive tape to the back of the Ethernet power supply
board JZ030.
x2
6. Remove the adhesive and position the power supply board with the cable 21 on the
input CN1 facing the column.
1 : +15V
2 : +5.2V
3 : +5.2V
4 : GND
5 : GND
6 : -15V
2.5 mm
A
3 mm
x3
4. Cut all the cable ties attached to the PCI sensor cables on the unit head.
5. Disconnect and remove the following PCI sensor cables from the unit head:
7. Cut the cable ties that attache cable CJ669/17/J1 either on the CJ669 or CJ699
board.
8. Attach the Ethernet cable connector with adhesive tape to cable CJ669/17/J1. Make
sure that the Ethernet cable connector is well protected by the adhesive tape to
avoid it being damaged.
9. Insert from above, the cables into the cable opening on the unit head and pull from
below to help pass the Ethernet cable through the opening.
10. Once the Ethernet cable has gone completely through the opening, remove the
adhesive from the cables.
12. Attache the sensor cables with 2 connectors with adhesive tape to cable
CJ669/17/J1. Make sure that the attached connectors are attached at different
level to avoid thickness.
13. Insert from above, the cables into the cable opening on the unit head and pull from
below to help pass them through the opening.
14. Once the cables have gone completely through the opening, remove the adhesive
from the cables.
16. Pass all the cables through the hole above the rotative arm and tie with cable tie.
x2
B
A
x2
3 mm
5 mm
Before replacing the CJ664 board make sure that the unit is turned off.
5. Make sure that the 2 EEPROMS are mounted on the CJ664 board.
J3 J4 J2 J1
C80 K1 DZ1 J1
L1
J4 L2
1
C7
1
1
Q1 R19
R1 R2 DZ2
C1 L3
J2 D3
R4 D1 D2 R5 R6 R7 D4 C5
LD1
J3 R10 R11 C4 C8 C9
E1
P1
C64
U3 C12 R12 C11 C2
U5 ST 1
C82
C6 U4 U6 Q5 C13 D7 C3 R17
U39 C10 LD17
D8 D9
R18 P3 R20 LD15 R 22
J5
E3 R21 R25
U1 U11 C21 R27 R28 R29 R30 C17 R34 R35
J20 R37 R38 U15
C20 R43 C25
R42 C23 C31 C32 C26 R45 R46 C29
X31 R39 U12
R15 B1
R8 R9 U10 C30 U9 C24
J 6 J5 D6 Z1 D14
R13 R47 C27 C28
R49D18 R48 U19B X8
J6
R201 D 17
X7 C37
D21 D22 R56
D19 C35 C36 R54
D10 R26 U16
J8
R3 D23 C34 R52 R50
R14 R200 D16
R32 R55
R63 R64 C38 U19 C40
C22 D13 U8 R66 C41
R67
C42
U14 R 69
J9
R57 R58 R74
R81
J9 U24 C79
D31
J 11 LD19 LD18 DZ3 R83 R85
D29 D30 U22 Q3 Q4
R86 U27
C52 U28 R96
R62 C53
BUZ1 C54
11
R90
D32 D33 U26 R87
R88 R89 R97 R98 U30 R92 DZ5
E2
D45 C51
J10 D24 R59 D37 R 108
U17
J10
D25 X5 R118
D46 R123 R120
R100 R124 R137
U36 DZ4 C65
U21 R122 C73
R110 R126 R136 R130 R131 R127 C57
D28 R135 R129 SW2
R128 R132
C 67 D58
C83 R121
D27
R208 C66
R79 U29 U31 U33 U34 U32
U43 J 12
1 D43 C63 C69 D44 LD13D52 C76 D48 C68
PT 4 R141 R205 Q9 C59 C60 C61 C84 D53
R93 LD6 LD9 U41
J19 D34 C55 J13 C78 C77 R207 J14 R142 J17 J 18 R151 J16 C62 D39 D41 D57
D56
D45
SW2 D46
R122
R132 SW2
2.5 mm
CJ664
6. Position the DIP switch (SW2) in the same position as the replaced board.
7. Position the CJ664 board with the connections J25, J10 and J11 facing the column
and screw the 6 fixing screws.
WARNING
You must replace the EEPROMS with the new EEPROMS if
the version is lower than v3.4. In this case, make sure that you
do the following calibrations:
1. Insert the tip of a screw driver gently under each side of the EEPROM to raise it.
WARNING
You MUST remove the EEPROM from both sides without
breaking the integrated circuits.
CJ732
x2
LOW HIGH
U28 U27
1
3. Position correctly and plug in the EEPROM U27 with the indentation facing “HIGH”
and the EEPROM U28 with the indentation facing “LOW”.
1. Make sure that the unit is turned off and disconnected from the mains.
2. Insert one side of the Ethernet cable RJ45 from above the opening in the column.
3. Make sure that it slides down all the way to the base of the column.
4. Pull the cable through the opening at the base of the column.
5. Attach the cable with the cable tie to the other cables in the column near the unit
head.
WARNING
ONLY if you have not changed the EEPROMS.
1. On the control panel, press and hold , at the same time and turn ON the
unit.
2. Error message E54 or E56 is displayed, click Restore 2 times to clear the screen.
The saved system and calibration parameters are restored in the new EEPROMS.
3. Access the Panoramic Window. In the Panoramic Window, click and select
Maintenance > Manufacturing Parameters. The Manufacturing Parameters dialog
box appears.
4. Click Save.
You need to check that the x-ray beam is correctly centered on the CCD after replacing
the collimator.
4. Launch an x-ray using the remote control. Check that the x-ray beam represented in
black is centered correctly in the white frame.
If the x-ray beam is not centered correctly, you must adjust the collimator and relaunch an
x-ray until the x-ray beam is centered correctly.
1. Switch ON the PC. Wait for the connection to be established between the PC and
the unit.
2. Place the tool base on the chin rest base. Position the ball phantom on the tool base.
Raise the chin rest base to the maximum height.
3. Access the Imaging Window. Click in the Imaging Window to access the
Panoramic Acquisition Window.
4. Adjust the ball phantom to the mid-sagittal and canine laser beams. On the control
panel, press :
The mid-sagittal laser beam must be positioned on the central line (A).
The canine laser beam must be positioned on the horizontal line (B).
Make sure that (A) and (B) are perpendicular.
A B
5. Leave the x-ray room and using the x-ray remote control, launch an x-ray.
6. Use the Control Panel of the Imaging Window for an appropriate contrast of the
image. Check for the following results:
B C
7. If the result is unsatisfactory, check to see if the mid-sagittal laser beam is correctly
positioned on the ball phantom, then redo the operation.
If the result is still unsatisfactory, align the chin rest, mid-sagittal laser and the canine
laser and redo the operation until satisfaction (see KODAK 8000 Service Guide
(SM724):
All the screws in the wall and the floor have been tightened.
All the screws attaching the head and the fixed arm to the column have been
tightened.
D
2x M5x20
B
2x M5x50
2x M5x15
C
E
2x M5x50
2x
M5x12
3 mm
1. Fit the upper cover (A) on the column. Fit the cover on the wall-bracket (B). Fit the
cover on the column connecting terminal (C) and tighten the fixing screws.
2. Fit the head cover (D) of the unit to the head chassis and tighten the fixing screws.
3. Fit under the head cover (E) and tighten the fixing screws.
4. Fit the inside column cover (F) and position and attach securely the ON/OFF button
to the cover with the fixing screws. Screw and tighten the fixing screws of the column
cover.