Professional Documents
Culture Documents
Analitical Error
Analitical Error
Analitical Error
Abstract
Although generally robust, immunoassays remain vulnerable to occasional analytical errors that may have serious
implications for patient care. Sporadic errors that occur as a result of properties of the specimen are particularly difficult
to detect. They may be due to the presence of cross-reacting substances, antianalyte antibodies or antireagent
antibodies, all of which may lead to erroneously high or low results. Low results may be observed for tumour markers due
to high-dose hooking in the presence of very high analyte concentrations. Erroneous results can occur unexpectedly with
any specimen and there is no practical means of identifying specimens likely to cause problems in immunoassays. The
possibility of interference should always be considered when results do not appear to be in accord with the clinical
picture. Errors can occur in even the best-managed laboratories and their early investigation is always desirable. If there is
any doubt whatsoever about a result, clinical staff should be encouraged to contact the laboratory. Investigations for
possible interference that can be undertaken in most laboratories include testing for linearity on dilution, recovery
experiments, treatment with heterophilic blocking tubes and confirmation using a different method. It may be desirable to
consult specialist laboratories if more complex studies are necessary. Informing clinical and laboratory staff of the ever-
present possibility of unexpected interference, ensuring brief clinical details are available to laboratory staff, and above
all facilitating excellent communication between laboratory and clinical staff are key to minimizing the risk of clinical
mismanagement due to unsuspected interference.
international efforts are being made to address this for ana- recognize both intact hCG and its free beta-subunit
lytes as diverse as insulin, parathyroid hormone (PTH), (hCGb) as some tumours may produce only the subunit.9
thyroid hormones, troponin and vitamin D.21 In a prototype In contrast, immunoassays which recognize only intact
International Federation of Clinical Chemistry and hCG are adequate for the diagnosis and monitoring of
Laboratory Medicine (IFCC) project, six International pregnancy.
Reference Reagents for clinically important forms of hCG A clearer understanding of what forms of an analyte are
were prepared to enable better characterization of what present in physiological specimens will be much facilitated
current hCG immunoassays measure.22 Availability of by the increasing availability of well-validated mass spectro-
these Reference Reagents, together with complementary metric methods. Already in use in many routine laboratories
antibody mapping studies,23 should facilitate development for measurement of relatively simple molecules such as
of clinically more relevant assays in the future.22 For use steroids, drugs of abuse and therapeutic drugs, technologi-
as a tumour marker, for example, immunoassays should cal advances mean that mass spectrometry is increasingly
being applied to study complex heterogeneous molecules.
A recent report describes the identification of new clinical
(a) variants of PTH and their measurement, together with
Correct result that of previously identified forms of PTH, using quantitat-
ive mass spectrometric immunoassays.24
(b) While availability of highly specific assays is undoubtedly
Positive cross-reaction necessary to improve understanding of the effect of disease
Cross-reactant- processes on circulating analyte concentrations, in clinical
two epitopes shared practice, use of assays with broader specificity is sometimes
advantageous. Table 2 provides some examples illustrating
(c) the importance of sound understanding of the specificity
Negative cross-reaction of the immunoassay used. For some analytes, observed
Cross-reactant- cross-reactions may be clinically beneficial (e.g. cross-
one epitope shared reaction of hCG in an assay for luteinising hormone [LH]
may be helpful in identifying unsuspected pregnancy in a
Figure 1 Cross-reactions in two-site immunoassay. (a) Antibodies bind to
specific analyte – correct result obtained; (b) cross-reactant sharing two epi-
woman with amenorrhoea1 or of recombinant insulin ana-
topes in common with analyte – positive cross-reaction; (c) cross-reactant logues in assays for insulin25) provided the characteristics
sharing only one epitope in common with analyte – negative cross-reaction. of the immunoassay are appreciated by the user, while for
(Reprinted with permission from Seth J, Sturgeon CM. Pitfalls in immunoas-
other analytes they highlight a need for immunoassays
says in endocrinology. Endocrinology and Metabolism In-Service Training and
Continuing Education. 1993;11(4):89–99. # American Association for Clinical with improved specificity (e.g. digoxin, growth hormone
Chemistry, Inc.) [GH]) or better standardization (e.g. PSA, PTH).
Sturgeon and Viljoen. Immunoassay error and interference 421
................................................................................................................................................
Table 2 Examples of analytes for which the analytical specificity of the immunoassay method used is likely to affect clinical interpretation
Effect of presence of
cross-reactant on analytical
Analyte Potential cross-reactants result obtained Clinical implications References
Chorionic hCG beta-subunit (hCGb) Potentially higher hCG results Use of an immunoassay recognizing 9
gonadotrophin obtained in immunoassays hCG þ hCGb essential for oncology
(hCG) recognizing hCGß applications as some testicular
cancers may produce only hCGb and
not intact hCG
Digoxin Aldosterone antagonists – e.g. Falsely elevated or lowered digoxin Not possible to assume 70,71
spironolactone, canrenoate concentrations interchangeability of immunoassays.
Analytical interferences in digoxin
immunoassays (both from
digoxin-like immunoreactive
substances and drug metabolites) are
a real problem, likely to become even
more important as lower therapeutic
ranges are recommended.
Confirmation in a physicochemical
method such as HPLC is highly
desirable
Growth hormone GH receptor antagonist – e.g. Falsely elevated or lowered GH Only certain immunoassays can be 72
(GH) Pegvisomant (Somavertw) concentrations used to measure GH in the presence
of Pegvisomant
Insulin Novel insulin analogues – e.g. Differences in cross-reactivity of Knowledge of such differences may be 73
Insulin Lispro (Humalogw), novel insulin analogues are critical for adequate assessment, e.g.
Insulin Detemir (Levemirw) reflected in discordant insulin of factitious hypoglycaemia
concentrations as measured in
different methods
Luteinising hCG Apparently measurable LH in early Failure to appreciate this when using an 74
hormone pregnancy, when LH and FSH are LH assay which cross-reacts with
both suppressed hCG may delay pregnancy diagnosis
in an amenorrhoeic patient
Parathyroid N-truncated fragments – e.g. Different results for patients with Not possible to assume 75
hormone (PTH) 7 –84 PTH chronic renal failure depending on interchangeability of immunoassays.
the assay used Establishment of appropriate
method-specific reference intervals
or cut-off values essential for clinically
effective application of clinical
guidelines
Prostate specific PSA complexed to Measurement of ‘total’ PSA Not possible to assume 26,76
antigen (PSA) a1-antichymotrypsin inhibitor influenced by relative recognition interchangeability of immunoassays.
of free and complexed forms of For non-equimolar assays,
PSA establishment of appropriate
method-specific decision points
essential for screening applications
Testosterone Other steroids – e.g. Potentially higher testosterone Inappropriate clinical referral and 77,78
dehydroepiandrosterone results in immunoassays cross- unnecessary further investigations
sulphate (DHAS) reacting with other steroids.
Most likely to be a problem in direct
non-extraction methods
Encouragingly, the latter is being proactively addressed for commercially available tests. Manufacturers’ advice should
PSA under the auspices of the NHS Cancer Screening always be followed, and documented procedures for treat-
Programme.26 Comparability of PTH results should be simi- ing specimens identified as haemolysed or lipaemic
larly improved by calibration of PTH immunoassays in should be available in all laboratories. Turbidity in a lipae-
terms of the recently established International Standard mic sample can usually be visually detected and, if
for PTH(1-84), IS 95/646.27 required, triglycerides removed from the sample either by
precipitation or by ultracentrifugation before repeating the
analysis on the clarified sample. Haemolysis is also readily
detectable provided specimens are visually examined
Lipids, haemoglobin and other serum
before analysis. Automated detection of haemolysis,
constituents icterus and lipaemia is possible: the advantages of this
There is considerable literature relating to the effect of excess approach, including a helpful algorithm, have been reported
concentrations of normal serum constituents on immunoas- for 28 analytes including some proteins.30 Immunoassay
says.20 This has been comprehensively collated by several results for some analytes (e.g. adrenocorticotrophic
authors20,28,29 and is also presented in data sheets for most hormone, gastrin, insulin and PTH) are particularly likely
422 Annals of Clinical Biochemistry Volume 48 September 2011
................................................................................................................................................
to be affected adversely by haemolysis20 and a repeat speci- interference in these specimens was found to be 4%, a pro-
men should always be requested. The possible effect of portion that could be reduced to 0.10% by removing the Fc
additives in specimen collection tubes should also be fragments from the capture antibody.7 This was reduced
considered.28 even further by adding heat-treated, non-specific murine
immunoglobulin to the assay buffer. Some manufacturers
of commercial immunoassays do their best to minimize
Antibody- and antibody-like interferences the risk of such interference, usually by adding non-immune
Since immunoassays rely on interactions between the animal serum samples or other blocking agents to the reac-
analyte being measured and reagent antibodies to produce tion buffer. While appreciating that no measure can be
the assay signal, it is not surprising that they are particularly relied upon to be effective in removing interference in
susceptible to interference from other unrelated antibodies every specimen – reflecting the biological variability
that may be present in the reaction mixture. It is important described above – more could certainly be done by some
to recognize that interfering antibodies may be present only manufacturers to reduce the risk of interference.
transiently in a patient’s serum, and that their characteristics The majority of published reports relate to interference
and reactivity may vary, such that no immunoassay can be from antireagent antibodies3,8,31 although other complex
considered to be completely robust to all possible interfer- formations can occur, as described below.
ence. Such interference, which again may be positive or
negative, is more likely to occur in two-site ‘sandwich’
assay formats, in which interfering antibodies may either Interference due to antireagent antibodies
interact with the analyte to prevent binding or may form Some patient serum samples contain antibodies capable of
a bridge between the signal and capture antibodies in the binding to animal immunoglobulins such as those used in
absence of analyte (Figure 2). In a large study of 11,261 immunoassays, although estimates of the frequency of
serum samples, carcinoembryonic antigen (CEA) was such antibodies vary widely (,1% to 80% is suggested in
measured using a two-site, two-step immunometric assay one review,31 0.4– 4% in another2), reflecting differences in
using mouse monoclonal antibodies.7 The frequency of detection methods and populations studied.31 However,
Anti-reagent Ab present
binding capture & labeled Ab
Anti-reagent Ab present
blocking either capture or labeled Ab
(d)
Correct result
Suppression of interference by
addition of non-immune IgG
Figure 2 Interference from antireagent antibodies in two-site immunoassays. (a) Interfering antibodies absent – correct result obtained; (b) cross-linking of
capture and labelled antibody in the presence of anti-reagent antibody – incorrectly high result (analyte may or may not be bound as well, depending on
steric factors, but in either case the result will be falsely high); (c) interfering antibodies binding to either capture or labelled antibody only, reducing sandwich
formation – incorrectly low result obtained; (d) effect of adding non-immune animal IgG to reduce interference – correct result obtained. (Reprinted with per-
mission from Seth J, Sturgeon CM. Pitfalls in immunoassays in endocrinology. Endocrinology and Metabolism In-Service Training and Continuing Education.
1993;11(4):89–99. # American Association for Clinical Chemistry, Inc.)
Sturgeon and Viljoen. Immunoassay error and interference 423
................................................................................................................................................
the mechanism of interference and its severity depend both likely to cause problems in two-step procedures incorporat-
on assay design and on the nature of the interfering anti- ing a wash step, but thyroid hormone results can still be
body (Table 3). In general, interferences caused by human misleading. In patients with suspected FDH, molecular
anti-mouse antibody (HAMA) are likely to be more pro- genetic testing is comparatively simple and provides an
nounced than those due to heterophilic antibody, while unambiguous diagnosis.33 Similarly, although their preva-
antibodies in serum samples from patients with rheumatoid lence is difficult to assess, the presence of antibodies to thy-
disease may not always be of sufficiently high affinity to roxine or tri-iodothyronine may affect some thyroid hormone
cause the bridging observed with HAMA or heterophilic methods, giving high or low results, while apparently not
antibodies (Figure 2). altering results in others.34 Antibodies to insulin may
cause very high insulin results.35
Subjects with differentiated thyroid carcinoma perimenopausal symptoms is due to antibody interference.
In patients with treated differentiated thyroid carcinoma, Failure to identify heterophilic antibody interference in
undetectable serum thyroglobulin concentration following serum from a patient with a favourable prognosis germ
TSH stimulation is considered the most reliable marker of cell tumour, for example, recently resulted in him receiving
cure.11,42 However, the presence of thyroglobulin antibodies several courses of toxic and unnecessary second- and
or heterophilic antibodies, respectively, may lead to false third-line chemotherapy.46
decreases and increases in measured concentration, as has Even when resources are limited, once interference is sus-
recently been described in detail.11,42,43 Thyroglobulin anti- pected, there are several simple steps that should always be
bodies are often present in patients with autoimmune taken before the need for more specialized investigations is
thyroid disease, with 10% of healthy individuals having considered. Possible steps are shown in the flow chart in
measurable levels.5 Quantification of thyroglobulin by Figure 3 and discussed below.
tandem mass spectrometry may ultimately circumvent the
problem of antibody interference.44 In the meantime, effec- Initial procedures as soon as interference
tive multidisciplinary collaboration is essential to minimize
is suspected
the risk of error, as has recently been comprehensively
described.43 As soon as an erroneous result is suspected, the specimen
identity should be confirmed before repeating the assay to
exclude the possibility of an analytical error. Where
Identifying results requiring investigation samples are subaliquoted (even by automated procedures),
for possible interference it is highly desirable to carry out the repeat assay on
While considerable information is now available about the serum from the original correctly labelled primary tube. If
nature of possible endogenous interferences that may be the result is confirmed on repeat analysis, relevant clinicians
encountered in clinical specimens, identifying such interfer- should be alerted by telephone, all relevant results relating
ence still relies almost entirely on a high index of clinical to the specimen removed from, and the reason flagged in,
suspicion, i.e. the perception that a result does not fit the the laboratory computer with an explanatory note, and a
clinical picture. The random and sporadic nature of most further specimen urgently requested. Depending on the
interferences means that unless a cohort of patients at par- timescale in which this specimen can be acquired, primary
ticular risk can be identified (as is the case for subjects investigations for possible interference should be performed
with confirmed hyperprolactinaemia whose serum may on the first or second specimen.
contain macroprolactin45), screening for possible interfer-
ence is unlikely to be either feasible or cost-effective.
Interference should always be suspected if results do not Preliminary investigations for suspected
correlate with the clinical picture. This is facilitated by a interference
good understanding of the limitations of immunoassays in
While unfortunately there are no tests that unequivocally
general as well as of the specific analytical characteristics
exclude the possibility of interference, several relatively
of the method used. Laboratory staff are well placed to con- simple investigations may confirm it (Figure 3).
tribute to this process, which is aided considerably by pro-
vision of brief clinical details on request forms. Clinical
colleagues should always be actively encouraged to Confirmation of results in other immunoassay methods
contact the laboratory if they have any doubt about a It is usually desirable for the investigating laboratory to send
result. When clinical staff are the first to suspect that an aliquot of the specimen to other laboratories for confir-
results may be incorrect, early dialogue is essential so that mation of the result by one or more different methods.
relevant specimens can be retrieved before they are dis- Ideally, at least one of these would depend on an alternative
carded and appropriate investigation instituted. methodology (e.g. radioimmunoassay), but in practice this is
usually not feasible, except for hCG and thyroglobulin (see
Appendix). Reference methods (e.g. equilibrium dialysis for
Investigating suspected antibody
free hormones) are clearly desirable, but are well-established
or antibody-like interference for relatively few analytes. If, after taking account of
The extent of investigations undertaken once interference is method-related differences in bias as reflected in EQA and
suspected is likely to be influenced to some extent by how other available data, results differ significantly, this pro-
much time and serum is available, but the most critical vides convincing evidence of interference although it will
consideration should be the possible implications for not necessarily identify which result is ‘correct’.
patient management. Establishing whether interference is
the cause of a clinically unexpected increase in a tumour
marker such as hCG – where erroneous results can and Dilution and recovery studies
have led to unnecessary chemotherapy or other treatment Serial dilution and recovery studies can be informative, as
and also have ongoing implications for serial monitoring – lack of linearity of results on dilution of the specimen (in
is much more critical than demonstrating whether an the diluent provided by the manufacturer) or low recovery
apparently high LH result in a woman with of known amounts of analyte (e.g. the assay standard)
Sturgeon and Viljoen. Immunoassay error and interference 425
................................................................................................................................................
(e
])
Figure 3 Flow chart showing sequence of investigations that might be undertaken to investigated suspected immunoassay interference
added to the serum being investigated are suggestive of other reports, it highlights the need to ensure that a sufficient
interference. Determining the limits of acceptability for concentration of HBR is present to suppress interference. To
such studies is difficult, however, and apparently satisfac- address this, some laboratories routinely treat specimens
tory linearity and/or recovery may be observed even in sequentially with more than one HBR tube when investi-
the presence of an interfering substance. Conversely, some gating potential interference. If this is done, the analyte
specimens that are apparently free from interfering sub- should be measured after each treatment until the same con-
stances do not dilute out linearly, particularly in some of centration is obtained in two consecutive determinations.
the more complex tumour marker assays (e.g. CA19-920),
making interpretation of results even more difficult.
Addition of non-immune animal serum
Adding non-specific immunoglobulins to the reaction mixture
Treatment with heterophilic blocking reagents
may reduce interference if the human antibodies bind to these
Many laboratories keep a supply of commercially available instead of binding the assay antibodies.7 Murine and bovine
heterophilic antibody blocking tubes, and treating the antibodies reduce interference in the highest percentage of
specimen with these according to the manufacturers’ patient’s samples and also have the highest avidity for hetero-
instructions is also desirable. The tubes are coated with a philic antibodies. Ingestion of bovine IgG in milk has been
heterophilic blocking reagent (HBR) directed against suggested to be one of the factors inducing heterophilic anti-
human heterophilic antibody, which is claimed to block bodies in some subjects, which might explain the effectiveness
interference by steric hindrance.47 Results that differ from of bovine IgG as a blocking agent.7 Polymerized IgG prep-
the original result following treatment with these tubes arations have been shown to be superior to native IgG.49
suggest interference, but obtaining the same result does Trial and error is required in determining the optimal concen-
not necessarily exclude it. tration to be added with appropriate controls included.
HBR is also commercially available in solution and is rec-
ommended for use at a concentration of 400 mg/mL.47
Interestingly, it has recently been shown that lower concen- Polyethylene glycol precipitation
trations of HBR appear to enhance interactions in a model At neutral pH, immunoglobulins are not particularly
system rather than blocking them.48 Whatever the expla- soluble and it is relatively easy to precipitate them.20
nation for this observation, which is in accord with some When investigating possible antibody interference in
426 Annals of Clinical Biochemistry Volume 48 September 2011
................................................................................................................................................
they bind to different IgG antibodies, both between species vulnerable to the high-dose hook effect at very high
and between different antibody subclasses from the same analyte concentration.64,65 Free analyte and bound analyte
species, so using both (Protein A/G) is attractive. Neither compete for the limited number of solid-phase antibody-
Protein A nor Protein G reacts with other classes of immu- binding sites, yielding inappropriately low results which
noglobulin (IgA, IgM, etc.) so they are not helpful in identi- may even hook back to within the reference interval
fying interfering antibodies of these classes. (Figure 4). The risk of reporting an incorrectly low result
Protein A/G is a gene fusion product that contains four Fc is greatest for analytes with wide physiological concen-
binding domains from Protein A and two from Protein G.61 tration ranges, e.g. AFP, CA125, CEA, hCG, PSA, prolactin
As it binds to all human IgG subclasses, it can be particu- and thyroglobulin. Although manufacturers usually
larly helpful in purifying IgG antibodies where the subclass specify a concentration below which hooking is unlikely to
identity is not known. Chromatography on immobilized occur in a particular immunoassay, the only means of com-
Protein A/G has recently been used successfully as a pre- pletely eliminating this risk is dilution of all specimens. This
treatment procedure to eliminate non-specific assay interfer- is likely to be both costly and over-cautious, particularly in
ences in individual serum samples from rheumatoid the absence of good data as to the frequency with which
arthritis patients62 (see Appendix for sample protocol). such high concentrations may occur.
While it is recommended that laboratories have in place a
defined procedure for minimizing the risk of an unidenti-
Assessment in an unprotected immunoassay fied hook,66 some hooked specimens are probably never
Commercially available immunoassays are protected identified.64 In practice, failure to recognize extremely high
against potential interference by addition of blocking AFP and hCG concentrations in conditions that are poten-
agents. Consequently, access to pairs of immunoassays, tially fatal but curable (e.g. AFP in childhood hepatoblas-
one of which is protected and one of which is not, is now toma and hCG in gestational trophoblastic neoplasia)
limited. However, paired immunoassays are still available constitutes a critical clinical error64 which, at best, is likely
in some specialist laboratories.7 Different results in such to delay treatment.67 Ensuring that such specimens are
assays provide convincing evidence of interference, but as always checked at dilution would therefore seem highly
with most other approaches, obtaining the same result in desirable. In some laboratories, all tumour marker results
both methods does not necessarily exclude interference. above a certain threshold (usually somewhat lower than
the highest standard concentration) are checked at dilution.
This requires a rather empirical assessment of the risk the
Assessment in an ‘interference’ or ‘non-sense’ assay laboratory is willing to take of missing an unexpectedly
An early ‘interference’ assay for the detection of multivalent high result.
antibody-binding substances in serum used the same anti- As always, provision of relevant clinical information and
body as both capture and signal antibody and was used to good communication with clinical staff facilitates identifi-
investigate interference in a two-site immunoradiometric cation of specimens that are likely to be problematic and
assay for hCG.63 The method relies on the principle that a
signal will only be obtained in the presence of a substance
that could bind two or more molecules of the one antibody,
which would not be the case for an analyte with only one
antibody-binding site. The same principle can be refined
using a combination of two unrelated assays, such as
solid phase hCG and tracer anti-a-fetoprotein (AFP).63
Non-specific binding must be reduced by using a buffer con-
taining a reasonably high concentration of added protein Excess analyte present
(bovine serum albumin and bovine IgG are recommended)
and the analyte must not contain repeating epitopes, so anti-
bodies to mucin antigens such as CA15-3 or CA19-9 would
not be suitable. The essential requirement for a non-sense
assay is that it measures nothing but interference.
Low result
for which additional dilutions would be desirable, for patients’ medical records and ideally in the laboratory infor-
example, when prolactin is measured for the first time in a mation system. In some cases it may simply not be possible
patient presenting with a visual field defect.64 to report a result and it is necessary to attribute this to tech-
nical difficulties. Recording this is essential.
investigations described above, is that awareness of how to Antibodies; Approved Guideline. Document I/LA30-A. Wayne, PA:
identify problem specimens as well as the required practical CLSI, 2009;28
6 Jones AM, Honour JW. Unusual results from immunoassays and the role
skills are then actively maintained at local level. For more of the clinical endocrinologist. Clin Endocrinol (Oxf ) 2006;64:234 – 44
complex investigations, referral to a specialist laboratory 7 Bjerner J, Nustad K, Norum LF, et al. Immunometric assay interference:
may be desirable (see Appendix). incidence and prevention. Clin Chem 2002;48:613 –21
8 Ismail AA, Walker PL, Barth JH, et al. Wrong biochemistry results: two
case reports and observational study in 5310 patients on potentially
misleading thyroid-stimulating hormone and gonadotropin
Conclusion immunoassay results. Clin Chem 2002;48:2023 –9
9 Sturgeon CM, Duffy MJ, Stenman UH, et al. National Academy of
There have been major advances in understanding the Clinical Biochemistry laboratory medicine practice guidelines for use of
mechanisms by which substances present in the occasional tumor markers in testicular, prostate, colorectal, breast, and ovarian
routine clinical specimen can cause false-positive or false- cancers. Clin Chem 2008;54:e11 – 79
negative results in immunoassays, and this understanding 10 Preissner CM, Dodge LA, O’Kane DJ, et al. Prevalence of heterophilic
antibody interference in eight automated tumor marker immunoassays.
has guided the development of in-house and commercially
Clin Chem 2005;51:208 –10
available assays with improved reliability. Nevertheless, 11 Giovanella L, Ghelfo A. Undetectable serum thyroglobulin due to
clinically misleading results due to endogenous interferents negative interference of heterophile antibodies in relapsing thyroid
in the test specimen do continue to occur sporadically. They carcinoma. Clin Chem 2007;53:1871 – 2
present a unique and serious threat to patient care because 12 Wu A, Collinson P, Jaffe A, et al. High-sensitivity cardiac troponin assays:
what analytical and clinical issues need to be addressed before
they are not detectable by normal laboratory quality
introduction into clinical practice? Interview by Fred S. Apple. Clin Chem
control procedures, are reproducible within the test 2010;56:886 –91
system, may be clinically plausible and are relatively rare. 13 British Thyroid Association. UK Guidelines for the use of thyroid
While the first line of defence in countering such threats is function tests. 2006. See http://www.british-thyroid-association.org/
the inclusion of assay robustness in the criteria for choice info-for-patients/Docs/TFT_guideline_final_version_ July_2006.pdf (last
checked 16 May 2011)
of test system, the importance of maintaining a high index 14 Surks MI, DeFesi CR. Normal serum free thyroid hormone
of suspicion in inspecting assay results cannot be overstated. concentrations in patients treated with phenytoin or carbamazepine. A
This is a responsibility of both laboratory and clinical staff, paradox resolved. JAMA 1996;275:1495 –8
and requires critical review of the test result in relation to 15 Dasgupta A. The effects of adulterants and selected ingested
other clinical features, especially where clinical interven- compounds on drugs-of-abuse testing in urine. Am J Clin Pathol
2007;128:491 – 503
tions are planned on the basis of the test result alone. 16 O’Kane M. The reporting, classification and grading of quality failures in
Procedures are available for the laboratory to check the medical laboratory. Clin Chim Acta 2009;404:28 –31
whether a suspicious result might be due to endogenous 17 Plebani M. The detection and prevention of errors in laboratory
interferents. medicine. Ann Clin Biochem 2010;47:101 – 10
18 Westgard JO, Westgard QC. Madison, Wisconsin, 2009. See http://www.
westgard.com/ (last checked 16 May 2011)
DECLARATIONS 19 Ungerer JP, Pretorius CJ, Dimeski G, et al. Falsely elevated troponin I
results due to outliers indicate a lack of analytical robustness. Ann Clin
Competing interests: None. Biochem 2010;47:242 –7
Funding: None. 20 Wild DG. The Immunoassay Handbook. 3rd edn. Oxford: Elsevier Ltd, 2005
21 Klee GG. Harmonization and standardization of thyroid function tests.
Ethical approval: Not applicable. Clin Chem 2010;56:879 –80
Guarantor: CMS. 22 Sturgeon CM, Berger P, Bidart JM, et al. Differences in recognition of the
Contributorship: CMS wrote the draft and AV contributed 1st WHO international reference reagents for hCG-related isoforms by
to the final manuscript. diagnostic immunoassays for human chorionic gonadotropin. Clin Chem
2009;55:1484 – 91
Acknowledgements: We would like to thank George Klee,
23 Berger P, Sturgeon C, Bidart JM, et al. The ISOBM TD-7 Workshop on
Kjell Nustad, John Seth and Joanna Sheldon for their hCG and related molecules. Towards user-oriented standardization of
careful reading of the manuscript and most helpful and pregnancy and tumor diagnosis: assignment of epitopes to the
much appreciated suggestions. three-dimensional structure of diagnostically and commercially relevant
monoclonal antibodies directed against human chorionic gonadotropin
and derivatives. Tumour Biol 2002;23:1– 38
REFERENCES 24 Lopez MF, Rezai T, Sarracino DA, et al. Selected reaction
monitoring-mass spectrometric immunoassay responsive to parathyroid
1 Seth J, Sturgeon C. Pitfalls in immunoassays in endocrinology. hormone and related variants. Clin Chem 2010;56:281 – 90
Endocrinology and Metabolism In-service Training and Continuing Education 25 Heald AH, Bhattacharya B, Cooper H, et al. Most commercial insulin
[AACC] 1993;11:89– 99 assays fail to detect recombinant insulin analogues. Ann Clin Biochem
2 Ismail AA. Interference from endogenous antibodies in automated 2006;43:306 –8
immunoassays: what laboratorians need to know. J Clin Pathol 26 Sturgeon CM, Ellis AR. Improving the comparability of immunoassays
2009;62:673 –8 for prostate-specific antigen (PSA): progress and problems. Clin Chim
3 Klee GG. Interferences in hormone immunoassays. Clin Lab Med Acta 2007;381:85 –92
2004;24:1 –18 27 Burns C, Moore M, Sturgeon C, et al. WHO International Collaborative
4 Sturgeon CM. Limitations of assay techniques for tumor Study of the Proposed 1st International Standard for Parathyroid
markers. Chapter 4. In: Diamandis EP, Fritsche HA, Lilja H, Chan DW, Hormone 18 –84, Human, Recombinant. Geneva: WHO, 2009.
Schwartz MK, eds. Tumor Markers: Physiology, Pathobiology, Technology See http://whqlibdoc.who.int/hq/2009/WHO_BS_09.2115_eng.pdf
and Clinical Applications. Chicago: AACC Press, 2002 (last checked 16 May 2011)
5 Howanitz JH, Bjerner J, Chace NM, et al. Clinical and Laboratory 28 Young DS. Effects of Preanalytical Variables on Clinical Laboratory Tests. 3rd
Standards Institute (CLSI). Immunoassay Interference by Endogenous edn. Washington, DC: AACC, 2007
430 Annals of Clinical Biochemistry Volume 48 September 2011
................................................................................................................................................
29 Guder WG, Narayanan S, Wisser H, et al. Diagnostic Samples: from the 55 Heald AH, Butterworth A, Kane JW, et al. Investigation into possible
Patient to the Laboratory: the Impact of Preanalytical Variables on the Quality causes of interference in serum testosterone measurement in women.
of Laboratory Results. 4th edn. Weinheim: Wiley-Blackwell, 2009 Ann Clin Biochem 2006;43:189 –95
30 Vermeer HJ, Thomassen E, de Jonge N. Automated processing of serum 56 Wood P. Salivary steroid assays – research or routine? Ann Clin Biochem
indices used for interference detection by the laboratory information 2009;46:183 –96
system. Clin Chem 2005;51:244 –7 57 Cawood ML, Field HP, Ford CG, et al. Testosterone measurement by
31 Kricka LJ. Human anti-animal antibody interferences in immunological isotope-dilution liquid chromatography-tandem mass spectrometry:
assays. Clin Chem 1999;45:942 –56 validation of a method for routine clinical practice. Clin Chem
32 Cartwright D, O’Shea P, Rajanayagam O, et al. Familial dysalbuminemic 2005;51:1472 – 9
hyperthyroxinemia: a persistent diagnostic challenge. Clin Chem 58 Maunsell Z, Wright DJ, Rainbow SJ. Routine isotope-dilution liquid
2009;55:1044 – 6 chromatography-tandem mass spectrometry assay for simultaneous
33 Halsall DJ, Fahie-Wilson MN, Hall SK, et al. Macro thyrotropin –IgG measurement of the 25-hydroxy metabolites of vitamins D2 and D3. Clin
complex causes factitious increases in thyroid-stimulating hormone Chem 2005;51:1683 –90
screening tests in a neonate and mother. Clin Chem 2006;52:1968 –9 59 Wong T, Shackleton C, Covey T, et al. Identification of the steroids in
34 Zouwail SA, O’Toole AM, Clark PM, et al. Influence of thyroid hormone neonatal plasma that interfere with 17 alpha-hydroxyprogesterone
autoantibodies on 7 thyroid hormone assays. Clin Chem 2008;54:927 –8 radioimmunoassays. Clin Chem 1992;38:1830 – 7
35 Paiva ES, Pereira AE, Lombardi MT, et al. Insulin autoimmune syndrome 60 Taylor PJ. Matrix effects: the Achilles heel of quantitative
(Hirata disease) as differential diagnosis in patients with high-performance liquid chromatography-electrospray-tandem mass
hyperinsulinemic hypoglycemia. Pancreas 2006;32:431 –2 spectrometry. Clin Biochem 2005;38:328 –34
36 Fahie-Wilson MN, John R, Ellis AR. Macroprolactin; high molecular 61 ThermoScientific. Comparison of Protein A G, A/G and L. Comparison
mass forms of circulating prolactin. Ann Clin Biochem 2005;42:175 – 92 of Protein A, G, A/G and L. See http://www.piercenet.com (last checked
37 Jassam NF, Paterson A, Lippiatt C, et al. Macroprolactin on the Advia 16 May 2011)
Centaur: experience with 409 patients over a three-year period. Ann Clin 62 McCutcheon KM, Quarmby V, Song A. Development and
Biochem 2009;46:501 –4 optimization of a cell-based neutralizing antibody assay using a sample
38 Bates KJ, Hall EM, Fahie-Wilson MN, et al. Circulating immunoreactive pre-treatment step to eliminate serum interference. J Immunol Methods
cardiac troponin forms determined by gel filtration chromatography 2010;358:35 –45
after acute myocardial infarction. Clin Chem 2010;56:952 – 8 63 Boscato L, Stuart M. Incidence and specificity of interference in two-site
39 Webster R, Fahie-Wilson M, Barker P, et al. Immunoglobulin interference immunoassays. Clin Chem 1986;32:1491 –5
in serum follicle-stimulating hormone assays: autoimmune and 64 Sturgeon C. Expecting the unexpected – the continued need for
heterophilic antibody interference. Ann Clin Biochem 2010;47:386 –9 vigilance in tumour marker assays. Ann Clin Biochem 2006;43:247 –8
40 Sakai H, Fukuda G, Suzuki N, et al. Falsely elevated thyroid-stimulating 65 Al-Mahdili HA, Jones GR. High-dose hook effect in six automated
hormone (TSH) level due to macro-TSH. Endocr J 2009;56:435 –40 human chorionic gonadotrophin assays. Ann Clin Biochem 2010;47:383 – 5
41 Jeffery J, Millar H, Mackenzie P, et al. An IgG complexed form of vitamin 66 Sturgeon CM, Hoffman BR, Chan DW, et al. National Academy of
B12 is a common cause of elevated serum concentrations. Clin Biochem Clinical Biochemistry laboratory medicine practice guidelines for use of
2010;43:82 –8 tumor markers in clinical practice: quality requirements. Clin Chem
42 British Thyroid Association. Guidelines for the Management of Thyroid 2008;54:e1 –e10
Cancer. 2nd edn. London: Royal College of Physicians, 2007 67 Jassam N, Jones CM, Briscoe T, et al. The hook effect: a need for constant
43 Clark PM. Laboratory services for thyroglobulin and implications for vigilance. Ann Clin Biochem 2006;43:314 – 7
monitoring of differentiated thyroid cancer. J Clin Pathol 2009;62:402 –6 68 abcNews. Misdiagnosis of cancer. See http://abcnews.go.com/
44 Hoofnagle AN, Becker JO, Wener MH, et al. Quantification of Primetime/story?id=132213&page=1 (last checked 16 May 2011)
thyroglobulin, a low-abundance serum protein, by immunoaffinity 69 Gorovits B. Antidrug antibody assay validation: industry survey results.
peptide enrichment and tandem mass spectrometry. Clin Chem AAPS J 2009;11:133 – 8
2008;54:1796 – 804 70 Steimer W, Muller C, Eber B. Digoxin assays: frequent, substantial, and
45 Beltran L, Fahie-Wilson MN, McKenna TJ, et al. Serum total prolactin and potentially dangerous interference by spironolactone, canrenone, and
monomeric prolactin reference intervals determined by precipitation other steroids. Clin Chem 2002;48:507 – 16
with polyethylene glycol: evaluation and validation on common 71 Cobo A, Martin-Suarez A, Calvo MV, et al. Clinical repercussions of
immunoassay platforms. Clin Chem 2008;54:1673 – 81 analytical interferences due to aldosterone antagonists in digoxin
46 Ballieux BE, Weijl NI, Gelderblom H, et al. False-positive serum human immunoassays: an assessment. Ther Drug Monit 2010;32:169 –76
chorionic gonadotropin (HCG) in a male patient with a malignant germ 72 Bidlingmaier M, Freda PU. Measurement of human growth hormone by
cell tumor of the testis: a case report and review of the literature. immunoassays: current status, unsolved problems and clinical
Oncologist 2008;13:1149 –54 consequences. Growth Horm IGF Res 2010;20:19– 25
47 Scantibodies. Scan-Brief: Scantibodies Heterophilic Blocking Reagent 73 Krull I, Sahli R, Diem P, et al. Variability in cross-reactivity of novel
(HBR). Santee, CA: Scantibodies. See http://www.scantibodies.com/ insulin analogues in immunometric insulin assays. Diabet Med
ScanBriefV1N1.pdf (last checked 16 May 2011) 2009;26:1075 – 6
48 Koshida S, Asanuma K, Kuribayashi K, et al. Prevalence of human 74 Vivekanandan S, Andrew CE. Cross-reaction of human chorionic
anti-mouse antibodies (HAMAs) in routine examinations. Clin Chim Acta gonadotrophin in Immulite 2000 luteinizing hormone assay. Ann Clin
2010;411:391 – 4 Biochem 2002;39:318 –9
49 Bjerner J. Human anti-immunoglobulin antibodies interfering in 75 Souberbielle J-C, Cavalier E, Jean G. Interpretation of serum parathyroid
immunometric assays. Scand J Clin Lab Invest 2005;65:349 – 64 hormone concentrations in dialysis patients: what do the KDIGO
50 Fahie-Wilson M, Halsall D. Polyethylene glycol precipitation: proceed guidelines change for the clinical laboratory? Clin Chem Lab Med
with care. Ann Clin Biochem 2008;45:233 –5 2010;48:769 –74
51 Ellis MJ, Livesey JH. Techniques for identifying heterophile antibody 76 Stephan C, Kopke T, Semjonow A, et al. Discordant total and free
interference are assay specific: study of seven analytes on two automated prostate-specific antigen (PSA) assays: does calibration with WHO
immunoassay analyzers. Clin Chem 2005;51:639 –41 reference materials diminish the problem? Clin Chem Lab Med
52 Ram S, Harris B, Fernando JJ, et al. False-positive polyethylene glycol 2009;47:1325 – 31
precipitation tests for macroprolactin due to increased serum globulins. 77 Warner MH, Kane JW, Atkin SL, et al. Dehydroepiandrosterone sulphate
Ann Clin Biochem 2008;45:256 –9 interferes with the Abbott Architect direct immunoassay for testosterone.
53 Smith TP, Fahie-Wilson MN. Reporting of post-PEG prolactin Ann Clin Biochem 2006;43:196 –9
concentrations: time to change. Clin Chem 2010;56:484– 5 78 Middle JG. Dehydroepiandrostenedione sulphate interferes in many
54 ACOG Committee on Gynecologic Practice. Avoiding inappropriate direct immunoassays for testosterone. Ann Clin Biochem 2007;44:173– 7
clinical decisions based on false-positive human chorionic gonadotropin 79 Sigma-Aldrich. Protein A antibody purification kit. Technical bulletin.
test results. Int J Gynaecol Obstet 2003;80:231 –3 St Louis, MO: Sigma-Aldrich
Sturgeon and Viljoen. Immunoassay error and interference 431
................................................................................................................................................
80 Boscato L, Stuart M. Heterophilic antibodies: a problem for all there is insufficient sample to analyse without dilution,
immunoassays. Clin Chem 1988;34:27 –33 the specimen can be diluted with PBS;
(4) Multiply the prolactin concentrations by 2 to correct for
(Accepted 14 April 2011)
the dilution with PEG. Additional multiplication will be
required for diluted specimens.
bicarbonate/carbonate coating buffer [3.03 g the points being made. Their inclusion is not intended in
NA2CO3 þ 6.0 g NAHCO3 in 1.0 L distilled water, any way to constitute a recommendation or otherwise.)
adjusted to pH 9.6]);
(2) Incubate overnight at 48C and then wash three times Specialist laboratories
with wash solution (0.05% v/v Tween-20 in PBS Contact details for specialist laboratories with particular
[1.16 g Na2HPO4, 0.1 g KCl, 0.1 g K3PO4 and 4.0 g expertise in investigation of samples giving apparently aty-
NaCl in 0.5 L distilled water, adjusted to pH 7.4]); pical results include the following:
(3) Add 200 mL of blocking buffer (1% w/v gelatin in PBS)
and incubate for one hour at room temperature on an † Human chorionic gonadotrophin (hCG) – Professor M Seckl,
orbital shaker; Medical Oncology Department, Charing Cross Hospital,
(4) Discard blocking buffer; Fulham Palace Road, London W6 8RF, UK;
(5) Add 100 mL of sample. Incubate for two hours on an † Peptide hormones and tumour markers – Dr C Sturgeon, UK
orbital shaker at room temperature; National External Quality Assessment Service (UK
(6) Wash wells three times with wash solution as above; NEQAS [Edinburgh]), Department of Clinical Biochemistry,
(7) Add 100 mL of horse radish peroxidase (HRP)-labelled Royal Infirmary, Edinburgh EH16 4SA, UK;
rabbit polyclonal anti-human hCG antibody (Cat No. † Specialist immunology – Dr J Sheldon, Protein Reference
AB30451; Abcam; 1:4000) or HRP-labelled mouse Unit, St George’s Hospital, Blackshaw Road, London
monoclonal anti-human AFP antibody (Cat No. SW17 0NH, UK;
AB10072; Abcam; 1:16,000) in blocking buffer and † Steroid hormones – Dr J Barth, Department of Clinical
incubate for one hour on orbital shaker at room Biochemistry & Immunology, Leeds General Infirmary,
temperature; Leeds LS1 3EX, UK;
(8) Wash wells three times with wash solution as above; † Thyroid hormones (including equilibrium dialysis for free T4
(9) Add 100 mL of tetramethylbenzidine solution (Cat No. and gel filtration chromatography for TSH) – Dr D Halsall,
T8665; Sigma-Aldrich Company Ltd, Dorset, UK) and Department of Clinical Biochemistry, Addenbrooke’s
incubate for 10 min at room temperature; Hospital, Hill Road, Cambridge CB2 2QQ, UK;
(10) Add 50 mL of sulphuric acid (1 mol/L); † Thyroglobulin, atypical thyroid hormone results (including
(11) Using a plate reader, measure absorbance at 450 nm. screening for familial dysalbuminaemic hyperthyroxinaemia,
anti-T4 and anti-T3 antibodies) and a-subunit assays – Dr PMS
Clark, Regional Endocrine Laboratory, Department of
(Where suppliers have been mentioned in this review, their Clinical Biochemistry, University Hospital Birmingham
products have been selected solely because they illustrate NHS Foundation Trust, Birmingham B29 6JD, UK.