TERUFUSION ® Infusion Pump TE-172

Specifications
Product Name TERUFUSION Infusion Pump TE-172 Operating Conditions Temperature: +5 to +40oC, humidity: 20 to 90%
Model TE-172 Storage Conditions Temperature: –20 to + 45oC, humidity: 10 to 95% (No condensation)

TERUFUSION Infusion Pump TE-172


Pumping System MIDPRESS peristaltic finger system (Provides accurate infusion without Power Suppy AC 100-120V/AC 100 – 240V, 50/60Hz DC12 –15V

Accuracy
pressing the tubing completely.)
Within +3% [Conditions] Infusion set: Specified TERUFUSION Infusion sets
Internal Nickel-Cadmium battery
Operating time: approx.3 hours
®
(Nominal value) type E for pump, solution: water/saline, test time: second hour, delivery Conditions: new battery, delivery rate: 25mL/h,
rate: 1.0mL/h or more, temperature: 25oC temperature: 25oC, with the power off, after 15-hour charging
Within +5% [Conditions] Infusion set: Specified TERUFUSION Infusion sets Power Consumption 24VA (AC 100 – 120V)/28VA (AC 100 – 240V), 11W (DC power)
for pump, solution; water/saline, test time: second hour, delivery rate: Classification Class I & internal power supply: TypeCF/IPX1 continuous operation
1.0mL/h or more, temperature: 25oC Dimensions 100 (W) x 215 (H) x 208 (D) mm (Excl. protrusions)
Delivery Rate 0.1 – 1,200mL/h (0 – 99.9mL/h: 0.1mL/h step, 100 – 1,200mL/h: Weight 3.7kg (TE-172)
1mL/h step). The delivery rate can be set in 1mL/h steps throughout the Body Weight Mode Dose rate: 0.01 — 99.99μg/kg/min. (0.01μg/kg/min. step)
range by using the internal mode select switch. (Dose Rate Calculation) Body weight: 0.1 — 300.0kg (0.1 kg step)
Delivery Limit 1 – 9,999mL/h (1mL step) or “----” (delivery limit free) Drug mass: 0.1 — 999.9mg (0.1 mg in step)
Volume Delivered 0.0 – 9,999mL (0.0 – 99.9mL: 0.1mL steps, 100 – 9,999mL/h: Solution volume: 0.1 — 999.9mL (0.1mL step)
1mL steps). The volume delivered reading is in 1 mL steps from 1 to Memory Function The latest five patterns of the above four parameters are automatically
9,999mL by using the internal mode select switch. saved and can be displayed.
Purging Higher than 500mL/h Specifications and design are subject to change without notice.
Alarms AIR/OCCLUSION/FLOW ERR./DOOR/LOW BATT. *Computer interface is available on the specified product.
*Operating indicator flashes red, and the alarm message comes on.
Pump stops except during the “LOW BATT.” Alarm.
Occlusion Detection 13.3 – 133.3kPa (100 – 1,000mmHg, 0.1 – 1.4kgf/cm2) Units available
Pressure The detection pressure can be adjusted in 3 levels (L, M, H) with
Reorder Voltage Plug type Socket type Body weight Computer
in the above range.
number mode interface
Completion Function When the total volume delivered reaches the preset delivery limit, the
TE]172CX1 100 to 120V American yes yes
indicator and buzzer notify the operator.
TE]172CW2 100 to 240V UK yes yes
Nurse Call Alarm Outputs an alarm relay contact. (Can be connected to the nurse call system)
TE]172CW3 100 to 240V European yes yes
Contact rating: DC12V 1.0A
TE]172CW3* 100 to 240V European yes yes
Computer Interface* RS-232C/RS-485
Special Functions • “Keep vein open” function (After the delivery limit has been reached, *Separate setup procedures are required for products with voltage-plug combinations other than those listed above.
delivery continues at 1mL/h for a set flow rate of > 1mL/h: or at the delivery
rate setting < 1nL/h.)
• Tubing clamp function (When the door is opened, tubing is automatically
clamped.)
• Volume delivered clear function
• Time remaining display
• Battery capacity indicator (in 3 levels)
• Adjustable display brightness (Can be switched to 2 levels.)
• Adjustable buzzer volume (in 3 levels)
• Operation history function (Storage/display of operation history)
• Standby function
• Repeat alarm function
• Start reminder function
The following functions can be selected via internal switches:
• Volume memory fucntion
• Delivery rate 1mL/h steps

0197

The outstanding reliability of Terumo’s MIDPRESS puts you at ease


TM
User-friendly features and easy operation
TERUFUSION Infusion Pump TE-172
®
in a single space-saving unit

Terumo developed an innovative MIDPRESS system that uses an improved peristaltic finger method.
TM
Selectable three-step (L, M and H) Informative alarms and added safety
occlusion detection pressures functions
With the conventional peristaltic finger method, solutions are administered by occluding As versatile as they are reliable, these infusion User-friendly alarms and safety functions
Conventional Body weight mode
the solution-administration tube until it becomes completely flat. peristaltic finger method pumps provide a variety of pressure level include: air-in-line, occlusion, irregular flow,
The TE-172 automatically calculates and sets settings (three in all) to match virtually any door alarm, low battery, re-alarm, start
the proper flow rate when the dose rate, body
With the MIDPRESS method, however, solutions are administered with
TM
requirement. For earlier alarm notification reminder alarms and self check, and a front
Completely flat weight, drug mass and solution volume are during occlusion, the detection pressure can panel lock-out safety functions.
high accuracy by pressing the solution-administration tube only halfway to closure. entered. This eliminates the need for time- be set at the lowest setting.
consuming calculations and having to refer to
Easy-to-see operation indicators
By precisely controlling the finger and stroke through the MIDPRESS , a stable volume is
TM
Completely open a conversion table.
1) Press the [DISPLAY ON/OFF] switch and For easy confirmation from any angle, the
delivered and a more accurate flow rate is achieved. the body weight mode screen lights up. indicator lamp protrudes from the unit.
2) Press [SELECT] switch to select between A green lamp blinks to indicate that operation is
MIDPRESS System
TM
the dose rate (þg/kg/min.), body weight in progress, while a flashing red warning lamp
(kg), drug mass (mg), and solution volume informs the user that a problem has occurred.
(mL).
3) The flow rate is automatically calculated
Tube doesn’t open Tube is pressed and displayed after entering the specified
completely. halfway to closure. dose rate, body weight, drug mass and
solution volume.
A A
Dose rate (þg/kg/min.) x Body weight (kg) x Solution volume (mL)
B B Flow rate (mL/h) = x 60 (*1)
Drug mass (mg) x 1000 (*2)

C C *1 Coefficient to convert “minute Umin.I” to “hour (h)”


*2 Coefficient to convert “mg” to “µg”
D D
E
E

Connects to the drip sensor (option)


Inflow Inflow stops Outflow Outflow Outflow stops Pulse flow adjustment For safety, this drip sensor (option) can double-
Solution flows into the tube. A certain amount of Outflow starts as part D Outflow continues as part C Part D presses the tube fully. Part E presses the tube and
check irregular flow rates.
solution accumulates in opens and part B presses presses the tube. completes outflow cycle.
the tube as part A presses the tube.
against the tube.

• High accuracy
High flow-rate accuracy is achieved: within +3% using the specified
infusion sets type E for pump and within +5% using conventional
infusion sets for pump. The MIDPRESSTM method succeeds in minimizing
flow-rate changes even over long periods, particularly for the delivery of
vasopressor or depressor drugs.
Conventional time remaining indication History function
function Historical data of infusion records: warnings,
• Wide range of flow rate settings is possible —
After the flow rate and volume to be delivered settings, starts and stops of infusion, etc.
from very low to high.
have been input, the time required to complete
Flow rates can be set within a wide range (from 0.1mL/h to 1,200mL/h). Nurse call connector available
the delivery is automatically indicated.
This makes it possible to perform infusions in a wide variety of situations
and applications, such as small infusions for newborn babies, rapid Options

infusion for emergency care and the delivery of parental nutrition.

• Transfusion TERUFUSION Solution


Administration Set for pump
Since the MIDPRESSTMsystem doesn’t press the tube completely flat, (TI*PU 200 L/300L)

damage to hemocytes is reduced. This means the TE-172 can be


used for blood transfusions in pediatric or ICU cases with the specified
extension tube for pumps.
TERUFUSION Volumetric
Solution Administration Set for
• Solid aluminum die-cast frame Convenient battery indicator pump (TS*PM series)
AC/DC
Easy-to-read, large display The TE-172 features a series of
The TERUFUSION Infusion Pump TE-172 has a BATTERY
informative battery indicator lamps for
precision machined solid aluminum die-cast frame for the Brightness adjustable display easy confirmation of the onboard battery F E
base of the MIDPRESSTM finger part. This structure helps power* (three levels).
The brightness of the display can be lowered to Battery operation time reference: SAFEED Extension Tube for
use with pumps
ensure highly accurate flow rates. allow a patient to enjoy undisturbed sleep. *Operates for approximately there hours Approx. 160 minutes or longer XX*BPH165L
in continuous use.
Approx. 80 minutes or longer
Door light for easy nighttime use Approx. 30 minutes or longer**
The light inside the door is convenient for **When the low-battery alarm is not on.
nighttime use.