Download as pdf or txt
Download as pdf or txt
You are on page 1of 3

LOT No.

MEK5D235N
HEMATOLOGY CONTROL   NORMAL EXP. DATE 2023-08-05
MEK-5DN for WBC 5 part differential 3.0mL   ASSAY SHEET OPEN VIAL STABILITY 14 DAYS

MEK MEK MEK MEK CAUTION :

-7222
*2
-7300 *2
-8222 *1*2
-9100 *2 Do not use the hematology control in
PARAMETER UNIT RANGE hematology analyzers not listed here.
*2
-9200
ASSAY VALUES ASSAY VALUES ASSAY VALUES ASSAY VALUES The hematology control must be stored at
temperatures between 2 and 8゚C.
WBC 103/μL ±0.8 8.1 8.0 7.9 7.96
NE% % ±15.0 50.5 50.5 50.5 50.69 If the hematology control is stored out of this
temperature range, LY% and LY values may
LY% % ±15.0 30.9 30.9 30.9 32.82 become high.
MO% % ±5.0 13.3 13.3 13.3 10.66
EO% % ±5.0 3.9 3.9 3.9 4.86 Please note that the assay value ranges are
reviewed lot by lot and are subject to change.
BA% % ±4.0 1.4 1.4 1.4 0.97
NE 3
10 /μL ±1.5 4.1 4.0 4.0 4.03 *1 Use MEK-8222 software version 02-03 or
later.
LY 103/μL ±1.5 2.5 2.5 2.4 2.61
MO 3
10 /μL ±0.5 1.1 1.1 1.1 0.85 *2 Assay values of WBC 5 part differentials
are for the case of optical adjustment with
3
EO 10 /μL ±0.5 0.3 0.3 0.3 0.39 MEK-CAL hematology calibrator.
3
BA 10 /μL ±0.4 0.1 0.1 0.1 0.08
*3 RDW is displayed as RDW-CV on MEK-
6
RBC 10 /μL ±0.20 4.55 4.48 4.53 4.45 7300/-9100,9200.
HGB g/dL ±0.5 13.1 13.0 12.9 13.04
HCT % ±3.0 42.0 40.2 39.6 45.0
MCV fL ±6.0 92.3 89.8 87.4 101.1
MCH pg ±2.8 28.8 29.0 28.5 29.3
MCHC g/dL ±3.0 31.2 32.3 32.6 29.0
*3
RDW % ±4.0 14.2 15.2 14.4 14.8
3
PLT 10 /μL ±50 254 269 248 285.8 The accompanying handy bar-code reader
MPV fL ±2.0 4.5 6.9 4.8 9.6 to MEK-9100,9200 can read the above QR
code and load assay values into its
ISOTONAC・3 hematology analyzer.
DILUENT
ISOTONAC・4 The word “QR Code” is registered trademark
of DENSO WAVE INCORPORATED in
HEMOLYNAC・3N HEMOLYNAC・310 Japan and other countries.
HEMOLYSING REAGENT
HEMOLYNAC・5 HEMOLYNAC・510

NIHON KOHDEN CORPORATION


1-31-4 NISHIOCHIAI, SHINJUKU-KU

ebdaigcecagibejejbagcakebjegiae
MEK-5DN ebdfjafajejbakelcagiae
235N ebdfjabif jafajbejbifibfbejbifejbgcagiae
2023-08-05
TOKYO 161-8560 JAPAN

0694-901691K
LOT No. MEK5D235H
HEMATOLOGY CONTROL   HIGH EXP. DATE 2023-08-05
MEK-5DH for WBC 5 part differential 3.0mL   ASSAY SHEET OPEN VIAL STABILITY 14 DAYS

MEK MEK MEK MEK CAUTION :

-7222
*2
-7300 *2
-8222 *1*2
-9100 *2 Do not use the hematology control in
PARAMETER UNIT RANGE hematology analyzers not listed here.
*2
-9200
ASSAY VALUES ASSAY VALUES ASSAY VALUES ASSAY VALUES The hematology control must be stored at
temperatures between 2 and 8゚C.
WBC 103/μL ±2.2 24.0 23.9 23.4 24.31
NE% % ±10.0 70.9 70.9 70.9 71.14 If the hematology control is stored out of this
temperature range, LY% and LY values may
LY% % ±10.0 14.3 14.3 14.3 15.52 become high.
MO% % ±5.0 10.2 10.2 10.2 8.05
EO% % ±5.0 3.6 3.6 3.6 4.41 Please note that the assay value ranges are
reviewed lot by lot and are subject to change.
BA% % ±4.0 1.0 1.0 1.0 0.88
NE 3
10 /μL ±2.8 17.1 17.0 16.7 17.30 *1 Use MEK-8222 software version 02-03 or
later.
LY 103/μL ±2.8 3.4 3.4 3.3 3.77
MO 3
10 /μL ±1.0 2.4 2.4 2.4 1.96 *2 Assay values of WBC 5 part differentials
are for the case of optical adjustment with
3
EO 10 /μL ±1.0 0.9 0.9 0.8 1.07 MEK-CAL hematology calibrator.
3
BA 10 /μL ±0.8 0.2 0.2 0.2 0.21
*3 RDW is displayed as RDW-CV on MEK-
6
RBC 10 /μL ±0.25 5.64 5.60 5.59 5.63 7300/-9100,9200.
HGB g/dL ±0.7 18.3 18.1 18.2 18.79
HCT % ±4.0 56.3 54.8 53.2 61.5
MCV fL ±8.0 99.9 97.8 95.2 109.2
MCH pg ±2.8 32.4 32.3 32.6 33.4
MCHC g/dL ±3.0 32.5 33.0 34.2 30.6
*3
RDW % ±4.0 13.9 14.8 14.1 15.4
3
PLT 10 /μL ±70 625 615 623 650.6 The accompanying handy bar-code reader
MPV fL ±4.0 5.1 5.8 5.2 6.8 to MEK-9100,9200 can read the above QR
code and load assay values into its
ISOTONAC・3 hematology analyzer.
DILUENT
ISOTONAC・4 The word “QR Code” is registered trademark
of DENSO WAVE INCORPORATED in
HEMOLYNAC・3N HEMOLYNAC・310 Japan and other countries.
HEMOLYSING REAGENT
HEMOLYNAC・5 HEMOLYNAC・510

NIHON KOHDEN CORPORATION


1-31-4 NISHIOCHIAI, SHINJUKU-KU

ebdaigcecagibejejbagcgachabgiae
MEK-5DH ebdfjafajejbgacdfagiae
235H ebdfjabif jafajbejbifibfbejbifejbgcagiae
2023-08-05
TOKYO 161-8560 JAPAN

0694-901692K
LOT No. MEK5D235L
HEMATOLOGY CONTROL   LOW EXP. DATE 2023-08-05
MEK-5DL for WBC 5 part differential 3.0mL   ASSAY SHEET OPEN VIAL STABILITY 14 DAYS

MEK MEK MEK MEK CAUTION :

-7222
*2
-7300 *2
-8222 *1*2
-9100 *2 Do not use the hematology control in
PARAMETER UNIT RANGE hematology analyzers not listed here.
*2
-9200
ASSAY VALUES ASSAY VALUES ASSAY VALUES ASSAY VALUES The hematology control must be stored at
temperatures between 2 and 8゚C.
WBC 103/μL ±0.6 3.0 3.0 3.0 2.99
NE% % ±15.0 36.7 36.7 36.7 38.41 If the hematology control is stored out of this
temperature range, LY% and LY values may
LY% % ±15.0 42.3 42.3 42.3 46.36 become high.
MO% % ±5.0 15.2 15.2 15.2 10.07
EO% % ±5.0 2.4 2.4 2.4 2.92 Please note that the assay value ranges are
reviewed lot by lot and are subject to change.
BA% % ±4.0 3.4 3.4 3.4 2.24
NE 3
10 /μL ±0.5 1.0 1.0 1.0 1.14 *1 Use MEK-8222 software version 02-03 or
later.
LY 103/μL ±0.5 1.3 1.3 1.3 1.39
MO 3
10 /μL ±0.2 0.5 0.5 0.5 0.30 *2 Assay values of WBC 5 part differentials
are for the case of optical adjustment with
3
EO 10 /μL ±0.2 0.1 0.1 0.1 0.09 MEK-CAL hematology calibrator.
3
BA 10 /μL ±0.2 0.1 0.1 0.1 0.07
*3 RDW is displayed as RDW-CV on MEK-
6
RBC 10 /μL ±0.15 2.66 2.64 2.67 2.53 7300/-9100,9200.
HGB g/dL ±0.4 6.5 6.5 6.4 6.63
HCT % ±2.0 22.2 21.4 21.0 22.9
MCV fL ±7.0 83.5 80.9 78.8 90.6
MCH pg ±2.4 24.4 24.6 24.0 26.2
MCHC g/dL ±3.0 29.3 30.4 30.5 29.0
*3
RDW % ±4.0 15.1 16.5 15.4 15.1
PLT 103/μL ±30 48 54 48 56.0 The accompanying handy bar-code reader
MPV fL ±3.0 6.1 7.6 6.6 10.8 to MEK-9100,9200 can read the above QR
code and load assay values into its
ISOTONAC・3 hematology analyzer.
DILUENT
ISOTONAC・4 The word “QR Code” is registered trademark
of DENSO WAVE INCORPORATED in
HEMOLYNAC・3N HEMOLYNAC・310 Japan and other countries.
HEMOLYSING REAGENT
HEMOLYNAC・5 HEMOLYNAC・510

NIHON KOHDEN CORPORATION


1-31-4 NISHIOCHIAI, SHINJUKU-KU

ebdaigcecagibejejbagc eib ggiae


MEK-5DL ebdfjafajejbceielegiae
235L ebdfjabif jafajbejbifibfbejbifejbgcagiae
2023-08-05
TOKYO 161-8560 JAPAN

0694-901690K