Professional Documents
Culture Documents
C4. Performance Evaluation Test - Test
C4. Performance Evaluation Test - Test
C4. Performance Evaluation Test - Test
1. Purpose
2. Samples used for determining cut-off value
3. Methods of determining cut-off value
4. Specific test data about determining cut-off value
5. Verification and confirmation of determining cut-off value
6. Summary of determining cut-off value
1. Purpose
The Six Respiratory Pathogens Nucleic Acid Diagnostic Kit
(PCR-Fluorescence Probing) produced by our company (hereinafter referred
as this kit) use multiple fluorescent quantitative PCR methods to detect
whether the throat swab samples of patients with similar symptoms of
respiratory tract infection caused by unknown reason, such as fever, swelling
and pain, cough, headache and dyspnea, contain the specific targets of nucleic
acid of influenza A virus (Flu A), influenza B virus (Flu B), Respiratory
Syncytial Virus (RSV), Adenoviruses (Adv), Human rhinovirus (HRV) and
Mycoplasma Pneumonia (MP), so as to determine whether the suspected
patients are infected with the above 6 respiratory pathogens. In order to
ensure the validity of detection results of targets (targeted nucleic acid
sequences), an internal standard (a positive internal control) is added to PCR
system to monitor the validity of PCR reaction and to avoid false negative.
The following is the basis for determining the cut-off value of the target gene
and internal standard detection through experimental research and statistical
analysis, which is used as the basis for judging the results of clinical samples.
normal saline to the detection limit concentration and to 48 parts; The positive
sample for Human rhinovirus with a concentration of 2.95E+07 copies/mL
were diluted with normal saline to the detection limit concentration and to 48
parts; The positive sample for Mycoplasma pneumoniae with a concentration
of 6.24E+07 copies/mL were diluted with normal saline to the detection limit
concentration and to 48 parts.
3) 2076 clinical throat swab samples (Clinical trail)
The clinical samples are all from clinical research. There were a total of
2076 throat swab samples collected for detection. Among them, 422 samples
were collected from Wuhan infectious disease hospital, 420 samples were
collected from Jiangxi provincial children’s hospital, 410 samples were
collected from the affiliated hospital of Hangzhou normal university, 414
samples were collected from Guangzhou women and children medical center,
and 410 samples were collected from Hunan provincial people’s hospital.
Figure 1 Amplification results of the clinically strong positive throat swab sample for
influenza A virus
25 30.75 28.06
2.5 34.69 31.02
0.5 Undetermined 36.96
0.2 Undetermined Undetermined*
*Note:Being an internal standard, it will be diluted with the dilution of
samples, so there is the detection result of negative internal.
It can be seen from the above graph that the sample with 2.5 TCID50/mL
virus was detected as positive, while 0.5 TCID50/mL was detected as negative.
Because of the difference in concentration, it is necessary to make gradient
dilution by small intervals for the sample with 2.5 TCID50/mL virus, to
determine the measurable concentration between 2.5 TCID50/mL and 0.5
TCID50/mL.
The above samples were diluted to 2.0 TCID50/mL, 1.5 TCID50/mL and 1.0
TCID50/mL and were repeatedly detected 8 times. The test results are shown
in the following Table 2 and Figures 3-5.
Table 2. Sensitivity detection
Detected positive
Diluted samples Positive rate(%)
samples/Sample size
2.0 TCID50/mL 8/8 100%
1.5 TCID50/mL 8/8 100%
1.0 TCID50/mL 7/8 87.5%
Reed-Muench method, was detected by the qualified kit in our company. The
results are shown in Figure 6.
Figure 6 Amplification results of the clinically strong positive throat swab sample for
influenza B virus
4.1.2.2 Serial dilution, detection sensitivity determination
Firstly, the clinically strong positive throat swab sample for influenza B
virus (5.0 × 103 TCID50/mL) was diluted with Hanks solution to 2.5 × 103
TCID50/mL、2.5 × 102 TCID50/mL, 25 TCID50/mL, 2.5 TCID50/mL, 0.5 TCID50/mL
and 0.2 TCID50/mL, and then was detected with this kit. Detection results are
shown in the following Figure 7.
25 30.16 28.28
2.5 33.30 31.18
0.5 Undetermined 39.20
0.2 Undetermined 37.64
It can be seen from the above graph that the sample with 2.5 TCID50/mL
virus was detected as positive, while 0.5 TCID50/mL was detected as negative.
Because of the difference in concentration, it is necessary to make gradient
dilution by small intervals for the sample with 2.5 TCID50/mL virus, to
determine the measurable concentration between 2.5 TCID50/mL and 0.5
TCID50/mL.
The above samples were diluted to 2.0 TCID50/mL, 1.5 TCID50/mL and 1.0
TCID50/mL and were repeatedly detected 8 times. The test results are shown
in the following Table 4 and Figures 8-10.
Table 4. Sensitivity detection
Detected positive
Diluted samples Positive rate(%)
samples/Sample size
2.0 TCID50/mL 8/8 100%
1.5 TCID50/mL 8/8 100%
1.0 TCID50/mL 7/8 87.5%
Figure 10. Amplification results of samples with 1.0 TCID50/mL influenza B virus
It can be seen from Table 4 and Figures 7-9 that, the detection rate of
samples with 2.0 TCID50/mL and 1.5 TCID50/mL virus is 100%, while the
detection rate of samples with 1.0 TCID50/mL is 87.5%. In order to ensure the
accuracy of detection, the sensitivity of this kit to detect clinical samples is
determined to be 2.0 TCID50/mL, which is consistent with the test results of
the analytical sensitivity of the registration application file CH3.5.05.
Figure 11 Amplification results of the clinically strong positive throat swab sample for
RSV
4.1.3.2 Serial dilution, detection sensitivity determination
Firstly, the clinically strong positive throat swab sample for RSV
(3.03E+07 copies/mL) was diluted with Hanks solution to 2.0E+06 copies/mL,
2.0E+05 copies/mL, 2.0E+04 copies/mL, 2.0E+03 copies/mL, 200 copies/mL, 100
copies/mL, and then was detected with this kit. Detection results are shown in
the following Figure 12.
It can be seen from the above graph that the sample with 2000 copies/mL
virus was detected as positive, while 200 copies/mL was detected as negative.
Because of the difference in concentration, it is necessary to make gradient
dilution by small intervals for the sample with 200 copies/mL virus, to
determine the measurable concentration between 200 copies/mL and 2000
copies/mL.
The above samples were diluted to 600 copies/mL, 500 copies/mL and 400
copies/mL and were repeatedly detected 8 times. The test results are shown in
the following Table 6 and Figures 13-15.
Table 6. Sensitivity detection
Detected positive
Diluted samples Positive rate(%)
samples/Sample size
600 copies/mL 8/8 100%
500 copies/mL 8/8 100%
400 copies/mL 7/8 87.5%
4.1.4 Adenovirus
4.1.4.1 Choose one clinically strong positive throat swab sample for
Adenovirus
One clinically strong positive throat swab sample for Adenovirus, whose
concentration was 3.77E+07 copies/mL determined by the Reed-Muench
method, was detected by the qualified kit in our company. The results are
Figure 16. Amplification results of the clinically strong positive throat swab sample
for Adenovirus
It can be seen from the above graph that the sample with 2000 copies/mL
virus was detected as positive, while 200 copies/mL was detected as negative.
Because of the difference in concentration, it is necessary to make gradient
dilution by small intervals for the sample with 200 copies/mL virus, to
determine the measurable concentration between 200 copies/mL and 2000
copies/mL.
The above samples were diluted to 600 copies/mL, 500 copies/mL and 400
copies/mL and were repeatedly detected 8 times. The test results are shown in
the following Table 8 and Figures 18-20.
Table 8. Sensitivity detection
Detected positive
Diluted samples Positive rate(%)
samples/Sample size
600 copies/mL 8/8 100%
500 copies/mL 8/8 100%
400 copies/mL 7/8 87.5%
Figure 21 Amplification results of the clinically strong positive throat swab sample for
Human Rhinovirus
It can be seen from the above graph that the sample with 2000 copies/mL
virus was detected as positive, while 200 copies/mL was detected as negative.
Because of the difference in concentration, it is necessary to make gradient
dilution by small intervals for the sample with 200 copies/mL virus, to
determine the measurable concentration between 200 copies/mL and 2000
copies/mL.
The above samples were diluted to 600 copies/mL, 500 copies/mL and 400
copies/mL and were repeatedly detected 8 times. The test results are shown in
the following Table 10 and Figures 23-25.
Table 10. Sensitivity detection
Detected positive
Diluted samples Positive rate(%)
samples/Sample size
600 copies/mL 8/8 100%
500 copies/mL 8/8 100%
400 copies/mL 7/8 87.5%
Figure 23. Amplification results of samples with 600 copies/mL Human rhinovirus
Figure 24. Amplification results of samples with 500 copies/mL Human rhinovirus
Figure 25. Amplification results of samples with 400 copies/mL Human rhinovirus
It can be seen from Table 10 and Figures 23-25 that, the detection rate of
samples with 600 copies/mL and 500 copies/mL virus is 100%, while the
detection rate of samples with 400 copies/mL is 87.5%. In order to ensure the
accuracy of detection, the sensitivity of this kit to detect clinical samples is
determined to be 500 copies/mL, which is consistent with the test results of
the analytical sensitivity of the registration application file CH3.5.05.
Figure 26 Amplification results of the clinically strong positive throat swab sample for
Mycoplasma pneumoniae
It can be seen from the above graph that the sample with 2000 copies/mL
Mycoplasma pneumoniae was detected as positive, while 200 copies/mL was
detected as negative. Because of the difference in concentration, it is necessary
to make gradient dilution by small intervals for the sample with 200
copies/mL Mycoplasma pneumoniae, to determine the measurable
concentration between 200 copies/mL and 2000 copies/mL.
The above samples were diluted to 600 copies/mL, 500 copies/mL and 400
copies/mL and were repeatedly detected 8 times. The test results are shown in
the following Table 12 and Figures 28-30.
Table 12. Sensitivity detection
Detected positive
Diluted samples Positive rate (%)
samples/Sample size
600 copies/mL 8/8 100%
500 copies/mL 8/8 100%
400 copies/mL 7/8 87.5%
It can be seen from Table 12 and Figures 28-30 that, the detection rate of
samples with 600 copies/mL and 500 copies/mL Mycoplasma pneumoniae is
100%, while the detection rate of samples with 400 copies/mL is 87.5%. In
order to ensure the accuracy of detection, the sensitivity of this kit to detect
clinical samples is determined to be 500 copies/mL, which is consistent with
the test results of the analytical sensitivity of the registration application file
CH3.5.05.
was 38.16.
According to the detection sensitivity of clinical samples, the above
mentioned clinical sample for RSV was diluted with Hanks solution to 500
copies/mL and to 48 parts (positive samples with detection limit
concentration), and the detection results are shown in Table 15:
Table 15. Ct value of targeted gene detection of 48 positive samples for RSV at
the detection limit concentration
Sam CY5/ Sampl CY5/C Samp CY5/ Sam CY5/ Sam CY5/
ple Ct e t le Ct ple Ct ple Ct
1 36.36 11 35.21 21 36.48 31 36.43 41 37.01
2 35.34 12 35.79 22 34.97 32 35.68 42 35.41
3 35.52 13 37.66 23 35.68 33 35.02 43 39.46
4 35.83 14 35.60 24 36.10 34 34.87 44 34.58
5 34.95 15 34.49 25 35.62 35 36.69 45 36.24
6 38.44 16 35.50 26 34.77 36 34.62 46 36.24
7 38.71 17 35.07 27 35.51 37 36.53 47 36.99
8 35.21 18 36.57 28 35.28 38 35.54 48 36.20
9 35.56 19 35.76 29 37.54 39 38.37 / /
10 34.76 20 35.10 30 36.16 40 35.20 / /
Results showed that the Ct values of 48 positive samples for RSV at the
detection limit concentration were subject to normal distribution, and were all
less than 40. The upper limit (X+2.33 S) of 99% reference range was 38.61.
According to the detection sensitivity of clinical samples, the above
mentioned clinical sample for Adenovirus was diluted with Hanks solution to
500 copies/mL and to 48 parts (positive samples with detection limit
concentration), and the detection results are shown in Table 16:
Table 16. Ct value of targeted gene detection of 48 positive samples for
Adenovirus at the detection limit concentration
Sam FAM Sampl FAM/ Samp FAM/ Sam FAM Sam FAM
ple /Ct e Ct le Ct ple /Ct ple /Ct
1 35.92 11 37.14 21 36.01 31 37.52 41 36.85
2 36.72 12 36.53 22 36.85 32 37.01 42 36.93
3 36.21 13 38.31 23 36.10 33 36.99 43 36.38
4 36.33 14 35.80 24 36.48 34 36.83 44 36.23
5 36.38 15 36.21 25 37.05 35 36.60 45 36.76
6 36.09 16 36.01 26 36.89 36 36.59 46 36.62
7 35.66 17 37.37 27 36.91 37 36.90 47 37.06
8 36.29 18 36.45 28 37.38 38 36.73 48 36.53
9 37.20 19 38.60 29 37.37 39 36.72 / /
10 36.13 20 36.66 30 37.28 40 36.72 / /
Results showed that the Ct values of 48 positive samples for Adenovirus
at the detection limit concentration of Adenovirus were subject to normal
distribution, and were all less than 40. The upper limit (X+2.33 S) of 99%
. e.
Undete Undet
15 33 Undete. 51 Undete. 69 Undete. 87
. e.
Undete Undet
16 34 Undete. 52 Undete. 70 Undete. 88
. e.
Undete
17 35 Undete. 53 Undete. 71 Undete. / /
.
Undete HRV:
18 36 54 Undete. 72 Undete. / /
. 44.92
The six respiratory pathogens nucleic acid diagnostic kit that has passed
the quality inspection was used and the strict operation was conducted in
accordance with the product instructions, so as to detect 92 confirmed clinical
positive samples for influenza B virus. Detection results are shown in Table 22
below.
Table 22. Detection results of 92 clinical positive samples for influenza B
Sam HEX/C Sam HEX/C Sam HEX/C Sam HEX/C Sam HEX/
ple t ple t ple t ple t ple Ct
1 21.57 20 29.25 39 32.41 58 26.37 77 21.75
2 28.36 21 21.95 40 30.42 59 28.54 78 21.94
3 33.58 22 31.41 41 22.04 60 24.71 79 23.01
4 30.28 23 28.35 42 25.21 61 30.20 80 26.56
5 28.55 24 28.23 43 24.15 62 22.36 81 26.65
6 31.71 25 33.77 44 30.51 63 26.31 82 24.96
7 30.20 26 30.18 45 22.97 64 28.60 83 26.37
8 24.47 27 22.18 46 28.42 65 33.03 84 29.18
9 22.36 28 23.54 47 22.35 66 25.93 85 34.34
10 28.08 29 34.77 48 23.92 67 26.63 86 28.51
11 28.24 30 24.86 49 28.64 68 25.61 87 22.18
12 33.52 31 28.22 50 23.98 69 32.76 88 22.13
13 33.87 32 29.28 51 26.02 70 32.91 89 28.85
14 34.66 33 28.77 52 22.17 71 31.18 90 29.03
15 22.39 34 24.23 53 24.60 72 32.43 91 34.26
16 33.66 35 23.62 54 28.32 73 34.14 92 29.23
17 34.30 36 31.36 55 28.63 74 28.39 / /
18 33.73 37 30.28 56 22.93 75 22.20 / /
19 30.12 38 21.95 57 24.25 76 31.13 / /
The six respiratory pathogens nucleic acid diagnostic kit that has passed
the quality inspection was used and the strict operation was conducted in
accordance with the product instructions, so as to detect 91 confirmed clinical
positive samples for RSV. Detection results are shown in Table 23 below.
Table 23. Detection results of 91 clinical positive samples for RSV
Sam CY5/C Sam CY5/C Sam CY5/C Sam CY5/C Sam CY5/
ple t ple t ple t ple t ple Ct
1 29.10 20 33.68 39 28.92 58 23.19 77 25.29
2 28.31 21 32.73 40 22.99 59 32.07 78 24.52
3 30.78 22 24.10 41 23.74 60 32.87 79 30.40
4 28.47 23 31.12 42 22.91 61 28.79 80 23.13
5 25.37 24 28.52 43 24.22 62 34.74 81 31.56
6 34.31 25 28.85 44 33.88 63 30.51 82 28.38
7 23.48 26 22.88 45 32.63 64 24.96 83 26.55
8 26.24 27 27.20 46 33.06 65 28.46 84 32.77
9 32.08 28 22.97 47 30.37 66 23.80 85 24.28
The six respiratory pathogens nucleic acid diagnostic kit that has passed
the quality inspection was used and the strict operation was conducted in
accordance with the product instructions, so as to detect 92 confirmed clinical
positive samples for Adenovirus. Detection results are shown in Table 24
below.
Table 24. Detection results of 92 clinical positive samples for Adenovirus
Sam FAM/ Sam FAM/ Sam FAM/ Sam FAM/ Sam FAM/
ple Ct ple Ct ple Ct ple Ct ple Ct
1 36.16 20 30.31 39 27.84 58 34.81 77 37.16
2 23.41 21 30.40 40 36.30 59 30.36 78 37.38
3 25.09 22 32.32 41 37.18 60 34.34 79 35.01
4 29.85 23 36.78 42 28.11 61 38.07 80 32.47
5 28.68 24 25.89 43 32.99 62 30.19 81 32.40
6 29.48 25 32.51 44 35.51 63 34.97 82 32.81
7 29.87 26 36.51 45 36.62 64 25.99 83 37.40
8 29.68 27 25.87 46 23.83 65 36.58 84 35.10
9 27.98 28 35.28 47 37.97 66 34.85 85 34.43
10 33.90 29 28.51 48 23.50 67 32.78 86 30.19
11 31.98 30 37.39 49 30.06 68 32.31 87 36.73
12 34.91 31 36.80 50 32.38 69 38.20 88 34.74
13 29.74 32 32.69 51 36.26 70 34.43 89 38.34
14 37.36 33 32.75 52 36.32 71 36.90 90 34.57
15 32.29 34 34.21 53 38.57 72 35.32 91 30.24
16 31.90 35 38.58 54 35.29 73 36.85 92 25.67
17 34.35 36 34.25 55 37.44 74 36.97 / /
18 32.21 37 34.40 56 34.88 75 37.37 / /
19 32.37 38 30.14 57 27.89 76 27.88 / /
The six respiratory pathogens nucleic acid diagnostic kit that has passed
the quality inspection was used and the strict operation was conducted in
accordance with the product instructions, so as to detect 92 confirmed clinical
positive samples for Human rhinovirus. Detection results are shown in Table
25 below.
Table 25. Detection results of 92 clinical positive samples for Human
rhinovirus
Sam HEX/ Sam HEX/ Sam HEX/ Sam HEX/ Sam HEX/
ple Ct ple Ct ple Ct ple Ct ple Ct
1 33.57 20 24.04 39 31.28 58 30.67 77 30.32
2 28.52 21 28.21 40 28.00 59 29.08 78 28.54
3 23.91 22 26.65 41 30.83 60 32.15 79 32.35
4 30.15 23 26.70 42 32.50 61 26.25 80 29.16
5 26.77 24 30.35 43 28.15 62 28.31 81 31.84
6 28.15 25 21.72 44 33.20 63 32.55 82 22.17
7 27.52 26 33.48 45 28.73 64 31.68 83 28.10
8 24.15 27 31.55 46 27.17 65 31.51 84 28.66
9 29.76 28 22.16 47 29.66 66 29.04 85 30.30
10 29.12 29 33.89 48 30.17 67 30.51 86 30.99
11 30.28 30 24.07 49 31.32 68 29.79 87 32.95
12 26.47 31 21.42 50 29.22 69 26.20 88 25.61
13 25.76 32 28.70 51 30.23 70 29.09 89 28.56
14 26.41 33 23.63 52 34.85 71 23.82 90 29.56
15 33.23 34 28.58 53 26.11 72 30.58 91 30.83
16 30.38 35 32.13 54 30.39 73 26.25 92 30.63
17 31.03 36 22.21 55 25.91 74 30.47 / /
18 30.31 37 30.89 56 28.34 75 29.59 / /
19 28.69 38 24.40 57 26.13 76 26.34 / /
The six respiratory pathogens nucleic acid diagnostic kit that has passed
the quality inspection was used and the strict operation was conducted in
accordance with the product instructions, so as to detect 92 confirmed clinical
positive samples for Mycoplasma pneumoniae. Detection results are shown in
Table 26 below.
Table 26. Detection results of 92 clinical positive samples for Mycoplasma
pneumoniae
Sam CY5/C Sam CY5/C Sam CY5/C Sam CY5/C Sam CY5/
ple t ple t ple t ple t ple Ct
1 23.12 20 31.27 39 33.40 58 33.34 77 29.83
2 24.99 21 29.44 40 27.67 59 34.19 78 25.65
3 29.89 22 35.01 41 27.21 60 33.14 79 31.45
4 25.38 23 29.78 42 35.39 61 31.38 80 31.55
5 31.86 24 25.09 43 29.05 62 31.11 81 29.70
6 33.53 25 29.36 44 35.64 63 31.51 82 31.74
7 33.82 26 23.13 45 31.16 64 34.56 83 32.11
Result 1-2 Related indicators such as the area under the ROC curve of influenza A
virus
The coordinates of the curve
The test of results: Ct value
If larger than or equal to
Sensitivity 1-Specificity
the value, then be positivea
22.9500 1.000 1.000
Result 2-2 Related indicators such as the area under the ROC curve of influenza B
virus
Result 3-1 Basic information of ROC curve analysis on Respiratory syncytial virus
Result 3-2 Related indicators such as the area under the ROC curve of respiratory
syncytial virus
The coordinates of the curve
The test of results: Ct value
If larger than or equal to
Sensitivity 1-Specificity
the value, then be positivea
21.8800 1.000 1.000
22.8950 1.000 .989
22.9400 1.000 .978
22.9800 1.000 .967
23.0600 1.000 .956
23.1400 1.000 .945
23.1700 1.000 .934
23.2250 1.000 .923
23.2900 1.000 .912
23.3950 1.000 .901
23.4750 1.000 .890
23.5000 1.000 .868
23.6300 1.000 .857
23.7700 1.000 .846
23.9500 1.000 .835
24.1600 1.000 .824
24.2500 1.000 .813
24.3300 1.000 .802
24.4500 1.000 .791
24.6600 1.000 .780
24.8800 1.000 .769
25.0150 1.000 .758
25.1800 1.000 .747
Result 4-2 Related indicators such as the area under the ROC curve of Adenovirus
The coordinates of the curve
The test of results: Ct value
If larger than or equal to the
Sensitivity 1-Specificity
value, then be positivea
22.4100 1.000 1.000
23.4550 1.000 .989
23.6650 1.000 .978
24.4600 1.000 .967
25.3800 1.000 .957
25.7700 1.000 .946
25.8800 1.000 .935
25.9400 1.000 .924
26.9150 1.000 .913
27.8600 1.000 .902
27.8850 1.000 .891
27.9350 1.000 .880
28.0450 1.000 .870
28.3100 1.000 .859
28.5950 1.000 .848
29.0800 1.000 .837
29.5800 1.000 .826
29.7100 1.000 .815
29.7950 1.000 .804
29.8600 1.000 .793
29.9650 1.000 .783
30.1000 1.000 .772
30.1650 1.000 .761
30.2150 1.000 .739
30.2750 1.000 .728
30.3350 1.000 .717
30.3800 1.000 .707
31.1500 1.000 .696
31.9400 1.000 .685
32.0950 1.000 .674
32.2500 1.000 .663
32.3000 1.000 .652
32.3150 1.000 .641
32.3450 1.000 .630
Result 5-2 Related indicators such as the area under the ROC curve of Human
rhinovirus
Result 6-2 Related indicators such as the area under the ROC curve of Mycoplasma
pneumoniae
Sample No.
Sample No.
Sample No.
No
12 31.26 31.93 34 15.81 29.09 56 26.20 31.29 78 25.47
Ct
13 32.01 27.49 35 29.73 32.91 57 22.00 30.30 79 27.30 28.85
14 22.08 32.35 36 22.67 32.97 58 27.08 28.17 80 35.46 31.48
15 22.29 21.36 37 19.95 32.48 59 24.01 25.18 81 33.65 31.61
No
16 31.24 19.19 38 27.42 32.83 60 20.98 28.83 82 32.20
Ct
17 26.47 15.94 39 25.31 34.83 61 20.96 22.11 83 30.62 23.63
18 28.57 30.63 40 20.11 35.56 62 35.67 33.66 84 28.05 23.37
19 31.26 18.15 41 35.77 36.91 63 33.18 33.38 85 25.53 20.13
20 29.77 29.87 42 35.78 37.70 64 32.07 32.05 86 26.60 34.12
21 30.43 29.78 43 33.51 32.56 65 31.13 33.33 87 33.48 32.23
22 28.81 26.43 44 23.16 23.48 66 30.30 33.86 88 21.72 22.07
Accuracy study 1
Final Final
Sample Sample
Inhibitor name concentrati Inhibitor name concentrati
No. No.
on on
Oxymetazoline
F1 100 μg/mL F9 SDS 100 μg/mL
hydrochloride
F2 Dexamethasone 50 μg/mL F10 EDTA 10 μg/mL
Cefmenoxime
F3 50 μg/mL F11 Urea 100 μg/mL
Hydrochloride
F4 Menthol 50 μg/mL F12 Heme 10 μg/mL
Purified
F5 Zanamivir 100 μg/mL F13 20 μg/mL
Mucin
F6 Ribavirin 100 μg/mL F14 FeCl3 100 μg/mL
Absolute
F7 Azithromycin 100 μg/mL F15 20%(v/v)
ethanol
Human whole
F8 NaCl 60 μg/mL F16 20%(v/v)
blood
Figure 1-1 The effect test curve of influenza A virus anti-inhibitor/interfering substance
(2017001 batch)
Figure 1-2 The effect test curve of influenza B virus anti-inhibitor/interfering substance
(2017001 batch)
Accuracy study 2
Figure 1-3 The effect test curve of respiratory syncytial virus anti-inhibitor/interfering
substance (2017001 batch)
Figure 1-5 The effect test curve of Human rhinovirus anti-inhibitor/interfering substance
(2017001 batch)
Accuracy study 3
Figure 1-6 The effect test curve of Mycoplasma pneumoniae anti-inhibitor/interfering
substance (2017001 batch)
Accuracy study 4
30.9 31.3 31.2 30.8 31.7 31.1 31.2 31.9 31.1 31.0 31.4 31.9
F14
8 4 0 9 7 7 5 0 0 6 9 0
30.9 31.3 31.2 31.0 31.4 31.0 31.3 32.2 31.3 31.2 31.6 32.0
F15
8 1 0 5 7 6 4 6 6 3 0 4
30.9 30.9 30.6 30.6 30.7 31.4 31.2 31.3 30.9 30.5 31.6 31.5
F16
3 1 7 7 9 4 3 2 3 5 7 2
Accuracy study 5
PCR mix A PCR mix B
FAM Ct HEX Ct CY5 Ct FAM Ct HEX Ct CY5 Ct
value value value value value value
Con 30.7 30.7 31.0 30.9 31.5 30.9 30.5 30.9 31.2 31.0 31.3 31.1
trol 9 9 7 5 9 0 8 3 7 6 3 6
307 31.0 31.3 31.1 31.5 31.4 30.7 31.1 31.2 31.1 31.2 31.2
F1
7 9 6 7 0 3 6 8 7 8 1 4
30.8 31.0 31.2 31.1 31.3 30.7 31.3 31.3 31.4 31.5 31.3 31.4
F2
0 7 0 2 9 8 4 4 2 8 1 0
31.0 31.1 31.0 31.4 31.5 31.6 31.8 31.7 31.2 31.4 31.3 31.3
F3
5 9 0 3 0 1 4 5 9 3 9 4
30.9 31.0 31.6 31.4 31.5 31.3 31.4 31.5 31.6 31.2 31.6 31.1
F4
0 8 3 0 1 8 6 9 2 3 0 7
30.9 31.2 31.5 31.5 31.8 31.5 31.4 31.8 31.6 31.4 31.5 31.3
F5
8 4 1 6 3 9 2 4 2 6 8 8
31.1 31.3 31.4 31.3 31.1 31.6 31.5 31.2 31.4 31.4 31.2 31.2
F6
4 6 3 6 1 8 0 5 6 8 7 5
31.3 31.2 31.3 31.3 31.4 31.4 31.4 31.1 31.3 31.2 31.1 31.4
F7
0 1 4 7 1 8 5 9 3 4 3 2
30.9 30.9 30.9 31.0 31.4 31.2 31.3 31.0 31.2 31.0 31.2 31.4
F8
7 1 9 2 9 6 5 9 3 0 7 8
31.2 31.3 31.0 31.1 31.2 31.5 31.2 31.3 31.3 31.0 31.2 31.3
F9
3 1 1 9 4 8 8 5 3 0 3 0
31.1 31.1 31.5 31.7 31.4 31.3 31.3 31.3 30.6 31.1 31.3 31.2
F10
4 6 1 1 8 9 7 3 1 6 1 0
31.0 31.1 31.2 31.5 31.3 31.3 31.5 31.3 31.2 31.2 31.3 31.3
F11
6 3 8 0 8 8 5 9 4 4 2 4
31.2 31.2 31.4 31.4 31.5 31.6 31.5 31.4 31.3 31.5 31.3 31.5
F12
0 4 4 5 9 5 7 8 0 1 1 7
31.1 31.3 31.9 31.8 31.8 31.5 31.4 31.4 31.3 31.4 31.3 31.7
F13
9 8 0 3 1 3 6 0 6 9 6 0
31.2 31.3 31.5 31.3 31.5 31.1 31.5 31.3 31.2 31.2 31.2 31.6
F14
3 2 9 6 5 6 8 5 2 6 9 1
31.1 31.2 31.8 31.6 31.5 31.6 31.4 31.2 30.0 31.2 31.3 30.7
F15
9 1 1 7 3 5 0 0 7 8 1 2
30.8 31.0 30.6 30.7 30.5 31.5 31.3 31.4 30.9 30.9 31.1 30.9
F16
4 1 3 9 2 0 5 7 2 1 2 8
Accuracy study 6
instrument were not significantly different from the control samples that do
not contain interfering substances. It shows that potential PCR
inhibitors/interfering substances have no interference effect on the PCR
reaction, and have no effect on the detection of respiratory pathogens in this
kit.
1.2 PCR inhibitor removal ability test
The positive throat swab samples with low concentration of influenza A
virus, influenza B virus, respiratory syncytial virus, adenovirus, Mycoplasma
pneumoniae, and human rhinovirus were used as the test samples. The test
sample was divided into several parts, and each part was added with the
possible PCR inhibitors/interfering substances listed in Table 1-1, and the
detection was carried out according to the operating method in the manual of
this kit. The magnetic bead method was used for nucleic acid extraction and
the test sample without interfering substances was used as the control. The
test results are shown in Table 1-5 to Table 1-7 and Figure 1-7 to Figure 1-12
(only the batch of 2017001 is taken as an example for screenshot display).
Accuracy study 7
Figure 1-9 The test curve of respiratory syncytial virus anti-inhibitor/interfering
substance removal effect (2017001 batch)
Accuracy study 8
Figure 1-12 The test curve of Mycoplasma pneumoniae anti-inhibitor/interfering
substance removal effect (2017001 batch)
Table 1-5 Statistical results of Ct value of inhibitor/interfering substance removal
effect test(2017001 batch)
Influenza A Influenza B
RSV Adenovirus MP HRV
virus virus
Test Test Test Test Test Test Test Test Test Test Test Test
1 2 1 2 1 2 1 2 1 2 1 2
Contr
28.04 27.42 30.71 28.95 28.05 27.64 31.88 30.57 31.12 32.09 31.39 30.58
ol
F1 27.02 27.34 30.29 28.98 28.10 27.90 30.63 30.82 31.85 32.08 31.18 30.84
F2 27.27 27.37 29.30 29.11 28.20 28.04 31.05 31.02 31.93 31.71 31.39 30.94
F3 27.28 27.06 29.96 29.38 28.22 27.85 31.32 31.12 31.91 31.71 31.22 31.09
F4 27.44 27.35 29.96 29.19 28.57 27.82 31.45 31.24 31.72 32.58 31.19 31.10
F5 27.32 27.31 30.10 30.02 28.79 28.19 31.01 31.22 31.89 32.11 30.94 31.04
F6 27.38 27.18 30.26 29.29 26.58 27.81 31.24 31.30 31.90 32.09 30.83 31.01
F7 27.81 26.94 30.04 29.37 27.10 28.25 30.91 31.17 31.86 31.73 31.20 31.11
F8 26.78 27.66 31.27 28.47 26.34 26.59 30.60 30.53 31.90 32.40 30.24 30.38
F9 26.71 26.89 28.82 29.22 26.76 26.76 30.76 30.93 31.98 32.06 30.76 30.78
F10 27.54 27.52 29.09 29.26 27.49 26.96 30.94 31.10 32.19 32.36 30.84 30.75
F11 27.33 26.64 28.88 29.21 28.12 27.31 31.27 31.45 32.04 31.45 31.06 30.91
F12 26.34 26.64 29.21 28.16 28.11 27.06 31.31 31.41 32.23 32.28 31.07 31.16
F13 26.53 26.70 28.53 29.08 27.83 27.65 30.86 31.32 32.00 32.21 30.86 31.04
F14 25.83 26.98 28.29 28.22 28.24 26.58 31.30 31.24 32.83 32.16 31.30 31.01
F15 26.96 27.07 29.64 28.97 28.32 26.33 30.93 31.22 32.10 31.81 30.65 30.76
F16 26.82 28.50 29.68 29.67 27.37 26.27 30.42 30.66 32.19 32.38 30.90 30.73
Influenza A Influenza B
RSV Adenovirus MP HRV
virus virus
Test Test Test Test Test Test Test Test Test Test Test Test
1 2 1 2 1 2 1 2 1 2 1 2
Contr
26.46 26.53 29.70 29.38 27.14 26.89 29.73 29.92 31.50 32.07 30.81 30.11
ol
Accuracy study 9
F1 26.66 27.13 29.61 29.65 27.20 27.45 30.09 30.02 31.53 31.67 30.18 30.15
F2 27.19 27.65 29.67 29.91 27.39 27.33 30.26 30.38 31.80 31.96 30.54 30.53
F3 27.41 27.54 29.31 29.29 27.57 27.60 30.54 30.41 31.57 31.13 30.44 30.33
F4 27.54 27.67 29.59 29.25 27.65 27.72 30.16 31.02 31.85 31.91 30.62 30.30
F5 27.44 27.63 29.51 29.61 27.93 27.65 30.27 30.27 31.57 31.75 29.85 30.63
F6 27.31 27.15 29.16 29.51 27.67 27.50 30.29 30.25 31.37 31.70 30.37 30.23
F7 27.32 27.53 28.98 29.43 27.49 27.65 30.04 30.30 31.60 31.38 30.31 30.69
F8 27.34 27.40 29.44 29.07 27.20 27.03 29.88 30.19 31.79 31.94 30.08 30.11
F9 27.26 27.48 29.73 29.47 27.53 27.23 30.19 30.24 32.35 31.43 30.00 30.37
F10 27.41 27.62 29.61 29.39 27.31 27.30 30.39 30.27 31.35 31.22 30.38 30.57
F11 27.35 27.40 29.30 28.75 27.67 27.34 30.54 30.56 31.66 32.08 30.30 30.56
F12 27.59 27.42 29.40 29.63 27.43 27.59 30.53 30.80 32.12 31.07 30.30 30.64
F13 27.72 27.56 29.71 29.57 27.75 27.83 30.31 30.44 32.04 31.83 30.35 30.36
F14 27.43 27.43 29.48 29.20 27.77 27.33 30.29 30.31 31.73 31.90 30.45 30.66
F15 27.16 27.37 29.67 29.37 27.43 27.46 29.90 30.24 31.35 31.54 30.25 29.97
F16 27.50 27.45 29.46 29.21 26.77 26.98 30.07 30.14 31.82 31.50 30.07 30.01
Accuracy study 10
2. Measurement accuracy
2.1 Test of reference samples
The reference samples P1-P12 required in the technical requirements of
the six respiratory pathogens nucleic acid diagnostic kit (PCR-fluorescence
probing) were used as the test samples. Reference samples P1-P12 consist of
12 positive samples for the six inactivated respiratory pathogens with
different concentrations. Among them, P1 is influenza A/H1N1 (2009) virus,
P2 is influenza A H3 subtype virus, P3 is influenza A H5 subtype virus, P4 is
influenza A H7N9 subtype virus, and P5 is influenza B virus Victoria subtype,
P6 is influenza B virus Yamagata subtype, P7 is respiratory syncytial virus, P8
is adenovirus 3 subtype, P9 is adenovirus 4 subtype, P10 is adenovirus 7
subtype, P11 is Mycoplasma pneumoniae, and P12 is human Rhinovirus. Each
reference sample is diluted with Hanks virus preservation solution (pH
7.2-7.6) or normal saline. 3 batches of qualified kits were used to detect the
samples on the same ABI 7500 instrument, and each sample were repeatedly
tested 3 times to evaluate the measurement accuracy of this kit.
Test results are as follows:
Accuracy study 11
Figure1-15 Amplification curve of reference samples P1-P12 by using PCR mix A
(2017003 batch)
Accuracy study 12
Figure1-18 Amplification curve of reference samples P1-P12 by using PCR mix B
(2017003 batch)
Table 1-8 Test results of reference samples P1-P12(2017001 batch)
Test results PCR mix A PCR mix B
Reference STDEV
sample Test1 Test2 Test3 Test1 Test2 Test3
P1 24.52 23.37 23.73 No Ct No Ct No Ct 0.59
P2 27.16 27.21 27.06 No Ct No Ct No Ct 0.08
P3 19.80 19.83 19.88 No Ct No Ct No Ct 0.04
P4 29.66 29.68 29.67 No Ct No Ct No Ct 0.01
P5 30.32 29.96 29.80 No Ct No Ct No Ct 0.27
P6 34.02 33.72 34.06 No Ct No Ct No Ct 0.19
P7 28.23 28.36 28.31 No Ct No Ct No Ct 0.07
P8 No Ct No Ct No Ct 21.41 21.24 21.79 0.28
P9 No Ct No Ct No Ct 25.48 25.74 26.02 0.27
P10 No Ct No Ct No Ct 28.35 28.33 28.26 0.05
P11 No Ct No Ct No Ct 24.57 24.69 24.60 0.06
P12 No Ct No Ct No Ct 32.30 32.50 32.30 0.12
Table 1-9 Test results of reference samples P1-P12(2017002 batch)
Test results PCR mix A PCR mix B
Reference STDEV
sample Test1 Test2 Test3 Test1 Test2 Test3
P1 24.48 23.63 23.62 No Ct No Ct No Ct 0.49
P2 27.15 27.21 27.07 No Ct No Ct No Ct 0.07
P3 19.82 19.62 19.49 No Ct No Ct No Ct 0.17
P4 29.87 29.57 29.59 No Ct No Ct No Ct 0.17
P5 29.90 30.28 30.13 No Ct No Ct No Ct 0.19
P6 34.15 34.15 34.36 No Ct No Ct No Ct 0.12
P7 28.27 28.19 28.39 No Ct No Ct No Ct 0.10
P8 No Ct No Ct No Ct 21.24 21.14 21.50 0.19
P9 No Ct No Ct No Ct 25.57 25.77 25.76 0.11
P10 No Ct No Ct No Ct 28.32 28.36 28.27 0.05
P11 No Ct No Ct No Ct 24.43 24.52 24.27 0.13
P12 No Ct No Ct No Ct 32.16 32.15 32.14 0.01
Accuracy study 13
Table 1-10 Test results of reference samples P1-P12(2017003 batch)
Test results PCR mix A PCR mix B
Reference STDEV
sample Test1 Test2 Test3 Test1 Test2 Test3
P1 24.52 23.34 23.64 No Ct No Ct No Ct 0.61
P2 27.19 27.32 27.22 No Ct No Ct No Ct 0.07
P3 19.96 19.51 19.54 No Ct No Ct No Ct 0.25
P4 29.65 29.76 29.81 No Ct No Ct No Ct 0.08
P5 30.01 30.00 30.16 No Ct No Ct No Ct 0.09
P6 33.84 33.97 33.80 No Ct No Ct No Ct 0.09
P7 28.15 28.24 28.20 No Ct No Ct No Ct 0.05
P8 No Ct No Ct No Ct 21.16 21.19 20.78 0.23
P9 No Ct No Ct No Ct 25.60 25.71 25.83 0.12
P10 No Ct No Ct No Ct 28.53 28.43 28.32 0.11
P11 No Ct No Ct No Ct 24.60 24.56 24.14 0.25
P12 No Ct No Ct No Ct 31.70 32.03 31.83 0.17
Conclusion: It can be seen from the above that, the test results of reference
samples P1-P12 by using the 3 batches of reagents (2017001, 2017002, 2017003)
were all positive, and there was no non-specific amplification. The standard
deviation (STDEV) of the Ct values of each sample is between 0.01 and 0.61.
Therefore, this kit has good detection accuracy.
Figure 1-19 The test results of samples for influenza A virus, influenza B virus and
respiratory syncytial virus by using PCR mix A (2017001 batch)
Accuracy study 14
Figure 1-20 The test results of samples for influenza A virus, influenza B virus and
respiratory syncytial virus by using PCR mix B (2017001 batch)
Figure 1-21 The test results of samples for influenza A virus, influenza B virus and
respiratory syncytial virus by using PCR mix A (2017002 batch)
Figure 1-22 The test results of samples for influenza A virus, influenza B virus and
respiratory syncytial virus by using PCR mix B (2017002 batch)
Accuracy study 15
Figure 1-23 The test results of samples for influenza A virus, influenza B virus and
respiratory syncytial virus by using PCR mix A (2017003 batch)
Figure 1-24 The test results of samples for influenza A virus, influenza B virus and
respiratory syncytial virus by using PCR mix B (2017003 batch)
Figure 1-25 The test results of samples for Adenovirus, Mycoplasma pneumoniae and
Human rhinovirus by using PCR mix A (2017001 batch)
Accuracy study 16
Figure 1-26 The test results of samples for Adenovirus, Mycoplasma pneumoniae and
Human rhinovirus by using PCR mix B (2017001 batch)
Figure 1-27 The test results of samples for Adenovirus, Mycoplasma pneumoniae and
Human rhinovirus by using PCR mix A (2017002 batch)
Figure 1-28 The test results of samples for Adenovirus, Mycoplasma pneumoniae and
Human rhinovirus by using PCR mix B (2017002 batch)
Accuracy study 17
Figure 1-29 The test results of samples for Adenovirus, Mycoplasma pneumoniae and
Human rhinovirus by using PCR mix A (2017003 batch)
Figure 1-30 The test results of samples for Adenovirus, Mycoplasma pneumoniae and
Human rhinovirus by using PCR mix B (2017003 batch)
Figure 1-31 The test results of negative samples by using PCR mix A (2017001 batch)
Accuracy study 18
Figure 1-32 The test results of negative samples by using PCR mix B (2017001 batch)
Figure 1-33 The test results of negative samples by using PCR mix A (2017002 batch)
Figure 1-34 The test results of negative samples by using PCR mix B (2017002 batch)
Accuracy study 19
Figure 1-35 The test results of negative samples by using PCR mix A (2017003 batch)
Figure 1-36 The test results of negative samples by using PCR mix B (2017003 batch)
Table 1-11 Test results of samples for influenza A virus
Ct value of samples
Batch Reaction mix
1 2 3 4 5 6 7 8
PCR mix A 26.77 20.97 18.84 29.08 32.12 23.92 25.64 22.78
2017001
PCR mix B No Ct No Ct No Ct No Ct No Ct No Ct No Ct No Ct
PCR mix A 26.75 21.06 19.41 30.17 33.47 25.21 27.02 23.86
2017002
PCR mix B No Ct No Ct No Ct No Ct No Ct No Ct No Ct No Ct
PCR mix A 26.58 20.98 18.23 29.31 32.91 24.44 26.23 22.67
2017003
PCR mix B No Ct No Ct No Ct No Ct No Ct No Ct No Ct No Ct
Accuracy study 20
PCR mix B No Ct No Ct No Ct No Ct No Ct No Ct No Ct No Ct
Table 1-17 Test results of negative samples (No amplification curve was observed for
Accuracy study 21
FAM, HEX and CY5, only the Ct value of internal standard was collected)
Ct value of internal standard
Batch Reaction mix
1 2 3 4 5 6 7 8
PCR mix A 27.13 27.27 27.06 26.87 27.04 26.85 27.67 27.56
2017001
PCR mix B 28.02 27.49 26.83 26.85 27.16 27.20 27.24 27.87
PCR mix A 27.97 27.46 27.37 27.52 27.88 27.51 27.44 27.78
2017002
PCR mix B 27.55 27.24 27.54 27.56 28.29 28.67 28.77 28.91
PCR mix A 27.27 27.73 27.43 27.92 27.50 27.76 27.54 27.41
2017003
PCR mix B 27.46 27.67 27.43 27.57 27.67 27.52 27.57 27.67
Conclusion: It can be seen from the above that, the test results of positive
throat swab samples for influenza A virus, influenza B virus, Respiratory
syncytial virus, Adenovirus, Mycoplasma pneumoniae and Human
rhinovirus by using the 3 batches of reagents (2017001, 2017002, 2017003)
were all positive, and there was no non-specific amplification. Negative
samples were all negative, and the test results of internal standard were all
within the range of reference value. Therefore, this kit has good detection
accuracy.
Figure 1-37 Test results of PC01 and PC02 by using PCR mix A (2017001 batch)
Accuracy study 22
Figure 1-38 Test results of PC01 and PC02 by using PCR mix B (2017001 batch)
Figure 1-39 Test results of PC03—PC06 by using PCR mix A (2017001 batch)
Accuracy study 23
Figure 1-40 Test results of PC03—PC06 by using PCR mix B (2017001 batch)
Figure 1-41 Test results of PC01 and PC02 by using PCR mix A (2017002 batch)
Accuracy study 24
Figure 1-42 Test results of PC01 and PC02 by using PCR mix B (2017002 batch)
Figure 1-43 Test results of PC03—PC06 by using PCR mix A (2017002 batch)
Accuracy study 25
Figure 1-44 Test results of PC03—PC06 by using PCR mix B (2017002 batch)
Figure 1-45 Test results of PC01 and PC02 by using PCR mix A (2017003 batch)
Accuracy study 26
Figure 1-46 Test results of PC01 and PC02 by using PCR mix B (2017003 batch)
Figure 1-47 Test results of PC03—PC06 by using PCR mix A (2017003 batch)
Accuracy study 27
Figure 1-48 Test results of PC03—PC06 by using PCR mix B (2017003 batch)
Table 1-18 Test results of national reference samples PC01~PC06 (2017001 batch)
Test results PCR mix A PCR mix B
Reference
samples Test1 Test2 Test1 Test2
PC01 32.03 31.77 No Ct No Ct
PC02 31.27 30.76 No Ct No Ct
PC03 30.47 30.24 No Ct No Ct
PC04 29.39 29.34 No Ct No Ct
PC05 29.38 29.36 No Ct No Ct
PC06 29.23 29.09 No Ct No Ct
Table 1-19 Test results of national reference samples PC01~PC06 (2017002 batch)
Test results PCR mix A PCR mix B
Reference
samples Test1 Test2 Test1 Test2
PC01 30.93 30.89 No Ct No Ct
PC02 31.30 30.72 No Ct No Ct
PC03 30.13 29.65 No Ct No Ct
PC04 30.50 30.17 No Ct No Ct
PC05 29.53 29.39 No Ct No Ct
PC06 28.79 28.74 No Ct No Ct
Table 1-20 Test results of national reference samples PC01~PC06 (2017003 batch)
Test results PCR mix A PCR mix B
Reference
samples Test1 Test2 Test1 Test2
PC01 30.93 31.26 No Ct No Ct
PC02 31.33 31.17 No Ct No Ct
PC03 29.60 29.38 No Ct No Ct
PC04 30.08 29.73 No Ct No Ct
Accuracy study 28
PC05 29.29 29.02 No Ct No Ct
PC06 28.53 28.44 No Ct No Ct
Conclusion: It can be seen from the above that, when testing the positive
reference samples PC01~PC06 in the national reference samples of influenza
A/B virus nucleic acid detection reagent by using the 3 batches of reagents
(2017001, 2017002, 2017003), PC01 and PC02 were all positive with influenza B
virus, PC03—PC06 were all positive with influenza A virus, and there was no
non-specific amplification. The results met the requirements of national
reference sample test for influenza A/B virus nucleic acid detection reagents.
Therefore, this kit has good detection accuracy.
Accuracy study 29
mixed with equal volumes to form the test sample. The test sample was
divided into several parts, and the possible PCR inhibitors/interfering
substances listed in Table 2-1 were added to each part, and the detection was
carried out according to the operating method in the manual of this kit. The
test sample without interfering substances was used as the control. The test
results of three batches of the kit on the SLAN-96P PCR instrument are shown
in Table 2-2 to Table 2-4 and Figure 2-1 to Figure 2-6 (only the batch of 2017001
is taken as an example for screenshot display).
Table 2-1 PCR inhibitor/interfering substance added and its concentration
Figure 2-1 The test curve of influenza A virus anti-inhibitor/interfering substance effect
(2017001 batch)
Accuracy study 30
Figure 2-2 The test curve of influenza B virus anti-inhibitor/interfering substance effect
(2017001 batch)
Figure 2-3 The effect test curve of respiratory syncytial virus anti-inhibitor/interfering
substance (2017001 batch)
Accuracy study 31
Figure 2-5 The effect test curve of Human rhinovirus anti-inhibitor/interfering substance
(2017001 batch)
Accuracy study 32
30.2 30.1 31.2 31.4 30.8 31.0 28.9 28.9 30.7 30.5 31.1 31.7
F6
9 5 0 7 2 8 1 1 5 5 3 4
30.2 30.1 31.3 31.5 31.0 30.0 29.0 28.0 30.7 30.6 31.1 31.1
F7
0 0 9 2 7 8 2 5 9 9 6 8
30.1 30.1 31.2 31.5 31.4 31.3 28.9 28.9 30.8 30.8 31.4 31.3
F8
0 9 8 1 9 0 8 1 8 1 1 4
30.0 30.3 31.6 31.3 31.2 31.5 28.4 28.6 30.6 30.7 31.3 31.4
F9
3 0 1 0 9 7 5 7 1 8 5 2
30.2 30.1 31.6 31.5 31.2 31.5 29.0 28.8 30.8 30.8 31.5 31.5
F10
7 8 6 8 9 1 0 7 1 3 4 4
30.3 30.2 31.4 31.6 31.3 31.3 28.9 28.9 30.6 30.8 31.5 31.4
F11
4 8 4 3 3 5 3 8 5 1 8 8
30.4 30.4 31.3 31.5 31.7 31.4 28.9 28.5 30.8 30.7 31.5 31.4
F12
0 3 1 5 0 9 9 4 3 5 7 6
30.3 30.4 31.6 31.6 30.8 31.0 28.8 28.7 30.7 30.7 31.4 31.5
F13
1 5 1 8 8 1 0 8 7 6 3 0
30.4 30.3 31.7 31.6 30.9 31.1 28.9 29.0 30.7 30.8 31.6 31.6
F14
9 0 8 6 4 5 6 7 0 6 4 1
30.3 30.3 31.1 31.7 30.8 31.0 28.8 28.9 30.6 30.7 30.9 31.2
F15
1 0 5 2 8 1 4 1 0 4 5 3
30.4 30.1 31.6 31.4 31.2 30.9 28.8 28.8 30.7 30.8 31.5 31.3
F16
4 7 0 6 6 1 9 6 4 7 5 3
Accuracy study 33
8 4 0 6 0 3 8 8 0 3 7 0
28.9 28.8 31.0 30.7 31.3 31.3 28.0 27.9 31.1 30.3 32.1 31.9
F10
0 7 5 8 0 9 0 9 6 2 3 1
28.7 28.9 31.0 31.0 31.0 31.0 28.1 28.0 31.1 31.1 32.1 31.9
F11
9 4 6 4 3 6 1 1 3 9 6 6
29.0 28.8 31.0 31.2 31.1 31.0 28.0 28.0 30.9 31.0 32.1 32.3
F12
3 8 9 8 7 5 6 4 2 6 9 7
28.8 28.8 31.0 31.1 30.8 31.0 28.1 28.0 31.0 31.1 31.7 31.8
F13
0 5 1 2 7 9 3 2 2 3 6 5
29.0 28.8 31.0 31.0 30.8 31.3 28.0 28.0 31.1 31.0 31.9 31.7
F14
6 0 7 5 8 7 2 2 7 2 8 5
28.9 28.8 31.1 31.2 30.7 31.6 27.9 28.0 31.0 31.1 32.1 32.1
F15
3 8 0 3 3 5 4 4 3 8 9 8
28.9 28.9 30.8 31.0 31.2 31.1 28.0 28.1 31.2 31.0 32.2 32.1
F16
3 7 9 8 4 4 0 7 0 5 6 4
Accuracy study 34
28.7 28.7 30.9 31.0 31.2 31.1 28.1 28.0 30.6 30.8 31.6 31.7
F13
9 9 6 5 0 0 4 8 1 4 3 3
28.7 28.6 31.1 31.2 30.9 31.3 28.1 28.2 30.6 31.2 31.6 31.6
F14
1 6 5 1 7 1 1 7 0 3 6 9
28.8 28.8 31.3 31.2 31.0 31.0 28.0 28.0 30.7 31.0 32.0 31.5
F15
7 3 3 3 8 9 6 9 3 8 3 5
28.6 28.6 31.1 30.9 31.3 30.8 28.0 28.8 30.5 30.4 32.1 31.7
F16
5 3 4 6 1 8 7 7 1 7 1 6
Accuracy study 35
Figure 2-8 The test curve of influenza B virus anti-inhibitor/interfering substance
removal effect (2017001 batch)
Accuracy study 36
Figure 2-11 The test curve of Mycoplasma pneumoniae anti-inhibitor/interfering
substance removal effect (2017001 batch)
Accuracy study 37
F13 28.45 28.49 31.13 30.97 30.22 30.14 30.08 29.94 33.10 32.75 31.03 31.00
F14 28.38 28.37 30.94 30.95 30.39 30.30 30.6 30.03 32.61 33.33 30.89 31.15
F15 28.31 28.50 31.12 30.99 30.16 30.35 30.12 29.84 32.57 32.64 30.86 30.94
F16 28.40 28.41 31.07 31.01 30.36 30.12 30.00 29.85 32.91 31.96 30.88 30.98
Accuracy study 38
F8 28.14 28.12 31.18 31.07 30.41 30.35 30.01 29.96 31.61 31.87 30.11 30.00
F9 28.26 29.00 30.94 31.27 30.41 30.72 29.99 30.13 31.83 31.44 29.86 30.48
F10 28.20 28.16 31.03 31.27 30.53 30.49 30.11 29.93 31.82 31.96 30.15 29.98
F11 28.19 28.14 31.38 31.17 30.37 30.46 29.96 30.06 32.52 31.93 30.14 29.63
F12 28.12 28.26 31.32 31.17 30.46 30.12 30.16 29.95 32.12 32.19 30.12 29.94
F13 28.15 28.19 31.22 31.18 30.64 30.59 29.98 29.94 32.00 32.18 30.02 30.39
F14 28.15 28.24 31.17 31.19 31.57 31.23 29.93 29.99 31.65 32.14 30.21 30.16
F15 28.17 28.19 31.30 30.92 30.81 30.50 29.57 29.71 31.94 32.66 29.93 30.13
F16 28.22 28.32 31.09 31.10 30.71 30.58 30.11 29.94 32.66 31.11 30.11 30.19
Conclusion: It can be seen from the above experimental results that, by using
the clinically positive throat swab sample as the test sample, the test results of
samples with different concentrations of various PCR inhibitors/interfering
substances have no significant differences after nucleic acid extraction via the
magnetic bead method. It shows that the magnetic bead method is beneficial
for removing PCR inhibitors/interfering substances.
2. Measurement accuracy
2.1 Test of reference samples
The reference samples P1-P12 required in the technical requirements of
the six respiratory pathogens nucleic acid diagnostic kit (PCR-fluorescence
probing) were used as the test samples. Reference samples P1-P12 consist of
12 positive samples for the six inactivated respiratory pathogens with
different concentrations. Among them, P1 is influenza A/H1N1 (2009) virus,
P2 is influenza A H3 subtype virus, P3 is influenza A H5 subtype virus, P4 is
influenza A H7N9 subtype virus, and P5 is influenza B virus Victoria subtype,
P6 is influenza B virus Yamagata subtype, P7 is respiratory syncytial virus, P8
is adenovirus 3 subtype, P9 is adenovirus 4 subtype, P10 is adenovirus 7
subtype, P11 is Mycoplasma pneumoniae, and P12 is human Rhinovirus. Each
reference sample is diluted with Hanks virus preservation solution (pH
7.2-7.6) or normal saline. 3 batches of qualified kits were used to detect the
samples on the same SLAN-96P instrument, and each sample were repeatedly
tested 3 times to evaluate the measurement accuracy of this kit.
Test results are as follows:
Figure 2-13 Amplification curve of reference samples P1-P12 by using PCR mix A
(2017001 batch)
Accuracy study 39
Figure 2-14 Amplification curve of reference samples P1-P12 by using PCR mix A
(2017002 batch)
Figure 2-15 Amplification curve of reference samples P1-P12 by using PCR mix A
(2017003 batch)
Figure 2-16 Amplification curve of reference samples P1-P12 by using PCR mix B
(2017001 batch)
Accuracy study 40
Figure 2-17 Amplification curve of reference samples P1-P12 by using PCR mix B
(2017002 batch)
Figure 2-18 Amplification curve of reference samples P1-P12 by using PCR mix B
(2017003 batch)
Table 2-8 Test results of reference samples P1-P12 (2017001 batch)
Test results PCR mix A PCR mix B
Reference STDEV
sample Test1 Test2 Test3 Test1 Test2 Test3
P1 24.69 23.69 23.03 No Ct No Ct No Ct 0.84
P2 26.50 26.75 26.73 No Ct No Ct No Ct 0.14
P3 19.38 19.31 18.93 No Ct No Ct No Ct 0.24
P4 29.34 29.21 29.37 No Ct No Ct No Ct 0.09
P5 28.36 28.26 28.22 No Ct No Ct No Ct 0.07
P6 31.92 32.61 32.39 No Ct No Ct No Ct 0.35
P7 25.73 25.77 25.42 No Ct No Ct No Ct 0.19
P8 No Ct No Ct No Ct 20.36 19.97 20.26 0.20
P9 No Ct No Ct No Ct 24.85 25.11 24.96 0.13
P10 No Ct No Ct No Ct 27.47 27.33 27.75 0.21
P11 No Ct No Ct No Ct 21.86 22.07 22.12 0.14
P12 No Ct No Ct No Ct 30.30 30.31 30.38 0.04
Accuracy study 41
Test results PCR mix A PCR mix B
Reference STDEV
sample Test1 Test2 Test3 Test1 Test2 Test3
P1 24.00 23.91 24.09 No Ct No Ct No Ct 0.09
P2 27.35 27.66 27.52 No Ct No Ct No Ct 0.16
P3 20.12 19.91 19.98 No Ct No Ct No Ct 0.11
P4 29.70 29.97 29.91 No Ct No Ct No Ct 0.14
P5 27.13 27.18 27.17 No Ct No Ct No Ct 0.03
P6 31.35 31.50 31.21 No Ct No Ct No Ct 0.15
P7 24.63 24.60 24.70 No Ct No Ct No Ct 0.05
P8 No Ct No Ct No Ct 21.27 20.96 21.27 0.18
P9 No Ct No Ct No Ct 26.12 26.12 25.84 0.16
P10 No Ct No Ct No Ct 28.55 28.40 28.48 0.08
P11 No Ct No Ct No Ct 21.16 21.07 21.14 0.05
P12 No Ct No Ct No Ct 29.42 29.22 29.43 0.16
Conclusion: It can be seen from the above that, the test results of reference
samples P1-P12 by using the 3 batches of reagents (2017001, 2017002, 2017003)
were all positive, and there was no non-specific amplification. The standard
deviation (STDEV) of the Ct values of each sample is between 0.02 and 3.29.
Therefore, this kit has good detection accuracy.
Figure 2-20 The test results of samples for influenza A virus, influenza B virus and
respiratory syncytial virus by using PCR mix B (2017001 batch)
Figure 2-21 The test results of samples for influenza A virus, influenza B virus and
respiratory syncytial virus by using PCR mix A (2017002 batch)
Accuracy study 43
Figure 2-22 The test results of samples for influenza A virus, influenza B virus and
respiratory syncytial virus by using PCR mix B (2017002 batch)
Figure 2-23 The test results of samples for influenza A virus, influenza B virus and
respiratory syncytial virus by using PCR mix A (2017003 batch)
Figure 2-24 The test results of samples for influenza A virus, influenza B virus and
respiratory syncytial virus by using PCR mix B (2017003 batch)
Accuracy study 44
Figure 2-25 The test results of samples for Adenovirus, Mycoplasma pneumoniae and
Human rhinovirus by using PCR mix A (2017001 batch)
Figure 2-26 The test results of samples for Adenovirus, Mycoplasma pneumoniae and
Human rhinovirus by using PCR mix B (2017001 batch)
Figure 2-27 The test results of samples for Adenovirus, Mycoplasma pneumoniae and
Human rhinovirus by using PCR mix A (2017002 batch)
Accuracy study 45
Figure 2-28 The test results of samples for Adenovirus, Mycoplasma pneumoniae and
Human rhinovirus by using PCR mix B (2017002 batch)
Figure 2-29 The test results of samples for Adenovirus, Mycoplasma pneumoniae and
Human rhinovirus by using PCR mix A (2017003 batch)
Figure 2-30 The test results of samples for Adenovirus, Mycoplasma pneumoniae and
Human rhinovirus by using PCR mix B (2017003 batch)
Accuracy study 46
Figure 2-31 The test results of negative samples by using PCR mix A (2017001 batch)
Figure 2-32 The test results of negative samples by using PCR mix B (2017001 batch)
Figure 2-33 The test results of negative samples by using PCR mix A (2017002 batch)
Accuracy study 47
Figure 2-34 The test results of negative samples by using PCR mix B (2017002 batch)
Figure 2-35 The test results of negative samples by using PCR mix A (2017003 batch)
Figure 2-36 The test results of negative samples by using PCR mix B (2017003 batch)
Table 2-11 Test results of samples for influenza A virus
Ct value of samples
Batch Reaction mix
1 2 3 4 5 6 7 8
PCR mix A 25.65 19.60 17.98 28.07 31.31 22.96 24.52 22.17
2017001
PCR mix B No Ct No Ct No Ct No Ct No Ct No Ct No Ct No Ct
PCR mix A 25.90 20.07 17.68 28.97 31.21 23.70 25.19 21.88
2017002
PCR mix B No Ct No Ct No Ct No Ct No Ct No Ct No Ct No Ct
2017003 PCR mix A 26.18 20.01 18.31 28.55 31.37 23.72 25.36 21.91
Accuracy study 48
PCR mix B No Ct No Ct No Ct No Ct No Ct No Ct No Ct No Ct
Accuracy study 49
Table 2-16 Test results of samples for Human rhinovirus
Ct value of samples
Batch Reaction mix
1 2 3 4 5 6 7 8
PCR mix A No Ct No Ct No Ct No Ct No Ct No Ct No Ct No Ct
2017001
PCR mix B 23.17 26.76 28.25 34.54 35.93 25.13 30.30 21.18
PCR mix A No Ct No Ct No Ct No Ct No Ct No Ct No Ct No Ct
2017002
PCR mix B 23.50 26.94 28.73 34.49 36.41 25.62 30.59 21.46
PCR mix A No Ct No Ct No Ct No Ct No Ct No Ct No Ct No Ct
2017003
PCR mix B 23.12 26.56 28.35 33.66 34.71 24.95 30.38 20.99
Table 2-17 Test results of negative samples (No amplification curve was observed for
FAM, HEX and CY5, only the Ct value of internal standard was collected)
Ct value of samples
Batch Reaction mix
1 2 3 4 5 6 7 8
PCR mix A 27.38 27.09 27.45 27.11 27.80 27.11 26.94 27.32
2017001
PCR mix B 27.49 27.14 27.26 27.24 27.09 27.13 27.07 27.37
PCR mix A 28.56 28.49 28.27 28.53 28.44 28.45 28.71 28.58
2017002
PCR mix B 28.38 28.49 28.51 28.45 28.67 28.99 28.77 28.58
PCR mix A 27.40 27.63 27.55 27.98 27.79 27.94 27.73 27.52
2017003
PCR mix B 27.33 27.54 27.74 27.79 27.94 27.89 27.77 27.34
Conclusion: It can be seen from the above that, the test results of positive
throat swab samples for influenza A virus, influenza B virus, Respiratory
syncytial virus, Adenovirus, Mycoplasma pneumoniae and Human
rhinovirus by using the 3 batches of reagents (2017001, 2017002, 2017003)
were all positive, and there was no non-specific amplification. Negative
samples were all negative, and the test results of internal standard were all
within the range of reference value. Therefore, this kit has good detection
accuracy.
Accuracy study 50
Figure 2-37 Test results of PC01 and PC02 by using PCR mix A (2017001 batch)
Figure 2-38 Test results of PC01 and PC02 by using PCR mix B (2017001 batch)
Accuracy study 51
Figure 2-39 Test results of PC03—PC06 by using PCR mix A (2017001 batch)
Figure 2-40 Test results of PC03—PC06 by using PCR mix B (2017001 batch)
Figure 2-41 Test results of PC01 and PC02 by using PCR mix A (2017002 batch)
Accuracy study 52
Figure 2-42 Test results of PC01 and PC02 by using PCR mix B (2017002 batch)
Figure 2-43 Test results of PC03—PC06 by using PCR mix A (2017002 batch)
Figure 2-44 Test results of PC03—PC06 by using PCR mix B (2017002 batch)
Accuracy study 53
Figure 2-45 Test results of PC01 and PC02 by using PCR mix A (2017003 batch)
Figure 2-46 Test results of PC01 and PC02 by using PCR mix B (2017003 batch)
Accuracy study 54
Figure 2-47 Test results of PC03—PC06 by using PCR mix A (2017003 batch)
Figure 2-48 Test results of PC03—PC06 by using PCR mix B (2017003 batch)
Table 2-18 Test results of national reference samples PC01~PC06 (2017001 batch)
Test results PCR mix A PCR mix B
Reference
samples Test1 Test2 Test1 Test2
PC01 28.97 29.37 No Ct No Ct
PC02 29.12 29.21 No Ct No Ct
PC03 28.16 27.96 No Ct No Ct
PC04 28.12 28.17 No Ct No Ct
PC05 28.13 28.73 No Ct No Ct
PC06 28.05 28.02 No Ct No Ct
Table 2-19 Test results of national reference samples PC01~PC06 (2017002 batch)
Test results PCR mix A PCR mix B
Reference
samples Test1 Test2 Test1 Test2
PC01 29.15 29.15 No Ct No Ct
PC02 29.11 29.10 No Ct No Ct
PC03 28.06 27.97 No Ct No Ct
PC04 27.99 28.20 No Ct No Ct
PC05 28.08 28.09 No Ct No Ct
PC06 28.06 28.08 No Ct No Ct
Table 2-20 Test results of national reference samples PC01~PC06 (2017003 batch)
Test results PCR mix A PCR mix B
Reference
samples Test1 Test2 Test1 Test2
PC01 29.47 29.12 No Ct No Ct
Accuracy study 55
PC02 29.18 29.13 No Ct No Ct
PC03 28.01 28.02 No Ct No Ct
PC04 27.97 28.06 No Ct No Ct
PC05 28.05 28.05 No Ct No Ct
PC06 28.13 28.18 No Ct No Ct
Conclusion: It can be seen from the above that, when testing the positive
reference samples PC01~PC06 in the national reference samples for influenza
A/B virus nucleic acid detection reagent by using the 3 batches of reagents
(2017001, 2017002, 2017003), PC01 and PC02 were all positive with influenza B
virus, PC03—PC06 were all positive with influenza A virus, and there was no
non-specific amplification. The results met the requirements of national
reference sample testing for influenza A/B virus nucleic acid detection
reagents. Therefore, this kit has good detection accuracy.
Accuracy study 56
Six Respiratory Pathogens Nucleic Acid Diagnostic Kit
(PCR-Fluorescence Probing)
Precision Study
Table 1-1 Ct value of imprecision test of reagents within the batch of 2017001
Proj R1 R2 R3 R4 R5 R6
ect FA FA FA FA
HEX HEX CY5 CY5 HEX CY5 HEX CY5
No. M M M M
22.3 24.0 25.5 27.4 25.2 29.1 23.1 24.5 25.7 26.1 27.6 28.9
1
6 2 7 8 9 2 3 3 9 9 9 0
22.0 23.6 25.5 27.6 25.2 29.3 22.9 24.4 25.6 25.6 27.2 28.4
2
2 4 9 3 2 5 3 0 3 7 1 2
22.0 23.6 25.4 27.3 25.3 29.2 22.8 24.3 25.5 25.7 27.2 28.3
3
7 7 3 0 4 3 8 4 6 7 7 4
22.0 23.7 25.5 27.4 25.3 29.0 22.7 24.2 25.4 25.8 27.2 28.5
4
7 1 7 4 2 2 1 5 5 8 9 3
21.7 23.5 25.7 27.4 25.4 28.9 22.5 24.0 25.3 25.9 27.3 28.5
5
8 0 3 8 2 7 5 9 2 4 3 9
21.9 23.5 25.8 27.6 25.3 28.7 22.7 24.2 25.5 26.0 27.4 28.5
6
8 3 9 4 5 8 5 9 2 1 2 7
22.0 23.6 25.8 27.5 25.2 28.5 22.9 24.4 25.6 26.1 27.7 28.8
7
8 7 1 7 5 2 2 0 0 6 1 2
22.3 23.9 25.5 27.4 25.4 29.2 23.0 24.4 25.6 26.0 27.5 28.5
8
2 1 4 1 3 0 2 9 9 4 0 6
22.3 24.0 26.0 27.7 25.2 29.1 23.0 24.5 25.7 25.9 27.4 28.4
9
1 0 2 4 2 7 4 1 2 2 0 4
22.2 23.7 25.4 27.2 25.4 29.2 22.9 24.2 25.5 25.9 27.4 28.5
10
1 4 4 9 3 4 1 9 5 0 1 7
0.81 0.77 0.77 0.54 0.33 0.86 0.75 0.56 0.53 0.61 0.61 0.60
CV
% % % % % % % % % % % %
Table 1-2 Ct value of imprecision test of reagents within the batch of 2017002
Proj R1 R2 R3 R4 R5 R6
ect FA FA FA FA
HEX HEX CY5 CY5 HEX CY5 HEX CY5
No. M M M M
22.8 23.9 26.4 27.8 25.0 28.7 23.4 24.7 24.9 26.2 27.2 27.5
1
6 8 9 5 8 7 1 6 6 2 6 4
Table 1-3 Ct value of imprecision test of reagents within the batch of 2017003
Proj R1 R2 R3 R4 R5 R6
ect FA FA FA FA
HEX HEX CY5 CY5 HEX CY5 HEX CY5
No. M M M M
22.0 23.3 25.1 26.6 24.3 27.7 22.6 23.7 24.5 25.4 26.7 27.3
1
2 3 3 9 4 2 5 3 8 3 3 6
21.7 22.8 25.1 26.6 24.0 28.1 22.5 23.8 24.3 25.1 26.3 27.1
2
3 8 0 9 2 5 1 7 9 3 9 9
22.1 23.4 25.1 26.7 24.1 28.2 22.5 23.9 24.4 25.2 26.6 27.3
3
1 3 4 8 8 3 7 4 4 0 5 4
21.7 23.0 25.2 26.7 24.1 28.2 22.3 23.5 24.2 25.4 26.7 27.4
4
2 2 4 2 8 1 6 4 9 8 3 3
21.4 22.7 25.2 26.9 23.8 28.3 22.1 23.3 24.1 25.3 26.7 27.3
5
1 3 3 5 5 2 7 3 7 7 1 5
21.7 22.8 26.3 27.1 24.2 28.3 22.2 23.3 24.2 25.3 26.7 27.3
6
5 8 8 0 2 6 1 1 2 0 6 9
21.8 23.0 25.3 27.0 24.1 28.0 22.4 23.6 24.3 25.4 26.8 27.3
7
2 8 2 0 2 9 7 7 7 0 8 8
21.6 23.0 25.3 26.9 24.4 28.1 22.2 23.4 24.0 25.2 26.7 27.4
8
8 8 1 4 9 3 1 8 9 8 2 4
21.8 23.0 25.2 26.7 24.1 27.9 22.2 23.5 24.2 25.1 26.4 27.2
9
8 3 1 6 7 5 7 9 0 6 3 2
22.0 23.2 25.4 26.9 24.3 28.1 22.1 23.3 24.1 25.2 26.4 27.3
10
1 0 3 9 1 0 8 6 2 7 8 0
0.93 0.92 1.48 0.56 0.73 0.66 0.79 0.94 0.64 0.47 0.60 0.31
CV
% % % % % % % % % % % %
Conclusion: It can be seen from the above results that, the positive
detection rate of the three batches of kits for normal negative control are all
0% and the positive detection rate of the positive control samples R1-R6 are all
100%. For the precision reference samples R1-R6, the CV of detected Ct values
within the three batches of kits range from 0.31% to 1.48%, the CV of detected
Ct values between batches range from 1.55% to 2.29%, and both of the values
are less than 5%. Therefore, it can be considered that this kit has good
detection precision within and between batches.
Table 1-5 Ct value of intra-day imprecision test of reagents within the batch of 2017001
Proj R1 R2 R3 R4 R5 R6
ect FA FA FA FA
HEX HEX CY5 CY5 HEX CY5 HEX CY5
No. M M M M
22.3 24.0 25.5 27.4 25.2 29.1 23.1 24.5 25.7 26.1 27.6 28.9
1
6 2 7 8 9 2 3 3 9 9 9 0
22.0 23.6 25.5 27.6 25.2 29.3 22.9 24.4 25.6 25.6 27.2 28.4
2
2 4 9 3 2 5 3 0 3 7 1 2
22.0 23.6 25.4 27.3 25.3 29.2 22.8 24.3 25.5 25.7 27.2 28.3
3
7 7 3 0 4 3 8 4 6 7 7 4
22.0 23.7 25.5 27.4 25.3 29.0 22.7 24.2 25.4 25.8 27.2 28.5
4
7 1 7 4 2 2 1 5 5 8 9 3
21.7 23.5 25.7 27.4 25.4 28.9 22.5 24.0 25.3 25.9 27.3 28.5
5
8 0 3 8 2 7 5 9 2 4 3 9
21.9 23.5 25.8 27.6 25.3 28.7 22.7 24.2 25.5 26.0 27.4 28.5
6
8 3 9 4 5 8 5 9 2 1 2 7
22.0 23.6 25.8 27.5 25.2 28.5 22.9 24.4 25.6 26.1 27.7 28.8
7
8 7 1 7 5 2 2 0 0 6 1 2
22.3 23.9 25.5 27.4 25.4 29.2 23.0 24.4 25.6 26.0 27.5 28.5
8
2 1 4 1 3 0 2 9 9 4 0 6
22.3 24.0 26.0 27.7 25.2 29.1 23.0 24.5 25.7 25.9 27.4 28.4
9
1 0 2 4 2 7 4 1 2 2 0 4
22.2 23.7 25.4 27.2 25.4 29.2 22.9 24.2 25.5 25.9 27.4 28.5
10
1 4 4 9 3 4 1 9 5 0 1 7
22.3 22.9 25.4 26.2 25.1 28.4 23.0 23.5 25.4 25.3 26.2 28.0
11
5 6 3 5 1 9 4 2 6 8 1 8
22.2 22.8 25.5 26.4 25.0 28.8 22.9 23.4 25.3 25.5 26.2 28.0
12
7 1 6 3 3 2 7 4 8 1 5 2
22.2 22.8 25.7 26.6 24.8 29.0 22.8 23.3 25.3 25.7 26.3 28.2
13
1 4 1 4 8 9 6 2 1 3 9 6
22.0 22.4 25.8 26.7 24.9 29.0 22.5 23.1 24.9 26.1 26.8 28.5
14
3 3 9 4 6 8 1 1 7 8 0 3
Table 1-6 Ct value of intra-day imprecision test of reagents within the batch of 2017002
Proj R1 R2 R3 R4 R5 R6
ect FA FA FA FA
HEX HEX CY5 CY5 HEX CY5 HEX CY5
No. M M M M
22.8 23.9 26.4 27.8 25.0 28.7 23.4 24.7 24.9 26.2 27.2 27.5
1
6 8 9 5 8 7 1 6 6 2 6 4
22.7 23.9 26.4 27.6 24.8 28.8 23.6 24.7 25.0 26.2 27.4 27.7
2
0 2 2 9 5 9 4 5 8 4 5 5
22.8 24.0 26.4 28.1 24.7 28.3 23.6 24.7 24.9 26.4 27.5 28.0
3
1 1 5 0 2 9 2 9 9 2 7 0
22.8 24.1 26.4 28.0 24.6 28.5 23.8 24.6 24.9 26.5 27.7 27.9
4
1 7 0 6 6 0 2 3 2 5 2 6
22.3 23.4 26.6 28.2 25.0 28.1 23.2 24.2 24.7 26.4 27.4 27.7
5
5 2 0 4 4 6 7 7 8 8 5 0
22.7 23.7 26.5 27.9 25.2 28.6 23.3 24.3 25.0 26.4 27.4 27.6
6
3 4 1 5 8 8 1 5 3 8 5 9
23.0 24.2 26.4 28.1 25.1 28.3 23.4 24.7 24.9 26.5 27.6 27.6
7
2 5 0 9 8 6 7 4 0 1 2 9
23.2 24.3 26.4 28.0 24.9 28.4 23.5 24.7 25.1 26.6 27.9 27.5
8
1 7 0 4 1 0 4 4 3 2 0 9
23.2 24.5 26.6 28.1 24.9 28.5 23.4 24.6 25.0 26.0 27.0 27.4
9
5 0 2 0 5 4 5 1 4 9 3 6
23.1 24.3 26.6 27.8 25.1 28.7 23.6 24.8 25.0 26.1 27.3 27.7
10
8 6 0 1 8 6 8 7 6 7 2 7
22.3 23.4 26.7 27.7 25.5 28.6 23.2 24.0 25.3 26.3 27.2 28.4
11
6 3 0 6 9 9 0 3 6 3 4 7
22.7 23.7 26.7 27.6 25.5 28.9 23.3 24.1 25.2 26.2 27.4 28.6
12
8 5 4 0 5 8 1 8 8 6 5 8
22.6 23.7 27.0 28.0 25.4 28.6 23.5 24.3 25.3 26.1 27.3 28.5
13
2 7 0 3 6 8 9 6 9 9 6 2
22.7 23.7 26.8 28.0 25.3 28.8 23.4 24.2 25.2 26.3 27.4 28.5
14
0 0 9 2 2 9 4 9 6 0 1 1
Table 1-7 Ct value of intra-day imprecision test of reagents within the batch of 2017003
Proj R1 R2 R3 R4 R5 R6
ect FA FA FA FA
HEX HEX CY5 CY5 HEX CY5 HEX CY5
No. M M M M
22.0 23.3 25.1 26.6 24.3 27.7 22.6 23.7 24.5 25.4 26.7 27.3
1
2 3 3 9 4 2 5 3 8 3 3 6
21.7 22.8 25.1 26.6 24.0 28.1 22.5 23.8 24.3 25.1 26.3 27.1
2
3 8 0 9 2 5 1 7 9 3 9 9
22.1 23.4 25.1 26.7 24.1 28.2 22.5 23.9 24.4 25.2 26.6 27.3
3
1 3 4 8 8 3 7 4 4 0 5 4
21.7 23.0 25.2 26.7 24.1 28.2 22.3 23.5 24.2 25.4 26.7 27.4
4
2 2 4 2 8 1 6 4 9 8 3 3
21.4 22.7 25.2 26.9 23.8 28.3 22.1 23.3 24.1 25.3 26.7 27.3
5
1 3 3 5 5 2 7 3 7 7 1 5
21.7 22.8 26.3 27.1 24.2 28.3 22.2 23.3 24.2 25.3 26.7 27.3
6
5 8 8 0 2 6 1 1 2 0 6 9
21.8 23.0 25.3 27.0 24.1 28.0 22.4 23.6 24.3 25.4 26.8 27.3
7
2 8 2 0 2 9 7 7 7 0 8 8
21.6 23.0 25.3 26.9 24.4 28.1 22.2 23.4 24.0 25.2 26.7 27.4
8
8 8 1 4 9 3 1 8 9 8 2 4
21.8 23.0 25.2 26.7 24.1 27.9 22.2 23.5 24.2 25.1 26.4 27.2
9
8 3 1 6 7 5 7 9 0 6 3 2
22.0 23.2 25.4 26.9 24.3 28.1 22.1 23.3 24.1 25.2 26.4 27.3
10
1 0 3 9 1 0 8 6 2 7 8 0
21.7 23.2 25.1 26.5 24.6 27.8 22.0 23.2 24.6 24.9 26.3 27.7
11
3 0 7 7 8 4 5 3 2 0 7 1
21.7 23.1 25.2 26.9 24.7 27.9 22.0 23.1 24.5 24.9 26.3 27.8
12
0 0 7 0 4 2 9 8 0 7 9 6
21.7 23.2 25.3 26.7 24.7 28.0 22.0 23.2 24.6 24.8 26.3 27.6
13
2 0 4 4 2 7 1 9 0 2 4 7
21.5 22.9 25.3 26.8 24.4 28.0 21.5 22.8 24.2 26.4 27.1 28.0
14
9 3 9 7 7 6 3 7 5 4 1 0
Figure 1-9 Inter-day imprecision test results of negative control (2017001 batch)
Table 1-8 Ct value of inter-day imprecision test of reagents within the batch of 2017001
Proj R1 R2 R3 R4 R5 R6
ect FA FA FA FA
HEX HEX CY5 CY5 HEX CY5 HEX CY5
No. M M M M
22.3 24.0 25.5 27.4 25.2 29.1 23.1 24.5 25.7 26.1 27.6 28.9
1
6 2 7 8 9 2 3 3 9 9 9 0
22.0 23.6 25.5 27.6 25.2 29.3 22.9 24.4 25.6 25.6 27.2 28.4
2
2 4 9 3 2 5 3 0 3 7 1 2
22.0 23.6 25.4 27.3 25.3 29.2 22.8 24.3 25.5 25.7 27.2 28.3
3
7 7 3 0 4 3 8 4 6 7 7 4
22.0 23.7 25.5 27.4 25.3 29.0 22.7 24.2 25.4 25.8 27.2 28.5
4
7 1 7 4 2 2 1 5 5 8 9 3
21.7 23.5 25.7 27.4 25.4 28.9 22.5 24.0 25.3 25.9 27.3 28.5
5
8 0 3 8 2 7 5 9 2 4 3 9
21.9 23.5 25.8 27.6 25.3 28.7 22.7 24.2 25.5 26.0 27.4 28.5
6
8 3 9 4 5 8 5 9 2 1 2 7
22.0 23.6 25.8 27.5 25.2 28.5 22.9 24.4 25.6 26.1 27.7 28.8
7
8 7 1 7 5 2 2 0 0 6 1 2
Table 1-9 Ct value of inter-day imprecision test of reagents within the batch of 2017002
Proj R1 R2 R3 R4 R5 R6
ect FA FA FA FA
HEX HEX CY5 CY5 HEX CY5 HEX CY5
No. M M M M
22.8 23.9 26.4 27.8 25.0 28.7 23.4 24.7 24.9 26.2 27.2 27.5
1
6 8 9 5 8 7 1 6 6 2 6 4
22.7 23.9 26.4 27.6 24.8 28.8 23.6 24.7 25.0 26.2 27.4 27.7
2
0 2 2 9 5 9 4 5 8 4 5 5
22.8 24.0 26.4 28.1 24.7 28.3 23.6 24.7 24.9 26.4 27.5 28.0
3
1 1 5 0 2 9 2 9 9 2 7 0
22.8 24.1 26.4 28.0 24.6 28.5 23.8 24.6 24.9 26.5 27.7 27.9
4
1 7 0 6 6 0 2 3 2 5 2 6
22.3 23.4 26.6 28.2 25.0 28.1 23.2 24.2 24.7 26.4 27.4 27.7
5
5 2 0 4 4 6 7 7 8 8 5 0
22.7 23.7 26.5 27.9 25.2 28.6 23.3 24.3 25.0 26.4 27.4 27.6
6
3 4 1 5 8 8 1 5 3 8 5 9
23.0 24.2 26.4 28.1 25.1 28.3 23.4 24.7 24.9 26.5 27.6 27.6
7
2 5 0 9 8 6 7 4 0 1 2 9
Table 1-10 Ct value of inter-day imprecision test of reagents within the batch of 2017003
Proj R1 R2 R3 R4 R5 R6
ect FA FA FA FA
HEX HEX CY5 CY5 HEX CY5 HEX CY5
No. M M M M
22.0 23.3 25.1 26.6 24.3 27.7 22.6 23.7 24.5 25.4 26.7 27.3
1
2 3 3 9 4 2 5 3 8 3 3 6
21.7 22.8 25.1 26.6 24.0 28.1 22.5 23.8 24.3 25.1 26.3 27.1
2
3 8 0 9 2 5 1 7 9 3 9 9
22.1 23.4 25.1 26.7 24.1 28.2 22.5 23.9 24.4 25.2 26.6 27.3
3
1 3 4 8 8 3 7 4 4 0 5 4
21.7 23.0 25.2 26.7 24.1 28.2 22.3 23.5 24.2 25.4 26.7 27.4
4
2 2 4 2 8 1 6 4 9 8 3 3
21.4 22.7 25.2 26.9 23.8 28.3 22.1 23.3 24.1 25.3 26.7 27.3
5
1 3 3 5 5 2 7 3 7 7 1 5
21.7 22.8 26.3 27.1 24.2 28.3 22.2 23.3 24.2 25.3 26.7 27.3
6
5 8 8 0 2 6 1 1 2 0 6 9
21.8 23.0 25.3 27.0 24.1 28.0 22.4 23.6 24.3 25.4 26.8 27.3
7
2 8 2 0 2 9 7 7 7 0 8 8
Conclusion: It can be seen from the above results that, the positive
detection rate of the three batches of kits for normal negative control are all
0% and the positive detection rate of the positive control samples R1-R6 are all
100%. For the precision reference samples R1-R6, the CV of detected Ct values
of the three batches of kits by different operators and within one day range
from 0.55% to 2.52%, the CV of detected Ct values of the three batches by the
same operators and between days range from 0.38% to 3.54%, and both of the
values are less than 5%. Therefore, it can be considered that this kit has good
detection precision under different operators and at different times.
Figure1-13 Imprecision test results of positive samples for influenza A virus within
batches (2017001 batch)
Figure1-14 Imprecision test results of positive samples for influenza B virus within
batches (2017001 batch)
Figure1-16 Imprecision test results of positive samples for Adenovirus within batches
(2017001 batch)
Table 1-11 Ct value of imprecision test of reagents within the batch of 2017001
Sample Respiratory
Influenza Influenza Human Mycoplasma
No. Syncytial Adenovirus
A virus B virus Rhinovirus pneumoniae
virus
1 29.38 31.57 30.93 29.95 31.48 30.30
2 29.99 31.60 30.58 30.21 30.94 29.30
3 29.15 31.67 30.72 29.85 31.19 29.31
4 29.43 31.62 30.50 29.84 30.78 30.28
5 28.87 31.59 31.32 29.83 31.00 30.54
Table 1-12 Ct value of imprecision test of reagents within the batch of 2017002
Sample Respiratory
Influenza Influenza Human Mycoplasma
No. Syncytial Adenovirus
A virus B virus Rhinovirus pneumoniae
virus
1 29.83 30.43 30.29 29.51 30.39 29.93
2 30.14 30.49 29.73 29.96 30.35 29.86
3 30.17 30.45 28.95 29.30 29.97 29.24
4 29.81 30.14 30.40 28.94 29.85 30.07
5 29.50 30.24 30.41 29.41 30.04 29.98
6 29.98 30.06 29.86 29.28 30.08 29.99
7 30.33 30.40 29.82 29.07 30.27 28.80
8 29.64 30.01 30.45 29.21 30.12 29.41
9 29.93 30.15 30.48 29.23 30.04 30.31
10 30.13 30.53 30.56 29.24 30.34 30.25
CV
0.85% 0.63% 1.68% 0.94% 0.61% 1.62%
value
Table 1-13 Ct value of imprecision test of reagents within the batch of 2017003
Sample Respiratory
Influenza Influenza Human Mycoplasma
No. Syncytial Adenovirus
A virus B virus Rhinovirus pneumoniae
virus
1 29.93 31.27 29.85 29.79 31.02 29.63
2 29.75 31.05 29.34 30.10 30.26 29.10
3 29.71 31.21 29.95 29.74 30.78 29.51
4 29.58 30.95 29.64 29.34 30.43 29.11
5 28.81 30.82 29.65 29.65 31.00 29.44
6 28.84 30.26 30.18 29.11 30.81 29.02
7 28.89 30.41 29.52 28.88 30.24 29.20
8 28.98 30.25 29.77 28.81 30.16 29.23
9 28.87 30.68 29.79 29.04 30.15 29.34
10 29.09 30.11 29.66 29.03 30.21 29.37
CV
1.51% 1.38% 0.78% 1.51% 1.17% 0.67%
value
Sample Respiratory
Influenza Influenza Human Mycoplasma
No. Syncytial Adenovirus
A virus B virus Rhinovirus pneumoniae
virus
1 29.38 31.57 30.93 29.95 31.48 30.30
2 29.99 31.60 30.58 30.21 30.94 29.30
3 29.15 31.67 30.72 29.85 31.19 29.31
4 29.43 31.62 30.50 29.84 30.78 30.28
5 28.87 31.59 31.32 29.83 31.00 30.54
6 29.05 31.31 31.35 29.41 30.96 29.89
7 29.02 30.73 30.71 29.19 30.92 30.33
8 28.86 30.92 30.86 28.93 30.66 30.44
9 29.10 31.07 30.81 29.16 30.87 30.15
10 28.88 30.95 30.71 29.00 30.48 30.53
11 29.83 30.43 30.29 29.51 30.39 29.93
12 30.14 30.49 29.73 29.96 30.35 29.86
13 30.17 30.45 28.95 29.30 29.97 29.24
14 29.81 30.14 30.40 28.94 29.85 30.07
15 29.50 30.24 30.41 29.41 30.04 29.98
16 29.98 30.06 29.86 29.28 30.08 29.99
17 30.33 30.40 29.82 29.07 30.27 28.80
18 29.64 30.01 30.45 29.21 30.12 29.41
19 29.93 30.15 30.48 29.23 30.04 30.31
20 30.13 30.53 30.56 29.24 30.34 30.25
21 29.93 31.27 29.85 29.79 31.02 29.63
22 29.75 31.05 29.34 30.10 30.26 29.10
23 29.71 31.21 29.95 29.74 30.78 29.51
24 29.58 30.95 29.64 29.34 30.43 29.11
25 28.81 30.82 29.65 29.65 31.00 29.44
26 28.84 30.26 30.18 29.11 30.81 29.02
27 28.89 30.41 29.52 28.88 30.24 29.20
28 28.98 30.25 29.77 28.81 30.16 29.23
29 28.87 30.68 29.79 29.04 30.15 29.34
30 29.09 30.11 29.66 29.03 30.21 29.37
CV
1.68% 1.74% 1.94% 1.35% 1.39% 1.73%
value
It can be seen from the above results that, the accuracy of the three
batches of kits to detect clinical samples is 100%, the CV of detected Ct values
within the three batches of kits range from 0.31% to 1.68%, the CV of detected
Ct values between batches range from 1.35% to 1.94%, and both of the values
are less than 5%. Therefore, it can be considered that this kit has good
detection precision within and between batches.
Figure 1-19 Intra-day imprecision test results of positive samples for influenza A virus
(2017001 batch)
Figure 1-20 Intra-day imprecision test results of positive samples for influenza B virus
(2017001 batch)
Figure 1-22 Intra-day imprecision test results of positive samples for Adenovirus
(2017001 batch)
Figure 1-24 Intra-day imprecision test results of positive samples for Mycoplasma
pneumoniae (2017001 batch)
Table 1-15 Ct value of intra-day imprecision test of reagents within the batch of 2017001
Sample Respiratory
Influenza Influenza Human Mycoplasma
No. Syncytial Adenovirus
A virus B virus Rhinovirus pneumoniae
virus
1 29.38 31.57 30.93 29.95 31.48 30.30
2 29.99 31.60 30.58 30.21 30.94 29.30
3 29.15 31.67 30.72 29.85 31.19 29.31
4 29.43 31.62 30.50 29.84 30.78 30.28
5 28.87 31.59 31.32 29.83 31.00 30.54
Table 1-16 Ct value of intra-day imprecision test of reagents within the batch of 2017002
Sample Respiratory
Influenza Influenza Human Mycoplasma
No. Syncytial Adenovirus
A virus B virus Rhinovirus pneumoniae
virus
1 29.83 30.43 30.29 29.51 30.39 29.93
2 30.14 30.49 29.73 29.96 30.35 29.86
3 30.17 30.45 28.95 29.30 29.97 29.24
4 29.81 30.14 30.40 28.94 29.85 30.07
5 29.50 30.24 30.41 29.41 30.04 29.98
6 29.98 30.06 29.86 29.28 30.08 29.99
7 30.33 30.40 29.82 29.07 30.27 28.80
8 29.64 30.01 30.45 29.21 30.12 29.41
9 29.93 30.15 30.48 29.23 30.04 30.31
10 30.13 30.53 30.56 29.24 30.34 30.25
11 29.98 31.42 29.86 29.18 30.77 29.42
12 29.10 31.31 29.55 29.51 30.42 29.24
13 29.68 31.16 30.28 29.13 30.63 28.89
14 28.99 30.32 29.66 29.28 30.51 28.32
15 28.54 31.32 30.11 29.31 30.58 29.38
16 28.25 30.90 30.44 28.83 30.27 28.52
17 28.57 29.53 29.61 28.59 29.79 28.33
18 28.76 30.18 30.11 28.56 30.25 29.26
19 28.46 30.13 29.84 28.54 29.97 29.39
20 28.78 30.07 29.58 28.79 30.19 29.56
CV
2.29% 1.67% 1.40% 1.22% 0.87% 2.06%
value
Table 1-17 Ct value of intra-day imprecision test of reagents within the batch of 2017003
Figure 1-26 Inter-day imprecision test results of positive samples for influenza B virus
(2017001 batch)
Figure 1-28 Inter-day imprecision test results of positive samples for Adenovirus
(2017001 batch)
Figure 1-30 Inter-day imprecision test results of positive samples for Mycoplasma
pneumoniae (2017001 batch)
Table 1-18 Ct value of inter-day imprecision test of reagents within the batch of 2017001
Sample Respiratory
Influenza Influenza Human Mycoplasma
No. Syncytial Adenovirus
A virus B virus Rhinovirus pneumoniae
virus
1 29.38 31.57 30.93 29.95 31.48 30.30
2 29.99 31.60 30.58 30.21 30.94 29.30
3 29.15 31.67 30.72 29.85 31.19 29.31
4 29.43 31.62 30.50 29.84 30.78 30.28
5 28.87 31.59 31.32 29.83 31.00 30.54
Table 1-19 Ct value of inter-day imprecision test of reagents within the batch of 2017002
Sample Respiratory
Influenza Influenza Human Mycoplasma
No. Syncytial Adenovirus
A virus B virus Rhinovirus pneumoniae
virus
1 29.83 30.43 30.29 29.51 30.39 29.93
2 30.14 30.49 29.73 29.96 30.35 29.86
3 30.17 30.45 28.95 29.30 29.97 29.24
4 29.81 30.14 30.40 28.94 29.85 30.07
5 29.50 30.24 30.41 29.41 30.04 29.98
6 29.98 30.06 29.86 29.28 30.08 29.99
7 30.33 30.40 29.82 29.07 30.27 28.80
8 29.64 30.01 30.45 29.21 30.12 29.41
9 29.93 30.15 30.48 29.23 30.04 30.31
10 30.13 30.53 30.56 29.24 30.34 30.25
11 29.30 30.63 29.93 28.32 30.44 29.91
12 29.24 30.91 30.12 28.54 30.56 30.07
13 29.63 30.56 30.01 28.29 29.98 29.79
14 28.72 30.22 29.88 28.04 30.47 29.79
15 29.06 30.47 30.02 28.53 30.15 29.65
16 28.98 30.85 30.29 28.73 30.43 28.39
17 29.79 30.94 29.83 28.58 30.61 28.84
18 29.23 30.60 29.67 28.76 30.14 29.22
19 28.75 30.52 30.00 28.68 30.36 29.41
20 29.48 31.02 29.99 29.17 30.78 29.66
CV
1.62% 1.00% 1.24% 1.66% 0.81% 1.74%
value
Table 1-20 Ct value of inter-day imprecision test of reagents within the batch of 2017003
Conclusion: It can be seen from the above results that, the accuracy of the
three batches of kits to detect clinical samples is 100%, the CV of detected Ct
values of inter-day imprecision test among the three batched of kits by
different operators range from 0.84% to 3.59%; the CV of detected Ct values of
intra-day imprecision test among the three batched of kits by the same
operators range from 0.81% to 2.43%, and both of the values are less than 5%.
Therefore, it can be considered that this kit has good detection precision
under different operators and at different times.
Figure 1-32 The test of national precision reference sample CV2 (2017001-2017003 batch)
Table 1-21 Ct value of precision test of the three batches of kits on the national reference
samples CV1 and CV2
Conclusion: It can be seen from the above results that, the accuracy of the
three batches of kits to detect the precision reference samples CV1 and CV2 in
the national reference samples of nucleic acid detection reagents for influenza
A/B virus is 100%, and the CV of detected Ct values of precision between
batches are 2.83% and 2.28%, which are both lower than 5%. It meets the
requirements of national reference sample detection. Therefore, it can be
considered that this kit has good detection precision on national precision
reference samples.
Figure 2-1 Imprecision test results of negative control within batches (2017001 batch)
Figure 2-3 Imprecision test results of R3 and R4 within batches (2017001 batch)
Figure 2-4 Imprecision test results of R5 and R6 within batches (2017001 batch)
Table 2-1 Ct value of imprecision test of reagents within the batch of 2017001
Proj R1 R2 R3 R4 R5 R6
ect FA FA FA FA
HEX HEX CY5 CY5 HEX CY5 HEX CY5
No. M M M M
20.9 23.0 24.9 27.2 24.3 27.9 21.5 23.5 23.3 24.9 27.1 27.6
1
7 9 0 0 5 1 1 6 3 4 0 8
2 20.9 23.0 24.8 27.2 24.2 27.9 21.4 23.5 23.2 24.7 26.9 27.5
Table 2-2 Ct value of imprecision test of reagents within the batch of 2017002
Proj R1 R2 R3 R4 R5 R6
ect FA FA FA FA
HEX HEX CY5 CY5 HEX CY5 HEX CY5
No. M M M M
20.4 23.1 24.6 27.0 23.7 27.6 21.4 23.4 23.2 26.9 26.7 26.9
1
6 1 8 3 0 2 8 7 0 7 6 7
20.4 23.0 24.6 27.0 23.7 27.5 21.5 23.4 23.2 27.0 26.7 27.0
2
4 4 3 1 0 3 1 1 4 0 3 0
20.5 23.0 24.6 27.0 23.7 27.4 21.4 23.3 23.1 27.0 26.8 27.0
3
1 6 4 0 5 6 9 6 9 2 1 2
20.5 23.0 24.6 27.0 23.5 27.5 20.9 23.3 23.1 26.9 26.7 26.9
4
9 4 1 0 6 5 2 6 3 4 8 4
20.3 22.4 24.6 27.0 23.7 27.5 21.5 23.4 23.2 26.9 26.7 26.9
5
3 7 3 4 0 9 3 7 8 4 6 4
20.7 23.0 24.6 27.0 23.6 27.5 21.0 23.3 23.1 26.9 26.7 26.9
6
9 4 1 3 3 9 1 2 9 3 1 3
20.7 23.0 24.6 26.9 23.5 27.6 21.4 23.3 23.2 26.8 26.6 26.8
7
0 2 2 9 7 4 6 4 0 8 2 8
20.7 23.0 24.6 27.0 23.5 27.6 21.5 23.4 23.2 27.0 26.7 27.0
8
1 4 4 1 7 5 2 5 0 2 6 2
20.4 23.0 24.6 26.9 23.7 27.5 21.4 23.4 23.2 27.0 26.8 27.0
9
3 5 8 9 1 1 7 7 4 3 6 3
20.4 23.0 24.6 27.0 23.6 27.4 21.5 23.4 23.3 26.9 26.8 26.9
10
9 3 6 3 7 8 8 4 0 7 1 7
0.72 0.80 0.10 0.07 0.29 0.24 1.08 0.25 0.21 0.18 0.24 0.18
CV
% % % % % % % % % % % %
Proj R1 R2 R3 R4 R5 R6
ect FA FA FA FA
HEX HEX CY5 CY5 HEX CY5 HEX CY5
No. M M M M
21.6 22.8 25.5 26.3 23.6 26.9 21.4 22.7 23.3 25.3 26.2 27.2
1
3 7 7 7 2 6 6 1 2 4 9 0
21.6 22.8 25.6 26.4 23.6 27.1 21.4 22.7 23.2 25.3 26.3 27.2
2
2 9 1 0 2 2 8 2 4 2 0 6
21.6 22.9 25.5 26.3 23.6 27.1 21.4 22.7 23.2 25.3 26.3 27.2
3
4 0 6 8 2 8 6 0 6 6 2 5
21.6 22.8 25.5 26.3 23.6 27.1 21.4 22.7 23.2 25.5 26.3 27.2
4
1 6 6 7 1 4 6 2 1 3 1 3
21.6 22.8 25.6 26.4 23.6 27.0 21.4 22.7 23.2 25.4 26.3 27.3
5
2 9 7 2 2 4 5 3 9 7 2 2
21.6 22.8 25.5 26.3 23.5 27.0 21.4 22.7 23.3 25.4 26.2 27.1
6
4 5 6 4 9 9 8 2 0 4 8 6
21.6 22.8 25.5 26.3 23.6 27.1 21.4 22.7 23.2 25.4 26.3 27.1
7
2 9 7 2 3 0 7 5 9 1 0 9
21.5 22.8 25.5 26.2 23.6 27.0 21.5 22.7 23.2 25.3 26.2 27.2
8
7 6 8 6 3 5 1 5 9 9 5 0
21.6 22.9 25.5 26.3 23.5 27.0 21.4 22.7 23.3 25.4 26.5 27.3
9
5 1 4 1 9 6 7 2 3 9 8 9
21.6 22.8 25.3 26.3 23.6 27.1 21.4 22.7 23.3 25.3 26.2 27.2
10
2 9 9 2 0 4 9 3 5 0 9 4
0.10 0.09 0.28 0.18 0.06 0.23 0.08 0.07 0.18 0.30 0.35 0.25
CV
% % % % % % % % % % % %
Proj R1 R2 R3 R4 R5 R6
ect FA FA FA FA
HEX HEX CY5 CY5 HEX CY5 HEX CY5
No. M M M M
20.9 23.0 24.9 27.2 24.3 27.9 21.5 23.5 23.3 24.9 27.1 27.6
1
7 9 0 0 5 1 1 6 3 4 0 8
20.9 23.0 24.8 27.2 24.2 27.9 21.4 23.5 23.2 24.7 26.9 27.5
2
4 8 5 0 1 0 7 0 1 0 5 0
21.0 23.0 24.7 26.9 24.2 28.0 21.4 23.4 23.1 24.8 27.1 27.4
3
2 8 2 9 6 5 5 7 7 7 1 3
21.3 23.0 24.8 27.1 24.2 27.7 21.3 23.4 23.1 24.8 26.8 27.3
4
0 0 0 9 6 7 7 8 7 4 7 8
21.0 23.1 24.8 27.2 24.0 27.6 21.4 23.5 23.2 24.9 27.0 27.6
5
9 4 7 8 3 8 4 9 4 4 3 9
Conclusion: It can be seen from the above results that, the positive
detection rate of the three batches of kits for normal negative control are all
0% and the positive detection rate of the positive control samples R1-R6 are all
100%. For the precision reference samples R1-R6, the CV of detected Ct values
within the three batches of kits range from 0.06% to 1.08%, the CV of detected
Ct values between batches range from 0.24% to 3.50%, and both of the values
are less than 5%. Therefore, it can be considered that this kit has good
detection precision within and between batches.
Figure 2-5 Intra-day imprecision test results of negative control (2017001 batch)
Table 2-5 Ct value of intra-day imprecision test of reagents within the batch of 2017001
Proj R1 R2 R3 R4 R5 R6
ect FA FA FA FA
HEX HEX CY5 CY5 HEX CY5 HEX CY5
No. M M M M
20.9 23.0 24.9 27.2 24.3 27.9 21.5 23.5 23.3 24.9 27.1 27.6
1
7 9 0 0 5 1 1 6 3 4 0 8
20.9 23.0 24.8 27.2 24.2 27.9 21.4 23.5 23.2 24.7 26.9 27.5
2
4 8 5 0 1 0 7 0 1 0 5 0
21.0 23.0 24.7 26.9 24.2 28.0 21.4 23.4 23.1 24.8 27.1 27.4
3
2 8 2 9 6 5 5 7 7 7 1 3
21.3 23.0 24.8 27.1 24.2 27.7 21.3 23.4 23.1 24.8 26.8 27.3
4
0 0 0 9 6 7 7 8 7 4 7 8
21.0 23.1 24.8 27.2 24.0 27.6 21.4 23.5 23.2 24.9 27.0 27.6
5
9 4 7 8 3 8 4 9 4 4 3 9
21.0 23.1 24.8 27.2 24.1 27.6 21.4 23.5 23.2 24.8 27.0 27.4
6
6 4 2 2 7 4 2 8 4 7 2 1
21.0 23.1 24.9 27.1 24.1 27.6 21.5 23.5 23.2 24.9 26.9 27.4
7
8 8 1 7 9 2 6 9 7 2 8 5
8 20.9 23.1 24.9 27.1 24.2 27.5 21.5 23.5 23.2 25.1 27.1 27.5
Table 2-6 Ct value of intra-day imprecision test of reagents within the batch of 2017002
Proj R1 R2 R3 R4 R5 R6
ect FA FA FA FA
HEX HEX CY5 CY5 HEX CY5 HEX CY5
No. M M M M
20.4 23.1 24.6 27.0 23.7 27.6 21.4 23.4 23.2 26.9 26.7 26.9
1
6 1 8 3 0 2 8 7 0 7 6 7
20.4 23.0 24.6 27.0 23.7 27.5 21.5 23.4 23.2 27.0 26.7 27.0
2
4 4 3 1 0 3 1 1 4 0 3 0
20.5 23.0 24.6 27.0 23.7 27.4 21.4 23.3 23.1 27.0 26.8 27.0
3
1 6 4 0 5 6 9 6 9 2 1 2
20.5 23.0 24.6 27.0 23.5 27.5 20.9 23.3 23.1 26.9 26.7 26.9
4
9 4 1 0 6 5 2 6 3 4 8 4
20.3 22.4 24.6 27.0 23.7 27.5 21.5 23.4 23.2 26.9 26.7 26.9
5
3 7 3 4 0 9 3 7 8 4 6 4
20.7 23.0 24.6 27.0 23.6 27.5 21.0 23.3 23.1 26.9 26.7 26.9
6
9 4 1 3 3 9 1 2 9 3 1 3
20.7 23.0 24.6 26.9 23.5 27.6 21.4 23.3 23.2 26.8 26.6 26.8
7
0 2 2 9 7 4 6 4 0 8 2 8
8 20.7 23.0 24.6 27.0 23.5 27.6 21.5 23.4 23.2 27.0 26.7 27.0
Table 2-7 Ct value of intra-day imprecision test of reagents within the batch of 2017003
Proj R1 R2 R3 R4 R5 R6
ect FA FA FA FA
HEX HEX CY5 CY5 HEX CY5 HEX CY5
No. M M M M
21.6 22.8 25.5 26.3 23.6 26.9 21.4 22.7 23.3 25.3 26.2 27.2
1
3 7 7 7 2 6 6 1 2 4 9 0
21.6 22.8 25.6 26.4 23.6 27.1 21.4 22.7 23.2 25.3 26.3 27.2
2
2 9 1 0 2 2 8 2 4 2 0 6
21.6 22.9 25.5 26.3 23.6 27.1 21.4 22.7 23.2 25.3 26.3 27.2
3
4 0 6 8 2 8 6 0 6 6 2 5
21.6 22.8 25.5 26.3 23.6 27.1 21.4 22.7 23.2 25.5 26.3 27.2
4
1 6 6 7 1 4 6 2 1 3 1 3
21.6 22.8 25.6 26.4 23.6 27.0 21.4 22.7 23.2 25.4 26.3 27.3
5
2 9 7 2 2 4 5 3 9 7 2 2
21.6 22.8 25.5 26.3 23.5 27.0 21.4 22.7 23.3 25.4 26.2 27.1
6
4 5 6 4 9 9 8 2 0 4 8 6
21.6 22.8 25.5 26.3 23.6 27.1 21.4 22.7 23.2 25.4 26.3 27.1
7
2 9 7 2 3 0 7 5 9 1 0 9
8 21.5 22.8 25.5 26.2 23.6 27.0 21.5 22.7 23.2 25.3 26.2 27.2
Figure 2-9 Inter-day imprecision test results of negative control (2017001 batch)
Table 2-8 Ct value of inter-day imprecision test of reagents within the batch of 2017001
Proj R1 R2 R3 R4 R5 R6
ect FA FA FA FA
HEX HEX CY5 CY5 HEX CY5 HEX CY5
No. M M M M
20.9 23.0 24.9 27.2 24.3 27.9 21.5 23.5 23.3 24.9 27.1 27.6
1
7 9 0 0 5 1 1 6 3 4 0 8
20.9 23.0 24.8 27.2 24.2 27.9 21.4 23.5 23.2 24.7 26.9 27.5
2
4 8 5 0 1 0 7 0 1 0 5 0
3 21.0 23.0 24.7 26.9 24.2 28.0 21.4 23.4 23.1 24.8 27.1 27.4
Table 2-9 Ct value of inter-day imprecision test of reagents within the batch of 2017002
Proj R1 R2 R3 R4 R5 R6
ect FA FA FA FA
HEX HEX CY5 CY5 HEX CY5 HEX CY5
No. M M M M
20.4 23.1 24.6 27.0 23.7 27.6 21.4 23.4 23.2 26.9 26.7 26.9
1
6 1 8 3 0 2 8 7 0 7 6 7
20.4 23.0 24.6 27.0 23.7 27.5 21.5 23.4 23.2 27.0 26.7 27.0
2
4 4 3 1 0 3 1 1 4 0 3 0
3 20.5 23.0 24.6 27.0 23.7 27.4 21.4 23.3 23.1 27.0 26.8 27.0
Table 2-10 Ct value of inter-day imprecision test of reagents within the batch of 2017003
Proj R1 R2 R3 R4 R5 R6
ect FA FA FA FA
HEX HEX CY5 CY5 HEX CY5 HEX CY5
No. M M M M
21.6 22.8 25.5 26.3 23.6 26.9 21.4 22.7 23.3 25.3 26.2 27.2
1
3 7 7 7 2 6 6 1 2 4 9 0
21.6 22.8 25.6 26.4 23.6 27.1 21.4 22.7 23.2 25.3 26.3 27.2
2
2 9 1 0 2 2 8 2 4 2 0 6
3 21.6 22.9 25.5 26.3 23.6 27.1 21.4 22.7 23.2 25.3 26.3 27.2
Conclusion: It can be seen from the above results that, the positive
detection rate of the three batches of kits for normal negative control are all
0% and the positive detection rate of the positive control samples R1-R6 are all
100%. For the precision reference samples R1-R6, the CV of detected Ct values
of the three batches of kits by different operators and within one day range
from 0.11% to 4.17%, the CV of detected Ct values of the three batches by the
same operators and between days range from 0.15% to 4.33%, and both of the
values are less than 5%. Therefore, it can be considered that this kit has good
detection precision under different operators and at different times.
Figure 2-13 Imprecision test results of positive samples for influenza A virus within
batches (2017001 batch)
Figure 2-15 Imprecision test results of positive samples for respiratory syncytial virus
within batches (2017001 batch)
Figure 2-16 Imprecision test results of positive samples for Adenovirus within batches
(2017001 batch)
Figure 2-18 Imprecision test results of positive samples for Mycoplasma pneumoniae
within batches (2017001 batch)
Table 2-11 Ct value of imprecision test of reagents within the batch of 2017001
Sample Respiratory
Influenza Influenza Human Mycoplasma
No. Syncytial Adenovirus
A virus B virus Rhinovirus pneumoniae
virus
1 29.64 30.20 30.18 29.73 30.47 29.71
2 29.68 30.26 30.21 29.69 30.42 30.21
Table 2-12 Ct value of imprecision test of reagents within the batch of 2017002
Sample Respiratory
Influenza Influenza Human Mycoplasma
No. Syncytial Adenovirus
A virus B virus Rhinovirus pneumoniae
virus
1 29.24 30.65 30.75 29.38 30.81 30.62
2 29.23 30.43 30.32 29.34 30.42 30.20
3 29.20 30.77 30.74 29.46 30.70 30.37
4 29.25 30.84 30.87 29.12 30.32 30.24
5 29.21 30.42 30.39 29.43 30.45 30.50
6 29.24 30.45 30.93 29.43 30.51 30.51
7 29.34 30.75 30.46 29.60 30.85 31.19
8 29.29 30.48 30.43 29.61 30.41 31.00
9 29.33 30.73 30.79 29.30 30.74 31.18
10 29.38 30.74 30.60 29.38 30.92 30.56
CV
0.21% 0.53% 0.71% 0.48% 0.70% 1.19%
value
Table 2-13 Ct value of imprecision test of reagents within the batch of 2017003
Sample Respiratory
Influenza Influenza Human Mycoplasma
No. Syncytial Adenovirus
A virus B virus Rhinovirus pneumoniae
virus
1 29.28 30.31 30.85 29.33 30.24 30.50
2 29.43 30.33 30.72 29.41 30.39 30.74
3 29.39 30.11 30.61 29.52 30.34 30.64
4 29.39 30.27 30.83 29.31 30.44 30.76
5 29.37 30.50 30.81 29.36 30.40 30.63
6 29.46 30.24 30.98 29.46 30.38 30.88
7 29.49 30.34 30.81 29.40 30.37 30.83
8 29.27 30.48 30.79 29.46 30.33 30.76
9 29.40 30.21 30.67 29.44 30.46 30.59
10 29.47 30.07 31.04 29.58 30.36 30.83
CV
0.25% 0.46% 0.42% 0.28% 0.20% 0.40%
value
Sample Respiratory
Influenza Influenza Human Mycoplasma
No. Syncytial Adenovirus
A virus B virus Rhinovirus pneumoniae
virus
1 29.64 30.20 30.18 29.73 30.47 29.71
2 29.68 30.26 30.21 29.69 30.42 30.21
3 29.61 30.39 30.17 29.81 30.27 30.19
4 29.66 30.42 30.28 29.66 30.25 30.26
5 29.68 30.35 29.72 29.73 29.91 30.30
6 29.69 30.07 30.37 29.76 30.34 30.01
7 29.71 30.37 30.09 29.90 30.48 30.51
8 29.83 30.30 30.17 29.95 30.37 30.28
9 29.65 30.17 29.97 29.83 30.22 30.11
10 29.73 30.33 30.21 29.72 30.32 30.11
11 29.24 30.65 30.75 29.38 30.81 30.62
12 29.23 30.43 30.32 29.34 30.42 30.20
13 29.20 30.77 30.74 29.46 30.70 30.37
14 29.25 30.84 30.87 29.12 30.32 30.24
15 29.21 30.42 30.39 29.43 30.45 30.50
16 29.24 30.45 30.93 29.43 30.51 30.51
17 29.34 30.75 30.46 29.60 30.85 31.19
18 29.29 30.48 30.43 29.61 30.41 31.00
19 29.33 30.73 30.79 29.30 30.74 31.18
20 29.38 30.74 30.60 29.38 30.92 30.56
21 29.28 30.31 30.85 29.33 30.24 30.50
22 29.43 30.33 30.72 29.41 30.39 30.74
23 29.39 30.11 30.61 29.52 30.34 30.64
24 29.39 30.27 30.83 29.31 30.44 30.76
25 29.37 30.50 30.81 29.36 30.40 30.63
26 29.46 30.24 30.98 29.46 30.38 30.88
27 29.49 30.34 30.81 29.40 30.37 30.83
28 29.27 30.48 30.79 29.46 30.33 30.76
29 29.40 30.21 30.67 29.44 30.46 30.59
30 29.47 30.07 31.04 29.58 30.36 30.83
CV
0.64% 0.70% 1.10% 0.69% 0.67% 1.13%
value
It can be seen from the above results that, the accuracy of the three
batches of kits to detect clinical samples is 100%, the CV of detected Ct values
within the three batches of kits range from 0.20% to 1.19%, the CV of detected
Ct values between batches range from 0.64% to 1.13%, and both of the values
are less than 5%. Therefore, it can be considered that this kit has good
Figure 2-19 Intra-day imprecision test results of positive samples for influenza A virus
(2017001 batch)
Figure 2-21 Intra-day imprecision test results of positive samples for respiratory
syncytial virus (2017001 batch)
Figure 2-22 Intra-day imprecision test results of positive samples for Adenovirus
(2017001 batch)
Figure 2-24 Intra-day imprecision test results of positive samples for Mycoplasma
pneumoniae (2017001 batch)
Table 2-15 Ct value of intra-day imprecision test of reagents within the batch of 2017001
Sample Respiratory
Influenza Influenza Human Mycoplasma
No. Syncytial Adenovirus
A virus B virus Rhinovirus pneumoniae
virus
1 29.64 30.20 30.18 29.73 30.47 29.71
2 29.68 30.26 30.21 29.69 30.42 30.21
3 29.61 30.39 30.17 29.81 30.27 30.19
4 29.66 30.42 30.28 29.66 30.25 30.26
Table 2-16 Ct value of intra-day imprecision test of reagents within the batch of 2017002
Sample Respiratory
Influenza Influenza Human Mycoplasma
No. Syncytial Adenovirus
A virus B virus Rhinovirus pneumoniae
virus
1 29.24 30.65 30.75 30.75 30.81 30.62
2 29.23 30.43 30.32 30.32 30.42 30.20
3 29.20 30.77 30.74 30.74 30.70 30.37
4 29.25 30.84 30.87 30.87 30.32 30.24
5 29.21 30.42 30.39 30.39 30.45 30.50
6 29.24 30.45 30.93 30.93 30.51 30.51
7 29.34 30.75 30.46 30.46 30.85 31.19
8 29.29 30.48 30.43 30.43 30.41 31.00
9 29.33 30.73 30.79 30.79 30.74 31.18
10 29.38 30.74 30.60 30.60 30.92 30.56
11 29.40 31.22 30.83 29.31 30.85 30.68
12 29.41 30.18 30.66 29.47 30.17 30.62
13 29.26 31.23 30.63 29.37 30.01 30.71
14 29.24 31.19 30.79 29.41 28.52 30.55
15 29.40 31.32 30.47 29.45 30.29 30.76
16 29.41 31.33 30.84 29.68 31.15 30.87
17 29.27 31.18 30.98 29.57 31.20 30.74
18 29.49 31.39 30.68 29.61 31.03 30.57
19 29.53 30.07 30.90 29.56 30.51 30.76
20 29.52 30.25 30.62 29.61 29.69 30.56
CV
0.36% 1.36% 0.63% 2.00% 1.97% 0.85%
value
Sample Respiratory
Influenza Influenza Human Mycoplasma
No. Syncytial Adenovirus
A virus B virus Rhinovirus pneumoniae
virus
1 29.28 30.31 30.85 29.33 30.24 30.50
2 29.43 30.33 30.72 29.41 30.39 30.74
3 29.39 30.11 30.61 29.52 30.34 30.64
4 29.39 30.27 30.83 29.31 30.44 30.76
5 29.37 30.50 30.81 29.36 30.40 30.63
6 29.46 30.24 30.98 29.46 30.38 30.88
7 29.49 30.34 30.81 29.40 30.37 30.83
8 29.27 30.48 30.79 29.46 30.33 30.76
9 29.40 30.21 30.67 29.44 30.46 30.59
10 29.47 30.07 31.04 29.58 30.36 30.83
11 29.73 30.35 30.66 29.56 30.35 30.73
12 29.65 30.48 30.99 29.64 30.19 30.92
13 29.72 30.12 30.68 29.67 30.40 30.75
14 29.70 30.25 30.56 29.69 30.38 30.70
15 29.70 30.48 30.69 29.76 30.41 30.39
16 29.61 30.39 30.75 29.62 30.28 30.38
17 29.61 30.52 30.96 29.61 30.34 30.77
18 29.66 30.47 30.91 29.67 30.52 30.69
19 29.78 30.12 30.69 29.55 30.38 30.76
20 29.59 29.34 30.66 29.52 30.35 30.66
CV
0.54% 0.86% 0.47% 0.44% 0.24% 0.47%
value
Figure 2-26 Inter-day imprecision test results of positive samples for influenza B virus
(2017001 batch)
Figure 2-28 Inter-day imprecision test results of positive samples for Adenovirus
(2017001 batch)
Figure 2-30 Inter-day imprecision test results of positive samples for Mycoplasma
pneumoniae (2017001 batch)
Table 2-18 Ct value of inter-day imprecision test of reagents within the batch of 2017001
Sample Respiratory
Influenza Influenza Human Mycoplasma
No. Syncytial Adenovirus
A virus B virus Rhinovirus pneumoniae
virus
1 29.64 30.20 30.18 29.73 30.47 29.71
2 29.68 30.26 30.21 29.69 30.42 30.21
3 29.61 30.39 30.17 29.81 30.27 30.19
4 29.66 30.42 30.28 29.66 30.25 30.26
Table 2-19 Ct value of inter-day imprecision test of reagents within the batch of 2017002
Sample Respiratory
Influenza Influenza Human Mycoplasma
No. Syncytial Adenovirus
A virus B virus Rhinovirus pneumoniae
virus
1 29.24 30.65 30.75 29.38 30.81 30.62
2 29.23 30.43 30.32 29.34 30.42 30.20
3 29.20 30.77 30.74 29.46 30.70 30.37
4 29.25 30.84 30.87 29.12 30.32 30.24
5 29.21 30.42 30.39 29.43 30.45 30.50
6 29.24 30.45 30.93 29.43 30.51 30.51
7 29.34 30.75 30.46 29.60 30.85 31.19
8 29.29 30.48 30.43 29.61 30.41 31.00
9 29.33 30.73 30.79 29.30 30.74 31.18
10 29.38 30.74 30.60 29.38 30.92 30.56
11 29.27 30.40 30.51 29.19 30.47 30.41
12 29.26 30.60 30.65 29.18 30.42 30.51
13 29.22 30.51 30.59 29.17 30.37 30.35
14 29.22 30.60 30.44 29.22 30.51 30.38
15 29.24 30.48 30.42 29.18 30.39 30.37
16 29.23 30.40 30.65 29.16 30.30 30.52
17 29.25 30.64 30.63 29.22 30.26 30.63
18 29.21 30.38 30.47 29.21 30.24 30.49
19 29.22 30.39 30.54 29.18 30.38 30.41
20 29.18 30.27 30.68 29.16 30.49 30.59
CV
0.17% 0.53% 0.55% 0.51% 0.66% 0.90%
value
Sample Respiratory
Influenza Influenza Human Mycoplasma
No. Syncytial Adenovirus
A virus B virus Rhinovirus pneumoniae
virus
1 29.28 30.31 30.85 29.33 30.24 30.50
2 29.43 30.33 30.72 29.41 30.39 30.74
3 29.39 30.11 30.61 29.52 30.34 30.64
4 29.39 30.27 30.83 29.31 30.44 30.76
5 29.37 30.50 30.81 29.36 30.40 30.63
6 29.46 30.24 30.98 29.46 30.38 30.88
7 29.49 30.34 30.81 29.40 30.37 30.83
8 29.27 30.48 30.79 29.46 30.33 30.76
9 29.40 30.21 30.67 29.44 30.46 30.59
10 29.47 30.07 31.04 29.58 30.36 30.83
11 29.69 30.56 30.80 29.76 30.36 30.76
12 29.68 30.31 30.53 29.75 30.30 30.57
13 29.66 30.47 30.69 29.78 30.37 30.81
14 29.71 30.70 30.49 29.66 30.25 30.61
15 29.74 30.38 30.48 29.79 30.25 30.49
16 29.76 30.38 30.54 29.94 30.21 30.57
17 29.71 30.41 30.85 29.83 30.15 30.57
18 29.85 30.43 30.77 29.97 30.22 30.59
19 29.78 30.28 30.80 29.79 30.18 30.73
20 29.76 30.29 30.73 29.79 30.61 30.66
CV
0.63% 0.49% 0.49% 0.72% 0.36% 0.38%
value
Conclusion: It can be seen from the above results that, the accuracy of the
three batches of kits to detect clinical samples is 100%, the CV of detected Ct
values of inter-day imprecision test among the three batched of kits by
different operators range from 0.17% to 2.00%; the CV of detected Ct values of
intra-day imprecision test among the three batched of kits by the same
operators range from 0.17% to 1.10%, and both of the values are less than 5%.
Therefore, it can be considered that this kit has good detection precision
under different operators and at different times.
Figure 2-32 The test of national precision reference sample CV2 (2017001-2017003 batch)
Table 2-21 Ct value of precision test of the three batches of kits on the national reference
samples CV1 and CV2
Conclusion: It can be seen from the above results that, the accuracy of the
three batches of kits to detect the precision reference samples CV1 and CV2 in
the national reference samples of nucleic acid detection reagents for influenza
A/B virus is 100%, and the CV of detected Ct values of precision between
batches are 0.32% and 0.23%, which are both lower than 5%. It meets the
requirements of national reference sample detection. Therefore, it can be
considered that this kit has good detection precision on national precision
reference samples via SLAN-96P PCR instrument.
Table 1-1 Pathogen information and dilution gradient for testing (Influenza A, Influenza
B)
Diluted detection
Pathogen No. Type
concentration(TCID50/mL)
StrainInf A-A A/H1N1 4.0 2.0 1.5 1.0
StrainInf A-B A/H3 4.0 2.0 1.5 1.0
StrainInf A-C A/H5 4.0 2.0 1.5 1.0
StrainInf B-A B/Victoria 4.0 2.0 1.5 1.0
StrainInf B-B B/Victoria 4.0 2.0 1.5 1.0
StrainInf B-C B/Yamagata 4.0 2.0 1.5 1.0
Figure 1-1 Test results of Strain Inf A-A at a concentration of 2.0 TCID50/mL (2017001
batch)
Figure 1-2 Test results of Strain Inf A-B at a concentration of 2.0 TCID50/mL (2017001
batch)
Figure 1-4 Test results of Strain Inf B-A at a concentration of 2.0 TCID50/mL (2017001
batch)
Figure 1-5 Test results of Strain Inf B-B at a concentration of 2.0 TCID50/mL (2017001
batch)
Figure 1-7 Test results of Strain RSV-A at a concentration of 500 copies/mL (2017001
batch)
Figure 1-8 Test results of Strain RSV-B at a concentration of 500 copies/mL (2017001
batch)
Figure 1-10 Test results of Strain ADV-A at a concentration of 500 copies/mL (2017001
batch)
Figure 1-11 Test results of Strain ADV-B at a concentration of 500 copies/mL (2017001
batch)
Figure 1-13 Test results of Strain ADV-D at a concentration of 500 copies/mL (2017001
batch)
Figure 1-14 Test results of Strain ADV-E at a concentration of 500 copies/mL (2017001
batch)
Figure 1-16 Test results of Strain ADV-G at a concentration of 500 copies/mL (2017001
batch)
Figure 1-18 Test results of Strain ADV-I at a concentration of 500 copies/mL (2017001
batch)
Figure 1-19 Test results of Strain HRV-A at a concentration of 500 copies/mL (2017001
Figure 1-20 Test results of Strain HRV-B at a concentration of 500 copies/mL (2017001
batch)
Figure 1-21 Test results of Strain HRV-C at a concentration of 500 copies/mL (2017001
batch)
Figure 1-22 Test results of Strain MP-A at a concentration of 500 copies/mL (2017001
batch)
Figure 1-24 Test results of Strain MP-C at a concentration of 500 copies/mL (2017001
batch)
Table 1-3 The detection rate of the minimum detection limit determination test for
influenza A virus
Pathogen No. Strain Inf A-A Strain Inf A-B Strain Inf A-C
Detection rate
Concentration Positive Positive Positive Positive Positive Positive
(TCID50/mL) rate percentage rate percentage rate percentage
4.0 60/60 100% 60/60 100% 60/60 100%
2.0 60/60 100% 60/60 100% 60/60 100%
1.5 48/60 80% 50/60 83.3% 48/60 80%
1.0 42/60 70% 36/60 60% 39/60 65%
Table 1-4 The detection rate of the minimum detection limit determination test for
influenza B virus
Pathogen No. Strain Inf B-A Strain Inf B-B Strain Inf B-C
Detection rate
Concentration Positive Positive Positive Positive Positive Positive
(TCID50/mL) rate percentage rate percentage rate percentage
4.0 60/60 100% 60/60 100% 60/60 100%
2.0 60/60 100% 60/60 100% 60/60 100%
Table 1-5 The detection rate of the minimum detection limit determination test for
respiratory syncytial virus
Pathogen No. Strain RSV-A Strain RSV-B Strain RSV-C
Detection rate
Concentration Positive Positive Positive Positive Positive Positive
(copies/mL) rate percentage rate percentage rate percentage
1000 60/60 100% 60/60 100% 60/60 100%
500 60/60 100% 60/60 100% 60/60 100%
400 46/60 76% 47/60 78.3% 44/60 73.3%
300 26/60 43% 33/60 55% 36/60 60%
Table 1-6 The detection rate of the minimum detection limit determination test for
Adenovirus
Pathogen No. Strain ADV-A Strain ADV-B Strain ADV-C
Detection rate
Concentration Positive Positive Positive Positive Positive Positive
(copies/mL) rate percentage rate percentage rate percentage
1000 60/60 100% 60/60 100% 60/60 100%
500 60/60 100% 60/60 100% 60/60 100%
400 48/60 80% 49/60 81.7% 48/60 80%
300 37/60 61.7% 26/60 43.3% 38/60 63.3%
Table 1-7 The detection rate of the minimum detection limit determination test for
Adenovirus
Pathogen No. Strain ADV-D Strain ADV-E Strain ADV-F
Detection rate
Concentration Positive Positive Positive Positive Positive Positive
(copies/mL) rate percentage rate percentage rate percentage
1000 60/60 100% 60/60 100% 60/60 100%
500 60/60 100% 60/60 100% 60/60 100%
400 60/60 100% 60/60 100% 60/60 100%
300 55/60 91.67% 46/60 76.67% 58/60 96.67%
Table 1-8 The detection rate of the minimum detection limit determination test for
Adenovirus
Pathogen No. Strain ADV-G Strain ADV-H Strain ADV-I
Detection rate
Concentration Positive Positive Positive Positive Positive Positive
(copies/mL) rate percentage rate percentage rate percentage
1000 60/60 100% 60/60 100% 60/60 100%
500 60/60 100% 60/60 100% 60/60 100%
400 60/60 100% 60/60 100% 60/60 100%
300 56/60 93.33% 58/60 96.67% 58/60 96.67
Table 1-10 The detection rate of the minimum detection limit determination test for
Human rhinovirus
Pathogen No. Strain HRV-A Strain HRV-B Strain HRV-C
Detection rate
Concentration Positive Positive Positive Positive Positive Positive
(copies/mL) rate percentage rate percentage rate percentage
1000 60/60 100% 60/60 100% 60/60 100%
500 60/60 100% 59/60 98.3% 60/60 100%
400 46/60 76.7% 48/60 80% 48/60 80%
300 36/60 60% 35/60 58.3% 35/60 58.3%
Table 1-11 Sample information for minimum detection limit verification (Influenza A,
Influenza B)
Concentration
Sample No. Type Source
(TCID50/mL)
Sichuan provincial disease
A/Qingyang/1272/201
Sample Inf A-A prevention and control 2100
4(H3)
center
A/Zhejiang/
Yiwu disease prevention and
Sample Inf A-B DTID-ZJU03/2009(H1 6500
control center
N1)
Loudi disease prevention
Sample Inf A-C A/H5 4400
and control center
Sichuan provincial disease
Sample Inf B-A B/Victoria prevention and control 6400
center
Sichuan provincial disease
Sample Inf B-B B/Victoria prevention and control 5500
center
Sichuan provincial disease
Sample Inf B-C B/Yamagata prevention and control 7100
center
Figure 1-25 Test results of sample Inf A-A at a concentration of 2.0 TCID50/mL (2017001
batch)
Figure 1-26 Test results of sample Inf A-B at a concentration of 2.0 TCID50/mL (2017001
batch)
Figure 1-28 Test results of sample Inf B-A at a concentration of 2.0 TCID50/mL (2017001
batch)
Figure 1-29 Test results of sample Inf B-B at a concentration of 2.0 TCID50/mL (2017001
batch)
Figure 1-31 Test results of sample RSV-A at a concentration of 500 copies/mL (2017001
batch)
Figure 1-32 Test results of sample RSV-B at a concentration of 500 copies/mL (2017001
batch)
Figure 1-34 Test results of sample ADV-A at a concentration of 500 copies/mL (2017001
batch)
Figure 1-35 Test results of sample ADV-B at a concentration of 500 copies/mL (2017001
batch)
Figure 1-37 Test results of sample HRV-A at a concentration of 500 copies/mL (2017001
batch)
Figure 1-38 Test results of sample HRV-B at a concentration of 500 copies/mL (2017001
batch)
Figure 1-40 Test results of sample MP-A at a concentration of 500 copies/mL (2017001
batch)
Figure 1-41 Test results of sample MP-B at a concentration of 500 copies/mL (2017001
batch)
Figure 1-44 Test result of the national reference sample S2 at a concentration of 2.0
TCID50/mL (2017001 batch)
Figure 1-46 Test result of the national reference sample S4 at a concentration of 2.0
TCID50/mL (2017001 batch)
Table 1-19 The detected Ct value of the national sensitivity reference samples
Influenza B virus Influenza A virus
Batch No.
S1 S2 S3 S4 S5
1 34.54 34.44 34.00 33.18 32.43
2 34.02 34.16 33.52 32.80 31.84
3 34.01 34.31 33.34 32.51 32.36
4 34.56 34.23 33.47 32.80 29.47
5 34.25 33.92 33.13 32.69 32.80
6 34.21 33.76 32.94 32.43 31.93
7 34.19 34.40 33.15 33.90 31.34
8 34.65 33.80 33.18 32.67 32.61
9 34.26 34.85 34.92 32.12 32.55
10 34.11 33.83 34.84 33.27 32.57
2017001
11 34.28 34.59 34.78 32.52 32.77
12 34.32 33.57 34.97 31.75 32.04
13 34.49 35.21 33.33 32.95 32.63
14 33.94 34.61 34.12 32.69 32.21
15 34.34 35.04 33.76 32.51 31.07
16 34.44 34.45 34.15 32.52 32.29
17 34.20 33.57 33.94 32.48 32.62
18 34.45 36.08 33.61 32.67 32.78
19 33.60 35.55 34.12 32.86 32.91
20 33.98 34.93 33.67 32.36 33.05
21 33.96 35.25 31.93 33.05 31.42
22 33.96 34.92 31.67 31.86 31.47
23 33.59 34.48 31.74 32.17 31.51
2017002
24 33.67 34.07 31.72 32.28 31.49
25 34.00 34.54 31.62 32.11 32.64
26 33.41 33.84 31.47 31.81 32.17
Table 2-1 Pathogen information and dilution gradient for testing (Influenza A, Influenza
Diluted detection
Pathogen No. Type
concentration(TCID50/mL)
StrainInf A-A A/H1N1 4.0 2.0 1.5 1.0
StrainInf A-B A/H3 4.0 2.0 1.5 1.0
StrainInf A-C A/H5 4.0 2.0 1.5 1.0
StrainInf B-A B/Victoria 4.0 2.0 1.5 1.0
StrainInf B-B B/Victoria 4.0 2.0 1.5 1.0
StrainInf B-C B/Yamagata 4.0 2.0 1.5 1.0
B)
Figure 2-1 Test results of Strain Inf A-A at a concentration of 2.0 TCID50/mL (2017001
batch)
Figure 2-2 Test results of Strain Inf A-B at a concentration of 2.0 TCID50/mL (2017001
batch)
Figure 2-4 Test results of strain Inf B-A at a concentration of 2.0 TCID50/mL (2017001
batch)
Figure 2-5 Test results of strain Inf B-B at a concentration of 2.0 TCID50/mL (2017001
batch)
Figure 2-7 Test results of Strain RSV-A at a concentration of 500 copies/mL (2017001
batch)
Figure 2-8 Test results of Strain RSV-B at a concentration of 500 copies/mL (2017001
batch)
Figure 2-10 Test results of Strain ADV-A at a concentration of 500 copies/mL (2017001
batch)
Figure 2-11 Test results of Strain ADV-B at a concentration of 500 copies/mL (2017001
batch)
Figure 2-13 Test results of Strain ADV-D at a concentration of 500 copies/mL (2017001
batch)
Figure 2-14 Test results of Strain ADV-E at a concentration of 500 copies/mL (2017001
batch)
Figure 2-16 Test results of Strain ADV-G at a concentration of 500 copies/mL (2017001
batch)
Figure 2-17 Test results of Strain ADV-H at a concentration of 500 copies/mL (2017001
batch)
Figure 2-19 Test results of Strain HRV-A at a concentration of 500 copies/mL (2017001
batch)
Figure 2-20 Test results of Strain HRV-B at a concentration of 500 copies/mL (2017001
batch)
Figure 2-22 Test results of Strain MP-A at a concentration of 500 copies/mL (2017001
batch)
Figure 2-23 Test results of Strain MP-B at a concentration of 500 copies/mL (2017001
batch)
Table 2-3 The detection rate of the minimum detection limit determination test for
influenza A virus
Pathogen No. Strain Inf A-A Strain Inf A-B Strain Inf A-C
Detection rate
Concentration Positive Positive Positive Positive Positive Positive
(TCID50/mL) rate percentage rate percentage rate percentage
4.0 60/60 100% 60/60 100% 60/60 100%
2.0 60/60 100% 60/60 100% 60/60 100%
1.5 47/60 78.3% 49/60 81.7% 46/60 76.7%
1.0 41/60 68.3% 42/60 70% 35/60 58.3%
Table 2-4 The detection rate of the minimum detection limit determination test for
influenza B virus
Pathogen No. Strain Inf B-A Strain Inf B-B Strain Inf B-C
Detection rate
Concentration Positive Positive Positive Positive Positive Positive
(TCID50/mL) rate percentage rate percentage rate percentage
4.0 60/60 100% 60/60 100% 60/60 100%
2.0 60/60 100% 59/60 98.3% 60/60 100%
1.5 48/60 80% 48/60 80% 48/60 80%
1.0 42/60 70% 19/60 31.7% 43/60 71.7%
Table 2-5 The detection rate of the minimum detection limit determination test for
respiratory syncytial virus
Pathogen No. Strain RSV-A Strain RSV-B Strain RSV-C
Detection rate
Concentration Positive Positive Positive Positive Positive Positive
(TCID50/mL) rate percentage rate percentage rate percentage
1000 60/60 100% 60/60 100% 60/60 100%
500 60/60 100% 58/60 96.7% 60/60 100%
400 46/60 76.7% 49/60 81.7% 48/60 80%
300 24/60 40% 30/60 50% 28/60 46.6%
Table 2-7 The detection rate of the minimum detection limit determination test for
Adenovirus
Pathogen No. Strain ADV-D Strain ADV-E Strain ADV-F
Detection rate
Concentration Positive Positive Positive Positive Positive Positive
(copies/mL) rate percentage rate percentage rate percentage
1000 60/60 100% 60/60 100% 60/60 100%
500 60/60 100% 60/60 100% 60/60 100%
400 60/60 100% 60/60 100% 60/60 100%
300 55/60 91.67% 42/60 70.00% 59/60 98.33%
Table 2-8 The detection rate of the minimum detection limit determination test for
Adenovirus
Pathogen No. Strain ADV-G Strain ADV-H Strain ADV-I
Detection rate
Concentration Positive Positive Positive Positive Positive Positive
(copies/mL) rate percentage rate percentage rate percentage
1000 60/60 100% 60/60 100% 60/60 100%
500 60/60 100% 60/60 100% 60/60 100%
400 60/60 100% 60/60 100% 60/60 100%
300 60/60 100% 60/60 100% 60/60 100%
Table 2-9 The detection rate of the minimum detection limit determination test for
Mycoplasma pneumoniae
Pathogen No. Strain MP-A Strain MP-B Strain MP-C
Detection rate
Concentration Positive Positive Positive Positive Positive Positive
(copies/mL) rate percentage rate percentage rate percentage
1000 60/60 100% 60/60 100% 60/60 100%
500 60/60 100% 60/60 100% 60/60 100%
400 48/60 80% 48/60 80% 49/60 81.7%
300 37/60 61.7% 36/60 60% 37/60 61.7
Table 2-10 The detection rate of the minimum detection limit determination test for
Human rhinovirus
Table 2-11 Sample information for minimum detection limit verification (Influenza A,
Influenza B)
Figure 2-25 Test results of sample Inf A-A at a concentration of 2.0 TCID50/mL (2017001
batch)
Figure 2-26 Test results of sample Inf A-B at a concentration of 2.0 TCID50/mL (2017001
batch)
Figure 2-27 Test results of sample Inf A-C at a concentration of 2.0 TCID50/mL (2017001
batch)
Figure 2-29 Test results of sample Inf B-B at a concentration of 2.0 TCID50/mL (2017001
batch)
Figure 2-30 Test results of sample Inf B-C at a concentration of 2.0 TCID50/mL (2017001
batch)
Figure 2-32 Test results of sample RSV-B at a concentration of 500 copies/mL (2017001
batch)
Figure 2-33 Test results of sample RSV-C at a concentration of 500 copies/mL (2017001
batch)
Figure 2-35 Test results of sample ADV-B at a concentration of 500 copies/mL (2017001
batch)
Figure 2-36 Test results of sample ADV-C at a concentration of 500 copies/mL (2017001
batch)
Figure 2-38 Test results of sample HRV-B at a concentration of 500 copies/mL (2017001
batch)
Figure 2-39 Test results of sample HRV-C at a concentration of 500 copies/mL (2017001
batch)
Figure 2-41 Test results of sample MP-B at a concentration of 500 copies/mL (2017001
batch)
Figure 2-42 Test results of sample MP-C at a concentration of 500 copies/mL (2017001
batch)
Figure 2-43 Test result of the national reference sample S1 at a concentration of 2.0
TCID50/mL (2017001 batch)
Figure 2-44 Test result of the national reference sample S2 at a concentration of 2.0
TCID50/mL (2017001 batch)
Figure 2-46 Test result of the national reference sample S4 at a concentration of 2.0
TCID50/mL (2017001 batch)
Figure 2-47 Test result of the national reference sample S5 at a concentration of 2.0
TCID50/mL (2017001 batch)
Table 2-19 The detected Ct value of the national sensitivity reference samples
1 Analytical specificity
1.1 Verification of clinical samples
Negative reference samples N1 (measles virus), N2 (mumps virus), N3
(rubella virus), N4 (Staphylococcus aureus), N5 (Escherichia coli), N6
(Pseudomonas aeruginosa), N7 (human parainfluenza virus type I), N8
(normal throat swab)and N9 (normal throat swab) required in the product
technical requirements of the six respiratory pathogens nucleic acid
diagnostic kit (PCR-fluorescence probing), and other pathogens that have
homology in their nucleic acid sequence and are likely to cause the same or
similar clinical symptoms, including: cytomegalovirus, coxsackie virus type A,
human metapneumovirus, bacillus pertussis, chlamydia pneumoniae,
haemophilus influenzae, saliva Streptococcus, Streptococcus pneumoniae,
Neisseria meningitidis, Mycobacterium tuberculosis, human parainfluenza
type 2 and type 3 virus, Epstein-Barr virus, Cryptococcus, Aspergillus
fumigatus, Aspergillus flavus, Candida albicans, Legionella pneumophila and
Intestine Daovirus type 71 were used as the test samples. Three qualified kits
were used to test the samples on the same ABI7500 instrument. And the
specificity of the kit was tested by detecting the negative and positive
coincidence rate. Test results are as follows (only the amplification curves of
the 2017001 batch of reagents are shown):
Figure 1-2 HEX channel amplification curve of samples for analytical specificity (2017001
batch of reagents)
Figure 1-3 CY5 channel amplification curve of samples for analytical specificity (2017001
batch of reagents)
Table 1-1 Test results of variable samples by using three batches of reagents
Batch&Result 2017001 batch 2017002 batch 2017003 batch
Sample FAM HEX CY5 FAM HEX CY5 FAM HEX CY5
Positive control 28.43 29.83 29.80 28.47 28.59 30.14 28.50 29.83 29.98
(Mix A, B) 28.68 30.24 30.08 29.03 28.72 30.73 28.77 30.24 29.85
N1 No Ct No Ct No Ct No Ct No Ct No Ct No Ct No Ct No Ct
N2 No Ct No Ct No Ct No Ct No Ct No Ct No Ct No Ct No Ct
N3 No Ct No Ct No Ct No Ct No Ct No Ct No Ct No Ct No Ct
N4 No Ct No Ct No Ct No Ct No Ct No Ct No Ct No Ct No Ct
N5 No Ct No Ct No Ct No Ct No Ct No Ct No Ct No Ct No Ct
N6 No Ct No Ct No Ct No Ct No Ct No Ct No Ct No Ct No Ct
N7 No Ct No Ct No Ct No Ct No Ct No Ct No Ct No Ct No Ct
N8 No Ct No Ct No Ct No Ct No Ct No Ct No Ct No Ct No Ct
N9 No Ct No Ct No Ct No Ct No Ct No Ct No Ct No Ct No Ct
Cytomegaloviru No Ct No Ct No Ct No Ct No Ct No Ct No Ct No Ct No Ct
s
Conclusion: It can be seen from Figure 1-1 to Figure 1-3 and Table 1-1 that,
the positive rate of negative reference samples N1-N8 and other pathogens
that have homology in nucleic acid sequence and are likely to cause the same
or similar clinical symptoms detected by using this detection kit developed by
our company is all 0%. Therefore, this kit has good specificity, and no clinical
cross-reaction occurs.
Figure 1-4 FAM channel amplification curve of national reference samples NC01-NC06
(2017001 batch of reagents)
Figure 1-5 HEX channel amplification curve of national reference samples NC01-NC06
(2017001 batch of reagents)
Table 1-2 Test results of variable samples by using three batches of reagents
Batch&Result 2017001 batch 2017002 batch 2017003 batch
Sample FAM HEX CY5 FAM HEX CY5 FAM HEX CY5
Positive control 32.71 32.92 33.32 33.18 33.04 33.71 33.21 33.10 33.93
(Mix A, B) 32.70 32.83 33.06 32.27 32.91 33.72 31.92 32.87 33.59
NC01 No Ct No Ct No Ct No Ct No Ct No Ct No Ct No Ct No Ct
NC02 No Ct No Ct No Ct No Ct No Ct No Ct No Ct No Ct No Ct
NC03 No Ct No Ct No Ct No Ct No Ct No Ct No Ct No Ct No Ct
NC04 No Ct No Ct No Ct No Ct No Ct No Ct No Ct No Ct No Ct
NC05 No Ct No Ct No Ct No Ct No Ct No Ct No Ct No Ct No Ct
NC06 No Ct No Ct No Ct No Ct No Ct No Ct No Ct No Ct No Ct
Oxymetazoline
F1 100 μg/mL F9 SDS 100 μg/mL
Hydrochloride
Cefmenoxime
F3 50 μg/mL F11 Urea 100 μg/mL
Hydrochloride
F4 Menthol 50 μg/mL F12 Heme 10 μg/mL
Purified
F5 Zanamivir 100 μg/mL F13 20 μg/mL
mucin
F6 Ribavirin 100 μg/mL F14 FeCl3 100 μg/mL
Absolute
F7 Azithromycin 100μg/mL F15 20%(v/v)
ethanol
Human
F8 NaCl 60 μg/mL F16 20%(v/v)
whole blood
Figure 1-7 Test curve of the effect of internal and external interfering substances on
influenza A virus (2017001 batch)
Figure 1-8 Test curve of the effect of internal and external interfering substances on
Figure 1-9 Test curve of the effect of internal and external interfering substances on
respiratory syncytial virus (2017001 batch)
Figure 1-10 Test curve of the effect of internal and external interfering substances on
Adenovirus (2017001 batch)
Figure 1-11 Test curve of the effect of internal and external interfering substances on
Human rhinovirus (2017001 batch)
Table 1-4 Ct value of the tested effects of anti-internal and external interfering
substances (2017001 batch)
PCR mix A PCR mix B
FAM Ct HEX Ct CY5 Ct FAM Ct HEX Ct CY5 Ct
value value value value value value
Cont 30.7 30.7 31.0 30.8 31.2 31.4 31.4 30.9 31.2 31.1 31.6 31.6
rol 6 9 3 0 3 0 2 8 3 0 2 1
30.7 30.8 30.9 30.9 30.8 31.9 31.0 31.3 31.5 31.2 31.4 31.5
F1
4 5 6 8 5 6 3 8 8 4 9 4
30.8 30.9 31.1 31.0 31.6 32.2 31.4 31.4 31.4 31.4 31.4 31.6
F2
9 3 0 8 2 6 6 5 8 4 5 2
31.0 30.7 31.3 30.6 32.0 31.6 31.7 31.5 31.4 31.2 32.2 31.9
F3
8 2 5 4 5 6 4 0 4 2 3 0
31.2 31.1 31.3 31.1 31.7 31.7 31.7 31.8 31.3 31.2 32.0 31.7
F4
1 8 1 8 5 6 0 2 4 5 0 1
31.1 31.2 31.4 31.0 31.6 31.7 31.6 31.5 31.5 31.3 31.9 31.4
F5
1 6 1 6 0 6 9 9 8 9 4 7
30.7 31.3 31.5 31.3 31.5 31.3 31.8 31.5 31.5 31.3 31.7 31.5
F6
3 0 5 6 2 0 2 4 0 3 9 6
31.1 31.1 31.0 31.0 31.7 30.8 32.0 31.5 31.3 31.2 31.7 32.0
F7
2 8 4 6 1 2 0 8 6 6 3 0
30.8 30.9 30.5 30.7 31.7 31.4 31.3 31.4 31.0 30.6 31.3 31.6
F8
9 1 7 7 6 8 5 9 8 7 3 2
31.4 31.0 31.0 30.8 31.1 31.8 31.4 31.8 31.1 31.0 31.5 31.8
F9
0 1 7 6 7 5 2 3 8 6 1 9
30.9 31.0 31.1 31.2 31.4 31.8 31.4 31.3 31.3 31.1 31.7 31.4
F10
8 1 6 1 0 4 0 6 0 7 0 8
31.2 31.0 31.3 31.4 31.2 31.8 31.3 31.4 31.1 30.9 31.4 31.3
F11
4 3 1 0 7 1 2 9 7 7 0 0
31.1 30.7 31.2 31.1 30.7 31.2 31.8 31.6 31.2 31.1 31.8 31.4
F12
1 0 8 5 5 2 2 5 4 8 8 0
30.9 31.2 30.6 31.4 31.4 31.0 31.6 31.5 31.2 31.1 31.6 31.9
F13
3 9 7 0 3 7 4 3 2 2 4 0
Table 1-5 Ct value of the tested effects of anti-internal and external interfering
substances (2017002 batch)
PCR mix A PCR mix B
FAM Ct HEX Ct CY5 Ct FAM Ct HEX Ct CY5 Ct
value value value value value value
Cont 31.0 30.6 31.1 31.1 31.0 31.2 31.0 30.9 31.2 30.2 31.1 31.3
rol 4 4 0 9 6 5 3 9 7 6 6 8
30.3 31.0 31.2 31.5 31.2 31.5 31.1 31.0 31.2 31.2 30.9 31.2
F1
5 2 0 7 5 7 5 0 0 1 4 2
30.6 30.7 31.2 31.4 31.5 31.6 31.0 31.2 31.2 31.2 31.2 31.2
F2
5 6 9 3 6 1 8 4 5 9 7 4
30.8 30.5 31.0 31.1 31.7 31.2 31.5 31.4 31.4 31.2 31.8 31.5
F3
7 0 6 4 8 9 1 3 9 8 5 8
30.9 30.9 30.9 31.3 30.4 31.7 31.2 31.4 31.5 31.1 31.4 31.6
F4
1 9 3 5 1 4 5 3 8 5 3 5
30.7 30.6 31.1 31.4 31.6 31.6 31.3 31.2 31.4 31.6 31.3 30.9
F5
0 5 5 2 9 8 9 9 2 1 1 6
30.6 30.7 31.6 31.8 31.5 31.1 31.3 31.4 31.5 31.1 31.5 31.1
F6
2 9 6 8 5 6 0 2 0 7 6 3
30.7 30.9 31.8 31.6 31.9 31.8 31.3 31.3 31.2 31.4 31.9 31.4
F7
6 9 0 8 3 1 9 9 2 1 0 5
30.3 30.4 31.4 31.3 31.2 32.1 31.0 31.2 31.0 31.0 32.0 31.7
F8
5 4 8 6 9 3 5 5 7 6 1 1
31.0 30.6 31.3 31.5 31.2 31.5 31.1 31.1 31.1 31.0 31.7 31.6
F9
2 7 3 6 3 7 5 1 3 7 5 5
31.0 30.9 31.3 31.5 31.9 31.6 31.1 31.2 31.2 31.2 31.5 31.6
F10
8 0 4 0 1 0 7 9 7 1 2 2
30.8 30.9 30.6 31.1 31.4 32.0 31.1 31.4 31.2 31.3 32.0 31.3
F11
6 6 4 8 2 7 3 1 0 2 4 5
31.0 30.6 30.6 31.0 31.4 31.6 31.3 31.3 31.4 31.3 31.8 31.4
F12
3 8 2 1 4 5 3 7 1 7 2 7
30.9 30.9 31.3 31.3 31.1 31.6 31.2 31.6 31.2 31.5 31.4 31.3
F13
9 5 8 4 6 6 3 8 4 6 5 6
30.3 30.5 31.6 32.0 31.6 31.9 31.3 31.1 31.3 31.3 31.2 31.4
F14
0 8 4 2 7 1 5 7 8 0 8 3
30.3 30.7 31.2 32.0 31.7 32.4 31.1 31.2 31.3 31.5 31.1 31.3
F15
8 0 0 1 7 0 4 6 7 9 6 1
30.7 30.6 31.3 31.6 31.4 31.1 30.9 31.1 30.7 31.1 31.6 32.0
F16
0 2 0 8 7 6 9 1 8 2 0 0
Table 1-6 Ct value of the tested effects of anti-internal and external interfering
Conclusion: It can be seen from the test results of the above mentioned
three batched of reagents that, on the ABI7500 instrument, the test results of
weak positive samples containing the above-mentioned internal and external
interfering substances are not significantly different from those of the control
group (weak positive samples that do not contain interfering substances).
Therefore, it can be shown that this kit has a good anti-interference ability
against these internal and external interfering substances.
Figure 1-13 Verified curve of the effect of internal and external interfering substances on
influenza A virus (2017001 batch)
Figure 1-14 Verified curve of the effect of internal and external interfering substances on
influenza B virus (2017001 batch)
Figure 1-16 Verified curve of the effect of internal and external interfering substances on
Adenovirus (2017001 batch)
Figure 1-17 Verified curve of the effect of internal and external interfering substances on
Human rhinovirus (2017001 batch)
Figure 1-18 Verified curve of the effect of internal and external interfering substances on
Mycoplasma pneumoniae (2017001 batch)
Table 1-7 Ct value of the verified effects of internal and external interfering
substances (2017001 batch)
Table 1-8 Ct value of the verified effects of internal and external interfering
substances (2017002 batch)
Mycoplasm
Respiratory
Influenza A Influenza B Respiratory a Human
syncytial
virus virus Adenovirus pneumonia rhinovirus
virus
e
Test Test Test Test Test Test Test Test Test Test Test Test
1 2 1 2 1 2 1 2 1 2 1 2
Contr
26.46 26.53 29.70 29.38 27.14 26.89 29.73 29.92 31.50 32.07 30.81 30.11
ol
F1 26.66 27.13 29.61 29.65 27.20 27.45 30.09 30.02 31.53 31.67 30.18 30.15
F2 27.19 27.65 29.67 29.91 27.39 27.33 30.26 30.38 31.80 31.96 30.54 30.53
F3 27.41 27.54 29.31 29.29 27.57 27.60 30.54 30.41 31.57 31.13 30.44 30.33
F4 27.54 27.67 29.59 29.25 27.65 27.72 30.16 31.02 31.85 31.91 30.62 30.30
F5 27.44 27.63 29.51 29.61 27.93 27.65 30.27 30.27 31.57 31.75 29.85 30.63
F6 27.31 27.15 29.16 29.51 27.67 27.50 30.29 30.25 31.37 31.70 30.37 30.23
F7 27.32 27.53 28.98 29.43 27.49 27.65 30.04 30.30 31.60 31.38 30.31 30.69
F8 27.34 27.40 29.44 29.07 27.20 27.03 29.88 30.19 31.79 31.94 30.08 30.11
F9 27.26 27.48 29.73 29.47 27.53 27.23 30.19 30.24 32.35 31.43 30.00 30.37
F10 27.41 27.62 29.61 29.39 27.31 27.30 30.39 30.27 31.35 31.22 30.38 30.57
F11 27.35 27.40 29.30 28.75 27.67 27.34 30.54 30.56 31.66 32.08 30.30 30.56
F12 27.59 27.42 29.40 29.63 27.43 27.59 30.53 30.80 32.12 31.07 30.30 30.64
Table 1-9 Ct value of the verified effects of internal and external interfering
substances (2017003 batch)
Mycoplasm
Respiratory
Influenza A Influenza B Respiratory a Human
syncytial
virus virus Adenovirus pneumonia rhinovirus
virus
e
Test Test Test Test Test Test Test Test Test Test Test Test
1 2 1 2 1 2 1 2 1 2 1 2
Contr
26.79 26.78 29.17 29.22 26.90 27.20 29.95 30.11 31.19 30.66 31.10 30.15
ol
F1 26.85 27.08 29.39 29.41 27.16 27.35 30.48 30.31 30.48 30.73 30.41 30.64
F2 27.24 27.23 28.35 29.58 27.27 27.62 30.26 30.47 30.24 30.62 30.46 30.43
F3 27.32 27.24 29.61 29.16 27.52 27.67 30.80 30.80 30.61 30.31 30.77 30.67
F4 27.64 27.22 29.52 29.36 27.67 27.86 30.58 30.78 30.94 30.76 30.36 31.50
F5 27.29 27.36 29.46 29.83 27.91 27.83 30.67 30.60 30.84 30.35 30.51 30.41
F6 27.57 27.17 29.50 29.47 27.47 27.41 30.39 30.67 31.25 31.03 29.66 30.22
F7 26.98 27.35 29.27 30.02 27.51 27.80 30.34 30.53 30.70 30.87 30.43 30.26
F8 27.00 27.14 28.94 29.49 27.65 27.63 30.29 30.10 30.30 30.29 30.24 30.22
F9 27.23 27.17 29.23 29.39 27.12 27.76 30.58 30.63 30.56 30.85 30.14 30.32
F10 27.56 27.10 29.47 29.41 27.53 27.75 30.49 30.72 30.90 30.32 30.13 30.45
F11 27.13 27.30 30.01 29.47 27.66 27.69 30.69 30.84 30.79 30.26 30.53 30.53
F12 27.63 27.48 29.78 29.59 27.81 27.33 30.81 30.89 31.16 30.84 30.39 30.49
F13 27.18 27.57 29.48 29.83 27.82 27.56 30.56 31.00 30.86 30.35 30.29 30.61
F14 27.60 27.41 29.29 29.61 27.68 27.40 30.62 30.55 30.45 30.49 29.73 30.27
F15 27.29 27.24 29.25 29.78 27.48 27.26 30.19 30.46 30.50 30.65 30.25 30.28
F16 27.09 27.06 29.10 29.44 27.60 27.04 30.09 30.37 31.10 29.41 30.06 30.22
Conclusion: It can be seen from the test results of the above mentioned
three batched of reagents that, on the ABI7500 instrument, the test results of
weak positive samples containing the above-mentioned internal and external
interfering substances are not significantly different from those of the control
group (weak positive samples that do not contain interfering substances).
Therefore, it can be shown that, under the experimental conditions of this kit,
these internal and external substances below the above-mentioned
concentration have no effect on the test results of the kit.
1 Analytical specificity
1.1 Verification of clinical samples
Negative reference samples N1 (measles virus), N2 (mumps virus), N3
(rubella virus), N4 (Staphylococcus aureus), N5 (Escherichia coli), N6
(Pseudomonas aeruginosa), N7 (human parainfluenza virus type I), N8
(normal throat swab)and N9 (normal throat swab) required in the product
technical requirements of the six respiratory pathogens nucleic acid
diagnostic kit (PCR-fluorescence probing), and other pathogens that have
homology in their nucleic acid sequence and are likely to cause the same or
similar clinical symptoms, including: cytomegalovirus, coxsackie virus type A,
human metapneumovirus, bacillus pertussis, chlamydia pneumoniae,
haemophilus influenzae, saliva Streptococcus, Streptococcus pneumoniae,
Neisseria meningitidis, Mycobacterium tuberculosis, human parainfluenza
type 2 and type 3 virus, Epstein-Barr virus, Cryptococcus, Aspergillus
fumigatus, Aspergillus flavus, Candida albicans, Legionella pneumophila and
Intestine Daovirus type 71 were used as the test samples. Three qualified kits
were used to test the samples on the same SLAN-96P instrument. And the
specificity of the kit was tested by detecting the negative and positive
coincidence rate. Test results are as follows (only the amplification curves of
the 2017001 batch of reagents are shown):
Figure 2-1 FAM channel amplification curve of samples for analytical specificity (2017001
batch of reagents)
Figure 2-3 CY5 channel amplification curve of samples for analytical specificity (2017001
batch of reagents)
Table 2-1 Test results of variable samples by using three batches of reagents
Batch&Result 2017001 batch 2017002 batch 2017003 batch
Sample FAM HEX CY5 FAM HEX CY5 FAM HEX CY5
Positive control 28.51 30.13 31.17 30.79 30.74 31.23 28.12 31.10 31.02
(Mix A, B) 28.37 30.07 30.19 31.08 30.63 31.77 28.02 30.79 30.47
N1 No Ct No Ct No Ct No Ct No Ct No Ct No Ct No Ct No Ct
N2 No Ct No Ct No Ct No Ct No Ct No Ct No Ct No Ct No Ct
N3 No Ct No Ct No Ct No Ct No Ct No Ct No Ct No Ct No Ct
N4 No Ct No Ct No Ct No Ct No Ct No Ct No Ct No Ct No Ct
N5 No Ct No Ct No Ct No Ct No Ct No Ct No Ct No Ct No Ct
N6 No Ct No Ct No Ct No Ct No Ct No Ct No Ct No Ct No Ct
N7 No Ct No Ct No Ct No Ct No Ct No Ct No Ct No Ct No Ct
N8 No Ct No Ct No Ct No Ct No Ct No Ct No Ct No Ct No Ct
N9 No Ct No Ct No Ct No Ct No Ct No Ct No Ct No Ct No Ct
Cytomegaloviru No Ct No Ct No Ct No Ct No Ct No Ct No Ct No Ct No Ct
s
Coxsackie virus No Ct No Ct No Ct No Ct No Ct No Ct No Ct No Ct No Ct
type A
Human No Ct No Ct No Ct No Ct No Ct No Ct No Ct No Ct No Ct
metapneumovir
Figure 2-4 FAM channel amplification curve of national reference samples NC01-NC06
(2017001 batch of reagents)
Figure 2-5 HEX channel amplification curve of national reference samples NC01-NC06
(2017001 batch of reagents)
Table 2-2 Test results of variable samples by using three batches of reagents
Batch&Result 2017001 batch 2017002 batch 2017003 batch
Sample FAM HEX CY5 FAM HEX CY5 FAM HEX CY5
Positive control 33.83 33.89 34.60 34.91 34.22 35.19 33.83 33.57 34.75
(Mix A, B) 34.64 34.31 34.77 34.50 34.17 34.90 34.91 33.88 35.34
NC01 No Ct No Ct No Ct No Ct No Ct No Ct No Ct No Ct No Ct
NC02 No Ct No Ct No Ct No Ct No Ct No Ct No Ct No Ct No Ct
NC03 No Ct No Ct No Ct No Ct No Ct No Ct No Ct No Ct No Ct
NC04 No Ct No Ct No Ct No Ct No Ct No Ct No Ct No Ct No Ct
NC05 No Ct No Ct No Ct No Ct No Ct No Ct No Ct No Ct No Ct
NC06 No Ct No Ct No Ct No Ct No Ct No Ct No Ct No Ct No Ct
Oxymetazoline
F1 100 μg/mL F9 SDS 100 μg/mL
Hydrochloride
F2 Dexamethasone 50 μg/mL F10 EDTA 10 μg/mL
Cefmenoxime
F3 50 μg/mL F11 Urea 100 μg/mL
Hydrochloride
F4 Menthol 50 μg/mL F12 Heme 10 μg/mL
F5 Zanamivir 100 μg/mL F13 Purified mucin 20 μg/mL
F6 Ribavirin 100 μg/mL F14 FeCl3 100 μg/mL
Oxymetazoline
F7 100 μg/mL F15 SDS 20%(v/v)
Hydrochloride
F8 Dexamethasone 60 μg/mL F16 EDTA 20%(v/v)
Figure 2-7 Test curve of the effect of internal and external interfering substances on
influenza A virus (2017001 batch)
Figure 2-8 Test curve of the effect of internal and external interfering substances on
influenza B virus (2017001 batch)
Figure 2-10 Test curve of the effect of internal and external interfering substances on
Adenovirus (2017001 batch)
Figure 2-11 Test curve of the effect of internal and external interfering substances on
Human rhinovirus (2017001 batch)
Table 2-4 Ct value of the tested effects of anti-internal and external interfering substances
(2017001 batch)
PCR mix A PCR mix B
FAM Ct HEX Ct CY5 Ct FAM Ct HEX Ct CY5 Ct
value value value value value value
Cont 29.9 30.2 31.5 31.7 31.3 31.0 29.0 29.0 30.5 30.8 31.6 31.1
rol 6 4 7 1 3 7 4 2 0 1 6 5
30.0 30.2 31.6 31.8 31.2 31.6 28.9 28.8 30.7 30.6 31.4 31.6
F1
6 5 9 0 1 0 6 2 6 8 7 0
30.0 30.0 31.6 31.6 31.3 31.3 28.9 28.8 33.6 30.7 31.4 31.3
F2
9 8 3 0 6 9 5 8 6 9 5 9
30.1 30.0 31.2 31.4 31.1 31.2 28.9 28.7 30.7 30.6 31.8 31.2
F3
5 9 4 9 2 3 1 6 5 2 1 8
30.3 30.2 31.5 31.7 31.4 31.1 28.9 28.7 30.9 30.8 31.3 31.5
F4
3 8 1 0 5 5 0 4 7 2 6 9
30.3 30.1 31.4 31.6 30.8 31.2 28.9 28.8 30.8 30.9 31.5 31.7
F5
8 8 0 7 7 9 7 0 2 7 4 1
30.2 30.1 31.2 31.4 30.8 31.0 28.9 28.9 30.7 30.5 31.1 31.7
F6
9 5 0 7 2 8 1 1 5 5 3 4
30.2 30.1 31.3 31.5 31.0 30.0 29.0 28.0 30.7 30.6 31.1 31.1
F7
0 0 9 2 7 8 2 5 9 9 6 8
30.1 30.1 31.2 31.5 31.4 31.3 28.9 28.9 30.8 30.8 31.4 31.3
F8
0 9 8 1 9 0 8 1 8 1 1 4
30.0 30.3 31.6 31.3 31.2 31.5 28.4 28.6 30.6 30.7 31.3 31.4
F9
3 0 1 0 9 7 5 7 1 8 5 2
30.2 30.1 31.6 31.5 31.2 31.5 29.0 28.8 30.8 30.8 31.5 31.5
F10
7 8 6 8 9 1 0 7 1 3 4 4
30.3 30.2 31.4 31.6 31.3 31.3 28.9 28.9 30.6 30.8 31.5 31.4
F11
4 8 4 3 3 5 3 8 5 1 8 8
30.4 30.4 31.3 31.5 31.7 31.4 28.9 28.5 30.8 30.7 31.5 31.4
F12
0 3 1 5 0 9 9 4 3 5 7 6
30.3 30.4 31.6 31.6 30.8 31.0 28.8 28.7 30.7 30.7 31.4 31.5
F13
1 5 1 8 8 1 0 8 7 6 3 0
Table 2-5 Ct value of the tested effects of anti-internal and external interfering
substances (2017002 batch)
PCR mix A PCR mix B
FAM Ct HEX Ct CY5 Ct FAM Ct HEX Ct CY5 Ct
value value value value value value
Cont 28.9 29.1 31.0 30.3 30.6 30.9 28.0 28.0 31.0 31.2 31.8 31.6
rol 1 9 6 7 0 5 8 5 4 2 1 6
29.0 29.1 30.9 31.0 30.2 31.2 28.0 28.0 31.1 31.1 32.0 31.7
F1
0 0 7 4 4 1 7 5 7 3 8 1
28.7 28.8 31.1 31.2 30.6 31.2 28.0 28.0 31.2 31.0 32.3 32.6
F2
0 7 4 2 2 1 4 6 8 5 7 2
28.8 28.8 31.0 31.0 30.9 31.2 28.0 28.1 31.1 31.1 32.1 31.9
F3
9 8 6 3 2 4 4 3 8 5 3 7
29.0 28.8 30.9 30.8 31.1 31.3 28.1 28.0 30.9 31.2 32.1 31.8
F4
3 5 9 4 6 2 6 9 6 1 5 5
29.0 28.8 30.9 30.9 30.7 31.3 28.1 28.0 30.9 31.2 31.9 31.9
F5
0 3 5 5 5 1 7 8 5 0 6 9
28.8 29.0 30.8 30.8 30.8 31.4 28.1 28.1 30.9 31.2 32.2 32.3
F6
8 0 4 9 8 3 2 1 6 0 1 5
28.9 28.9 31.0 31.2 30.9 31.5 28.0 28.3 30.9 31.0 32.0 31.9
F7
0 5 5 6 5 1 4 0 6 7 3 7
28.9 28.8 30.8 30.9 31.0 31.1 27.9 28.0 31.0 31.1 32.0 31.6
F8
8 7 3 5 2 3 3 0 0 6 6 4
28.8 28.8 31.0 30.9 31.3 31.2 28.0 27.9 31.1 31.0 32.2 32.2
F9
8 4 0 6 0 3 8 8 0 3 7 0
28.9 28.8 31.0 30.7 31.3 31.3 28.0 27.9 31.1 30.3 32.1 31.9
F10
0 7 5 8 0 9 0 9 6 2 3 1
28.7 28.9 31.0 31.0 31.0 31.0 28.1 28.0 31.1 31.1 32.1 31.9
F11
9 4 6 4 3 6 1 1 3 9 6 6
29.0 28.8 31.0 31.2 31.1 31.0 28.0 28.0 30.9 31.0 32.1 32.3
F12
3 8 9 8 7 5 6 4 2 6 9 7
28.8 28.8 31.0 31.1 30.8 31.0 28.1 28.0 31.0 31.1 31.7 31.8
F13
0 5 1 2 7 9 3 2 2 3 6 5
29.0 28.8 31.0 31.0 30.8 31.3 28.0 28.0 31.1 31.0 31.9 31.7
F14
6 0 7 5 8 7 2 2 7 2 8 5
28.9 28.8 31.1 31.2 30.7 31.6 27.9 28.0 31.0 31.1 32.1 32.1
F15
3 8 0 3 3 5 4 4 3 8 9 8
28.9 28.9 30.8 31.0 31.2 31.1 28.0 28.1 31.2 31.0 32.2 32.1
F16
3 7 9 8 4 4 0 7 0 5 6 4
Conclusion: It can be seen from the test results of the above mentioned
three batched of reagents that, on the SLAN-96P instrument, the test results of
weak positive samples containing the above-mentioned internal and external
interfering substances are not significantly different from those of the control
group (weak positive samples that do not contain interfering substances).
Therefore, it can be shown that this kit has a good anti-interference ability
against these internal and external interfering substances.
Figure 2-13 Verified curve of the effect of internal and external interfering substances on
influenza A virus (2017001 batch)
Figure 2-14 Verified curve of the effect of internal and external interfering substances on
influenza B virus (2017001 batch)
Figure 2-16 Verified curve of the effect of internal and external interfering substances on
Adenovirus (2017001 batch)
Figure 1-17 Verified curve of the effect of internal and external interfering substances on
Mycoplasma pneumoniae (2017001 batch)
Contr
28.13 28.28 31.08 31.11 29.87 30.24 29.96 29.52 31.65 32.24 30.02 30.10
ol
F1 28.14 28.15 31.02 31.13 30.30 30.27 30.11 29.88 31.62 32.42 30.17 30.18
F2 28.26 28.24 30.89 31.15 30.17 30.10 30.10 30.09 31.94 32.64 30.36 30.18
F3 28.59 25.54 30.93 31.03 30.33 30.11 29.99 30.14 31.89 32.04 30.11 30.13
F4 28.40 28.28 31.07 30.89 30.67 30.83 29.91 30.05 32.08 32.23 30.21 30.11
F5 28.34 28.38 30.96 31.10 30.15 30.03 29.98 29.89 32.02 32.16 30.00 29.93
F6 28.18 28.32 30.87 31.07 30.37 30.14 29.94 30.14 32.63 31.61 30.12 30.14
F7 28.16 28.21 31.13 31.41 30.36 30.10 30.05 29.90 31.94 31.93 30.01 29.89
F8 28.25 28.27 31.40 31.07 30.20 30.21 29.93 29.94 31.37 31.55 30.31 30.09
F9 28.21 28.42 31.07 31.00 30.33 30.27 29.97 29.81 31.72 31.69 30.14 30.37
F10 28.59 28.55 31.18 31.10 29.45 30.49 29.96 29.95 32.04 32.43 30.29 30.10
F11 28.42 28.36 31.22 31.04 30.27 30.20 29.93 29.98 31.65 31.66 29.94 30.40
F12 28.50 28.43 31.19 32.11 30.32 30.20 29.89 30.07 31.75 32.70 30.30 30.13
F13 28.45 28.53 31.92 30.70 30.31 30.22 29.89 30.00 31.95 32.04 30.30 28.46
F14 28.33 28.56 31.13 31.07 30.58 30.69 29.93 29.91 31.78 32.08 30.10 30.27
F15 28.31 28.18 31.02 30.98 30.14 30.25 29.91 30.02 31.81 31.59 30.06 30.11
F16 28.14 28.14 31.12 31.23 30.20 30.25 30.07 30.24 32.24 31.61 30.46 30.26
Conclusion: It can be seen from the test results of the above mentioned
three batched of reagents that, on the SLAN-96P instrument, the test results of
weak positive samples containing the above-mentioned internal and external
interfering substances are not significantly different from those of the control
group (weak positive samples that do not contain interfering substances).
Therefore, it can be shown that, under the experimental conditions of this kit,
these internal and external substances below the above-mentioned
concentration have no effect on the test results of the kit.