Test preparation
Instrument-specific information
Table 1 shows all of the instruments that have the program for this test. The table also
shows sample cell and orientation requirements for specific instruments.
To use the table, select an instrument, then read across to find the applicable information
for this test.
Table 1 Instrument-specific information
Instrument Sample cell orientation Sample cell
DR6000 The fill line is to the right. 2495402
DR3800
DR2800
DR2700
DR1900
DR5000 The fill line is toward the user.
DR3900
Before starting
Analyze the samples immediately. The samples cannot be preserved for later analysis.
Time and temperature are very important to this test. The sample should be 25 ± 1 °C (77 ± 2 °F) and the times specified in
the procedure steps must be followed precisely. A temperature-controlled water bath is recommended for best accuracy.
Prepare formaldehyde-free water as follows: Distill water from alkaline permanganate (4 g sodium hydroxide, 2 g potassium
permanganate per 500 mL of water). Discard the first 50–100 mL of distillate.
Do not use the Pour-Thru Cell or sipper module (for applicable instruments) with this test.
Review the Safety Data Sheets (MSDS/SDS) for the chemicals that are used. Use the recommended personal protective
equipment.
Dispose of reacted solutions according to local, state and federal regulations. Refer to the Safety Data Sheets for disposal
information for unused reagents. Refer to the environmental, health and safety staff for your facility and/or local regulatory
agencies for further disposal information.
Items to collect
Description Quantity
1
Refer to Consumables and replacement items on page 6 for order information.
Sample collection
• Analyze the samples immediately. The samples cannot be preserved for later
analysis.
• Collect samples in clean glass or plastic bottles with tight-fitting caps. Completely fill
the bottle and immediately tighten the cap.
• Prevent agitation of the sample and exposure to air.
Test procedure
Start
1. Start program 200 2. Prepare the blank: Pour 3. Add the contents of one 4. Start the instrument
Formaldehyde. For 25 mL of formaldehyde-free MBTH powder pillow to the timer. A 17-minute reaction
information about sample water into a 50‑mL mixing blank. Put the stopper on time starts.
cells, adapters or light cylinder. the mixing cylinder. Immediately proceed with
shields, refer to Instrument- the next step after the timer
specific information starts.
on page 1.
5. Immediately after the 6. Prepare the sample: Fill 7. When the timer shows 8. Put the stopper on the
reaction period starts, shake a mixing cylinder to the 25- "15:00", add the contents of mixing cylinder. Shake
the blank sample vigorously mL line with sample. one MBTH powder pillow to vigorously for 20 seconds.
for 20 seconds. Do not wait the prepared sample.
for the timer to expire.
Zero
13. Just before the timer 14. Immediately clean the 15. Insert the blank into the 16. When the timer shows
shows "2:00", pour 10 mL blank sample cell. cell holder. "2:00" push ZERO. The
of the blank solution into a display shows 0 µg/L CH2O.
sample cell. Pour the
solution slowly to prevent
bubbles on the cell walls. If
bubbles form, swirl to
remove the bubbles.
Read
17. Fill a second sample 18. Clean the prepared 19. Insert the prepared 20. When the timer expires,
cell with 10 mL of the sample cell. sample into the cell holder. push READ. Results show
prepared sample. in µg/L CH2O.
Interferences
Interfering substance Interference level
Acetate More than 1000 mg/L
Aldehydes (other) Positive interference at all levels
Ammonium (as N) More than 10 mg/L
Aniline More than 10 mg/L
Accuracy check
Standard additions method (sample spike)
Use the standard additions method (for applicable instruments) to validate the test
procedure, reagents and instrument and to find if there is an interference in the sample.
Items to collect:
• 4000-mg/L Formaldehyde Standard Solution
• 50-mL mixing cylinder, graduated (3x)
• Flask, volumetric, Class A, 100-mL glass
• Pipet, TenSette®, 0.1–1.0 mL and tips
• Formaldehyde-free water
Method performance
The method performance data that follows was derived from laboratory tests that were
measured on a spectrophotometer during ideal test conditions. Users can get different
results under different test conditions.
Program Standard Precision (95% confidence interval) Sensitivity
Concentration change per 0.010 Abs change
200 320 µg/L CH2O 312–328 µg/L CH2O 3 µg/L CH2O
Summary of Method
Formaldehyde reacts with MBTH (3-methyl-2-benzothiazoline hydrazone) and a
developing solution to form a blue color. The color intensity is proportional to the
formaldehyde concentration. The measurement wavelength is 630 nm.