2 3 2
2 3 2
Purpose of test
Materials
Source holder
Procedure
1. To assess precision, calculate for each source the percentage differences between the
individual measured activities, A£, and their mean, A, that is:
Record results of calculations
2. To assess accuracy, calculate for each source the percentage difference between the
mean measured activity, A, and the certified activity of the source corrected for
radioactive decay to the day of measurement, C, that is
Record results of calculations
Observations
If the operating conditions for the radionuclide concerned can be selected either by means of
a selector switch, selector push button or plug-in module or by means of a potentiometer with
a dial, the relevant procedure should be carried out twice, first with one than the other
enabled. The results should agree. The procedure described tests the accuracy of the
instrument in measurements on the radionuclides used in the test, but not necessarily in
measurements on other radionuclides. This may be a significant limitation, especially with
instruments having selector switches, selector push-buttons or plug-in modules for different
radionuclides. When therapy with unsealed radionuclides is carried out, it is especially
recommended that the accuracy of the instrument used for activity measurements be
additionally tested annually with certified sources of the radionuclides concerned. If a
certified source for a radionuclide of particular interest is not available, an estimate of the
accuracy of the instrument in measurements on this radionuclide may be obtainable from
results with another radionuclide through knowledge of the decay schemes of the two
radionuclides and the energy response function of the ionization chamber. Thus, a certified
"Co source may be used to estimate the accuracy of the instrument in measurements on
""Tcm. The advice of the manufacturer may be sought in this regard. The expedient is likely
to involve significant errors, however, when the radiation energies involved are low.
Interpretation of results
The results may reveal imprecision (random errors), bias (systematic errors), or both.
Limits of acceptability
The limits of acceptability for the results of the test are determined by the precision and
accuracy of the instrument specified by the manufacturer. In general, however, for
measurements on sources such as those specified, the precision should be such that all
individual measured activities on any source are within + 5% of the mean measured activity,
provided that radioactive decay has a negligible effect over the measurement period, and the
accuracy should be such that the mean measured activity is within +_ 10% of the certified
value corrected for radioactive decay to the day of measurement.
Conclusion
Record whether or not the results confirm acceptable performance. If not, indicate follow-up
action taken.