Professional Documents
Culture Documents
BCI000002 Master T User Manual 1.1
BCI000002 Master T User Manual 1.1
Master T
BCI000002
USER MANUAL 1.1
Related
Rev. Date Description
Section
Changed manual layout, some
1.1 03/2018 All
corrections
Master T – User Manual
INDEX
INTRODUCTION ......................................................................................3
1 DESCRIPTION OF THE ANALYSER ..................................................3
2 INSTALLATION ...................................................................................6
2.1 Rear panel setting .................................................................................................... 6
2.2 Location of the instrument ........................................................................................ 7
2.3 Power supply connection and label .......................................................................... 7
2.4 Caution .................................................................................................................... 8
2.5 Incubator.................................................................................................................. 8
2.6 Printer settings ......................................................................................................... 9
3 MAIN MENU......................................................................................10
4 SETUP MENU ...................................................................................13
5 ARCHIVES MANAGEMENT .............................................................18
5.1 Edit method............................................................................................................ 22
5.2 Edit Control ............................................................................................................ 33
5.3 Graphic QC ............................................................................................................ 33
5.4 Delete Archives ...................................................................................................... 37
5.5 Delete Archiv. Method ............................................................................................ 37
6 OPERATING TIPS ............................................................................38
6.1 How to aspirate the liquids into the flow cell ........................................................... 38
6.2 Manual reading mode ............................................................................................ 38
6.3 Washing................................................................................................................. 40
6.4 Reset and switching off .......................................................................................... 40
7 RUN METHOD ..................................................................................41
7.1 Recall method ........................................................................................................ 41
7.2 Blank (Zero-setting) ............................................................................................... 45
7.3 Production of Standards ........................................................................................ 46
7.4 Standard confirm procedure ................................................................................... 46
7.5 Sample measuring ................................................................................................. 49
7.6 Flag Messages ...................................................................................................... 52
8 ABS MENU .......................................................................................54
9 WASHING .........................................................................................55
10 MAINTENANCE .............................................................................56
10.1 Cleaning procedures .............................................................................................. 56
10.2 First installation ...................................................................................................... 56
10.3 Everyday cleaning.................................................................................................. 56
10.4 Special cleaning ..................................................................................................... 56
10.5 Further advice ........................................................................................................ 57
2
Master T – User Manual
INTRODUCTION
Thank you for choosing Hospitex Master T, a new instrument designed with powerful
features in a compact unit for biochemical investigations, enzymes, turbidimetry and
drug tests. With such characteristics as the easy to operate integrated touch screen,
the Master T is the perfect solution for you laboratory.
3
Master T – User Manual
It performs:
- End point
- Kinetic
- Fixed Time
- Multistandard assays
Results are shown by graphic display and printed on paper by built-in printer.
4
Master T – User Manual
Technical specifications:
Minimal
measurement 400 L
volume
5
Master T – User Manual
2 INSTALLATION
6
Master T – User Manual
- Please check the setting of the power supply switch according to your country’s
electrical network.
- Connect the power plug to a good grounded AC wall outlet, preferably one that
is not shared with other electric appliances and with low fluctuation of line
voltage compared to the standard voltage specified (10-15%).
- Keep the instrument away from other appliances that generate high frequency
electrical noises (e.g. radiological instruments).
- Before connecting the power cord, check that the AC power supply corresponds
to the value that is stated on the instrument’s label. Set the voltage selector
properly.
7
Master T – User Manual
2.4 Caution
Do not connect the instrument to a power supply different from the value indicated on
the label.
- Before connecting the power and finishing the installation section, make sure
that the instrument is turned off (check the power switch located on the rear
part of the instrument).
- Make sure that your AC main line has an efficient ground line. A bad ground
line connection may compromise analysis results and damage the
instrument.
- After turning on the instrument, pay attention not to spill liquids or micro solid
substances on the surface around the instrument.
- Keep the instrument away from young children.
If the above procedures are carefully followed, you are now ready to TURN ON
the instrument by using the switch located on the rear panel.
2.5 Incubator
The incubator temperature is brought up to 37°C by the software and it will remain
Constant until the instrument is turned off.
Note: It is very important to heat the macro cuvettes to the proper temperature for
obtaining the most accurate analysis results.
8
Master T – User Manual
Step 1 Step 2
Step 3 Step 4
9
Master T – User Manual
3 MAIN MENU
- Switch on: Turn on the instrument from the main power switch located behind
the instrument.
- First screen: User name screen, digit the operator code by pressing the keys
then press SAVE, now the software will proceed to main menu. The operator
code will be displayed at the beginning of each method analysis. To skip this
option and proceed to main menu just press EXIT
Main menu: The main menu is composed of 6 options, which can be selected by
scrolling over the messages. Every option is linked to a function of the instrument:
10
Master T – User Manual
MESSAGE DESCRIPTION
Printing options.
11
Master T – User Manual
Feed button
Feeds paper out from the printer.
Exit Button
Press this button to exit from all menus
and return to main menu.
12
Master T – User Manual
4 SETUP MENU
From main menu select Setup, the following screen will appear:
From the set up menu it is possible to customize the instrument features to best fit
the operator’s needs. Scroll the menu by using the up and down arrows.
MESSAGE DESCRIPTION
13
Master T – User Manual
14
Master T – User Manual
Example:
15
Master T – User Manual
16
Master T – User Manual
Save button
Press this button to SAVE changes.
17
Master T – User Manual
Exit Button
Press this button to exit from all
menus and return to main menu.
5 ARCHIVES MANAGEMENT
Database dimensions
The Master T database memory is in two different hardware parts: EEPROM and
Flash memory.
The Master T archives are:
QC: Up to 16 different QC types can be memorized, each one holds 200 results.
18
Master T – User Manual
MESSAGE DESCRIPTION
19
Master T – User Manual
- Levey-Jennings graph
20
Master T – User Manual
- Westgard Rules
- Centred value
- Mean
- SD: Standard Deviation
- CV: Coefficient of Variation
Exit
Press this button to exit from all menus and return
to main menu.
QC Print Format:
Print button
------------------------------------
11/01/2005
015 - GLUCOSE
Date ABS Result
09/02 0.345 1 01.4 mg/dl
10/02 0.341 100.2 mg/dl
12/02 0.347 101.9 mg/dl
N. Samples 3
Average: 101.1
CV 1.6
21
Master T – User Manual
Graph button
Use this button to print the Levey-Jennings graph
- Delete method
- Delete Arch. Method
Editing the assay parameter files incorrectly may affect the calculation of the results
and may produce erroneous results. Verify edits to the assay parameter files against
the Master T assay applications.
Press Archive Management menu and the following screen will appear:
Choose Edit method, the list of assays present in memory will appear.
22
Master T – User Manual
- Use the up and down arrows to move to different options over the screen.
The different option will be highlighted once selected.
- Use the Page up and down buttons to move to browse between the list
pages.
- Use the Home and End buttons to move to the limits of the list.
Select one method and the relevant parameters will appear as follow:
MESSAGE DESCRIPTION
- KIN: KINETIC
Method Type - FXT: FIXED TIME
- EIA: ELISA
23
Master T – User Manual
- WATER
- BLANK
Zero
- SAMPLE BLANK
25°C
Temperature 30°C
37°C
Factor calibration:
- Select: Calibration NO
Calibration
- Then press down arrow
- The Factor box will be displayed.
- Insert the numeric value for the factor.
Use the number keys to enter the
numeric value, press C to delete.
24
Master T – User Manual
Standard calibration:
25
Master T – User Manual
keys.
26
Master T – User Manual
27
Master T – User Manual
28
Master T – User Manual
Save button
Press this button to SAVE changes.
Exit Button
Press this button to exit from all menus and
return to main menu.
29
Master T – User Manual
Note:
In order to avoid errors in method running (during a set series of experiments), it is
important not to modify the CALIBRATION and the method TYPE once the series of
experiments has begun. Any modification to one specific methods parameter will
automatically erase the calibration in memory, a new calibration procedure will be
request before being able to run the method.
31
Master T – User Manual
- Look for the first free position stated as “NEW METHOD” in the assay list
using the Up and Down arrow buttons.
- Select and enter inside the method parameters.
- Name method and set assay parameters following assay producer
specification.
- Save and Exit.
Edit method:
Erase method:
32
Master T – User Manual
- Press DEL and then SAVE to delete from list a control list.
5.3 Graphic QC
In Graphic menu it is possible to view and edit method values for each control
serums previously memorized.
The different control defined on the Edit Control page, will be then be displayed in the
method parameters in order select QC (Control Quality) procedure.
Westgard rules
Levey
Jennings
graph
QC
Statistic
values
33
Master T – User Manual
When QC results are completed with no errors, the results are saved in the Graphic
QC menu. The QC values and statistics can be viewed from this menu.
The Quality menu will report quality assay status, Levey-Jennings graph, QC Data
List and Westgard Rules.
Expected concentration and SD can be defined for the methods parameters, and
Westgards Rules can be view to evaluate the QC. If a Westgard Rule is violated the
OK message will disappear.
The QC Summary Table displays the Westgard QC Rules and Statistics for all the
assay and levels of QC.
BUTTON DESCRIPTION
CODES:
34
Master T – User Manual
“ * “: Excluded QC value.
MESSAGE DESCRIPTION
The following table defines each column of the Westgard Rules information:
RULE WESTGARD RULES DESCRIPTION
35
Master T – User Manual
DESCRIPTION
36
Master T – User Manual
Note: The Master T patient database has a total memory capability of 1000 tests.
When the reserved memory space is full, new test results will superimpose the older.
Because of this, we suggest printing the patient test results at the end of each
working session (Batch or Profile print), or at the same time as the test is performed.
How to delete:
- Press “Del. Arch. Method”
- The list of tests will appear
- Scroll (highlight) with the up and down arrows the method from the database
that must be deleted.
- Press No to exit
37
Master T – User Manual
6 OPERATING TIPS
Figure A
38
Master T – User Manual
- To take measurement press the aspiration switch or touch the help window on
display.
- Read the sample.
- Repeat same operation for the following samples.
When using the cuvette disposable Macrocuvette, make sure there is at least 1000
µL of total volume of Reagent + Sample in the manual cuvette.
Then insert the manual cuvette in the cuvette position and press the aspiration switch
or touch the help window on display.
Make sure to position the manual Macrocuvette correctly, THE READING SURFACE
MUST BE FROM LEFT TO RIGHT of instrument, as follows:
Macrocuvette
Reading well
39
Master T – User Manual
6.3 Washing
After each session of measurements, it is recommended that the flow cell be washed
with distilled water.
For washing the flow cell press the ‘Washing’ command so the pump will drain away
the water at high speed into the flow cell.
After washing, it is necessary to repeat the operation without water so that the water
remaining inside the instrument can be expelled completely. It is very important to not
alter the next measure.
Remember that the instrument must only be reset if the software does not work.
40
Master T – User Manual
7 RUN METHOD
From Main screen select Run Method, the list of all tests present in memory will be
displayed.
After a method has been chosen, the run screen will appear as follow:
41
Master T – User Manual
The Run screen is divided in 12 windows, some of them are just for data visualization
(the touch screen feature is not active in such zones), the other windows show
information and the touch screen feature is active:
MESSAGE/BUTTON DESCRIPTION
ID assignment criterion:
42
Master T – User Manual
43
Master T – User Manual
44
Master T – User Manual
Kinetics methods:
Once the method is loaded, its parameter will be automatically printed. To decide
whether or not to print automatically the method parameters, please refer to setup
menu for information about printing format.
The Blank is performed each time the absorbance zero is required from method
parameters:
- When water or reagent blank is chosen, the zero protocol is performed only
once, at the beginning of the running method.
- When sample blank zero is chosen, the zero protocol is performed at the
beginning of each measurement (standard or sample measurement) of the
relevant running session.
Before pressing the aspiration switch, after every standard or sample
measurement, it is necessary to fill the flow cell with zero solution again, after
this operation it is possible to proceed to a new standard or sample
measurement.
45
Master T – User Manual
If the current method has been performed at least once, the old standard results will
be retained in the memory and must be reviewed, confirmed, or modified.
If instead the current method has not been previously performed, the standard value
will not appear and the calibration procedure in the help window will automatically
be displayed.
The first time that the program runs the old factor is the current factor.
The first time that the program runs the old standard session is the current standard
session.
46
Master T – User Manual
Procedure:
Once a method is recalled from the list, it will ask for calibration in the help window
box as follow:
1) CALIBRATING METHOD? YES/NO
2)
NEW BLANK? YES/NO
3) INSERT BLANK
Set up (according to the zero type) the water or reagent blank test tube and fill the
flow cell by pressing the aspiration switch.
Wait for a delay time (5 sec.) for the environment to stabilize.
ABS blank value will be acquired.
4) INSERT STANDARD
Set up the standard solution tube and fill the flow cell by pressing the aspiration
switch.
Calibration value will be acquired, the K factor will be displayed in the relevant box,
the instrument will wait for confirmation order as follow:
47
Master T – User Manual
Press NO to refuse last calibration procedure, the help window will ask the
following:
In case no calibration factor has been stored in the memory, the instrument will
require the calibration as mandatory, starting from point 3.
All calibrations, single point or multipoint will be graphically visualized on the screen.
48
Master T – User Manual
For End Point and Elisa methods, after the delay time, the result of the ABS
standard measurement and the concentration value are displayed. The result of the
sample concentration has a supervisor control (see Flag messages):
49
Master T – User Manual
During the incubation time and reading time for Kinetic and Fixed Time methods the
program will display the following screen:
During the reading time for Kinetic and Fixed Time methods the screen shows:
50
Master T – User Manual
ID assignment criterion:
The ID number is set automatically to 1 when the first sample of the first assay is run
for the first time of the day.
The first available ID for each Assay will appear automatically once the test has been
recalled.
In the patient database each ID correspond to only one patient name; it is not
possible to have the same patient name linked to a different ID number.
To change the patient ID, touch the ID box and a virtual keyboard will open, insert the
required ID exit and save. Now the next patient will be identified with the
programmed ID number.
- Just after a patient measurement it is possible to repeat again with the same
solution in the cuvette. In order to do this press the REP button on the screen
and read, the new measurement value will superimpose over the old one.
- If the sample repetition will be performed a second time, touch the ID box and
a virtual keyboard will open, insert the required ID exit and save. Now the
results of the recalled patient for that analysis will be displayed, press the REP
button on the screen and read again, the new measurement value will
superimpose the old one.
Note: To perform repetition of readings using the flow cell, it is necessary to disable
the automatic waste function. In this way the peristaltic pump will be not activated
after patient measurement and the sample will remain in the flow cell for repetition.
To disable automatic serum discharge, go in Setup menu and set to NO automatic
empty, save and exit.
To perform repetition of readings using the manual cuvette, it is not necessary to
disable the automatic waste function.
51
Master T – User Manual
Processing QC controls
In order to insert in the QC database a control serum value for a specific assay
proceed as follow:
- During the run session, it is possible to insert 3 control types (QC1 – QC2 –
QC3) for each method type.
- To process a QC value, touch the ID box and the virtual keyboard will open.
- Now it is possible to choose between 3 different QC (QC1 – QC2 – QC3)
stated in the keyboard keys.
- Once the selection is done, the QC will appear automatically in the Nm patient
name box.
- Once the QC reading measurement has been processed the result will be
saved in the database.
- The QC database can memorize up to 25 archives containing 100 results
each; any archives with more than 100 results will substitute the oldest for the
newest (circular buffer).
MESSAGE
MEANING DESCRIPTION
FLAG
Probable Cause
Pathological evidence
Patient result
value is higher Corrective Action
H of normality
Review the result and report as over
range defined
for test. the normality high range value.
Dilute the sample and rerun.
Probable Cause
Pathological evidence
Patient result Corrective Action
value is lower
L of normality Review the result and report as under
range defined the normality low range value.
for test.
Concentrate sample and rerun.
52
Master T – User Manual
53
Master T – User Manual
Probable Cause
8 ABS MENU
ABS mode is a service function, useful to test the ABS of a sample with a
programmable wavelength. To use this function, you must set the instrument to zero
with a reference (for example water or reactive).
When the program arrives to this screen, the instrument will set up the optical filter
(showing the current wavelength), select a different one if necessary using the
arrows, press the filter wheel icon to set the new wavelength.
Set the test tube of water (or other substances) in the inlet pipe and press the
aspiration button, after a few seconds, continuous evolution of the ABS value will
appear on the display. During this phase the following options could be chosen:
- Press 0 key to have zero on the current ABS value (water or other
substances). You will choose this option when you will set to zero the
instrument.
- Prepare a new sample (reagents, potassium-dichromate or others) and press
the aspiration switch to start a new ABS measurement session.
- Press the aspiration switch to fill the flow cell with the sample.
- Press 0 key to choose the zero value.
- Press Exit button to return to main menu.
54
Master T – User Manual
9 WASHING
After each session of measurements, it is recommended that the flow cell be washed
with distilled water.
- MAIN menu
- RUN Mode
- ABS Mode.
For washing the flow cell set the tube with water in the inlet pipe and press the Wash
button so the pump will aspirate water and wash it.
After the washing, it is necessary to repeat the operation without water so that the
water remaining inside the instrument can be expelled completely. It is very important
to not alter the following measure.
55
Master T – User Manual
10 MAINTENANCE
Perform at least 10 complete washing cycles by using the Washing procedure. Every
washing cycle lasts approx. 1 minute and stops automatically.
At the end of each working session, it is recommended to rinse the instrument with
detergent solution or alcohol via the inlet pipe. Place a cup of detergent solution
under the inlet pipe and carry out the Washing procedure.
Empty the waste tank: The waste bottle and all tubings contain sample and reagent
material. This material should be treated as potentially bio hazardous. Appropriate
waste management should be observed!
56
Master T – User Manual
- If the instrument is not being used for a long period of time, empty the
hydraulic circuit and disconnect the tubes.
- When the instrument is in use, keep the two little doors of the instrument
(measurement block and printer doors) firmly closed.
- Clean the external surface of the instrument every month with a non-abrasive
detergent.
- Clean the surface near the fan grids every month in order to take off
dangerous dust.
57