Velocity Type Strong

Motion Seismometer

General
The TSM-1 Velocity Type Strong-Motion
TSM - 1
Seismometer incorporates three built-in seismic
sensors including two for horizontal direction and
one for vertical direction. The TSM-1’ s wide
frequency and dynamic range capability can
handle the full spectrum of seismic disturbances,
from microtremors to large earthquakes.
The Seismometer’ s highly sensitive, all-purpose
enclosed triaxial design provides superior mobility,
which makes it easy to set up at temporary
observation stations.

Applications

Strong earthquake and aftershock observation

Earthquake measurement, monitoring, alarm
Sensor unit
warnings

Monitoring of structural sway caused by Frequency Response
long-period ground motion 20 dB

Volcano observation
0 dB

Features
Gain[dB]

−20 dB
1) Triaxial design simplifies installation.
2) Force balance torquer arrangement requires no on-site −40 dB

adjustment.
−60 dB

3) Capability to cope with long-period, large amplitude 360 deg

motion with room to spare.


180 deg
4) High output voltage simplifies data logging and signal
Phase[Degree]

processing. 0 deg

5) Simple construction, high reliability.


−180 dB

External Dimensions (Unit : mm) −360 dB


0.001 Hz 0.01 Hz 0.1 Hz 1 Hz 10 Hz 100 Hz 1k Hz
Frequency[Hz]
25

Measurement Range
NS 300 kine
308

Horizontal
300

Component
30 kine
EW
3 kine

300m kine
Velocity

30m kine

3m kine
23

200 300μ kine

UD
161

30μ kine

3μ kine Microseism
120

0.01 Hz 0.1 Hz 1 Hz 10 Hz 100 Hz


Frequency

246
29

10

Sensor unit Signal distributor


Velocity Type Strong–Motion Seismometer TSM-1

Specifications
Composition Horizontal direction, 2 + Vertical direction, 1
Output signal Velocity(Temperature range 25℃±15℃ stable temperature) Acceleration(reference)
Measurement range ±3 m/S(±300kine) [see Note 1] ±20 m/S(±2000gal)
2

Dynamic range 145dB@1Hz Typ. −


Frequency response(±3dB) 0.01∼80Hz±10% −
Output sensitivity 6 V/(m/S)±5%[60mV/kine](Equilibrium output) 1 V/(m/S2)±5%[10V/(9.8m/S2)] ( Equilibrium output)
Scalefactor temperature coefficient 700ppm/℃ or less (Design value) 200ppm/℃ or less
Output sensitivity linearity
0.02% 0.02%
(9.8m/S2 standard)
±1×10−3 m/S or less[±100×10−3 kine] ±50×10−3 m/S2 or less[±5gal]
Bias
Typ.±0.3×10 m/S[±30×10 kine]
−3 −3

40×10−6 m/S/℃ or less[4×10−3 kine/℃] (Design value) ±1.5×10−3 m/S2/℃ or less[150×10−3 gal/℃]
Bias temperature coefficient
Typ.4×10 m/S/℃[0.4×10 kine/℃]
−6 −3

180×10-9 m/S@1Hz Typ.[18×10−6 kine]


Resolution −
(−135dB@1Hz Typ.)
Cross-sensitivity 0.5% or less
Operating temperature −20℃∼+60℃
Enclosure rating JIS C 0920(IP65): dust-tight; water jet protection 
Power AC100V[see Note 2]or DC12 to 24V
Power consumption AC100V : 100mA or less DC12 to 24V : 10W or less
Mass (Unit : mm) Sensor unit : 10kg or less Signal distributor : 6kg or less
Note 1: upper limit of measurement range guaranteed by electrical current input to torquer equivalent to the maximum amplitude.
Note 2: The power customized to any voltage.

Composition 20
Options
⑤ AC100V
0 No. Name Model No. Description
⑤ Power cable ⑥ External TSM-1-T Available options
③ Sensor cable
−20 diagnostics *Test signal generation
② Signal distributor
① Sensor unit equipment *Integrated response check
−40 ⑦ Diagnostics cable *External signal input
*Electric power from
−60 ⑥ External ④ External signal distributer
0.001 0.01 0.1 1 10 100 1000
diagnostics
Frequency[Hz] signal cable ⑦ Diagnostics TSM-1-W4 Used for ⑥
⑧ Setup fixture equipment
360 equipment cable Standard length 2m
⑧ Setup fixture TSM-1-M* Used for ①
180 *Nos 1、2、3 etc
No. Name Model No. Description
① Sensor0
unit TSM-1-S
② Signal distributor TSM-1-D
③ Sensor
−180
cable TSM-1-W1 Standard 10m
④ External signal cable TSM-1-W2 Standard 2m
⑤ Power
−360
cable TSM-1-W3 Standard 2m
Cable customized
0.001
to0.01
any length
0.1 1 10 100 1000

Measurement Range
1000kine
Design and specifications are subject to change without prior notice, and without any obligation on the part of the manufacturer.
100kine

10kine

1kine
CAUTION Before operating this equipment, you should first throughly read the operator's manual.
Velocity[m/s]

Horizontal component
100mkine
Vertical component
10mkine

1mkine

100μkine

10μkine

1μkine
0.01 0.1 1 10 100
Frequency[Hz]
Communication & Control Systems Company Sensing Control Systems Dept.
Head Office 2-16-46, Minami-Kamata, Ohta-ku, Tokyo 144-8551, JAPAN
TEL.+81-3-3731-2631 FAX.+81-3-3738-8670

December2016 Cat.No.1432-2-1-RG

You might also like