Professional Documents
Culture Documents
Analyzing & Testing. Dilatometry. Method, Instruments, Applications From - 180 C To 2800 C. Leading Thermal Analysis
Analyzing & Testing. Dilatometry. Method, Instruments, Applications From - 180 C To 2800 C. Leading Thermal Analysis
Dilatometry
Method, Instruments, Applications – From -180°C to 2800°C
Dilatometry
2
Table of Contents
Dilatometry Accessories 20
High-Temperature Applications 28
Shrinkage Dilatometry
Softening
Expansion
Polymers
Ceramics Composites
Porcelain CTE
Glass
3
Dilatometry – Method and Principle of Operation
Method
Principle of Operation
4
Results and Measuring Accuracy
dL/Lo / %
Literature (1500°C): 1.307 %
Test #1 (1500°C): 1.309 % The thermal expansion in percent is
Sample: Polycryst. Alumina Test #2 (1500°C): 1.312 %
1.400 Test #3 (1500°C): 1.311 % an important value for evaluating a
1.200 material’s performance.
Literature (1000°C): 0.799 %
Test #1 (1000°C): 0.800 %
1.000
Test #2 (1000°C): 0.802 % In this diagram, the length changes of
Test #3 (1000°C): 0.802 %
0.800 3 test runs (lines) on polycrystalline
alumina (Al2O3) are presented together
Literature (500°C): 0.350 %
0.600 Test #1 (500°C): 0.349 % with the corresponding data from
Test #2 (500°C): 0.350 %
0.400
Test #3 (500°C): 0.351 % literature (crosses).
0.200
The differences between these values at
0.000 500°C, 1000°C and 1500°C generally
amount to less than 1%, demonstrating
200 400 600 800 1000 1200 1400 1600
Temperature / °C the outstanding accuracy of the
NETZSCH DIL 402 PC which was used
Comparison of the expansion data of polycrystalline alumina with literature values for these measurements.
The superior thermal uniformity of a Wide temperature range Linear thermal expansion
horizontal pushrod dilatometer can be Easy handling Coefficient of thermal expansion (CTE)
attributed to very low temperature High accuracy Volumetric expansion
gradients. A horizontal furnace is not Homogeneous temperature profile Shrinkage steps
subject to convection parallel with the Flexible sample geometry Glass transition temperature
sample. Low risk of contamination (LVDT) Phase transitions
Sintering temperature/sintering step
Particularly for long samples, a Density change
horizontal dilatometer system is Softening points
essential in ensuring that the temper- Influence of additives/raw materials
ature distribution is reasonably uniform. Decomposition temperature –
e.g., of organic binders
Anisotropic behavior
Optimizing of the firing process
Rate-controlled sintering (RCS)
Caloric effects by using c-DTA®
Thermokinetics
5
Dilatometer 402 PC – The Reliable Workhorse for
Glass and Ceramics
The DIL 402 PC is specially tailored to Design The forced-air cooling system of the
the needs of the ceramic and glass 1600°C furnace makes it possible to
industry. High resolution and stability, a The instrument has a horizontal design cool from maximum sample temper
wide measurement range, and a robust with an easy-to-operate furnace. A large ature down to room temperature.
and compact design are only some of recess in the tube-type sample carrier
the many advantages of this facilitates the placement of samples
cost-effective instrument. This even when their geometries are less Sample Holder Systems and Sample
dilatometer combines easy operation, than ideal. A thermocouple in direct Dimensions
high adaptability to different applica- proximity to the sample ensures repro-
tions, and outstanding performance in ducible temperature measurement. Exchangeable sample holder systems
one instrument. The optimized design made of alumina and fused silica are
of the measurement system with available. The maximum sample dimen-
inductive transducer compensates for Furnaces sions are 50 mm in length with a
temperature fluctuations and yields diameter of up to 12 mm (optionally
highly reproducible data. A chiller is not The interchangeable furnaces with 19 mm).
required. temperature ranges up to 1200°C
and 1600°C allow for the analysis
The DIL 402 PC operates in accordance of expansion control in solids, green
with nearly all national and international bodies, powders and pastes, for
standards (e.g., DIN 51045, or production and quality assurance
JIS R 2207-contact version). purposes across a variety of applications.
6
Technical Key Data for the DIL 402 PC At a Glance – The Characteristics
of the DIL 402 PC
RT to 1200°C
Temperature range/furnaces
RT to 1600°C High flexibility with two
Heating rate 0.1 to 50 K/min interchangeable furnaces
Temperature precision 0.1 K Various accessories for special
applications (sample holders
Measuring range 500/5000 μm
and containers)
Load at the sample 15 cN to 45 cN (optionally 45 cN to 100cN) Large sample dimensions
∆l resolution 8 nm/digit Ease of operation
Sample diameter Up to 12 mm (optionally 19 mm) Highly reproducible results
Up to 25 mm
Sample length
Up to 50 mm
Alumina
Sample holders
Fused silica
3-point bending mode (determination
of the firing stability, set made of Al2O3)
Special sample holder
Fibers in tension
systems
Thin metals and ceramic foils
Penetration
Powders
Sample containers Pastes
Metals and glasses in the liquid state Special sample holder for the determination of
the firing stability
Oxidizing static and dynamic
Atmosphere
Inert dynamic
For temperature calibration and detection of
c-DTA®
endo- and exothermal effects
Gas flow controller For purge gas (optional) thermocouple furnace
sample
sample
inductive pushrod carrier
transducer
adjustment screw
base unit DIL 402 PC – 3-dimensional sectional view
7
Dilatometer 402 C – Perfect for Sophisticated Demands
The Dilatometer DIL 402 C features an Measurement System Motorized Pushrod and Automatic
excellent set of capabilities optimized Zeroing
to fulfill any potentially occurring appli- The high-resolution displacement trans-
cation need. These include flexibility of ducer is employed in a Vacodil® (Invar) The motorized pushrod and the easy-
temperature range, sample holder, measuring system. The maximum to-move furnace make it quite simple
and gas atmosphere, along with high measuring range is 5000 μm. The to insert samples, even when sample
accuracy and easy handling – the extremely low drift exhibited by the geometries are less than ideal. The
result of over 50 years of experience in system guarantees measurements pushrod is brought into contact with
building dilatometers. with very high repeatability, accuracy the sample via a software command
and long-term stability for application which also automatically brings the
The state-of-the-art design is compact temperatures up to 2000°C. LVDT to a zero/central position. Should a
and extremely flexible; it uses a number different starting position be required for
of interchangeable furnaces with large expansion or shrinkage, this can
horizontal movement and provides Wide Temperature Range and be accomplished with a simple software
easy access to the sample holder. It is Exchangeable Furnaces command as well.
the perfect instrument for applications
in the areas of not only traditional Six easily exchangeable furnaces are
ceramics and glass, but also advanced available to cover a temperature range
ceramics and powder metallurgy (e.g., of -180°C to 2000°C. This not only
when carrying out sintering studies on provides flexibility for different appli-
reactive powders). cations, but also ensures the smallest
temperature gradient along the sample
for both low- and high-temperature
applications.
DIL 402 C
8
At a Glance – The Characteristics
of the DIL 402 C vacuum
flange sample
tube sample
Wide temperature range thermocouple
carrier
6 exchangeable furnaces pushrod sample
Vacuum-tight by design heating shut-off valve
Large sample dimensions element (gas outlet)
front furnace
panel
flow-through
protective tube
The vacuum-tight design of the DIL In the DIL 402 C, the contact force of The thermal analysis system controller,
402 C allows for careful control of the pushrod can be adjusted between TASC 414, combines a multi-step
the atmosphere and pure gas condi- 15 cN and 45 cN, and optionally programmer and PID controller with a
tions. For example, the system can be between 45 cN and 100 cN. An high-resolution data acquisition system.
evacuated and back-filled with pure almost frictionless ball-bearing design A Sample Temperature Controller (STC)
inert or non-reactive gas. Static or supporting the pushrod guarantees unit provides for excellent sample
dynamic reactive gas atmospheres can smooth detection of changes in sample temperature control.
be applied as well. dimension.
Materials sensitive to oxygen can be
studied under pure inert gas conditions. Evolved Gas Analysis
To provide such conditions, an evacu- Thermostatic Control
ation system (optionally available) with The vacuum-tight design of the DIL
a two-stage rotary pump can be hooked To prevent any thermal influence from 402 C is ideally suited for connection
up to the standard vacuum flange. For the furnace or from changes in room to a mass spectrometer via a capillary
high-vacuum requirements (10-4 mbar), temperature, the LVDT system is kept at coupling or to an FT-IR via a transfer
a turbo molecular pump system is a constant temperature using a precise line. The outgassing of additives,
available. thermostat. This guarantees repro- organic binders and decomposition
ducible measurements even in the most products can thus be studied.
sensitive measuring range.
9
Dilatometer 402 C – Perfect for Sophisticated Demands
The tube-type sample carrier is standard for the DIL 402 C. The sample is placed
within the tube. Supports can be used for centering and to prevent direct contact
(sticking) between the sample and carrier. The sample carriers are available in fused
silica (max. 1100°C), alumina (max. 1680°C), or graphite (max. 2000°C).
The fused silica or alumina tube-type sample carriers can optionally be replaced with
rod-type sample carriers for ideal heat transfer to the sample. These consist of three
support rods, leaving the sample open to the heat source and the gas atmosphere.
This results in an increase in accuracy.
Sample Containers
For measurements on pastes, powders, molten metals and other material configura-
tions, special sample containers made of fused silica, alumina, sapphire or graphite
are optionally available.
10
Technical Key Data for the DIL 402 C
Temperature range -180°C to 2000°C
Metal: -180°C to 500°C (two versions available; closed or open to one side)
Stainless steel: -150°C to 1000°C (with extended heating zone)
Tube furnace: RT to 1000°C (with protective tube made of fused silica)
User-interchangeable furnaces
SiC: RT to 1600°C
Rh: RT to 1600°C (especially for use in glove box)*
Graphite: RT to 2000°C
Depending on furnace, available as open or closed tubes; made of:
Alumina
Interchangeable protective tubes
Fused silica
Glassy carbon
Heating rate 0.01 to 50 K/min
Temperature precision 0.1 K
Liquid nitrogen (low-temperature furnaces)
Cooling systems Forced air integrated in the SiC furnace and tube furnace
Water cooling (graphite furnace)
Inert
Oxidizing (graphite furnace up to 1680°C with special protective tube)
Atmosphere Reducing
Vacuum (<10-4 mbar (10-2 Pa))
Static and dynamic operation
Gas flow control supplement for 1 gas path
Gas flow control options
Mass flow controller (MFC), box for 3 gases
c-DTA® For temperature calibration and detection of endo- and exothermal effects
Measuring range 500/5000 μm
Standard: 15 cN to 45 cN
Load at the sample
Optionally: 45 cN to 100 cN
∆l resolution 0.125 nm/digit, 1.25 nm/digit
Sample diameter Up to 12 mm (optionally 19 mm)
Up to 25 mm (standard)
Sample length
Up to 50 mm (optionally for low-temperature furnaces)
Tube-type: fused silica, alumina, graphite
Interchangeable sample holders
Rod-type: fused silica or alumina
Thin metal, ceramic and glass foils
Penetration
Special sample holder systems
Bending (especially for determination of firing stability, set made of Al2O3)
Rod-type sample holder in combination with the humidity generator and humidity sensor
For pastes, liquids, waxes and molten metals (incl. aggressive samples) in fused silica,
Sample containers
alumina, sapphire, graphite and steel
Optional for measurements under a relative humidity of between 5% and 90%;
Humidity generator
temperature range from 30°C to 80°C
11
DIL 402 CD – Dual and Differential Excellence
The Dual and Differential Dilatometer DIL 402 CD – Two Instruments in One
The DIL 402 CD offers the highest level Advantages of the Double Sample Measurement System
of performance in thermal expansion Arrangement
measurements. In the double sample The high-resolution displacement trans-
arrangement, the system is capable of Corrected tests in one run ducer (LVDT, 0.125 nm/1.25 nm/digit) is
high sample throughput. In the differ- Compensation of drift effects for employed in an Invar measuring system.
ential arrangement, the system offers long-term measurements The maximum measuring range is
calibration-corrected measurements in Corrected RCS (Rate-Controlled 5000 μm. The extremely low drift
one run and improved long-term drift Sintering, Super-Res®) exhibited by the system guarantees
stability. measurements with very high repeat-
ability, accuracy and long-term stability
Furnaces for application temperatures up to
Design 2000°C.
Five easily exchangeable furnaces cover
The horizontal, vacuum-tight instrument the temperature range from -180°C to
construction offers specific advan- 2000°C. The forced-air, liquid-nitrogen
tages, especially in the dual sample and water-cooled furnaces offer fast
arrangement: homogeneous heating heating and cooling times along with a
of both samples, easy sample insertion, homogeneous temperature profile over
and effective protection of the the sample.
measuring system by gas flow.
In addition, no convective heat transfer
takes place between the measuring
system (two high-resolution inductive
displacement transducers) and the
sample area, thus eliminating increased
drift and noise.
DIL 402 CD
12
thermostatically- sample
At a Glance – The Characteristics controlled support thermocouple
of the DIL 402 CD displacement
transducer (2x) vacuum sample pushrod sample purge gas outlet
Wide temperature range (LVDT) flange carrier (2x) (2x) furnace (FTIR, MS)
5 exchangeable furnaces
Two measurement positions
Higher sample throughput
Calibration-corrected results in
one run
Accurately corrected Rate-
Controlled-Sintering (RCS) tests
Adjustable contact pressure
Extremely low drift and high
resolution
Coupling to evolved gas analysis purge gas inlet base plate/ flow-through
control panel protective tube
The vacuum-tight (<10-4 mbar (10-2 Pa)) To prevent any thermal influence from
design of the DIL 402 CD allows for the furnace or from changes in room
careful control of the atmosphere and temperature, the LVDT system is kept at
pure gas conditions. Measurements in a constant temperature using a precise
inert, oxidizing or reducing atmospheres thermostat. This guarantees repro-
can be performed under static or ducible measurements even in the most
dynamic conditions. sensitive measuring range.
The gas flow can be controlled by a
simple gas control supplement or a mass
flow controller system (MFC). In order to
provide pure inert conditions for oxygen-
sensitive materials, various evacuation
systems are optionally available.
13
DIL 402 CD – Dual and Differential Excellence
In the DIL 402 CD, the contact force of Fused silica (max. 1100°C)
the pushrod can be adjusted between Alumina (max. 1680°C)
15 cN and 45 cN. An almost frictionless Graphite (max. 2000°C)
ball-bearing design supporting the
pushrod guarantees smooth detection of In addition, a special sample carrier for
changes in sample dimension. measurements on thin metal, ceramic
and glass foils can be used with the
DIL 402 CD.
Sample Geometries
14
Technical Key Data for the DIL 402 CD
Temperature range -180°C to 2000°C
Five user-interchangeable furnaces:
Metal: -180°C to 500°C (two versions available; closed or open to one side)
Stainless steel: -150°C to 1000°C (with extended heating zone)
Furnaces
SiC: RT to 1600°C
Rh: RT to 1600°C (for use in glove box, Ar atmosphere)
Graphite: RT to 2000°C
Depending on furnace, available as open or closed tubes; made of:
Fused silica
Interchangeable protective tubes
Alumina
Glassy carbon
Heating rate 0.01 to 50 K/min
Temperature precision 0.1 K
Liquid nitrogen (low-temperature furnaces)
Cooling systems Forced-air integrated in SiC furnace and tube furnace
Water cooling (graphite furnace)
Inert
Oxidizing (graphite furnace up to 1680°C with special protective tube)
Atmosphere Reducing
Vacuum (<10-4 mbar (10-2 Pa))
Static and dynamic operation
Gas flow control supplement for 1 gas path
Gas flow control options
MFC gas control box for 3 gases
c-DTA® For temperature calibration and detection of endo- and exothermal effects
Measuring range 500/5000 μm
Load at the sample 15 cN to 45 cN, adjustable
∆l resolution 0.125 nm/digit, 1.25 nm/digit
Sample diameter Up to 6 mm
Sample length Up to 25 mm
Fused silica
Interchangeable tube-type
Alumina
sample holders
Graphite
Special sample holder system Slotted rod for measurements on thin metal, ceramic and glass foils
15
DIL 402 E – Precision at the Highest of Temperatures
For dilatometer measurements in Depending on the furnace type, various Safety System
the highest temperature range, the cooling facilities (e.g., forced air or
versatile, modular DIL 402 E is the water) are available which offer fast A comprehensive safety system
instrument of choice. The design of heating and cooling rates along with a continuously checks the flows of
this instrument is based upon the well- homogeneous temperature profile at cooling water and protective gas during
proven dilatometer series, but it can the sample. The water-cooled graphite the measurement. The power supply to
be equipped to achieve two additional tube furnace allows for heating rates the furnace would immediately be cut
upper temperature limits. of up to 50 K/min and short cooling off if trouble should arise.
periods to room temperature.
DIL 402 E
16
At a Glance – The Characteristics
of the DIL 402 E micrometer thermostatically thermo- water
screw controlled area pushrod couple sample cooling
Widest temperature range gas vacuum
User-interchangeable furnaces glass fibre to
pyrometer
Highly vacuum-tight
High accuracy up to 2800°C
Thermocouple or pyrometer gas
operation
Adjustable contact pressure
Short cooling times down to RT
Rate-Controlled-Sintering
(RCS) tests inductive insulation graphite
Extremely low drift and high transducter heating element
resolution
The DIL 402 E can be operated under The mechanics of the pushrod and The transducer’s measuring signal is
inert, reducing, static or dynamic transducer (LVDT) are not influenced by evaluated in the TA system controller.
atmospheres. Various vacuum pumps the furnace and sample holder. They The rapid analog-to-digital conversion
are available, including turbo molecular remain at a constant temperature due is carried out at 4 million digits and
pumps for creating the purest of to the thermostatic control. The pushrod reaches a high resolution of
atmospheres. can apply a load as low as 15 cN to the 0.125/1.25 nm/digit.
This vacuum-tight dilatometer allows sample.
for careful control of the atmosphere
and pure gas conditions by using a mass
flow controller.
17
DIL 402 E – Precision at the Highest of Temperatures
18
Technical Key Data for the DIL 402 E
Temperature range Room temperature to 2800°C
RT to 2000°C (thermocouple operation) to 650°C to 2400°C (pyrometer operation)
User-interchangeable furnaces
RT to 2000°C (thermocouple operation) to 650°C to 2800°C (pyrometer operation)
Interchangeable protective tubes Al2O3 for tests under oxidative atmospheres up to 1680°C
for the 2400°C furnace* Glassy carbon
Heating rate 0.01 to 50 K/min
Cooling systems Water cooling or cooling thermostat
Inert (He > 2000°C)
Oxidizing (graphite furnace up to 1680°C with special Al2O3 protective tube)
Atmosphere
Reducing
Static and dynamic operation
Vacuum <10-4 mbar (10-2 Pa)
Gas flow control supplement for 1 gas path
Gas flow control options
MFC gas control box for 3 gases
c-DTA® For temperature calibration and detection of endo- and exothermal effects
Measuring range 500/5000 μmmax.
Load at the sample 15 cN to 45 cN
∆l resolution 0.125 nm/digit, 1.25 nm/digit
12 mm in thermocouple operation
Sample diameter
6 mm in pyrometer operation
Sample length Up to 25 mm in thermocouple or pyrometer operation up to 2800°C
Exchangeable sample holders and For 2800°C furnace: Graphite
push rods For 2400°C furnace: Graphite, glassy carbon, alumina (<1680°C)
Sample containers for pastes, liquids and molten metals made of
Special sample holder systems
fused silica, alumina and sapphire
Accessories Sample supports and protective sleeves
* The 2800°C furnace is operated without protective tube
19
Dilatometer Accessories
Dilatometry Accessories
20
getter ring
carrier
Grinding machines for the prepa- For calibration of the expansion, manu- The presence of residual oxygen can
ration of plan-parallel samples and facturer-certified standards are available. be critical in certain applications (e.g.,
vernier calipers for the online input of These come in different lengths and ma- metals, metal alloys) because possible
sample length can be ordered with the terials (fused silica, alumina, sapphire, oxidation of the sample would lead to
instrument. steel, platinum, tungsten, and POCO undesired results and false interpre-
graphite). tations. The OTS® system allows for
All calibration routines are supported by effective reduction of the oxygen partial
the standard dilatometer software. pressure in the vicinity of the sample. A
ceramic substrate bearing a getter ring is
mounted on the sample carrier or in the
sample carrier tube. The residual oxygen
content remaining after evacuation is
thereby almost entirely eliminated
(< 1ppm).
Additional information
21
Dilatometer Software – Software Specialties
All dilatometers run under Proteus® software on a Windows® operating system. User-
friendly menus combined with automated routines make Proteus® very easy to use
while still providing sophisticated analysis. The Proteus® software is licensed with the
instrument and can of course be installed on other computer systems.
22
Key Features of the Measurement Software
23
Dilatometer Software – Software Specialties
24
Length/µm
98
94
90
Model: Three -Step Reaction A B C D
86 st
1 step: diffiusion
nd
2 step: n -th order
rd
82 3 step: n -th order
Investigation of the sintering behavior of Aragonite (CaCO3) on the base of a 3-step reaction model using the
Thermokinetics software
Software Specialties
25
Performance and Low-Temperature Applications
Unmatched Reproducibility
1.60
1 Run (Heat)
st
1.00
room temperature and 1550°C. The
0.80 heating/cooling rates were 5 K/min.
0.60 The atmosphere was helium. The
comparison clearly demonstrates the
0.40
outstanding reproducibility of the DIL
0.20 402 C. As can be seen, the heating and
0.0 cooling data are almost identical. The
0 200 400 600 800 1000 1200 1400 1600
difference between the four test results
Temperature / °C
is typically less than 0.3%.
Thermal expansion of a sapphire sample during heating and cooling cycles.
2.25
Outstanding Accuracy
Platinum/
Copper 10 % Rhodium In this example, the linear thermal
1.75
Aluminum expansion behaviors of pure aluminum,
Thermal Expansion / %
Rhodium
copper and electrolytic iron are
1.25 Platinum
compared with literature values. Clearly
Iron there is excellent agreement. The differ-
0.75 ences are generally less than 1%. In
addition, Pt/10%Rh was compared with
Literature values
0.25 (error bars: 3 %) the values for platinum and rhodium.
Iron The test results are in the expected
-0.25 range between the pure metals.
-250 0 250 500 750 1000 1250 1500 1750
Temperature / °C
2
by SCHOTT, Germany. It is designed for
Temp. / °C T. Alpha / 1/K zero thermal expansion around room
0.0 ... 50.0 0.014E-06
0 temperature. This material is often
Temp. / °C T. Alpha / 1/K
0.0 ... 50.0 0.011E-06 used for high-performance terrestrial
-2 telescopes. The figure shows the linear
Literature: Temp./ °C T. Alpha / 1/K thermal expansion between -150°C
0.0 ... 50.0 0.00 +0.05E-06
-4
and 100°C. The sample was measured
twice at a heating rate of 3 K/min in a
-150 -100 -50 0 50 100 helium atmosphere. The measured CTEs
Temperature / °C
between 0°C and 50°C are in excellent
High-precision measurements in very low temperature ranges are easy with the agreement with the literature values
DIL 402 C – even for materials with zero thermal expansion. (SCHOTT brochure) for this material.
26
Fiber-Reinforced Polymer
Roof Tiles
0.10
-1.8 °C
A fired roof tile was exposed to water
for 24 hours and then tested with the 0.05 Heat
DIL 402 C between -20°C and 25°C.
Thermal Expansion / %
27
High-Temperature Applications
0.600
Coefficients of thermal expansion
Glass transition
temperature: (CTE), glass transition temperatures and
549,4°C
0.500 554,5°C
softening points are crucial parameters
554,7°C for characterization of glass materials.
0.400 This plot shows three tests on the same
Temperature CTE / 1/K
20 ... 300°C: 8.918E-06
type of glass but from different batches.
0.300 20 ... 300°C: 8,902E-06 It can clearly be seen that the coeffi-
20 ... 300°C: 8.910E-06
cients of the thermal expansion are
0.200
in good agreement within the instru-
ment’s uncertainty boundaries. The
0.100
glass transition temperatures and the
softening point of sample no. 3 (blue
0.000
100.0 200.0 300.0 400.0 500.0 600.0 curve) show lower values, indicating
Temperature / °C a slightly different composition. The
Glass samples of the same type but from different batches show slight deviations in composition which can thermal expansion tests were, of course,
be well observed in the softening point (DIL PC measurement). switched off automatically by the
softening point detection in order to
protect the system from contamination.
Brick Clay
Thermal Expansion / % Phys. Alpha 10-6 1/K
2
Brick and tile products contribute
400
1 878.8 °C
-0.26 %
Sintering significantly to the development and
98.3 °C 576.5 °C quality of construction projects all
0
200 over the world. This figure of a brick
-2.59 %
-1 Dehydration 587.3 °C clay measurement exhibits a step in
the thermal expansion curve (red) at
-2 0
118.7 °C 576°C which can clearly be seen in the
919.8 °C
-3 Quartz formation and -2.02 % physical α-curve as a peak at 587°C.
dehydration of clays 1305.9 °C
-200 This change in thermal expansion is
-4
1116.3 °C due to the overlapping dehydration of
-5 1250.9 °C clays and phase transition of quartz.
-400
Melting Above 800°C, sintering starts. During
-6
200 400 600 800 1000 1200 1400
the sintering process, a melting effect
Temperature / °C can be observed which could be due
to an inorganic component of the raw
Dehydration, quartz transition and sintering of a brick clay measured with the DIL 402 C material.
28
High Sample Throughput –
Fired Tiles with Different Expansion dL/Lo / %
Behavior 0.8
0.7
Two colored tiles were simultaneously
red tile
0.6
heated at 3 K/min to 1100°C using the 971.6 °C
DIL 402 CD. The curves for the relative 0.5
brown tile
expansion of these fired tiles show 0.4 567.0 °C
DIL 402 CD measurement on two different fired tiles. The dual sample arrangement ensures a high sample
throughput for routine applications.
29
High-Temperature Applications
30
Silicon Nitride
1200.6 °C
Because of its excellent thermal and 0 -1.05 % 0.10
mechanical properties, silicon nitride 1424.5 °C
-2 -0.0
is used increasingly often in high-tech
applications (e.g., valves in automobile -4 1278.9 °C
engines). Of course, the properties of -0.10
Thermal Expansion / %
1789.5 °C
Rate of Expansion / %
-6
the final parts are heavily influenced
by the production/sintering process. -0.20
-8 -14.33 %
Depicted in this plot are the thermal
-10 -0.30
expansion and rate of expansion of a
1511.0 °C
silicon nitride green body. The sintering -12
-0.40
step starting at 1201°C is due to the
influence of sintering additives. The -14 1760.3 °C
-0.50
major shrinkage step occurred at -16
1424°C (extrapolated onset). The effect
-0.60
above 1760°C is most probably due to 0 500 1000 1500 2000
the evaporation of additives. Temperature / °C
DIL 402 C allows measurements under well-defined atmospheres up to 2000°C. This allows observation of
the complete sintering behavior of many ceramic materials.
Natural Carbon
Thermal Expansion / %
The reduction of impurities in natural
2602 °C
carbon can be seen during heating
under a helium atmosphere. The 2.00
sintering behavior changes during
reduction, which can be observed 2165 °C
1.50
in the effects between 1150°C and
1600°C. After these effects, the thermal
expansion increases almost linearly up 1.00
to 2600°C. Reliable measurements can 1146 °C 1441 °C
be performed at the highest tempera-
tures with the DIL 402 E up to 2800°C. 0.05 1602 °C
The purest of gas atmospheres and highest of temperatures can be achieved with the DIL 402 E.
31
c-DTA® and Other Specialties – Deep Insight into Material Behavior
c-DTA® Feature
dL/dt /(%/min)
dL/Lo / % c-DTA / K
481.4 °C 894.1 °C exo
20 0.05
0.000
522.2 °C -0.891 %
10 0.00
296.2 °C
348.1 °C
-0.500 190.5 °C 517.3 °C
-0.05
0
218.3 °C
-1.000 273.0 °C 966.6 °C -0.10
-10
Oxidation and burnout
of organics Sintering and -0.15
-1.500
formation of the -20
cordierite phase
-0.20
-2.324 %
-2.000
-30
1412.8 °C -0.25
-2.500 -40
-0.30
356.8 °C
1178.4 °C -50
-3.000 -0.35
200.0 400.0 600.0 800.0 1000.0 1200.0 1400.0
Temperature / °C
Simulation of the production process of cordierite by using the DIL 402 PC in combination with c-DTA®
(sample length 24.83 mm, heating rate 5 K/min)
During the production of cordierite the organic additives are burnt out.
(an alumina-silica ceramic carrier for Using the c-DTA® software, the burnout
catalysts), various raw materials are of the additives can be observed as an
ground and mixed to form a green exothermic effect between 190°C and
body. This plot shows the simulation 525°C (blue curve). At high tempera-
of the firing process of a cordierite raw tures (> 890°C), sintering starts and the
material. Under oxidizing atmospheres, cordierite phase is formed.
32
Volumetric Expansion and Bulk Density of Powders, Pastes and Liquid Metals
Dilatometers are generally used for liquid metals can now be accurately
measurement of solids. However, determined. Knowledge of the thermo-
by using the liquid metal containers physical properties such as volumetric
(made of fused silica, graphite, alumina expansion and bulk density of metals
or sapphire) and a special software during melting is of paramount impor-
extension, the volumetric expansion and tance for the simulation of casting
bulk density of powders, pastes and processes using finite element models.
7.9
8 Sample holder: Sapphire
was employed for this test. The
Density Change / g/cm3
The NETZSCH Density Determination software allows calculation of the density as a function of
temperature based on the thermal expansion.
33
c-DTA® and Other Specialties – Deep Insight into Material Behavior
(extrapolated endset).
Created with NETZSCH Proteus software
Density /(g/cm^3) dV/Vo /%
The volumetric expansion and temper-
14.0
ature-dependent density change were Density /(g/cm^3) dV/Vo /%
calculated using the software module 7.800 12.0
14.0
Density Determination. During melting, Density Onset: 1331 °C
10.0
12.0
the sample experienced an expansion 7.800
7.600
step of more than 4% within a temper- Density Onset: 1331 °C 8.0
10.0
ature difference of 40°C. 7.600
7.400 6.0
8.0
7.400 4.0
6.0
7.200
Volume Onset: 1331 °C
2.0
4.0
7.200
7.000
Volume Onset: 1331 °C
0.0
2.0
34
Dilatometry under Humidity
35
The NETZSCH Group is an owner-managed, internationally operating technology
company headquartered in Germany.
The three Business Units – Analyzing & Testing, Grinding & Dispersing and Pumps &
Systems – provide tailored solutions for highest-level needs. Over 2,500 employees
at 130 sales and production centers in 23 countries across the globe guarantee that
expert service is never far from our customers.
NETZSCH-Gerätebau GmbH
Wittelsbacherstraße 42
95100 Selb
Germany
Tel.: +49 9287 881-0
Fax: +49 9287 881 505
at@netzsch.com www.netzsch.com