Find the Spectra-Tech Nic-Plan at our website: Click HERE


DRAFT Nicolet Nic-Plan IR Microscope DRAFT
• Never put papers or other items on the back ledge of the instrument cabinet. The
IR will quickly overheat.
• The microscope optics must be carefully aligned manually before use. Your
instructor will verify the alignment.
1) USE PROTECTIVE INSULATED GLOVES and SAFETY GOGGLES when handling
liquid nitrogen. Fill the microscope Dewar chamber with liquid N2. Open the flip
top, and slowly pour in liquid nitrogen from a 250 mL silvered Dewar. When first
beginning to fill, pour in only enough liquid nitrogen to fill the funnel (you will see
it start to bubble against the plastic screen), and then stop pouring for at least 60
seconds and let that amount drain down the funnel. Repeat this two or three more
times to fully cool down the funnel and tubing system. Then slowly pour the rest of
the liquid nitrogen into the funnel making sure to never overflow the funnel. It will
require about 750 mL (about three small Dewars) to fill the chamber. This should
allow the microscope to run for about 8 hours.

2) Turn on the main power switch (black rocker switch on the back of the microscope,
not the IR itself).

3) Boot the computer.

4) Launch the OMNIC software using the desktop icon.


[NOTE: the response time of the computer is very slow.]

3) Click Collect/Experiment Setup… (above), which will bring up the following


screen.

Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
4) Under the Collect Tab, click the Open button and select the method called
nicplan8 LJS that contains settings appropriate for the IR microscope.

5) Under the Bench Tab, verify these settings:


Sample compartment => Left µScope %T (transmission) or %R (reflection)
Detector => MCT-B
Beamsplitter => KBr
Source => IR
Accessory => None
Range => 4000-400 cm-1
Gain => Autogain

Verify the interferogram in the left-hand window, with an amplitude in


the range of about +/- 8 to 3 depending on what is in the microscope field.

6) If the interferogram is weak, the microscope optics are probably misaligned.


Consult your instructor.

Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
7) Click the OK button at the bottom of the Experiment Setup window.

8) Using the mechanical stage controls, move the sample slide to an area not
covered by the sample. To verify the field of view, press the View button on
the microscope base. For a substrate that reflects some light, turn up the
intensity on the upper lamp (right-hand roller knob at front base of
microscope). Through the eyepieces, you should see a blue rectangle
illuminating the substrate. Focus the microscope to get the clearest rectangle,
although it will not be super-sharp because the IR optics aren’t optimal for
visible light. When finished focusing, turn down the lamp intensity again.

To verify a clear path through the pinhole on the slide, turn up the intensity
on the substage lamp (left-hand roller knob) and look for the green rectangle.

When finished aligning the slide, always turn down both lamps.

9) From the Collect menu, select Collect Sample (see step 2). The instrument
will notify you that it is about to collect a background spectrum; click OK.

10) When finished collecting the background, the instrument will notify you to
insert your sample. Again push the View button and turn up the lamp
intensity. Turn the “Upper Aperture View” knob to the right of the eyepieces
to the down position. You will see a blue spot projected onto the slide surface.
Use the mechanical stage controls to move the sample under the microscope
objective so that the blue spot is on the sample area you wish to study.

11) When finished using the instrument, go to Collect/Experiment Setup…, click


the Bench Tab, and set the Source to OFF.

12) Turn off the main power switch (black rocker switch on the back of the
microscope, not the IR itself).

Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Artisan Technology Group is an independent supplier of quality pre-owned equipment

Gold-standard solutions We buy equipment Learn more!


Extend the life of your critical industrial, Planning to upgrade your current Visit us at artisantg.com for more info
commercial, and military systems with our equipment? Have surplus equipment taking on price quotes, drivers, technical