Professional Documents
Culture Documents
Prosedur T, K, Y
Prosedur T, K, Y
Prosedur T, K, Y
OPERATING PROCEDURE
WRITTEN PRACTICE
O01/ WP-TKY/AWS-NDT-IINDT
THIS AND ALL OPERATING PROCEDURES/STANDARD PRACTICE INSTRUCTIONS CONTAINED IN THE COMPANY QUALITY
ASSURANCE SYSTEM ARE CONTROLLED BY THE QUALITY ASSURANCE MANAGER AND MAY NOT BE AMENDED,
REVISED, OR IN ANY WAY ALTERED WITHOUT HIS AUTHORITY.
The signatures below indicate authorisation of all numbered pages in this document up to and including the date of authorisation.
06/03/10 0 Brillian Wiraprima Sardi Hendra Wijaya Hendra Wijaya Original Issued
CONTENTS
1. SCOPE
2. REFERENCING CODE
3. GENERAL METHOD
6. PERSONNEL QUALIFICATION
10. REPORTING
1. SCOPE
This procedure with Specific UT Reference Code is intended to be describing of Ultrasonic Examination of welds that will be carried
out by Indonesian Institute of NDT Ultrasonic Testing Personnel. This procedure is complying with the requirements /
recommendation of AWS D1.1 2008 edition.
2. REFERENCING CODE
2.2 ASNT SNT – TC 1A, 2006 Edition and NDT Written Practice for Qualification and Certification of NDT Personnel
2.7 IINDT Written Practice 001/OP – WP/NDT-IINDT NDT Personnel Qualification & Certification
3. GENERAL METHOD
3.1.1 All Examination shall be of pulse echo method. A – Scan presentation units with test frequency of 2.0 to 10.0 MHz shall be
used e.g. Models USM 35X, Sonatest Sitescan 150S or equivalent.
3.1.2 All welds and parent material shall be examined with direct contact method.
3.1.3 Probes shall be of either single or twin crystal, with polarized ceramics. The probe angle and beam exit point shall be clearly
marked on the angle probes. Place the probe toward to radius of IIW V1 block or V2 block, maximize the signal to 50% - 75%
FSH by moving the probe backward and forward parallel to the block sides. When the signal is at maximum, the probe index
will correspond to the engraved line on the block which marks the geometrical centre of the quadrant.
The probe index measurement should be repeatable to within +/- 1mm. If the measured position differs from the existing mark
by more than 1mm, the new position should be marked on the probe side and shall be used subsequent probe check and
defect plotting.
The refraction angle shall be 45º to 80º within material.
3.1.4 Suppression/ Reject of the CRT time base will not be permitted during examination. (i.e. suppression / reject control shall be
turned – off ).
3.1.5 Electronic alarm and gate system will not be utilized in examination of structural welded joints.
3.1.6 Viewing hood over the CRT screen shall be used when examination is to be carried out in bright lighting condition.
3.1.7 Wallpaper paste, cellulose, starch or other water based couplant shall be used on site examination. Grease or oil may be use
for Calibration. Metylan will be used for T, K, Y joints.
3.1.8 The surface where scanning is to be carried out shall be free of scale, weld spatters, pitting, thick coating and other
irregularities which may interfere the transmission of sound wave. Uniform coating of paint may be accommodated by the use
of transfer correction (AWS D.1.1. Sect. 6) but tight layers of paint shall not exceeding 10 mils (0.25 mm) in thickness.
As a minimum the following equipment accessories shall be available to all technicians to fulfill the requirement of this procedure.
1,5 to 2 meter – long with suitable connection to fit apparatus and probes to be used, i.e. lemo to lemo, lemo to BNC, or lemo
to microdot / subvis connecting.
The pulse echo instrument shall be serviced and calibrated at regular intervals at least once in six months. Performance check of
Ultrasonic test units and probes as per AWS D1.1 Clause 6.29 and 6.30 shall be carried out periodically for variation in functions.
Minimum frequency of calibration and standardization are listed below:
The UT Equipment shall have been calibrated by independent laboratory, and it will calibrated by the inspector, since less
than six months.
Beam Exit Points and probe refracted angel will be Initial calibration and prior to testing of each individual component. Internal
Reflectors will be Initial calibration, after every change of shoe and weekly thereafter.
Calibration performed after determine the range (for example 50, 75, 100 …etc) selection. Place probe on side 25 mm
thickness of V1 block and adjust until back wall indication on exactly time base according to selected range. Set first back wall
echo indication 50 - 75 % FSH. Indication so completed is to be after re-checking by 50 mm Perspex or 100 mm Thickness of
V1 block.
Calibration performed after determine the range (for example 100, 125, 150 …etc) selection. Direct angle beam (45, 60, 70)
toward to radius 25 mm of V2 block. Adjust until reflection indication from radius V2 on exactly time base according to
selected range. Calibration is completed after re-checked by radius 50 mm of V2 block or radius 100 mm of V1 block.
5.2.3 Distance Amplitude Correction use IOW A5 block ( see figure 3 in appendix I )
Position the probe for maximum response from 13 mm. Adjust the gain control to provide an 50 - 75 % of full screen indication
(100% DAC) from the hole, mark the peak of the indication on the screen. Do not change dB Sensitivity to construct 50% (6
dB Drop) DAC, 10% (20 dB Drop) DAC. Position the probe for maximum response from 19 mm, 25 mm, 32 mm hole distance.
Mark the peak every of indication from above hole to the screen. Connect the screen marks for the side drilled holes 13 mm,
19 mm, 25 mm, and 32 mm to provide the 100%, 50% and 10% DAC curve.
The DAC curve shall be re-constructed if a point on the DAC has decreased or increased or increased 20 % or 2 dB of its
amplitude.
6. PERSONNEL QUALIFICATION.
6.1 Personnel performing the ultrasonic testing shall be qualified in compliance/ in accordance with ASNT-TC-1A (2006 Edition
personnel qualification and certification in Non Destructive Testing) in Ultrasonic Testing Level II
6.2 Only certified and qualified level II personnel in tubular T, K, and Y Connections shall perform the testing independently for
AWS T.K.Y Connection.
6.3 PCN personnel certification in ultrasonic testing scheme for weldment inspection grade 2 or level II as appropriate groups 3.1
to 3.9 qualified.
7.1.1 Scanning shall be of zigzag pattern with a minimum interval of 10% of the 20dB isobar to ensure full coverage,
scanning speed 4 in/Sec overlap scanning is half of diameter probe with scanning movement A/B/C/D/E.
7.1.2 Sensitivity setting shall be that dB value whereby the first back wall echo off the material reaches 100% full screen
o
height. 0 compression wave probes within the range of 2.0 to 2.5 MHz frequencies shall be used. For thickness
measurement and base metal evaluation of materials below 20mm thickness, sound frequency of 4.0 to 5.0 MHz
may be used.
7.2.2 DAC curve shall be constructed for each angle probes to be used for AWS D1.1, Sect. 6 evaluation, side drilled hole
from Reference standard sensitivity block or IIW-V1 block shall be used with the reference level set to 50% to 75%
full screen height.
7.2.3 Transfer correction shall be constructed for each angle probes to be used, to compensate difference between
surface finish, contour, coating and acoustical attenuation characteristics, for AWS D1.1, Sect.6. Any surface
irregularities shall not be allowed for evaluation of weld.
7.2.4 Weld examination shall be as per the appended technique sheets, unless further investigation is deemed warranted
by the UT technicians.
7.2.5 Any restriction or limitation encountered that are not covered or dictated by a specific UT technique sheet
attachment shall be reported with a detailed technique sheet illustrating a scaled drawing and scanning pattern
used. Ranges utilized shall also be clearly detailed for each combination of angle probes and scanning pattern.
o
7.2.6 Scanning for longitudinal discontinuity shall be carried out with the angle probe at approximately 90 to the weld axis
from two directions where possible. Move the search unit laterally in a zigzag pattern with an interval of 10% within
20 dB isobars with scanning movement A/B.
Scanning for transverse discontinuity shall be carried out with the angle probe at approximately 15° to the weld axis
moving back and forward in a zigzag pattern with an interval of 10% within 20 dB isobars + scanning movement
pattern “D”. For SAW process or ground flush weld cap, scanning for transverse discontinuity may be carried out
directly over the cap with the angle probe aligned to the weld axis moving forward towards both directions i.e. 180°
apart with 45° probe.
7.2.7 Discontinuities shall be confirmed and position within the weld cross-section by establishing beam path length and
skip distance from a reference point (i.e. cap toe). Plotting of discontinuity may be ailed by comb type mimic / profile
gauges and trigonometrically computation (for constant configuration only).
Additional scanning pattern (i.e. transversal, lateral, orbital and rotational) will be carried out for the assessment and
evaluation of discontinuities detected.
7.2.8 Linear extremities of the discontinuity shall be determined by 6 dB / half intensity amplitude drop.
7.2.9 Cross sectional height of the continuity shall be determined initially by amplitude comparison to establish immediate
status. Further investigation for identity and recording purposes shall be carried out with the beam boundary
intercept (i.e. 20 dB drop) or maximum amplitude technique or combination of both.
7.2.10 Indications shall be sized using those probes that receive the optimum response. As necessary, indications will be
plotted with more than one angle to confirm location and type of discontinuity.
8.1 All measurements greater than 150 mm around or along the welds shall be carried out using soft non-stretchable tailor tapes.
For measurements less than 150 mm, 150 mm stainless steel precision rules shall be used.
8.2 All measurements shall be from an established reference point i.e. toeing centerline for TKY joints and end of the weld for all
other configurations.
8.3 Reference „O‟ datum and location of defect for repair within the weld axis shall be marked directly over the surface of the weld
and parent metal with indelible metal marker. A simple pictorial drawing representing the weld cross section indicating defect
orientation and depth will be sketched, along with brief instruction as appropriate, adjacent and arrowed to the defective weld
areas.
9.1 Repaired areas shall be re-examined taking into consideration the new fusion zone created by the excavation and the position
of the original defect.
9.2 The whole length of the repaired areas shall be re-examined plus 50 mm from both ends where possible.
10. REPORTING
All indications producing a signal of 50% full screen height or greater, shall be recorded and portrayed in full scale cross section,
except for AWS D1.1 Sect. 6, where only rejectable indications shall be reported. UT Report front sheet shall clearly identify status of
weld / joint tested.
Acceptance criteria shall be as accordance with AWS D1.1 class R for all kinds of joints.
T.K.Y tubular full penetration using API RP 2X class C and for back up weld evaluated using AWS D1.1 class X.
Appendix I
V1 Block
Figure 1
Appendix I
V2 Block
Figure 2
Appendix I
IOW A5 Block
Figure 3
Appendix I
Figure 4
Appendix II
1.0 GENERAL
1.1 A compression wave probe (single or twin crystal) and IIW V 1 block can be used for this check.
1.2 Range of check shall be at least equal to that which is to be used in subsequent testing.
2.0 PROCEDURE
2.1 Place the compression wave probe on position A, B, or C of IIW V 1 block. Adjust the time base so that the first and
fifth back wall echo indications coincide with particular graticule line.
2.2 Bring successive back wall echoes, in turn, to approximately the same height (e.g. 50% - 75% full screen). The leading
edge of each echo should line up with the appropriate graticule line. Record any deviation measured at approximately
half full screen height, from the ideal position.
2.3 For range less than 250 mm, place the probe etc. For range greater than 250 mm, place probe at A or B
3.0 TOLERANCE
Deviation of the base line linearity shall not exceed +/- 2 % of the full time base range, otherwise re-calibration is required.
The check shall be carried out at least once per week if the unit is used daily or prior to production testing.
A
B
Appendix III
Acceptance Criteria
Appendix III
Acceptance Criteria
Appendix III
Acceptance Criteria
b. These criteria are specifically tailored for examination of welds made from the outside only in tubular T, K,
and Y connections and for practical examination situations in which rejectable defects are subject to visual
confirmation during the weld removal and repair process.
c. Discontinuities are evaluated using a combination of beam boundary techniques and amplitude calibration
utilizing the transfer and distance amplitude corrections addressed in this recommended practice. The
reference level calibration for root reflectors is the 1/16-inch (1.6-millimeter) notches shown in the
reference standard (block) for Level C examination (Figure 46– API RP 2X, 2004). The reference level for
internal reflectors is the side of the 1/16-inch (1.6-millimeter) hole. Evaluation of a discontinuity is to be
based on the ultrasonic operator’s best determination of actual flaw dimensions, bearing in mind that
interior planar flaws which are poorly oriented with respect to the ultrasonic beam may return a smaller
echo than a surface (root) notch of the same size. Flaw size is to be given in terms of width and length as
illustrated in Figure 47 – API RP 2X: 2004. In addition, the reflector shall be classified as spherical,
cylindrical, or planar and its position within the weld cross section is accurately determined. Identification
as to type (such as, crack versus incomplete fusion) is not required.
d. All reflectors 6 decibels smaller than the reference level (that is, 50 percent DAC) may be disregarded
(DRL = disregard level). Reflectors above the DRL shall be further evaluated as described in the following
sections. The operator may elect to decrease the DRL based on the criticality of a particular component.
f. Clustered multiple reflectors with indications above the 50 percent DAC curve and two times the branch
thickness in length should be evaluated in terms of their encompassing width and length using the limits of
Figures 45 and 48– API RP 2X:2004.
g. Cylindrical or planar reflectors whose dimensions exceed the limits of Figures 45 and 48– API RP 2X:2004
(depending on location) are rejectable.
h. Internal reflectors near the fusion line of a weld shall be evaluated with various probe angles while making
an effort to obtain a beam path perpendicular to the fusion line.
i. Reflectors within the base metal of the main member (joint can) at welds shall be evaluated as described
in the following. In addition, any such reflector that exceeds the disregard level shall be reported to the
operator:
1. Individual reflectors exceeding the limits of Figure 45 – API RP 2X: 2004 are rejectable.
2. Accumulated reflectors exceeding 8 percent of the area under the weld in any 6-inch (150-millimeter)
or D/2length (whichever is less) are rejectable.
3. Rejectable base metal reflectors shall be reviewed by the operator prior to any excavation or attempted
repair. Consideration should be given to the risk of causing further problems (such as lamellar tearing)
with the additional thermal strain cycles such excavation and repair would entail.
j. Note that for welds made from one side only in tubular T, K, and Y connections, greatly relaxed criteria are
applied to the root area. Fatigue tests on tubular joints with gross root defects have shown that these do
not control failure as long as there is a sufficient weld area to keep the overall stresses from being
significantly affected. Not only are defects here less serious, but attempted repairs often creating a less
desirable condition.
k. Tighter acceptance criteria for all other weld areas still permit somewhat larger discontinuities than
traditional radiography- based codes—yet they are consistent with the proposed applications and do not
lead to any significant degradation in performance for class C welds. With appropriate overmatching of
weld tensile strength compared to base metal yield, static weld failure will not occur due to the minor loss
of area permitted.
i. A reasonable level of fracture toughness is provided by selecting base metal and qualifying welding
procedures to provide as-fabricated notch toughness in weld, parent material, and heat-affected zones
meeting the requirements of API Recommended Practices 2A-WSD or 2A-LRFD.
Appendix IV
TECHNIQUE DETAILS
1.0 INSTRUMENTATION
Flaw Detector Make: Sonatest or equiv. Model: Sitescan 150S or equiv.
Probes: Type Sonates Sonatest
Angle: 0 45 - 60 - 70° (see special instruction b)
Crystal Size: 10mm dia. 10mm Ø
Frequency: 4 / 5 MHz 2 / 2.5 MHz
Couplant: Polycell or equivalent
3.0 CALIBRATION
Time Base Range(s) IIW V1 or IIW V2 Calibration blocks. See special instruction (e).
Sensitivity: Echo off 1.5 mm Ø sdh from IIW V1 set to 50% ~75% FSH. See special Instruction (d)
Reference Block Material: IIW V1 Hole Size: 1.5 mm Ø SDH
Compensation for Attenuation: See special instruction (d)
Appendix IV
SPECIAL INSTRUCTIONS
(a) Twin crystal 0° probes are to be used on thickness less than 40mm.
(c) Probe testing angle and scanning shall be in accordance with those shown in Table 6.6 of AWS D1.1
(d) Indication Rating, “d”, shall be expressed as:
Where:
b – Reference level ; echo off 1.5 mm Ø side drilled hole in IIW V1 block set to 50% ~ 80% full screen
height. (dB)
a – Indication level ; the maximum attainable indication from the discontinuity adjusted to the reference
level. (dB)
c – Attenuation factor ; attained by subtracting one (1) inch from the sound path distance of the probe to the
discontinuity and multiplying the remainder by 2.
(Fractional values less than ½ dB shall be rounded out to the lower level and those of
½ or greater to be rounded out to higher level).
(e) Range calibration shall include the entire sound path distance to be used during the specific testing.
(f) Operators to be qualified as ASNT Level II or PCN equivalent.
(g) Figs. 1 - 3 present a number of scan diagrams for typical access situations.
Appendix IV
FIGURE 1
Face
If Accessible
Face
Face
Appendix IV
SKETCH
FIGURE 2
Face
If Accessible
Face
Appendix IV
SKETCH
FIGURE 3
Face
Face
TEST RESTRICTIONS :
Appendix V
TECHNIQUE DETAILS
1.0 INSTRUMENTATION
Flaw Detector Make: Sonates or equiv. Model: Sitescan 150S or equiv.
Probes: Type Sonatest Sonatest
Angle: 0 45 - 60 - 70° (see special instruction b)
Crystal Size: 10mm dia. 10mm Ø
Frequency: 4 / 5 MHz 4 MHz
Couplant: metylant or equivalent
3.0 CALIBRATION
Time Base Range(s) IIW V1 or IIW V2 blocks
Sensitivity: SDH From IIW V1 set to 75% FSH or Standard sensitivity block. See special Instruction (d)
Reference Block Material: IIW V1, IOW Standard sensitivity block Hole Size: 1.6 mm Ø SDH
Transfer Correction: See special instruction (e)
Determine Beam Spread an angle probe: See special instruction (f)
Determine Resolution: See special instruction (g)
Horizontal / Time Base Linearity: See appendix II
Appendix V
SPECIAL INSTRUCTIONS
(a) Twin crystal 0° probes are to be used on thickness less than 40mm.
Appendix V
(f) The beam spread at the 20 dB limit will be determine for each probe by the use of IOW A5 block. This beam spread can
be used for angle probe when sizing defect by the “20 dB drop” method.
Plot 20 dB beam profile for 45 deg, 60 deg 70 deg probe. With 70 deg probe, a 20 dB beam spread is so wide and
difficult to construct as to be almost useless.
Place the probe on the block at 13 mm as above, and maximize the signal to 50 – 75% Full Screen Height.
Mark where the index point of probe coincide on the block. Move the probe forward until the signal drops to 1/10 screen
height (20 dB drop). Mark on the block where the index point of probes stands on the block. Move the probe backwards
until the signal maximizes and then drop down to 1/10 screen height. Mark index point of probe on the block.
Draw a vertical line on the block from the hole to the upper edge. Measure stand off of the three index point marks and
note the down.
Repeat the process using the 19 mm, 25 mm, and 32 mm as applicable. Now draw horizontal lines on a beam plotting
charts at depths of 13 mm, 19 mm, 25 mm, 32 mm as applicable. Transfer the stand off of the index point
recorded in the above procedure to relevant lines on the chart. Join the mark up.
The centre line represents the main energy of the beam and the other two marks represent the leading and trailing
edges of the beam. Draw in the parallel near zone which is equal to D 2/4 on the beam plotting chart. Where D =
effective Diameter of probe crystal. For 70 deg probe at 10 dB beam spread, instead of dropping the signal to 1/10
FSH, use the 3/10 screen height. For all probe movements, avoid angular rotation. Straight edge or ruler can be
used as a guide.
Position for maximum echo
13 mm
Y 19 mm
X 25 mm
32 mm
(g) This performs to check the ability of probe unit combination to discriminate between signals from separate reflectors
lying at only slightly different ranges or near to surface.
Compression probe.
Place the probe at position A. Three signals from the 85 mm, 91 mm, and 100 mm surface Should be clearly displayed.
Position B and C is to determine the near surface resolution or the dead zone of the Probe. If the reflection from the
Perspex disc at 10 mm (dead zone) would not be detected.
Appendix V
C
8
5
m
m9
1
mm
1
0
0m
m
A B
D
1
9
m
m
2
5
m
m
3
2
m
m
The signal amplitude of subsequent signals should be clearly separated as indicated on the sketch below:
H H
Appendix V
(h) Fig. 1 Present a number of scan diagrams for typical access situations.
Appendix V
SKETCH
FIGURE 1
ss
ce
No Access
Ac
No
SECTION "A"
IF ACCESSIBLE
C A
B
e
ob
Pr
LE
ce
ss
gle CCE
SSIB
o Ac An IF A
N
b e
P ro SECTION "B"
o
0
TEST RESTRICTIONS – Scanning from external surface of the branch member only.