TECH-TIP - Ensuring Accurate Turbidity Measurements


through Turbidimeter Calibration and Verification

 December 9, 2017
Partnership Resources

Turbidity is a water quality monitoring parameter that is key to determining the amount of particulate
matter present in a water sample.  Lower filter effluent turbidity values have been shown to correlate with
a reduced risk of pathogens, such as Cryptosporidium and Giardia, passing through the filters to the
finished water.  Turbidity is also the primary water quality optimization parameter for the Partnership for
Safe Water’s treatment plant optimization program, and is a critical monitoring and process control
parameter for plants at all phases of the program.  Turbidity values can drive plant operational activities,
such as coagulation and filter backwash that are essential to producing a safe, high quality water.
The AWWA website uses cookies to provide you with an improved experience. By continuing to
OK
Because ofyou
use this site, theagree
critical nature
to our ofcookies.
use of turbidity
Tomeasurements andhow
learn more, including thetouse of turbidity
manage cookie as a key process control
parameter,
settings it isbrowser,
in your important to ensure
read our Cookie that
Policythat turbidity data collected by plant instrumentation and staff is
as accurate as possible.  This can be achieved in a variety of manners.  This article focuses on the use of
calibration, verification, and sampling techniques to help ensure the accuracy of turbidity data.  

Calibration
The USEPA requires that turbidimeters are calibrated on a quarterly basis.  The regulatory requirements
for turbidimeter calibration may vary in other areas, but it is important to note that turbidimeters may be
calibrated more frequently than required by regulation.  They may not be calibrated on a less frequent
basis.  

Turbidimeters are typically calibrated with a primary turbidity standard, such as formazin or a stabilized
form of the formazin standard.  Both types of primary standards are commercially available.  Different
instruments and manufacturers may recommend different ranges of standards and techniques for
turbidimeter calibration, so it is important to follow the calibration instructions for a given instrument, as
provided by the manufacturer.  Typically, online turbidimeters with a measurement range of 0-100 NTU will
be calibrated with a 20 NTU standard solution.  Although 20 NTU is a high value in comparison with
typical filter effluent turbidity measurements, calibration at this level minimizes the potential impact of
any error on the accuracy of the calibration process and subsequent measurements.  More sensitive laser
nephelometers are typically calibrated with lower level standards, according to manufacturer
instructions.  Benchtop turbidimeters are typically calibrated with a range of standards reflecting the
measurement range of the instrument.   

In addition to quarterly calibration, instruments should also typically be calibrated after performing
significant maintenance, such as changing the light source or cleaning the instrument body.  These
activities can impact the amount of light reaching the instrument’s detector, and therefore impact the
accuracy of turbidity measurements.  

Calibrations may be performed by plant operators, instrumentation technicians, or the manufacturer’s


service representative – the most important thing is that the instruments are calibrated properly at the
frequency required by the regulatory agency.  Calibration records should also be maintained according to
regulatory requirements.  A standard operating procedure (SOP) for instrument calibration is highly
recommended so that all staff calibrate instruments in a consistent manner.     

Verification
The accuracy of turbidimeter readings should also be periodically verified.  Many utilities will verify online
and benchtop turbidity measurements on a weekly or monthly basis.  Turbidity readings can be verified in
a variety of manners including use of a primary turbidity standard, use of a secondary turbidity standard,
or by comparison with another instrument.  To verify turbidimeter readings using a primary standard, a
volume of primary standard with a known turbidity value may be poured into a sample cell and read on a
benchtop instrument or poured into a calibration cylinder/chamber to be read using an online instrument. 
The instrument reading is then compared with the standard’s value.  Typically, if the instrument reading
matches the value of the standard within a given percent or NTU value (for example +/- 10% or 0.05 NTU
– this may vary depending regulatory requirements), the instrument can be considered to be reading
accurately.  Although calibration may be performed at higher turbidity levels, verification may be
performed closer to the range of sample measurements, and stabilized low-level turbidity standards are
generally commercially available.
The AWWA website uses cookies to provide you with an improved experience. By continuing to
use this site, you agree to our use of cookies. To learn more, including how to manage cookie
Benchtop andbrowser,
settings in your online instruments may
read our Cookie also be verified using secondary standards.  Secondary standards
Policy
for benchtop turbidimeters typically consist of gel-based standards.  Secondary standards for the
verification of online instruments may vary in design depending on the manufacturer but are typically
consist of a set of filters designed to represent various turbidity readings.  It is important to follow the
manufacturer’s instructions for the type of secondary standard and instrument being used.  Most
secondary standards will be read and assigned a value directly after completion of a primary standard
calibration.  The secondary standard is then measured periodically, between calibrations, and its value
compared with the assigned value.  A significant increase in deviation may indicate the need to recalibrate
the instrument.  Secondary standard verification can be a convenient way to ensure accuracy since it
does not require the use of a chemical or liquid reagent.  

Online turbidimeter performance may also be verified by comparison with a grab sample that is measured
in the lab.  If a grab sample comparison is used to verify turbidimeter performance, it is important that the
sample collected for lab measurement is representative of the sample measured by the online
turbidimeter.  The optimal place to collect this sample is from the turbidimeter drain.  Use caution when
collecting samples from laboratory sinks, particularly if they are located a long distance from the online
turbidimeter, as a significant lag time can mean that the sample collected in the lab is not representative
of that being measured by the online turbidimeter.  When verifying filter effluent turbidity samples at very
low NTU values, it is essential to use good lab technique for low-level turbidity samples.  For example;
ensure sample cells are excruciatingly clean, use silicone oil on the outside of the cell to mask minor
scratches, avoid introducing air bubbles to the sample, and minimize the impact of condensation on the
cell.  Standard should be set by the plant to determine the level of agreement between lab and online
turbidity values that is considered to be acceptable.    

If an instrument’s 4-20mA output is used to communicate with the plant’s SCADA system, the accuracy of
the instrument output and SCADA value should also be periodically verified.  This ensures that an
accurate reading is relayed to the plant’s SCADA system - which is important because it may be used for
process control.  Verifying this can be achieved by comparing the instrument local turbidity reading with
that displayed on the SCADA computer.  If the values do not agree, further troubleshooting may be
required.

No matter how turbidimeter verification is achieved, SOPs are recommended to ensure that verification
procedures are being carried out in a consistent manner among all plant staff.

Turbidity Sampling
The accuracy of turbidity readings is also dependent on sampling technique.  To obtain an accurate
measurement, the sample must be representative of the water being measured.  For online instruments,
this equates to installing turbidimeters as close as possible to the sample being measured.  For example,
consider locating turbidimeters directly in the filter gallery, rather than in a laboratory location that may be
located a significant distance from the sample.  Ensure that samples are collected from an area of the
pipe free from sediment buildup, and be sure that the flow rate provided to online instruments is
consistent with the manufacturer’s specifications (an air gap on the drain can be helpful for providing a
quick visual confirmation of flow).  

Although many online turbidimeters incorporate internal bubble traps in their design, a separate external
bubble trap may be required for samples that contain an excessive amount of entrained air, in order to
The AWWA website uses cookies to provide you with an improved experience. By continuing to
avoid
use thisair bubbles
site, causing
you agree a positive
to our use interference
of cookies. in the
To learn more, turbidity
including howmeasurements. 
to manage cookie It is also important to
ensure
settings that the
in your turbidimeter
browser, read ouritself is Policy
Cookie suitable for the application and expected turbidity levels.  For
example, a filter effluent turbidimeter may not be suitable for monitoring raw water turbidity, due to
limitations in range and the potential for maintenance issues caused by high levels of suspended solids
introduced into the turbidimeter body.

Sampling is also important for benchtop turbidity measurements.  As described previously, when
collecting grab samples or online instrument calibration, be sure that they samples are representative of
the water being measured by the online turbidimeter.  Also, when collecting samples for laboratory
analysis, consider collecting them in a clean sample cell to minimize any potential sources of
contamination as well as the introduction of air bubbles caused by sample transfer.  Finally, as with many
other parameters, it is recommended to measure turbidity sample as soon as possible after sample
collection.  If samples are to be stored for short durations of time, be sure to follow the guidance provided
in Standard Methods for sample storage techniques to minimize the impact of microbial activity on the
turbidity value.

Establishing consistent and accurate calibration, verification, and sampling techniques can help ensure
the accuracy of plant turbidity measurements.  At the heart of the Partnership for Safe Water’s treatment
plant optimization program, turbidity is an essential part of determining a plant’s optimization status,
quantifying plant performance, and the plant’s process control system.  Ensuring accurate turbidity
measurements helps to provide better data for driving plant decision making and – ultimately – achieving
optimization.
 

 
Advertisement

 Knowledge

Journal AWWA
Opflow
AWWA Water Science
Press Room
Policy Statements
Scholarships
Sourcebook
Partnership for Safe Water
Partnership for Clean Water

 Get Involved
Join AWWA
The AWWA website uses cookies to provide you with an improved experience. By continuing to
Renew Membership
use this site, you agree to our use of cookies. To learn more, including how to manage cookie
My Account
settings in your browser, read our Cookie Policy
Committees on Teams
Advertise with Us

 Jobs
Find a Job
Post a Job
Work for AWWA

 Follow Us

    
 Contact
AWWA Headquarters
6666 W. Quincy Ave.
Denver, CO 80235 USA
Phone: 303.794.7711 or 800.926.7337
Fax: 303.347.0804

AWWA Government Affairs Office


1300 Eye St. NW Suite 701
Washington, DC 20005-3314 USA
Phone: 202.628.8303

Contact AWWA

Copyright © 2023 American Water Works Association Privacy Policy Terms of Use Sitemap Cookie Policy
Login

The AWWA website uses cookies to provide you with an improved experience. By continuing to
use this site, you agree to our use of cookies. To learn more, including how to manage cookie
settings in your browser, read our Cookie Policy

You might also like