Professional Documents
Culture Documents
10 1016@j Aca 2020 01 024
10 1016@j Aca 2020 01 024
PII: S0003-2670(20)30059-3
DOI: https://doi.org/10.1016/j.aca.2020.01.024
Reference: ACA 237383
Please cite this article as: M. Karami, Y. Yamini, On-disc electromembrane extraction-dispersive
liquid-liquid microextraction: A fast and effective method for extraction and determination of ionic
target analytes from complex biofluids by GC/MS, Analytica Chimica Acta, https://doi.org/10.1016/
j.aca.2020.01.024.
This is a PDF file of an article that has undergone enhancements after acceptance, such as the addition
of a cover page and metadata, and formatting for readability, but it is not yet the definitive version of
record. This version will undergo additional copyediting, typesetting and review before it is published
in its final form, but we are providing this version to give early visibility of the article. Please note that,
during the production process, errors may be discovered which could affect the content, and all legal
disclaimers that apply to the journal pertain.
Department of Chemistry, Tarbiat Modares University, P.O. Box 14115-175, Tehran, Iran
1 ∗
Corresponding author: Tel.: +98 21 82883449; Fax: +98 21 82883460.
E-mail address: yyamini@modares.ac.ir (Y. Yamini).
1
Abstract
(EME-DLLME) was performed using a lab-on-a-disc device. It was used for sample
as model analytes in biofluids. The disc consisted of six extraction units for six parallel
extractions. First, 100 µL of a biofluid was used to extract the analytes by the drop-to-drop
EME to clean-up the sample. The extraction then was followed by applying the DLLME
method to preconcentrate the analytes and make them ready for being analyzed by gas
some significant advantages over the conventional methods, including saving space, cost, and
materials as well as low sample and energy consumption. In the designed device, centrifugal
force was used to move the fluids in the disc. Both sample preparation methods were
performed on the same disc without manual transference of the donor phases for doing the
two methods. Scalable centrifugal force made it possible to adjust the injection speed of the
organic solvent into the aqueous solution in the DLLME step by changing the spin speed.
Spin speed of 100 rpm was used in dispersion step and spin speed of 3500 rpm was used to
sediment organic phase in DLLME step. The proposed device provides effective and
reproducible extraction using a low volume of the sample solution. After optimization of the
determination of amitriptyline and imipramine in saliva, urine, and blood plasma samples.
The method provides extraction recoveries and preconcentration factors in the range of 43%-
70.8% and 21.5-35.5 respectively. The detection limits less than 0.5 µg L-1 with the relative
standard deviations of the analysis which were found in the range of 1.9%-3.5% (n = 5). The
method is suitable for drug monitoring and analyzing biofluids containing low levels of the
model analytes.
2
Keywords: Electromembrane extraction; Dispersive liquid-liquid extraction; Lab-on-a-Disc;
1. Introduction
Liquid-liquid extraction (LLE) is the most extensively used sample preparation technique
in chemistry and biochemistry [1]. Despite its widespread use, traditional LLE suffers from
consumption of large quantities of toxic organic solvents and being tedious as well as time
consuming. Due to global concerns about environmental pollution; the need to develop
Attempts in the field of analytical chemistry to develop methods using organic solvents as
(LPME). LPME methods also give rise to minimizing the overall extraction time [2, 3].
methods introduced in 2006 by Rezaee et al. [4]. Rapid injection of an appropriate mixture of
extraction and disperser solvents into an aqueous donor phase containing the desired
analyte/analytes makes a cloudy solution and a multitude of fine droplets of the extraction
solvent forms in the solution. Formation of these fine droplets is the most significant
advantage of DLLME, which causes the extraction to be obtained very fast [5]. Moreover, the
method is compatible with most analytical instruments. The features of being fast and capable
of reaching high preconcentration factors make DLLME receive analytical chemists’ interests
[5]. However, DLLME suffers from some disadvantages. Perhaps the most serious
disadvantage of the method is the lack of sample clean-up because of which this method can
migration of ionized analytes from an aqueous donor phase across a supported liquid
3
membrane to an acceptor solution. Since it was reported in 2006 [6], a considerable amount
of literature has been published on application of EME [7, 8]. Since EME is a membrane
based extraction, it provides a good sample clean-up, which makes this method capable of
The acceptor solution in EME is mostly an aqueous solution. Thus the main
While GC is faster, cheaper, and simpler than liquid chromatography (LC) and can be easily
conjugated with different kinds of detectors such as flame ionization detector (FID) and mass
spectrometer (MS). Some attempts have been made so far to transfer the extracted analytes of
EME to a GC-compatible phase [9-14]. DLLME is the method which can be properly
combined with EME. Combination of EME with DLLME provides a two-step extraction
method that benefits from great sample clean-up of EME as well as high preconcentration
factors of DLLME. Another advantage is that the extracts can be analyzed by GC. In 2012,
sample solution. [11]. The proposed method showed a good limit of detection and linearity
for extraction of trace levels of chlorophenols; however, the two-step extraction method with
manual transference of the donor phase can affect repeatability and precision of the extraction
and make the method tedious and hard to operate. Moreover, although the method provides
very good sample clean-up, it needs a large quantity of a donor phase. Therefore, it seems
that the method is not suitable for analysis of biofluids. Seidi et al. reported a method to
combine EME and DLLME to determine tricyclic antidepressants in biological samples [12].
They used 24 mL of a donor phase in an EME step with the acceptor solution volume of 10
µL.. However, as it was the case with the previously mentioned work, manual handling and
transference of phases can influence repeatability and precision of the method. Also using 24
4
mL of a donor phase still seems too much for some of biological fluids. Some other works
have been done to couple EME with DLLME [13, 14]. They had the same drawbacks as were
mentioned before.
General advantages of miniaturization including the capability to save time, space, cost,
and materials lead microfluidic systems to become a necessity in today’s world. Keeping
company with many areas of science, the last three decades have seen a growing interest in
miniaturization and automation of classical analytical operations. The main motivation for
such miniaturization is the development of micro total analysis systems (µTAS) or lab-on-a-
chip (LOC) technology. Furthermore, automation leads to the development of devices for
performing laboratory processes which are easy to handle without any need for a skillful
operator. To date plenty of attempts have been made to down-scale and implement LLE and
also LPME on microfluidic chip devices [15-25]. Research studies could successfully
platforms. There have been some reviews describing different microfluidic and on-chip LLEs
[26-30].
(LOAD), are one of the microfluidic platforms that provide specific advantages over other
on-chip devices [31-33]. LOADs use centrifugal force for flow control. They simply need a
rotary motor to create a centrifugal force. Therefore, there is no need for external pumps or
power supplies for liquid actuation, which can make a great potential for portability. Bubble-
free liquid handling without dead volume and scalable centrifugal forces for sedimentation
are some of the other advantages of such a platform. LOADs also offer closed fluidic systems
with the capability of parallel sample operation on the same disc [32]. To date several reports
have been published in the field of sample preparation using LOADs [34-36]. There are still
5
negligible reports using LOADs for implementation of LLE [37, 38] and using the inherent
In the present work, the on-disc EME-DLLME method has been introduced. The method
benefits from high clean-up ability of a drop-to-drop EME for extraction of target analytes
advantages over the conventional methods, including miniaturization and saving time, space,
and organic solvents. In addition, in the DLLME step, dispersion of an organic solvent
mixture into a donor phase and sedimentation of an extraction solvent can be performed
simply by changing the spin speed of the disc. In this work, we investigated the effect of the
injection speed of the organic solvent into the aqueous solution on extraction efficiency.
Finally, the applicability and performance of the system were demonstrated for human urine
2. Experimental
Amitriptyline (Ami) and imipramine (Imi) were kindly provided by Razak Laboratories
(Tehran, Iran). The structure of the drugs along with their physicochemical properties are
shown in Table1. NaOH, CCl4, and C2HCl3 were purchased from Merck (Darmstadt,
Germany). C2Cl4 and 2-nitrophenyl octylether (NPOE) were supplied from Acros (Geel,
Belgium) and Fluka (Buchs, Switzerland), respectively. Isopropanol was obtained from
Sigma-Aldrich (St. Louis, MO, USA). Methanol was purchased from Caledon (George
Town, Ont., Canada). All chemicals were of analytical reagent grade. Deionized water was
prepared by a Young Lin aquaMAx purification system 370 series (Seoul, Korea). The
6
Accurel PP 1E polypropylene membrane sheet with a wall thickness of 100 µm and a pore
Stock solutions of the drugs at 1 mg mL−1 as well as a standard solution of the mixture of
the drugs containing 50 µg mL−1 of each drug were prepared in methanol and stored in a
refrigerator at 4°C. The aqueous working solution was prepared daily by diluting the standard
solution at 0.5 µg mL−1 level in ultra-pure water and also in saliva, urine and blood plasma for
some investigations.
Urine sample. A drug-free urine sample was collected from a healthy volunteer. The sample
was stored in a glass vial which was carefully cleaned with hydrochloric acid and rinsed with
deionized water and stored at 4 °C to prevent bacterial growth and proteolysis. The urine
sample was spiked with the standard solution of the mixture of the drugs to obtain the desired
concentrations. A real urine sample was collected from a volunteer treated with Ami. The
collection was done 10 hours after taking an Ami tablet comprising 25 mg of the drug. The
sample was stored at 4 °C in a clean glass vial and analyzed a day after collection without any
further pretreatment or dilution. The sampling procedures were performed according to the
Plasma sample. Frozen drug-free human plasma samples (blood group +B) were obtained
from the Iranian Blood Transfusion Organization (Tehran, Iran) and stored at −4 °C. The
samples were allowed to thaw at room temperature, and then were shaken and diluted with
deionized water (1:4) before extraction. The plasma sample was spiked with the standard
7
Saliva sample. Drug-free saliva samples were collected from a healthy volunteer several
times. The sample was stored in a glass vial which was carefully cleaned with hydrochloric
acid, rinsed with deionized water, and stored at 4 °C to prevent bacterial growth and
proteolysis. The saliva samples were spiked with the standard solution of the mixture of the
drugs to obtain the desired concentrations. 100 µL of the saliva sample was exposed to the
on-disc EME-DLLME without any pretreatment or dilution. The sampling procedures were
Palo Alto, CA, USA) equipped with a FID detector. The injector was used in the splitless
mode at 300 °C. The column used for separation of the analytes was a Varian wall coated
fused silica capillary column (30 m, 0.25 mm, i.d., film thickness 0.25 µm). The FID
temperature was fixed at 300 °C. Ultrapure helium and nitrogen (>99.999%) were used as the
carrier and makeup gas at 1 mL.min-1 and 25 mL.min-1, respectively. Analysis of the
standards related to calibration curves and real samples was performed on an Agilent 7890B
performed using an HP-5MS capillary column (30 m × 0.25 mm) with a film thickness of
0.25 µm. The mass spectrometer was operated in the electron impact ionization mode with an
ionizing energy of 70 eV. The interface and ion source temperatures were both set at 300 °C.
The oven temperature was initially set at 100 °C for 1 min and then was programed to 200 °C
at 100 °C min-1 and was held at this temperature for 4 min. The temperature then increased to
260 ° C at 10 ° C min-1 followed by another increase to 280 ° C at 100 ° C min-1 and was held
for 4 min. Helium (99.999%) was used as the carrier gas at a flow rate of 1 mL min-1. The
solvent delay time was 10 min. The two model analytes were identified using the NIST
database as well. Quantitative determination of the analytes in the standards and samples
8
were performed in the selective ion monitoring (SIM) mode of MS. The monitored ions of
the analytes were selected on the basis of good selectivity and high sensitivity and were set as
follows: Ami, m/z 58.1, 202.1 and Imi, m/z 58.0, 208.1, 234.0.
The disc contains 5 layers and also 6 units for 6 parallel extractions. The disc was
designed using AutoCAD software (Autodesk, Inc.). A schematic view of the disc and the
fabricated disc is shown in Fig. 1A and C. A schematic of one extraction unit is also shown in
Fig. 1B. The structure of the disc was cut on a 1 mm-thick poly(methyl methacrylate)
(PMMA) sheet using a laser cutting device (Perfect laser Co., Ltd, China). Afterwards, all
layers were cleaned with isopropanol, rinsed with DI water, and dried with nitrogen gas. The
layers were fixed by chemical-thermal bonding. First, layers 1 and 2 were bonded. After that,
a hole (0.7 mm D.) was drilled in the reservoir I of all the 6 units to insert the electrodes. A
stainless steel wire (0.5 mm D.) as the electrode was mounted inside the reservoir I (Fig. 1B).
The assembly was washed again with isopropanol, rinsed with DI water, and dried with
nitrogen gas. A porous polypropylene membrane (about 1.5 × 1.5 cm) was placed above the
reservoir I. Then, layers 3, 4, and 5 were respectively bonded to other layers. Before bonding
layer 5 to layer 4, a stainless steel wire as an electrode was mounted to layer 5 through a hole
drilled in this layer. The bonding procedure was as follows: Several microliters of ethanol
was poured on the surface of the PMMA sheet. Then, the second layer was placed above it
and aligned to match the right position. The assembly was then placed in an oven with
pressure applied on it and was cured at 75 ºC for 40 min. Finally, a piece of a pressure
sensitive adhesive (PSA) film was placed on the hole II (Fig. 1) embedded on layer 3.
9
Initially, the disc was fixed to the rotor of a homemade centrifugal driver. Before starting
the extraction, NPOE was immobilized in the pores of the polypropylene membrane sheet.
100 µL of a 10 mM HCl solution, which is the acceptor solution during the EME process,
was injected through the hole I into the reservoir I, so that the solution was placed on the
electrode (cathode) mounted in this reservoir. Then, 100 µL of a donor phase (sample
solution) was placed in the reservoir II on the polypropylene membrane sheet. The donor
phase must be in contact with the electrode (anode) mounted in the last layer. The electrodes
were then connected to a DC power supply. The voltages of 85 V for urine and plasma
samples and 150 V for saliva samples were applied and extraction was performed for 15 min.
The extraction was run without any agitation or convection of the donor solution. During the
extraction, the target analytes migrated from the aqueous sample toward the acceptor solution
placed in the reservoir I through the SLM. This step provides a proper clean-up for extraction
of target analytes from complex matrices like biofluids. After the EME extraction was
completed, a DLLME was done in the disc. The dispersive liquid-liquid extraction was
implemented on the acceptor phase of the EME step. To do so, there was no need to manually
transfer the acceptor phase of the EME step to perform a DLLME step. By rotating the disc,
the acceptor phase of the EME step flowed to the extraction chamber for the next step. A
tetrachloroethene (the extraction solvent) was injected to the reservoir III. 5 microliter of a
NaOH solution (pH = 12) was also introduced to the reservoir IV to adjust the pH of the
donor phase in the DLLME step. Then, the disc rotated counter clockwise with the spin speed
of 100 rpm for 2 min. As the disc rotated, centrifugal force caused the fluids to exist in the
reservoirs I, III, and IV to flow toward the extraction chamber. It should be noted that the
acceptor solution of the EME step, which was placed in the reservoir I, was used as the donor
phase in the DLLME step. In this step, the extraction solvent was dispersed in the donor
10
phase (Fig. 2A). After 2 min, the spin speed increased to 3500 rpm and the device kept
rotating for 4 min (See the video in ESI). As a result, a phase separation occurred and the
extraction solvent, which was denser than the aqueous solution, sedimented at the conical end
of the extraction chamber (Fig. 2B). The settled phase (about 2 µL) was then drawn into a
Hamilton syringe by perforating the PSA film on the hole II and was injected manually into
the GC/MS device. Each extraction unit was used only once.
relative recovery
Preconcentration factor (PF) was defined as the ratio of the final concentration of the
analyte in the acceptor phase (Cf,a) to the initial concentration of the analyte (Ci,s).
,
PF = (1)
,
Where Cf,a was calculated from a calibration graph obtained from the direct injection of the
standard solutions of the model analyte and Ci,s is the initial concentration of the analyte.
The extraction recovery (ER) was defined as the percentage of the mole's number of the
analyte extracted to the acceptor phase (nf,a) with respect to the number of moles of the same
, , × ,
ER% = × 100 = × 100 (2)
, , × ,
RR% = × 100 (3)
Where Cfound, Creal, and Cadded are the concentration of the analyte after addition of a
known amount of the standard into the real sample, the concentration of the analyte in a real
11
sample, and the concentration of a known amount of the standard which was spiked into the
In this study, the on-disc EME-DLLME was developed. A schematic of the disc devise is
illustrated in Fig. 1. A chip device provides the facility to implement consecutive extractions
on a solitary chip with miniaturizing the whole procedure. Moreover, the extraction
procedure is easy to handle and could be performed automatically. To show the applicability
of the designed disc, two basic drugs were chosen to be extracted from biofluids.
Amitriptyline and imipramine were selected as model analytes for this study. To reach the
best extraction efficiency, the effective parameters of DLLME and EME were optimized.
3.1.1 Extraction solvent: The extraction solvent has a vital role in the extraction
efficiency of DLLME. The disc was designed in such a way that the extraction solvent could
be collected at the end of the conical shape of the extraction chamber. For this purpose, the
extraction solvent must have higher density than the aqueous solution. Therefore, chlorinated
solvents are good candidates. On the other hand, considering the chemical resistance of
PMMA, most of the chlorinated solvents damage the PMMA disc. Among chlorinated
tetrachloroethene) [39, 40]. As a result, C2Cl4 was chosen as the extraction solvent.
3.1.2 Volume of the extraction solvent: The critical importance of the volume of the
extraction solvent in the DLLME efficiency is undeniable. Using a small amount of the
extraction solvent could form a good emulsion, but this may not be sufficient to perform a
proper extraction. On the other hand, a large quantity of the extraction solvent does not result
12
emulsion formation and achieving good extraction efficiency. To investigate the effect of this
parameter, different volumes of C2Cl4 were examined. The results depicted in Fig. 3A show
that the response increased by increasing the volume of the extraction solvent up to 5 µL.
Further increase in the volume of the extraction solvents led to a decline in the responses,
which was probably due to the dilution effect. In volumes lower than 5 µL, the collection of
the settled phase was faced with some difficulties. In addition, repeatability of the results was
reduced by using a lower volume of the extraction solvent. A volume of 5 µL was used for
solvent system: the aqueous solution, the disperser solvent, and the extraction solvent. These
solvents can form an emulsion. A sufficient amount of the disperser solvent is important in
complete formation of emulsion, so the volume of the disperser solvent has a critical role in
any DLLME procedure. To investigate the effect of the volume of the disperser solvent, a
series of volumes were considered. As shown in Fig. 3B, by increasing the volume of the
further increase in the volume of the disperser solvent has no significant effect on the
responses.
3.1.4 pH: In DLLME, the best extraction can be done when the analytes are in their
neutral forms. Two worthy tricks to make analytes neutral are changing the pH of the donor
phase and adding a proper ion pair agent. In this work, as the EME step is performed first, the
donor solution (the acceptor phase of the EME step) was acidified. In this work, 5 microliters
of a NaOH solution (pH = 12) was introduced to the reservoir IV to adjust the pH of the
3.1.5 Spin speed: The main advantage of DLLME is formation of fine droplets.
Formation of the droplets leads to an increase in the contact area between an extraction
13
solvent and an aqueous solution, which causes faster distribution of the analytes between the
extraction solvent and the aqueous solution. As a result a very fast extraction of the analytes
into the extraction solvent can occur. To form such droplets, the organic solvent mixture (the
mixture of an extraction solvent and a disperser solvent) must be injected quickly into an
aqueous solution. The fast injection of the organic solvent mixture can affect the size of the
droplets and subsequently the extraction efficiency by increasing the contact area. In the
constructed disc, the mixture of the organic solvents are introduced into the reservoir III. The
reservoir is connected to the extraction chamber by a channel. As the disc rotates, centrifugal
force makes the organic solvent mixture flow to the extraction chamber. The aqueous
solution (the acceptor phase of the EME step, discussed in the on-disc EME-DLLME
procedure section) flows toward the extraction chamber as well. As a result, the organic
solvent mixture is injected into the aqueous solution and dispersion occurs (Fig. 2A). Faster
spin speed leads to faster dispersion of the organic solvent mixture into the aqueous solution
and the formation of finer droplets. Thus, the spine speed can affect the extraction
performance. To investigate this parameter, several spin speeds were examined. Fig. 3C
shows the results. Increasing the spin speed up to 100 rpm increases the extraction efficiency
as well. By increasing the spin speed more than 100 rpm, no significant increase in the
extraction efficiency was observed. To compare the size of the droplets in the dispersion step,
the droplets formed at different spin speeds were imaged using a light microscopy (at 40X
magnification). As can be seen in Fig. 4, by increasing the spin speed up to 100 rpm, the size
of the droplets decreased and no significant change in the size of the droplets was observed at
spin speeds higher than 100 rpm. As a result, 100 rpm was chosen as the optimum spin speed
14
In this study, the EME step is a drop-to-drop electromembrane extraction. Based on the
literature [12, 41], NPOE was used as the supported liquid membrane (SLM). NPOE was
directly injected by a syringe on the surface of polyprpylen sheet. 100 µL of the sample
solution (the donor phase) comprising the target analytes was placed on the polypropylene
membrane sheet and remained stagnant during the extraction. 100 µL of an aqueous solution
was also used as the acceptor phase. Other effective parameters were optimized as follows:
3.2.1 Donor and acceptor phase solutions: EME is a microextraction method which
uses an electrical field as a driving force in extraction of ionized compounds. The extraction
affected by the electrical field, so to extract target compounds, they must be in their ionic
form [6, 42]. By changing the pH value, acidic and basic compounds can be converted to
their ionic or neutral forms. Since the pH values for ionization of basic compounds are
different from the ones for acidic compounds, by adjusting a proper pH value, EME would be
able to selectively extract acidic or basic compounds. This is one of the advantages of the
phases plays a pivotal role in the EME extraction performance. Therefore, this parameter
To reach the optimum value of this parameter, the pH values of the donor and acceptor
phases were studied by changing the concentration of HCl in the range of 0-10 mM and 0-30
mM in the donor and acceptor phases, respectively. The results for Ami and Imi are presented
in Fig. 5A and Fig. S1. In this work, two basic drugs were used as the model analytes. The
sample solution (the donor phase) should be acidic enough so that the target analytes would
carry a net positive charge. Additionally, the acceptor solution should be acidic enough to
prevent deprotonation of drugs during the extraction and thus their back-diffusion toward the
donor phase [7, 24]. As can be seen in Fig. 5A, the best results were obtained when the
15
concentration of HCl in the acceptor phase was 10 mM and no HCl was added to the donor
phase (0 mM HCl). A decrease in the extraction efficiency was observed by increasing the
concentration of H+ in both the donor and acceptor phases. Using 10 mM HCl in the donor
phase and 20 and 30 mM HCl in the acceptor phase resulted in instability in the SLM and
bubble formation on the electrodes. All of these indicated that high concentration of H+ led to
3.2.2 Voltage and time: As already mentioned, an electrical field is the main driving
force in EME. It has been proven that the migration of analytes across the SLM is affected by
the magnitude of the applied voltage [43, 44]. Thus, this parameter has an important role that
must be investigated. In this work, first, a voltage in the range of 70-240V was applied to the
system and ultra-pure water was used to prepare the working solutions. It was observed that
220 V was the optimum voltage in that media (Fig. S2). The goal of this work is to extract
some model analytes from biological fluids. It is known that biological fluids like urine and
blood plasma contain a high concentration of salts. In EME, as voltage is applied across an
SLM, all ionic species in the donor and acceptor solutions move electrokinetically toward
their relative electrodes. As a result, a boundary layer of ions is formed at the interface on
both sides of the SLM. The existence of a high concentration of ionic species in biological
fluids leads to a significant increase in the number of ions migrating through SLM at a given
moment. So the thickness of the boundary layer and the current level increase. This also
results in Joule heating and therefore instability of SLM [7]. As the electrical field is the main
driving force in EME and increasing the magnitude of the applied voltage increases the
current level, the applied voltage used in the aqueous media is not usable in biological fluids.
Therefore, the optimization of the voltage was repeated in the urine, blood plasma, and saliva
media again. Since the mass transfer is time-dependent, a simultaneous investigation of time
16
and applied voltage was performed. The results of the effects of the applied voltage on the
extraction efficiency of Ami and Imi in the urine media are shown in Fig. 5B and Fig. S3,
respectively. The best extractability of the model analytes in the urine samples were obtained
at 85 V for 15 min. As can be seen, by increasing the applied voltage and time, the extraction
efficiency increased as well. At 85 V after 20 min, Joule heating and instability of the SLM
led to a decrease in the extraction efficiency. By performing the extraction at 105 V for 5
min, the same conditions were observed. Therefore, it was not possible to continue the
extraction at 105 V more than 5 min. The same results were observed in the blood plasma
media too. Thus, 85 V for 15 min was chosen as the optimum values in both the urine and
blood plasma media (Fig. S4). A set of experiments were conducted on the saliva samples
(Fig. S5) and 150 V for 15 min was selected as the optimum applied voltage and time.
conventional EME works, applying constant voltages, during relatively long extraction times
and/or under high applied voltages pose some problems [45, 46]. To overcome these
problems, using a pulsed voltage was proposed by Yamani’s research group in 2012 [47]. In
this work, the suitability of using pulsed voltage for a drop-to-drop EME was investigated. In
this work, outage duration cannot have any significant influence on the thickness of the
charged double layer, but it decreases the whole applied voltage time instead, and so the
extraction efficiency decreases (Fig. S6). The detailed results can be found in the supporting
information.
To validate the on-disc EME-DLLME system under optimized conditions (Table 2), Ami
and Imi were spiked in the drug-free urine, blood plasma, saliva, and also pure water at
different concentrations. Table 3 provides the summery of the method evaluation results.
17
Calculations were made based on equations 1-4. The results showed that the on-disc EME-
DLLME system can be effectively employed for analysis of model analytes even in
complicated matrices such as biofluids. A good linearity was obtained with a coefficient of
determination (R2) more than 0.9959. The limits of detection (LOD) of the on-disc device
were in the range of 0.1-0.5 µg L-1 for Ami and Imi in different biofluids with relative
standard deviations (RSDs) lower than 3.5%. A comparison of the analytical performance of
the proposed method with other analytical reports is presented in Table 4. Although the
proposed method used much less donor solution than other methods, it reached LODs which
were comparable with other methods. In addition, integration of two sample preparation
methods (EME and DLLME) in one chip provides a semiautomatic chip device with easy
handling, fast operating, good repeatability and reproducibility at a lower cost and with a
smaller amount of the donor solution and organic solvent as well. The method also benefits
from the advantages of both analytical methods. Consequently, the proposed method can be
used for extraction and determination of low levels of Ami and Imi in different biofluids.
To evaluate the practical suitability of the proposed method for determination of drugs in real
biological samples, Ami and Imi as model drugs were determined in three biofluids (urine,
Urine sample: The on-disc system was used for the analysis of the model drugs in the
urine sample of a patient undergoing therapy with Ami to assess the applicability of the
method. Sample preparation is already explained in the “real sample” section. In summary, a
urine sample was collected after 10 hours of taking an Ami tablet comprising 25 mg of the
drug. Accuracy was evaluated as the parameter of error %, so the urine sample was spiked
with 150 µgL-1 Ami and Imi. RSD% values indicated acceptable precision related to the
18
proposed system. Furthermore, as can be seen in Table 5, error% was less than +3.9%. Thus
accuracy was acceptable, and no significant matrix effect was observed for determination of
the model drugs using the proposed on-disc system. A typical chromatogram, obtained from
the extraction of the urine sample before and after the addition of a certain concentration of
Blood plasma: The validity of the on-disc system was assessed by extracting the
model drugs from a blood plasma sample. As described before, the blood plasma sample was
diluted 1:4 before extraction. The blood plasma samples were spiked with 50 µgL-1 of Ami
and Imi. The results shown in Fig. S7 and Table 5 indicate efficient extraction of the analytes.
During the EME step, it was seen that the SLM showed good stability.
Saliva: Applicability of the method was assessed in the saliva samples as well.
Sample preparation is already explained in the “real sample” section. A saliva sample was
spiked with 10 µg L-1 of Ami and Imi. As Table 3 shows, the method has acceptable accuracy
with error% less than +2.7. Acceptable precision based on RSD% values was also achieved.
A topical chromatogram of the blank and spiked saliva sample is shown in Fig. S8.
Furthermore, as can be seen in Table 5, error% was less than +2.9, accuracy was acceptable,
and no significant matrix effect was observed for determination of the model drugs using the
4. Conclusions
The present work demonstrated the design of a lab-on-a-disc device for performing a
combination of two LPME methods, EME and DLLME. The system was simple and the
component had a low cost, as each extraction unit was used only once. The system benefits
a lab-on-a-disc device. Using a drop-to-drop EME, the great clean-up potential of EME can
be used for extraction of target analytes from complex matrices. Utilizing a DLLME also
19
brings good preconcentration factors and GC compatibility of the final extracted phase.
advantages over the conventional method. Performing the extraction method on a chip device
makes it possible to use a small amount of the donor solution, which is not possible in
conventional extraction methods. EME and DLLME processes are performed on the same
disc, so there is no need to manually transfer the acceptor solution of EME to conduct a
DLLME step. Using a centrifugal platform, the organic solvent mixture can be automatically
injected into the donor solution in the DLLME step by simply rotating the disc. Thanks to the
scalable centrifugal force, injection can be performed with different speeds. In this work, we
showed that the injection speed could influence extraction efficiency by affecting the size of
the droplets. Finally, sedimentation of the extraction phase occurs by increasing the spin
speed. The system has the potential to run fully-automatically and be miniaturized more by
changing the detection method, which can be considered as the main limitation of such
systems for miniaturization. More studies focusing on fully automated on-disc extraction and
References
20
[7] Y. Yamini, S. Seidi, M. Rezazadeh, Electrical field-induced extraction and separation
techniques: Promising trends in analytical chemistry – A review, Anal. Chim. Acta 814
(2014) 1-22.
[8] A. Gjelstad, S. Pedersen-Bjergaard, Electromembrane extraction—Three-phase
electrophoresis for future preparative applications, Electrophoresis 35 (2014) 2421-2428.
[9] C. Basheer, J. Lee, S. Pedersen-Bjergaard, K.E. Rasmussen, H.K. Lee, Simultaneous
extraction of acidic and basic drugs at neutral sample pH: A novel electro-mediated
microextraction approach, J. Chromatogr. A 1217 (2010) 6661-6667.
[10] M. Rezazadeh, Y. Yamini, S. Seidi, B. Ebrahimpour, Electromembrane surrounded solid
phase microextraction: A novel approach for efficient extraction from complicated matrices,
J. Chromatogr. A 1280 (2013) 16-22.
[11] L. Guo, H.K. Lee, Electro membrane extraction followed by low-density solvent based
ultrasound-assisted emulsification microextraction combined with derivatization for
determining chlorophenols and analysis by gas chromatography–mass spectrometry, J.
Chromatogr. A 1243 (2012) 14-22.
[12] S. Seidi, Y. Yamini, M. Rezazadeh, Combination of electromembrane extraction with
dispersive liquid–liquid microextraction followed by gas chromatographic analysis as a fast
and sensitive technique for determination of tricyclic antidepressants, J. Chromatogr. B 913-
914 (2013) 138-146.
[13] L. Arjomandi-Behzad, Y. Yamini, M. Rezazadeh, Extraction of pyridine derivatives
from human urine using electromembrane extraction coupled to dispersive liquid–liquid
microextraction followed by gas chromatography determination, Talanta 126 (2014) 73-81.
[14] M. Khajeh, S. Pedersen-Bjergaard, M. Bohlooli, A. Barkhordar, M. Ghaffari-
Moghaddam, Maghemite nanoparticle-decorated hollow fiber electromembrane extraction
combined with dispersive liquid–liquid microextraction for determination of thymol from
Carum copticum, J. Sci. Food Agr. 97 (2017) 1517-1523.
[15] P.A.L. Wijethunga, Y.S. Nanayakkara, P. Kunchala, D.W. Armstrong, H. Moon, On-
chip drop-to-drop liquid microextraction coupled with real-time concentration monitoring
technique, Anal. Chem. 83 (2011) 1658-1664.
[16] N. Wang, S. Mao, W. Liu, J. Wu, H. Li, J.-M. Lin, Online monodisperse droplets based
liquid–liquid extraction on a continuously flowing system by using microfluidic devices,
RSC Adv. 4 (2014) 11919-11926.
[17] I. Vural Gürsel, S.K. Kurt, J. Aalders, Q. Wang, T. Noël, K.D.P. Nigam, N. Kockmann,
V. Hessel, Utilization of milli-scale coiled flow inverter in combination with phase separator
for continuous flow liquid–liquid extraction processes, Chem. Eng. J. 283 (2016) 855-868.
[18] S.K. Kurt, I. Vural Gürsel, V. Hessel, K.D.P. Nigam, N. Kockmann, Liquid–liquid
extraction system with microstructured coiled flow inverter and other capillary setups for
single-stage extraction applications, Chem. Eng. J. 284 (2016) 764-777.
[19] N.J. Petersen, J.S. Pedersen, N.N. Poulsen, H. Jensen, C. Skonberg, S.H. Hansen, S.
Pedersen-Bjergaard, On-chip electromembrane extraction for monitoring drug metabolism in
real time by electrospray ionization mass spectrometry, Analyst 137 (2012) 3321-3327.
[20] P. Mary, V. Studer, P. Tabeling, Microfluidic droplet-based liquid−liquid extraction,
Anal. Chem. 80 (2008) 2680-2687.
21
[21] M.D. Ramos Payán, H. Jensen, N.J. Petersen, S.H. Hansen, S. Pedersen-Bjergaard,
Liquid-phase microextraction in a microfluidic-chip – High enrichment and sample clean-up
from small sample volumes based on three-phase extraction, Anal. Chim. Acta 735 (2012)
46-53.
[22] P. Znidarsic-Plazl, I. Plazl, Steroid extraction in a microchannel system-mathematical
modelling and experiments, Lab Chip 7 (2007) 883-889.
[23] N.J. Petersen, S.T. Foss, H. Jensen, S.H. Hansen, C. Skonberg, D. Snakenborg, J.r.P.
Kutter, S. Pedersen-Bjergaard, On-chip electro membrane extraction with online ultraviolet
and mass spectrometric detection, Anal. Chem. 83 (2010) 44-51.
[24] Y.A. Asl, Y. Yamini, S. Seidi, B. Ebrahimpour, A new effective on chip
electromembrane extraction coupled with high performance liquid chromatography for
enhancement of extraction efficiency, Anal. Chim. Acta 898 (2015) 42-49.
[25] M. Karami, Y. Yamini, Y. Abdossalami Asl, M. Rezazadeh, On-chip pulsed
electromembrane extraction as a new concept for analysis of biological fluids in a small
device, J. Chromatogr. A 1527 (2017) 1-9.
[26] K. Wang, G. Luo, Microflow extraction: A review of recent development, Chem. Eng.
Sci. 169 (2017) 18-33.
[27] C. Xu, T. Xie, Review of microfluidic liquid–liquid extractors, Ind. Eng. Chem. Res. 56
(2017) 7593-7622.
[28] D. Ciceri, J.M. Perera, G.W. Stevens, The use of microfluidic devices in solvent
extraction, J. Chem. Technol. Biotechnol. 89 (2014) 771-786.
[29] N. Assmann, A. Ładosz, P. Rudolf von Rohr, Continuous micro liquid-liquid extraction,
Chem. Eng. Technol. 36 (2013) 921-936.
[30] B.C. Giordano, D.S. Burgi, S.J. Hart, A. Terray, On-line sample pre-concentration in
microfluidic devices: A review, Anal. Chim. Acta 718 (2012) 11-24.
[31] T. Vilkner, D. Janasek, A. Manz, Micro total analysis systems. Recent developments,
Anal. Chem. 76 (2004) 3373-3386.
[32] M. Madou, J. Zoval, G. Jia, H. Kido, J. Kim, N. Kim, LAB ON A CD, Annu. Rev.
Biomed. Eng. 8 (2006) 601-628.
[33] C. Li, X. Dong, J. Qin, B. Lin, Rapid nanoliter DNA hybridization based on
reciprocating flow on a compact disk microfluidic device, Anal. Chim. Acta 640 (2009) 93-
99.
[34] M. Vázquez, D. Brabazon, F. Shang, J.O. Omamogho, J.D. Glennon, B. Paull,
Centrifugally-driven sample extraction, preconcentration and purification in microfluidic
compact discs, TRAC,Trends Anal. Chem. 30 (2011) 1575-1586.
[35] E. Duffy, R. Padovani, X. He, R. Gorkin, E. Vereshchagina, J. Ducrée, E. Nesterenko,
P.N. Nesterenko, D. Brabazon, B. Paull, M. Vázquez, New strategies for stationary phase
integration within centrifugal microfluidic platforms for applications in sample preparation
and pre-concentration, Anal. Methods 9 (2017) 1998-2006.
[36] G. Duffy, I. Maguire, B. Heery, P. Gers, J. Ducrée, F. Regan, ChromiSense: A
colourimetric lab-on-a-disc sensor for chromium speciation in water, Talanta 178 (2018) 392-
399.
22
[37] A. Kazarine, M.C.R. Kong, E.J. Templeton, E.D. Salin, Automated liquid–liquid
extraction by pneumatic recirculation on a centrifugal microfluidic platform, Anal. Chem. 84
(2012) 6939-6943.
[38] L. Clime, D. Brassard, M. Geissler, T. Veres, Active pneumatic control of centrifugal
microfluidic flows for lab-on-a-chip applications, Lab Chip 15 (2015) 2400-2411.
[39] a.R.P.C. eplastics, USA.
[40] D.O.M.M. CORNELL UNIVERSITY USA, Sevier Lab.
[41] S.S.H. Davarani, A.M. Najarian, S. Nojavan, M.-A. Tabatabaei, Electromembrane
extraction combined with gas chromatography for quantification of tricyclic antidepressants
in human body fluids, Anal. Chim. Acta 725 (2012) 51-56.
[42] A. Gjelstad, K.E. Rasmussen, S. Pedersen-Bjergaard, Electrokinetic migration across
artificial liquid membranes: tuning the membrane chemistry to different types of drug
substances, J. Chromatogr. A 1124 (2006) 29-34.
[43] A. Gjelstad, K.E. Rasmussen, S. Pedersen-Bjergaard, Simulation of flux during electro-
membrane extraction based on the Nernst–Planck equation, J. Chromatogr. A 1174 (2007)
104-111.
[44] A. Gjelstad, T.M. Andersen, K.E. Rasmussen, S. Pedersen-Bjergaard, Microextraction
across supported liquid membranes forced by pH gradients and electrical fields, J.
Chromatogr. A 1157 (2007) 38-45.
[45] M. Balchen, L. Reubsaet, S. Pedersen-Bjergaard, Electromembrane extraction of
peptides, J. Chromatogr. A 1194 (2008) 143-149.
[46] M. Eskandari, Y. Yamini, L. Fotouhi, S. Seidi, Microextraction of mebendazole across
supported liquid membrane forced by pH gradient and electrical field, S. J. Pharm. Biomed.
Anal 54 (2011) 1173-1179.
[47] M. Rezazadeh, Y. Yamini, S. Seidi, A. Esrafili, Pulsed electromembrane extraction: A
new concept of electrically enhanced extraction, J. Chromatogr. A 1262 (2012) 214- 218.
[48] A. Gjelstad, K.E. Rasmussen, S. Pedersen-Bjergaard, Electromembrane extraction of
basic drugs from untreated human plasma and whole blood under physiological pH
conditions, Anal. Bioanal. Chem. 393 (2009) 921-928.
[49] M. Rezazadeh, Y. Yamini, S. Seidi, L. Arjomandi-Behzad, Voltage-step pulsed
electromembrane as a novel view of electrical field-induced liquid-phase microextraction, J.
Chromatogr. A 1324 (2014) 21-28.
[50] M. Bazregar, M. Rajabi, Y. Yamini, Z. Saffarzadeh, A. Asghari, Tandem dispersive
liquid–liquid microextraction as an efficient method for determination of basic drugs in
complicated matrices, J. Chromatogr. A 1429 (2016) 13-21.
[51] S.S.H. Davarani, A. Morteza-Najarian, S. Nojavan, A. Pourahadi, M.B. Abbassi, Two-
phase electromembrane extraction followed by gas chromatography-mass spectrometry
analysis, J. Sep. Sci. 36 (2013) 736-743.
23
Table 1. Chemical structures, pKa and log P of the model analytes
24
Table 2. Optimum conditions for extraction of model analytes by the on-disc EME-DLLME system
DLLME EME
Extractive solvent C2Cl4 Supported liquid membrane (SLM) NPOE
Water 220 V
Saliva 150 V
25
Table 3. Analytical performance of on-disc EME-DLLME for determination of Ami and Imi in water, saliva,
26
27
Table 4. Comparison of figures of merit of on-disc EME-DLLME with other methods for determination of Ami and Imi
28
c
Electromembrane solid phase microextraction
d
Electromembrane extraction- Dispersive liquid-liquid microextraction
e
Tandem dispersive liquid–liquid microextraction
29
Table 5. Determination of Ami and Imi in the real samples using on-disc EME-DLLME.
Sample Analyte Creal (µg L-1) Cadded (µg L-1) Cfound (µg L-1) RSD% Error%
30
Figure captions
Fig. 1. A) A schematic view of the designed LOD device consists of five layers. B) A
Fig. 2. Desperation (A) and sedimentation of the extraction solvent (B) in the DLLME step.
Fig. 3. Optimization of the effective parameters in the DLLME step. In this step, C2Cl4 and
methanol were used as the extraction and disperser solvents, respectively. A) Volume of the
extraction solvent. B) Volume of the disperser solvent. C) Spin speed. In the EME step, 180
V was applied for 15 min without adding any acid or base to the donor and acceptor phases.
Fig. 4. Effect of spin speed on the size of the formed droplets in the DLLME step.
Fig. 5. Optimization of the effective parameters in the EME step. A) Composition of the
donor and acceptor phases. B) Effect of applied voltage at different extraction times. For the
Fig. 6. A typical GC/MS chromatogram obtained from the extraction of the model analytes
from the plasma samples before and after spiking 50 µg L-1 of the analytes. The GC
31
Fig. 1.
32
Fig. 2.
33
Fig. 3.
34
Fig. 4.
35
Fig. 5.
36
37
Fig. 6.
38
Highlights
A lab-on-a-disc device was designed for sequential performing of EME and DLLME.
All of DLLME steps can be fulfilled on the chip by simply controlling the spin speed.
EME-DLLME was used for extraction of tricyclic antidepressants in biofluids.
The effective parameters on the EME and DLLME were investigated and optimized.
The chip is easy to handle without need to any skillful operator.
The device provides effective and reproducible extraction using a low volume of sample.
Declaration of interests
☒ The authors declare that they have no known competing financial interests or personal relationships
that could have appeared to influence the work reported in this paper.
☐The authors declare the following financial interests/personal relationships which may be considered
as potential competing interests: