Wiring Diagrams
English
Print No.: SPB7-250.844.01.04.02 Doc. Gen. Date: 01.05
Replaces: SPB7-250.844.01.03.02
0-2 Revision
Disclaimer
The installation and service of equipment described herein is to be performed by qualified personnel
who are employed by Siemens or one of its affiliates or who are otherwise authorized by Siemens or
one of its affiliates to provide such services.
Assemblers and other persons who are not employed by or otherwise directly affiliated with or autho-
rized by Siemens or one of its affiliates are directed to contact one of the local offices of Siemens or
one of its affiliates before attempting installation or service procedures.
Page
Document revision level . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0 - 2
Disclaimer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0 - 2
System overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 1
System overview - stand . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 2
Page
4 ______ Generator Overview ____________________________________________ 4 - 1
(1) (2)
(3)
NOV00132
Stand with rotating support arm with one FD (Flat Detector) object table and one
18 x 24 object table (used with Opdima, option) and Mo/W X-ray tube.
Stand display
Foot switch
R871
K7-K9 R805
Object table
D807 Right side S1, S2, M
D807 Left side S890
D801
S867
M5 S880
M4:1 S896
M4:2
M4:3 S891
M4:4 X891
D814 S869
Lamp, X894 S897 R861 X874 R863
M3 (Rotation)
M2 (Compression) X888
D802
Strain gauge S894
D808 Right side
D804 D808 Left side
M, S1, S2
R803
S3 S860
S861 X898, R806
Detector array S862 S896 S895 X858 D750
D803
X801 F801-808
Incoming
T802 T801 mains
Fig. 1 Front of the stand
PXCM unit
(Brick unit)
S882
X601
X602
X603
PB0852
HV-tank
PB0851
HV inverter
S881
PB0854
Tube power
PB0855
AC inverter
PB0850
Mains input converter
S883
S884
T803
X885
X804
NOV00127
Stand display
Foot switch
Stand with rotating support arm with one FD (Flat Detector) object table and one
18 x 24 object table (used with Opdima) and Mo/W X-ray tube.
PCB Overview 0
Fuses Overview 0
Switches 0
Potentiometers 0
Motors 0
Miscellaneous 0
*BRAKE RELAY Output to brake reversing relay, max brake if *BRAKE_RET active.
C_LIM Rotation and fast-acting current limit.
C_PWM Pulse Width Modulated power output to compression motor.
CASS_LOADED Cassette inserted in object table 1 (not used).
COLL_MOT_INT Collimator motor interrupt signal.
COLL_MOT_RX Collimator motor receive signal.
COLL_MOT_TX Collimator motor transmit signal.
COMP_LIMIT Power output via compression limit switch to compression motor.
*COMP_SPEED Pulse Width Modulated signal to control speed of compression motor (20kHz).
COMPRESS Operator control signal to move compression plate down.
D_PWM Pulse Width Modulated power output to rotation or lift motor.
DECOMP_LIMIT Power output via decompression limit switch to compression motor.
DECOMPRESS Operator control signal to move compression plate up.
*RELEASE Control signal to release the brakes from min or max to permanent magnet braking.
*RESET Reset signal to the CPU.
RESET Reset signal to the smart point interface.
*RESET_HW Reset signal from the CPU to the reset cicuit, caused by slave restart and CPU watchdog
reset.
RESET_SW Signal from the reset switch to the CPU via the reset circuit.
ROT_CW Operator control signal to rotate the x-ray system clockwise.
ROT_CCW Operator control signal to rotate the x-ray system counter clockwise.
RXD Input for receiving data via MPS.
*SPEED Pulse Width Modulated signal to control speed of rotation or lift motor (20 kHz).
STAND_REQ_HV_OFF Signal to terminate exposure. WIRED-OR line, see Master board D750.
(*KVA)
STEREO_LEVER Signal from stereo lever switch, indicating lever not in stereo position.
TABLE_CONFIG 4 inputs from object table, see table configuration .
TACHO Signal from rotation motor tachometer.
THICKNESS Analog signal indicating the compression thickness.
THICKNESS_B Buffered output to the thickness adapter. (used with Opdima)
THICKNESS_STEREO Isolated analog signal indicating compression thickness for stereo evaluation unit.
TILT Indicates that the stereo angle is 0 or ±10 . Visible through vent opening behind tube.
TUBE ANGLE Analog signal indicating the rotation angle.
TXD Output for sending data via MPS.
ZERO Used for adjusting offset of the compression force amplifiers.
Page Voltage Origin Supplied from Supply for Test LED Fuse
3-3/4E +5V D801 11V Digital IC, 5V V7 D801/F1 F801
+5V_REF
3-3/4F +5V_REF D801 +5V Pots. 5V_REF D801/F1 F801
3-12/3C D804 X843 pin 2
3-3/3E +5V_DSP D801 11V D803, Display 5V_DSP V6 D801/F1 F801
3-3/2E +7V D801 11V Analog IC, 7V V5 D801/F1 F801
3-12/3B D804 X843 pin1
3-6/4D +5V_MPS D801 11V_MPS Generator 5V_MPS V26 D801/F2 F802
interface
3-3/3D 0VD GND D801/X811 Digital IC 0VD1,OVD2
3-3/3D 0VA 0VD D801 Analog IC 0VA1
3-12/3B D804 X843 pin 3
3-6/5D 0V_MPS D750/GND Generator inter- D801 MPS
face ground
3-3/2C 11V T801 230VAC +5V, +5V_DSP, D801/F1 F801
+7V
3-6/5B 11V_MPS T801 230VAC +5V_MPS D801/F2 F802
3-4/4D +5V D802 14VAC Digital IC 5V V57 D802/F1 F803
3-4/4E +5V_M D802 +5V D801-D802 V57 D802/F1 F803
opto interface
3-2/5C 0V_W D801 D805/X856 Optointerface
3-2/4C 0V_M D801 D802/X822 Optointerface
3-4/3D +14V D802 14VAC +5V, D805 14V D802/F1 F803
3-4/2D +15V D802 +24VF Transistor 15V V16 D802/F2
drivers M1,M3
3-4/2C +24V D802 24VAC +15V D802/F3 F805
3-4/2D +24VF D802 +24V Switches, 24VF D802/F2
relays, etc
3-4/3A 14VAC T801 230VAC +14V D802/F1 F803
3-15/5B 17VAC T801 230VAC LAMP 17VACF D802/F5 F804
3-15/4E 17VACF D802 17VAC LAMP 17VAC_RET D802/F5 F804
3-15/4D 0V_L D802 D802 Lamp control
3-15/5D -V_L D802 17VACF Lamp control
3-4/2A 24VAC T801 230VAC D802/F3 F805
3-4/2E 24VACF D802 24VAC Fan D802/F4
3-4/3D 0V GND D802/X821 Ground 0V, 0V1
3-4/G2 +5V D805 D802/+14V Digital IC 5V V17 D802/F1 F803
3-4/2G +12V D805 +14V V_MOT D814 12V D802/F1 F803
3-4/2G +5V_SP D805 D802/+14V Digital IC 5V_SP D802/F1 F803
3-4/2G 8,2V_SP D805 D802/+14V Digital IC 8,2V_SP D802/F1 F803
1
X885
X847 X845
X846
X893 X847
X858 X849
X804
2
X857
WING BOARD X860
Page 1 of 19
SP PLM WH 5
TRANSFORMER X874
T801
X886
TUBE
TRANSFORMER
T802 X802
3
X895
X896
M2 M1 X897
TRANSFORMER
T803 DISPLAY
230 VAC ROTATION LIFT
Z801 D803 X826
T803 X888 X887
To Opdima
Filter X801
X803
X753
X755
D750
4
X756
Wiring Diagrams
Mammomat Novation
Incoming
DR
5
MAINS INPUT X706 X710 PB0855 X711 X723 X720 X731 PB0851 X730 X6
MAINS INPUT CONVERTER AC INVERTER TUBE POWER HV INVERTER HV TANK PXCM
(BRICK)
X705 X700 X712 X722 X724 X732
3-1
0
Medical Solutions
Siemens AG
A B C D E F G
3-2
INCOMING MAINS
STAND
GROUNDING CONNECTIONS
P.E.
GROUND RAIL
1
GENERATOR
GROUND RAIL
Door
CONTROL PANEL
X803 CURTAIN
D741 D740 8
HV
Common Tank X32 X5 X30
1 2 6 8 10
UPPER
Rev. 04
SPB7-250.844.01
X6 X5 3 3 6 3
X896
H3
X730 HV Inverter
1 2 6 8 10 D-subplate X897
17 1 5 6 7 12 18 19
TUBE TUBE FRONT
01.05
REAR
COVER
0V_sec LIFT CHASSIS PLATE COVER
2 4 8 10 16 20 14 D810 CARRIAGE
X731 COMPRESSION
X81 UNIT
0V
2
TUBE
Page 2 of 19
SP PLM WH 5
AC X713
HOUSING
Inverter 17 1 5 X804 6 7 12 18 19
PATIENT PATIENT
LOWER HANDLE HANDLE X895
X720 Tube K8
2 4 8 10 16 20 14 Power 3
Supply 5
X726
6 X858 K9
4 X883
7 4
6
X725 Ferrit sleeve
0V_sec 0VA X891
X755 X1 X2 X893 X885
X721 X758 3 4 8 22 22
8 1
3
2 4 6 8 10 2 23 23
X722 4 X756
24 24
5 X768
4 X860 X859 25 25
6 7 2 5 8 8 7
3
X752 6
8
X700 Mains Input X871
2 4 6 8 Conv. 10
D803 D801 X821 D805
X811 X812 X822 D802 WING 1
DISPLAY CPU 1 1 WING
X701 X760 MOTOR
6 6 2 2 BOARD
0V A 0V D BOARD
0V_sec 8 8 X801 X816 3 3 MAG TABLE
15 15 2 X860
4 4
10 10
0VA
0V A 16 16 5 5 WING 2 Frame left
12 12 X823 8 5
17 17 6 6
0VD D750 18 18 8 8 0V Detector
16
MASTER X855 Beam
0V D 0V D 0V_M 0V
0V_RS232 X765 BOARD 12
SHLD_1
3
6 X757 X815
1 1 Frame right
9 X813 X856
3 2 1 27 X840
10 X846
7 7 2 25 1
Service PC 0V D
4
8 8 33 29
X760
Wiring Diagrams
Mammomat Novation
16 34 30
0V_RS232 X830
X761 X602 X603 0V
X762 2 14
E 0V_MPS
COMPRESSION 2 2
Filter/mirror
8 8 9 9
17
0V_W UNIT disk switch
5 5 10 18 PRES_FORCE D804 X838
X805 GND_S_PC COMPRESSION 0V
12 19 R861 R863 2
X764 X814 THICK FORCE S867
A 24 25 14 10k 2k NESS TUBE_ANGLE
X874 X841 D814
B 20 23 16 7
GND_PTR 2 R803 Collimator
22 24 0V A
5 15 2 2k contr.board
PXCM X868
0
0VI_RS232 17
DR
5
(BRICK) 6
6
PRESET_ANGLE
X601 NOTE!
X763 R871
5 X855 DOTTED CONNECTIONS INDICATE
13 X809 10k
X769 3 SHLD SHIELDED CABLES WITHOUT
8 7
9 POT_RET 4 0V CONNECTION
X863
20 17 5 1
5
13
0V_A
GND_OPD_WS 0V_RS232
X041E
X041E
Medical Solutions
Siemens AG
A B C D E F G
D801
D801
2
Page 3 of 19
SP PLM WH 5
~
INVERTER 3 6 6 1 1 3 9
6 2 AT 2AT
7V
X713 11V 7V
230V AC 0VA1
V5
1 7 7 2 2 4 4 8
0VA
0VD2
7
SEE PAGE 6-2
0VA
11V_MPS 5V_DSP
X803-8
(SEE PAGE 3-6) +5V_DSP 0VA1
3
6 0VD
0VD2 V6 7V
0VD V5
X850 X850 7
3 T803 1 Z801 1
5V_REF 5V_REF
GND +5V_REF
5V_DSP
GND 6.8 V6 X811
5V F1
4 2 2
= 0VD1 V7
+5V
0VD1
5V
V7
2AT
0VD
V26 F2
4
X883-8 0V_MPS 5V_MPS
Wiring Diagrams
Mammomat Novation
1AT
0
DR
3-3
Medical Solutions
Siemens AG
A B C D E F G
3-4
POWER SUPPLY MOTOR CONTROL
1
D802
Rev. 04
SPB7-250.844.01
D805 14V
X886A/B X886C X821 X824
T801 13 F805 13 6 F4 9 12V
24VACF X830
6.3AT 1AT UREG 12V
EMERGENCY F3
01.05
18 X860 12V
24VACFR 10
24VAC STOP X823 4
S880 6.3AT +24VF
S882 12 +24VF 9
X886 X886 1
6.8Ω 5V 0V
14 14 14
2
+24VF 0V UREG 5V
Page 4 of 19
SP PLM WH 5
11 5 5V
X886
1
~ +24V
0V
24VF
17
+24VF
+24VF 0V
8.2V_SP 5V_SP
1
X821
2
= X822
7
MOTOR_VOLT
6
+14V +14V
UREG
8.2V +8.2V
+14V
UREG
5V +5V_S
0V
15V
5
F2
2 8 +15V
UREG
3.15AT V16
(SEE PAGE 3-3) 15V
0V
0V
X886A/B X886C X821 0V 0V1 D802
9 9
3
F803 4 F1 X823
X822
14VAC
5AT 5AT
14VACR
~ 0V
14V
8 0V D805
0V 5V 12V
10 10 3
= 0V
X823
9 +14V
X865
S880 +5V
5V X822 X824
X864
10 15V V16
+5V_M
UREG X866
V57 X824
5V 11
+5V
17VAC 0V 24VF
0V
(SEE PAGE 3-15) X823
4
5V_SP
Wiring Diagrams
Mammomat Novation
X844
1 2 14V
+ 5V
8,2V_SP
V57
X855 X868
22000 µF
X856 X869
0V1 X872
0
X857
DR
14VACR X871
5AT
3.15AT
6.3AT
1AT
5AT
X821 X860 X859
X873
F1 F2 F3 F4 F5
X041E
X041E
Medical Solutions
Siemens AG
A B C D E F G
D803
DISPLAY
1
X816 X801
D801 3 DATA x8
D0 3 D0 x8
0VD EN
D1 4 4 D1 C1
Rev. 04 01.05
SPB7-250.844.01
D2 5 5 D2 1D
D3 6 6 D3
DATA DATA
D4 7 7 D4
D5 8 8 D5
D6 9 9 D6
2
D7 10 10 D7
Page 5 of 19
SP PLM WH 5
OK
0VD1
15 POWER SUPPLY 15 OPCOM
0V 16 16 0V
CPU & PLD 17 17
FUNCTION 18 18
1 1
2 2 +5V_DSP
+5V_DSP
3
19 19
20 20
5V_DSP
BIN/OCT
STROBE 14 ADDRESS 14 STROBE 0
1
1
ADDRESS A3 11 11 A3 ADDRESS 2 2
A2 12 12 A2 4 3
4
A1 13 13 A1 & 5
6
7
4
Wiring Diagrams
Mammomat Novation
0VD
D801
D803
FORCE
0
X801 THICKNESS
DR
OPCOM OK
X816
5
ANGLE
3-5
5V_DSP
V6
0VD1
Medical Solutions
Siemens AG
A B C D E F G
3-6
D801
KVA X812
GRID_SP
COMP_SP
SPEED X813
*RESET
V60 COLL_MOT_RX
D750
01.05
FILT_MOT_RX_
*KVA STATUSLEDís MOT_INT
J1 UART_INT
SMART_RX
2
V1
X757 STAND_REQ_HV_OFF X815
Page 6 of 19
SP PLM WH 5
12MHz X1 0VD2
6 ( KVA) 6
* V2
PROM V60
X816
V3 V1
*AR V2
PLD
CPU & V4 V3
GEN_PREP_RDY PLD V4
5 (VH) 5 GRID_0 PLD
FUNC.
GRID_RUNNING TXD PROM
DATA AND ADDRESS
4 (AR) 4 CPU J1
RXD
0VA1
1 1
7V
3 X814
3 0V_MPS 0VD1 V5
3
0V_MPS 0VA
7 7
PROM RAM EEPROM 5V_REF
8 8 0VD 128k x 8 32k x 8 8k x 8
+5V_MPS
RXD 5V_DSP
I9
0VD 0V_MPS V6
X811
MPS 0VD1 F1
2 5V
2 MPS
V7
TXD
ROM MPS
X815
4
0V_MPS V26 F2
Wiring Diagrams
Mammomat Novation
D750 5V_MPS
S1
5V_MPS
RESET SWITCH
X811
1 F2
~ V26 0V_MPS
+5V_MPS
0
2 1AT
DR
= 0V_MPS
STATUS LEDís DURING STARTUP:
X886C
PROM/RAM TEST V1, V4 LIT.
5 6 FLOATING INTERFACE
5
GROUNDED BY MASTERBOARD D750 V2, V3 LIT IF PROM CHECKSUM ERROR.
F802
1AT
X757
V3, V4 LIT IF RAM ERROR.
X041E
X886A/B
11V_MPS
5 6
ALL TESTS PASSED, V1-V4 ARE DARK.
X3
HV TANK
1
TUBE
K9
2
h1M II
X726 H3
Page 7 of 19
SP PLM WH 5
6 GND FILAMENT W G
FILAMENT 2 SMALL_FOCUS_MO
M
YE
CONTROL
4 LARGE_FOCUS_MO
CIRCUITS X896
1 BIAS_VOLTAGE h1M K8 K9 K7
3 SMALL_FOCUS_W BN h3
GY h1W
3
D750 K8 START WARNING BLINK
ON PANEL
K9 INHIBIT EXPOSURE
D750
MASTER X755 X895/H11.X1
BOARD 3 K8 TEMP 4
4 K9 PRESS 6
P
K9
2 D802
0VD (K9) 5
ϑ
4
1 0VD (K8) 3
Wiring Diagrams
Mammomat Novation
X755
K8
X824 ϑ X895/H11.X1
18 1 2 X824
D802
K7
1 2
FAN
F4 9
0
DR
M6
1AT
S882 S880
X886A/B X821
F3
~
5
14 24VAC RN X886C 1 +24V
14
6.3AT
EMERGENCY
3-7
T801 0V
13 24VAC 13 6
STOP F3 F4 X821
F805
6.3AT
Medical Solutions
Siemens AG
A B C D E F G
3-8
D802
X825 X822
TACHO
1
ACT_SPEED
*DIR DRIVE_ERR
V24
*SPEED
X812 D801 D802
S1
Rev. 04
SPB7-250.844.01
2
TUBE_ANGLE 2
17 *SPEED 17
Page 8 of 19
SP PLM WH 5
3
MOTOR M
M3
+5V
L809
x825 X826 2 BK
LED 1 1 RD
2 2 BK
TACHO 0V
3 3 WH
+5V
4 4 BU TACHO
5 5 GN
0V
GND
M1/M3
D801 S883 X824 DRIVE_OK
2 17
X812 +24VF
15
1
4
COLLISION
Wiring Diagrams
Mammomat Novation
10 X824
LIFT_DOWN
5 8
X813 LIFT_UP
*SPEED ROT_CW
ROT_CCW X822 26 25 13 16 22
SEE PAGE ANGLE_MEM
3-10 LAMP
SEE PAGE
0
DR
0VD1 X898
1 2
5
RD BU
R806
BRAKE
X041E
M9
M10
X041E
Medical Solutions
Siemens AG
A B C D E F G
CPU LIFT MOTOR
SEE PAGE DRIVE_PROT
3-8 X887
1
1
D802 S881
21
2
X812 X822 PGM_ (OVER- +24VF
&
Page 9 of 19
SP PLM WH 5
17 17 SPEED CURRENT) K2
20 KHz PU
PWM 6 X888
0V 1 L808
0V 0V RD
0V1 0V 1 2 3 4 5
S1 +5V ROTATION
COMP
MOTOR M
Y
M3
0V 0V X X>Y
+3% +6% +12% +24% Servo 2 L809 BK TACHO
DRIVE ERROR
0V +5V (OVERCURRENT)
CURRENT LIMIT D V24
(FACTORY ADJUSTED) > C1
R
3
*SEL_ROT +5V
0V +5V
DRIVE_OK 22
11 11 S1 5
0V
SERVO
*DIR
14 14 0V C_LIM 24VF
V11 17
0V
+24VF
*RELEASE
18 18
BRAKE_RET
0V 15 *BRAKE_RETURN
*BRAKE
4
15 15
Wiring Diagrams
Mammomat Novation
0V X824
+5V +24VF *BRAKE_RELAY 8
+24VF
+24VF R805
0V 0V
0
DR
X824 X898
LIFT_DOWN 2 0V
1 2
LIFT_UP 1
ROT_CW 10 0V
RD BU
5
ROT_CCW 5 R806
BRAKE
3-9
M9
X822 26 25 13 16
CPU M10
SEE PAGE DRIVE_OK
3-8
Medical Solutions
Siemens AG
A B C D E F G
3-10
D807 D805 D807
1
+5V_R 1 1 +5V_R
X807
S1 LIFT_A 2 S1 S4
2 LIFT_DOWN
S5 S6
S2 LIFT_B 3 3 LIFT_UP
LEFT S2 S3
SIDE S3 ROT_CW 4 4 ROT_CW
Rev. 04
SPB7-250.844.01
SPARE 8 8 SPARE
S2 S3
X856
2
13
Page 10 of 19
SP PLM WH 5
15
16
+24VF
D807
X807 X865 14 +24VF
+5V_R 1 1 +5V_R
TO D801, D802, S883
S1 LIFT_A 2 2 LIFT_UP 10 LIFT_DOWN SEE PAGE 3-8
3
S1-S6 ON D807, D808
S6 LAMP 7 7 LAMP 7 LAMP
S1 S2 S3 S4 S5 S6
SPARE 8 8 SPARE 36 POSITION LIFT_A LIFT_B (NOT USED)
S1 LIFT_A 2 X889
6 S1 LIFT_A 3 6 3 LIFT_DOWN
4
Wiring Diagrams
Mammomat Novation
8 S3 ROT_CW 7 8 S3 7 4
ROT_CW ROT_CW
S4 ROT_CCW S4
10 9 10 ROT_CCW 9 5 ROT_CCW
12 SPARE 1 11 12 SPARE 1 11 6
14 SPARE 2 13 14 SPARE 2 13 7
0
8
DR
16 SPARE 3 15 16 SPARE 3 15
5
RIGHT SIDE BOARD LEFT SIDE BOARD X856
X041E
X041E 3-11
LEFT BLANK 0
1 2 3 4 5
This page is intentionally left blank.
G
F
E
D
C
B
A DR
Siemens AG SPB7-250.844.01 Page 11 of 19 Mammomat Novation
Medical Solutions Rev. 04 01.05 SP PLM WH 5 Wiring Diagrams
Medical Solutions
Siemens AG
A B C D E F G
3-12
X874
6
COMPRESSION UNIT D801
1
2 7 16 CPU &
D804 0VA 0VA
1 8 10 PLD
+7V +7V FUNC.
3 1 3 0VD
R860 +5V_REF +5V_REF
X842
Rev. 04
SPB7-250.844.01
F +5V_REF 1 1
+5V_REF 8
R R 4 COMP_FORCE
4
1
01.05
4
2
+5V_REF
R R 0VA +5V_REF GAIN
20k
2
OFFSET 1
3 (BR OPEN) 10k PRES_FORCE 3 6
20k
Page 12 of 19
SP PLM WH 5
A/D
R861
V2
STRAIN GAUGE 4 THICKNESS 5
R863
X816
2k
1-20
2 15
+5V_REF
0VA POT_RETURN
0.1V 0VA 20
POT_RETURN
SEE PAGE 3-14 POT_RETURN
0VA
0VA 1 THICKNESS_B
X843 7 2 1 3 8 4 5 6
3
+7V
1 7 13 X809
X885 +5V
BR_RET
+5V_REF
+7V
0VA
BR_OPEN
ZERO
COMP_FORC
11_OUT
5
TEST CONNECTOR
THICKNESS_STEREO 6 +
+
-
-
0V 7
1 13
X893 CMR SHIELD
D804 D801
4
SEE PAGE
Wiring Diagrams
Mammomat Novation
X843 3-17
X841 GAIN OFFSET X842
+7V
5V_REF
0VA
ZERO
COMP_FORCE
I1_OUT
BR_RET
BR_OPEN
X816
V2
D803 FORCE THICKNESS
0
DR
X801
X809 OPCOM OK 1-20
0VA1
7V
X814
5
5V_REF
X041E
X041E
Medical Solutions
Siemens AG
A B C D E F G
COMP_PROT
1
23
X812
23 +14V
CPU &
PLD
D805 COMPRESSION UNIT
FUNC. (LOW = COMPRESS)
Rev. 04 01.05
SPB7-250.844.01
0V
COMP_FORCE 8 *COMP_ +24VF X823 X860 2
S861
9 DIR 9 K4 1 COMP_LIMIT 3
1 K5 1
THICKNESS 6
PRES_FORCE 4 2 DECOMP_LIMIT 1 3
S862
COMP_SP *COMP_ COMP_SP 0V C_PWM 4
SPEED
20 20 DECOMP_LIMIT
0V1
+14V 9 6 +14V X873 L804
& GN
2
X813 COMPRESS 20KHz
2 RD
Page 13 of 19
SP PLM WH 5
PWM 0V
29 M2
0VD1 +5V PU (SEE PAGE 3-8) +24VF 12 4 1 M
BK L805
+15V 1 X844 +24VF
+5V 7 BK
0V
0VD 0V 0V 22000 µF
DECOMPRESS Y
COMP PU + * 8 X876 COMP
MOTOR
1
0VD2 X X>Y 2 X844 +24VF
27 8 5
2
COMP +5V
Y 0V 0V 3
0VD 0V X824 X856
X875
X812
COMP_OK (OVERCURRENT)
X X>Y
COMP ERROR 3 COMPRESS 33 +24VF 1
NOT USED
* SEE PAGE
3-2
(OVER CURRENT)
24 2
+5V V2 X822
D 3
3
0V 24
TABLE IN PLACE C1
X813
R
+24VF
(x4) 0V X856 32 31
X823 14 COMP_OK
TABLE_CONFIG
+24VF SEE PAGE
(SEE PAGE 3-17) DECOMPRESS 5-10
COMP_OK
DECOMPRESS
COMPRESS
X848
X803 1 2 3
D801 D802 1
WH
4
Wiring Diagrams
Mammomat Novation
RD 2
COMP_D 2 C802
RD 2
+15V
COMP_SP COMP_SP BK 3
X824
V16 C803
BK 3 3
0VD2 PU
COMP_PROT
0V D805
0
DR
5
X814
X856 X869 COMP.
3-13
X857 X872
X871
X860
X859 X873
0V1
Medical Solutions
Siemens AG
A B C D E F G
3-14
D801 D803 FORCE THICKNESS
5V_REF
+5V_REF
X814 OPCOM OK
1
D805 5 CPU &
X863
3
PLD
X855
+5V_REF 2 1 FUNC. ANGLE
2 18
X816 X801
10k 6
1 1-20 1-20
Rev. 04
SPB7-250.844.01
R871 PRESET_ANGLE 10
PRESET ANGLE 1
POT_RETURN 5 2
X868
TUBE ANGLE 8
R803 A/D
01.05
6 POT_RETURN 4
17 POT_RETURN
2k 7 TUBE_ANGLE 1 D805 D801 TESTPOINTS AND LEDís
9
5 3 0VA 0VA
2
15 20
Page 14 of 19
SP PLM WH 5
0VA
X863
0VA
+5V_REF
3 0VD X813
X813
X874
6
1
0VD2
(SEE PAGE
3-16)
COMPRESSION
R861 PRES_FORCE 3 8
UNIT TILT
10k (SEE PAGE 2-3) PROM X816
THICKNESS 5
R863
2k
3
2 POT_RETURN
PLD
X855 X868 PROM
1 X809
X856
CPU
4
V6
Wiring Diagrams
Mammomat Novation
BU 6 X811
0VD1 F1
S896 TILT X860 5V
0 , ±10 V8 X81 2 X83 X82 X81
X80
V7
LED
0
DR
S1
5
RESET SWITCH
X041E
X041E
Medical Solutions
Siemens AG
A B C D E F G
COLLIMATOR 1
D805 D801 D802
1
LEFT SIDE BOARD D805
X865
12V
D807 ** X812 X822
5V
0V
X807 X851 X851
1 1
+5V_R X813
X839 MIRROR
Rev. 04 01.05
SPB7-250.844.01
S6 X856
LAMP 7 7 LAMP 7
S6
5V_SP
8.2V_SP
5V_SP
LAMP 7 7 LAMP
X823
2
X831 SMARTPOINT
Page 15 of 19
SP PLM WH 5
SMART
RIGHT SIDE BOARD SMARTPOINT
1 INTERFACE 17VACF
POINT
INTERFACE 2
MEM
X860
4 TAG 17VAC_RET
8.2V_SP
+24V F5 X821
X856 X831
X869
X860
3
D801 D802 +24VF X823
X812 X822
1 1 12
2 2
3 3
X813
13
CPU & 4 4 MIRROR
LAMP 5
PLD 5
6 6
FUNC. 8
8
0V_M 0V
X812 X822 0V
4
*LAMP_CTRL 12
Wiring Diagrams
Mammomat Novation
12
0V
17VACF 0V
X821
V52
7 F5
5AT
COMPRESS 29
SET LAMP_V
R131
SEE PAGE 3-13
0V_L 0V_L
0
DR
COLLIMATOR
Phase
-V_L + angle LAMP
RMS X823 X894
control
/DC 16 1 1
5
URMS angle LAMP_V
ñ
3-15
-V_L L1 LAMP
X886A/B X886C
11 F804 11 0V_L 17VAC_RET
5AT 9 18 3
2
T801 17VAC
12 12
3-16
Medical Solutions
Siemens AG
A B C D E F G
COLLIMATOR 2
X813
D801 D805 SEE PAGE 3-4
SMART_RX
1
SMART_TX
1 15 EXT_COLL CPU & PLD +8.2V +5V_S
FUNCTION
SMARTPOINT INTERFACE
15 1
3 6 FILT_MOT_INT S897
X855 2 X839
RESET 20 11 RESET
1
+5V
01.05
4 24 COLL_MOT_RX
X855
COLL MOT_TX 21 21 COLL_MOT_RX COLL_MOT_INT 16 2
FILT_MOT_INT 14 3
2
5 12 FILT_MOT_RX
Page 16 of 19
SP PLM WH 5
SMART_
COLL_MOT_TX 13 4
FILT MOT_TX 22 35 FILT_MOT_RX
RX_
FILT_MOT_TX 8 5
6 5 SMART_RX
GRID_FRONT X856/19 7
X869/9
7 32 GRID_FRONT +5V 0V SEE PAGE 3-17
0V +12V
X830 1 2 4 7 6 3 8 5 9 10
X832
3
D814 +5V
+5V 0V 0V V MOT FILTER/MIRROR
DISK MICROSWITCH
X838 S867
1
D805 STEP MOTOR CONTROL
0V 5V
* Step motor M4:1 drives the right collimator plate, 0V
2
X813 X837
X866 Pin position for each collimator motor connector: 1 FM OUT 1A
4
STEP MOTOR DRIVE
Wiring Diagrams
Mammomat Novation
2 FM OUT 1B
1: OUT 1A
SMART_RX_ X855 2: OUT 2B
X868 3: 0V
SMART_RX
X856 4: QUAD B
X869
5: QUAD A 3 FM OUT 2A
6: 5V 4 FM OUT 2B M5
X857
X872 7: OUT 1B
X871
8: OUT 2A FILTER/MIRROR
X860
0
DISK STEP MOTOR
DR
X859
X873 OUT denotes motor winding signals,
QUAD denotes quadrature motor signals.
X836 X835 X834
X833
* * * *
5
M4:4 M4:3 M4:2 M4:1
X041E
X041E
Medical Solutions
Siemens AG
A B C D E F G
X849
OBJECT TABLE 1
OBJECT TABLE 2 (DIGITAL) X869
+24VF 2 10 D805 CONNECTOR
D801
11
CPU &
REAR VIEW
PLD
S1 M
OBJECT TABLE 1 + 2
FUNC. 1 3
GRID_OUT
1
GRID_MOTOR_N 4 7
1 8 X856 X813
1 TABLE_CONFIG_1 10
6
3 3 TABLE_CONFIG_3 3 1 2 TABLE_CONFIG_2 9
7 9
X829 TABLE X847
X845 1 6 3 TABLE_CONFIG_3 14
Rev. 04 01.05
SPB7-250.844.01
CONFIG
1.8 X845 (SEE 2 2 TABLE_CONFIG_2 2 4 TABLE_CONFIG_0 7
2
BELOW) GRID_O
X840 1 TABLE_CONFIG_1 1 5 CASS_LOADED 4
1.5 X846 2 1
1 8 GRID_OUT 16
MAG TABLE 0 4 TABLE_CONFIG_0 4
3 9 0VD2
GRID_FRONT
GRID_REAR X871
S3 1
0VD
2
19 GRID_FRONT 32
GRID_FRONT
Page 17 of 19
SP PLM WH 5
S2
3
8 GRID_OUT 8 SELECT_WING_1 11
9
6 AEC_POS
S890 18
GRID_MOTOR_N 11
X812 19
S891 *GRID_SPEED
7 10 X860 GRID_MOTOR_N
9 X859 1 7
4
X856
Wiring Diagrams
Mammomat Novation
3 2 1 0 1 3-12 12
S894 11 +24VF
NC NC NC NC NO TABLE LOCKED IN PLACE 2 SELECT_WING_1 0V GRID_MOTOR_N GRID_SP
STEREO
NC NC NC 24V AN: 18X24 WITH GRID (OPTION)
25
NC NC 24V NC AN: 24X30 WITH GRID +24VF
NC NC 24V 24V AN: 18X24 WITHOUT GRID
D801 D802
NC 24V NC NC AN: 24X30 WITHOUT GRID K8
X812
NC 24V 24V 24V AN: MAGNIFICATION 1.8 0V 0V
GRID_SP X824
24V NC NC NC DI: NORMAL (NO MAG) X813
X868
24V NC NC 24V DI: INVALID (TRANSITIONAL)
X855 1
GRID_M A 0V
X875 X876 X869
5
24V NC 24V NC DI: INVALID (TRANSITIONAL) 0V
0V
-1.25
X856 0V
24V NC 24V 24V DI: MAGNIFICATION 1.5
0V X857 X872
24V 24V NC NC DI: INVALID (TRANSITIONAL)
UREG
0VD2 0V
X871
3-17
24V 24V NC 24V DI: MAGNIFICATION 1.8
U
X860 X859
24V 24V 24V NC DI: INVALID GRID_0 X873
0V
I
24V 24V 24V 24V DI: INVALID
0V
3-18
Medical Solutions
Siemens AG
A B C D E F G
Fiber optic communications status:
1
Connection used during
service and manufacturing Likely causes are misconnected fiber optic lines
2
P EL power P
Page 18 of 19
SP PLM WH 5
3
C Data Shield C (BRICK unit)
D Data + D
J Data - J
N Serial Shield N
Xmit + A
Connection to mains
A X603
E Xmit - E
power.
M Recv + M
K Recv - K
F Bot EL ON F
X602
D A
4
J C B
E
Wiring Diagrams
Mammomat Novation
L H F
K
R
V
P N
M
G
V1
U T
X W V2 X606
D A V3
B See page 5-5
J C E X601
L H F
K X602
R M
DR
P N I
X604
V G
U T X761 0 X605
X W X603 X607
5
Master D750
X041E
3-19
X041E
LEFT BLANK
1 2 3 4 5
This page is intentionally left blank.
G
F
E
D
C
B
A DR
Siemens AG SPB7-250.844.01 Page 19 of 19 Mammomat Novation
Medical Solutions Rev. 04 01.05 SP PLM WH 5 Wiring Diagrams
X041E 4-1
MAINS VOLTAGE CONNECTION 0
The generator section of the mammomat requires a line impedance not exceeding the
maximum values according to the table below:
Connections: 0
X707
PE 1 2 3 4
NOV00123
Modules: 0
Fuses: 0
• F2 8A/660VAC
PC-boards: 0
PXCM unit
(Brick unit)
PB0852
HV Tank
X740 mAs-jumper
PB0851
X730 HV inverter
LED:
HV_ON
Potentiometers:
I_OUT X732
HV_READ X725
HV_WISH X726
0VA
X731 PB0854
X720 Tube power
Potentiometers:
Bias Mo Potentiometer:
Bias W Anode drive frequency
Testpoints:
0_VA X724
FIL_CURR_READ X713
MA_READ
PB0855
X721 AC inverter
X722
X723 X712
X705
LED:
AC_ON Fuse F2: Inside the module
X706
X710
X711 PB0850
Mains input converter
LEDs:
28V_prim PB0853
5V_RS232 Mains input
-15VA (Filter unit)
+15VA
28V_sec
5VD
NOV00122
15 16 17 18 19 20 21 22
Mo Mo W
kV mAs / mGy AEC Mo Rh Rh Program Auto Lim.
WARNING
NOV00124
1 2 3 4 5 6 7 8 9 10 11 12 1 13 14
Fig. 3
1 Exposure release buttons with green lights. Both buttons must be pressed during the entire exposure. This is
to ensure that the operator is standing behind the radiation shield during the exposure.
2 +/- kV selection and display
3 MAMMOMAT NovationDR ON/OFF
4 +/- mAs selection and display. In AEC mode (Automatic Exposure Control), the display remains dark before
exposure. After the exposure, the mAs value reached during radiography is indicated.
If the dose calculation option is enabled, the dose value is shown on the same display as the mAs value.
5 ìdrî (digital radiography) is displayed on the film density display to indicate that the MAMMOMAT NovationDR
is ready to be used. If the display is dark the communication with the DR-system is not working. The buttons
are not used.
6 AEC. These buttons are used to activate AEC mode (Automatic Exposure Control). H stands for low dose
exposure and D for high dose exposure.
7 Buttons for selection of anode/filter combination in manual mode.
8 Buttons for selection of one of four pre-programmed exposure parameter settings.
9 Decompression button.
10 Button for storing exposure parameter (see pos 8).
11 Button to select/deselect auto mode.
12 Button to select/deselect automatic decompression after exposure. When off, use foot switch or the knobs for
manual decompression or use the decompression button (see pos 9).
13 Limit button (acknowledgement). Serves to reset the system after errors, indicated by an error code, or faulty
exposures (limit errors), indicated also by the limit lamp and acoustic signal (e.g limit mAs, dose monitoring,
interrupted exposure).
14 Limit lamp. Flashing: Tube temperature is above normal. Lit: Limit error (see pos 13).
15 Not used for digital mammography.
16 Indicates an unknown table configuration. e.g. magnification table not fully inserted.
17 Risk of collision between tube head and floor (limit switch is activated).
18 Not used for digital mammography.
19 Collimator not in correct position or wrong combination of object table and compression plate.
20 Focal spot indication. Indicates focus in operation. Small focus is automatically selected, when a magnifica-
tion table is used.
21 Green light indicates that the unit is ready for exposure. This indicator has the same function as the green
lights on the stand display and the exposure release buttons.
22 Yellow light and an acoustic signal indicate radiation ON.
Switches 0
Test points
Test point Page Description
5VD_1 6-1 Logic supply voltage: 5 volt digital
0VD_1,2,3,4,5 6-1 Logic supply voltage: 0 volt digital
P15VA 6-1 Logic supply voltage: Positive 15 volt analog
0VA_1,2,3 6-1 Logic supply voltage: 0 volt analog
N15VA 6-1 Logic supply voltage: Negative 15 volt analog
5V_RS232 6-1 Logic supply voltage: 5 volt; for communication with PXCM unit
(Brick unit)
0V_RS232 6-1 Logic supply voltage: 0 volt; for communication with PXCM unit
(Brick unit)
5VI_RS232 6-1 Logic supply voltage: 5 volt; for communication with Service-PC,
Opdima WS and Printer
0VI_RS232 6-1 Logic supply voltage: 0 volt; for communication with Service-PC,
Opdima WS and Printer
M_WD_RESET_ 6-1 Signal that resets the master CPU and thus causes the entire Mam-
momat system to restart.
GEN_POWER_ON 6-1 Signal to the generator that turns on the generator and thus the
Mammomat system.
DC_BUS_OK 6-1 Signal from the generator, indicating the DC-bus voltage is within its
allowed voltage window.
HV_OFF_INT_ 6-1 Indicates that the signal M_PLD_HV_OFF_ from the PLD circuit
has detected an event that shall terminate or prevent high voltage
generation.
It causes the signal HV_ON from the Master board to go inactive,
which immediately prohibits high voltage generation.
REQ_HV_ON 6-1 Signal from PLD and CPU for generating high voltage and thus start
the exposure.
It will together with the signal GRID_O_HV_ON set the signal
HV_ON.
HV_ON 6-1 Signal set by the signals REQ_HV_ON and GRID_O_HV_ON out-
putted to the generator. One of two necessary signals, together with
the signal EN_HV_ON, to generate high voltage and thus start the
exposure.
This signal goes low at grid returns, due to GRID_O_HV_ON, thus
causing the intermittent loading mode during exposure.
EN_HV_ON 6-1 Signal from the exposure release buttons at the Control panel to the
generator. One of two necessary signals, together with the signal
HV_ON, to generate high voltage and thus start the exposure.
Note, the generator stops the acceleration power to the tube if this
signal is set longer than about 2.9 sec.
BRAKE_ANODE 6-1 Signal to the generator to stop the anode rotation by braking the
stator of the tube. The braking power (DC current) to the tube con-
tinues as long as the bit is set.
Note, the generator stops the braking power to the tube if this signal
is set longer than about 7.4 sec.
ANODE_FREQ 6-1 Signal from the generator for measurement of the anode rotation
speed. The frequency of the signal is equal to the anode rotation
frequency.
Note, PLD measurement of this signal starts 200 ms after the signal
ACC_ANODE goes low, mean value of 10 pulses is calculated.
Switches: 0
Potentiometers 0
Test points 0
LEDs: 0
Note: If LED H3 not is lit but H4 is lit, indicates the fuse F2 inside
the Mains input converter has blown and need to be replaced.
H4 4-5 Mains input The DC-bus voltage is present (>15 volt) to the HV inverter.
converter
AC_ON 4-5 AC inverter The AC output voltage is present.
HV_ON 4-5 HV inverter The EN_HV_ON signal is set and thus x-ray exposure is on.
Siemens AG
X041E
PE
V CONVERTER
H V INVERTER MODULE FILAMENT
Medical Solutions
1 INPUT CIRCUIT MAINS MODULE CURRENT
1
Rev. 04
BIAS VOLTAGE
POWER ON
01.05
SPB7-250.844.01
OVERVOLT. 0
COIL TRIP PROTECTION
H V-SET I
2
F1
II
F808 F807 28V_PRIM H V_ACT
0V_PRIM
T802
28V_SEC
FLYBACK
Page 1 of 2
BRICK CONVERTER
FLYBACK AUXILIARY
SP PLM WH 5
PXCM
CONVERTER POWER SUPPLY
BOARD 385VDC
0V_SEC
=
FILTER +
~
FOCUS SELECTOR
28V_SEC 5V_RS232 5VD +15VA -15VA FORWARD
0V_SEC 0V_RS232 0VD 0VA CONVERTER
H V_SET
+
3
GND
MA_ACT
AC INVERTER MODULE
T801 MA_SET
= DC POWER 385VDC
GENERATOR: BLOCK DIAGRAM
FIL_CURR_SET
230VAC
50 HZ
MA_ACT
~
FIL_CURR_ACT
TUBE POWER
H V_ACT MODULE
SINE WAVE
GENERATOR
A A
D µP D
4
INPUT
OUTPUT
GEN_POWER_ON
RS232
SERVICE PC
Z801
T803 µP
FILTER
TO OPDIMA
MASTER
Wiring Diagrams
STAND
Mammomat NovationDR
µP
CONTROL
PANEL BOARD
5
5-1
0
5-2 X041E
LEFT BLANK
1 2 3 4 5
Siemens AG
X041E
X707 BOOST-FLYBACK
PE CONVERTER 5 3
Medical Solutions
+DC-BUS
MAINS INPUT MODULE
1
Rev. 04
X705 X724
1 BLEADER
7 1 TUBE POWER
2 -DC-BUS
SUPPLY MODULE
Y Y
01.05
SPB7-250.844.01
3 1 4
8 1 AC INVERTER
2
X9 MODULE
X704 SENSE_CURRENT 7 3
Page 1 of 2
2 28V_SEC
TUBE
SP PLM WH 5
F808
1,3
F807
1,3
GENERATOR:
X802B 6 3 + 7 7
DC_POWER_OK 0VA
+ GEN. CONNECT
X802C 6 3 U REG -15VA 9 9
COMP
-15 V
X603 U REG 5VD
+ +5 V
0VD X701 X760
BRICK GEN. CONNECT
PXCM 14 14
U REG X1
28V_PRIM U REG 5V_RS232
+24V + + +5 V 13 13 MASTER
0V_PRIM 0V_RS232 +15VA +15VA BOARD
5 5
X706
MAINS INPUT & MAINS INPUT CONVERTER
0VA 0VA
28V_PRIM
4
6,8 6,8
1 -15VA -15VA
POWER_ON 5VD 7 7
2
5VD 5VD
9,11 9,11
+ DC_POWER_OK 0VD 0VD
SENSE_CURRENT POWER_ON 10,12 10,12
Wiring Diagrams
- DC_POWER_OK
5V_RS232 5V_RS232
DC_VOLTAGE 385V 15 15
Mammomat NovationDR
V1_GATE
0V_RS232 0V_RS232
RECTIFIED LINE V1_EMITTER AUXILIARY POWER SUPPLY BOARD 16 16
GEN_POWER_ON_P GEN_POWER_ON_P
5
UC 3854A 3 3
GEN_POWER_ON_N GEN_POWER_ON_N
4 4
0
DC_BUS_OK_P DC_BUS_OK_P
V2_GATE
1 1
V2_EMITTER DC_BUS_OK_N DC_BUS_OK_N
2 2
5-3
5-4 X041E
LEFT BLANK 0
1 2 3 4 5
1 2 3 4 5
X858
GENERATOR 1 MA_ACT_P 1 D750 MASTER BOARD 1 P15V_10W 1 1 5 DOUBLE DETECTOR
2 MA_ACT_N 2 2 N15V_10W 2 2 3
3 HV_ACT_P 3 3 RDLI 3 3 1
4 HV_ACT_N 4 X768 4 0V_RDL 4 4 4 X883
5 FIL_CURR_ACT_P 5 5 TRAI 5 5 2
6 FIL_CURR_ACT_N 6 6 SHIELD 6
G
7 HV_SET_P 7
8 HV_SET_N 8 1 DOOR_SWITCH_P DOOR
9 FIL_CURR_SET_P 9 2 DOOR_SWITCH_N SWITCH
10 FIL_CURR_SET_N 10 X754 3 GEN_POWER_AID_P POWER X-RAY TUBE
11 MA_SET_P 11 4 GEN_POWER_AID_N AID
12 MA_SET_N 12
0VD (K8)
X895
1 3 3 K8
13 HV_REGUL_ERR_P 13
2 X858
0VD (K9) 5 5
14 1 MA_ACT_P
HV_REGUL_ERR_N
GENERATOR 14 1 D750 MASTER BOARD DOUBLE DETECTOR K9
2 MA_ACT_N 2
X7551 P15V_10W
3 K8_TEMP
1 1
4
5
4
2 N15V_10W 2 2 3
15 MA_WERCURR_P 15
3 HV_ACT_P 3 RDLI 4 K9_PRESS 6
3 3 3 1 6
16 MA_WERCURR_N
4 HV_ACT_N 16 4 X768 4 0V_RDL 4 4 4 X883
17 GEN_CONNECT_P
5 FIL_CURR_ACT_P 17 5 5 TRAI 1 5 5 2
X804
PB_POWER_ON_ 15 15 1
6 FIL_CURR_ACT_N 6
18 GEN_CONNECT_N
7 HV_SET_P
18
7
6 SHIELD
2
6
PB_POWER_OFF_ 16 16 2 D741
19 GEN_IN_SPARE_1_P
8 HV_SET_N 19 8 1 DOOR_SWITCH_P
3 0VD DOOR 17 17 3
X32
F
2 DOOR_SWITCH_N SWITCH
20 GEN_IN_SPARE_1_N
9 FIL_CURR_SET_P 20 9
4 0VD 1 1 3
FIL_CURR_SET_N 10 X754 3 GEN_POWER_AID_P POWER
21
10
ACC_FREQ_P 21 X756
4 5 5VD
GEN_POWER_AID_N AID 2
X-RAY TUBE
2 1
D740
11 MA_SET_P 11
22 ACC_FREQ_N
12 MA_SET_N 22 12 6 PB_EXP_ON_
X895 3 3 2
23 STATUS_HV_ON_SIGNALS_P
13 HV_REGUL_ERR_P 23 13
1 0VD (K8) 3 3 K8 X5
2 7
0VD (K9) MPS
5 5 4 4 4
14 HV_REGUL_ERR_N 14 K9
24 STATUS_HV_ON_SIGNALS_N 24 X752 X755 K8_TEMP 0VD4
3 8 4 5 5 6
X721 25
15 MA_WERCURR_P
BIAS_VOLT_DETECTED_P 25
15
4 K9_PRESS 6 6
16 MA_WERCURR_N 16
BIAS_VOLT_DETECTED_N 1GRID_OUT_PX804 16
26 17 GEN_CONNECT_P 26 17 1 PB_POWER_ON_ 15 15 1 D801
27 SPARE_P18 GEN_CONNECT_N
27 18 X753
2 2 GRID_OUT_W
PB_POWER_OFF_ 16 16 1 2 X813 D741
19 GEN_IN_SPARE_1_P 19 3 0VD 17 17 3
X32
28 SPARE_N20 GEN_IN_SPARE_1_N
28 20 4 0VD 2 MPS 1 1 2 3
29 ACC_ANODE_P
21 ACC_FREQ_P 29 21 X756 5 5VD 2 2 1
D740
22 ACC_FREQ_N 22
4 GRID_RUNNING 4
30 ACC_ANODE_N 30 6 PB_EXP_ON_ 3 3 2
X5
23 STATUS_HV_ON_SIGNALS_P 23 X757 5 GEN_PREP_RDY 5
31 BRAKE_ANODE_P
24
31
STATUS_HV_ON_SIGNALS_N 24
7 MPS 4 4 4 X815
X721 X752 8 0VD 6 STAND_REQ_HV_OFF_
5 5 6 6
32 BRAKE_ANODE_N
25 BIAS_VOLT_DETECTED_P 32 25
E
BIAS_VOLT_DETECTED_N33 1 GRID_OUT_P 16
33 HV_ON_P26 26
1 D801
27 SPARE_P 27 X753 2 GRID_OUT_WEXT_POWER_ON_
1 X813
34 HV_ON_N 34 2 EXT_POWER_OFF_
28
FIL_CURR_ON_P
SPARE_N
35
28
2 MPS 2 PRODUCTION TEST
35 29 ACC_ANODE_P 29 6 REQ_HV_ON_
36 30 ACC_ANODE_N
FIL_CURR_ON_N 36 30
4
X758 GRID_RUNNING 4 EQUIPMENT
X757 5 7 +15VA
GEN_PREP_RDY 5
X815
31 BRAKE_ANODE_P 31 RL2
37 MA_REGUL_ON_P 37 6 STAND_REQ_HV_OFF_
8 +15VA 6
32 BRAKE_ANODE_N 32
38 MA_REGUL_ON_N
33 HV_ON_P 38 33 1 EXT_POWER_ON_
39 34 HV_ON_N
SEL_LARGE_FOCUS_P 39 34 X759
2 1 EXT_EXP_LAMP_P
EXT_POWER_OFF_
PRODUCTION TEST
40
35 FIL_CURR_ON_P
SEL_LARGE_FOCUS_N 40
35 6 2 EXT_EXP_LAMP_N
REQ_HV_ON_ EXTERN EXPOSURE LAMP
36 FIL_CURR_ON_N 36 X758 EQUIPMENT
7 +15VA
41 SEL_MO_ANODE_P 41 RL2
37 MA_REGUL_ON_P 37 8 +15VA
1 5V_RS232 1
42 MA_REGUL_ON_N
SEL_MO_ANODE_N
38 42 38 BRICK
39 SEL_LARGE_FOCUS_P 39 X759 1 2 0V_RS232
EXT_EXP_LAMP_P 14
43 EN_HU_ON_P 43 2 EXT_EXP_LAMP_N EXTERN EXPOSURE LAMP
40 SEL_LARGE_FOCUS_N 40 3 PREP_DR_DET_P 2
44 EN_HU_ON_N
41 SEL_MO_ANODE_P 44 41
4 PREP_DR_DET_N
45 GEN_OUT_SPARE_1_P
42 SEL_MO_ANODE_N 45 42
1 5V_RS232 1
BRICK15
0V_RS232
D
2 5 14
GEN_RDY_FOR_DR_EXP_P
43 EN_HU_ON_P 43 3
46 GEN_OUT_SPARE_1_N 46 3 PREP_DR_DET_P 2
44 EN_HU_ON_N 44 6 GEN_RDY_FOR_DR_EXP_N
PREP_DR_DET_N 16
47 GEN_OUT_SPARE_2_P 47 4 15
45 GEN_OUT_SPARE_1_P 45
5 7 DR_DET_RDY_FOR_EXP_P
GEN_RDY_FOR_DR_EXP_P 3 4
48 GEN_OUT_SPARE_2_N
46 GEN_OUT_SPARE_1_N 48 46
GEN_RDY_FOR_DR_EXP_N
6 8 16
DR_DET_RDY_FOR_EXP_N 17
47 GEN_OUT_SPARE_2_P 47
7 DR_DET_RDY_FOR_EXP_P 4
1 48 GEN_OUT_SPARE_2_N
DC_BUS_OK_P 1 48 9 DR_SYST_RDY_P 5
8 DR_DET_RDY_FOR_EXP_N 17
2 DC_BUS_OK_N
1 DC_BUS_OK_P 2 1 9 10 DR_SYST_RDY_N
DR_SYST_RDY_P 5 18
2 DC_BUS_OK_N 2 10 DR_SYST_RDY_N
11 18
DR_DET_ERR_P 6
3 GEN_POWER_ON_P 3
3 GEN_POWER_ON_P 3 11 DR_DET_ERR_P 6
4 GEN_POWER_ON_N 4 12 DR_DET_ERR_N 19
4 GEN_POWER_ON_N 4 12 DR_DET_ERR_N 19
5 +15VA 5 +15VA 5 5 X761X761
13 13 GEN_POWERED_ON_P
GEN_POWERED_ON_P 7 X602 7 X602
6 0VA 6 14 GEN_POWERED_ON_N
14 20
GEN_POWERED_ON_N 20
6 0VA 6
7 -15VA 7 15 X_RAY_ON_P 8
7 -15VA 7 15 X_RAY_ON_P 8
8 0VA 8 16 X_RAY_ON_N 21
8
X701
0VA 8 16 X_RAY_ON_N 21
X701 9 5VD 9 X760 17 SHIELD
10 0VD 10 18 22
12 0VD 12 20 0V_RS232 23
11 5VD 13 11
GEN_CONNECTION_LOOP 13 21 19
PXD_DR TXD_DR 11 10
12 0VD 14 12
GEN_CONNECTION_LOOP 14 22 20
0V_RS232 0V_RS232 24 23
15 5V_RS232 15 23 5V_RS232 12
13 GEN_CONNECTION_LOOP 13 21 PXD_DR 11
16 0V_RS232 16 24 0V_RS232 25
14 GEN_CONNECTION_LOOP 14 22 0V_RS232 24
HV_ACT 1 2 RXD_S_PC 2
15 5V_RS232 15 23 5V_RS232 12
X805
TXD_OPD_WS
PRODUCTION MA_AT 0VA 2 6 X765 3
3 TXD_S_PC OPDIMA WORKSTATION
3
RL1 X762
4 DTR_OPD_WS SERVICE - PC
TEST 0VA 3
7
X763 5 5
0VI_RS232 0VI_RS232 5
8
EQUIPMENT HV_SET 0VA 4 9
+15VA 6 DSR_OPD_WS
8 2 RXD_OPD_WS
GND_OPD_WS
MA_SET 0VA 5 10
3 TXD_OPD_WS
0VA 6 X765 2 RXD_PTR OPDIMA WORKSTATION
B
X769 RL1 3 4
TXD_PTR DTR_OPD_WS PRINTER/ID-FLASHER
7
T801 10V5_AC2 1 X764X763
5 0V_RS232
5 0VI_RS232
10V5_AC1 8 2 X772 A GND_PTR
+15VA 6
GND_PTR
DSR_OPD_WS
B
0VA 9
8 GND_OPD_WS
0VA 10
2 RXD_PTR
X769 3 TXD_PTR PRINTER/ID-FLASHER
T801 10V5_AC2 1 5 0V_RS232
X764
10V5_AC1 2 X772 A GND_PTR
B GND_PTR
A
Siemens AG
X041E
Medical Solutions
28V_sec V_Bias
1
0V_sec
Bias_W Bias_Mo
PWM R111 R109 X721
385VDC Filament converter IC det. Bias_volt_detected_p 25
T201
logic
26
Rev. 04
on on
off off
V_Bias_ret
+15V_prim
01.05
X726
SPB7-250.844.01
Bias_voltage 1 PK G
2
N201
Inv YE h1M
V204
UCC3804
Sel_Mo_anode_p Small_focus_W 3
0V_prim h2M
BU
Small_focus_Mo 2
Page 1 of 1
h1W
PRINCIPLE DIAGRAM
GY
Large_focus_W 5
SP PLM WH 5
L201 L202 L203
Large_focus_Mo 4
+ + WH h2W
385VDC_ret GND 6 BN h3
+15VA 7 BL
3
Generator
door
GENERATOR - TUBE POWER SUPPLY:
Voltage limit
TUBE FILAMENT AND FOCUS SELECTION
Current limit
9 Fil_curr_set_p Fil_curr_set
4
10
MA_regul_on_p TP
37
FIL_CURR_READ
38 X721
Fil_curr_act_p 5
11 MA_set_p
MA_set 6
12 MA_set_n
Wiring Diagrams
MA_act
MA_overcurr_p 15
Mammomat NovationDR
Logic 16
circuit TP
Fil_curr_act
MA_READ
X720
5 MA_sense MA_act_p 1
5
6 MA_sens_ret 2
TP
X721 0VA
39 Sel_large_focus_p*
40
41 Sel_Mo_anode_p*
* = active low
5-6
42
A B C D E F G
X722 X721
Siemens AG
X041E
9 Gen_connect_p 17
Medical Solutions
1
1, 3 28V_sec
2, 4 0V_sec
5 +15VA
6, 8 0VA
Rev. 04
385 VDC
7 -15VA
385 VDC_ret
To Filament Converter
01.05
SPB7-250.844.01
Rotation Inverter
2
X724
DC-bus Tube
1 V6 V5
X725
+ High High
Motor_II 2 WH II
Page 1 of 1
220mF & &
4 Low Low
SP PLM WH 5
3.3mF 2.2mF
side side L2
DC-bus_ret V13 V14
drive drive Motor_I YE I
4
Motor_0 1 BN 0
3
X723
28V_prim 15V_prim
1 15V_reg
To Filament Converter
2
V6+V14
28V_prim_ret 0V_prim PWM Switch
Circuit V13+V5 logic Square
Wave
circuit
GENERATOR -TUBE POWER SUPPLY:
Enable pin
PRINCIPLE DIAGRAM, TUBE ROTATION
Internal
freq_adj.
X720
Anode 28V_sec 1,3
4
drive freq.
0V_sec 2,4
X721
3 HV_act_p 11
28V_sec 4 HV_act_n 12
7 HV_set_p 13
Wiring Diagrams
8 HV_set_n 14
X721 Time-out 13 HV_regul_err_p 15
To To
29 Acc_anode_p acc. 14 16
D750 HV
Mammomat NovationDR
logic
30 Master 23 Status_hv_on_signals_p 18 inverter
board 24 20
Time-out
31 Brake_anode_p brake 33 HV_on_p 7
logic
5
32 34 8
43 En_hv_on_p 9
44 10
Rotation frequency
0
18 Gen_connect_n 19
21 Anode_freq_p
22
5-7
A B C D E F G
X730: 3 TP 5.1V_ref +24V
Siemens AG
I_OUT
X041E
Medical Solutions
X4
1
2 Overcurrent
X730:4 Comp
HV inv.
MA_SENSE_RET current
X731:6 1 ref
TP
HV_WISH
N4 UCC3806 OSC. 1 G1
Current
X731:13 HV_SET_P transformer
Rev. 04
Limit
+ A-out
7.5V
_
B-out
01.05
SPB7-250.844.01
X730:5 HV_act_+
E1
2
TP B-out
TP 0VA
HV_READ
Page 1 of 1
SHDWN CURSNS
SP PLM WH 5
X4
4
Interlock
RT CT
X731:7 HV_ON_P
3
E2
X731:9 EN_HV_ON_P HV_ON A-out
PRINCIPLE DIAGRAM, CONTROL LOGIC
+24V
X731:19 GEN_CONNECT
0
X730:9
OSC. 3 G3
4
HV_READ A_out
X731:18 STATUS_HV_ON_SIGNALS_P
E3
HV A-out
X731:15 HV_REGUL_ERR_P Time out
circuit
Wiring Diagrams
OSC. 1
Mammomat NovationDR
+24V
V-reg circuit OSC. 3
X731:3 Comp +24V
ref OSC. 4
X731: 2
0V_sec +12V
5
V-reg
X731: 4 OSC. 4 G4
0V_sec=0VA
X730: 1, 2, 6, 8, 10
E4
X731: 8, 10, 12, 14, 16, 20 B-out
5-8
A B C D E F G
Siemens AG
X041E
Medical Solutions
1
G1 G3
Rev. 04
E1 E3
01.05
SPB7-250.844.01
AC INVERTER
2
+ G2 G4
TUBE POWER SUPPLY
E2 E4
Page 1 of 2
SP PLM WH 5
M1
M2
PRINCIPLE DIAGRAM
OVERCURRENT
2 COMP
X4
1
REF
HV TANK
T1 6.6k X3 HV CABLE
C1 94M
X6 5 X730 HV INVERTER
HV_ACT
X-RAY
C2 18.8k TUBE
10c 1, 2, 6, 8, 10
GENERATOR - HV INVERTER AND HV TANK:
4 MA_SENSE_RET 0V_SEC
3 MA_SENSE
X5 X731
Wiring Diagrams
2, 4, 8, 10, 16, 20
X720
MAS CONNECTOR
Mammomat NovationDR
TUBE
POWER
0V_SEC X726
SUPPLY
0VA 6
5
GENERATOR DOOR
5-9
A B C D E F G
Siemens AG
STAND CONTROL PANEL
PB_POWER_ON
Medical Solutions
D750 D741 D741 S2 S1 D740
PB_POWER_OFF 1
1
2 16 16 2 LIM DECOMP AUTO PROG_4 PROG_3 PROG_2 PROG_1 AF_3 AF_1 AEC_D AEC_H
Rev. 04
EPROM
V2 V3 V4 V6
X27A X27B
S1 0V_P 5V_P RAM CPU VCC GND MPS
4
01.05
SPB7-250.844.01
D740
2
CAS
TBL
TILT PROG_1 AEC_H
CPU AEC_D OK
AEC PROG_2 XRAY
80C535 COLL PROG_3 AF_1
PB_EXP AF_2 LIM
FOC_L PROG_4
X5 J1 AF_3 DECOMP
Page 2 of 2
_ON FOC_S AUTO
RXD
6 3 3 2
SP PLM WH 5
MPS 12MHz X4 LED_3
MPS TXD LED_0 LED_1 LED_2
7 4 4 4
VCC S1
5VD
X27B.3
5 2 2 1 RESET
0VD GND OK EN
4 1 1 3 V3 V4 V6 UREG R113 X27A.3
24V_P +18V
S27A S27B
8 5 5 120R
6
3
PWR 0V_P
V2
PRINCIPLE DIAGRAM, CONTROL PANEL
X27B.4
DECOMP
LIM
2 S20
S27A
X27A
24V_P 1 MEM
1
S17
Wiring Diagrams
12
S27B 2
11
X27B
5
1 6
Mammomat NovationDR
6 6
5-10
2 7
7 7 8
12 12 9
18 18 10
5
19 19
X041E
X041E D750 MASTER BOARD: 6-1
CIRCUIT LOCATION 0
1 2 3 4 5
D750 MASTER BOARD
S4
AEC
CPU
PROM
G
AEC
5VD_1 PLD
PROM
0VD_1
0VA_1
F
X768
0VA_3
0VD_2
M_WD_RESET_
FIL_CURR_ACT
HV_ACT
E
X765
0VD_3 MA_ACT
HV_SET
N15VA
FIL_CURR_SET
MASTER P15VA
CPU
PROM MA_SET
0VA_2
D
ANODE_FREQ
X752
DC_BUS_OK HV_OFF_INT_
EN_HV_ON
C
ACC_ANODE
HV_ON
BRAKE_ANODE S3
0VD_5 X758 GEN_POWER_ON
S2
X754
X755 REQ_HV_ON
X757 0VD_4
MASTER X760
PLD
PROM X753
B
X756
DR_SYST_RDY
DR_DET_RDY_FOR_EXP
DR_DET_ERR_ X759
5V_RS232
0V_RS232
GEN_POWERED_ON_ 5VI_RS232
PREP_DR_DET_ X772
GEN_RDY_FOR_DR_EXP_
A
X_RAY_ON_
0VI_RS232
X761
X769 X764 X762
X763
Siemens AG SPB7-250.844.01 Page 1 of 2
DR
Mammomat Novation
Medical Solutions Rev. 04 01.05 SP PLM WH 5 Wiring Diagrams
GENERATOR:
6-2 CONTACT DIAGRAM 0
X041E
1 2 3 4 5
X5
X3
G
X6
X8
X1
X7
X2
F
E
D
S6
S7
X706 To
X760 Master
board
F1 F2
F2 X705 Z801
0.5AT 0.5AT Boost-Flyback T803
To
+ DC-bus Filter Opdima
X772
B
From
T801: S7
Brick PXCM
X9 F807 T802 X802c 2AT
X603
F808 X802b X802b 2AT
A
Incoming mains
Siemens AG SPB7-250.844.01 Page 2 of 2
DR
Mammomat Novation
Medical Solutions Rev. 04 01.05 SP PLM WH 5 Wiring Diagrams
X041E 7-1
Changes to previous version 0
2 9 Updated.
3 2 Updated.
3 6 Updated.
3 15 Updated.
3 16 Updated.
3 17 Updated.
6 1 Updated.