Technical specification
APPROVAL INFORMATION
Creator:
Yasemin 28/12/2021
Signature & Date
Approver:
Huyngtai KiM
28/12/2021
Signature & Date
Table of contents
TABLE OF CONTENTS ................................................................................................................................................. 3
1. OVERVIEW .......................................................................................................................................................... 4
1. Overview
1.1 Overview of device
This device consists of X-ray control, high-voltage generator, X-ray tube, X-ray tube support, irradiation equipment,
detector (detector), patient table, patient photography stand, digital imaging device, etc.
A high frequency inverter (Inverter) X-sensor voltage device designed to generate X-rays by combination of tube voltage,
tube current, irradiation time, etc. so that it can be filmed at various angles for diagnosis of the patient's skeletal, respiratory,
and urinary systems.
The digital imaging system is used to obtain the images taken by the X-ray unit as radiation from the X-ray unit passes
through the human body and is transmitted by the detector.
2.4 Contra-Indication
The GR10X-50K and GR10X-40K are not to be used for pediatric patients and not intended to be used for mammography.
2.5 Principle
When the external rated power is transferred to the X-ray control unit, set the tube current and time required for each
shooting and issue an X-ray inspection order, conduct the primary measurement of the high voltage generator.
When the set voltage is applied, a high pressure is generated to make X-rays on the secondary side.
When this high pressure is applied to the X-ray tube, an X-ray occurs when the electrons are released and hit the target
at high speed.
This X-ray is passed to the body part where you want to diagnose it.
X-rays passing through the subject form interactions through the selenium layer of the detector and convert them into
electrical signals, and the charges are collected through electrodes located in each pixel and output to the signal
The signal processing department performs the amplification and digital data conversion process, and the data (imaging
device) delivered to the workstation via cable is stored as a DICOM file, a medical imaging standard, through an
operational program.
The stored images are sent by the operating program to the Picture Archiving & Communication System (PACS) server
for use in image reading.
M-38
Light source (LED lamp 24 VAC 1A/24VDC 1A)
Inherent filtration 1.2 mmAl
Added filteration 0.5 mmAl (TBD)
Weight 5 kg
Max field size (less than 48cm x 48cm at SID 100 cm)
Collimator
Min field size (less than 1 cm x 1 cm at SID 100 cm)
24VAC, 50/60 Hz
Form of X-ray Beam Field (Rectangular)
Max Tube Voltage 150 kV
Light source (LED lamp 24 VAC 4A/24VDC 4A)
Crux 104i
Inherent filtration 1.0 mmAl
Weight 8 kg
Max field size (less than 43cm x 43cm at SID 100 cm)
Min field size (less than 2 cm x 2 cm at SID 100 cm)
Power input (220VAC, 50/60 Hz)
Tube Stand XETS-W1 Column Top LED display according to the detector
Automatic display of SID and tube rotation angle
2) Generator software
Manufacture : Delta ELECTRONICS, INC.
Project name: DMP-100R (Model: DMG-40KAX A, DMG-50KAX B)
Software name: Firmware Version: S00E25_Add40kVComp3kV_9A5E
GUI version: GUI Ver5.2.8d
3.4 Appearance
1) Tube Stand
2) Bucky Table
3) Bucky Stand
4) Console
Basic Console
Mini Console
4. Block diagram
4.1 Block Diagram
Bucky Stand
Bucky Stand
Board
Main Power
Tube Stand
Bucky
Line HV, Rotor
Tube
Base
DAP
Indicator
Tube Stand SYS CON
Power Board
Generator
AEC Chamber
Collimator
Trigger
Signal Detector
Handle Bar
Bucky
Indicator
Data Foot SW
Workstation
AEC Chamber
Detector
4.2 Description
Item Description
Adjusts the X-ray irradiation condition and controls the
Console
display device.
X-ray
Generator
Generator Generates a high voltage applied to the apparatus.
5. Environmental Specifications
Item Environmental Condition
Temperature 10°C ~ 40C°
Relative Humidity Range 30%RH ~ 65%RH
Atmospheric Pressure Range 700hPa ~ 1060hPa
Storage temperature -10°C ~ 55C°
FXRD-2530VAW FPD-2530VAW
1
FXRD-2530VAW PLUS FPD-2530VAW PLUS
FXRD-3643VAW FPD-3643VAW
2
FXRD-3643VAW PLUS FPD-3643VAW PLUS
FXRD-4343VAW FPD-4343VAW
3
FXRD-4343VAW PLUS FPD-4343VAW PLUS
FXRD-1717VA FPD-4343VA
4
FXRD-1717VB FPD-4343VB