Calibration of Disintegration Apparatus

You might also like

Download as docx, pdf, or txt
Download as docx, pdf, or txt
You are on page 1of 1

CALIBRATION OF DISINTEGRATION TEST APPARATUS

DEPARTMENT: _ FREQUENCY: ONCE IN A MONTH


EQUIPMENTID NO.: _

INSTRUMENT MAKE& MODEL DISINTEGRA TION TEST APPARA TUS


CALIBRATION DONE ON
CALIBRATION DUEON

PARAMETERS PARAMETERS TOLERANCE


PARAMETERS SET OBSERVED LIMITS
Left Right

-- 28 -32
Cycles I minute Cycles/minut
(Ref.-BP}

Distance traveled during --


raising & lowering of 50 to 60 mm
cells
Temperature Bath
With Beak 37.0°c 36 to 38°C (R
Thermometer er BP}
Temperature Bath
displayed on Beak 37.0°c
the screen er
Timer 30 minutes 29 to 31minu

Sieve Integrity Should be int

Remarks: TheInstrument is calibrated & qualified I out of calibration & not qualified for use.

DONE CHECKED BY: DATE: