Tools For Developing A Quality Management Program Proactive

You might also like

Download as pdf or txt
Download as pdf or txt
You are on page 1of 4

Int. J. Radiation Oncology Biol. Phys., Vol. 71, No. 1, Supplement, pp.

S187–S190, 2008
Copyright Ó 2008 Elsevier Inc.
Printed in the USA. All rights reserved
0360-3016/08/$–see front matter

doi:10.1016/j.ijrobp.2007.07.2385

QA FOR RT SUPPLEMENT

TOOLS FOR DEVELOPING A QUALITY MANAGEMENT PROGRAM: PROACTIVE


TOOLS (PROCESS MAPPING, VALUE STREAM MAPPING, FAULT TREE ANALYSIS,
AND FAILURE MODE AND EFFECTS ANALYSIS)

FRANK RATH, M.S.I.E.


Department of Engineering Professional Development, University of Wisconsin–Madison, Madison, WI

This article examines the concepts of quality management (QM) and quality assurance (QA), as well as the current
state of QM and QA practices in radiotherapy. A systematic approach incorporating a series of industrial engineer-
ing–based tools is proposed, which can be applied in health care organizations proactively to improve process
outcomes, reduce risk and/or improve patient safety, improve through-put, and reduce cost. This tool set includes
process mapping and process flowcharting, failure modes and effects analysis (FMEA), value stream mapping, and
fault tree analysis (FTA). Many health care organizations do not have experience in applying these tools and there-
fore do not understand how and when to use them. As a result there are many misconceptions about how to use
these tools, and they are often incorrectly applied. This article describes these industrial engineering–based tools
and also how to use them, when they should be used (and not used), and the intended purposes for their use. In
addition the strengths and weaknesses of each of these tools are described, and examples are given to demonstrate
the application of these tools in health care settings. Ó 2008 Elsevier Inc.

FMEA, Risk analysis, Quality assurance, Quality management.

INTRODUCTION Each of these reports prescribes a series of QA-related activ-


ities or approaches intended to verify the technical compo-
Although the terms ‘‘quality management’’ (QM) and ‘‘qual-
nents of various radiation therapy processes that are heavily
ity assurance’’ (QA) are used extensively, there is little com-
mon agreement on their exact meaning, and there are many based upon testing, inspection, or verification of hardware
misconceptions about QM/QA systems, procedures, and and software performance parameters.
tools. The key to a successful QM/QA is a relentless focus Looking at one of these reports in more detail gives some
on processes and on giving people involved in organizations’ insight into the general QA strategy of these AAPM task
processes the tools necessary, and also the responsibility, to groups, as well as some potential problems with that ap-
improve quality. These tools include failure mode and effects proach. The report from AAPM TG 53 concentrates on QA
analysis (FMEA), fault tree analysis (FTA), flowcharting, and for the treatment planning process. It addresses most of the
process mapping. Unfortunately this approach to QM/QA has technical components of radiation therapy planning (RTP).
not been widely adopted in the manufacturing or service in- The report calls for more than 170 tests or verifications to
dustries or in the health care field. There are several reasons be performed on several process components of RTP. Fol-
for this lack of success, but the main reason is nonuse or mis- lowing these guidelines would be a Herculean task for
use of the tools identified above. Both QM and QA strategies most radiation therapy departments. It would require a signif-
in most organizations heavily depend upon inspection and icant number of hours just to perform the tests or verifica-
removal of defective products (or rework) in manufacturing, tions, not to mention the QA management effort to plan
or waiting until there is a failure and then trying to identify and control the QA process. It would also almost certainly
the causes of those failures with the hope of preventing add cost and could tie up valuable equipment for testing
them from occurring again in service organizations. and verification.
Over the past few decades, the American Association of The approach used by TG53 focuses on the treatment plan-
Physicists in Medicine (AAPM) has convened several task ning system, and how the system interacts with the RTP pro-
groups to address the issue of QA in radiation therapy; these cess. This is an improvement over previous AAPM task
include Task Group (TG) 40 (1), TG 45 (2), and TG 53 (3). groups that looked primarily at the technical components of

Reprint requests to: Frank Rath, M.S.I.E., Department of Conflict of interest: none.
Engineering, Professional Development, University of Wisconsin– Received April 16, 2007, and in revised form July 18, 2007.
Madison, Madison, WI 53706. Tel: (608) 263-5989; Fax: (608) Accepted for publication July 19, 2007.
263-3160; E-mail: rath@engr.wisc.edu
S187
S188 I. J. Radiation Oncology d Biology d Physics Volume 71, Number 1, Supplement, 2008

radiation therapy such as the equipment and software. TG53 the realization of significant improvements in processes and
does not look at all of the components of the RTP process, the resulting increase in quality and productivity.
however. For example it does not look at the process steps in- Use of FMEA appears to have the highest potential for as-
volving the evaluation of information and the resulting deci- sisting the radiation therapy community in improving overall
sions made when developing an RTP. An effective QA processes and in reducing and controlling the risk of injury
system needs to address all of the components of a specific without overtaxing resources. The first step in using FMEA
process, not just the technical components or the operator– to analyze a process is to define that process. Process maps,
equipment interfaces. process trees (5), and process flowcharts (7, 10) are all effec-
Currently AAPM TG 100 (4, 5) is reviewing reports from tive tools for defining a process. Process flowcharts offer the
previous task groups and from several professional organiza- most flexibility and are one of the easiest to use tools to use in
tions. This group is also reviewing ISO guidelines in an effort defining a process. When using any of these tools, a system-
develop a suitable general QA approach that ‘‘balances pa- level top–down or a bottom—up systematic approach can be
tient safety and quality versus resources commonly available used. In a system-level top-down approach, the steps below
and strike[s] a good balance between prescriptiveness and should be followed:
flexibility.’’ The TG 100 initiative identifies three industrial
engineering–based tools as potential components of a QA 1. Flow chart the major steps in the process.
management system in radiation therapy: namely, process 2. Perform an FMEA on those steps. The FMEA will identify
trees, FMEA, and FTA. In addition TG 100 recommends those steps with the highest likelihood of failure or injury.
that these tools be used to assess risks or hazards in radiation 3. For those process steps, develop a detailed process
treatment processes. There are a few potential problems with flowchart.
these recommendations, however. The first is that the general 4. Perform an FMEA on the detailed process steps, resulting
radiation therapy community is concerned that the use of in the identification of only those detailed process steps
these tools to reduce the risks or hazards associated with ra- with the highest likelihood for failure or injury.
diation therapy will require a considerable amount of addi- When using a bottom–up systematic approach, the steps
tional resources. Although this should be a concern, below should be followed:
process mapping, flowcharting tools, and FMEA (5–10)
1. Develop a flowchart for the major steps in the process and
have been used for decades by the medical device and phar-
all of the detailed process steps nested beneath them (this
maceutical industries, among others, to reduce the level or
can be difficult and time consuming).
risks or hazards in products and processes with many positive
2. Perform an FMEA on all of the detailed process steps and
results. Also if the organizations are trained in the use of these
the higher level process steps they flow into, resulting in
tools and if these tools are applied with the assistance of ex-
risk assessments for all detailed and higher-level pro-
perienced facilitators, there are few if any additional re-
cesses, both those with a low likelihood for failure or in-
sources required, and the resulting process improvements
jury and those with a high likelihood for failure or injury.
will increase the effectiveness and productivity of the organi-
zation. A second potential problem is TG 100’s recommen- A system-level, top–down process flowchart for the radia-
dation that these tools be used for risk and hazard analysis tion treatment and delivery (RTPD) process is shown in
versus overall process improvement. This might prevent Fig. 1. This flowchart was developed with the assistance of

No

Is
Patient
tumor yes Physician
CT/MRI positioning
present evaluation
strategy

Initial RTP -
Position Verify RTP- Dosimetry - How much PET/CT
patient phantom + develop RTP radiation and imaging
where

Check equip
Deliver output and
treatment other
parameters

Fig. 1. A system level, top-down process flow chart for radiation treatment planning and delivery. CT = computed tomog-
raphy; MRI = magnetic resonance imaging; PET = positron emission tomography; RTP = radiation therapy planning.
Tools for developing a quality management program d F. RATH S189

Fig. 2. Standard process failure mode and effects analysis (FMEA) form.

the RTPD team at University of Wisconsin Hospitals. A The worst-case scenario should always be identified for
rough-cut FMEA of the process indicated that the step of each level of effect, and any safety or injury effects always
‘‘positioning the patient’’ had a high likelihood to fail, which take priority over all other effects. For the potential failure
could result in injury the patient. mode, ‘‘radiation treatment delivered to location outside of
Figure 2 shows is the standard form used in completing ei- specified location in the treatment plan,’’ there are several po-
ther a system-level top–down or a bottom–up, systematic tential end effects including tissue or organ damage, the tu-
FMEA (5, 6). For the RTPD process step ‘‘position the pa- mor not being treated with the specified radiation dose, or
tient,’’ the following steps should be adhered to when per- the patient experiencing irreparable damage and/or death. Ef-
forming an FMEA: fects should always be stated given the worst-case scenario.

1. Identify the intended objective or purpose of each process 5. Identify all of the process controls currently in place that
step: ‘‘Position the patient for optimal delivery of radia- achieve the following:
tion—both dose and delivery location.’’ a. Prevent the causes of failure modes from occurring.
2. Identify the potential failure modes for each process step: b. Detect failure modes when they do occur.
‘‘Radiation treatment delivered to location outside of c. Mitigate the severity of the effects resulting in failure
specified location from the treatment plan’’ would be modes when they do occur.
one, and there would likely be several more. Process controls are procedures, practices, policies and
3. Identify the causes of each potential failure mode: The methods put in place to improve the reliability, repeatability,
failure mode identified in the previous step will have sev- and quality outcomes of processes. They usually cover equip-
eral causes. One cause might be, ‘‘Incomplete or incorrect ment, technicians or operators, training, work instructions,
work instructions detailing the patient positioning pro- and so forth. The most effective process controls are those
cess, or inadequate training.’’ Causes of failure modes that significantly reduce the likelihood of potential causes of
come from the following general categories: hardware/ failure modes from ever occurring, or that result in the failure
equipment, software, physicists/technicians, patient, mode being detected before the process fails or a patient is in-
work instructions, and training. Each cause should specify jured. Technician or operator training, detailed work instruc-
how it led to the patient being in the wrong position. Each tions, process visual aids, preventive maintenance, process
potential failure mode will have many potential causes. verification, and equipment calibration are effective controls
4. Identify the local, downstream, and end effects of the fail- that can prevent the occurrence of potential causes or detect
ure mode. The local effect identifies what would happen a failure mode before there are any negative consequences.
during the process step to the patient or the process In most process FMEAs it is difficult to moderate the severity
when the failure mode occurred. The downstream effect of the resulting effects when a failure mode occurs. For the
identifies what would happen later if the failure mode oc- failure mode ‘‘radiation treatment delivered to location other
curred. The end effect identifies what would ultimately than specified in the treatment plan,’’ it would be virtually im-
happen to the patient if the failure mode occurred. For possible to reduce the severity of the effects on the patient.
the failure mode ‘‘radiation treatment delivered to location 6. Judge the effectiveness of the current process controls, us-
outside of specified location in the treatment plan’’: ing three independent ranking scales (5, 6), with each
a. There is likely no local or immediate effect on the patient. scale being from 1 to 10:
b. There is no likely downstream effect on the RTPD  The likelihood of an occurrence of a cause of a potential
process, but there is a likely downstream effect on failure mode (1 indicates that the cause will not occur,
the patient—the patient’s condition would worsen be- and 10 that the cause is inevitable; known as the occur-
cause of not treating the tumor and possibly destroying rence scale).
unintended tissue.  The likelihood of detecting a failure mode if it occurs (1
c. The end effect on the patient ultimately could be death. indicates that the failure mode will always be detected,
S190 I. J. Radiation Oncology d Biology d Physics Volume 71, Number 1, Supplement, 2008

and 10 that it will never be detected; known as the de- or opinions regarding the process, which in turn result in
tection scale). a lot of time spent debating occurrence, severity, and de-
 The severity of effects of a failure mode when it occurs tection rankings. It is important to keep the analysis mov-
(1 indicates no effect, 10 that the effect is very serious; ing; and when such debates do occur, it is usually best to
known as the severity scale) on both the end user (the pa- use an averaging technique. RPNs range from 1 to 1,000.
tient and/or hospital personnel) and the RTPD process. An RPN score of 125 or more indicates that some correc-
The overall risk level for process failure or injury is deter- tive action should be considered. The highest RPNs al-
mined by multiplying the occurrence score by the detection ways need to be addressed first.
score by local, downstream, and end effects severity scores. 8. Identify recommended actions (process controls) that will
Because there are three separate effects (local, downstream, reduce the risk of injury. The highest RPNs identify the pro-
and end effects), there are three risky priority numbers cess steps that are most likely to fail or result in injury, given
(RPNs) for each failure mode and cause combination. Never the existing process controls. The recommended actions
reduce the severity score of an end effect because the team should reduce the occurrence and detection scores for the
performing the FMEA judges that the cause of a failure highest RPNs, when they are implemented. The end result
mode resulting in that end effect almost never occurs. A se- should be better control of the process being analyzed, lead-
rious end effect score of 10 with a low occurrence score of ing to fewer process failures and fewer injuries or deaths.
2 and a relatively moderate detection score of 4 would result 9. Estimate new occurrence, severity, and detection scores
in an RPN of 80, which would indicate a low level of risk for for the process after implementing the recommended ac-
process failure or injury. However anytime a local, down- tions and calculate the resulting RPN.
stream, or end effect received a high severity score during 10. The last step in completing an FMEA is to identify the
an FMEA, that process step would need to be closely evalu- individuals responsible for implementing the recommen-
ated. Even though the occurrence score in the above example ded actions that will improve the process controls; a com-
is low, there is a distinct probability that the cause of the fail- pletion date also should be entered.
ure mode might occur. Also there is a high probability in the An FMEA, as a document, is fluid, and should be reviewed
above example that if the failure mode did occur, it would be and modified on a regular basis as processes or technologies
detected before something serious happend; however no de- change for effective risk assessment and management.
tection or inspection step is completely foolproof. The FMEA process should ideally be led by an experi-
7. Completing the scoring–ranking and RPN calculation enced facilitator, which can improve the consistency of an
phase of an FMEA can, and often does, lead to some spir- FMEA program and also minimize the number of hours it
ited discussions. It is highly recommended that a cross- takes to complete an FMEA. Formal FMEA training for
functional group familiar with the process being reviewed health care systems improves the quality of the FMEA effort
performs the FMEA. This often results in divergent views and can also reduce the time it takes to complete an FMEA.

REFERENCES
1. Kutcher GJ, Coia L, Gillin M, et al. Comprehensive QA for able at: http://www.oncologymeetings.org/quality_assurance/
radiation oncology: Report of AAPM Radiation Therapy Com- scientific_program.htm#sched. Accessed June 5, 2007.
mittee Task Group 40. Med. Phys 1994;21:581–618. 6. Stamatis DH. Failure mode and effect analysis: FMEA from
2. Nath R, Biggs P, Bova F, et al. AAPM code of practice for theory to execution. Milwaukee, WI: American Society for
radiotherapy accelerators: Report of AAPM Radiation Therapy Quality Control; 1995.
Task Group No. 45. Med. Phys 1994;21:1093–1121. 7. Fletcher CE. Failure mode and effects analysis: An interdisci-
3. Frass B, Doppke K, Hunt M, et al. AAPM Radiation Therapy plinary way to analyze and reduce medication errors. J Nurs
Committee TG53: Quality assurance program for radiotherapy
Admin 1997;27:19–26.
treatment planning. Med. Phys 1998;25:1773–1836.
8. Thomadsen B, Lin SW, Laemmrich MS, et al. Analysis of treat-
4. American Association of Physicists in Medicine Committee
ment delivery errors in brachytherapy using formal risk analysis
Tree Page for Task Group 100: Method for evaluating QA needs
in radiation therapy. American Association of Physicists in techniques. Int J Radiat Oncol Biol Phys 2003;571492–508.
Medicine web site. Available at: http://www/aapm.org/org/ 9. JCAHO. Failure mode and effects analysis in health care:
structure/default.asp?committee code=100. Accessed May 23, Proactive risk reduction, 2nd ed. Oakbrook Terrace, IL:
2007. JCAHO; 2005.
5. Huq MS. A Review of AAPM TG 100. Presentation at AS- 10. Hansen DA. Flowcharting help page [tutorial]. Available at:
TRO Quality Assurance of Radiation Therapy and Challenges http://home.att.net/dexter.a.hansen/flowchart/flowchart.htm.
of Advanced Technology Symposium, February 2007. Avail- Accessed May 28, 2007.

You might also like