Professional Documents
Culture Documents
URT180-Mantenance Manual
URT180-Mantenance Manual
1. Overview............................................................................................................................................................1
2. Common malfunction and remedy................................................................................................................ 1
2.1 Malfunctions due to improper installation environment...................................................................1
2.2 Common malfunction and remedy......................................................................................................1
2.3 Check whether power supply is normal............................................................................................. 3
3. Malfunction and remedy..................................................................................................................................4
3.1 No response when turn on instrument............................................................................................... 4
3.2 No display on the screen......................................................................................................................5
3.3 The test results cannot be printed out through the internal printer............................................... 6
3.4 Inactive reagent strip sensor................................................................................................................7
3.5 Time on the screen is abnormal.......................................................................................................... 8
3.6 Strip holder moves abnormally(move outward when turn on instrument)..............................8
3.7 Strip holder doesn’t move.....................................................................................................................8
3.8 Buttons don’t work................................................................................................................................. 8
3.9 TROUBLE-01、TROUBLE-04.............................................................................................................8
3.10 TROUBLE-02、TROUBLE-05.......................................................................................................... 8
3.11 TROUBLE-03、TROUBLE-06...........................................................................................................9
3.12 TROUBLE-10....................................................................................................................................... 9
VERSION:: 07/2016
1. Overview
URIT-180 urine analyzer is manufactured under strict supervision. This is a kind of precise analyzer,
improper operation or any damaged component may affect the results or make damage to the
instrument. If an operational or instrument problem occurs, an error number may display on the
Analyzer screen. This section of the manual lists various errors and messages, along with probable
causes and recommended remedies to quickly and easily correct the problem. If the problem persists
after following the recommended remedy, consult Table 2 for common malfunction and remedy and
table 3 for malfunction code and remedy.
Caution:
Be sure to turn off the power supply before opening the instrument cover or replacing parts.
Beware of the high voltage inside.
Consult table 1.
Table 1
No Installation Environment Malfunction
1 Strong light Incorrect result
1
Table 2: Common malfunction and remedy
2
Table 3: Malfunction code and remedy
3
U7
J5
U4
TP2/3/4/5
U2
Figure 2.1
Measure the voltage of TP2 as figure 2.1 shows, (normal range is 12±0.4V), check power adapter,
power input socket(J4), power switch, power input connector(J5), power input wire if the voltage not
within the range, and replace the defective parts.
4
U6 J4
J5
TP6
RP1
J16 J3
Figure 3.1
As shown in figure 3.1or 3.2, Check if the voltage of TP6 is -13±0.5V when the contrast grade is
adjusted to be >>>> on the screen (The voltage may vary as different definition). This voltage can be
adjusted through “LCD contrast adjust potentiometer (RP1)”. Replace the potentiometer or U6 if can
not be adjusted. Check whether LCD 20-core cable is well connected if the voltage is normal. Please
replace LCD module Problem still persists by re-plug LCD 20-core cable.
5
Printer connection cable
Panel socket
Figure 3.2
3.3 The test results cannot be printed out through the internal printer
6
J1 U3 U2 J4 U5 Q1 TP1/2/3/4/5
VP
8V
PS
GND
+5V
U4
Figure 3.3
As shown in figure 3.4, check the voltage of 1pin (ensure J3 on main board and JP1 on limit board are
well connected), the voltage of TP1should more than 4.2V if reagent strip is not placed on the holder,
adjust potentiometer W1 to make TP1 more than 4.2 if not up. Please replace strip sensor or limit
board if sensor ineffective.
Figure 3.4
7
3.5 Time on the screen is abnormal
As shown in figure 3.6, relocate the battery. Check whether the crystal oscillator (Y1 32.768K) and
U10 is normal.
Strip holder moving outward when turning on instrument, observe whether strip holder catch ward off
the optocoupler, check drive mechanism to find the reason. Replace optocoupler if warding off
normally. Consult 3.5.
Figure 3.5
Power supply is normal and motor wire socket J1 is well connected, main cause is that motor is
broken or motor control circuit defective. Motor control circuit U1 broken with high probability, consult
figure 3.6 for chip position.
As figure 3.2, re-plug panel socket if button do not work, replace panel if above is invalid.
3.9 TROUBLE-01、TROUBLE-04
Re-plug J16 and turn off instrument, if not solved please observe brightness value in debug interface:
a. When brightness value is near 0, make sure J16 is well plugged, test whether LED lamp is lighted
with button battery (in power off status), replace LED lamp if not lighted; replace test board if LED
lighted.
b. Brightness value is far more than 3200, replace test board.
3.10 TROUBLE-02、TROUBLE-05
Re-plug J16 and turn off instrument, if not solved please observe brightness value in debug interface:
a. When brightness value is near 0, make sure J16 is well plugged, test whether LED lamp is lighted
with button battery (in power off status), replace LED lamp if not lighted; replace test board if LED
lighted.
8
b. Brightness value is far more than 3200, replace test board.
3.11 TROUBLE-03、TROUBLE-06
Re-plug J16 and turn off instrument, if not solved please observe brightness value in debug interface;
a. When brightness value is near 0, make sure J16 is well plugged, test whether LED lamp is lighted
with button battery (in power off status), replace LED lamp if not lighted; replace test board if LED
lighted.
b. Brightness value is far more than 3200, replace test board.
NOTE:
When alarm TROUBLE-? and can not enter into system after turning on instrument, Press any
key to enter into system.
3.12 TROUBLE-10
As picture 3.6,check U10,Y2 on main board, replace relative chip if problem occurs.
U10
Y2
U1
J1
J16
Figure 3.6
NOTE:
Enter into debug interface: menu-manufacturer information-press F3 for 3 seconds, press F1,
F2, F3, F4 successively.