DustIQ calibration procedure


10 2 10 2
9 3 9 3
Example
8 15 min 4 8 Solar noon 4
7 5 7 1 PM 5
6 6
Finish the whole procedure!

Requirements: Important prerequisites:


It is not possible to restart
the procedure once (a part of) • a uniformly soiled DustIQ with a reported Transmission Loss of at least 5 to 10% For an accurate calibration it is necessary to have: the DustIQ in operational state for at least 15 minutes, clean distilled or demineralised water,
the DustIQ has been cleaned! • clear sky / bright sunlight for at least 15 minutes within two hours of local solar noon soap, sponge and a clean and dry cloth. In sunlight, to see a blue sensor LED blink, view it from the area shown within the red lines.

At the back of the DustIQ press and hold the calibration button for 5 seconds. Release to start calibration. Clean the complete DustIQ, including sensor 2, using soap and distilled or demineralised water.

This side up
Installed vertically
2 5 1 2 Clean DustIQ thoroughly, dry thoroughly

1
Blue LED ring follows blue sensor LEDS
5 sec

Sensor 2 Calibration Sensor 1


Daisy Chain -or- Temperature Sensor Calibration
Installed horizontally
Press the calibration button again. Both clean sensor signals will now be measured.
TOP

Host Solar Panel

1
Sensor 2 Sensor 1

1 2

DustIQ must have a clear


1 2
6
view of the sun and sky

DustIQ must have a clear


1 2 view of the sun and sky 50 sec
2 50 sec
Wait for 50 seconds, until the blue sensor LEDs blink steadily.

Wait for 50 seconds, until the blue sensor LEDs blink steadily. The soiled signals have now been measured.
Within 5 minutes after finishing the calibration procedure press the calibration button again to save the new

SAVE
calibration settings. The new calibration settings will be stored in the memory of the DustIQ and used immediately.
Clean the PV cell and sensor 1, using soap and distilled or demineralised water.
7
3 Clean thoroughly, dry thoroughly

CANCEL
Anywhere in the procedure, when the calibration button is not pressed within 5 minutes, the new calibration
1 2
Do not clean sensor 2 settings will not be saved. The DustIQ will then go back to normal operation using its original calibration settings.

Press the calibration button again. The clean sensor 1 signal will now be measured. Note Cleaning and drying thoroughly is essential for a good calibration result.

4
Note A short burst of rapid LEDs blinks indicate an error and the calibration will be aborted.
DustIQ must have a clear The LEDs will go off and then follow the normal 1 minute pattern.
1 2 view of the sun and sky
For manuals and software please visit
50 sec www.kippzonen.com/dustiq
Wait for 50 seconds, until the blue sensor LEDs blink steadily.