Professional Documents
Culture Documents
Mahuvava C Et Al. (2022)
Mahuvava C Et Al. (2022)
DOI: 10.1002/mp.15487
RESEARCH ARTICLE
1
Department of Physics and Astronomy, Abstract
University of Victoria, Victoria, British
Columbia, Canada
Purpose: To provide percentage depth dose (PDD) data along the central axis
2
for dosimetry calculations in small-animal radiation biology experiments per-
Department of Radiation Oncology,
University of Maryland School of Medicine,
formed in cabinet irradiators. The PDDs are provided as a function of source-
Baltimore MD, USA to-surface distance (SSD), field size, and animal size.
3
Department of Radiation Physics, University Methods: The X-ray tube designs for four biological cabinet irradiators, the
of Texas MD Anderson, Cancer Centre, RS2000, RT250, MultiRad350, and XRAD320, were simulated using the BEAM-
Houston, Texas, USA nrc Monte Carlo code to generate 160, 200, 250, and 320 kVp photon beams,
respectively. The 320 kVp beam was simulated with two filtrations: a soft F1 alu-
Correspondence
Courage Mahuvava, Department of Physics
minium filter and a hard F2 thoraeus filter made of aluminium, tin, and copper.
and Astronomy, University of Victoria, Victoria, Beams were collimated into circular fields with diameters of 0.5–10 cm at SSDs
BC V8P 5C2, Canada. of 10–60 cm. Monte Carlo dose calculations in 1–5-cm diameter homogeneous
Email: mahuvavac@uvic.ca
(soft tissue) small-animal phantoms as well as in heterogeneous phantoms with
Funding information
3-mm diameter cylindrical lung and bone inserts (rib and cortical bone) were
Compute Canada; Natural Sciences and performed using DOSXYZnrc. The calculated depth doses in three test-cases
Engineering Research Council of Canada were estimated by applying SSD, field size, and animal size correction factors
(NSERC); Canada Research Chair Program
to a reference case (40-cm SSD, 1-cm field, and 5-cm animal size), and these
results were compared with the specifically simulated (i.e., expected) doses to
assess the accuracy of this method. Dosimetry for two test-case scenarios of
160 and 250 kVp beams (representative of end-user beam qualities) was also
performed, whereby the simulated PDDs at two different depths were compared
with the results based on the interpolation from reference data.
Results: The depth doses for three test-cases calculated at 200, 320 kVp F1,
and 320 kVp F2 with half value layers (HVLs) ranging from ∼0.6 to 3.6 mm
Cu, agreed well with the expected doses, yielding dose differences of 1.2%,
0.1%, and 1.0%, respectively. The two end-user test-cases for 160 and 250 kVp
beams with respective HVLs of ∼0.8 and 1.8 mm Cu yielded dose differences
of 1.4% and 3.2% between the simulated and the interpolated PDDs. The dose
increase at the bone-tissue proximal interface ranged from 1.2 to 2.5 times the
dose in soft tissue for rib and 1.3 to 3.7 times for cortical bone. The dose drop-off
at 1-cm depth beyond the bone ranged from 1.3% to 6.0% for rib and 3.2% to
11.7% for cortical bone. No drastic dose perturbations occurred in the presence
of lung, with lung-tissue interface dose of >99% of soft tissue dose and <3%
dose increase at 1-cm depth beyond lung.
Conclusions: The developed dose estimation method can be used to translate
the measured dose at a point to dose at any depth in small-animal phantoms,
making it feasible for preclinical calculation of dose distributions in animals irra-
diated with cabinet-style irradiators. The dosimetric impact of bone must be
accurately quantified as dramatic dose perturbations at and beyond the bone
interfaces can occur due to the relative importance of the photoelectric effect
KEYWORDS
depth-dose, Monte Carlo, radiobiology dosimetry, small-animal radiotherapy, TG-319
F I G U R E 1 (a) Measured dose to a point (red circle) in the small animal. (b) The focus of this study, which is to determine the percentage
depth-dose (PDD) correction to convert this dose to the location where dose is of interest (green circle) in a small animal phantom mimicking
the experimental animal using Monte Carlo simulations. This is useful for dose calculation at a prescribed position in the organ of interest (target
or organ at risk)
large set of Monte Carlo dose calculations was carried remotely on the Compute Canada Cedar cluster22 using
out in this work for small-animal radiotherapy beams. Intel processors running at 2.1–2.4 GHz. BEAMnrc sim-
These will provide central-axis PDD data containing ulation times ranged from ∼60 to 100 h and DOSXYZnrc
information on a number of aspects that affect accurate simulation times ranged from ∼5 to 40 h.
dosimetry; namely the beam quality, SSD, field size,
animal size as well as the presence of lung and bone
heterogeneities in the animal m odel. This information 2.2.1 Source description
will facilitate the translation of the experimentally mea-
sured dose at a given point to the dose at the desired Table 1 summarizes the specifications of the irradiator
depth in the organ of interest, and therefore support models/beams, which were used in this work. For all of
improved dose calculations for preclinical studies. the X-ray units simulated, the SSD is varied using an
adjustable steel shelf, while circular beam shaping is
assumed to be achieved with a 2-mm thick lead shield
2.2 BEAMnrc simulations placed just above the animal. Also given are the speci-
fications for two cabinet irradiators (based on the Fax-
The Monte Carlo technique used throughout this work itron MultiRad350 and the Rad Source RS2000) used
involved a two-step process: Firstly, Monte Carlo source for end-user test-case dose calculations.
models based on four small-animal irradiators, namely
the PXi XRAD320 (PXi, North Branford, CT), the Fax-
itron MultiRad350 (Faxitron Bioptics, Tucson, AZ), the 2.2.2 Simulation geometry
Phillips RT250 (Philips Medical System, Cleveland, OH),
as well as the Rad Source RS2000 (Rad Source, Buford, BEAMnrc simulations were split into two stages, which
GA) units were used to simulate the production of 320, together comprised the full geometry of the X-ray
250, 200, and 160 kV X-ray beams in BEAMnrc.17 The source and the physical collimator setup, based on
XRAD320 has two different beam filters (F1 and F2), manufacturer specifications. Figure 2 shows the simu-
which were simulated separately. Table 1 gives the spec- lation geometry used to model the biological irradiators,
ifications of each filter. The irradiator models were com- including the source and phase space file (PSF) scor-
missioned or validated with experimental data from pre- ing planes Figure 2a, as well as the orientation of the
vious studies.18–20 homogenous cylindrical phantom relative to the beam
In the second simulation stage, DOSXYZnrc21 was Figure 2b. Dose calculations were also performed in
used to calculate 3D absorbed dose distributions 2-cm diameter cylindrical phantoms with 3-mm diame-
in cylindrical small-animal phantoms using BEAMnrc ter lung, rib, or cortical bone inserts located at 2.5-mm
phase-space files scored just below the beam collima- depth, as shown in Figure 2c.
tion system as the “source.” The global electron and In the first simulation stage, a mono-energetic elec-
photon energy cut-off values, ECUT and PCUT, were tron beam was simulated in a vacuum, impinging
respectively set to 0.521 and 0.01 MeV in all BEAMnrc upon a tungsten anode to generate Bremsstrahlung X-
and DOSXYZnrc simulations. All simulations were run rays, which were subsequently transported through the
4 DOSIMETRY FOR RADIOBIOLOGY STUDIES
TA B L E 1 Manufacturer specifications for the small-animal cabinet irradiators used in this study. These parameters were used for beam
production simulations in BEAMnrc. The half -value layers (HVL) of the five beams calculated in MATLAB based on the beam energy spectra in
Figure 3, are also shown
F I G U R E 2 Geometry of the X-ray unit and relative orientation of cylindrical small-animal phantoms for simulating beam production and
collimation in BEAMnrc (a) as well as dose calculations in DOSXYZnrc (b). Heterogeneous phantoms (2-cm diameter) with two different bone
materials as well as lung are shown in (c). Irradiator geometry and material compositions were provided by the manufacturers, while bone
compositions were obtained from the International Commission on Radiation Units and Measurements (ICRU)-44 tissue data.23 The
composition of lung (LUNG521ICRU) is available in the ICRU PEGS4 (preprocessor for Electron Gamma Shower v4.0) data in EGSnrc
inherent beryllium filtration and stored in the phase PSF2 (Figure 2a, collected 0.1 cm below the variable
space file PSF1 scored at 5 cm from the focal spot. collimator) for each beam size and SSD; these files were
The energy spectra for each beam in PSF1 are shown later used as sources for the final DOSXYZnrc dose cal-
in Figure 3. A total of 2 × 109 primary particles were culations.
simulated in the PSF1 X-ray phase space simulations. The total number of particles scored in PSF2 ranged
Directional bremsstrahlung splitting (without cross sec- from 7.5 × 106 , for the smallest (0.5 cm diameter) field
tion enhancement)21 with a splitting factor of 2000 and size in the heavily filtered 320 kVp F2 beam, up to 109
a splitting radius of 3.0 cm, defined at 5.0 cm from the for the largest (10-cm diameter) field in the 320 kVp
source, was employed to increase simulation efficiency. F1 beams. The 521ICRU PEGS4 material cross section
In the second BEAMnrc simulation stage, particle data were used. All other user-adjustable transport con-
transport through air and the variable aperture 2-mm trol parameters such as the maximum fractional energy
thick lead collimators was simulated. The output of this loss per step, Russian roulette, skin depth etc., were set
simulation was a set of phase-space files scored at to their default values or switched off.
DOSIMETRY FOR RADIOBIOLOGY STUDIES 5
TA B L E 3 Summary of test-case parameters for the reference and nonreference cabinet irradiators that were used in this study. The
nonreference cases represent end-user irradiators
TA B L E 4 Composition of lung, rib, and cortical bone used to generate PEGS4 cross-sectional data for Monte Carlo simulations in animal
models
The specific cross‑sectional data for the two bone 𝛿Dref , and DSSD be the dose for the test case SSD (first
materials were calculated in MATLAB based on ICRU- numerator in Equation (1)) with associated uncertainty
44 data23 while that of lung (LUNG521ICRU) is avail- 𝛿DSSD , etc.
able in the default 521ICRU PEGS4 data in EGSnrc.The √
√ (
material compositions of lung, rib, and cortical bone are √ 𝛿D )2 ( 𝛿D )2 (
𝛿DFS
)2 (
𝛿DAS
)2
√ ref SSD
𝛿Dcalc = |Dcalc | 2 + + +
summarized in Table 4. Dref DSSD DFS DAS
Dose to the medium (bone) was calculated (as
(2)
opposed to dose to water with bone density) as
per international recommendations for clinical trial
consistency.25,26 The propagated uncertainty for the calculated dose
(Dcalc ) is the quadratic sum in Equation (2). The
combined uncertainty of the dose difference (𝛿Ddiff )
between the calculated and simulated dose is found by
2.3.3 Uncertainty propagation adding their respective uncertainties (𝛿Dcalc and 𝛿Dsim )
in quadrature as shown in Equation (3).
Let the symbol 𝛿 represent the uncertainty for each of
√
the dose quantities in Equation (1). For simplicity, also let 2 2
Dref be the reference dose with associated uncertainty 𝛿Ddiff = (𝛿Dcalc ) + (𝛿Dsim ) (3)
DOSIMETRY FOR RADIOBIOLOGY STUDIES 7
F I G U R E 4 Reference-case percentage depth-doses (PDDs) as a function of animal size for the (a) PXi 320 kVp F1, (b) PXi 320 kVp F2,
and (c) RT250 200 kVp beams calculated for 40-cm source-to-surface distance (SSD) and 1-cm field size
Equations (2) and (3) assume that the original uncer- 3.1 Small-animal phantom PDDs for
tainties associated with the simulated doses in Equation reference cases
(1) are uncorrelated and random, which is a reasonable
assumption in this Monte Carlo study. Some general trends are evident from the plots, such
as the relatively high surface dose (>50% of the maxi-
mum dose) as compared to MV beams (≤15% PDD)27
2.4 Dosimetry formalism for end-users generally used in radiotherapy treatments. The simu-
(HVL interpolation) lated surface dose is, however, lower as compared to
typical values for kV beams (usually ∼90%–100%),27
Radiation units come with manufacturer-provided HVLs, and this may be due to partial volume averaging of the
and these values should also be measured by the user, calculation voxel size used (0.5 × 0.5 × 2 mm3 ). The
as will be prescribed in the TG-319 report. So far, the dose fall-off per centimetre can also be observed in
methodology for dose calculation and test case valida- Figure 4, with very small differences in doses at depth
tion has been provided for the reference beam quali- for different animal sizes, except near the beam exit
ties investigated, that is, the 320 kVp F1, 200 kVp, and regions of the phantom due to missing backscatter. The
320 kVp F2 beams with HVLs of 0.59,1.00,and 3.57 mm PDD decreases with increasing depth due to the inverse
Cu, respectively. square law and beam attenuation. In Figure 5, the dose
To perform dosimetry for nonreference irradia- increases in going from a 0.5-cm to 10-cm diameter field
tors/beam qualities, let the right-hand side of Equation size due to increased scatter contribution from the col-
(1), excluding the reference dose term, be the full CF to limator and the phantom. The effect is relatively pro-
rescale the dose from the measured to the prescribed nounced - dose differences up to 15% are apparent as a
position for any given cabinet irradiator. The doses at function of field size. The decrease in dose with increas-
3-cm and 2-cm depth for the 160 and 250 kVp beams ing SSD in Figure 6 is because of the inverse square
with HVLs 0.78 and 1.80 mm Cu, respectively, were sim- law, which dictates a reduction in beam intensity with
ulated in a 5-cm diameter phantom and compared with increasing distance from the source. This effect is more
the results based on the interpolation from reference pronounced for smaller SSDs.
data and the calculation formalism of Section 2.3.2. This Tables S1 and S2 (provided in the supplementary
was accomplished by calculating the full CFs for each of data) give PDD values at depths of 0.5–5.0 cm (in steps
the reference beam qualities investigated as a function of 0.5 cm) in a small-animal phantom as a function
of their corresponding HVLs and then interpolating of beam size and SSD, respectively, for beam qualities
the CFs for the end-user HVLs under investigation ranging from 0.6 to 3.6 mm Cu.
to rescale the reference dose to the treatment field
dose.
3.2 Dose estimation and validation for
test cases
3 RESULTS
Figure 7 shows test-case PDDs for the reference beams
The following figures (Figures 4–6) illustrate the effect (the PXi and RT250 beams) and nonreference end-user
of varying animal size, field size, or SSD on the PDD beams (the MultiRad350 and RS2000 beams) simu-
for cylindrical small-animal phantoms under reference lated for different dosimetric parameters (in Table 3)
conditions. in small-animal phantoms of different sizes. Dose
8 DOSIMETRY FOR RADIOBIOLOGY STUDIES
F I G U R E 5 Reference-case percentage depth-doses (PDDs) as a function of field size for the (a) PXi 320 kVp F1, (b) PXi 320 kVp F2, and
(c) RT250 200 kVp beams for 40-cm source-to-surface distance (SSD) and 5-cm animal size
F I G U R E 6 Reference-case percentage depth-doses (PDDs) as a function of source-to-surface distance (SSD) for the (a) PXi 320 kVp F1,
(b) PXi 320 kVp F2, and (c) RT250 200 kVp beams calculated for 1-cm field size and 5-cm animal size
TA B L E 5 Test-case dose calculation and comparison with the simulated dose for the (a) 320 kVp F1 beam; (b) 320 kVp F2 beam, and (c)
200 kVp beam. Uncertainties (δ) for the doses and for correction factors (CFs) are shown. Dcalc is the calculated dose, Dsim is the simulated
dose, and Ddiff is the dose difference (Dcalc − Dsim ). The CF is the corresponding ratio of the test and reference doses in Equation (1)
(a) PXi XRAD 320 kVp F1 (b) PXi XRAD 320 kVp F2 (c) RT250 200 kVp
Value PDD ± Value PDD ± PDD ±
(cm) δPDD (%) CF ± 𝜹CF (cm) δPDD (%) CF ± 𝜹CF Value (cm) δPDD (%) CF ± 𝜹CF
Depth 4 – – 2 – – 1 – –
Animal size 5 ǂ 38.9 ± 0.6 1.00 ± 0.02 4 ǂ 73.6 ± 0.6 1.00 ± 0.01 3 ǂ 84.7 ± 0.8 1.00 ± 0.01
SSD 60 41.3 ± 0.9 1.06 ± 0.03 30 70.5 ± 0.5 0.96 ± 0.01 10 74.1 ± 0.2 0.87 ± 0.01
Field size 3 44.4 ± 0.3 1.14 ± 0.02 10 82.8 ± 0.4 1.13 ± 0.01 0.5 82.2 ± 0.8 0.97 ± 0.01
Dcalc (%) 47.0 ± 1.7 79.8 ± 1.3 71.5 ± 1.4
Dsim (%) 47.1 ± 0.5 78.8 ± 0.3 70.3 ± 0.2
Ddiff ( %) –0.1 ± 1.8 1.0 ± 1.3 1.2 ± 1.4
ǂ
indicates the reference case for each beam.
TA B L E 6 Maximum dose (Dmax ) as well as dose at 1-cm depth (D1cm ) for reference beams with the associated dose difference, Ddiff (in
brackets) relative to dose in a homogenous (soft tissue) phantom
320 F1 100.0 80.9 367.3 69.2 (–11.7) 253.4 74.9 (–6.0) 98.6 83.5 +2.6
320 F2 100.0 87.0 133.0 83.8 (–3.2) 118.0 85.7 (–1.3) 99.7 89.2 +2.2
200 100.0 84.4 291.4 78.1 (–6.3) 208.2 81.5 (–2.9) 98.9 86.5 +2.1
*Ddiff = dose @ 1 cm depth with lung/bone −− dose @ 1 cm depth in soft tissue.
TA B L E 7 Dose calculation for the MultiRad350 beam (at 2-cm depth) and the RS2000 beam (at 3-cm depth) with simulation parameters
summarized in Table 3
Correction Correction
Irradiator HVL factor factor Dcalc
model (mm Cu) (2-cm depth) (3-cm depth) Dsim (%) (%) Ddiff (%)
challenge due to size constraints), these will not be strate that the results support the use of compiled PDD
representative of their experimental conditions. Some data sets, such as those provided in this work, to trans-
investigators were able to characterize their biological late from a reference field depth-dose to that expected
irradiators using Monte Carlo techniques and calculate in the treatment field.
animal-specific dose distributions.30,31 However, these Dose at the bone-tissue proximal interface ranged
techniques are difficult to implement by the average from 1.2 to 2.5 times higher than the dose in soft tissue
end-user, who often has limited access to resources.14,15 for rib and 1.3 to 3.7 times higher for cortical bone.These
With these PDDs, an investigator can now relate dose results are in agreement with a similar published study,33
from a point – which they measure experimentally – to which reported a dramatic bone dose-enhancement in a
dose at another depth for small animal experiments per- mouse model—up to three times higher when simulating
formed on kV X-ray cabinet irradiators. On their own, irradiation with 225 kV X-rays. In the present study, dose
the PDDs do not provide depth-dose information for a reductions at 1-cm depth beyond the bone interface, rel-
real animal model but provide the basis for depth dose ative to the case without bone, ranged from 1.3% to
estimation based on derived corrections to measured 6.0% for rib and 3.2% to 11.7% for cortical bone. These
doses, as long as heterogeneities are avoided. Dose cal- dose perturbations occur due to electron backscatter on
culations in three different test cases, using PDD CFs the bone surface and absorption inside the bone, owing
derived from the data, and comparisons with their cor- to the importance of photoelectric interactions in high
responding simulated doses were performed to evalu- atomic number materials for kV beams. On the other
ate feasibility for preclinical implementation of the calcu- hand, lung did not exhibit such drastic dose perturba-
lation methodology. The dosimetric impact of lung and tions, with lung-tissue interface dose >99% of the dose
bone was also evaluated. These data will facilitate the in a homogeneous phantom and <3% dose increase at
calculation of dose at the required depth in small-animal 1-cm depth beyond the lung. This is due to the lung’s
models, thereby increasing the accuracy of small animal similar effective atomic number with soft tissue.However,
dosimetry. the lower density lung (𝜌 = 0.26 g/cc) receives a slightly
The depth doses for the three test-cases at 320 F1, higher dose, due to the increased scattering of electrons
320 F2, and 200 kVp agreed well with expected doses, from the higher density soft tissue (𝜌 = 1.0 g/cc).
yielding dose difference magnitudes of 0.1%, 1.0%, The proportion of photons interacting via the photo-
and 1.2%, respectively. Taking into account the prop- electric effect depends on the energy spectrum (and
agated uncertainties, the maximum dose differences consequently the HVL) of the X-ray beam, with more
increase to 2.3%, 2.6%, and 1.7%, respectively, which interactions occurring in beams, which contain a larger
is small compared to the uncertainty generally tolerated number of low-energy photons.34 The calculated HVLs
in radiotherapy.32 The dose estimation technique used of the 320 kVp F1, 200 kVp, and 320 kVp F2 beams
in this work assumes that the CFs for SSD, field size, based on the X-ray spectra in Figure 3 were found to be
and animal size are uncorrelated; therefore, the over- 0.6, 1.0, and 3.6-mm Cu, respectively. The 200 kVp beam
all PDD CF can be obtained by simply multiplying CFs HVL agreed with measured data.20 It must be noted that
for these three terms. However, this still propagates the since we did not have access to the geometry of the mir-
uncertainty in the calculated dose.It is interesting to note ror and ionization chamber, these components were not
that the uncertainties in the dose differences (Ddiff ) of included in the Monte Carlo model of the PXi XRAD unit,
Table 5a-c can be of the same order of magnitude or which might account for the softer beam compared to
larger than the dose difference itself,especially when the what was measured experimentally (with HVL values of
doses in comparison are very close to each other. Never- 0.8–1.12 mm Cu for F1 and 3.7–4.0 mm Cu for F2).30
theless, the small percentage dose differences demon- In Figure 8, the dose perturbation in the presence of
DOSIMETRY FOR RADIOBIOLOGY STUDIES 11
F I G U R E 8 Effect of cortical bone, rib, and lung on small-animal percentage depth-doses (PDDs) for the (a) PXi 320 kVp F1, (b) PXi 320 kVp
F2, and (c) RT250 200 kVp beams calculated for a 40-cm source-to-surface distance (SSD) and 1-cm field size in a 2-cm diameter cylindrical
phantom. The graphs were plotted on the same y-axis range to easily compare changes in dose perturbations with changing beam quality
bone for the 320 kVp F1 beam is highest, owed to hav- HVL interpolation methodology for cabinet irradiators
ing the lowest HVL of the three beams, followed by the with different HVLs, two test-case scenarios, one for
200 kVp and then the 320 kVp F2 beam. The 320 kVp a 160 kVp beam and another for a 250 kVp beam
F2 beam exhibits a smaller dose perturbation, due to with respective HVLs of 0.78 and 1.80 mm Cu, were
the increased X-ray filtration in the beam, which attenu- simulated and the PDDs at two different depths com-
ates most of the lower-energy photons. The high sensi- pared with the interpolated, formalism-derived results
tivities of kV beams to tissue heterogeneities demon- based on the reference data. The results demonstrate a
strate the importance of including these structures in reasonable agreement, at least for our test-case cabinet
the dose estimation framework. In small animals, this irradiators, with respective dose differences of 1.4%
sensitivity can be reflected in treatment planning uncer- and 3.2% between the simulated and the interpolated
tainties within the organs of interest (targets or organs doses.
at risk). An expanded body of work could include the cre-
These estimated doses were only validated in the ation of a web-based program to calculate dose at
absence of bone. In cases where dose calculations any depth in the animal for an arbitrary beam. Further-
are desired near heterogeneities (especially bone inter- more, the current work concerns itself only with the
faces), additional attention (e.g., Monte Carlo) is needed distribution of physical absorbed dose – the effects
for accurate dose determination. CFs are unlikely to of beam quality, attenuation, and heterogeneities on
be a feasible approach – first, due to difficulties in radiobiological effectiveness (RBE) are not investigated.
quantifying the effective thickness of bone (particularly Previous research in microdosimetry has shown there
the dense cortical bone) in a given animal in the path is a high dependence on the absorbed dose delivered
of the beam without CT image guidance. Second, due in bone marrow immediately adjoining cortical bone.
to the very high dependence of the attenuation of kV It is speculated that this may be driving much of the
beams on the soft X-ray component of the spectrum, RBE differences in bone marrow effects with beam
which leads even modest changes in beam quality to quality.38
have a disproportionate impact on the attenuation in
cortical bone.35,36 Thankfully, in irradiations performed
using cabinet irradiators, targets tend to be relatively 5 CONCLUSIONS
large (≥1 cm), and bones seldom obscure the entire
treatment area outside the cranium. In the absence of The developed dose estimation method can be used
3D dose computation techniques, investigators typically to translate measured doses in the reference point to
calculate dose to water or lung, ignoring the confounding any depth along the central beam axis in a small-animal
effects of bone. A recently published paper by Dewert experiments performed on kV X-ray cabinet irradiators.
and Kunugi37 addresses the difficulty of performing The dosimetric impact of bone must be accurately quan-
reference dosimetry on orthovoltage machines. tified and included in the dose calculation framework as
The presented work gives data for three beams dramatic dose perturbations at and beyond the bone
with HVL values between 0.6 and 3.6 mm Cu. For interfaces can occur due to the relative importance of
a system with different a beam energy or HVL, it is the photoelectric effect at kV beam energies. These
likely possible to interpolate the data presented in results will help improve dosimetric accuracy in preclin-
Figures 4–6 to beams with other HVL values. To test the ical experiments.
12 DOSIMETRY FOR RADIOBIOLOGY STUDIES
38. Belley MD, Ashcraft KA, Lee CT, et al. Microdosimetric and bio-
logical effects of photon irradiation at different energies in bone How to cite this article: Mahuvava C, Esplen
marrow. Radiat Res. 2015;184(4):378-391.
NM, Poirier Y, Kry SF, Bazalova-Carter M. Dose
calculations for preclinical radiobiology
experiments conducted with single-field cabinet
S U P P O R T I N G I N F O R M AT I O N irradiators. Med Phys. 2022;1-13.
Additional supporting information may be found in the https://doi.org/10.1002/mp.15487
online version of the article at the publisher’s website.