Professional Documents
Culture Documents
Analytical Methods
Analytical Methods
Analytical Methods
This Volume is the property of Stamicarbon bv and, except where otherwise agreed in writing, it is put at the receiver's
A4-00000-A
disposal without consideration other then the receiver's agreement that it shall not be reproduced, copied, lent or disposed of
directly or indirectly nor used for any purpose other than that for which it is specifically furnished, and that it is to be
returned to Stamicarbon bv on demand together with any copies made.
size
revision -
A4 00000
STAMICARBON CONFIDENTIAL INFORMATION Sheet 1 of 165
Stamicarbon bv, Mauritslaan 49, Urmond P.O. Box 53, 6160 AB Geleen, The Netherlands
project ZURE
A4-00000-A
project ZURE
A4-00000-A
Authorisation :
(signature)
1. Scope
2. Safety
Chemicals to be used:
Acetic acid
Acetonitril
Acetylacetone
Ammonium acetate
Formaldehyde
Phosphoric acid
Sodium hydroxide
2-Propanol
Work safely, especially with hazardous chemicals, as in this method
formaldehyde (suspected carcinogen) with a MAC value of 1 ppm (1.5 mg/m3)
MAC TGG-15 min.of 2 ppm (3 mg/m3). Never use hydrochloric acid in the
neighbourhood of formaldehyde, because formaldehyde reacts with
hydrochloric acid giving off very toxic “bis(chloromethyl)ether”, which can
cause lung cancer.
Use the guidelines of e.g. Safety Data Sheets (SDS) or a local equivalent.
A HPLC apparatus operates under high pressure (100 - 400 bar). In spite of the
use of small volumes, always wear safety glasses when working with HPLC.
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
3. Principle
4. Apparatus
5. Chemicals
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
6. Procedure
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
- Clean the sparge filter (see 8.3) of the buffer solution A (6.1) once a week.
- Before starting the HPLC (4.1), the eluent solutions must be degassed for 30
minutes with He gas: flow approx. 100 ml/min. This is especially important for
acetonitril.
- Purge if necessary solutions (A), (B), (C) and the Hantz reagent. Perform a purge
when changing the eluent solutions by means of opening the purge valve on the
HPLC pump. Close this purge valve firmly. Purge with 5 ml/min.
- Switch on the autosampler (4.1.3) and adjust temperature to 20°C
- Switch on the fluorescence detector (4.1.8).
- Switch on the Waters column heater module (4.1.6). Press ‘menu/enter’ button
and select with the arrow key up ‘ON’. Press ‘menu’ again. The oven heats up till
30°C.
- Switch on the LKB Bromma oven (4.1.7). Gradually increase the temperature to
90°C.
- The reversed phase HPLC column (4.1.5) needs to be activated when storing
overnight at 100% buffer solution or storing for a prolonged time at 50%
acetonitril/50% buffer. This is done by running a gradient till 80 % acetonitril and
returning till 100% buffer solution. Activate the column by running a wash gradient
method 1 or method 2, depending on the stop conditions of the former analysis.
Press METHOD’, ‘1’ and ‘ENTER’ (See attachment 11.1 for further details).
- Increase flow from 0.2 ml/min till 1.0 ml/min in steps of 0.2 ml/min by pressing
‘FLOW’ followed by 0.2 till 1.0 and ‘ENTER’.
- After 30 minutes this wash gradient can be stopped by pressing ‘STOP’ and
‘ENTER’ key.
- Load method 4 by pressing ‘METHOD’ , ‘4’ and ‘ENTER’. The system is now
running isocratic at 1.0 ml/min with 100 % buffer solution A.
- Start the pump of the Hantz reagent with a flow of 1.0 ml/min.
- Allow the system to stabilize for at least 30 minutes.
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
7
7. Calculation
Cm
fm = -----------
Am
7.2 Calculate the formaldehyde concentration of the sample in mg/kg resin, using the
formula:
Cs = f * As * b / a
8. Remarks
8.1 Prepare a fresh buffer solution(6.1) as often as possible and discard the old
buffer solution because of the possibility of bacterial growth in this buffer
solution.The buffer solution remains stable for approximately 1 week.
8.2 Mix the Hantz reagent carefully for at least 30 minutes. This is necessary
because the miscibility especially of the acetylaceton is difficult. The Hantz
reagent is stable for maximum 1 week.
8.3 Clean sparge filters every week. The filter is situated in bottle A at the end of the
tubing from the bottle A to the gradient mixer. There must be a free flow from
buffer solution to the gradient mixer when disconnecting line A from the gradient
mixer. Remove the filter from the tubing and place with 2-propanol (5.11) for 5
minutes into a ultrasonic bath (4.3). Rinse with water (5.5), also for 5 minutes in
an ultrasonic bath (4.3).
8.4 Replace the seals once every year.
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
9. References
11. Attachments
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
Three methods are used to run the methylol analysis. Method 1 and 2 are used to
activate the column and method 3 is used to run the real methylol analysis.
Method 1 is used when starting at 100% buffer and ending at 100% buffer, while
method 2 is used when starting at 50% buffer and ending at 100% buffer.
So depending on the stop conditions of the former analysis (the last eluent
composition can be seen by pressing ‘%’ key on the HPLC pump), method 1 or 2
should be used. The gradients used in method 1, 2 and 3 are summarised in table 1.
Increase the flow over the column gradually, start with a flow of 0.2 ml/min and
increase with steps of 0.2 ml/min till 1.0 ml/min. Avoid sudden pressure pulses and
abrupt changes in eluent compositions.
Note: Method 2 stops with another eluent composition than it starts. Be sure to
stop the run before the end of this run. The initial ‘%’ must be manually reset
to 0% (B).
* Rest is (A)
** This method stops with another eluent composition than it starts. Do not stop the
gradient run before initial ‘%’ is reset to 0% (B).
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
Authorisation :
(signature)
2. Safety
- Use the required personal protective devices during sampling (extra long rubber
gloves and face shield).
- Sampling should be done by two persons.
- Take into account the temperature and pressure in the line.
- Always follow local safety regulations.
Work safely, especially with hazardous chemicals. Use the guidelines of e.g. Safety
Data Sheets (SDS) or a local equivalent.
3. Principle
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
4. Apparatus
4.1 Natural rubber bladder ( diameter about 35 cm), provided with a PTFE adapter,
with PTFE -stopcock for example BOLA-stopcock-hose connector, ( Cat.No. E
650-12, connector outer ø 11.0 mm, stopcock bore 6.0 mm) and thick-walled
transparent silicon tubing ( length approx. 15 cm), usable specially at high
temperatures ( above 75 °C ), see figure 1
4.2 Natural rubber bladder ( diameter about 35 cm), provided with a PVC adapter,
with PVC-stopcock for example COMER-stopcock, (D.20, PN16, DN15-½"PVC,
stopcock bore 14 mm) and thick-walled transparent silicon tubing ( length
approx. 15 cm), usable until a maximum temperature of 75 °C, see figure 1
4.3 Bladders e.g. Glassworks: A.J.Janssen, Geleen, The Netherlands
4.4 Balance e.g. Mettler Toledo PG-2002.
4.5 Faceshield, e.g. type Perfo B from MS Safety Shop Schippers Bladel BV
4.6 Rubber gloves extra long, e.g. Emperor 108 from MS Safety Shop Schippers
Bladel BV
4.7 Hose clambs, 10-14 mm
5. Chemicals
5.1 Demi-water
5.2 Hydrochloric acid (2 mol/L), e.g. titrisol Merck
6. Procedure
6.1.1 Method a:
- Inspect the bladder by inflating with N2 and immersing in a bin filled with water.
- Wash the inside of the bladder with water (5.1) to remove any impurities.
- Remove the water thoroughly as possible and weigh the dry empty bladder
(m 0).
- After that, transfer 750 ml of water (5.1) into the bladder and shake for 1 min.
- Fill a 250 ml flask with water from the bladder and close the flask (= solution to
be used in the determination of the blank).
- Remove the air from the bladder and weigh (m 1).
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
6.1.2 Method b:
- Inspect the bladder by inflating with N2 and immersing in a bin filled with water.
- Wash the inside of the bladder with water (5.1) to remove any impurities.
- Remove the water thoroughly as possible and weigh the dry empty bladder
(m 0).
- Transfer approx. 500 ml of water (5.1) into the bladder, remove the air from
the bladder and weigh (m 1).
6.2 Sampling:
- Use the required protective devices: rubber gloves (4.6) and face shield (4.5).
- If the sampling valve is clogged, heat it gently, taking care not to overheat one
point of the valve.
- After that, carefully open the sampling valve and purge the sampling line for a
few moments to the chemical sewer or in the open air.
- Close the valve and, if necessary, clean the sampling point properly.
- Connect the bladder to the sampling point and open the bladder cock.
- Proceed as follow:
- One person holds the connection between the bladder and the sampling point
with one hand (with glove (4.6)) and handles the sampling valve with the other
hand. The other person shakes the bladder in the mean time.
- Open carefully the sampling valve and introduce a smooth constant stream of
the product into the bladder (check flow visual through the silicon tube). Stop
sampling as soon as the required amount of the product has been collected in
the bladder.
- First close the sampling valve and then the bladder cock and immediately
remove the bladder from the sampling point.
- Allow the bladder to cool and weigh (m 2).
6.3.1 Method a:
- Transfer the content of the bladder into a 2-litre measuring flask.
- Rinse the bladder several times with water (5.1).
- Discharge the water into the measuring flask.
- Fill up to 2000.0 ml with water (5.1) and mix (= solution A).
- For calculation of the dilution factor see 7.
6.3.2 Method b:
- Transfer the content of the bladder into a flask and mix.
- Depending on the amount of the sample and the probably composition, take
aliquot portions of the resulting solution for use in the determinations specified.
- For calculation of the dilution factor see 7.
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
7. Calculation
For method a:
2000
f= ---------------
( m2- m 1)
for method b:
( m2- m 0 )
f= ---------------
( m2- m 1)
Where:
f : dilution factor
m2 : weight bladder + water + sample in g
m1 : weight bladder + water in g
m0 : weight bladder in g
8. Reference
G:\methods\doc\400v2.doc
DSM method M 0003-E
CHEMLAB 1921 MN
10. Attachments
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
Attachment 1:
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
Authorisation :
(signature)
Sampling:
a. the urea solution of the outlet of the reactor
b. the urea solution of the outlet of the stripper before the expansion valve
c. the urea melt from the evaporation unit
d. the urea solution of the outlet recirculation separator
e. the carbamate solution from the low pressure carbamate condensor
f. the liquid to the desorber
g. the reflux condensor level tank
h. the urea solution from the buffertank
i. The liquid from the desorber to the sewer
2. Safety
- Use the required personal protective devices during sampling (extra long rubber
gloves and face shield).
- Sampling should be done by two persons.
- Take into account the temperature and pressure in the line.
- Always follow local safety regulations.
Work safely, especially with hazardous chemicals. Use the guidelines of e.g. Safety Data
Sheets (SDS) or a local equivalent.
3. Chemicals
3.1 Demi-water
3.2 Hydrochloric acid (2mol/L) e.g. titrisol Merck
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
4. Procedure
4.1.1 Method a:
- Inspect the bladder by inflating with N2 and immersing in a bin filled with water.
- Wash the inside of the bladder with water (3.1) to remove any impurities.
- Remove the water thoroughly as possible and weigh the dry empty bladder (m 0 ).
- Transfer approx. 500 ml of water (3.1) into the bladder, remove the air from the
bladder and weigh (m 1 ).
4.2. Sampling:
Follow procedure a (see sampling method SCM-400 under 6.2) for the products as
described above under 1.a. - 1.e. and 1.g.
Follow procedure b (see sampling method SCM-400 under 6.2) for the products as
described above under 1.f., 1.h. and 1.i.
In this way usually collect about 400 g of the products in the bladder.
5. Reference
G:\methods\doc\401v1.doc
G:\methods\doc\400v1.doc
DSM method M 0003 31-E
Upgrade.
Point i of the sampling in paragraph 1, Scope and Field of Application, was missed in
paragraph 4.1., preparation of the bladder and in paragraph 4.2., sampling.
7. Attachments
No attachments.
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
Authorisation :
(signature)
1. Scope
2. Safety
Chemicals to be used:
Copper(II) sulphate pentahydrate
Potassium sodium tartrate tetrahydrate
Potassium hydroxide
Biuret
Methanol
Ammonium hydroxide
Acetone.
Work safely, especially with hazardous chemicals. Use the guidelines of e.g. Safety Data
Sheets (SDS) or a local equivalent.
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
3. Principle
In an alkaline medium biuret reacts with copper(II) sulphate to form a violet complex
compound. Excess of copper is kept in solution by means of potassium sodium tartrate.
The extinction of the coloured solution is measured at a wavelength of 550 nm.
4. Apparatus
4.1 Ultra sonic bath e.g. Cole Parmer type 8893, set at 20°C
4.2 Drying oven, temperature set at 105°C e.g. Hereaus type VT 6130M
4.3 Spectrophotometer or filterphotometer e.g. Hitachi. In case of a filterphotometer it
should have a filter with a maximum transmission at app. 550 nm.
4.4 Steam bath e.g. Type LT 6 from Labo Tech
4.5 Magnetic stirrer e.g. IKA combi mag. RCT
4.6 1G3 filter crucible
4.7 Diaphragm vacuum pump, e.g. Vacuubrand type MZ-2C
4.8 Rotavapor, e.g. Büchi type R-124
4.9 Pressure gauge, e.g. Vacuubrand type DVR-1
4.10 Round bottom boiling glask, e.g. Duran, 100 ml with joint 24/29
5. Chemicals
5.1 Demi water, free from carbon dioxide: boil demi water under reflux for 15 minutes,
cool down to 20°C. Cover this water with a nitrogen gas blanket.
5.2 Copper (II) sulphate pentahydrate (CuSO4.5H2O) cryst. extra pure e.g. Merck
5.3 Copper (II) sulphate solution: dissolve in demi water (5.1) in a 1L measuring flask
6.00 g of copper(II)sulphate pentahydrate (5.2). Fill up to 1000.0 mL with demi
water (5.1) and mix (see 9.3).
5.4 Potassium hydroxide (KOH) pellets 85% p.a. e.g. Merck
5.5 Potassium sodium tartrate tetrahydrate (KNaC4H4O6.4H2O) cryst. extra pure e.g.
Merck.
5.6 Potassium sodium tartrate solution: dissolve in demi water (5.1) in a 1L measuring
flask 32.0 g of potassium hydroxide (5.4) and 20.0 g of potassium sodium tartrate
(5.5). Fill up to 1000.0 mL with demi water (5.1) and mix. Allow this solution to stand
for 2 days before using it or place the solution for 10 minutes in an ultra sonic bath
(4.1) at 20°C.
5.7 Biuret (C2H5N3O2 ) ± 99% e.g. Fluka, dried for 3 hours at 105°C in a drying oven
(4.2) (see 9.5).
5.8 Methanol 99% e.g. Merck.
5.9 Ammonium hydroxide, 28% NH3 in water e.g. Sigma
5.10 Ammonium hydroxide 10 wt%, dilute 35.71 g of ammonium hydroxide (5.9) in
64.29 g demi water in a 200 ml measuring flask and mix.
5.11 Acetone, 99% e.g. Merck
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
6. Calibration
– Dissolve in demi water in a 1L measuring flask, 4.000 g of biuret (5.7), fill up with
demi water and mix (= solution I).
– Dissolve in demi water in a 500 ml measuring flask, 2.000 g of biuret (5.7), fill up with
demi water and mix (= solution II).
– Pipette 100.00 ml of solution I into a 500 ml measuring flask, fill up with demi water
and mix (= solution G). (1 ml º 0.8 mg biuret)
– Pipette 50.00 ml of solution II into a 250 ml measuring flask, fill up with demi water
and mix (= solution K). (1 ml º 0.8 mg biuret)
– Take two series of eight 50 ml measuring flasks.
– Pipette into the flasks of each series 0, 5.00, 11.00, 17.00 ml of solution G and 2.00,
8.00, 14.00, 20.00 ml of solution K.
– Introduce into the eight flasks of the first series:
10.00 ml potassium sodium tartrate solution (5.6)
10.00 ml copper (II) sulphate solution (5.3)
Fill up with demi water and mix (= standard solutions).
– Introduce into the eight flasks of the second series:
10.00 ml potassium sodium tartrate solution (5.6)
Fill up with demi water and mix (= reference solutions).
– Perform the following measurements at a wavelength of 550 nm and an optical path
length of 4 or 5 cm using a spectrophotometer (4.3).
– Adjust the photometer to 100% transmission relative to each reference solution, and
measure the extinction (E0= extinction of standard solution containing no biuret, Ei=
extinction of standard solution containing biuret) of the corresponding standard
solution.
7. Sample preparation
– Weigh a g (accurate to 1 mg) of sample into a 200 ml measuring flask (see 9.1 and
9.7). Usually ± 10 g of sample is used so that 1 – 15 mg of biuret will be present in
the extinction measurement.
– Fill up with demi water and mix. If necessary, filter through a folded filter (= solution
A).
– Pipette 20.00 ml of solution A into a 50 ml measuring flask and add:
10.00 ml potassium sodium tartrate solution (5.6)
10.00 ml copper (II) sulphate solution (5.3)
Fill up with demi water and mix (= solution B).
– Pipette 20.00 ml of solution A into a second 50 ml measuring flask and add:
10.00 ml potassium sodium tartrate solution (5.6)
Fill up with demi water and mix (= solution C).
– Add into a third 50 ml measuring flask:
10.0 ml potassium sodium tartrate solution (5.6)
10.00 ml copper (II) sulphate solution (5.3)
Fill up with demi water and mix (= solution D) (see 9.4).
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
8 Calculation
Ci
fi= -----------
Ei – E0
Calculate the average calibration factor, or use linear regression to calculate the
calibration curve (see attachment 1).
b * 200 * 100 b
wt% biuret= ---------------------- = ----
1000 * a * 20 a
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
9. Remarks
9.1 If the solution to be measured with the spectrophotometer contains more than 1.5
mg of NH3 proceed as follows:
– Transfer a g of sample into a 250 ml beaker.
– Add 5 ml of demi water.
– Add 50 ml of methanol (5.8).
– Add a few boiling chips.
– Heat to boiling temperature on a steam bath (4.4) (fume-cupboard) and evaporate
until 10 ml of liquid is left.
– Cool down to ± 25 °C.
– Transfer the content of the beaker to a 200 ml measuring flask. Fill up with demi
water and mix. If necessary filter through a folded filter (= solution A).
– Proceed as described under chapter 7.
9.2 Water- insoluble, non-filterable substances produce turbidity. This affects the
extinction in such way that too high biuret concentrations may be found. By
measuring with respect to a reference solution this hazard is eliminated.
9.3 Copper(II)sulphate is not stable in solutions containing carbon dioxide. After
a while a precipitate of basic copper carbonate will form.
9.4 To check the condition of the copper(II)sulphate solution, run a blank with this
solution (=D) each time an analysis is performed. The extinction ED should remain
constant. If it does not, adjust the copper(II)sulphate solution to the concentration
prescribed.
9.5 If biuret is contaminated with cyanuric acid and/or urea, it should be purified as
follows:
– Transfer 5 g of biuret into a 100 ml beaker.
– Add 15 ml of ammonium hydroxide 10 wt% (5.10).
– Add a magnetic stirring rod and stir (4.5) for 15 minutes. Cyanuric acid and urea
dissolve while biuret is practically insoluble.
– Pass the solution through a 1G3 filter crucible (4.6) and wash the crucible twice
with 5 ml of demi water followed by three times with 10 ml of acetone (5.11).
– Finally dry (4.2) for 3 hours at 105°C and keep the biuret in a brown, wide-mouth
bottle with ground stopper.
9.6 Urea causes an increase in extinction. Per g of urea in 50 ml of the solution to be
measured with the spectrophotometer the extinction increases with 0.0034, which
corresponds to 0.17 mg of biuret.
9.7 For the determination of biuret in urea solutions prepare the sample as follows:
– Place a round bottom boiling flask (4.10) for 1 hour in a drying oven (4.2) at 105 °C.
Allow the flask to cool in a desiccator.
– Weigh a g (accurate to 0.01 g) of the sample to this flask. Use such an amount of
sample that a residue of 3 – 8 g will be obtained.
– Connect the flask to the rotavapor (4.8).
– Set the water bath of the rotavapor at a temperature of 40 °C (In case of urea
solutions with small amounts of biuret never use temperatures higher than 50°C) and
the stirrer speed to 150 rpm.
– Use a vacuum pump (4.7) to create a vacuum of 50 mbar, read on a pressure gauge
(4.9).
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
– Stop evaporating when the content of the flask has solidified, allow the flask to stand
for 25 minutes.
If ammonia and/or ammonium salts are present in the urea solution in amounts
interfering with the determination of biuret (more than 1.5 mg of NH3 in the solution
to be measured with the spectrophotometer) proceed as follows:
After evaporation add 50 ml of methanol (5.8) to the content of the flask.
Again evaporate to dryness
– After that, place the flask for 1 hour in a drying oven (4.2) at 105 °C. Allow the flask to
cool in a desiccator.
– Dissolve the content of the flask in water.
– Transfer the solution into a 200 ml measuring flask, fill up with demi water and mix.
If necessary, filter through a folded filter (= solution A).
– Proceed as described under chapter 7 using this solution as solution A.
10. Reference
G:\methods\doc\402v1.doc
G:\Os99\990120020.xls
DSM method 135-E
DSM method 382-E
AGRO method 0042-E
Upgrade.
12. Attachments
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
Attachment 1: Biuret calibration curve (wavelength 550 nm, optical path length 5 cm)
0,400
0,350
y = 0,0218x
0,300 R2 = 0,9961
extinction (550 nm)
0,250
0,200
0,150
0,100
0,050
0,000
0,00 2,00 4,00 6,00 8,00 10,00 12,00 14,00 16,00 18,00
concentration biuret m g/50 m l
(*): Photometer adjusted to 100% transmission relative to the reference solution and
the extinction of the corresponding standard solution was measured.
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
Authorisation :
(signature)
1. Scope
2. Safety
Chemicals to be used:
Urea
Hydrochloric acid
Sodium hydroxide
Formaldehyde
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
3. Principle
4. Apparatus
5. Chemicals
6. Procedure
– Perform all the operations while passing nitrogen (5.7) over the solution.
– Pipette 15.00 ml of sodium hydroxide into a 250 ml beaker.
Use sodium hydroxide solution (5.3) if 20 – 100 mg/L of carbon dioxide is present.
Use sodium hydroxide solution (5.4) if 100 – 600 mg/L of carbon dioxide is present.
– Add vc ml of sample, accurate to 0.01 ml. Use such an amount of sample that 20 –
600 mg of carbon dioxide will be present in 100 ml (see 8.1).
– Dilute to 150 ml with demi water (5.6).
– Add a magnetic stirring bar.
– Add 25.0 ml of formaldehyde solution (5.5).
– Slowly titrate (4.1, 4.2) to pH 8.7 with hydrochloric acid (5.1) if sodium hydroxide
solution (5.3) has been used.
– Slowly titrate (4.1, 4.2) to pH 8.7 with hydrochloric acid (5.2) if sodium hydroxide
solution (5.4) has been used.
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
– Stop the titration if the pH remains 8.7 for 5 seconds. Titrant consumption= v ml.
– Run a blank. Titrant consumption = v0 ml.
7. Calculation
7.1 Calculation of the carbon dioxide content in mg/L using the formula.
(v0 - v) * c * 44 * 1000
--------------------------------
vc
8. Remarks
8.2 For solutions with a carbon dioxide concentration < 20 ml/L, a infrared detection
method should be used.
9. Reference
G:\methods\doc\403v1.doc
DSM 0648-E
DSM 0648 10-E
SCM 024
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
Upgrade.
11. Attachments
No attachments
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
Authorisation :
(signature)
1. Scope
2. Safety
Work safely. Use the guidelines of e.g. Safety Data Sheets (SDS) or a local
equivalent.
3. Principle
4. Apparatus
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
5. Chemicals
No chemicals necessary.
6. Procedure
- Place the round bottom boiling flask (4.5) for 1 hour in a drying oven (4.4) at 105
°C. Allow the flask to cool in a desiccator and weigh the flask (m0 g) (accurate to
0.01 g).
- Weigh a g (accurate to 0.01 g) of the sample into this flask. Use such an amount
of sample that a residue of 3 – 8 g will be obtained.
- Connect the flask to the rotavapor (4.2).
- Set the water bath of the rotavapor at a temperature of 40 °C and the stirrer
speed to 150 rpm.
- Use a vacuum pump (4.1) to create a vacuum of 50 mbar, read on a pressure
gauge (4.3).
- Stop evaporating when the content of the flask has solidified (see 8.2), allow the
flask to stand for 25 minutes.
- After that, place the flask for 1 hour in a drying stove at 105 °C. Allow the flask to
cool in a desiccator and weigh the flask (m1 g) (see 8.1).
7. Calculation
( m1 - m0 )
-------------- * 1000
a
Where:
m0 = weight of flask, empty and dried, in g
m1 = weight of flask with evaporation residue after drying, in g
a = the sample amount of the original sample in g
(weigh accuracy 0.01 gram)
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
8. Notes
9. Reference
G:\methods\doc\404v1.doc
DSM method 0382-E
Upgrade.
11. Attachments
No attachments.
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
DETERMINATION OF UREA
(spectrophotometric method, diacetyl monoxime method)
Authorisation :
(signature)
1. Scope
2. Safety
Chemicals to be used:
Urea
Sulphuric acid
Phosphoric acid
Acetic acid
Diacetyl monoxime
Thiosemicarbazide
Iron(III)chloride hexahydrate
Diacetyl monoxime and thiosemicarbazide are very toxic, take the appropriate
precautions (fume-cupboard, gloves and goggels).
Work safely, especially with hazardous chemicals. Use the guidelines of e.g. Safety Data
Sheets (SDS) or a local equivalent.
3. Principle
Reaction of urea with diacetyl monoxime in a phosphoric acid medium to form a violet-
coloured compound (using iron (III) as a catalyst).
The extinction of the coloured solution is measured at a wavelength of 525 nm.
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
4. Apparatus
5. Chemicals
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
6. Calibration
7. Sample preparation
– Grind the granules (4.2) to a particle size of less than 0.5 mm. This should be done
in a room with a relative humidity of 40-60% and a temperature of 22 ± 2°C.
– Weigh a g (accurate to 1 mg) of sample into a 250 ml measuring flask. Usually ± 10 g
of sample is used.
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
– If the urea solution is turbid, filter through a double fluted filter (4.6).
– Pipette v ml of the clear solution into a 50 ml measuring flask. Perform the analysis
on max. 10 ml of solution which may contain at most 0.1 mg of urea.
– Dilute to 10 ml if necessary.
– Add to a second 50 ml measuring flask 10 ml of demi water (=blank).
– Add 5.0 ml of colour reagent (5.9) and mix.
– Add 20.0 ml of acid reagent (5.13) and mix.
– Place the flasks in the thermostat (4.3) and allow to stand at 90°C for 1 hour.
– Cool to room temperature.
– Make up to the mark with demi water and mix.
– Determine the absorbance (4.1) at a wavelength of 525 nm and a path length of 1 or
5 cm.
8 Calculation
Ci
fi = -----------
Ei – E0
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
Calculate the average calibration factor, or use linear regression to calculate the
calibration curve (see attachment 1).
8.2 Calculation of the urea content of the NP and NPK granules in mg/kg using formula:
b * 250 * 1000
mg/kg urea = -----------------------
a*v
8.3 Calculation of the urea content of the urea solution in mg/L using formula:
b * 1000
mg/L urea = ------------------
v
9. Remarks
9.1 The use of clean glassware is very important. Therefore it is recommended to wash
the glassware with acid reagent (5.13) and colour reagent (5.9) prior to
measurement.
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
10. Reference
G:\methods\doc\405v1.doc
DSM-DMG 0209-E
Note ABR 87-10: Determination of trace quantities of urea (1)
Note ABR 88-21: Determination of trace quantities of urea (2)
Upgrade.
12. Attachments
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
0,4
0,35
extinction (525 nm)
0,3 y = 4,2749x
2
R = 0,9954
0,25
0,2
0,15
0,1
0,05
0
0 0,02 0,04 0,06 0,08 0,1
concentration urea in mg/50 ml
[Urea] Extinction Es – E0
(mg/50 ml) Es
0,00 0,0187 (= E0) 0
0.0101 0.0606 0.0419
0.0305 0.1624 0.1437
0.0506 0.2282 0.2095
0.0708 0.3347 0.3160
0.0915 0.3990 0.3803
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
Authorisation :
(signature)
1. Scope
Determination of water in urea. This method is also applicable to urea melt and to other
solid intermediate products from the urea plant.
2. Safety
Chemicals to be used:
Hydranal Composite 5K
Hydranal Working medium K
Hydranal Water standard
Methanol
Urea
Work safely, especially with hazardous chemicals. Use the guidelines of e.g. Safety Data
Sheets (SDS) or a local equivalent.
3. Principle
- Dissolution of the urea directly in the Hydranal Working medium K in the titration
vessel. Or dissolution of the urea melt into methanol.
- Titration of water, in the urea or urea melt/methanol solution, with the Hydranal
Composite 5K titrant.
- Biamperometric detection of the titration endpoint, with a Pt/Pt-electrode.
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
4. Apparatus
5. Chemicals
6. Procedure
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
- Calculate the water concentration in wt% according to the formula given in 7.2.
- The water concentration must be between the given borders of the Hydranal-Water
standard.
7. Calculation
(b1 * a1 * 0.01) + a2
Tn = -------------------------- * 100 * T0
(a1 + a2) * b2
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
7.2 Calculation of the water concentration in the Hydranal Water standard in wt%
v*T
------------- * 100
e * 1000
v*T
------------- * 100
w * 1000
(c3 – c1)
Dilution factor: f = ---------------
(c3 – c2)
f*v*T
wt% H2O in urea melt = -------------- * 100
c4 * 1000
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
8. Reference
G:\methods\doc\020v1.doc
G:\methods\doc\406v1.doc
DSM 805-E
DMG 309
DMG 0281-2B-E
9. Remarks
9.1 Always close the titration vessel immediately to prevent moisture contamination.
11. Attachments
No attachments
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
Authorisation :
(signature)
1. Scope
Determination of water in urea. This method is also applicable to urea melt and to other
solid intermediate products from the urea plant.
2. Safety
Chemicals to be used:
Hydranal Composite 5K
Hydranal Working medium K
Hydranal Water standard
Methanol
Urea
Work safely, especially with hazardous chemicals. Use the guidelines of e.g. Safety Data
Sheets (SDS) or a local equivalent.
3. Principle
- Dissolution of the urea directly in the Hydranal Working medium K in the titration
vessel. Or dissolution of the urea melt into methanol.
- Titration of water, in the urea or urea melt/methanol solution, with the Hydranal
Composite 5K titrant.
- Biamperometric detection of the titration endpoint, with a Pt/Pt-electrode.
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
4. Apparatus
5. Chemicals
6. Procedure
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
- Calculate the water concentration in wt% according to the formula given in 7.2.
- The water concentration must be between the given borders of the Hydranal-Water
standard.
7. Calculation
(b1 * a1 * 0.01) + a2
Tn = -------------------------- * 100 * T0
(a1 + a2) * b2
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
7.2 Calculation of the water concentration in the Hydranal Water standard in wt%
v*T
------------- * 100
e * 1000
v*T
------------- * 100
w * 1000
(c3 – c1)
Dilution factor: f = ---------------
(c3 – c2)
f*v*T
wt% H2O in urea melt = -------------- * 100
c4 * 1000
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
8. Reference
G:\methods\doc\020v1.doc
G:\methods\doc\406v1.doc
DSM 805-E
DMG 309
DMG 0281-2B-E
9. Remarks
9.1 Always close the titration vessel immediately to prevent moisture contamination.
11. Attachments
No attachments
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
Authorisation :
(signature)
1. Scope
Determination of water in liquid ammonia, in concentrations between 0,005 and 1,0 wt%
2. Safety
Chemicals to be used:
Hydranal Composite 5K
Hydranal Water standard
Methanol
Liquid Ammonia
Ethylene glycol
Acetic acid
Work safely, especially with hazardous chemicals.
Liquid ammonia is highly corrosive and toxic. Liquid ammonia irritates the eyes
and the skin and causes serious chemical burns by frostbite.
Its vapour is strongly irritant to the mucous membrane and eyes, and produces a
suffocating effect on the respiratory tract.
Another person should always accompany a person taking samples.
When sampling, use thick rubber gloves, a facial shield and a helmet.
A gas mask with an ammonia filter should be within reach and ready for use.
Never work solitaire with the liquid ammonia at the laboratory till after the
evaporation of the liquid ammonia.
Never use materials made of polyethylene.
Hydranal Composite 5K contain toxic substances. Avoid inhalation and skin
contact.
Use the guidelines of e.g. Safety Data Sheets (SDS) or a local equivalent.
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
3. Principle
4. Apparatus
5. Chemicals
6. Procedure
6.1 Sampling
- Place two dry conical flasks (4.4) in the rack (4.5), one to purge the sample line and
one to take the sample.
- Introduce 5,00 ml of ethylene glycol (5.4) into the sampling flask (4.4)
- Close the inlet and the outlet of the two flasks (4.4) with rubber stoppers pierced
through with a capillary (e.g. a syringe needle) having a inside diameter of 1 mm (see
9.2). Close all glass connections with clamps.
- Install (when necessary) a sample line with a control valve.
- Remove the capillary-pierced stoppers from the inlet and the outlet of the flasks (4.4).
- Connect the inlet of the purge flask A (4.4) to the sampling point and the outlet to the
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
purge line.
- Cautiously open the valve of the liquid-ammonia line so that in about 1 minute 500 ml
of liquid ammonia flows into the purge flask A (4.4).
- (Note: If 500 ml is not sufficient to purge the sample line, adjust the purging procedure
to the operating conditions).
- Close the valve, disconnect the purge flask A (4.4) and now collect the sampling flask
B (4.4) to the sampling point and the purge line.
- At the sampling point, record the pressure or the temperature of the liquid ammonia to
be sampled, in order to be able to determine the evaporation factor during sampling;
see table appendix 11.2.
- Cautiously open the valve and take a sample by allowing 500 ml (=V1 ml) of liquid
ammonia to flow into the sample flask B (4.4) in about 1 minute.
- Close the valve, disconnect the sampling flask B (4.4) and place the capillary-pierced
stoppers on the inlet and outlet of the two flasks A and B (4.4).
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
7. Calculation
(b1 * a1 * 0.01) + a2
Tn = -------------------------- * 100 * T0
(a1 + a2) * b2
7.2 Calculation of the water concentration in the Hydranal Water standard in wt%
v * Tn
------------- * 100
e * 1000
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
(va-vb) * Tn * F * 10
--------------------------
V2 * 0,68 * V1
8. Reference
G:\methods\doc\407v1.doc
AGRO 0019-E
(ISO 7105)
9. Remarks
9.1 Always close the titration vessel immediately to prevent moisture contamination.
9.2 The capillary is necessary because pressure build-up in the flask owing to
evaporation ammonia must be prevented and the release of ammonia vapour during
transportation of the sample from the sampling point to the fume-hood of the
laboratory should be a minimum.
9.3 The methanolic solution of ethylene glycol to which acetic acid has been added
should be processed immediately, since extra water is formed as a result of the
esterification which occurs.
9.4 After some time the endpoint determination may become insensitive. In such a case
clean the platinum electrode by immersing 1 minute in a mixture (1:1) of nitric acid
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
(5.6) and hydrochloric acid (5.7) . Rinse the electrode with water. Dry the electrode
with a tissue and rinse the electrode with methanol (5.2).
11. Attachments
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
Appendix 11.2: Relation between the pressure, temperature and evaporation factor of
liquid ammonia
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
Authorisation :
(signature)
1. Scope
2. Safety
Chemicals to be used:
Chromotropic acid
Sulphuric acid
Formaldehyde = suspected carcinogen with a MAC TGG-15 min. value of 3 mg/m3
and a MAC TGG 8h value of 1.5 mg/m3.
Never use Hydrochloric acid in the neighbourhood of formaldehyde, because
formaldehyde reacts with Hydrochloric acid giving off very toxic
“bis(chloromethyl)ether” , which can cause lung cancer.
Work safely, especially with hazardous chemicals. Use the guidelines of e.g. Safety
Data Sheets (SDS) or a local equivalent.
3. Principle
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
4. Apparatus
5. Chemicals
5.1 Chromotropic acid disodium salt dihydrate C10H6Na2O8S2.2H20 p.a. e.g. Merck
5.2 Chromotropic acid solution: transfer 2.0 g chromotropic acid (5.1) in a 25 ml flask,
add 20 ml of demi water and stir for 15 minutes with a magnetic stirrer (4.2).
Filter through a hardened filter (4.4). This solution is stable for at least 1 week.
5.3 Sulphuric acid, 96% e.g. Merck
5.4 Formaldehyde solution 37 - 38 wt% e.g. Merck, r = 1.09 kg/L
Determine the formaldehyde content (Cf) according to ISO 2227
6. Calibration
– Transfer 3.6 g (m1) or 3,3 ml formaldehyde solution (5.4) to a 1 liter measuring flask.
Fill up to the mark with demi water and mix (= solution I).
– Transfer 1.8 g (m2) or 1,65 ml formaldehyde solution (5.4) to a 500 ml measuring
flask. Fill up to the mark with demi water and mix (= solution II).
– Pipette 20.00 ml of solution I in a 1 liter measuring flask and weigh the exact
amount (m3). Fill up with demi water and mix (= solution A).
– Pipette 10.00 ml of solution II into a 500 ml measuring flask and weigh the exact
amount (m4). Fill up with demi water and mix (= solution B).
– Pipette into 50 ml measuring flasks respectively: 0 (=blank) , 1.00, 3.00 and 5.00 ml
(Vi) of solution A. If necessary dilute to 5 ml with demi water.
– Pipette into 50 ml measuring flasks respectively: 2.00 and 4.00 ml of solution B (Vi).
If necessary dilute to 5 ml with demi water.
– Divide the solution over the entire bottom of the 50 ml measuring flask.
– Add 1.0 ml chromotropic acid solution (5.2) and mix.
– Carefully add 30 ml sulphuric acid (5.3) and mix.
– Heat for 30 minutes on a closed steam bath (4.3) or electric at a temperature of
125°C.
– Remove from the steam bath and cool down to ambient temperature.
– Carefully transfer the contents of the evaporating dish into a 500 ml measuring flask
containing already 200 ml of demi water.
– Fill up to the mark with demi water and mix.
– Perform the extinction measurements at a wavelength of 570 nm and an optical
path length of 5 cm using a spectrophotometer (4.1).
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
7. Sample preparation
8 Calculation
for solution I :
m1 * Cf * m3 * Vi
Ci= -----------------------
100 * 1000
for solution II :
m2 * Cf * m4 * Vi
Ci= -----------------------
100 * 500
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
Ci
fi= -----------
Ei – E0
Calculate the average calibration factor, or use linear regression to calculate the
calibration curve (see attachment 1).
b * 100
-------------
a*V
9. Remark
10. Reference
G:\methods\doc\408v3.doc
SBB 0071-02-E
ISO 2227, Formaldehyde solutions for industrial use, Determination of formaldehyde
content.
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
Upgrade.
Change the safety data of formaldehyde.
In paragraph 6, calibration: add weighing of the pipetted amount.
Add electric heating in paragraph 6 and 7, with the temperature.
Change the reference method for the determination of formaldehyde in ISO 2227 (5.4).
12. Attachments
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
0,7
y = 5,4105x
0,6 R2 = 0,9954
Extinction (570 nm)
0,5
0,4
0,3
0,2
0,1
0
0 0,02 0,04 0,06 0,08 0,1 0,12 0,14 0,16
concentration form aldehyde mg/ 500 m l
[Formaldehyde] Extinction
(mg/500 ml)
0.00 0,000
0.0269 0.151
0.0537 0.300
0.0806 0.464
0.1343 0.705
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
Authorisation :
(signature)
1. Scope
Determination of the free ammonia content (NH3) of urea and other intermediate
products obtained in the production of urea. By the term, ‘free ammonia’ is meant the
ammonia present in the sample. Any other ammonia compounds are not included in
the determination.
2. Safety
Chemicals to be used:
Hydrochloric acid
Phenolphthalein
Ethanol
Work safely, especially with hazardous chemicals. Use the guidelines of e.g. Safety Data
Sheets (SDS) or a local equivalent.
3. Principle
4. Apparatus
4.1 Burette, 5 ml
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
5. Chemicals
6. Procedure
7. Calculation
V * T * 17.03 * 1000
ppm free NH3= ----------------------------
a
8. Remarks
8.1 This method is also applicable for e.g. urea melt. Urea melt is a melt of 95-99 wt%
urea and 1 - 5 wt% water with original temperature of 130°C. Only take care that
solution A, the measuring solution, contains about 15 g urea/200 ml.
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
9. Reference
G:\methods\doc\410v1.doc
Chemlab 1605-OV-E
11. Attachments
No attachments.
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
Authorisation :
(signature)
1. Scope
Determination of the free ammonia content (NH3) of urea and other intermediate
products obtained in the production of urea. By the term, ‘free ammonia’ is meant the
ammonia present in the sample. Any other ammonia compounds are not included in
the determination.
2. Safety
Chemicals to be used:
Hydrochloric acid
Phenolphthalein
Ethanol
Work safely, especially with hazardous chemicals. Use the guidelines of e.g. Safety Data
Sheets (SDS) or a local equivalent.
3. Principle
4. Apparatus
4.1 Burette, 5 ml
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
5. Chemicals
6. Procedure
7. Calculation
V * T * 17.03 * 1000
ppm free NH3= ----------------------------
a
8. Remarks
8.1 This method is also applicable for e.g. urea melt. Urea melt is a melt of 95-99 wt%
urea and 1 - 5 wt% water with original temperature of 130°C. Only take care that
solution A, the measuring solution, contains about 15 g urea/200 ml.
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
9. Reference
G:\methods\doc\410v1.doc
Chemlab 1605-OV-E
11. Attachments
No attachments.
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
Authorisation :
(signature)
1. Scope
2. Safety
Chemicals to be used:
Hydrochloric acid
Salicylic acid sodium salt
Tri-sodium citrate-2-hydrate
Disodiumpentacyanonitrosoferrate(III))
Sodiumhydroxide
Dichloroisocyanuric acid sodium salt dihydrate
Ammoniumchloride
Bromothymolblue
Magnesiumoxide
Work safely, especially with hazardous chemicals. Use the guidelines of e.g. Material
Safety Data Sheets (MSDS) or a local equivalent.
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
3. Principle
4. Apparatus
5. Chemicals
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
6. Calibration
7. Sample preparation
8. Calculation
Fi= Ci
---------
Ei - E0
Calculate the average calibration factor, or use linear regression to calculate the
calibration curve (see attachment 1).
Cs = b
-----
a
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
9. Remarks
10. Reference
G:\methods\doc.411v1.doc
DSM 0103-E
NEN 6472
12. Attachments
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
0.9
0.8
y = 0.0853x
extinction (655 nm)
0.7 2
R =1
0.6
0.5
0.4
0.3
0.2
0.1
0
0 2 4 6 8 10
concentration ammonia, ug/50 ml
0 0
0.975 0.081
2.403 0.205
4.698 0.397
7.242 0.620
9.673 0.825
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
Authorisation :
(signature)
1. Scope
Potentiometric determination of the ammonia content (NH3) of gas (e.g. off gas from
the prilling tower) obtained in the production of urea, by detection of equivalence
points.
2. Safety
Chemicals to be used:
Sodium hydroxide
Sulphuric acid
Bromphenol blue
Work safely, especially with hazardous chemicals. Use the guidelines of e.g. Safety
Data Sheets (SDS) or a local equivalent.
3. Principle
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
4. Apparatus
4.1 Apparatus for the determination of ammonia in gas from the prilling tower, see
appendix 1.
4.2 Potentiometric titrator, e.g. Mettler DL 25 or Metrohm Titrino
4.3 Combi pH electrode, e.g. Orion type 81-72 BN.
4.4 Washing bottle, capacity 250 ml, with sintered glass filter
4.5 Manometer, filled with demiwater
4.6 Gasmeter
4.7 Thermometer
4.8 Pressure regulator, filled with demiwater. The water level depends on the
resistance in t the washing bottle and the rate of sample introduction
4.9 Measure flask, 250 mL.
4.10 Vacuumpomp
5. Chemicals
5.1 Sodium hydroxide solution, c(NaOH) = 0,1 mol/L, e.g. Titrisol from MERCK
5.2 Sulphuric acid, c(H2SO4) = 0.05 mol/L e.g. Titrisol from MERCK.
5.3 Bromphenol blue:
5.4 Bromphenol blue solution: triturate 40 mg of bromphenol blue with 1 ml of sodium
hydroxide solution (5.1), dilute to 100 mL and mix.
6. Procedure
6.1 Sampling
- The quantity of air passed through should be such (usually 1 m3) that 2 – 200 mg of
ammonia is absorbed.
- Transfer 50,00 mL of sulphuric acid (5.2) to wash bottle A (4.4) and so much water
that the wash height is 8 cm.
- Set up the apparatus as indicated in the figure; at wash bottle A (4.4), connect it to
the sample point, and at stopcock 2 connect it to the vacuum pump.
- Close stopcock 1, switch on the vacuum pump and open stopcock 2 to such an
extent that a moderate stream of air is aspirated through pressure controlling device
E (4.8).
- Record the readings of the gas meter C (4.6) (= V0 L), the thermometer D (4.7) (=
T1 °C) and the barometer B (4.5) (= B1 mbar).
- Open stopcock 1 and, by adjusting stopcock 2 and controller E (4.8), pass air
through at a rate of 200 litres per hour.
- Record the under pressure indicated by manometer B (4.5) (U1 mbar).
- At the end of the sampling procedure, once more record the under pressure
indicated by manometer B (4.5) (U2 mbar), the temperature indicated by
thermometer D (4.7) (T2 °C) and the barometric pressure (B2 mbar).
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
- Close stopcock 1 and record the reading of the gas meter C (4.6) (V1 L)
- Close stopcock 2 and disconnect the apparatus from the sample point and the
vacuum pump.
- Switch off the vacuum pump.
- Transfer the contents of the wash bottle A (4.4) to a 250 ml measuring flask (4.9),
make up to the mark and mix (Solution I).
6.2 Determination
- Calibrate the titrator (4.2) at pH 7.00 and 4.00 according to the instruction manual of
the titrator.
- Pipette 100,00 mL of solution I into a 300 mL conical flask.
- Add three drops of bromphenol blue (5.3) and titrate with sodium hydroxide
solutions (5.1) until the colour changes (consumption v1 ml).
- Or perform an endpoint titration with sodium hydroxide solution (5.1) to pH 4.5.
- Determine the consumption of the sodium hydroxide solution (5.1) at the
equivalence point in mL (v1 mL).
- Run a blank by titrating 50,00 ml of sulphuric acid (5.2) with sodium hydroxide
solution (5.1), (consumption v0 mL).
7. Calculation
7.1 Calculation of the amount of air passed through wash bottle A, normalised to 0 °C,
1013 mbar and dry, using formula:
Where: V0 = the gas meter reading at the beginning of the sampling procedure,
in L
V1 = the gas meter reading at the end of the sampling procedure, in L
f = the correction factor of the gas meter
B = the average barometric pressure during the sampling procedure, in
mbar (= ½ B1 + ½ B2)
t = the average temperature during the sampling procedure, in °C
(= ½ t1 + ½ t2)
U = the average under-pressure during the sampling procedure, in
mbar (= ½ U1 + ½ U2)
Wt = the water vapour pressure at the average temperature T °C, in
mbar, see table (appendix 2)
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
8. Remarks
9. Reference
DSM 0465-E
DSM 381-34-E
G:\methods\doc\412v1.doc
11. Attachments
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
Authorisation :
(signature)
1. Scope
2. Safety
Work safely.
Use the guidelines of e.g. Safety Data Sheets (SDS) or a local equivalent.
3. Principle
4. Apparatus
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
5. Chemicals
No Chemicals necessary.
6. Procedure
6.1 Sample
At least 10 round granules are required to determine the crushing or deformation
strength. These granules are obtained as follows:
- Of a granulated product analyze the sieve fraction containing granules of average
diameter d (see method SCM 416).
To this end, sieve the “as supplied” product through the woven wire sieve which
has an aperture size that most closely approaches the average diameter of the
product from both sides. Take the roundest granules from the sieve fraction, if
necessary select granules with the apparatus used for determining the rolling
tendency according to method SCM 430.
Assume their diameter d to be the average of the apertures of the two sieves
used.
- Of a prilled product, analyze prills which diameters that come closest to the
average diameter d.
To this end, sieve the “as supplied” product through the woven wire sieve which
aperture size most closely approaches this average diameter.
Use only sound, round prills, which are retained in the meshes when the sieve is
turned upside down.
Their diameter d is the size of the apertures of the sieve used.
6.2. Determination
- With a pair of tweezers, place one granule of sample in the centre of the pressing
table of the crushing tester (4.1)
- Press the start button to raise the pressure and start the recording. If the granule
is crushed, the crushing tester (4.1) will relieve the pressure.
- Remove the remains of the granule with a brush (see 8.3).
- Repeat the above operations with at least nine granules,
7. Calculation
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
8. Remarks
8.2 The crushing or deformation strength is dependent on, among other factors, the
nature and composition of the product and, within the accuracy op the
determination depends only little on the diameter of the granule.
It is desirable to determine the crushing strength of the individual components of
bulk blends (where possible).
8.3 Keep the surface of the pressing table clean to permit proper contact.
9. Reference
11. Attachment
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
Authorisation :
(signature)
1. Scope
2. Safety
3. Principle
Dividing of the sample into a number of fractions with the help of a sieve shaker and
a set of wired sieves.
Weighing of the amount of sample retained on each sieve and calculation of the
different fractions.
4. Apparatus
4.1 Sieve shaker (see 8.1) giving vertical and horizontal motion to the granules, e.g.
Retsch AS 200 control G
4.2 Stainless steel woven wire test sieves provided with calibration certificates (with
diameters of 20 cm and 5 cm wide vertical rims) and having square apertures
according to ISO-565, e.g. Retsch (see attachment 2).
4.3 Stainless steel bottom pan with a diameter of 20 cm, e.g. Retsch.
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
5. Chemicals
No chemicals necessary.
6. Procedure
- Select a series of test sieves (4.2). If necessary, add one or more sieves from
the list of supplementary sizes for an accurate division of the sample into
fractions of different particle sizes.
- Place the bottom pan (4.3) on the sieve shaker (4.1) and stack the selected
sieves on top in order of increasing aperture size (see 8.2).
- Use the product as received, in an amount such that, after sieving the maximum
permissible volume of sample on each sieve used is not exceeded (see 8.1 and
attachment 2).
Use either the entire sample or a representative portion obtained with a sample
divider (4.5).
- Weigh the sample (see 8.3), accurate to 0.1 g (= m g) and place it on the top
test sieve.
- Place the cover (4.4) on the top sieve, fix the test sieve assembly and start the
sieve shaker (4.1).
- Sieve for 10 minutes (see 8.4).
- Remove the cover.
- Carefully lift the first sieve from the stack and invert it over a collecting bin.
Gently tap the side of the sieve a few times, if necessary remove granules which
are retained in the apertures by cautiously pressing them loose by hand (see
8.5).
- Weigh the sieve residue, accurate to 0.1 g (= m1 g).
- Proceed in the same way with the other sieves and with the bottom pan (= m2,
m3, m4, ..., mz g).
- Check that the maximum amount of residue permitted on each sieve has not
been exceeded (see attachment 2). If it has, the analysis should be repeated
with a smaller amount of sample or with a different set of sieves.
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
7. Calculation
7.1 Calculate M, the sum of the weights of the different sieve fractions:
M = m1 + m2 + … + mz
Where: m1: the weight of the sieve fraction on the first sieve, in g
m2: the weight of the sieve fraction on the second sieve, in g
mz: the weight of the sieve fraction on the bottom pan, in g
M: the sum of the sieve fractions, in g.
7.2 Determine the difference between sample weight and sum of sieve fractions:
M–m
This difference should be £ 0.005 g. If this condition is not met, the sieve
analysis must be repeated with a fresh sample portion.
7.3 Calculate F, the magnitude of each sieve fraction in wt %, using the formula:
mi
F= ------- * 100
M
Where: F: the sieve fraction between sieve i and the sieve above that , in wt %
(round off to the nearest 0.1 %).
mi: the weight of the sieve fraction on sieve i, in g
M: the sum of the sieve fractions, in g.
8. Remarks
8.1 The sieve shaker must preferably be installed in a conditioned area with a
temperature of 22 ± 2 °C and a relative humidity between 40 and 45 %. All
operations must be carried out in this area.
8.2 If the sieve assembly is too high for the shaker, longer guide rods or some
‘half-height’ sieves (= 2.5 cm) must be used.
8.3 Depending on the average particle size and the type of fertilizer a set of sieves is
recommended. (see attachment 1)
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
8.4 The sieving time depends on the nature of the sample and on the type and
setting (amplitude) of the shaker.
8.5 Never use a hard brush to remove granules that are retained in the apertures.
Regularly clean the sieves by immersion in an ultrasonic bath (4.6) filled with
water followed by drying in an oven (4.7) at a temperature of 100°C.
8.6 Regularly check (at least once a year) the accuracy of the sieving apertures with
a standard sample of glass beads or by means of a duplicate analysis with a
spare set of new sieves with a calibration certificate.
9. Reference
G:\methods\doc\SCM427v1
AGRO 1011
Upgrade.
11. Attachments
Sieves, Aperture sizes and an example for the calculation and presentation of the
analytical results.
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
Attachment 1: Sieves
Mean particle ±3 ±2 ±2 ±1
size (mm)
Set of sieves 4.75 2.8 4.75 2.0
with aperture 4.0 2.36 4.0 1.7
sizes (mm) 3.35 2.0 3.35 1.4
2.8 1.7 2.8 1.0
2.36 1.4 2.36 0.71
2.0 1.0 2.0 0.5
1.4 0.5 1.4
0.5 1.0
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
(*): The mass of the sample is calculated by multiplying the volume given in the table
by the bulk density (g/cm3) of the sample.
The principal sizes are also given in the standards series of DIN-4188, AFNOR X 11-
501, BS-410, ASTM E 11-81 and NEN 2560.
The supplementary sizes are listed in the standard series of ASTM 11-70 and BS-
410 only.
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
Attachment 3: Example
100
60
Q3(x) / %
50
40
30
20
10
0
400 600 800 1000 2000 4000
Particle Size / µm
particle size Urea 46%N
grain class Q3(x)
mm %
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
Authorisation :
(signature)
1. Scope
Determination of the total nitrogen content of urea. Nitrogen present as ammonium salts,
biuret, cyanuric acid etc. is included in the determination.
The nitrogen content can also be calculated from the water and formaldehyde content of
urea. For the determination of these two contents use analytical method SCM 406 and
SCM 408.
2. Safety
Chemicals to be used:
Sulphuric acid
Sodiumhydroxide
Work safely, especially with hazardous chemicals. Use the guidelines of e.g. Safety Data
Sheets (SDS) or a local equivalent.
3. Principle
- Conversion of the nitrogen present in urea to ammonia by heating with sulphuric acid.
- Determination of ammonia by the distillation method:
- Releasing of ammonia with sodium hydroxide, distillation and collection of the
released ammonia in a known amount of sulphuric acid.
- Titration of the excess amount of acid with sodium hydroxide, as a measure of the
amount of ammonia nitrogen.
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
4. Apparatus
5. Chemicals
5.1 Sulphuric acid, 98 wt%, c(H2SO4) = 18 mol/L, e.g. MERCK, for the determination of
nitrogen.
5.2 Sulphuric acid, c(H2SO4) = 0.5 mol/L, Titrisol e.g. MERCK.
5.3 Sodium hydroxide solution, c(NaOH) = 10 mol/L or min. 32% GR for analysis (for the
determination of nitrogen e.g. MERCK.
5.4 Sodium hydroxide solution, c(NaOH) = 0,5 mol/L, Titrisol e.g. MERCK
5.5 Methyl orange solution 0.1 wt%, e.g. MERCK
6. Procedure
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
7. Calculation
(Vo – V) * c * f * 14.01
-----------------------------
10 * a
Where:
Vo : the amount of sodium hydroxide solution (0,5 mol/L), consumed in the blank, in mL
V : the amount of sodium hydroxide solution (0,5 mol/L), consumed in the
determination, in mL.
c : the concentration of the sodium hydroxide solution (0,5 mol/L)
f : the dilutionfactor, in this case = 10.
a : the amount of sample used, in g
7.2 Calculation of the total nitrogen content in wt% from the water- and formaldehyde
content (SCM 406 and SCM 408)
(100 – X - Y) * 28
-----------------------
60
Where:
X : the water content determined by SCM 406, in wt-%
Y : the formaldehyde content determined by SCM 408, in wt-%
Accuracy of the analysis method determined from the relative standard deviations of the
two analytical methods. The relative standard deviation = 0.02% (based on 1 times s)
8. Reference
G:\methods\doc\417v1.doc
SCM 406
C5-0224/2003, European Parlement, 14-5-2003, Common stand.
DSM 0201-E
DSM 0130-E
9. Remarks
none
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
11. Attachments
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
Authorisation :
(signature)
1. Scope
2. Safety
Work safely.
Use the guidelines of e.g. Safety Data Sheets (SDS) or a local equivalent.
3. Principle
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
4. Apparatus
4.1 Equipment for the determination of the shock resistance (see attachment 1)
4.1.1 Metering device
4.1.2 Manometer
4.1.3 Reducing valve
4.1.4 Shooting pipe (glass)
4.1.5 Metal plate
4.1.6 Holder (Perspex) for the metal plate (4.1.5)
4.1.7 Receiver
4.1.8 Plastic stopper with seal (rubber ring)
4.2 Sieve shaker, woven wire sieves and bottom pan according to method SCM 416
4.3 Equipment for the determination of the rolling tendency of fertilizer granules/prills
according to SCM 430
4.4 Dry gas meter, calibrated, with a capacity of 25 m3/hour
4.5 Stopwatch
4.6 Compressed air with a gauge pressure of at least 2 bar (200 kPa)
5. Chemicals
No chemicals necessary.
6. Procedure
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
- After all the sample has been shot, close the valve in the compressed air line,
remove the stopper (4.1.8) and, using a brush, transfer the contents of the
receiver (4.1.7) to a sieve with apertures of 1.0 mm or 1.4 mm depending on the
d50 (see 6.1), equipped with a bottom pan (4.2).
- Remove any grit formed by sieving and introduce the sieve fraction into the funnel
of the metering device of the apparatus for determining the rolling tendency (4.3)
- After the entire sample has been fed, weigh the product in the receiver for the
round prills (= b g).
7. Calculation
b
B = --------- * 100
a
B
S = --------- * 100
A
Where: A= the rolling tendency of the prills in the sieved sample before the test
B= the rolling tendency of the prills in the sieved sample after the test
- Carefully remove the plate holder (4.1.6) from the shooting pipe (4.1.4) and
connect the gas meter (4.4) to this end of the shooting pipe (4.1.4) by means of a
flexible tube with a length of at most 5 cm.
- Connect the shooting pipe (4.1.4) to the compressed air system and use the
reducing valve (4.1.3) to set the pressure of the compressed air at 0.5 bar (as
indicated by the manometer (4.1.2)).
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
- Record the setting on the counter of the gas meter (S1) and simultaneously start
the stopwatch (4.5).
- Record the position of the counter of the manometer (S2) after, for example
exactly 300 seconds.
- Calculate the throughput in m 3 /hour (Q1) from the amount of air passing through
in m3 (S2 –S1) and the time in seconds.
- Using the gas meter, measure the amount of air at pressures of 1.0 and 1.5 bar
on the manometer (4.1.2) and then calculate the amount of air in m3/hour (Q2
and Q3 respectively)
- Close off the compressed air pipe, carefully remove the gas meter and connect
the plate holder back (4.1.6) to the shooting pipe (4.1.4).
- Plot the amount of air (m3/hour) against the pressure (bar).
- Calculate the amount of air (Q) required to create the specified air velocity of 21
m/s in the shooting pipe (4.1.4) on the basis of the inside diameter of the shooting
pipe (4.1.4) using the formula:
9. Reference
10. Remarks
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
12. Attachments
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
1: metering device
2: manometer
3: reducing valve
4: shooting pipe (glass)
5: metal plate
6: holder (perspex) for the metal plate
7: receiver
8: plastic stopper with seal (rubber ring)
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
Authorisation :
(signature)
1. Scope
2. Safety
Chemicals to be used:
n-Hexadecane
Tetrachloroethylene
Sulphuric acid
Liquid Ammonia
Work safely, especially with hazardous chemicals.
Liquid ammonia is highly corrosive and toxic. Liquid ammonia irritates the eyes
and the skin and causes serious chemical burns by frostbite.
Its vapour is strongly irritant to the mucous membrane and eyes, and produces a
suffocating effect on the respiratory tract.
Another person should always accompany a person taking samples.
When sampling, use thick rubber gloves, a facial shield and a helmet.
A gas mask with an ammonia filter should be within reach and ready for use.
Never work solitaire with the liquid ammonia at the laboratory till after the
evaporation of the liquid ammonia.
Never use materials made of polyethylene.
Handling with tetrachloroethylene always in a fume hood.
Use the guidelines of e.g. Safety Data Sheets (SDS) or a local equivalent.
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
3. Principle
4. Apparatus
4.1 PTFE outside coated conical flask of about 750 ml, provided with a normal-ground-
joint adapter tube and a 500 ml mark, see appendix 11.1. Use clamps to connect
the loose components with the conical flask.
4.2 Rack for at least 2 flasks (see 4.1).
4.3 Water bath with circulation-pump thermostat, capable of being set to 30°C.
4.4 Electronic thermometer with a long measuring probe and suitable for measuring
temperatures till –30°C.
4.5 Infrared spectrophotometer, e.g. Perkin Elmer 1310
4.6 Recorder paper
4.7 Seal able quartz cuvettes (e.g. quality Infrasil – QI – fa. Helma), with an optical path
length of 10 mm (see 9.2).
4.8 Separating funnel of at least 250 mL with ground stopper, provided with PTFE
stopcock (see 9.2).
4.9 Measuring flask, 50 and 100 mL made of glass (see 9.2).
4.10 Cotton wool, oil-free
5. Chemicals
5.1 Tetrachloroethylene,
5.2 Sulphuric acid, c(H2SO4) = 0.5 mol/L
5.3 n-Hexadecane,
5.4 Calibration solution: introduce 100 mg n-Hexadecane (5.3) (melting point 18.2°C;
heat to approx. 30°C on a water bath (4.3) and mix well before use) in a dry oil-free
measuring flask of 100 ml (4.1), make up to the mark with tetrachloroethylene (5.1)
and mix (1 ml = 1 mg oil)
6. Procedure
6.1 Sampling
- Place two dry conical flasks (4.1) (see 9.2) in the rack (4.2), one to purge the sample
line and one to take the sample.
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
- Close the inlet and the outlet of the two flasks (4.1) with rubber stoppers pierced
through with a capillary (e.g. a syringe needle) having a inside diameter of 1 mm (see
9.1). Close all glass connections with clamps.
- Install (when necessary) a sample line with a control valve.
- Remove the capillary-pierced stoppers from the inlet and the outlet of the flasks (4.1).
- Connect the inlet of the purge flask (4.1) to the sampling point and the outlet to the
purge line.
- Cautiously open the valve of the liquid-ammonia line so that in about 1 minute 500 ml
of liquid ammonia flows into the purge flask (4.1).
- (Note: If 500 ml is not sufficient to purge the sample line, adjust the purging procedure
to the operating conditions)
- Close the valve, disconnect the purge flask (4.1) and now collect the sampling flask
(4.1) to the sampling point and the purge line.
- At the sampling point, record the pressure or the temperature of the liquid ammonia to
be sampled, in order to be able to determine the evaporation factor during sampling;
see table appendix 11.2.
- Cautiously open the valve and take a sample by allowing 500 ml (=V1 ml) of liquid
ammonia to flow into the flask (4.1) in about 1 minute.
- Close the valve, disconnect the sampling flask (4.1) and place the capillary-pierced
stoppers on the inlet and outlet of the two flasks (4.1).
- Measure the band height, in mm, at 2925 cm-1 and read the corresponding amount
of oil from the calibration graph (= a mg). See appendix 11.2.
- Calculate the amount of oil using the formula given in 7.1.
7. Calculation
b * f * 1000
----------------
v * 0.68
Where:
b : the amount of n-hexadecane, in mg, read from the calibration graph.
f : the evaporation factor (see table, appendix 11.3) which is a correction factor for the
amount of liquid ammonia which evaporates during sampling.
v : the amount of liquid ammonia used, in mL.
0.68: The density of liquid ammonia at –33°C and 1 bar.
8. Reference
G:\methods\doc\419v1.doc
AGRO 0013-E
POLYCHEMLAB 1549v3
(ISO 7106)
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
9. Remarks
9.1 The capillary is necessary because pressure build-up in the flask owing to
evaporation ammonia must be prevented and the release of ammonia vapour during
transportation of the sample from the sampling point to the fume-hood of the laboratory
should be a minimum.
9.2 Glasswork: all the glasswork has to be cleaned with tetrachloroethylene (5.1) till oil-
free, before used.
9.3 The calibration solution of n-hexadecane with tetrachloroethylene has an infrared
absorption pattern which is virtually the same as the absorption patterns of the types of
oil which are used in UKF plants.
11. Attachments
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
Appendix 11.3: Relation between the pressure, temperature and evaporation factor of
liquid ammonia
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
Authorisation :
(signature)
1. Scope
2. Safety
Chemicals to be used:
Methanol
Xanthydrol
Sodium hydroxide
Hydrochloric acid
Acetic acid
Bromophenol blue
Urea
Work safely, especially with hazardous chemicals. Use the guidelines of e.g. Material
Safety Data Sheets (MSDS) or a local equivalent.
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
3. Principle
4. Apparatus
5. Chemicals
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
6. Procedure
6.2 If the amount of sample is more than 8 mL, but at most 25 mL, and contains
10-40 mg of urea.
- Transfer a suitable amount (usually between 8 and 25 mL) of sample to a 250 mL
beaker.
- Add 2 drops of bromophenol blue (5.7; see 8.2) and whilst stirring, neutralize drop
wise with hydrochloric acid (5.5)
Careful; possible carbon dioxide release.
- Cool, if necessary, and add 1 mL of hydrochloric acid (5.5) in excess.
- Allow the solution to stand for 5 minutes and then neutralize with sodium
hydroxide solution (5.3).
- Dilute to 30 mL with demi water, add 10 mL of methanol (5.9) and mix.
- Whilst stirring, first add 80 ml of acetic acid (5.8) and then xanthydrol/methanol
slurry (5.2) in 5 portions of 4.5 mL each; wait 5 minutes after each addition.
- After the last addition allow the beaker to stand for 1 hour (see 8.1) and proceed
as described under 6.3.
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
- Dry a glass crucible (G4, pore 4) for 1 hour in a drying oven at 105 °C, allow to
cool the crucible in a desiccator and weigh the crucible (m0 mg).
- Filter the contents of the beaker through the crucible (do not draw off till dryness).
- Wash beaker and crucible with about 20 mL of washing liquid (5.10).
- Finally, suck off till dryness, dry the crucible with its contents for 1 hour at 105 °C,
allow to cool the crucible with its contents in a desiccator and weigh (m1 mg)
7. Calculation
( m1 - m0 ) Murea
-------------- * ------------- * 1000
a Murea-xanthydrol-precipitate
Where:
m0 = weight of crucible, empty and dried, in mg
m1 = weight of crucible with its contents after drying, in mg
a = the sample amount of the original sample in ml
Murea = molar mass of urea (60 g/mol)
Muea-xanthydrol-precipitate = molar mass of urea-xanthydrol-precipitate (420 g/mol)
8. Remarks
8.1 Biuret may also give a precipitate with xanthydrol. However, biuret reacts much
more slowly than urea. An amount of less than 5 mg of biuret does not
interfere, provides the length of time between the last addition of reagent and
the filtration is not longer than 1 hour.
8.2 The treatment with hydrochloric acid can be omitted if no ammonia, carbamate,
carbonate and cyanate are present in the sample.
8.3 The quality of the xanthydrol/methanol slurry (10 % in methanol) shall be
checked by means of a determination with urea p.a. The solution is stable for 3
months, if stored in a properly closing brown flask.
8.4 Remove the precipitate from the crucible with a spatula. Clean the crucible with
methanol.
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
8.5 The method is also applicable for the determination of the urea content of the
liquid from desorber to sewer.
8.5.1 Sample
Use the sample solution obtained after sampling the product and pretreating the
sample, as described in sampling method SCM 401v2.
- Pipette vi mL (usually 25 mL) of solution A (for this sample solution see sampling
method SCM 400v1.
8.5.2 Procedure
- Pipette vi mL of the sample into a 250 mL beaker.
- Proceed as described under 6.2.
8.5.3 Calculation
Calculate the urea content in % (m/m), using the formula:
where :
m0 = mass of the empty crucible, in mg
m1 = mass of crucible plus precipitate, in mg
vi = pipetted amount of solution A, in mL
m3 = mass of the bladder before sampling (see sampling method 400v1, DSM
Melamine Skill Centre), in g
m4 = mass of the bladder after sampling (see sampling method 400v1, DSM
Melamine Skill Centre), in g
9. Reference
G:\methods\doc\ 420v1.doc
G:\methods\doc\ 400v2.doc
G:\methods\doc\ 401v2.doc
DSM method 0207-E
DSM method 0207 24-E
Upgrade.
11. Attachments
No attachments.
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
Authorisation :
(signature)
1. Scope
2. Safety
Chemicals to be used:
Acetone
Work safely, especially with hazardous chemicals. Use the guidelines of e.g. Safety Data
Sheets (SDS) or a local equivalent.
3. Principle
- Isokinetic sampling of an amount of dust-laden off gas from the gas duct (in-stack).
- Measurement of the volume of the collected gas.
- Separation of the dust from the off gas with a previously dried and weighed filter.
- Determination of the mass of the dust by weighing.
- Calculation of the gas particulate concentration.
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
4. Reference
G:\methods\doc\421v1.doc
DSM 1209-E
ISO 9096:2003
EPA, method 5
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
Authorisation :
(signature)
1. Scope
2. Safety
Chemicals to be used:
Mercury(II)thiocyanate,
Iron(III)nitrate
Methanol
Dimethylsulfoxid
Nitric acid
Sodiumhydroxide
Sodiumchloride
Work safely, especially with hazardous chemicals. Use the guidelines of e.g. Safety Data
Sheets (SDS) or a local equivalent.
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
3. Principle
4. Apparatus
4.1 Spectrophotometer or filter photometer e.g. Dr. Lange Lico 300. In case of a filter
photometer it should have a filter with a maximum transmission at app. 460 nm.
4.2 Glassware: Use dedicated glassware. Before it is used, the glassware should first be
rinsed with nitric acid (1 mol/L) (5.4) and then a number of times with millipore water
(5.1).
4.3 Plastic rectangular cuvettes with a path length of 5 cm e.g. LZM130, Dr. Lange.
4.4 pH indicator strips, pH 0-2,5, e.g. MERCK
5. Chemicals
5.1 Millipore water, free from chloride (<0.01 mg/L), conductivity 0.1 µS
5.2 Spectroquant chloride, 1.14755. e.g. MERCK. This reagent kit contains two reagents
5.2.1 Reagent Cl-1A, contains nitric acid (5-20 %) and Iron(III)nitrate (10-20 %)
5.2.2 Reagent Cl-2A, contains mercury(II)thiocyanate (0,5-2 % as Hg), methanol
(0,1-3 %) and dimethylsulfoxid (>50 %).
5.3 Nitric acid 65 %, e.g. MERCK
5.4 Nitric acid solution c(HNO3)= 1 mol/L: dilute 97,0 g nitric acid (5.3) into 900 ml
millipore water (5.1) in a 1L measuring flask. Mix and fill up to 1000,0 mL with
millipore water (5.1) and mix again.
5.5 Sodiumhydroxide (NaOH) pellets p.a. e.g. MERCK
5.6 Sodiumhydroxide solution c(NaOH) = 1 mol/L: dissolve 40,0 g sodiumhydroxide (5.5)
into 1000 ml millipore water (5.1).
5.7 Sodiumchloride (NaCl), pa, ± 99,5 % e.g. MERCK
6. Calibration
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
7. Sample preparation
– Measure the pH of the sample with the pH indicator strips (4.4). Not with a pH-meter
and an electrode because of the risk of contamination with the chloride of the KCl-
solution of the electrode.
– Adjust the pH with nitric acid (1 mol/L) (5.4) or with sodiumhydroxide solution
(1 mol/L) (5.6) to a pH of app. 1,0. Weight the sample before (m1) and after (m2) the
adjustment to correct later for the dilution.
– Pipette a ml (accurate to 0,01 ml) of sample into a plastic cuvette (4.3), with a
maximum of 5,00 ml. Usually 5,00 ml of sample is used so that 0 – 0,025 mg of
chloride will be present in the extinction measurement. Because of the high risk of
contamination with chloride, perform from each sample a triplet.
– Pipette so much Millipore water (5.1) in each cuvette that each cuvette contains 5,00
mL liquid.
– Introduce into the cuvettes:
6 drops of reagent CL-1A (5.2.1)
6 drops of reagent CL-2A (5.2.2)
Mix carefully with a stirring rod.
– Perform the following measurements at a wavelength of 460 nm and an optical path
length of 5 cm using a spectrophotometer (4.1).
– Adjust the photometer to 100% transmission relative to Millipore water, and measure
the extinction of the sample solution (Ei= extinction of sample solution).
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
8 Calculation
Cs
fs= -----------
Es – E0
Calculate the average calibration factor, or use linear regression to calculate the
calibration curve (see attachment 1). Leave out results that look contaminated with
chloride.
(Ei - E0) * fs * 5 m1
mg/L chloride= --------------------- * ----
a m0
Where:
a= amount of sample taken into investigation into the cuvette
(ml)
5= max volume in cuvette (ml)
m0= weight of sample before pH adjustment (g)
m1= weight of sample after pH adjustment (g)
9. Remarks
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
10. Reference
12. Attachments
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
Attachment 1: Chloride calibration curve (wavelength 460 nm, optical path length 5
cm)
6.000
Extinction (460 nm, 5 cm)
y = 1.0482x
5.000
R2 = 0.9982
4.000
3.000
2.000
1.000
0.000
0.000 1.000 2.000 3.000 4.000 5.000 6.000
Concentration chloride (mg/L)
(*): Photometer adjusted to 100% transmission relative to Millipore water and the
extinction of the standard solutions was measured and corrected for the extinction of
the standard solutions that contains no chlorid.
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
Authorisation :
(signature)
1. Scope
2. Safety
Work safely.
Use the guidelines of e.g. Safety Data Sheets (SDS) or a local equivalent.
3. Principle
- The loose bulk density is the mass of the sample, in grams, filled into a 1-litre
vessel under specified conditions.
- The tapped bulk density is the mass, in grams, of 1-litre of sample under
continuous vibration compacted to constant volume.
4. Apparatus
4.1 Apparatus for the determination of bulk density preferably made of stainless steel
(see attachment 1).
- Funnel with valve, fixed with 3 legs on the bottom plate
- Movable 1 litre measuring vessel without spout
4.2 Vibrator (220 V/45W), amplitude 0.10 – 0.15 mm, under load e.g. Retsch AS 200
control G
4.3 Stainless steel or plastic ruler, about 120 * 20 * 2 mm
4.4 Glass stick with a diameter of about 3 to 4 mm
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
5. Chemicals
No additional chemicals.
6. Procedure
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
7. Calculation
7.1 Calculate the loose bulk density in g/L using the formula:
where:
m0 = mass of the empty measuring vessel, in g
m1 = mass of the filled measuring vessel, in g (first determination)
m2 = mass of the filled measuring vessel, in g (second determination)
m3 = mass of the filled measuring vessel, in g (third determination)
v = volume of the measuring vessel, in L
7.1 Calculate the tapped bulk density in g/L using the formula:
where:
m0 = mass of the empty measuring vessel, in g
ma = mass of the filled measuring vessel after compaction, in g (first determination)
mb = mass of the filled measuring vessel after compaction, in g (second
determination)
mc = mass of the filled measuring vessel after compaction, in g (third determination)
v = volume of the measuring vessel, in L
8. Remarks
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
8.5 The waiting time is about 1 minute for prills, about 5 minutes for granulated
products.
The waiting time for unknown products must be found experimentally.
8.6 If moisture content of the sample is too high as result of moisture absorption
during storage or transport, the sample should be dried prior to analysis.
Sample is therefore stored in a room with constant low water vapour pressure.
9. Reference
G:\methods\doc\424v1.doc
UKF 1062-E
AGRO 1061
Upgrade
11. Attachments
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
Authorisation :
(signature)
1. Scope
Determination of the ammonia content (NH3) in urea and other intermediate products
obtained in the production of urea.
Besides ammonia present as ammonia also ammonia bound as carbonate, hydrogen
carbonate, carbamate and other ammonia salts are determined.
All components together are considered “total ammonia”, defined by an equivalence
point of pH = 5.2.
2. Safety
Chemicals to be used:
Hydrochloric acid
Methyl Red
Methylene Blue
Work safely, especially with hazardous chemicals. Use the guidelines of e.g. Safety Data
Sheets (SDS) or a local equivalent.
3. Principle
4. Apparatus
4.1 Burette, 5 ml
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
5. Chemicals
6. Procedure
7. Calculation
V * T * 17.03 * 1000
ppm NH3= ----------------------------
a
8. Remarks
8.1 This method is also applicable for e.g. urea melt. Urea melt is a melt of 95 % urea
and 5 % water with original temperature of 130°C. Only take care that solution A,
the measuring solution, contains about 15 g urea/100 ml.
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
9. Reference
G:\methods\doc\410v1.doc
G:\methods\doc\425v1.doc
DSM 206
Chemlab 1605-OV-E
11. Attachments
No attachments.
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
Authorisation :
(signature)
1. Scope
Determination of the ammonia content (NH3) in urea and other intermediate products
obtained in the production of urea.
Besides ammonia present as ammonia also ammonia bound as carbonate, hydrogen
carbonate, carbamate and other ammonia salts are determined.
All components together are considered “total ammonia”, defined by an equivalence
point of pH = 5.2.
2. Safety
Chemicals to be used:
Hydrochloric acid
Methyl Red
Methylene Blue
Work safely, especially with hazardous chemicals. Use the guidelines of e.g. Safety Data
Sheets (SDS) or a local equivalent.
3. Principle
4. Apparatus
4.1 Burette, 5 ml
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
5. Chemicals
6. Procedure
7. Calculation
V * T * 17.03 * 1000
ppm NH3= ----------------------------
a
8. Remarks
8.1 This method is also applicable for e.g. urea melt. Urea melt is a melt of 95 % urea
and 5 % water with original temperature of 130°C. Only take care that solution A,
the measuring solution, contains about 15 g urea/100 ml.
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
9. Reference
G:\methods\doc\410v1.doc
G:\methods\doc\425v1.doc
DSM 206
Chemlab 1605-OV-E
11. Attachments
No attachments.
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
Authorisation :
(signature)
1. Scope
2. Safety
Chemicals to be used:
Hydrochloric acid
Work safely, especially with hazardous chemicals. Use the guidelines of e.g. Safety
Data Sheets (SDS) or a local equivalent.
3. Principle
4. Apparatus
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
5. Chemicals
6. Procedure
- Calibrate the titrator at pH 7.00 and 4.00 according to the instruction manual of the
titrator.
- Add into a 400 ml beaker a g, accurate to 0.01 g, of urea (usually ± 15g), or a g of
solution A (see 8.1)
- Add 200 ml of demi water and dissolve the sample (=solution B)
- Perform an endpoint titration with hydrochloric acid 0.01 mol/l (5.1) to pH 4.5.
- Determine the consumption of hydrochloric acid (5.1) at the equivalence points in
ml.
7. Calculation
7.1 Calculation of the free ammonia content as defined by the equivalence point at
pH ± 7.5, using formula:
V1 * T * 17.03 * 1000 * f
ppm NH3= -------------------------------
a
7.2 Calculation of the total ammonia content as defined by the equivalence point at
pH ± 5.2, using formula:
V2 * T * 17.03 * 1000 * f
ppm NH3= -------------------------------
a
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
8. Remarks
8.1 This method is also applicable for the following samples (solution A):
a. the urea solution of the outlet of the reactor.
b. the urea solution of the outlet of the stripper before the expansion valve.
c. the urea melt from the evaporation unit. Urea melt is a melt of approximately
95 - 99 wt-% urea and approximately 1 - 5 wt-% water with original
temperature of 130°C.
d. the urea solution of the outlet recirculation separator.
e. the carbamate solution from the low pressure carbamate condenser.
f. the liquid to the desorber.
g. the reflux condensor level tank.
h. the urea solution from the buffertank.
i. the liquid from desorber to sewer.
Use the sample solution obtained after sampling the product and pretreating the
sample, as described in method SCM 400 and SCM 401.
- Pipet into a 400 ml beaker a ml of solution A.
- For sample a, e and g this is 5 ml.
- For sample b and f this is 25 ml.
- For sample c, d, h and i this is 100 ml.
- Continue as mentioned in par. 6.
Make sure that solution B, the measuring solution, contains about 15 g urea/200 ml.
9. Reference
SCM 400
SCM 401
DSM 381-34-E
G:\methods\doc\427v2.doc
11. Attachments
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
Authorisation :
(signature)
1. Scope
2. Safety
Chemicals to be used:
Hydrochloric acid
Work safely, especially with hazardous chemicals. Use the guidelines of e.g. Safety
Data Sheets (SDS) or a local equivalent.
3. Principle
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
4. Apparatus
5. Chemicals
6. Procedure
- Calibrate the titrator at pH 7.00 and 4.00 according to the instruction manual of the
titrator.
- Add into a 400 ml beaker a g, accurate to 0.01 g, of urea (usually ± 15g), or a g of
solution A (see 8.1)
- Add 200 ml of demi water and dissolve the sample (=solution B)
- Perform an endpoint titration with hydrochloric acid 0.01 mol/l (5.1) to pH 4.5.
- Determine the consumption of hydrochloric acid (5.1) at the equivalence points in
ml.
7. Calculation
7.1 Calculation of the free ammonia content as defined by the equivalence point at
pH ± 7.5, using formula:
V1 * T * 17.03 * 1000 * f
ppm NH3= -------------------------------
a
7.2 Calculation of the total ammonia content as defined by the equivalence point at
pH ± 5.2, using formula:
V2 * T * 17.03 * 1000 * f
ppm NH3= -------------------------------
a
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
8. Remarks
8.1 This method is also applicable for the following samples (solution A):
a. the urea solution of the outlet of the reactor.
b. the urea solution of the outlet of the stripper before the expansion valve.
c. the urea melt from the evaporation unit. Urea melt is a melt of approximately
95 - 99 wt-% urea and approximately 1 - 5 wt-% water with original
temperature of 130°C.
d. the urea solution of the outlet recirculation separator.
e. the carbamate solution from the low pressure carbamate condenser.
f. the liquid to the desorber.
g. the reflux condensor level tank.
h. the urea solution from the buffertank.
i. the liquid from desorber to sewer.
Use the sample solution obtained after sampling the product and pretreating the
sample, as described in method SCM 400 and SCM 401.
- Pipet into a 400 ml beaker a ml of solution A.
- For sample a, e and g this is 5 ml.
- For sample b and f this is 25 ml.
- For sample c, d, h and i this is 100 ml.
- Continue as mentioned in par. 6.
Make sure that solution B, the measuring solution, contains about 15 g urea/200 ml.
9. Reference
SCM 400
SCM 401
DSM 381-34-E
G:\methods\doc\427v2.doc
11. Attachments
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
Authorisation :
(signature)
Keywords:
1. Scope
2. Safety
Chemicals to be used:
Formic acid
Sodium hydrogencarbonate
Sodium carbonate
A HPLC apparatus operates under high pressure (100 - 400 bar). In spite of the
use of small volumes, always wear safety glasses when working with HPLC.
Work safely, especially with hazardous chemicals. Use the guidelines of
Material Safety Data Sheets (MSDS) or a local equivalent.
3. Principle
- Injection of the sample solution into the continuous eluent flow, via a sample loop.
- Separation of formic acid on a column packed with a low-capacity anion
exchanger based on divinylbenzene-polystyrene.
Volgnummer: fscm042
Aanmaakdatum: 17august1997
Dokument filename: j:\quality\formul\fscm034.doc
4. Apparatus
4.2 Filter: with pores of 0.2 or 0.45 µm, e.g. Schleicher & Schull with nylon
membrane, item no 10463020.
4.3 Glassware: only use glassware especially reserved for ion chromatography
analyses and even clean this glassware before using it by rinsing it twice with
water (5.1) or the sample solution. If possible don’t use glass stoppers
because they are always contaminated. If possible close the volumetric flasks
with the same syringe that will be used to fill the vials of the auto sampler.
5. Chemicals
Volgnummer: fscm042
Aanmaakdatum: 17august1997
Dokument filename: j:\quality\formul\fscm034.doc
6. Procedure
Starting up procedure
- Pump eluent through, successively, the injector, the columns, the ASRS
suppressor (eluent in/out), the cell of the conductivity detector and again the
ASRS suppressor (regenerant in/out), at a rate of 2 ml per minute.
- Adjust flow through the column to 2.0 ml/min and immediately switch on the
ASRS. Be sure to switch the ASRS off when the flow over the column is
stopped.
- The base conductivity is about 5 µS/cm.
-Weigh of stock solution A1; 0.05 – 0. 5 g (W a1) into measuring flasks of 200 mL.
-Fill up to 200 mL with water (5.1) and mix.
-Weigh of stock solution A2; 0.05 – 0. 5 g (W a2) into measuring flasks of 200 mL.
-Fill up to 200 mL with water (5.1) and mix.
-Calculate the concentration of formic acid from the calibration solutions (for
calculation see 8.1).
- Inject 50 – 100 µl (4.1) of each calibration standard onto the columns.
- Determine the peak surface of the component (Ac) (for an example of a
chromatogram see attachment 2).
- Prepare these solutions always freshly.
6.3 Determination
- If necessary, dilute a g of sample solution with water (5.1) to a mass of b g (=
solution B). The concentration of formic acid in solution B needs to be between
0.2 and 2.5 mg/L.
- Inject 50 – 100 µl (4.1) of solution B onto the columns.
- Determine the peak surface of the component (As).
- Inject the next sample as soon as the base conductivity has returned to its original
level and no interfering peaks of slowly eluting components are to be expected.
Volgnummer: fscm042
Aanmaakdatum: 17august1997
Dokument filename: j:\quality\formul\fscm034.doc
- Apply the washing procedure (7) if it is found that on account of the presence of
irreversible components in the column or very slow elution of the components the
next determination would be interfered with.
7. Washing procedure
8. Calculation.
f = Cc / Ac
Calculate the average factor, or use linear regression caculation (see attachment 3).
Cs = f * As * b / a
Volgnummer: fscm042
Aanmaakdatum: 17august1997
Dokument filename: j:\quality\formul\fscm034.doc
9. Remarks
10. Reference.
New method.
12. Attachments.
Volgnummer: fscm042
Aanmaakdatum: 17august1997
Dokument filename: j:\quality\formul\fscm034.doc
Composition of eluent
40 mg/L NaHCO3
Ion k’
fluoride <1
Acetate 1.5
Formiate 2.5
Chloride > 10
Nitrite > 10
Nitrate > 10
bromide > 10
Phosphate > 10
sulfate > 10
t r - t ro
k' =
t ro
Volgnummer: fscm042
Aanmaakdatum: 17august1997
Dokument filename: j:\quality\formul\fscm034.doc
Volgnummer: fscm042
Aanmaakdatum: 17august1997
Dokument filename: j:\quality\formul\fscm034.doc
3000
2500 y = 1110.4x
2
R = 0.998
area (counts)
2000
1500
1000
500
0
0 0.5 1 1.5 2 2.5 3
concentration formic acid (mg/L)
Volgnummer: fscm042
Aanmaakdatum: 17august1997
Dokument filename: j:\quality\formul\fscm034.doc
Authorisation :
(signature)
1. Scope
Determination of the rolling tendency of fertilizer granules and prills. This method
provides a measure of the roundness of the granules and prills.
2. Safety
Work safely.
Use the guidelines of e.g. Safety Data Sheets (SDS) or a local equivalent.
3. Principle
- Feeding an even and slow flow of sample onto a rotating, slightly inclined, flat disc
in such a way that the particles do not hinder each other’s movement.
- The particles roll down and change direction because of the rotary motion of the
disc. The particles with the highest rolling tendency will deviate the least from their
original direction.
- Catching the particles that roll off the disc within an arbitrarily chosen area.
- Weighing the particles caught and calculating the rolling tendency from the mass
of the particles caught.
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
4. Apparatus
4.1 Equipment (see 9.1) for the determination of the rolling tendency of fertilizers
see attachments 1 and 2
Equipment consists of:
4.1.1 Stainless steel funnel
4.1.2 V-shaped feed chute, the drop by the particles from the chute onto the
disc is
10 mm
4.1.3 Vibrator, e.g. Retsch AS 200 Control G
4.1.4 Melamine coated Pertinax disc (trade name Trespa) inclined at 7°30’
from the
horizontal
4.1.5 Disc drive set at 16 rpm
4.1.5 Compressed air supply
4.2 Level (calibrated)
4.3 Stopwatch
4.4 Protractor (calibrated)
4.5 “Eindmaten”, stainless steel blocks with fixed (calibrated) dimensions with sizes
1, 2, 3, 4 and 5 mm
4.6 Stainless steel woven wire sieve of 1.4 mm e.g. Retsch
5. Chemicals
No chemicals necessary.
6. Procedure
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
- At the end of the test, weigh, to the nearest 0.1 g, the mass (= m g) of the
particles caught in the tray for round product.
- Clean the funnel, feed chute and disc with alcohol if necessary.
7. Calculation
Calculation the rolling tendency of the sample in wt% using the formula
m
--------- * 100
a
8. Reference
G:\methods\doc\430v1.doc
AGRO 1081
03162001.doc
9. Remarks
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
Upgrade.
11. Attachments
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
Authorisation:
(signature)
1. Scope
Determination of the pH of a 25% solution of urea in water, the buffer value and the
amount of ammonium salts of urea. Under buffer value in this method is
understood: the amount of mmol acid/kg urea that is required to bring the pH of a
25% solution of urea in water to 5.4.
The method is only applicable to well hang urea: In the production area (e.g. silo)
urea should be stored for at least 7 days at a temperature of 35 – 40°C.
A laboratory equivalent will be the next treatment of a fresh produced urea sample:
Stabilize a sample in a closed bottle (filled up to 80%) for 24 hours at 60°C.
2. Safety
Chemicals to be used:
Formaldehyde
Sodium hydroxide
Sulphuric acid
Potassium chloride
Urea
Work safely. Use the guidelines of e.g. Material Safety Data Sheets (MSDS) or a
local equivalent.
3. Principle
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
4 Apparatus
4.1 pH meter, e.g. Mettler Toledo DL25 provided with a combi electrode e.g. Thermo
Orion Sure flow electrode type 91-72BN
5. Chemicals
6. Procedure
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
7. Calculation
7.1 Calculate the buffer value in mmol/kg urea, using the formula:
(v - v0) * T2 * 1000
---------------------------
100
7.1 Calculate the amount of ammonium salts in mmol/kg urea, using the formula:
8. Reference
New method.
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
Authorisation :
(signature)
Keywords: Urea, granules, ball mill, friability, abrasion and shatter resistance
1. Scope
2. Safety
Work safely.
Use the guidelines of e.g. Safety Data Sheets (SDS) or a local equivalent.
3. Principle
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
4. Apparatus
4.1 Steel drum provided with six flights and a sealable cover. Its axis of rotation is
at an angle of 30° to the horizontal. The drum is driven by a DC motor whose
speed is set to 60 ± 1 r.p.m. (see 6.2) (Reference: shop drawing CL 14534
d.d. 21-06-1989).
4.2 25 stainless steel balls, diameter 7.925 ± 0.025 mm, mass 2.015 ± 0.005 g
(see 6.2) (see 9.3).
4.3 Timer set to 15 minutes or stopwatch.
4.4 Sieve shaker, woven wire sieves and bottom pan according to method SCM
416.
5. Chemicals
No chemicals necessary.
6. Procedure
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc
7. Calculation
Calculate the friability in wt% of the analysis material using the following formula:
b
--------- * 100
a
8. Reference
9. Remarks
New method.
Volgnummer: fscm042
Aanmaakdatum: 12 augustus, 1997
Dokument filename: j:\quality\formul\fscm042.doc