Professional Documents
Culture Documents
Flexview Installation
Flexview Installation
Appendix
19. Appendix
The content of this memory stick should be updated at the first opportunity e.g.
during attendance of Wärtsilä service engineer.
Fig. 128
· Select Next in the following screen:
Fig. 129
· Carefully read the readme file and click Next again (Fig. 130).
Fig. 130
· In the Select Components window select the item according to your
requirements.
Fig. 131
· For a new installation it is required to install the VCI drivers first (refer to
chapter 19.2). Only then continue with the flexView Base-Installation
(chapter 19.4) and flexView Plant setup (chapter 19.5).
Caution!
Do not select the VCI driver installation, if they are already installed on your
computer, unless you need to reinstall the drivers for repair (chapter 19.7)
· Select the VCI driver in the Select Components window and click Next
(Fig. 131, Fig. 132).
Fig. 132
Caution!
The USB-to-CAN adapter must not be connected before the drivers are installed
and the PC has been restarted.
Fig. 133
Fig. 134
Fig. 135
· Click Finish when the installation process is finished.
Fig. 136
· Press OK to restart
Fig. 137
1 2
Hardware installation
Fig. 139
Fig. 140
After the successfull installation the “USB” LED on the USB-to-CAN adpter has
to be green.
Step 1
Fig. 141
Fig. 142
Step 3
Fig. 143
The correct drivers` version can be checked also here (for Windows XP):
· Go to Startè Control Panel è Add or Remove Programs menu
· There should be the following records shown in Fig. 144.
Fig. 144
Fig. 145
Caution!
Do not select the VCI driver installation if they are already installed on your
computer.
Installation window
The new flexView program and all necessary system files are installed on the
computer.
Fig. 147
Note:
It is not necessary to reboot.
Fig. 148
Click Yes.
Fig. 149
Click Install.
Fig. 150
Click OK.
Fig. 151
Fig. 152
Installation window
Fig. 153
Fig. 154
Fig. 155
Click Next.
Enter the vessel’s plant ID (Hull No.) as indicated in the Manual Control Panel
in the control room or the local control stand, when you press the
SOUND OFF – ALM ACKN. button for more than 5 seconds Fig. 156.
Fig. 156
Fig. 157
Plant ID input
Fig. 158
The vessel related files are created by the flexView installation program.
Installation window
Fig. 159
Installation is finished.
Note:
The flexView program screen layout is designed for the standard Windows
appearance layout.
Fig. 160
To change these settings, click with the right mouse button on the desktop and
select Properties Fig. 160.
Fig. 161
Select the card Appearance and change the Windows and buttons from
Fig. 162
Now the flexView screen is properly displayed by Windows.
Fig. 163
Fig. 164
flexView in demo-mode
If the CAN adapter is not connected flexView is only allowed to run in demo-
mode.
Fig. 165
Fig. 166
Fig. 167
If the USB-to-CAN adapter failed to initialize, the CAN LED remains off. Unplug
and re-plug the adapter into the USB port to fix the problem.
Single indications showing question marks appear, if other FCM modules are
switched off.
If all indication fields and menu bars changed to a light blue color and data is
not updated anymore, then a trend is running in the background with maximum
possible scan time.
This is not a failure but a normal condition: to prevent CAN bus overload the
scanning of all other flexView indications is paused, until the trend is stopped.
Fig. 168
After mechanical work on the CAS and CAS drive, the crank angle offsets must
be adjusted.
1. Turn the engine with the turning gear until the flywheel marks show TDC,
cylinder number one = 0.0°.
- Marks (1) and (2) should be aligned as indicated in Fig. 169.
- TDC of cylinder number one can be cross checked from the crank case.
But safety measures must be taken accordingly before opening of the
crankcase – see “Warning labels” located near the crank case doors on
the engine.
Note:
For negative deviations, e.g. 359.8<°, use negative values, e.g. -0.2<°,
see Fig. 169.
2 1 2
T.D.C. 1
2. Move the engine CCW by using the turning gear device until the LED on
the TDC pickup is switched off. Now turn the engine CW until the LED on
the TDC pickup lights up. Read the current crank angle position of the
engine by using the angle scale on the fly wheel and enter the value in
flexView, card ADJUST. See figure Fig. 170 pos.2.
3. Move the engine further by turning gear device CW until the LED on the
TDC pickup switches off. Then move the engine CCW until the LED of the
TDC pickup switches on. Read the current crank angle position of the
engine by using the angle scale on the fly wheel and enter the value in
flexView, card ADJUST. See figure Fig. 170 pos.3.
4. The mean TDC offset angle will be calculated automatically. The result is
displayed on “flex View”. See figure Fig. 170 pos.4.
4 2 3
8 6
7 5
Fig. 170
1. Pickups “A”, “B”, “C” and “REF” are wired with ACM-20. The associated
LEDs on the ACM-20 are either off or light up yellow.
2. Turn the engine CW by turning gear device. The pickup sequence on the
ACM-20 has to be in the correct order: “A”, “B”, “C” etc indicated by the
yellow LEDs. See figure Fig. 171
Fig. 171
3. Turn the engine CCW by using the turning gear device until the reference
mark is detected by the ACM-20 module. The “REF” LED lights up.
1. Turn the engine CCW past the reference mark – the LED switches off again
– until the engine is turned away more than 45° from the reference mark.
2. Switch on the dedicated calibration mode with “START” in the flexView card
ADJUST and then start turning CW the engine by turning gear device. From
now on, the engine is interlocked and cannot be started until the following
step has been completed. See figure Fig. 170 pos.5.
Remark: When the engine is turning the turning gear device, it must not
be stopped until the reference mark has been detected.
4. Turn the engine again CW or CCW over the same reference mark. The
ACM-20 module begins with the continuous transmission of the current
crank angles over SSI to WECS. The LED “CA” is switched on. See figure
Fig. 171.
2. Speed up the engine with air-run. During this phase the offsets “Meas offset
angles” are determined by the software (minimal speed is required
>20RPM). At standstill the values are displayed on flexView, card ADJUST.
See figure Fig. 170 pos.7
3. Put the values “Meas offset #1/#2” in field “ACM CA #1/#2 Offset” and the
values are now equal. See figure Fig. 170 pos.8. The calibration process
for the ACM-20 units is now completed.
Hardware Requirements:
Software Requirements:
Recommendation:
To prevent the execution of flexView with problems we highly recommend the
usage of a graphic adapter with dedicated memory (min. 256MB).
Remark:
To protect the PC from virus attacks it is essential that only flexView is installed
and executed on the PC. Wärtsilä cannot guarantee a proper execution of
flexView if two or more programs will be executed at the same time. Due to the
high amount of computer resources that flexView needs Wärtsilä proposes its
own PC.
Important:
It should be considered that a suitable computer is part of the yard supply.
Fig. 172