Professional Documents
Culture Documents
Case-Based Pace, ICDs, and CRT Volume 1
Case-Based Pace, ICDs, and CRT Volume 1
RESYNCHRONIZATION
PACEMAKERS, ICDs, AND CARDIAC
,
Edited by
Paul A. Friedman MD, FACC, FHRS | Melissa A. Rott RN
Anita Wokhlu MD | Samuel J. Asirvatham MD, FACC, FHRS
David L. Hayes MD, FACC, FHRS
A Case-Based Approach to Pacemakers, ICDs, and Cardiac Resynchronization
A Case-Based Approach
TO Pacemakers, ICDs, AND
Edited by
Paul A. Friedman MD, FACC, FHRS | Melissa A. Rott RN
Anita Wokhlu MD | Samuel J. Asirvatham MD, FACC, FHRS
David L. Hayes MD, FACC, FHRS
© 2011 Mayo Foundation for Medical Education and Research
First paperback edition, 2013
www.cardiotextpublishing.com
Any updates to this book may be found at: www.cardiotextpublishing ments and are urged to review the latest instructions and warnings for
.com/titles/detail/9781935395812. any medicine, equipment or medical device. Readers should consult
with a specialist or contact the vendor of any medicine or medical device
Comments, inquiries, and requests for bulk sales can be directed to the where appropriate.
publisher at: info@cardiotextpublishing.com.
Except for the publisher’s website associated with this work, the publish-
All rights reserved. No part of this book may be reproduced in any form er is not affiliated with and does not sponsor or endorse any websites,
or by any means without the prior permission of Mayo Foundation for organizations or other sources of information referred to herein.
Medical Education and Research. Direct permissions questions to Scien-
tific Publications, Plummer 10, Mayo Clinic, 200 First Street SW, Roch- The publisher and the author specifically disclaim any damage, liability,
ester, MN 55905. or loss incurred, directly or indirectly, from the use or application of any
of the contents of this book.
All trademarks, service marks, and trade names used herein are the
property of their respective owners and are used only to identify the Unless otherwise stated, all figures and tables in this book are used cour-
products or services of those owners. tesy of the authors.
This book is intended for educational purposes and to further general Cover and book design: Ann Delgehausen, Trio Bookworks
scientific and medical knowledge, research, and understanding of the
conditions and associated treatments discussed herein. This book is not Library of Congress Control Number: 2011925451
intended to serve as and should not be relied upon as recommending or ISBN: 978-1-935395-81-2
promoting any specific diagnosis or method of treatment for a particular
condition or a particular patient. It is the reader’s responsibility to deter- Printed in the United States of America
mine the proper steps for diagnosis and the proper course of treatment
for any condition or patient, including suitable and appropriate tests, 16 15 14 13 1 2 3 4 5 6 7 8 9 10
medications or medical devices to be used for or in conjunction with any
diagnosis or treatment.
Due to ongoing research, discoveries, modifications to medicines,
equipment and devices, and changes in government regulations, the in-
formation contained in this book may not reflect the latest standards,
developments, guidelines, regulations, products or devices in the field.
Readers are responsible for keeping up to date with the latest develop-
Editors and Other Contributors v
Editors
Paul A. Friedman MD, FACC, FHRS Samuel J. Asirvatham MD, FACC, FHRS
Consultant, Division of Cardiovascular Diseases Consultant, Divisions of Cardiovascular Diseases
Mayo Clinic, Rochester, Minnesota and Pediatric Cardiology
Professor of Medicine Mayo Clinic, Rochester, Minnesota
College of Medicine, Mayo Clinic Professor of Medicine and of Pediatrics
College of Medicine, Mayo Clinic
Melissa A. Rott RN
Heart Rhythm Services David L. Hayes MD, FACC, FHRS
Division of Cardiovascular Diseases Consultant, Division of Cardiovascular Diseases
Mayo Clinic, Rochester, Minnesota Mayo Clinic, Rochester, Minnesota
Professor of Medicine
Anita Wokhlu MD College of Medicine, Mayo Clinic
Fellow in Electrophysiology, Mayo School of Graduate
Medical Education
College of Medicine, Mayo Clinic, Rochester, Minnesota
Assistant Professor of Medicine, College of Medicine,
Mayo Clinic
vi Editors and Other Contributors
Contributors
Craig S. Cameron MD, FACC, Oklahoma Heart Institute, Tulsa, Oklahoma (Cases 52 and 53)
Gregory A. Cogert MD, FACC, Oklahoma Heart Institute, Tulsa, Oklahoma (Cases 52 and 53)
Connie M. Dalzell RN, Mayo Clinic, Rochester, Minnesota
Joseph J. Gard MD, College of Medicine, Mayo Clinic, Rochester, Minnesota
Michael Glikson MD, FACC, FESC, Leviev Heart Center, Sheba Medical Center, Tel Hashomer, Israel (Case 54)
Michael J. Hillestad RN, Mayo Clinic, Rochester, Minnesota
Nancy Y. Lexvold RN, Mayo Clinic, Rochester, Minnesota
Madhavan Malini MBBS, College of Medicine, Mayo Clinic, Rochester, Minnesota
Marjorie L. Martin RN, Mayo Clinic, Rochester, Minnesota
David A. Sandler MD, FACC, FHRS, Oklahoma Heart Institute, Tulsa, Oklahoma (Cases 52 and 53)
Matthew J. Swale MBBS, College of Medicine, Mayo Clinic, Rochester, Minnesota
K. L. Venkatachalam MD, Mayo Clinic, Jacksonville, Florida
Tracy L. Webster RN, Mayo Clinic, Rochester, Minnesota
Preface vii
The book that you hold in your hands, A Case-Based Approach to Pace- presented. In light of the growing use of pacemakers, defibrillators, and
makers, ICDs, and Cardiac Resynchronization: Questions for Examination resynchronization devices, we are confident that readers will find this
Review and Clinical Practice, is a compilation of our favorite teaching practical means of self-assessment and education useful. Although the
cases that were seen at or sent to Mayo Clinic. As our device practice has questions are designed in a multiple-choice format that may be particu-
grown, we have found that one of the best ways to remain current and larly useful for self-assessment for test-takers, they are not formally vali-
to educate incoming physicians and nurses is the review of interesting dated board questions. This book is for any individual who sees patients
“unknown” clinical cases. Consequently, we established a morning con- with implantable devices, or who will be taking an examination related
ference in 2008 for the purpose of presenting and discussing interest- to device management.
ing or uniquely educational cases. Since learners ranged from cardiology
fellows who were new to the device practice to experienced nurses and How to Use This Book
physicians, group discussion brought out facets of interest at all levels. The cases generally progress from simpler to more complex, understand-
Cases for this book were selected based on clinical relevance and their ing that there will be individual variation in what constitutes a difficult
usefulness for illustrating general principles, practical tips, or interesting case.
findings in device practice. Occasionally, manufacturer-specific features There is no table of contents because the case numbers are clearly
are discussed, but always with a goal of advancing general concepts in marked at the top of each page and we specifically did not want to
device management. include in the beginning of the book a listing of the “diagnosis” for each
The cases in this book are presented as a case history, an image case and therefore limit the ability for the reader to approach the cases
when pertinent, and a multiple-choice question. The answer and a as unknowns.
detailed explanation is presented on subsequent pages. We’ve adopted For the reader interested in reviewing a specific type of case (such
this format to encourage the reader to think through the differential as “T-wave oversensing” or “inappropriate shock”), two resources are
diagnosis and approach the clinical problem based on the information offered. An appendix is provided that identifies the major diagnostic
viii Preface
dilemma presented by each case, and the index will direct the reader to greatly from friends and colleagues at other institutions who have kindly
cases and discussions focusing on specific issues. However, we encour- shared interesting cases with us, and permitted us to include them in
age readers to progress sequentially through cases as unknowns to maxi- this work. We are grateful for their generosity. If you come across an
mize learning and interest. interesting case that you would like included in a future edition of this
This book is one of two volumes. The first volume includes intro- book, we would love to discuss it with you. E-mail addresses are listed
ductory and intermediate cases. The second volume includes additional below for that purpose. Please enjoy the cases! We look forward to your
intermediate cases as well as advanced cases. There are more multipart feedback and future contribution.
cases in volume 2, to delve more deeply into important concepts.
In various electronic versions of this book, hypertext links and —Paul Friedman MD and David Hayes MD
linked indices have been added to facilitate navigation. Also, a combined
index that covers both volumes is available at www.cardiotextpublishing
.com/titles/detail/9781935395812. Samuel Asirvatham: asirvatham.samuel@mayo.edu
This text includes a collective wisdom of numerous physicians, Paul Friedman: friedman.paul@mayo.edu
nurses, technicians, educators, and practitioners. We are indebted to the David Hayes: dhayes@mayo.edu
entire Heart Rhythm services team at Mayo Clinic for identifying and Melissa Rott: rott.melissa@mayo.edu
discussing cases, and educating us with them. We have also benefitted Anita Wokhlu: woklhu.anita@mayo.edu
Abbreviations ix
in the presence of ischemic heart disease. Four years later he received his first shock
during a visit to his chiropractor. He had sought chiropractic treatment due to chronic
shoulder discomfort.
Device settings:
• Mode: DDD
• Mode switch: on
• Pacing rate: 40 bpm (lower rate), 120 bpm (upper rate)
• Amplitude: 2.6 V (right ventricular), 2.0 V (atrial)
• Pulse width: 0.50 ms (right ventricular), 0.40 ms (atrial)
• Sensitivity: 0.18 mV (right ventricular), 0.18 mV (atrial)
• Dynamic AV: off
• Refractory after pace: 250 ms
• Antitachycardia therapies: initial burst ATP followed by 33-J shock
Q
:
As a result of the EGMs and therapies delivered,
what would be your next step?
1. Initiate new or additional antiarrhythmic medications
2. Consider ablation of the ventricular ectopic focus
3. Reprogram ventricular sensitivity
4. Look for source of EMI
4 A Case-Based Approach to Pacemakers, ICDs, and Cardiac Resynchronization
ventricular pace/sense lead when the chronic ventricular pace/sense lead was
damaged during recent cardiac surgery. He is pacemaker-dependent.
Device settings:
• Mode: VVIR
• Pacing rate: 60 to 120 bpm
• Ventricular lead: output 5.0 V at 0.4 ms; sensitivity 2.5 mV
• Pace/sense polarity: bipolar
The rhythm strip shown in Figure 2.1 was obtained following the procedure.
Friedman, Rott, Wokhlu, Asirvatham, Hayes 7
Q
: What is the etiology of these multiple pacing spikes?
1. Artifact from hospital monitor
2. Ventricular safety pacing
3. Biventricular pacemaker with significant VV offset
4. Presence of temporary pacemaker undersensing permanent pacing
8 A Case-Based Approach to Pacemakers, ICDs, and Cardiac Resynchronization
Device settings:
• Mode: DDDR
• Pacing rate: 70 to 120 bpm
• Paced AV delay: 150 ms
• Sensed AV delay: 120 ms
• Rate adaptive AV: on
Friedman, Rott, Wokhlu, Asirvatham, Hayes 11
Q
:
What phenomenon is captured on the rhythm strip
in Figure 3.1?
1. Rate response accelerates rhythm and lengthens AV delay in response to activity
2. Programming head brief loss of communication with device
3. Ventricular safety pacing
4. Mode switch in response to an atrial arrhythmia
12 A Case-Based Approach to Pacemakers, ICDs, and Cardiac Resynchronization
Figure 3.2 The first 7 events are presenting rhythm of AP-VP with rate-adaptive AV delay. The next 3
events, the rate increases to 100 bpm with shortened AV delay of 100 ms. In the last 6 events the rate
decreases to 85 bpm with programmed paced AV delay of 150 ms.
Case 4
RESYNCHRONIZATION
14 An 81-year-old male presents for routine dual-chamber pacemaker interrogation
PACEMAKERS, ICDs, AND CARDIAC
,
(Figure 4.1). His device was implanted 6 years ago for sick sinus syndrome with
syncope. He also has a history of paroxysmal atrial fibrillation.
Device settings:
• Mode: DDDR
• Pacing rate: 55 to 130 bpm
• Mode switch: on at atrial detect rate of 175 bpm
• Paced AV delay: 180 ms; sensed AV delay: 150 ms
• Atrial lead parameters: 3.5 V at 0.4 ms and 0.25 mV
• Ventricular lead parameters: 2.5 V at 0.4 ms and 4.0 mV
Friedman, Rott, Wokhlu, Asirvatham, Hayes 15
Q
:
between the surface ECG and both the marker channel
and the ventricular EGM?
1. Device undersensing atrial flutter
2. Surface ECG displaying artifact
3. Noise reversion from TENS unit
4. Antitachy pacing being delivered
16 A Case-Based Approach to Pacemakers, ICDs, and Cardiac Resynchronization
Q
:
Which of the following best describes the tracing
in Figure 5.1?
1. Normal device function
2. Intermittent output failure
3. Atrial lead fracture
4. Inappropriate mode switch
20 A Case-Based Approach to Pacemakers, ICDs, and Cardiac Resynchronization
Figure 5.2 A threshold test with a pacing rate that is close to the
intrinsic sinus rate, showing a ventricular threshold of 0.7 V.
Case 6
RESYNCHRONIZATION
22 A 52-year-old female undergoes dual-chamber pacemaker implantation for intermittent
PACEMAKERS, ICDs, AND CARDIAC
,
AV block. Right atrial and right ventricular leads are placed. Shortly after implantation
the ECG shown in Figure 6.1 is obtained.
Friedman, Rott, Wokhlu, Asirvatham, Hayes 23
Q
: What is the most logical first step after seeing this ECG?
1. Shorten the AV interval
2. Chest x-ray
3. MRI of chest
4. Reprogram atrial sensitivity
24 A Case-Based Approach to Pacemakers, ICDs, and Cardiac Resynchronization
6 2. Chest x-ray
The ECG demonstrates an RBBB, which usually indicates activation of the level of the atrial lead as opposed to being more inferior and cross-
the left ventricle. With a lead placed in the coronary sinus either inten- ing the tricuspid valve before being directed to the left. On the lateral
tionally or inadvertently, RBBB could be seen but the intention in this film the lead is directed posteriorly, consistent with a left ventricular
patient was to place the ventricular lead in the right ventricle. lead position.
RBBB could also be a result of the lead being placed across an When recognized acutely, the lead should be repositioned (Figure
atrial or ventricular septal defect (or patent foramen ovale) or by perfo- 6.3). Prior to doing so, it would be prudent to perform an ECG to deter-
ration of the right ventricle and advancement of the lead through the mine the location and size of the defect and whether there is any signifi-
pericardial space to the left ventricular surface. cant degree of right-to-left shunting. If significant right-to-left shunting is
A chest x-ray should be obtained to definitively determine the posi- present, the potential embolic risk of chronic transvenous pacing should
tion of the right ventricular lead. In the chest x-ray shown in Figure 6.2, be considered.
the lead has a high “take-off” on the PA film, ie, it crosses to the left at
Figure 6.2 Patient’s chest x-ray after implantation. Figure 6.3 Patient’s chest x-ray after repositioning
of the ventricular lead.
Friedman, Rott, Wokhlu, Asirvatham, Hayes 25
Following repositioning of the right ventricular lead, the ECG and present. Nothing would be accomplished by reprogramming any param- 6
chest x-ray shown in Figure 6.4 were obtained. On the chest x-ray the eters, answers 1 and 4.
“take-off” for the ventricular lead on the PA film is now considerably MRI, answer 3, was not necessary to make the diagnosis in this pa-
lower, and on the lateral film the lead is now directed anteriorly. tient, and generally MRI is a relative contraindication unless the patient
Other than a RBBB morphology of the paced QRS, no pacing has an MRI-resistant pacing system in place.
abnormalities are noted on the original ECG. Lead change artifacts are
and dilated cardiomyopathy. His second dual-chamber ICD, a Medtronic model 7271,
is currently implanted. The RV lead is a Medtronic model 6943-65 true bipolar lead
implanted with his initial ICD 7 years ago.
Device settings:
• Mode: DDDR
• Pacing rate: 70 bpm (lower), 120 bpm (upper)
• AV delay: 230 ms (paced and sensed)
• Amplitude: 2.0 V (atrial), 3.0 V (ventricular)
• Pulse width: 0.6 ms (atrial), 0.6 ms (ventricular)
• Sensitivity: 0.45 mV (atrial), 0.6 mV (ventricular)
Q
: Why are there 2 “VS” markers for each QRS?
1. P-wave oversensing
2. T-wave oversensing
3. Double-counting of the QRS complex
4. Bigeminal PVCs
28 A Case-Based Approach to Pacemakers, ICDs, and Cardiac Resynchronization
7 1. P-wave oversensing
Markers show that the atrial lead first senses the P wave labeled AS, and Differential diagnosis for this phenomenon would include:
then the ventricular lead senses both the P wave and the R wave as VS. • RV lead dislodged to the tricuspid annulus or right atrium
Note that this oversensing of the P wave by the ventricular lead does not near the valve
occur with every intrinsic beat, as the fifth and sixth events are sensed • RV lead implanted in the coronary sinus
appropriately. This oversensing occurred at the programmed sensitivity • Lead-to-lead interaction with the atrial lead hitting the
of 0.6 mV (Figure 7.2). The device was then reprogrammed to 0.9 mV ventricular lead during valve motion
(Figure 7.3) and the far-field P-wave oversensing by the ventricular lead • RV lead distal coil crossing tricuspid valve, thus sensing
was not seen at this setting. A chest x-ray (Figure 7.4) showed normal RV atrial activity due to integrated bipolar versus true bipolar
lead placement and integrity. Close monitoring of this lead was advised. sensing; this is more common with proximally placed lead
Eventually, the oversensing occurred at even the least-sensitive program- or in children (with smaller hearts)
mable value of 1.2 mV, so the lead was capped and a new lead implanted. • RV lead insulation break near atrial lead allowing for
sensing of atrial activity
device interrogation. Implant indication showed sinus node dysfunction with syncope.
Device settings:
• Mode: DDDR 60 to 130 bpm
• Mode switch: on at atrial tachycardia detection rate of 170 bpm
• Paced AV delay: 250 ms; sensed AV delay: 200 ms; search AV: on
• PVARP: 250 ms; rate-responsive PVARP high; PVAB 150 ms
• Atrial lead parameters: output 5.0 V at 0.4 ms; sensitivity: auto
• Ventricular lead parameters: output 5.0 V at 0.3 ms; sensitivity: auto
• Pace/sense polarity: bipolar
Diagnostics reveal 2189 mode switches occurring over the past 3 months since
implant. Total time of mode switching accounts for less than 1%, and the AT/AF
burden graph shows an overall time of 0%. Lead testing results in normal sensing,
impedance, and threshold values for both leads. Saved EGM for these episodes appears
in Figure 8.1
Friedman, Rott, Wokhlu, Asirvatham, Hayes 31
8
Figure 8.1 Patient’s EGM.
From top to bottom are
shown: atrial electrogram,
ventricular electrogram,
markers, and intervals.
Q
:
Which is the best explanation for these mode switch
occurrences?
1. Newfound paroxysmal atrial arrhythmia
2. Far-field R-wave oversensing
3. Retrograde VA conduction due to loss of atrial capture
4. Fractured atrial lead
32 A Case-Based Approach to Pacemakers, ICDs, and Cardiac Resynchronization
Figure 8.2 ECG and EGM recording during PVAB reprogramming. From top to bottom are shown: surface
ECG lead II, markers, atrial unipolar tip electrogram, and ventricular unipolar tip electrogram.
Case 9
RESYNCHRONIZATION
34 An 85-year-old male is seen in hospital complaining of intermittent chest pounding/
PACEMAKERS, ICDs, AND CARDIAC
,
Device settings:
• Mode: DDDR
• Pacing rate: 60 to 110 bpm
• AV delay: 180 ms
• PV delay: 150 ms
• Dynamic AV delay: on
• PVARP: dynamic on; maximum 320 ms
• Ventricular output: 2.5 V at 0.4 ms
• Atrial output: 2.0 V at 0.4 ms
Q
:
Assuming that the patient should be seen, which of the
following would likely be the most appropriate action?
1. Assess and reprogram the rate-adaptive sensor parameters
2. Lengthen the PV interval
3. Lengthen the PVARP
4. Reprogram to DDIR
36 A Case-Based Approach to Pacemakers, ICDs, and Cardiac Resynchronization
Figure 9.2 Stored event retrieved during clinic visit. From top to bottom are shown: atrial electrogram, ventricular electrogram, surface ECG, and markers.
Friedman, Rott, Wokhlu, Asirvatham, Hayes 37
the PV interval, answer 2, would not reliably prevent PMT. Although ever, in the case, the APC altered AV synchrony and allowed retrograde 9
programming to DDIR, answer 4, would prevent PMT, it is less appropri- conduction that led to PMT.
ate for the patient because atrial tracking would no longer be possible. One clinical question would be whether the initial tracing repre-
Any event that disrupts AV synchrony could potentially initiate a sented an atrial tachycardia. If this question persists in the clinic, if ap-
PMT. The most common cause would be premature ventricular contrac- plication of a magnet during the tachyarrhythmia fails to terminate the
tions, and the second most likely cause is atrial failure to capture. How- tachycardia, then the etiology is not pacemaker-mediated.
Figure 9.3 PMT termination. From top to bottom are shown: atrial electrogram, ventricular electrogram, surface ECG, and markers.
Case 10
RESYNCHRONIZATION
38 A patient presents for a preoperative evaluation of his permanent pacemaker, which
PACEMAKERS, ICDs, AND CARDIAC
,
Device settings:
• Mode: DDDR
• Pacing rate: 60 to 120 bpm
• Atrial output: 2.0 V at 0.4 ms
• Ventricular output: 2.5 V at 0.4 ms
• Sensitivity: A = 0.5 mV; V = 2.5 mV
• Accelerometer and minute ventilation sensors: on
Initial interrogation noted a single stored event of a high-rate episode (Figure 10.1).
The patient denied any symptoms; no awareness of palpitations or tachyarrhythmias.
Friedman, Rott, Wokhlu, Asirvatham, Hayes 39
10
Figure 10.1 Tracing from initial interrogation. From top to bottom are shown: atrial electrogram, ventricular electrogram, markers.
Q
:
Given this stored high-rate atrial episode,
what is your next step?
1. Program atrial sensitivity to a less-sensitive value
2. Activate mode switch
3. Turn off minute ventilation sensor
4. Attempt to correlate stored EGM with patient activities
40 A Case-Based Approach to Pacemakers, ICDs, and Cardiac Resynchronization
10
Figure 10.2 Stored event. From top to bottom are shown: atrial electrogram, ventricular electrogram, markers.
Case 11
RESYNCHRONIZATION
42 A 27-year-old woman received a pacemaker 2 weeks prior to the transtelephonic
PACEMAKERS, ICDs, AND CARDIAC
,
Device settings:
• Mode: DDD
• Lower rate limit: 60 bpm
• Upper tracking rate limit: 110 bpm; upper sensor rate limit: 120 bpm
• Mode switch: on
• Paced AV delay: 180 ms; sensed AV delay: 150 ms; search AV: on
• Atrial lead parameters: output 5.0 V at 0.4 ms; sensitivity 0.5 mV
• Ventricular lead parameters: output 5.0 V at 0.52 ms; sensitivity 2.8 mV
Friedman, Rott, Wokhlu, Asirvatham, Hayes 43
11
Q
:
What rhythm is captured on this nonmagnet
transtelephonic monitoring strip?
1. Upper rate limit behavior
2. Tracking of intermittent bursts of atrial tachycardia
3. Atrial undersensing
4. Intermittent ventricular failure to output
44 A Case-Based Approach to Pacemakers, ICDs, and Cardiac Resynchronization
11
12
Q
:
What does the tracing in Figure 12.1 include?
1. Intrinsic beat
2. Paced beat
3. Fusion beat
4. Pseudofusion beat
5. All of the above
48 A Case-Based Approach to Pacemakers, ICDs, and Cardiac Resynchronization
phology will vary depending on the relative contribution of each type one chamber is superimposed on a deflection arising from the depolar- 12
of depolarization. To make the diagnosis of a fusion beat, it is helpful ization of the other chamber.
to have a template of the intrinsic beat or a paced beat morphology for What is the clinical relevance of these findings? First, it is impor-
comparison. tant to recognize that fusion and pseudofusion beats can be normal phe-
Pseudofusion arises when a stimulus artifact is superimposed on nomena. Clear fusion confirms capture. However, when pseudofusion
an intrinsic P wave or QRS complex, giving the false appearance of cap- or questionable fusion occurs, it may be difficult to determine if the
ture. Although the stimulus artifact is present, the morphology of the P device failed to sense intrinsic or if the stimulus failed to cause depo-
wave or QRS is unchanged from the intrinsic morphology. The pacing larization. In some cases, it is helpful to confirm capture by increasing
stimulus is ineffective because it has been delivered when the tissue is the pacing rate to assess for a morphology consistent with capture and
refractory. to perform a capture threshold test. It is also important to confirm sens-
Pseudo-pseudofusion, not shown, occurs in the setting of dual- ing, which can be done by decreasing or inhibiting the basal pacing rate.
chamber pacing (most often DVI mode) when a stimulus artifact from
Case 13
RESYNCHRONIZATION
50 A question of pacemaker malfunction has arisen in a patient with a pacemaker
PACEMAKERS, ICDs, AND CARDIAC
,
programmed to DDDR, 60 to 140 bpm, with paced AV delay at 220 ms. You are called
to evaluate the rhythm strip shown in Figure 13.1. In some cycles, 2 pacing artifacts
are seen, but in other cycles, there are no pacing artifacts.
Friedman, Rott, Wokhlu, Asirvatham, Hayes 51
13
Q :
intervening QRS complex (circled in Figure 13.1) occurs
in which portion of the device timing cycle?
1. Ventricular blanking period
2. Crosstalk sensing window
3. Alert period
4. Postventricular atrial blanking period
52 A Case-Based Approach to Pacemakers, ICDs, and Cardiac Resynchronization
13
Q :
Which term is appropriate for this ECG appearance
(circled in Figure 13.1)?
1. Fusion beat
2. Pseudo-pseudofusion beat
3. Pseudofusion beat
4. Safety pacing
54 A Case-Based Approach to Pacemakers, ICDs, and Cardiac Resynchronization
13 2. Pseudo-pseudofusion beat
This case tests your understanding of pacemaker timing and pseudo- Some form of pseudofusion or pseudo-pseudofusion appears to be
pseudofusion. going on. To make the distinction, you need to discern whether the
In Figure 13.1, all the QRS complexes are narrow, consistent with artifact that precedes the QRS is derived from the ventricular chamber
intrinsic conduction. Every alternate QRS complex is preceded by stimu- or the atrial chamber. When a pacing artifact from a given chamber fails
lus artifact without any change in QRS morphology in 2 leads. In addi- to contribute to depolarization because the chamber is refractory from
tion, every T-wave complex is preceded by a second stimulus, giving the an intrinsic depolarization, that is pseudofusion. When a pacing arti-
appearance of attempted dual-chamber pacing. fact arising from another chamber coincidentally precedes an intrinsic
Friedman, Rott, Wokhlu, Asirvatham, Hayes 55
depolarization, producing the same appearance of pacing noncapture Answer 3 is less likely because the pacing artifact in front of the T wave 13
coincident with intrinsic depolarization, that is pseudo-pseudofusion. is timed to 1000 ms from the previous QRS complex, suggesting they
Distinguishing the 2 would be simple if channel markers were are ventricular; conversely, the artifact in front of the QRS is timed 220
available. In the absence of such information, one needs to march out ms in front, suggesting that the atrial pacing artifact precedes the QRS.
pacemaker timing cycles. Answer 4 is incorrect because ventricular safety pacing usually occurs
Answer 1 is incorrect because although there is a pacing stimulus after crosstalk, which is not seen here. In addition, the safety pacing
preceding the QRS, the QRS morphology is no different than intrinsic. stimulus will usually occur with a very short AV delay (approximately
100 ms) rather than a long AV delay.
Case 14
RESYNCHRONIZATION
56 A 14-year-old female presents with intermittent, symptomatic complete heart block
PACEMAKERS, ICDs, AND CARDIAC
,
and a DDDR pacemaker is placed. The pacemaker results in complete resolution of her
symptoms. Two years later she develops recurrent symptoms of light-headedness. At
this time she is essentially pacemaker-dependent. She notes that it is not as profound
as it was at the time the original diagnosis was made.
Pacemaker interrogation reveals the programmed parameters are identical to what had
been recorded at the time of her most recent visit to the pacemaker clinic. On testing,
the pacemaker threshold was 0.5 V at 0.5 ms, below which there was complete loss of
ventricular capture. However, when the rhythm strip was observed for longer periods
of time at 0.5 V and 0.5 ms the abnormality in the tracing in Figure 14.1 was noted.
Even when programmed to the highest programmable output values, the tracing
appeared identical. Ventricular impedance was checked repetitively with the patient
sitting quietly, with manipulation of the pacemaker pocket, and with the patient
performing various arm maneuvers. The impedance remained in the range of 550 to
600 ohms on all checks.
A chest x-ray was obtained and with close inspection no abnormalities of the pulse
generator or leads could be detected.
Friedman, Rott, Wokhlu, Asirvatham, Hayes 57
14
Q
: What is the most appropriate clinical approach?
1. Program to maximum voltage amplitude and pulse width and reevaluate in 1 month
2. Place a new epicardial ventricular lead
3. Place a new ventricular lead and avoid future competitive sports
4. Place a new ventricular lead via alternate venous route
58 A Case-Based Approach to Pacemakers, ICDs, and Cardiac Resynchronization
14
Device settings:
• Mode: AAIR
• Pacing rate: 60 to 140 bpm
• Atrial lead parameters: output 5.0 V at 0.4 ms; sensitivity 0.3 mV
Friedman, Rott, Wokhlu, Asirvatham, Hayes 61
15
Q
:
What is the best explanation for the rhythm shown in
Figure 15.1?
1. Atrial output programmed below threshold
2. P wave diminished and sensitivity needs adjusting
3. Atrial lead dislodgment
4. Undersensing of atrial tachycardia
62 A Case-Based Approach to Pacemakers, ICDs, and Cardiac Resynchronization
Figure 15.2 Patient x-ray of lead dislodgment. Figure 15.3 Lateral x-ray of lead dislodgment.
Friedman, Rott, Wokhlu, Asirvatham, Hayes 63
cause it must accumulate evidence for the atrial tachyarrhythmia. To can be used toward the criteria for an atrial tachyarrhythmia. With the 15
meet the criteria for an atrial tachyarrhythmia, the device must see more ventricular output programmed to minimal values, the drain on the bat-
atrial sensed events than ventricular events. If the device is programmed tery should be negligible. With the sensitivity at the least-sensitive set-
AAIR with no ventricular lead in place, it will sense the rapid atrial ting, any noise should be eliminated.
events above the atrial detection rate and annotate them as “TS” for Caution should be taken when the device is nearing the elective
“Tachy Sense,” but it will not treat these events because there is no ven- replacement indicator because this Medtronic pacemaker will switch
tricular event to compare them with, so detection is never met. If the from the programmed mode to the usual elective replacement indicator
device is programmed to AAIR with a ventricular lead implanted, the V parameters of VVI at 65 bpm. A patient who was reliant on the atrial
sensed events will count toward meeting the criteria for an atrial tachyar- pacing would be left without any pacing support.
rhythmia and treatment will occur. In this case, when no V lead is im- Although this was felt to be the option for this particular patient
planted, the device was programmed to either DDDR or DDIR, with the and in her best interest, and documented as such in the medical record,
ventricular outputs at the lowest programmable values and the sensitiv- the configuration utilized would be considered an off-label use of the
ity at the least-sensitive value so the “pseudo” ventricular paced events device.
Case 16
RESYNCHRONIZATION
64 A 73-year-old man with ischemic cardiomyopathy, QRS duration of 120 ms, and
PACEMAKERS, ICDs, AND CARDIAC
,
16
Q
: What is the most likely cause of these shocks?
1. Lead fracture
2. P-wave oversensing
3. T-wave oversensing
4. R-wave double-counting
66 A Case-Based Approach to Pacemakers, ICDs, and Cardiac Resynchronization
16 3. T-wave oversensing
This case illustrates the problem of inappropriate shocks for T-wave over- multaneous, alternating sets of VV intervals. The top row of dots sug-
sensing in the setting of a relatively small sensed R wave. gests a rate of approximately 600 ms (100 bpm). The lower row of dots
The first step is to recognize that the shocks are inappropriate. suggests a ventricular rate of 250 ms (240 bpm). Physiologically, such
The interval plot in Figure 16.1 provides the first clue. There are 2 si- alternation seems unlikely for ventricular tachycardia or ventricular fi-
brillation. This observation is sometimes referred to “tramtracking” or notice the small R-wave amplitude on the near-field. On the near-field, 16
“railroad tracks” and is consistent with double-counting or oversensing R waves are much smaller in amplitude overall—often less than 1 mV.
of alternating intervals (in this case, the QRS–T-wave and T-wave–QRS In fact, there is phasic variation to R-wave size (perhaps due to respira-
intervals). tion). As R waves diminish, T waves become more prominent. Recall
The next step is to determine the source of oversensing. It is im- that ventricular sensing is beat-to-beat function of Decay Delay curves
portant to look at all available EGMs to determine if P-wave oversensing, that depend on the size of the preceding R wave (as illustrated in Figure
R-wave double-counting, or T-wave oversensing is contributing to the 16.3). The diminished R to T ratio impairs the ability of the device to
inappropriate assessment of ventricular rates. “ignore” the T wave. Instead, the T wave is miscounted as a second R
Reviewing the EGMs and the corresponding sensing on the mark- wave. This is reported on the marker channel as ventricular fibrillation
er channel on Figure 16.3 clarifies the etiology of the abnormal interval detect followed by ventricular fibrillation sense (“FS”).
plots. The high-voltage or shock, far-field EGM is similar to a surface Answer 1 is incorrect because irregular R-R intervals and high-
EKG because it uses widely spaced, large electrodes (can/proximal coil frequency noise on the near-field EGM known as make-break potentials
and RV coil) and, therefore, can be used to identify P, QRS, and T waves. (often nonphysiologic, without QRS correlate) would be seen on the
In this case, the underlying rhythm is regular and normal in rate (600 near-field EGM. Answer 2 is incorrect because the double-counting was
ms or 100 bpm). The sensing, or near-field EGM is different. Acquisition not occurring at the time of expected P waves. Answer 4 is incorrect,
of signal comes from RV tip to ring, and therefore, only ventricularly because the R wave was not particularly wide (QRS duration 120 ms)
based signals are typically seen. Note that the scales on the near-field and and the double-counting occurred on the T wave.
far-field are automatically gained differently such that one might fail to
Q :
What is the next best step to manage these
inappropriate shocks?
1. Change the ventricular sensitivity threshold from 0.3 to 1.0 mV and repeat VF induction
2. Increase ventricular blanking following ventricular pacing
3. Force constitutive ventricular pacing to reduce T-wave amplitude
4. Lead revision
68 A Case-Based Approach to Pacemakers, ICDs, and Cardiac Resynchronization
16 4. Lead revision
Only revision of the ventricular lead or certain programming changes antitachycardia pacing, but not in inappropriate shocks. If the R wave is
(not listed among the answers) would be appropriate in this case. larger than the T wave, reprogramming can generally eliminate T-wave
T-wave oversensing can be categorized based on the near-field oversensing. However, if the R-wave amplitude is small (generally less
EGM as occurring (1) only during ventricular pacing, (2) with a large than 3 mV, although no specific value has been defined), reprogram-
R wave that is larger than the T wave (with R-wave amplitude >>T-wave ming is more difficult and increases the risk of undersensing. In our ex-
amplitude), or (3) with a small R wave that is smaller than the T wave. perience, programming options such as raising the minimum threshold
When T-wave oversensing occurs only following paced beats, the sensitivity value (ie, so-called decreasing the sensitivity, all manufactur-
pacing rate may be slowed (since sensing occurs off of the T wave rather ers) or adjusting Decay Delay (St. Jude Medical) are more effective when
than the QRS complex). This may result in bradycardia or ineffective the R to T ratio is greater than one. It has been suggested that stability
Friedman, Rott, Wokhlu, Asirvatham, Hayes 69
can be turned on if there is sufficient difference in the RT and TR inter- sensing incurs the risk of undersensing of R waves and possibly ven- 16
vals (so that therapy in the VT zone is inhibited due to interval variabil- tricular fibrillation. Increasing the ventricular blanking after ventricular
ity). However, when the R wave is smaller than the T wave, lead revision paced events (answer 2) might mitigate the T-wave oversensing that oc-
provides the most robust solution. Replacement with a true bipolar lead curs with pacing but will not affect T-wave oversensing of ventricular
(as opposed to an integrated bipolar lead) may minimize oversensing sensed QRS complexes. Forced ventricular pacing (answer 3) is not the
risk. Alternatively, a separate pace-sense lead can be added, as was done best answer. Although it can alter and reduce T-wave morphology and af-
in this patient. fect refractory periods, chronic RV pacing can negatively impact cardiac
Decreasing sensitivity (answer 1) is not ideal because the T wave is function in this patient with low ejection fraction. Additionally, prema-
larger than the R wave. Adjusting sensitivity to eliminate T-wave over- ture atrial or ventricular complexes may interrupt pacing and permit
oversensing to occur.
Case 17
RESYNCHRONIZATION
70 The same 73-year-old man from case 16, with a history of single-chamber defibrillator
PACEMAKERS, ICDs, AND CARDIAC
,
17
Figure 17.1 Arrhythmia episode. The top row demonstrates far-field signal
derived from the high-voltage EGM (shock). The bottom row demonstrates
near-field signal derived from a separate RV sense-pace lead.
Q
: What is the most likely cause of the shock episode?
1. Lead dislodgment
2. Dual atrial and ventricular tachycardias
3. Electromagnetic interference
4. Oversensing of diaphragmatic myopotentials
72 A Case-Based Approach to Pacemakers, ICDs, and Cardiac Resynchronization
17 1. Lead dislodgment
This case illustrates the problem of inappropriate shocks from ventricu- with ventricular oversensing of atrial fibrillation. Administration of
lar oversensing of atrial signal due to ventricular lead dislodgment. a defibrillatory shock clears up much of the high-frequency signal on
Comparison of the near-field and far-field channels in Figure 17.2 the near-field, likely terminating the atrial fibrillation. A wide complex
clarifies the problem. Recall that the far-field signal is derived from the rhythm is then seen on the far-field, and irregular complexes (probably
shock circuit (pulse generator can/superior vena cava coil and RV coil) atrial in origin) are sensed on the near-field. Most likely, the rhythm is
and provides morphology information. The near-field EGM, in this case, an idioventricular rhythm (as the ventricular rate exceeds the atrial rate),
is derived from a separately implanted RV pace-sense lead, and is used by although intermittent ventricular sensing of a slowed atrial dysrhythmia
the device to identify and count ventricular signals. with aberrancy or postshock morphology changes cannot be excluded.
The far-field channel shows irregular, but distinct, R-R intervals These findings are consistent with dislodgment of the ventricular
and no apparent P waves, which is consistent with atrial fibrillation. lead into or near the atrium, resulting in inadvertent misclassification of
The near-field signal, in contrast, looks erratic and is also consistent atrial fibrillation as ventricular fibrillation. Figure 17.3 shows the real-
time EGM that was transmitted following the shock. The ventricular ventricular signals. Atrial arrhythmias, in particular, may be sensed in- 17
EGMs on the near-field channel correspond to P waves rather than QRS correctly as ventricular arrhythmias.
complexes, indicating that the ventricular pace-sense lead was dislodged There is no ventricular tachycardia present, making answer 2 in-
to the atrium. The smaller, dedicated pace-sense lead may be more prone correct. Electromagnetic interference (answer 3) is incorrect because
to dislodgment into the atrium than a defibrillator’s lead. it would be present, and typically larger in amplitude, on the far-field
Oversensing of atrial signals on the ventricular lead can occur with EGM as well as the near-field EGM. Answer 4 is incorrect because dia-
dislodgment of the ventricular lead to the annulus or to the atrium (as in phragmatic myopotentials, although also near-field, present as bursts of
the present case). Integrated bipolar leads, particularly with proximally lower-amplitude, high-frequency noise that varies with respiration and
placed leads or in children, may also lead to oversensing of atrial EGMs occur most commonly after pacing or long diastolic intervals. The pat-
since the large distal coil is part of the sensing circuit. If it is placed at tern should be similar after a shock.
or near the level of the tricuspid valve annulus, it may record atrial and
receives a dual-chamber pacemaker. The next day, the telemetry strip shown in Figure
18.1 is recorded.
Friedman, Rott, Wokhlu, Asirvatham, Hayes 75
18
Q
: What does this tracing most likely represent?
1. Nonsustained VT
2. Atrial fibrillation with aberrant conduction
3. Normal pacemaker function
4. Pacemaker mediated tachycardia
76 A Case-Based Approach to Pacemakers, ICDs, and Cardiac Resynchronization
18
19
Q
:
What intervention is required to address
the ventricular tachycardia episodes?
1. Reposition the RV lead away from the tricuspid annulus
2. Turn on antitachycardia pacing at this heart rate
3. Adjust the atrial sensitivity to correct undersensing
4. Correct lead reversal in the header
80 A Case-Based Approach to Pacemakers, ICDs, and Cardiac Resynchronization
ated through the pacing system analyzer. As illustrated in Figure 19.3, • Failure to capture the ventricle with VVIR pacing 19
the timing on the A and V channels suggests that intrinsic events on • Failure to mode switch in a patient with known atrial
the “atrial” channel (actually sensed R complexes) intermittently follow arrhythmias
from events on the “ventricular” channel after 90 ms (actually atrial in
nature and consistent with AV conduction). Nevertheless, in the absence The ways to avoid lead reversal during dual-chamber implants
of surface EGMs, the presence of dual-chamber pacing at 60 bpm against include:
the backdrop of intrinsic sinus bradycardia with pauses complicates the • Verify by serial number or atrial lead labeling that leads are
diagnosis of lead reversal. properly inserted. Always connect the ventricular lead first.
Answer 1 is incorrect because moving the lead would not correct • Recheck all parameters when the leads are connected to the
the problem of header reversal. Answer 2 is incorrect because there is device. Dramatic changes in sensing or threshold should
no evidence of failure to pace consistent with an open circuit. Answer 3 catch your attention.
is incorrect because a defibrillator is not indicated. This is atrial flutter, • Ventricular pacing can be tested in VVI mode rather than
and mode switching is the more appropriate device response. Impor- DDI to allow for failure to capture in the event of lead
tantly, however, if this device were a defibrillator, the patient might have reversal. This will minimize the chance that atrial pacing
received inappropriate therapies for ventricular tachycardia. will not be mistaken for ventricular.
• The person who interrogates or a colleague should verify
Signs of lead reversal in the header are as follows: that surface EGMs correspond to the ventricular EGM,
• Surface ECG demonstrates events in the opposite chamber particularly if new, unexplained abnormalities arise.
from what is demonstrated by intracardiac ECGs
• Ventricular EGM amplitude is smaller than expected from
interrogation through the pacemaker system analyzer; atrial
EGM amplitude may be more robust than expected
Case 20
RESYNCHRONIZATION
82 A 53-year-old female with a permanent pacemaker (St. Jude Identity XL DR) recently
PACEMAKERS, ICDs, AND CARDIAC
,
The next afternoon, you are called to the patient’s bedside because it occurs
again while she is watching television. A 12-lead ECG confirms that the patient has
transitioned to a junctional, nonpaced rhythm. Device interrogation demonstrates
that the ventricular lead has a normal impedance and threshold value, and there is no
evidence of T-wave oversensing. Figure 20.2 demonstrates the heart rate histogram for
the 2 days prior to your assessment.
Friedman, Rott, Wokhlu, Asirvatham, Hayes 83
20
Device settings:
• Mode: VVI
• Pacing rate: 90 to 120 bpm
• V threshold lead: 0.75 V at 0.4 ms
• V impedance: 310 ohms
• Battery voltage: 2.76 V (ERI 2.5 V)
• V sensitivity: 4.0 mV
• Sensor: off
• Rest rate: off
• Maximum sensor rate: 120 bpm
• Hysteresis rate: 50 bpm
• Search interval: off
• AutoCapture: on
84 A Case-Based Approach to Pacemakers, ICDs, and Cardiac Resynchronization
20
20
Q
:
What is the likely explanation for the repeated failure to
pace below the lower rate limit?
1. Heart rate hysteresis
2. AV interval hysteresis
3. T-wave oversensing
4. Automated capture algorithm
86 A Case-Based Approach to Pacemakers, ICDs, and Cardiac Resynchronization
onstrates improvement of the heart rate histogram after programming incorrect because there is a sensing of a premature ventricular complex, 20
hysteresis to “off.” rather than T-wave oversensing. This results in inhibition of output and
Answer 2 is incorrect because the patient is in a VVI mode with initiates the hysteresis escape interval. Answer 4 is incorrect because
a silent atrium that is not being sensed. Search positive AV interval hys- automated capture algorithms usually work in concert with a backup
teresis is a specific type of hysteresis in which there is periodic prolon- safety pulse when capture fails. A marked and lasting change to a slow
gation of the PR interval after a ventricularly paced event. Answer 3 is intrinsic rate would be inconsistent with automated capture algorithms.
20
Q
:
What is the consequence of empirically changing the
ventricular sensitivity level from 1.5 to 4.0 mV?
1. Higher likelihood of T-wave oversensing
2. Atrial oversensing
3. Ventricular undersensing
4. Inappropriate activation of hysteresis
90 A Case-Based Approach to Pacemakers, ICDs, and Cardiac Resynchronization
20 3. Ventricular undersensing
Making the ventricular sensing channel less sensitive may be beneficial A potentially lethal consequence of ventricular undersensing is pacing
to limit the potential for sensing noise and double counting of T waves, on the T wave, which can induce ventricular tachyarrhythmias. R on T
but in this case, where the R wave is small, reprogramming resulted in resulting in a ventricular tachyarrhythmia is more likely to occur in the
ventricular undersensing. Figure 20.6 illustrates how this occurs. First, patient with significant LV dysfunction.
there is failure to sense the intrinsic ventricular event (R), followed by a Answer 1 is incorrect because raising the sensitivity level reduces
pacing artifact, which at times occurs coincident with the T wave. Pacing the risk of T-wave oversensing. Answer 2 is incorrect because raising the
artifacts without capture represent functional failure to capture when sensitivity level reduces the risk of sensing far-field atrial events. Answer
the ventricle is refractory. When the pacing is attempted once sufficient 4 is incorrect because undersensing predisposes to overpacing in the
time elapses after an intrinsic ventricular event, ie, when the ventricular ventricle. Ventricular hysteresis is usually triggered by situations of in-
myocardium is no longer refractory, ventricular depolarization occurs. trinsic ventricular activity.
Friedman, Rott, Wokhlu, Asirvatham, Hayes 91
20
Compass report.
Friedman, Rott, Wokhlu, Asirvatham, Hayes 93
21
A 71-year-old female was implanted with a dual-chamber ICD 6 years ago due to
inducible monomorphic ventricular tachycardia and ischemic cardiomyopathy. A
routine remote transmission reveals the Medtronic Cardiac Compass report shown in
Figure 21.1.
Q
:
On the basis of this report, what events occurred
in late May 2010?
1. Multiple shocks
2. Increase in patient activity
3. AT/AF episodes with elevated ventricular rate
4. Increase in nonsustained episodes
94 A Case-Based Approach to Pacemakers, ICDs, and Cardiac Resynchronization
21
What is the most likely cause for both the day and night
Q
:
average ventricular rate remaining at 60 bpm from the
end of May 2010 to early September 2010?
1. Postoperative reprogramming failed to turn rate response on
2. Frequent ventricular arrhythmia
3. Device in persistent mode switch due to AT/AF
4. Magnet response
96 A Case-Based Approach to Pacemakers, ICDs, and Cardiac Resynchronization
21
Case 22
RESYNCHRONIZATION
98 A 77-year-old male sends an unscheduled remote transmission, as he believes
PACEMAKERS, ICDs, AND CARDIAC
,
he received a shock from his ICD while asleep. He first received a dual-chamber
pacemaker 11 years ago due to atrial fibrillation status post AV node ablation. Four
years later, the pacemaker was upgraded to a CRT-P device due to nonischemic
dilated cardiomyopathy, and then 1 year later the device was upgraded to a CRT-D
device due to ventricular dysfunction and heart failure. Over the past 4 years, he
developed worsening heart failure and ventricular tachycardia requiring ICD shocks,
necessitating medication adjustments and hospitalizations. In addition, his quality of
life diminished due to difficulty participating in his favorite pastimes of hunting and
fishing.
Friedman, Rott, Wokhlu, Asirvatham, Hayes 99
22
22
Figure 22.2 Cardiac Compass report.
Friedman, Rott, Wokhlu, Asirvatham, Hayes 101
22
Q
:
On the basis of the reports outlined in Figures 22.1 and
22.2, what happened to the patient in early June 2010?
1. Increase in percent pacing
2. Sudden increase in fluid volume
3. Increase in patient activity
4. Termination of AT/AF
102 A Case-Based Approach to Pacemakers, ICDs, and Cardiac Resynchronization
22
artery disease, congestive heart failure, aortic valve replacement, and chronic atrial
fibrillation. He has received appropriate shocks for presyncopal tachyarrhythmia.
Device settings:
• Tachy VF zone: 188 bpm
• Therapies: ATPx1 during charging; 35-J shock x 6
• Brady mode: VVIR
• Pacing rate: 60 to 120 bpm
• Mode switch: on.
• RV lead parameters: output 2.5 V at 0.4 ms; sensitivity 0.3 mV
• LV lead parameters: output 2.5 V at 0.4 ms; pace polarity: LV tip to LV ring
Friedman, Rott, Wokhlu, Asirvatham, Hayes 105
23
Figure 23.1 Ventricular
fibrillation episode EGM.
Q
:
Why did the device deliver a shock after ATP slowed
the ventricular rhythm (Figure 23.1)?
1. Oversensing of ventricular rhythm
2. Far-field sensing of atrial arrhythmia
3. All shocks are committed once charging begins
4. Ventricular fibrillation resumed just prior to the charge ending
106 A Case-Based Approach to Pacemakers, ICDs, and Cardiac Resynchronization
23
Device settings:
• VF zone: 185 bpm
• Therapies: one sequence of ATP followed by eight 41-J shocks
• VT zone: 150 bpm
• Therapies: 3 sequences of ATP followed by six 41-J shocks
• VT zone detection enhancements: on
• Initial: V rate > A rate
• AFib rate threshold: 170 bpm
• Stability: 20 ms AND Onset 9%
• Sustained rate duration: 3 minutes
Friedman, Rott, Wokhlu, Asirvatham, Hayes 109
24
24
Q
: Why was therapy delivered?
1. Rhythm monomorphic VT
2. Episode fell into VF zone with no detection enhancements
3. Sustained rate duration timed out
4. ATP treating atrial tachycardia
112 A Case-Based Approach to Pacemakers, ICDs, and Cardiac Resynchronization
24
original pacemaker was placed 10 years earlier, and over the years the patient had
become increasingly pacemaker-dependent and rarely displayed intrinsic rhythm
on the transtelephonic transmission. At the time of the transmission (Figure 25.1),
the patient was residing in a nursing home and assisted by a nursing aide. Vitals
immediately prior to the transmission documented a pulse rate of 65 bpm.
Device settings:
• Mode: VVIR
• Pacing rate: 65 to 130 bpm
• Amplitude: 2.5 V
• Pulse width: 0.5 ms
• Sensitivity: 2.8 mV
• VRP: 330 ms
Friedman, Rott, Wokhlu, Asirvatham, Hayes 115
25
Q
:
Given the apparent discrepancy with past transtelephonic
transmissions, what would be the most logical next step?
1. Have patient schedule an in-clinic visit to evaluate and reprogram sensitivity
2. Discuss with the nursing home aide and then repeat the transmission
3. Unless elective replacement indicators are present, ignore the transmission
4. Immediate ER assessment to avoid R-on-T-induced tachyarrhythmia
116 A Case-Based Approach to Pacemakers, ICDs, and Cardiac Resynchronization
25
Device settings:
• Mode: DDD
• Pacing rate: 60 to 110 bpm
• PAV delay: 175 ms
• SAV delay: 150 ms
• V output: 0.4 ms at 4.0 V
• A output: 0.4 ms at 4.0 V
• AV hysteresis options: off
The 12-lead EGM without magnet application shown in Figure 26.1 was obtained. Next,
a magnet is placed to force DOO pacing, and an additional tracing is obtained, shown
in Figure 26.2.
26
Q :
Based on your EGM diagnosis and the programmed
parameters, what could explain these findings?
1. Ventricular lead dislodgment
2. Connector pin not fully inserted in the header
3. Crosstalk in the absence of safety pacing
4. Ventricular avoidance pacing algorithm
120 A Case-Based Approach to Pacemakers, ICDs, and Cardiac Resynchronization
pacing artifact is not apparent on the corresponding surface EGM and Answer 1 is incorrect because with lead dislodgment there are still 26
the actual AR interval is significantly greater than the programmed AV ventricular pacing artifacts present and the presentation is failure to
interval. capture, not failure to output. Answer 3 is incorrect because crosstalk,
A chest x-ray was obtained (Figure 26.4). The close-up of the pulse or oversensing of atrial signal on the ventricular lead, results in failure
generator is seen, and on inspection of the header of the device, the to output in nonmagnet mode, but not in a nonsensing magnet mode.
ventricular lead is not completely inserted into the header. This is most Forced ventricular pacing artifacts would still be present. Similarly, an-
evident when the tips of the atrial and ventricular connector pins are swer 4 is incorrect because ventricular avoidance algorithms rely on sens-
compared. An invasive approach to secure the lead in the connector ing and, therefore, are not operative during magnet mode.
block is required in this situation.
dysfunction. The pacemaker was programmed to a DDDR pacing mode at the end of
the procedure with AV interval of 220 ms. Thresholds obtained during the implant
via the pacing systems analyzer with cables connected directly to the leads were
excellent: A = 0.5 ms and 0.7 V, impedance 574 ohms; V = 0.5 ms and 0.5 V, impedance
690 ohms.
Within an hour of the patient returning to her hospital room, the pacemaker service
was called because of the tracing obtained on telemetry shown in Figure 27.1.
Friedman, Rott, Wokhlu, Asirvatham, Hayes 123
27
Q
: What would be your next step?
1. Activate the PMT algorithm
2. Turn off “sudden-brady” response
3. Invasively revise the pacing system
4. Shorten the AV interval
124 A Case-Based Approach to Pacemakers, ICDs, and Cardiac Resynchronization
are seen as AS, or atrial sensed, events. The ( ) indicate that all the AS equal to or near the maximum tracking rate. A Sudden Brady Response 27
events occur in a refractory period. This is consistent with lead reversal, (answer 2) is incorrect because one would observe a sudden increase in
ie, the atrial lead has been plugged in the ventricular port and vice versa. paced rate (AP-VP), which is not present. Shortening the AV interval
System revision was required. (answer 4) might seem to be a possibility when reviewing the original
There are no noninvasive options that would correct this prob- tracing without markers, but the markers clearly identify a lead reversal,
lem. Activating the PMT algorithm (answer 1) assumes the problem is so that answer 4 is incorrect.
that of PMT. With PMT, the markers would demonstrate AS-VP at rates
Device settings:
• Mode: AAIR
• Pacing rate: 75 to 160 bpm
• Output: 2.2 V at 0.5 ms
• Sensitivity: 1 mV
• Refractory: 280 ms
• VRR: on
• Bipolar
• Atrial threshold: 1.4 V at 0.5 ms
Interrogation reveals the arrhythmia
logbook shown in Figure 28.1.
28
Q
:
labeled as tachyarrhythmias, far-field sensing, and no
alteration in the pacing rate?
1. Far-field events have occurred during the atrial refractory period
2. Tachycardia events are resetting timing cycle
3. Tachycardias are real
4. Initial atrial pace fails to capture and a safety pace is delivered in refractory
128 A Case-Based Approach to Pacemakers, ICDs, and Cardiac Resynchronization
28
Figure 28.3 Patient tracing (top: surface ECG, bottom: intracardiac electrogram) and key to markers.
Case 29
RESYNCHRONIZATION
130 A 73-year-old male returned for routine follow-up of a DDD device. Although he had
PACEMAKERS, ICDs, AND CARDIAC
,
29
29 3. Pacemaker syndrome
Rate drop (RDR) is a programmable feature that detects a predefined feeling light-headed, attributable to the pacemaker syndrome, followed
dramatic drop in heart rate and results in pacing at a significantly faster by the fast heart rate, attributable to the RDR.
rate for a defined period. The algorithm is used in patients with neu- During this episode the patient spontaneously noted that he was
rocardiogenic syncope with significant cardioinhibition. Rapid pacing experiencing the symptoms previously described, ie, racing heart and
may help to minimize the blood pressure fall that often accompanies the mild light-headedness.
cardioinhibitory response. (RDR is a trademark of Medtronic. Several No definite evidence suggests that there may be a loss of integrity
manufacturers have variations of this algorithm, eg, the Sudden Brady on the atrial lead, ie, repetitive impedance measurements were normal
Response [SBR] of Boston Scientific). and there is nothing to support a diagnosis or lead insulation on conduc-
In this case, without the benefit of EGMs for review an actual tor coil abnormality.
RDR episode, it is impossible to know if the RDR episodes were respon- The options would be to turn off the RDR algorithm or to ad-
sible for his symptoms and whether or not they were appropriate, ie, just the sensitivity on the atrial lead, ie, less sensitive, retest with the
were they initiated by a bradyarrhythmia that met RDR criteria. isometric maneuvers, and see if detection of noise could be avoided. In
The nurse specialist seeing the patient documented acceptable this patient it was possible to alter the atrial sensitivity to avoid noise
atrial and ventricular pacing thresholds with adequately programmed detection but still have adequate sensing of intrinsic atrial events. Once
pacing margin of safety (answer 1) and also performed isometric maneu- this was accomplished the patient had no further symptomatic episodes.
vers, and the tracing in Figure 29.3 was obtained. No spontaneous atrial or ventricular tachyarrhythmias had been docu-
Note that the noise artifact that occurred during isometrics result- mented (answer 4).
ed in the pacemaker falsely recognizing noise on the atrial lead as atrial The patient’s symptoms of palpitations could have been explained
activity, which was tracked, resulting in a rate that is significantly higher by appropriate RDR and awareness of faster pacing or native tachyar-
than the patient’s actual intrinsic rate. At the end of isometrics, noise rhythmias. Options would include turning the RDR off or reprogram-
artifact ceases and there is a relative bradycardia. This change in rate was ming the RDR criteria and observation.
enough to meet the RDR criteria, so that the pacing rate was increased, As noted in the initial description, the pacing mode was DDD;
leading to symptoms. This sequence is consistent with his symptoms of therefore, inappropriate rate-adaptive sensor response would be incor-
rect (answer 2).
Friedman, Rott, Wokhlu, Asirvatham, Hayes 133
29
Device settings:
• Mode: DDD
• Pacing rate: 60 to 170 bpm (185 is maximum programmable tracking
and sensor rate)
• PVARP: 250 ms
• Paced AVI = 170 ms; sensed AVI = 140 ms
• V output: 5.0 V and 0.5 ms
• A output: 5.0 V and 0.4 ms
Four months later the patient’s local cardiologist reported that the patient, who
was very active and a competitive tennis player, would develop sudden and marked
fatigue during heavy exertion. The cardiologist had reprogrammed the pacemaker to
a maximum tracking rate of 185 bpm, the highest available on the device, but the
patient remained symptomatic.
Friedman, Rott, Wokhlu, Asirvatham, Hayes 135
The patient returned and a treadmill exercise assessment was performed. The patient 30
suddenly noted his typical exercise limiting symptoms at 9 minutes, 32 seconds and
exercised to maximum duration of 9 minutes, 47 seconds. The peak exercise tracing is
shown in Figure 30.1.
Figure 30.1
Patient’s tracing
at peak exercise.
Q
:
Given the patient’s quality of life impairment and the
exercise ECG findings, what is your next step?
1. Program the rate-adaptive sensor to a more sensitive threshold setting
2. Replace the device with one that has a higher programmable maximum tracking rate
3. Shorten the AV interval and PVARP to yield shorter TARP
4. Restrict patient’s activity level
136 A Case-Based Approach to Pacemakers, ICDs, and Cardiac Resynchronization
Shortening the TARP, answer 3, would not make a difference even The only viable option for this patient was to replace the device 30
if the programming option exists because the maximum achievable track- with one capable of achieving higher maximum tracking rates. The new
ing rate in this device is 185 bpm. pulse generator was programmed to its maximum programmable value
In a patient who is physically active and able to achieve high ven- of 200 bpm. At this setting the patient was able to complete competitive
tricular rates that are appropriate for the level of activity, and in whom tennis games without becoming symptomatic.
there is no contraindication for strenuous activity, eg, a patient with hy-
pertrophic cardiomyopathy and a history of sudden death, it is difficult
to justify or enforce lifelong limitation of a desired activity (answer 4).
Case 31
RESYNCHRONIZATION
138 An elderly male receives a dual-chamber pacemaker for intermittent AV block.
PACEMAKERS, ICDs, AND CARDIAC
,
The tracings shown in Figure 31.1 were obtained the day following the pacemaker
implantation.
31
Figure 31.1 Patient tracings, day 1 (upper tracing without Figure 31.2 Three-month follow-up magnet tracing.
magnet; lower tracing with magnet application).
31
block and symptomatic bradycardia. At the time she presented for consideration of
pacing, an ambulatory monitor demonstrated a predominant underlying rhythm of
atrial flutter with intermittent normal sinus rhythm. The patient recorded symptoms
of “light-headedness” and “feeling like I could pass out,” at which times tracings
demonstrated atrial flutter with ventricular response rates of 30 to 38 bpm. A
pacemaker was implanted.
The patient now presents to your institution without her pacemaker identification
card or any other information regarding the device or current programming. The
tracing shown in Figure 32.1 was obtained in the device clinic.
Without the benefit of the programmed parameters, and by analyzing only this single
tracing, you need to attempt to determine the underlying rhythm, the programmed
pacing mode, and if the tracing represents normal or abnormal device function.
Friedman, Rott, Wokhlu, Asirvatham, Hayes 143
32
Q
: Based on your ECG diagnosis, what is your next step?
1. Make ventricular channel more sensitive
2. Increase ventricular output
3. Lengthen ventricular blanking period
4. Do nothing
144 A Case-Based Approach to Pacemakers, ICDs, and Cardiac Resynchronization
32
33
A 76-year-old pacemaker-dependent man underwent CRT-D placement during the
day. You get a call at 2 am because his nurse is concerned about repeated failures
to capture. You interrogate the Medtronic D224TRK CRT-D pacemaker. The device
is programmed to DDDR 60 to 120 bpm. Both left ventricular and right ventricular
lead thresholds and impedances are normal, and provocative arm maneuvers do not
demonstrate system malfunction. With the left ventricular pacing configuration of
LV tip to LV ring, the threshold is 2 V at 0.5 ms. Because it is a new implant, the LV
output was programmed empirically to 5 V at 0.4 ms. Sensed AV delay is 100 ms and
programmed AV delay is 130 ms. Figure 33.1 demonstrates telemetry strips of concern.
in repeated or aborted attempts at automated capture, which could re- branch block configuration is not anodal stimulation. It represents sup- 33
sult in a frequent sense of skipped beats. port pacing that is part of a capture management algorithm. Answer 3
Answer 1 is incorrect because the presence of pacing spikes in the is incorrect. There are no data presented to suggested an elevated LV
absence of capture is not failure to output or inhibition of pacing as threshold or risk of inability to capture the ventricle; the noncaptured
might occur during oversensing on V channel due to crosstalk from the beats are part of the automated threshold test. Importantly in the ab-
atrial channel. During crosstalk, either there would be no ventricular sence of LV pacing, there would still be right ventricular pacing if the
output (and hence, no spike), or a safety pacing spike (which occurs 110 RV lead is functional, so that urgent temporary pacemaker placement
ms after the atrial spike). Answer 2 is incorrect because the left bundle would not be needed.
34
Q
: Figure 34.1 is compatible with which of the following?
1. Ventricular oversensing
2. An algorithm to minimize ventricular pacing
3. Dynamic AV delay
4. Safety pacing due to crosstalk
152 A Case-Based Approach to Pacemakers, ICDs, and Cardiac Resynchronization
34
35
Q
:
The programmed paced AV interval is 240 ms. Where does
the labeled QRS complex (*) in Figure 35.1 occur?
1. PVARP
2. Crosstalk sensing window
3. Atrial blanking period
4. Ventricular blanking period
156 A Case-Based Approach to Pacemakers, ICDs, and Cardiac Resynchronization
the timing of the QRS that results from its relation to the atrial pacing result in an absolute atrial blanking period preventing sensing of the 35
artifact. An annotated version of the tracing in Figure 35.2 demonstrates afterdepolarization of the pacing stimulus. In some devices, even sensed
2 sinus complexes with intrinsic ventricular conduction. Next, there is atrial events may initiate an atrial blanking period. Answer 4 is incor-
an atrial pacing artifact. Because of baseline noise in the tracing, it is rect; during the ventricular blanking period, nothing can be sensed on
difficult to tell if this pacing artifact is delivered as a result of atrial the ventricular channel.
undersensing or if it is occurring simultaneously with an intrinsic atrial
event. An intrinsic QRS follows, and it occurs during a ventricular pe-
riod known as the crosstalk sensing window (Figure 35.3). The typical
response to sensing during this time period is ventricular safety pacing,
which is readily identified here by the foreshortened atrioventricular in-
terval (100 to 120 ms).
This question assesses your knowledge of event timing on the ven-
tricular channel. On the ventricular channel, a ventricular sensed or
paced event creates a ventricular refractory period. On the ventricular
channel, after an atrial paced event, there is a cross-channel ventricular
blanking period, which is usually followed by a crosstalk safety pacing
window and an alert period. Signals sensed during the crosstalk sensing
window (usually <120 ms from the atrial pacing output) are considered
nonphysiologic due to the close coupling interval and may be caused by
atrial pacing afterpotentials. Additional events sensed in the crosstalk
sensing window may be noise, premature ventricular complexes, or ven-
tricular complexes resulting from undersensed atrial events.
Answer 1, PVARP, is incorrect; this represents the period of time
after a paced or sensed ventricular event during which atrial events are
refractory sensed events and do not affect the timing of events. It is re-
lated to sensing of events on the A channel, rather than the V channel.
Answer 3, atrial blanking period, is incorrect. Again, it reflects a time
window on the atrial channel. On the atrial channel, a paced event may Figure 35.3 The crosstalk sensing window.
Case 36
RESYNCHRONIZATION
158 A patient with a Medtronic single-chamber defibrillator receives a shock, and you
PACEMAKERS, ICDs, AND CARDIAC
,
36
Q
:
What is the best explanation for the cause of the shock
delivery after arrhythmia termination?
1. Prolonged rapid atrial fibrillation
2. Lead fracture
3. SVT-VT discrimination algorithm error
4. Oversensing during reconfirmation
160 A Case-Based Approach to Pacemakers, ICDs, and Cardiac Resynchronization
36
Figure 37.1 Tracing obtained during implant testing. From top to bottom are shown
surface, marker channels, and near-field ventricular channel.
Friedman, Rott, Wokhlu, Asirvatham, Hayes 163
37
Q
:
Based on this tracing, which of the following is most
accurate?
1. The rhythm is normal and artifact is present
2. Underdetection of ventricular fibrillation is present
3. The defibrillation threshold is elevated
4. The ICD is functioning properly
164 A Case-Based Approach to Pacemakers, ICDs, and Cardiac Resynchronization
37
Figure 37.2 Tracing obtained during implant testing. From top to bottom are shown surface, marker channels, and near-field ventricular
channel. Blue circles indicate inappropriate detection in fast VT zone instead of VF zone. The rectangle shows inappropriate delivery of
antitachycardia pacing (TP) during polymorphic VT.
166 A Case-Based Approach to Pacemakers, ICDs, and Cardiac Resynchronization
37 Figure 37.3 magnifies the EGMs on the bottom panel of the trac- We have no information about the defibrillation threshold, which
ing. Despite continued easily visible far-field ventricular EGMs, the is the amount of shock energy needed to terminate VF by the device, as a
near-field EGM amplitude decays and is undersensed. This suggests a device shock was not delivered (so answer 3 is incorrect). This is not arti-
different lead position, placing a new electrode, or using a different sens- fact (making answer 1 incorrect), and device function is not functioning
ing configuration (if available) is preferable to programming to increas- properly (answer 4), since VF was not properly detected!
ing sensitivity. Generally, VF is not undersensed when the sinus rhythm
R wave is greater than 5 to 7 mV.
Friedman, Rott, Wokhlu, Asirvatham, Hayes 167
37
Figure 37.3 Magnification of the EGMs in the bottom panels. The arrows indicate the near-field EGM (sharp signal). Note that it is getting
progressively smaller, until it is undersensed (asterisk), although the far-field EGMs are unchanged. This suggests the EGM at the lead
site is not suitable for sensing VF. Continued undersensing results in delivery of a pacing impulse at the +. Despite the fact that a pacing
impulse is delivered, a VS marker is placed. That results since the ventricular paced event at + is actually a safety pace event (note that it is
immediately preceded by an AP event). This indicates that the ventricular EGM actually was sensed, but during the safety pace window that
follows every atrial paced event.
Case 38
RESYNCHRONIZATION
168 The following fluoroscopic images were obtained at the the time of ICD implant.
PACEMAKERS, ICDs, AND CARDIAC
,
Friedman, Rott, Wokhlu, Asirvatham, Hayes 169
38
Figure 38.1 Left anterior oblique (LAO) and right anterior oblique (RAO) images of lead positions.
Q
:
Where is the defibrillator lead (arrow) shown in
Figure 38.1?
1. In the right ventricle
2. Dislodged from the ventricle to the atrium
3. In the coronary sinus
4. In the azygous vein
170 A Case-Based Approach to Pacemakers, ICDs, and Cardiac Resynchronization
38
Tachy parameters:
• VT zone: 140 bpm (430 ms); initial detection: 16 intervals; therapies: ATP x 9,
35-J CV x 3
• FVT zone via VT: 167 bpm (360 ms); therapies: ATP x 6, 5-J CV x 1, 35-J CV x 3
• VF zone: 188 bpm (320 ms); initial detection: 18/24 intervals; therapies: 35-J
shock x 6
• PR logic: AFib/Aflutter and sinus tach: on; other 1:1 SVTs off, SVT limit 290 ms
• Stability: 40 ms; 1:1 VT-ST boundary: 50%; high rate timeout: off
Brady parameters:
• Mode: DDIR
• Pacing rate: 60 to 85 bpm
Friedman, Rott, Wokhlu, Asirvatham, Hayes 173
39
Figure 39.1 Patient’s remote transmission of nonsustained episode interval plot. Legend: V-V: interval (ms) from one
ventricular event to the next. A-A = interval (ms) from one atrial event to the next; FVT = fast ventricular tachycardia zone
interval; TS = interval in VT zone; VF = ventricular fibrillation zone interval; VS = V-V interval below detection zones; VT =
ventricular tachycardia zone interval. If the V-V interval measures 430 ms, the event will not count towards VT detection;
the V-V interval must be <430 ms.
174 A Case-Based Approach to Pacemakers, ICDs, and Cardiac Resynchronization
39
B
Friedman, Rott, Wokhlu, Asirvatham, Hayes 175
39
Q
:
The episode lasted 20 seconds. Why was therapy
not delivered?
1. The rhythm was not ventricular tachycardia or fibrillation
2. Initial detection was not met due to the stability detection enhancement
3. High rate timeout was programmed off
4. Therapy was not programmed as on in the VT zone
176 A Case-Based Approach to Pacemakers, ICDs, and Cardiac Resynchronization
counter from reaching the required 16 consecutive events to declare a timeout forces therapy to be delivered after a VT has been detected and 39
VT episode. To treat this VT, the programmed stability value should be sustained for the duration timer. Since initial detection was not met,
increased, or stability turned off. the high rate timeout was never started, so answer 3 is incorrect. In this
Stability is nominally set for 30 msec. However, in the setting of case it would not have timed out and allowed therapy, as it was not pro-
antiarrhythmic drugs, there may be greater variability in R-R intervals grammed on. Therapy was indeed programmed as on in the VT zone,
during VT, leading to undertreatment (as in the present case) with the so answer 4 is incorrect.
use of stability. Looking at the episode plot, it appears the VT hovered right below
Answer 1 is incorrect, since the rhythm is ventricular tachycardia the VT zone lower rate limit for 15 seconds before the EGM recording be-
with A:V dissociation with the atrial median cycle length of 800 ms gan. Besides reprogramming the stability interval, the VT zone detection
and a much faster ventricular median cycle length of 390 ms. High rate rate should be lowered if delivering therapy to this slower VT is desired.
Case 40
RESYNCHRONIZATION
178
PACEMAKERS, ICDs, AND CARDIAC
,
Figure 40.1 Tracing recorded during a tachycardia. From top to bottom are shown
the atrial channel, ventricular channel, and markers.
Friedman, Rott, Wokhlu, Asirvatham, Hayes 179
40
Q
:
In the tracing shown in Figure 40.1, which of the following
is true?
1. Sinus tachycardia is appropriately rejected
2. Sinus tachycardia is inappropriately shocked
3. Ventricular tachycardia is inappropriately rejected
4. Ventricular tachycardia is appropriately shocked
180 A Case-Based Approach to Pacemakers, ICDs, and Cardiac Resynchronization
40
Figure 40.2 Sinus tachycardia appropriately rejected by the ICD. The vertical arrows indicate a PAC that is tracked by the ventricle. The horizontal arrows
are the same length, showing that the ventricular complex marked by the asterisk came earlier than expected, making this a premature ventricular (or rarely,
junctional) complex. The morphology is unchanged with the early ventricular complex (note score of 100, indicating perfect match with baseline template),
but morphology discrimination using the near-field may fail to distinguish SVT and VT in up to 10% of arrhythmias. The third from last complex has a lower
match score (76) and a PR interval shorter than other complexes, suggesting another PVC, possibly fused with the sinus complex.
Case 41
RESYNCHRONIZATION
182
PACEMAKERS, ICDs, AND CARDIAC
,
Figure 41.1 Morphology analysis. Adapted with permission from Swerdlow CD,
Friedman PA. Advanced ICD troubleshooting: part I. PACE. 2005;28:1322-1346.
Friedman, Rott, Wokhlu, Asirvatham, Hayes 183
41
The morphology analysis during device interrogation is shown in Figure 41.1
for a patient with a St. Jude Atlas dual-chamber defibrillator who has a history
of atrial arrhythmias.
Q
:
What is the best explanation for the findings on the strip
in the right panel of Figure 41.1?
1. Alignment error in the setting of intermittent pacing
2. Rate-related bundle branch aberrancy
3. Morphology mismatch by using paced beats as a morphology template
4. Failure to identify template match during ventricular tachycardia
184 A Case-Based Approach to Pacemakers, ICDs, and Cardiac Resynchronization
41
at the beginning of the ventricular EGM. The differences in the “attack ment error and may be preferable. In patients with rate-related bundle 41
rates” of the sensitivity following paced and sensed ventricular events branch block, pacing at faster heart rates (ie, 120 bpm in an AAI mode)
results in detecting the beginning of the QRS using the small predeflec- during template acquisition and disabling the automatic template fea-
tion in one case, and not in the other. This leads to different starting ture permit use of the morphology SVT-VT discriminator.
points and thus alignments for the EGMs following paced complexes. Answer 2 is incorrect because there is no evidence of rate-related
The consequence of 2 different starting points, in this case, is a shift bundle branch aberrancy. There, wide beats are paced. The narrow beats
in the EGM and thus total malalignment (0% match) despite identical are intrinsically conducted. Answer 3 is incorrect because narrow beats
EGMs. from the intrinsic conduction system are the morphology templates. The
Because the alignment error is observed only following paced com- paced beats do not impact the template; they impact the sensing of the
plexes, it has uncertain clinical significance during sensed tachycardia. QRS of interest. Answer 4 is incorrect because there is no evidence of
However, reprogramming sensitivity settings may eliminate the align- ongoing tachycardia.
Device settings:
Tachy parameters:
• VT zone: 185 bpm; detection duration: 2.5 s; detection enhancements: off
• Therapies: ramp ATP x 1 sequence of 8 pulses at 88%, 41-J shock x 5
• VF zone: 205 bpm; detection duration: 1.0 s
• Therapies: 41-J shock x 8
189
42
Brady parameters:
• Mode: DDDR
• Pacing rate: 50 to 120 bpm
• Ventricular pacing: RV + LV.
• LV pace and sense configuration: LV tip to coil
Device interrogation revealed the EGM and daily measurement trends shown in Figures
42.1, 42.2, and 42.3.
190 A Case-Based Approach to Pacemakers, ICDs, and Cardiac Resynchronization
42
42
Figure 42.2 Daily measurements data table.
192 A Case-Based Approach to Pacemakers, ICDs, and Cardiac Resynchronization
42
42
Q
:
impedances in late May 2010? (Of note, pacing thresholds
unchanged from January to September 2010.)
1. RV lead failure
2. Loose set screw
3. Exposure to electromagnetic interference
4. Excessive fibrosis
194 A Case-Based Approach to Pacemakers, ICDs, and Cardiac Resynchronization
42 1. RV lead failure
The RV defibrillation lead had failed. As a result of the LV lead utiliz- noise deflections are seen briefly on the far-field signal—suspicious for
ing the RV lead to both pace and sense, impedance increases were seen a process affecting several conductors, eg, lead damage or contact with
on both the RV and LV pacing leads. Pocket manipulation also repro- an abandoned lead. This series of observations raises the possibility that
duced noise on the shock (far-field) EGM. There was question as to why the patient was moving her arm (leading to myopotentials on the far-
the shock EGM also showed noise, especially since the shock imped- field EGM, which includes the pulse generator as electrode), and that the
ance had not changed. The low-amplitude noise on the far-field (shock) mechanical motion subsequently triggered the make-break potentials on
EGM that begins a few cycles before the noise on the near-field (RV the near-field and far-field EGMs.
pace) EGM most likely reflects myopotentials. In addition, some large
195
The RV lead was removed and a new lead was implanted. The consistent values, a make-break connection was not thought to be the 42
failure was believed to be in the RV ring electrode, since both the RV problem, ie, answer 2 is incorrect. Whereas exposure to electromagnetic
and LV pacing utilized this electrode, and the RV pace (near-field) EGM interference (answer 3) may result in nonsustained episodes, it would
consistently demonstrated noise. not cause the impedance changes that were then maintained. Answer 4
Chest x-ray demonstrated that all pins were fully inserted into the is incorrect, for excessive fibrosis at the lead or myocardial interface is
header of the pulse generator, and verified with fluoroscopy prior to the the cause of exit block, which presents as higher pacing thresholds, not
lead revision. As the impedances rose suddenly but then plateaued at a change in impedance.
Case 43
RESYNCHRONIZATION
196 A 68-year-old male received dual-chamber pacemaker sinus node dysfunction
PACEMAKERS, ICDs, AND CARDIAC
,
with syncope.
The transtelephonic pacemaker tracing shown in Figure 43.1 was obtained 3 weeks
postimplant.
Device settings:
• Mode: AAIRóDDDR
• Pacing rate: 60 to 130 bpm
• Paced AV delay: 200 ms
• Sensed AV delay: 170 ms
Friedman, Rott, Wokhlu, Asirvatham, Hayes 197
43
Figure 43.1 Panel A:
Nonmagnet TTM rhythm
strip.
Q
:
Why are there variations in the paced and sensed
AV delays?
1. AV search hysteresis
2. Magnet application
3. Intermittent ventricular failure to capture
4. Ventricular undersensing while in Managed Ventricular Pacing (MVP) mode
198 A Case-Based Approach to Pacemakers, ICDs, and Cardiac Resynchronization
43
43 Answer 1 is incorrect, as the device goes from immediately sensing ufacturer. However, answer 2 is incorrect for this Medtronic EnRythm
the intrinsic QRS in events 1, 3 to 5, and 7 to a shortened AV delay in P1501DR device would respond to magnet application by AV pacing
events 2, 6, and 8. Then, once the consistent pattern of AV-sequential at 85 bpm with the programmed AV delay of 200 ms, and this rhythm
pacing with the programmed paced AV delay of 200 ms begins in events strips shows AV pacing at the sensor-driven rate of 68 bpm.
9 to 13, the AV delay remains at the programmed AV delay and does not Answer 3 is incorrect, as each event shows either atrial pacing with
extend out to search for the intrinsic QRS complex to promote ventricu- intrinsic conduction of the QRS complex, or AV pacing with ventricular
lar sensing. capture or fusion of the ventricular pacing output and intrinsic QRS
Transtelephonic rhythm strips usually do show varying AV delays complex.
due to magnet application, dependent upon the specific model and man-
201
43
Figure 43.3 Panel A: MVP mode: Ventricular pacing only as needed in the presence of transient loss of
conduction. Reproduced with permission of Medtronic, Inc. Panel B: Continued MVP mode: Ventricular
support if loss of AV conduction is persistent. Reproduced with permission of Medtronic, Inc.
Case 44
RESYNCHRONIZATION
202 A 53-year-old man with a history of atrial fibrillation and a favorable response to
PACEMAKERS, ICDs, AND CARDIAC
,
44
44
44
Q
:
What is the likely reason for a new failure to respond
to cardiac resynchronization therapy?
1. Poor LV lead site selection at the time of original implant
2. Insufficient LV-RV offset to account for postsurgical slow exit from the LV pacing site
3. The RV lead is dislodged into the coronary sinus
4. RV anodal capture is occurring due to LV lead dislodgment
206 A Case-Based Approach to Pacemakers, ICDs, and Cardiac Resynchronization
density, increasing the likelihood of capturing the surrounding RV tis- suggests a very apical position. This is confirmed by the chest x-ray. The 44
sue. In addition, anodal stimulation may be more prominent if LV pac- postrevision chest x-ray portable PA view is shown in Figure 44.4. The
ing occurs significantly later than a paced or sensed RV event. LV lead is now lateral and leftward.
Note that if an LV tip-to-ring configuration is used, both anode Answer 1 is incorrect because preoperatively the pacing vector was
and cathode are LV leads so that anodal stimulation is not clinically consistent with LV capture (see Figure 44.1) and the patient responded
important. When pacing in an LV-to-RV configuration, if both cathodal to therapy. Answer 2 is incorrect because although poor offset can in-
and anodal stimulation occur, it is generally of no clinical consequence. crease the RV pacing component of the vector, there was no previous
The main concern is when anodal stimulation occurs alone (without mention of LV exit block, and failure of LV pacing was due to frank lead
capture at the cathode, or LV site). Recognition of isolated anodal stimu- dislodgement into the RV.
lation often requires a 12-lead ECG to assess pacing morphology. Answer 3 is incorrect because the RV lead is in place, as confirmed
In this case, the LV pacing lead was dislodged into the right ven- by the x-ray and the RV paced QRS morphology. The LV lead is dis-
tricle. It is possible that there was direct capture of the RV tissue from lodged into the right ventricle.
the LV lead tip. However, a compelling argument for anodal capture is
that despite significant separation of the RV and displaced LV lead tips
in the right ventricle (as seen in the chest x-ray), the 12-lead pacing con-
figuration is identical in RV and LV pacing. Also, the observation that
the capture only occurred in one configuration (LV tip to RV coil) is
consistent with anodal capture, as direct capture from the LV lead (the
cathode) was never demonstrated.
This case is a good illustration of the importance of careful post-
surgical evaluation of device function, particularly the LV lead.
Postrevision RV pacing, LV pacing, and biventricular pacing vec-
tors are now seen in Figure 44.3. Note that lead I is now negative with
LV pacing and has a hybrid appearance with biventricular pacing. Also
note that LV pacing postrevision results in an RBBB pattern (very posi-
tive QRS in V1) as opposed to the biphasic pattern with RV pacing. The
fact that RV pacing is biphasic in V1 (as opposed to negative/LBBB)
208 A Case-Based Approach to Pacemakers, ICDs, and Cardiac Resynchronization
44
44
dual-chamber pacemaker 3 months after pulse generator change. His pacing leads
have been in place for 9 years. The initial indication for pacemaker implant was
postoperative high-grade AV block with syncope.
Device settings:
• Mode: DDDR
• Pacing rate: 60 to 130 bpm
• Mode switch: on
• Atrial lead: output 2.5 V at 0.3 ms; sensitivity: 0.1 mV;
AutoSense: off
• Ventricular lead: output 1.0 V at 0.4 ms (AutoCapture: on);
sensitivity: 2.5 mV; AutoSense: off
211
45
At the time of the device clinic visit the patient was in atrial fibrillation and current
interrogation reveals a P wave of 0.7 mV and impedance of 390 ohms. Prior atrial lead
measurements were:
• Initial implant (9 years prior): P wave 2.1 mV in sinus rhythm, 529 ohms
• Pregenerator change via device: P wave 0.5 mV, impedance 426 ohms
• During generator change lead testing via analyzer: P wave 0.3 mV, impedance
469 ohms
• Immediately postgenerator change via device: P wave 0.4 mV, impedance
350 ohms
45
45
Q
:
What is the best explanation for these 255 atrial noise
reversion episodes?
1. Atrial lead sensing failure after pack change
2. Atrial lead pin not fully inserted into header of new device
3. Frequent exposure to electromagnetic interference
4. Oversaturation of the atrial sensing amplifier
214 A Case-Based Approach to Pacemakers, ICDs, and Cardiac Resynchronization
events in the atrium are ignored during the episode. In the ventricle, into the header would most likely result in a variable impedance. Inter- 45
these algorithms are protective against oversensing of noise and inap- rogation revealed stable lead measurements and trending since the pulse
propriate pauses. Once a ventricular noise reversion event is triggered, generator change. Answer 3 is incorrect because frequent exposure to
there is “reversion” to an asynchronous ventricular pacing mode, such EMI would most likely result in noise on both leads and only atrial noise
as DOO or VOO. If noise reversion occurs in an ICD, ventricular noise reversion is occurring and stored EGMs fail to demonstrate any noise on
reversion would disable the charge for VT/VF therapy, and detection the ventricular EGM.
enhancements during atrial noise reversion are classified as V>A so VT/
VF therapy would not be withheld.
Answer 1 is incorrect because the atrial sensing threshold has actu-
ally improved slightly since pack change, from 0.3 to 0.4 mV to 0.7 mV.
Answer 2 is incorrect, because a lead pin that was not fully inserted
Appendix 217
Spoiler alert! This appendix identifies the cases in this book by diagnosis, which may suggest or reveal the answers to questions
in the cases. Because we want to encourage readers to approach the cases as unknowns, we are presenting this list as an
appendix rather than as a table of contents and we recommend that you use the appendix only after working through the cases.
1 Diathermy, artifact, oversensing, electromagnetic interference, 2 14 Diagnosis and management of intermittent failure to capture, 56
2 Pacing artifacts, temporary pacemakers, device-device 15 Atrial lead dislodgement, failure to sense, failure to capture,
interaction, 6 programming ATP in single-chamber (atrial only) device, 60
3 Magnet response, magnet mode, 10 16 Inappropriate shocks, lead dislodgement, 64
4 Artifact on surface ECG, 14 17 P-wave oversensing, V lead dislodgement, 70
5 Threshold test and digital surface ECG missing spikes, 18 18 Mode switch surface ECG, pacemaker wide complex
6 Inadvertent LV pacing via patent foramen ovale, normal and tachycardia analysis, 74
abnormal ECG and chest x-ray following pacemaker implant, 22 19 Lead reversal in header,
7 Far-field oversensing, P-wave oversensing, 26 A-V reversal, 78
8 Far-field oversensing, R-wave oversensing, inappropriate 20 Hysteresis, ventricular undersensing, differential diagnosis
mode switch, 30 of pacing at rates below the lower rate limit, 82
9 Pacemaker mediated tachycardia, 34 21 Analysis of diagnostics in CRT device, 92
10 Approach to high rate episode, lead noise, EMI, 38 22 Analysis of diagnostics in CRT device, 98
11 Pacemaker Wenckebach, upper rate behavior, 42 23 Reconfirmation, committed vs. noncommitted shock
12 Surface ECG recognition of fusion, pseudofusion, delivery, 104
and capture, 46 24 SVT-VT discrimination enhancements, shock for sinus
13 Pacemaker timing cycles and blanking periods, distinguishing tachycardia, sustained rate duration, inappropriate shock, 108
pseudofusion from pseudo-pseudofusion, 50 25 Complex transtelephonic monitoring tracing analysis, 114
218 Appendix
26 Ventricular failure to output, 118 36 T-wave oversensing, reconfirmation, inappropriate shock, 158
27 Lead reversal, 122 37 Undersensing of VF by an ICD, 162
28 Far-field R-wave oversensing, AAI pacing mode, 126 38 Management of high defibrillation threshold, recognition
29 Rate-Drop Response, assessment of pacemaker pacing faster of alterative defibrillation lead positions (including coronary
than expected, 130 sinus and azygous vein) by chest x-ray, 168
30 Upper rate behavior, programming to treat chronotropic 39 SVT-VT discriminators (stability), VT underdetection, 172
insufficiency, 134 40 SVT-VT discriminators (onset and morphology), 178
31 Atrial lead dislodgment, 138 41 SVT-VT descriminators (morphology), 182
32 AutoCapture, unexpected pacing artifacts on surface ECG, 42 Differential diagnosis of lead noise, lead failure, 188
algorithmic pacing, 142 43 Managed ventricular pacing, algorithmic pacing, 196
33 LV capture management, algorithmic pacing, 146 44 ECG and chest x-ray analysis of LV lead position, anodal
34 Managed ventricular pacing, algorithmic pacing, 150 stimulation, 202
35 Ventricular safety pacing, atrial undersensing, crosstalk window, 45 Noise reversion, atrial sensing oversaturation, 210
blanking or refractory intervals, pacemaker timing cycles, 154
Index 219
AAIR pacing mode, 60–63, 126–28, automated capture management algorithms, cardiomyopathy
150–53, 198–99 146–49 dilated, 26, 98, 172
AAISafeR algorithm, 152 Automatic Sensitivity Control, 186 ischemic, 72–73, 92–96
AF (atrial fibrillation), 94, 210–15 automatic switched mode (AMS), 32 chest x-ray
algorithm minimizing ventricular pacing, AV block atrial lead dislodgment, 62
150–53 intermittent, 138–41 determining lead position, 24–25, 170–71
amiodarone, 94 pacemaker for, 10–13, 22–25, 34–37 loose connector pin, 121
anodal capture, 202–9 postoperative, 42 postoperative, 204, 207, 209
antitachycardia pacing (ATP), 4, 104–7 AV interval hysteresis, 86–87 connector pins, incomplete insertion,
AT (atrial tachycardia), 94 AV node conduction, amiodarone and, 94 122–23
atrial arrhythmias, 60–63 AV synchrony, 37 crosstalk, 154–57, 156
atrial contraction, premature, 36 axillary vein, 58 CRT. See cardiac resynchronization therapy
atrial far-field events, 126–28 azygous vein, 168–71 (CRT)
atrial fibrillation (AF), 94, 210–15
atrial flutter, 16, 142–45 bundle branch block, right, 22–25 DDD pacing mode, 2–5, 42–45, 112–13,
atrial lead 118–23, 130–33, 134–37
dislodgment, 62–63, 138–41 cardiac resynchronization therapy (CRT) DDDR pacing mode, 14–17, 26–29, 30–33,
far-field R-wave oversensing, 30–33 automated capture management 34–37, 50–55, 122–25, 146–49,
reversal, 80–81, 124–25 algorithms, 146–49 188–95, 198–99, 214–15
atrial refractory period, 36–37, 214 fluid volume measurements, 98–103 DDIR pacing mode, 172–77
atrial sensing amplifier, 210–15 LV lead dislodgment and RV anodal diathermy, 4
atrial tachycardia (AT), 94 capture, 202–9 dilated cardiomyopathy, 26, 98, 172
atrioventricular terms. See AV terms right ventricular lead failure, 188–95
autointrinsic conduction search, 152 VF resumption, 104–7
220 Index
EGM. See electrogram (EGM) heart block, complete, 56–59, 134–37 loss of programmer communication, 12
electrocardiogram (ECG) LVAD (left ventricular assist device), 102–3
algorithm minimizing ventricular implantable cardioverter-defibrillator (ICD)
pacing, 150–53 appropriate rejection of sinus magnet response, 12
anodal capture, 202–9 tachycardia, 178–81 minimized ventricular pacing (MVP)
atrial flutter, 142–45 AT/AF episodes, 92–94 avoidance algorithms
QRS complex in crosstalk sensing DDD pacing mode, 108–13 AAIR to DDDR mode switching, 198–99
window, 154–57 defibrillator lead in azygous vein, 168–71 automated capture management
right bundle branch block, 22–25 electromagnetic interference, 2–5 timing, 152–53
sinus tachycardia vs. VT, 178–81 increase in thoracic fluid volume, 98–103 ventricular undersensing, 196–201
ventricular failure to output, 118–23 lead dislodgment, 72–73 morphology algorithms, EGM, 182–87
electrogram (EGM) lead revision, 68–69
algorithms minimizing ventricular morphology algorithm error, 182–87 noise reversion algorithm, 214–15
pacing, 150–53 nonsustained episode, 172–77
electromagnetic interference, 2–5 oversensing during reconfirmation, OptiVol fluid index, 99, 102
lead reversal diagnosis, 80–81, 124–25 158–61 oversensing. See also T-wave oversensing
morphology analysis, 182–87 postoperative reprogramming failure, lead integrity loss, 4
near-field, 167 95–96 P-wave, 28
normal pacemaker function, 20–21 sustained rate duration timed out, 112–13 R-wave, far-field, 30–33
oversensing during reconfirmation, T-wave oversensing, 66–67, 158–61 ventricular, atrial events, 156–57
158–61 ventricular sensed markers, 26–29
stored, correlation with patient VF underdetection, 162–67 PAC (premature atrial contraction), 36
activities, 40–41 intrinsic beat, surface ECG, 47–48 paced beat, 47– 48
T-wave oversensing, 67, 158–61 ischemic cardiomyopathy, 72–73, 92–96 pacemaker syndrome, 130–33
underdetection of ventricular isometrics, noise artifacts, 132 pacemaker-dependent patients
fibrillation, 162–67 avoidance of welding, 40
ventricular failure to output, 118–23 leads discrepancies in transtelephonic
electromagnetic interference (EMI), 2–5, 40 dislodgment, 202–9 transmissions, 114–17
exertion, 136–37 failure, 4, 188–95 lightheadedness in, 56–59
placement, 24–25, 58 management after implant, 8
failure to capture, atrial, 62 reversal, 80–81, 124–25 pacemaker-mediated tachycardia (PMT),
far-field events, 126–28, 156 left ventricular assist device (LVAD), 102–3 36–37
far-field R wave (FFRW) oversensing, 30–33 Left Ventricular Capture Management, pacemakers
far-field sensing, atrial, 126–28 148, 149 biventricular, 206
Fontan procedure, 60–63 lightheadedness, 56–59, 130–33 connector pin not inserted in header,
fusion beat, 47–50, 76 118–23
Index 221
DDDR pacing mode, 10–13, 14–17, postventricular atrial refractory period stability detection enhancement,
30–33, 50–55, 78–81 (PVARP), 36–37 ICD, 172–77
dual-chamber premature atrial contraction (PAC), 36 Sudden Brady Response (SBR)
AAIR pacing mode, 60–63 premature ventricular contraction algorithm, 132
algorithms minimizing ventricular (PVC), 180 supraventricular tachycardia (SVT), 180
pacing, 150–53 programming head, loss of communication sustained rate duration, 112–13
atrial lead dislodgment, 138–41 with device, 12
atrial sensing amplifier oversaturation, pseudofusion beat, 47, 48, 49, 76 tachycardia. See also ventricular
214–15 pseudo-pseudofusion beat, 49, 54–55 tachycardia (VT)
far-field R-wave oversensing, 30–33 pseudo-Wenckebach behavior, 136–37 antitachycardia pacing (ATP), 4, 104–7
magnet response, 12 P-synchronous pacing, 138 atrial, 94
mode switch occurrences, 30–33 pulmonary edema, 102–3 pacemaker-mediated, 36–37
normal function, 74–77 pulse generator, 122–23 sinus vs. ventricular, 178–81
surface ECG vs. marker channel/ PVARP (postventricular atrial refractory supraventricular, 180
ventricular EGM, 14–17 period), 36–37 telemetry, loss of communication
normal device function, 18–21 PVC (premature ventricular with device, 12
permanent, preoperative evaluation, contraction), 180 thoracic impedance, 102–3
38–41 trace markers, far-field events, 128
single-chamber, 6–9 railroad tracks (tramtracking), 67 tramtracking (railroad tracks), 67
stored EGM correlation with patient rate duration, sustained, 112–13 transtelephonic monitoring, 42–45, 114–17
activities, 40–41 rate smoothing, sensor-driven, 136 transvenous pacing, 24
temporary, undersensing permanent Rate-Drop Response (RDR) algorithm, T-wave oversensing
pacing, 8 130–33 programming to avoid, 67, 68
transtelephonic monitoring, 42–45 R-wave, far-field oversensing, 30–33 during reconfirmation, 158–61
ventricular failure to output, 118–23 single-chamber ICD, 66–67
VVIR pacing mode, 6–9 SBR (Sudden Brady Response) ventricular lead dislodgment, 72–73
VVIR pacing with AutoCapture, 142–45 algorithm, 132 ventricular lead revision, 68–69
pacing, transvenous, 24 Search AV hysteresis, 152 ventricular sensitivity threshold, 82–91
Parkinson’s disease and atrial flutter, 16 sick sinus syndrome with syncope, 14–17
patients. See pacemaker-dependent patients sinus node dysfunction undersensing
PMT (pacemaker-mediated dual-chamber pacemaker, 74–77, 122–25 atrial, 62
tachycardia), 36–37 with syncope, 30–33 temporary pacemakers, 8
post-cardiac surgery ventricular undersensing in MVP ventricular, 90–91, 196–201
AT/AF episodes, 94 mode, 196–201 ventricular fibrillation, 162–67
device management, 8 sinus rhythm, 184–85 upper rate limit, programmed, 44
LV lead dislodgment, 202–9 sinus tachycardia, vs. VT ., 178–81
rate response reprogramming, 96
222 Index
ventricular blanking period, 52, 157 P wave oversensing, 28 sustained rate duration timed out,
ventricular complexes, surface ECG, 46–49 reversal, 80–81, 124–25 112–13
ventricular contraction, premature, 180 T-wave oversensing, 68–69 lead reversal, 80–81
ventricular failure to output, 118–23 ventricular pacing, 36, 150–53 sinus tachycardia vs., 178–81
ventricular fibrillation (VF) Ventricular Rate Regulation (VRR), ventricular threshold test, 20
nonsustained episodes, 188–95 73, 137–40 VF. See ventricular fibrillation (VF)
resumption prior to ATP confirmation, ventricular refractory period, 157 VT. See ventricular tachycardia (VT)
104–7 ventricular safety pacing, 156–57 VVI pacing mode, 82–91, 86–87, 90–91
undersensing by ICD, 162–67 ventricular sensed markers, 26–29 VVIR pacing mode, 6–9, 104–7,
ventricular lead ventricular tachycardia (VT) 114–17, 142–45
alternate venous routes, 58 implantable cardioverter-defibrillator
automatic capture algorithms, 148–49 dual-chamber, 2, 92–96 welding and EMI, 40
dislodgment, 72–73 stability detection enhancement, Wenckebach phenomenon, 44–45
failure, 188–95 176–77