ILase Handpiece Calibration Procedure

You might also like

Download as pdf or txt
Download as pdf or txt
You are on page 1of 2

Biolase Technology, Inc.

Document # Title Revision


- HANDPIECE UNIT CALIBRATION, iLase A
ECO: vvvvv Effective date: vvvvv Checked by:

1.0 PURPOSE

The purpose of this instruction is to document the Calibration Procedure for the iLase Handpiece unit.

2.0 SCOPE

N/A

3.0 ASSOCIATED DOCUMENTS

N/A

4.0 EQUIPMENT AND MATERIALS

N/A

5.0 CALIBRATION PROCEDURE

1. Set the Handpiece to any procedure layer 1

2. Enter calibration mode


Press the UP () button, and hold for 3 seconds until the
beeping sound is activated.
Press the UP () button, and hold for 3 seconds until the
beeping sound is activated.
Press the SELECT () button, and hold for 3 seconds until the
calibration appears on the display.
3. Press the SELECT () button until the <DAC1> is displayed on the
LCD.
Press the Finger Switch and read the power meter value.
Press + or- to adjust the number to match the power reading.
Press select enter key to advance to the DAC2 screen.

4. Repeat step 4 for DAC 2 to DAC 8.

5. Remove battery when screen shows DAC1 again.

6.0 END OF DOCUMENT

Page 1 of 2
FOR USE ONLY BY BIOLASE TECHNOLOGY, INC.
This document contains confidential and proprietary information.
It must not be reproduced or disclosed to others without prior written approval.
Biolase Technology, Inc.
Document # Title Revision
MI-XXX HANDPIECE UNIT CALIBRATION, iLase A
ECO: Effective date:

REVISION HISTORY
REVISION ECO # CHANGE SUMMARY
A Initial Release. Peter Pham

Page 2 of 2
FOR USE ONLY BY BIOLASE TECHNOLOGY, INC.
This document contains confidential and proprietary information.
It must not be reproduced or disclosed to others without prior written approval.