Discovery DSC Practical 2020 Part 1
Discovery DSC Practical 2020 Part 1
Calorimetry (DSC)
Practical Training Course
2020
Part 1
Calibration & Optimization & Trios
Els Verdonck
everdonck@tainstruments.com
1
Training Course Agenda (Part 1)
• Introduction
– Technique
– TzeroTM Technology vs. Heat flux DSC
• Experimental Method
Technique &
Instrumentation
3
DSC: The Technique
4
Definitions
5
Heat Flow in DSC
where:
dQ/dt = heat flow (W/g or mW)
W = J/sec
Cp = specific heat capacity (J/g.ºC)
Cp(T) = a + b.T + c.T2
dT/dt = heating rate (ºC/min)
f (t,T) = time dependent
(kinetic) components
e.g. melting, crystallisation, curing,...
6
DSC Thermogram
endothermic <- Heat Flow -> exothermic
Cold Oxidation
Cross-Linking
Crystallisation (Cure)
Glass
Transition
-> ΔCp
Temperature
7
How is Heat Flow Measured ?
DSC25xx with or without AS
8
Heat Flux DSC Instrument DSC 25
Note: schematics show Dynamic Sample Chamber
a 2920 heat flux DSC Reference Pan
Sample Pan
dQ/dt = Qs-Qr
Lid
dQ/dt = ∆T/R
Gas Purge Inlet ∆T = Ts - Tr
Assumes
symmetry
Chromel Chromel
Disc Disc between sample
and reference
Heating calorimeter
Block
T1
Thermocouple
Alumel Wire Junction
T0
DSC 2500 – DSC 250 – (DSC 25)
chromel T0
First generation Discovery: Ts
diffusion bonded sensors
Tr constantan
10
Heat Flow Equation T4
Tzero
∆T 1 1 dTs d∆T
q=− + ∆T0 − + (Cr − Cs ) − Cr
Rr Rs Rr dt dt
11
Baseline (= Empty Cell Run)
0.6
0.2
0.0
-0.2
-0.4
-100 0 100 200 300 400
Temperature (°C)
12
Baseline Curvature Hinders
Detection of Weak Transitions
Weak, broad Tg
Tg ???
13
Heat Flow Terms T4
Tzero
Tzero DSC’s
DSC 250
theory
15
Temperature / Heating Rate
16
Temperature / Heating Rate
T4P pan
sample
gas
sensor
17
Experimental Method
18
Heating/Cooling Method
• Heating Method
(NOTE: No equilibrate segment necessary
if starting at or near ambient temperature)
1) Ramp 10ºC/min to 300ºC
(NOTE: No equilibrate at RT necessary for
cooling down, defined in post test options)
• Cooling Method
1) Equilibrate at 300ºC
2) Ramp 10ºC/min to 25ºC
Exo Up
20
Method Design Rules
• Start Temperature
Generally, the baseline should have 2 minutes to
completely stabilise prior to the transition of
interest. Therefore, at 10ºC/min, start at least
20ºC below the transition onset temperature.
• End Temperature
Allow a 2 minute baseline after the transition of
interest in order to correctly select integration or
analysis limits.
But do not decompose the sample in the DSC cell !
21
Startup Hook in DSC
All DSC data will have an endothermic shift (on heating) at the beginning of the experiment.
This is commonly referred to as the “Startup Hook”, or more specifically, the “Approach to
Steady State”
Startup Hook
22
Why have 2 min of baseline after ?
6
238.55°C
4
Transition not yet
Heat Flow (W/g)
finished, so reaction
enthalpy always
2 underestimated
547.29J/g
629.94J/g
-2
160 180 200 220 240 260 280
Exo Up
Temperature (°C)
23
Calibration & Verification
Tzero or Baseline
Heat Flow
Temperature
24
Calibration Overview
∆T 1 1 dT d∆T
q = − + ∆ T 0 − + (C r − C s ) s − C r
Rr Rs Rr dt dt
26
Measuring the C’s & R’s
Performed using calibration setup
1. Run Empty Cell
2. Run Sapphire on both Sample & Reference side
27
Tzero Calibration – DSC 2500 & DSC 250
Might be good to cycle the cell several times between
min/max temp to be used for calibration (baseline
conditioning) prior to Tzero calibration
Choice of temperature range for Tzero calibration:
at least as broad as desired experimental range
Do NOT run it at a heating rate lower than 10ºC/min, by
preference 20ºC/min
After Tzero calibration the temperature and enthalpy
calibration is mandatory performed, e.g. with indium
28
Discovery DSC Tzero Calibration
?
T4P only
Baseline verification
29
Discovery DSC Calibration
for planning
in future
30
Discovery DSC Calibration
31
Discovery DSC Calibration
32
Viewing Tzero Calibration Results
Note that
calibration
results are
automatically
saved to the
equipment
33
Viewing Tzero Calibration Results
34
Viewing Existing Tzero Calibration
35
Viewing Existing Tzero Calibration
Highlight both files simultaneously, right click,
calibrations, Tzero calibration, C&R are calculated,
do not apply
36
Baseline Verification
not OK
Limits depend
upon DSC model good
excellent
37
Baseline Verification
38
Tzero Calibration & Baseline Verification
39
When to recalibrate ? NO
0.6
0.4
Heat Flow (mW)
0.2
––––––– sample A
––––––– empty cell
0.0
-0.2
50 100 150 200
Exo Up Temperature (°C) Universal V4.1D TA Instruments
40
When to recalibrate ? YES
-5.6
––––––– sample B
––––––– empty cell
-5.7
-5.8
Heat Flow (mW)
-5.9
-6.0
-6.1
-6.2
40 50 60 70 80 90 100
Exo Up Temperature (°C) Universal V4.1D TA Instruments
41
Baseline OK ? If not which action ?
0.5
0.4
0.3
0.2
Heat Flow (mW)
0.1
0.0
-0.1
OK
-0.2
-0.3
-0.4
-0.5
-100 0 100 200 300 400
Exo Up Temperature (°C) Universal V4.7A TA Instruments
42
Baseline OK ? If not which action ?
-11
-12
Recalibrate Tzero
Heat Flow (mW)
-13
-14
-15
-16
-100 0 100 200 300 400
Exo Up Temperature (°C) Universal V4.7A TA Instruments
43
Baseline OK ? If not which action ?
0
Heat Flow (mW)
-1
Check lid alignment
-2
-3
-4
-100 0 100 200 300 400
Exo Up Temperature (°C) Universal V4.7A TA Instruments
44
Baseline OK ? If not which action ?
0.2
Cycle 1
0.1
Heat Flow (mW)
Cycle 3
0.0
Lid moves ?
-0.1 Dirty cell ?
-0.2
-100 0 100 200 300 400
Exo Up Temperature (°C) Universal V4.7A TA Instruments
45
Baseline Calibration – DSC 25
46
Baseline Calibration – DSC 25
Note that
calibration result
is automatically
saved to the
equipment
If you want to make
changes to the analysis
(change temp region)
To see the analyzed graph
47
Baseline Calibration – DSC 25
48
Baseline Verification – DSC 25
49
DSC Calibrations: Cell Constant & Temperature
50
Heat Flow Calibration (Cell Constant)
51
Temperature Calibration
1 50
Extrapolated Onset is used as melting temperature
0
-1
30
-2
20
-3
HEATING RATE
10
-4
157.09ºC
PEAK
-5 0
150 152 154 156 158 160 162 164
°C)
Temperature (°
52
Influence of Sample Mass
0
Onset not
DSC Heat Flow (W/g)
Indium at
-2 10°C/minute influenced
Normalized Data 15mg by mass
10mg
4.0mg
-4
1.7mg
1.0mg
0.6mg
-6
150 152 154 156 158 160 162 164 166
Temperature (°C)
53
Temperature Calibration
• ASTM E967 method
– Pure metals typically used
– Calibration depends upon purge gas/flow rate and pan type
– Calibration changes a little with heating rate
– Temperature calibration (1 point) is automatically included
when calibrating cell constant
– Instead of 1 single temperature multiple calibration points
(up to a maximum of five) can be used (specify temperature
calib then)
– Choose the calibration standards so that their transition
temperatures span the interesting measuring region
– Use tin, lead, gallium and zinc one time only
54
Temperature Calibration
• If you want to calibrate temperature with more than 1
standard choose temperature calibration and include all
standards (also indium, despite the fact that indium has
already been run for enthalpy calibration)
• If you want to run calibration with indium at a lower rate than
10ºC/min, by preference first perform cell constant calib at
10ºC/min, and subsequently temp calib only at the lower
heating rate
55
Temperature and Enthalpy Standards
°C
Benzoic acid (147.3 J/g) Tm = 123°
Urea (241.8 J/g) Tm = 133°°C
Enthalpy °C
Indium (28.71 J/g) Tm = 156.6°
Anthracene (161.9 J/g) Tm = 216°°C
(cell constant)
Adamantane -65.54ºC 20.57 J/g
Cyclopentane* -150.77°°C
Cyclopentane* -135.09°°C
Cyclopentane* -93.43°°C
Cyclohexane# -83°°C * GEFTA recommended
Temperature Water# 0° °C Thermochim. Acta, 219 (1993) 333.
Gallium# 29.76°°C
# ITS 90 Fixed Point
Phenyl Ether# 30°°C
°C
p-NitrotolueneE 51.45° E Zone refined organic compound
NaphthaleneE 80.25°°C (sublimes)
Indium# 156.60°°C
Tin# 231.95°°C
Lead* 327.46°°C
Zinc# 419.53°°C see TN-11
56
Traceable Calibration Materials
• Certified materials used to establish traceability of instrument
calibration
58
Calibration & Verification of Heat Flow &
Temperature, All DSC’s
59
Calibration of Heat Flow & Temperature,
All DSC’s
Note that
calibration result
is automatically
saved to the
equipment
To see the
analyzed result
60
Calibration of Heat Flow & Temperature,
All DSC’s
61
Verification of Heat Flow & Temperature,
All DSC’s
62
Verification of Heat Flow & Temperature,
All DSC’s
63
Heat Flow & Temperature Calibration:
Temperature Calibration with Multiple Points
64
Reloading Older Cell Constant/Temperature
Calibration
65
Discovery DSC 2nd Generation:
Multiple Calibration Sets
66
Discovery DSC: Multiple Calibration Sets
67
Discovery DSC: Multiple Calibration Sets
68
Optimization of
Experimental Conditions
69
Instrument Preparation
• Cooling Accessories
– They should be operating and equilibrated prior to
runs (so do not turn off between runs !)
• Warm-up Time/Environment
– Electronics should be given at least one hour to
stabilize for important samples if the instrument has
been turned off (typically leave instrument always on)
– Electronics are effected by ambient temperature.
Avoid areas such as hoods or near an air conditioner
– Purge Gas
– Make sure cylinder is not empty or moisture
condensation will occur if cooler is on or still cold
70
Understanding Your Sample
71
Effective Sample Preparation
72
TA Instruments Tzero Pans
73
TA Instruments High Volume Pans
74
Sample Shape: Solids
75
Sample Shape: Solids
76
Sample Shape: Liquids
Sealing surface
77
DSC: Sample Size
78
Heating/Cooling Rate
79
Keeping the DSC Cell Clean
Software :
Case sensitive help, help topics, manual
E-training (basic)
www.tainstruments.com
Application library
www.tainstruments.com
81
Help and data analysis from your laptop ? Simply install the S/W from CD or download it !
www.tainstruments.com
82
Help on Desktop
83
Trios
84
www.tainstruments.com
85
www.tainstruments.com
or on You Tube !
• Baseline verification
• Powder preparation kit
• Integrating baselines
• .... and much more
86
www.tainstruments.com
87
www.tainstruments.com
88
Trios Software
89
PET (Heat-Cool-Heat)
C:/ProgramData/TAInstruments/Trios/Data/ExampleFiles/Thermal
90
PET (Heat-Cool-Heat) 1st Heat
C:/ProgramData/TAInstruments/Trios/Data/ExampleFiles/Thermal
91
PE (Low & High Density) 2nd Heat
C:/ProgramData/TAInstruments/Trios/Data/ExampleFiles/Thermal
92
PBT 2nd Heat
C:/ProgramData/TAInstruments/Trios/Data/ExampleFiles/Thermal
93
Liquid Sample MDSC 1st Heat
0.03
0.00
-0.03
-0.05
-0.08
-0.10
-0.13
-0.15
180 190 200 210 220 230 240 250 260
Exo Up (K)
Note: temperature in K
94
Indium Verification: Influence of Heating Rate
95
Two Indium Verification Test: Statistics
96
Reporting
https://triosvideos.s3.amazonaws.com/Report%20View.mp4
97
Thank You !