Certificate of Calibration

This certificate is issued without alteration, and in accordance with the conditions of accreditation granted by FLUKE. It is a correct record of
thesources made at the time of calibration. The values given in this certificate were correct at the time of calibration. Subsequently
the accuracy will depend on factors such as care exercised in handling the instrument and frequency of use. Recalibration should be performed
after a period which has been chosen to ensure that, under normal circumstances, the instruments accuracy remains within the desired limits.
The reported expanded uncertainties are based on a standard uncertainty multiplied by a coverage factor k = 2 providing a level of
confidence of approximately 95%. The uncertainties ofsource have been estimated in accordance with the principles defined in the
GUM. Guide to the expression of Uncertainty.

Certificate No : LAB-FL11B242A

Manufacturer : Fluke
Description : Multifunction Process Calibrator
Model No : 725 Ex
Serial No : 4596201
Plant No : AMB-UTIL-I-049
Departement : Utility Instrumentation

Calibrated for : Ambatovy Project


Address : Amboarikarivo Amboditandroho
Toamasina II - Region Atsinanana
Republique de Madagascar

Calibration
Environment
Temperature : 24,7 °C (±1°C)
Relative Humidity : 24,7 %rh (±5% rh)

Date of calibration : juin 08, 2023 Print Date juin 08, 2023
Expiry date : juin 08, 2024

Calibrated by : Felana V Esperence Checked by Nirinarivelona RABENJA

Signatory

Page 1 of 7
Certificate of Calibration
Certificate No: LAB-FL11B242A

1. Standards and equipment

Make Model Description Seriel Number Certificate Number Calibration due date
FLUKE 5502A Multi-product-calibrator 2714802 123593-2 17/03/2026
KEITHELY 2700 Multiple/Data Acquisition system 4005810 123593-3 10/03/2026

2. Procedure

2.1 - The UUT was Calibrated in accordance with FLUKE 725 Ex Calibration manual.

3. Results

3.1 Upper Display Voltage Measurement Tests

Range Input Tolerance UUT Uncertainty ofsource PASS/FAIL


(V) (V) (V) Display (V)
(V)
0 -0.002 to +0.002 0,000 ± 0,0007 PASS
30 V 15 14.995 to 15.005 14,997 ± 0,001 PASS
30 29.992 to 30.008 29,995 ± 0,002 PASS

3.2 Lower Display mV Measurement Tests

Range Input Tolerance UUT Uncertainty ofsource PASS/FAIL


(mV) (mV) (mV) Display (mV)
(mV)
75 0,00 -0.02 to +0.02 0,00 ± 0,008 PASS
50,00 49.97 to 50.03 49,99 ± 0,0095 PASS
89,00 88.96 to 89.04 88,98 ± 0,011 PASS

3.3 Lower Display Voltage Measurement Tests

Range Input Tolerance UUT Uncertainty ofsource PASS/FAIL


(V) (V) (V) Display (V)
(V)
0,00 -0.002 to 0.002 0,000 ± 0,0007 PASS
30V 10,00 9.996 to 10.004 9,997 ± 0,0009 PASS
20,00 19.994 to 20.006 0,000 ± 0,0013 FAIL

Technical Signatory

Page 2 of 7
Certificate of Calibration
Certificate No: LAB-FL11B242A

3.4 Upper Display mA Measurement Tests

Range Input Tolerance UUT Uncertainty ofsource PASS/FAIL


(mA) (mA) (mA) Display (mA)
(mA)
24 0 -0.002 to +0.002 0,000 ± 0,0007 PASS
12 11.995 to 12.005 11,998 ± 0,0016 PASS
24 23.993 to 24.007 23,995 ± 0,0027 FAIL

3.5 Lower Display mA Measurement Tests

Range Input Tolerance UUT Uncertainty ofsource PASS/FAIL


(mA) (mA) (mA) Display (mA)
(mA)
24 0 -0.002 to +0.002 0,022 ± 0,0007 FAIL
12 11.995 to 12.005 12,021 ± 0,0016 FAIL
24 23.993 to 24.007 24,018 ± 0,0027 FAIL

3.6 Lower Display Frequency Measurement Test

Range Input Tolerance UUT Uncertainty of PASS/FAIL


(Hz) (Hz) (Hz) Display source
(Hz) (Hz)
40 39.8 to 40.2 40,000 ± 0,0446 PASS
10 K 9.98 K to 10.02 K 10 K ± 0,0045 PASS

3.7 Lower Display Frequency Source Test

Range Input Tolerance UUT Uncertainty of PASS/FAIL


( Hz) ( Hz) ( Hz) Display source
( Hz) ( Hz)
40,00 39.98 to 40.02 40,00 ±0 PASS
1000,00 999.5 to 1000.5 999,98 ±0 PASS
1.0 K 997.5 to 1002.5 0,999 Hz ±0 FAIL
4.0 K 3.990 to 4.010 K 3,999 K ±0 PASS
10.0 K 9.975K to 10.025 K 9,999 K ±0 PASS

Technical Signatory

Page 3 of 7
Certificate of Calibration
Certificate No: LAB-FL11B242A

3.8 Lower Display 4-Wire Resistance Measurement Tests

Range Input Tolerance UUT Uncertainty ofsource PASS/FAIL


(Ω) (Ω) (Ω) Display (Ω)
(Ω)
3000 15,00 14.90 to 15.10 15,01 ± 0,0742 PASS
300,00 299.90 to 300.10 299,98 ± 0,0789 PASS
1500,0 1499.5 to 1500.5 1500,00 ± 0,754 PASS
3000,00 2999.0 to 3001.0 2999,80 ± 0,7891 PASS

3.9 Lower Display 4-Wire RTD Measurement

Input Tolerance UUT Uncertainty ofsource PASS/FAIL


(°C) (°C) Display (°C)
(°C)
-180 -179.7 to -180.3 -164,10 ± 0,0842 FAIL
100 99.7 to 100.3 117,80 ± 0,0953 FAIL
550 549.7 to 550.3 570,70 ± 0,1242 FAIL

3.10 Lower Display 3-Wire RTD Measurement

Input Tolerance UUT Uncertainty ofsource PASS/FAIL


(°C) (°C) Display (°C)
(°C)
-180 -179.5 to -180.5 -164,00 ± 0,08 FAIL
100 99.5 to 100.5 117,30 ± 0,1 FAIL
550 549.5 to 550.5 568,70 ± 0,12 FAIL

Technical Signatory

Page 4 of 7
Certificate of Calibration
Certificate No: LAB-FL11B242A

3.11 Lower Display Thermocouple Measurement Tests

Input Tolerance UUT Uncertainty ofsource PASS/FAIL


(°C) (°C) Display (°C)
(°C)
-180 -179.0 to -181.0 -179,40 ± 0,337 PASS
0 -0.7 to +0.7 0,20 ± 0,1758 PASS
750 749.3 to 750.7 750,10 ± 0,2694 PASS

3.12 Lower Display Thermocouple Source Tests

Input Tolerance UUT Uncertainty ofsource PASS/FAIL


(°C) (°C) Display (°C)
(°C)
-180 -179.0 to -181.0 -180,17 ± 0,057 PASS
0 -0.7 to +0.7 -0,07 ±0 PASS
750 749.3 to 750.7 750,05 ±0 PASS

3.13 Lower Display mA Source Tests

Range Input Tolerance UUT Uncertainty ofsource PASS/FAIL


(mA) (mA) (mA) Display (mA)
(mA)
4,00 3.9972 to 4.0028 3,943 ±0 FAIL
12,00 11.9956 to 12.0044 11,952 ±0 FAIL
24
20,0 19.9940 to 20.0060 19,962 ±0 FAIL
24,00 23.9932 to 24.0068 23,967 ±0 FAIL

3.14 Lower Display mV Source Tests

Range Input Tolerance UUT Uncertainty of PASS/FAIL


(mV) (mV) Display Measurement
0 -0.020 to +0.020 (mA)
0,000 (°C)
±0 PASS
200mV 50 49.970 to 50.030 50,000 ±0 PASS
100 99.960 to 100.040 100,010 ±0 PASS

Technical Signatory

Page 5 of 7
Certificate of Calibration
Certificate No: LAB-FL11B242A

3.15 Lower Display Voltage Source Tests

Range Input Tolerance UUT Uncertainty of PASS/FAIL


(mV) (mV) Display Measurement
20V 0 -0.002 to +0.002 (mA)
0,000 (°C)
±0 PASS
5 4.9970 to 5.0030 5,000 ±0 PASS
10 9.9960 to 10.0040 10,000 ±0 PASS

3.16 Lower Display Ohms Source Tests

Range Input Tolerance UUT Uncertainty of PASS/FAIL


(Ω) (Ω) (Ω) Display Measurement
(Ω) (Ω)

4-wires 15 14.9 to 15.1 15,0 ±0 PASS


360 359.9 to 360.1 360,0 ±0 PASS
1500 1499.5 to 1500.5 1499,9 ±0 PASS
3200,000 3199.0 to 3201.0 3199,170 ±0 PASS

4. Comments

4.1 - The UUT do not meet with the manufacturer's Accuracy specification.
4.2 - This temperature calibrator is not in good condition

5. Recommendation

5.1- Caution, must not be used

— End of Document —